Winter Arrhythmia School Blue Mountain February 6-8, 2015

Upload: others

Post on 17-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

Winter Arrhythmia School

Blue Mountain

February 6-8, 2015

Page 2: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

Myths and Facts: A Discussion of MRI Imaging for Pacemaker and ICD Patients

Matt Reitzel

BIOTRONIK

Canada

BIOTRONIK // WAS 2015: MRI in CIED

Page 3: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

3

Document identifier

Topics of discussion:

� Your experience with MR-conditional devices

� Overview of MR-conditional devices in the

field

� Components of MR-conditional systems

� CHRS recommendations for device/patient

management for MR scan

Page 4: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

4

Document identifier

Topics of discussion:

� Your experience with MR-conditional devices

� Overview of MR-conditional devices in the field

� Components of MR-conditional systems

� CHRS recommendations for device/patient management for MR scan

Page 5: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

5

Document identifier

Clicker Question: Does your centre currently implant MR-conditional CIED?

A) No, 0%

B) Yes, <5%

C) Yes, <20%

D) Yes, > 20%

Page 6: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

6

Document identifier

Clicker Question: Does your centre currently provide on-site MR scans for patients implanted with MR-conditional CIED?

A) No scanning, no protocol to scan

B) No scanning, protocol in place to scan if

required

C) Yes scanning preformed and protocol in

place

Page 7: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

7

Document identifier

Topics of discussion:

� Your experience with MR-conditional devices

� Overview of MR-conditional devices in the field

� Components of MR-conditional systems

� CHRS recommendations for device/patient management for MR scan

Page 8: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

8

Document identifier

BIOTRONIK ProMRI® - Developing the industry’s most complete MR-conditional CRM portfolio

2003 2008 2009 2010 2011 2012

BIOTRONIK starts

development of

MR-Conditional

systems

Approval of 1st MR

Cond. IPG System

Launch of 1st MR

Cond. IPG System

Launch MR Cond.

ICD System

MR Cond. CRT-

P Systems

BIOTRONIK’s

1st MR Cond.

IPG System

Expansion of

the MR Cond.

IPG Portfolio

Approval of 1st

MR Cond. ICD

System

2013

Expansion MR

Cond. ICD

System

MR Cond. Full

Body Scan IPG

System

BioMonitor

2014

MR Cond.

CRT-D Systems

Next generation

MR Cond. IPG

System

Eluna 8 DR-T

MR Cond. FBS &

3.0 T ICD System

Page 9: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

9

Document identifier

Topics of discussion:

� Your experience with MR-conditional devices

� Overview of MR-conditional devices in the field

� Components of MR-conditional systems– What’s the difference? Combination of leads/devices? Exclusion

zone or full body scan?

� CHRS recommendations for device/patient management for MR scan

� Future of MR-conditional devices

Page 10: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

10

Document identifier

Clicker Question: It is estimated that a patient has an X % chance of requiring an MR examination over his/her lifetime after CIED implantation. X = ?

A) 0-25%

B)26-50%

C) 51-75%

D) 75-100%

Page 11: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

11

Document identifier

Clicker Question: It is estimated that a patient has an X % chance of requiring an MR examination over his/her lifetime after CIED implantation. X = ?

A) 0-25%

B)26-50%

C) 51-75%

D) 75-100%

Kalin R, Stanton MS. Current clinical issues for MR scanning of pacemaker and defibrillator patients. Pacing ClinElectrophysiol 2005;28:326-8.

Page 12: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

12

Document identifier

50-75% of patients with a CIED are indicated for an MR scan

– Over 30 million MRI scans conducted in the US1.– 2.5 million US patients have a pacemaker, and this number is

growing by 80.000 per year2. – Estimated 50%-75% of patients with a device are indicated for an

MRI scan3.

0

50

100

150

200

250

300

350

1997 1998 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011

ICD

(10

00x)

0

10

20

30

40

50

60

MR

I (m

illi

on

s)

ICD

MRI

1 Benchmark Report: MRI 2007, IMV medical Information Devision 2008.

2 Sommer t et al, Circulation 2006, 114; 1285-1292

3 Kalin R, Stanton MS. Pacing Clin Electrophysiol 2005;28:326–8

Page 13: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

13

Document identifier

Utilization of MRI Scan

– Patients aged 70-74 years are most likely to receive an MRI scan*

MRI utilization increases by age

* BARMER Health insurance: Arztreport 2011

Page 14: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

14

Document identifier

Do the terms MRI Safe and MRI conditional have the same meaning?

MR- safe: An item that poses no known hazards in all MRI environments

MR- conditional: An item that has been demonstrated to pose no known hazards in a specified MR environment with specified conditions of use (field strength, spatial gradient etc.)

MR-unsafe: An item that is known to pose hazards in all MR environments

Page 15: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

15

Document identifier

MRI compatibility- what’s the big deal?

