spo monitoring - sle.co.uk brochure... · what do you get? when you opt for spo 2 monitoring on the...

4
SLE6000 module SpO 2 Monitoring SpO 2 monitoring with the SLE6000 ventilator

Upload: dinhtram

Post on 17-May-2019

217 views

Category:

Documents


0 download

TRANSCRIPT

SLE6000 module

SpO2 MonitoringSpO2 monitoring with the SLE6000 ventilator

What do you get?When you opt for SpO2 monitoring on the SLE6000 ventilator its software is upgraded to allow SLE’s uSpO2 cable (Masimo SET) to be plugged into the back of the ventilator.

This cable is designed to accept Masimo’s infant/neonatal LNCS sensors and to process and send the signals back to the ventilator for display and storage.

All of the settings and alarms are handled by the SLE6000. SpO2 and pulse rate values are stored and trended for 14 days.

OxyGenie SLE’s Auto O2 system, OxyGenie, needs a reliable source of SpO2 data to work effectively. The SpO2 upgrade software module and Masimo™ hardware in this brochure are required components to ensure effective operation. (Please see separate OxyGenie brochure for further details).

SLE6000: Monitor progress

SLE6000 SpO2 Settings screen

SLE6000 SpO2 waveform and digital value

Pulse OximetryPulse oximetry is used to measure the oxygen level (oxygen saturation) in the blood. It is a noninvasive, painless, indicator of oxygen delivery to the peripheral areas (such as the hand or foot).

Why add it to the SLE6000?Since the SLE6000 is the focus for all of the ventilation needs of the patient integrating SpO2 monitoring allows the caregiver to more easily control the inspired O2 whilst simultaneously monitoring all of the other ventilation parameters.

Having all of the data on a single screen allows the caregiver to focus on the patient’s ventilation. During therapy the caregiver can call up simultaneous trends of SpO2 and compare it with other parameters (such as Pressure or delivered Oxygen) allowing easier interpretation of linked events and better visualisation of the patient’s condition

Why Masimo?Masimo SET® (Signal Extraction Technology) has become the recognized leader in pulse oximetry monitoring, and has been proven to work very well on infants (1).

Studies have shown that Masimo SET performs better under challenging conditions of motion and low-perfusion better than conventional SpO2 technology, which is particularly important in neonates (2).

Part Number Description

L6000/SP2/KIT SLE uSpO2 cable (Masimo™ SET) (1.83 m) (includes LNCS® sample sensor starter kit).

Z6000/SPO SLE6000 SpO2 upgrade software module.

Sensors

LSPO2/2319 LNCS Inf-3 - Infant, SpO2 adhesive sensor, 3 ft, 20/box

LSPO2/2320 LNCS Neo-3 - Neonatal, SpO2 adhesive sensor, 3 ft, 20/box

LSPO2/2321 LNCS NeoPt-3 - Neonatal, SpO2 adhesive sensor, 3 ft, 20/box

Accessories

LSPO2/4089 RD to LNC Adapter cable, 3 ft. (to allow the use of Masimo RD sensors)

Part numbers and Specifications

Specifications (Neonatal/infant)

Displayed parameters Saturation (fraction of oxyhaemoglobin to functional haemoglobin), pulse rate, Signal IQ and plethysmogram

Trends Saturation and Pulse rate for previous 14 days

Measuring method Absorption spectrophotometry

Physical Specifications (uSpO2 cable)

Ventilator connector ODU-type plug. Powered from ventilator.

Size (H x W x D) 24 x 33 x 92 mm

Weight (excluding sensor) 0.122 kg

LSPO2/2320

LSPO2/2321

SLE uSpO2 cable (Masimo™ SET)

LSPO2/2319

Specifications correct at the time of writing.

Performance SpecificationsFractional SpO2 (%) Pulse Rate (BPM) Perfusion Index (%)

Display Range 0.0% - 100.0% 25 BPM - 240 BPM 0.02% - 20.0%

Calibration range 70% - 100% 25 BPM - 240 BPM 0.10% - 20.0%

No Motion Accuracy (rms) ± 2.0% ± 3.0 BPM -

Motion Accuracy (rms) ± 3.0% ± 5.0 BPM -

Resolution ≤ 0.1% ≤ 1 BPM ≤ 0.01%

Time to Display ≤ 8, ≤12 seconds ≤ 8, ≤12 seconds -

Asystole Detection Time ≤ 8 seconds ≤ 8 seconds ≤ 8 seconds

Delay ≤ 10 seconds ≤ 10 seconds ≤ 10 seconds

Response Time ≤ 20 seconds ≤ 20 seconds ≤ 20 seconds

Averaging Time (seconds) 2-4, 4-6, 8, 10, 12, 14, 16

- -

Environmental Operating Conditions

Temperature @ ambient humidity

5 ºC - 40 ºC

Humidity 15% - 95% (non-condensing)

Pressure 500 - 1060 mbar

tel: +44 (0)20 8681 1414fax: +44 (0)20 8649 [email protected]

SLE Limited Twin Bridges Business Park

232 Selsdon Road South Croydon

Surrey CR2 6PL

UK

Masimo, SET, and LNCS are federally registered trademarks of Masimo Corporation. | SLE and SLE6000 are trademarks of SLE Ltd. All other names, logos, and brands are property of their respective owners. | Copyright SLE Ltd ©2018 | Subject to Change | G0153/0UK/001 02/18

References1. A full citation list of featured clinical studies from Masimo is available at:http://www.masimo.com/evidence/featured-studies/feature/2. The link below features studies involving Masimo’s technologies:http://www.masimo.com/evidence/pulse-oximetry/set/3. Avoiding hyperoxemia during neonatal resuscitation: time to response of different SpO2 monitorsHernando Baquero, Ramiro Alviz, Armando Castillo, Fredy Neira, Augusto SolaActa Pædiatrica 2011 100, pp. 515–5184. Prevention of retinopathy of prematurity in preterm infants through changes in clinical practice and SpO2 technologyArmando Castillo, Richard Deulofeut, Ann Critz, Augusto SolaActa Pædiatrica 2011 100, pp. 188–1925. Reliability of Conventional and New Pulse Oximetry in Neonatal PatientsWilliam W. Hay, Jr., Donna J. Rodden, Shannon M. Collins, Diane L. Melara, Kathy A. Hale, Lucy M. FashawJournal of Perinatology 2002; 22:360 – 366