![Page 1: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/1.jpg)
Imagination at work.
Customer facing
Sievers M-Series TOC Analysers - DataPro 2
![Page 2: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/2.jpg)
Databases
Easier to access data
More secure data
Results, etc. are saved in the right place
Not easy to delete files on M9 as it was with DataPro
900
Network administrator can have good control on who
can access the data e.g. user cannot accidentally delete
data
Microsoft SQL is a type of database
![Page 3: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/3.jpg)
Database structure
DataPro 2 is a database engine: it queries the data
Good for 21 CFR Part 11 and EU Annex 11 compliance
Data of record should be documented as that on the PC
Both instrument and DataPro 2 have Part 11 compliance
Instrument and DataPro 2 communicate continuously so the data on the
instrument is simply like a backup
3
Client tier
Database engine (e.g. DataPro 2)
Database files (e.g. results)
![Page 4: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/4.jpg)
4
On SRC located in the “Instrument Software Installation Files\DataPro DVD” folder
![Page 5: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/5.jpg)
SQL Database Structure
5
Graphic courtesy of Microsoft Corp., MSDN Website 2014.
![Page 6: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/6.jpg)
Home Screen Overview
6
Same menu icons as on the
instrument
![Page 7: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/7.jpg)
Home Screen Overview
7
Up to 4 instruments can
be connected
![Page 8: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/8.jpg)
Home Screen Overview
8
Protocol and Vial information
![Page 9: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/9.jpg)
Home Screen Overview
9
Autosampler rack
display
![Page 10: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/10.jpg)
Home Screen Overview
10
Results populate in real-time
![Page 11: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/11.jpg)
Home Screen Overview
11
Method for the selected vial shown here
![Page 12: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/12.jpg)
Home Screen Overview
12
Consumables status
![Page 13: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/13.jpg)
Home Screen Overview
13
Errors and Warnings
![Page 14: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/14.jpg)
Setting Up Methods & Protocols
14
We’re in the Setup tab
![Page 15: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/15.jpg)
Creating and Saving Methods
Methods define how a vial will
be measured
• Values and options can be set
per customers’ needs
• Methods can be saved, if desired
Methods must be applied to
one or more vials
15
![Page 16: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/16.jpg)
Creating Protocols
16
One or more vials must be
“placed into” the rack by
highlighting them
• Some Excel® rules apply
– shift, ctrl, dragging
![Page 17: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/17.jpg)
Creating Protocols
17
One or more vials must be
“placed into” the rack by
highlighting them
• Some Excel® rules apply
– shift, ctrl, dragging
Apply the method to the
selected vials
![Page 18: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/18.jpg)
Creating Protocols
18
One or more vials must be
“placed into” the rack by
highlighting them
• Some Excel® rules apply
– shift, ctrl, dragging
Apply the method to the
selected vials
Vials will appear as grey,
stylized vials
![Page 19: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/19.jpg)
Creating Protocols
19
One or more vials must be
“placed into” the rack by
highlighting them
• Some Excel® rules apply
– shift, ctrl, dragging
Apply the method to the
selected vials
Vials will appear as grey,
stylized vials
The table below is editable
![Page 20: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/20.jpg)
Managing Favorites: User Protocols and Methods • Protocols and methods can be organized and ranked on Favorites
page.
• All Protocols/Methods will appear under favorites unless removed.
• Favorites can be organized into any order.
• Non favorites (Available) appear in the order they are removed from
favorites.
20
![Page 21: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/21.jpg)
Stacking Protocols
21
• Automated series of
protocols
• Allows stacking user-
protocols:
Suit – Samples - Suit
![Page 22: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/22.jpg)
DataGuard: Access Levels
22
- User can create a new Access level
by
clicking Add, typing in a level name
and
moving the desired roles to the
selected field.
- Highlighting an existing Access level
and selecting Edit a user can add or
delete roles to a an access levels.
- Highlighting an existing Access level
and selecting delete an access level
can be removed if it is not associated
with a User ID.
![Page 23: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/23.jpg)
DataGuard: Settings
The settings tab allows the user to modify the default
settings to meet required standards.
23
![Page 24: - DataPro 2€¦ · Database structure DataPro 2 is a database engine: it queries the data Good for 21 CFR Part 11 and EU Annex 11 compliance Data of record should be documented as](https://reader036.vdocuments.us/reader036/viewer/2022071508/6128df00cb16ec18464c3751/html5/thumbnails/24.jpg)