Page 16: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

16

Document identifier

Manufacturing MRI compatible devices- What’s the difference?

� Three Birdcages for 1.5 & 3.0 Tesla testing

� Shielded chambers

� RF antenna of clinical MRT is used

Page 17: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

17

Document identifier

Manufacturing MRI compatible devices- What’s the difference? Software

� Special MRI programming mode developed

� Asynchronous pacing maintains appropriate pacing support during MRI regardless of the noise induced on the pacing system

� Non-pacing mode recommended for non-pacemaker dependent patients

� Diagnostic data collection, automatic system monitoring measurements, atrial arrhythmia detection and atrialarrhythmia therapy are suspended during MRI mode operation

Page 18: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

18

Document identifier

Manufacturing MRI compatible devices- What’s the difference? Hardware

� Improved Electromagnetic Interference (EMI) susceptibility

� New chip developed to improve EMI rejection and minimize the number of external component

� Optimized component selection and layout for improved EMI rejection over 16Hz - 3000Mhz

� Significantly reduced the length of the electrical connections between the feedthrough and the electronic circuit

� Minimized EMI radiation inside pacemaker housing

Page 19: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

19

Document identifier

Clicker Question: Approved ProMRI systems are MR-safe.

A) True

B) False

Page 20: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

20

Document identifier

Clicker Question: Approved ProMRI systems are MR-safe.

A) True

B) False

No CIED system is “MR-safe”, but selected CIED systems are “MR-conditional”

Page 21: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

21

Document identifier

Clicker Question: Conditions for MR imaging of an MR-conditional CIED are consistent from vendor to vendor and device to device.

A) True

B) False

Page 22: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

22

Document identifier

Clicker Question: Conditions for MR imaging of an MR-conditional CIED are consistent from vendor to vendor and device to device.

A) True

B) False

Manufacturer-specified scanning parameters are always checked and followed; the MR scanning protocol will vary in accordance with the patient’s CIED system. Some systems are initially approved with exclusion zone and then retroactively approved for full body scanning. Recommended to check with vendor before scan to ensure parameters are know.

Page 23: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

23

Document identifier

Clicker Question: MRI for a device with an exclusion zone.

A) Yes

B) No

The manufacturer guidelines state that the MRI isocentrecan pass no lower than eye level and no higher than hip level. Are images of the C and L spine possible?

Page 24: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

24

Document identifier

Clicker Question: MRI for a device with an exclusion zone

A) Yes

B) No

Are images of the C and L spine possible?

Page 25: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

25

Document identifier

MR Exclusion Zone

Exclusion zone in daily practice

Actual Exclusion zone

Leng

ht1.

75 c

m

SchematicExclusion zone

permissible positioning zone

permissible positioning zone

exclusion zone

isocenter

isocenter

field of view 50* cm

field of view 50* cm

exclusion zone

* depends by MRI model

Page 26: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

26

Document identifier

MR Exclusion Zone

Exclusion zone in daily practice

Scan with isocenter on eye level

isocenter eye level

Page 27: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

27

Document identifier

MR Exclusion Zone

Exclusion zone in daily practice

Scan with isocenter on eye level

isocenter eye level

C2

C3

C4

C5

C6C7T1T2

T3

T4

T5

Real patient case with isocenter on eye level and Siemens MRI scanner.Patient length 1.75 m and scan taken within the ProMRI conditions.

Head, cervical spine and upper thoracic spine (T4) can be examined

Page 28: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

28

Document identifier

MR Exclusion Zone

Exclusion zone in daily practice

* depends by MRI model

Scan with iso-center on hip level

S1

L5

L4

L3

L2

Real patient case with isocenter on hip level and Siemens MRI scanner.Patient length 1.75 m and scan taken within the ProMRI conditions.

Lower lumbar spine can be examined

Page 29: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

29

Document identifier

Clicker Question: A device with an MRI exclusion zone limits what percentage of total indicated MRI Scans?

A) <5%

B) 10-15%

C) 15-25%

D) 25-50%

E) >50%

Page 30: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

30

Document identifier

Clicker Question: A device with an MRI exclusion zone (thoracic area) limits what percentage of total indicated MRI Scans?

A) <5%

B) 10-15%

C) 15-25%

D) 25-50%

E) >50%

BARMAR: Arztreport 2011. Barmar GEK- large Health Care provider data in Europe

Page 31: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

31

Document identifier

88% of MRI scans can be performed

– BARMER health insurance data show the MRI usage per scan location

– Data are verified by BIOTRONIK radiology consultants

Others: 1.7%

Head: 26,3%

Neck: 2%

Chest: 1.8%

Cervical & lumbar(L3-5) spine** : 26,7 %

Thoracic spine** : 3,7%

Abdomen: 4.5 %

Hip: 3.2 %

Extremities: 30.2 %

MRI procedures in % of total*

*BARMAR: Arztreport 2011** C, T and L spine percentages based on expert evaluation

Head 26,3%Neck 2.0%Cervical & Lumbar spine 26.7%Hip 3.2%Extremities 30.2%

88.4 %

Page 32: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

32

Document identifier

Clicker Question: Are radiopaque markers on devices and leads a reliable way to determine MR-compatibility?

A) Yes

B) No

Page 33: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

33

Document identifier

Clicker Question: Are radiopaque markers on devices and leads a reliable way to determine MR-compatibility?

A) Yes

B) No

How would we tell only with a marker if it was full body, exclusion zone, 1.5T or 3T etc. Also, as “legacy” leads are approved for MR-compatibility, these markers would not appear on these leads. It is unreliable due to this case and would also require checking with the manufacturer for the acceptable scanning parameters when these markers are present

Page 34: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

34

Document identifier

Clicker Question: Biotronik has “back dated” over 30 non-ProMRI lead types as MR Conditional (CE Mark)

With CE Mark, if a ProMRI pacemaker is

implanted with these leads, is the system MRI

compatible in Canada?

A) Yes

B) No

Page 35: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

35

Document identifier

Clicker Question: Biotronik has “back dated” over 30 non-ProMRI lead types as MR Conditional (CE Mark)

With CE Mark, if a ProMRI pacemaker is implanted with these leads, is the system MRI compatible in Canada?

A) Yes

B) No

Along with the current leads, BIOTRONIK has tested 34 current, non-ProMRI leads and approved them as MRI conditional (with CE Mark). At this time, Health Canada has not granted the back date and therefore, only the currently approved system is considered ProMRI

Page 36: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

36

Document identifier

Clicker Question: A patient is implanted with MR-compatible leads from Company X and an MR-compatible pacemaker from Company Y.

Is this system MR-compatible?

A) Yes

B) No

Page 37: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

37

Document identifier

Clicker Question: A patient is implanted with MR-compatible leads from Company X and an MR-compatible pacemaker from Company Y.

Is this system MR-compatible?

A) Yes

B) No

Labeling for Health Canada has approved “device systems”. Although individual components are MR-compatible, this has not been validated and stated as an MR-conditional system.

Page 38: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

38

Document identifier

Topics of discussion:

� Your experience with MR- conditional devices

� Overview of MR- conditional devices in the field

� Components of MR- conditional systems

� CHRS recommendations for device/patient management for MR scan

Page 39: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

39

Document identifier

Clicker Question: Which of the following are absolute contraindications for MR scanning?

A) MR scanning protocol outside of manufacturer specifications

B) Presence of abandoned pacing or ICD leads

C) Broken or fractured lead(s)- known or suspected

D) Permanent epicardial pacing or ICD leads

E) A,B and C

F) All of the above

Page 40: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

40

Document identifier

Clicker Question: Which of the following are absolute contraindications for MR scanning?

A) MR scanning protocol outside of manufacturer specifications

B) Presence of abandoned pacing or ICD leads

C) Broken or fractured lead(s)- known or suspected

D) Permanent epicardial pacing or ICD leads

E) A,B and C

F) All of the above

Page 41: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

41

Document identifier

Clicker Question: Which of the following methods is acceptable to monitor the patients vitals while undergoing MR scan?

A) pulse oximetry

B)Electrocardiographic (ECG)

C) Capability for verbal communication

between the MR scan operator and

patient

D) All of the above

Page 42: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

42

Document identifier

Clicker Question: Which of the following methods is acceptable to monitor the patients vitals while undergoing MR scan?

A) pulse oximetry

B)Electrocardiographic (ECG)

C) Capability for verbal communication between the MR scan operator and patient

D) All of the above

It is recommended to use either pulse oximetry or ECG in conjunction with verbal interaction according to CHRS consensus

Page 43: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

43

Document identifier

Clicker Question: Monitoring of the MR-conditional CIED patient during MR scanning may be preformed by:

A)MR Technologist

B)MR Radiologist

C) CIED Clinic nurse

D) Cardiologist with knowledge of CIED

E) All of the above

Page 44: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

44

Document identifier

Clicker Question: Monitoring of the MR-conditional CIED patient during MR scanning may be preformed by:

A)MR Technologist

B)MR Radiologist

C) CIED Clinic nurse

D) Cardiologist with knowledge of CIED

E) All of the above

According to CHRS consensus, all of the above are acceptable however, should be evaluated based on the patient’s medical status, functional status of the CIED and experience of the staff

Page 45: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

45

Document identifier

Clicker Question: Current approved devices require a CIED clinic staff to test and reprogram devices pre- and post-MR scan.

A)True

B) False

Page 46: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

46

Document identifier

Clicker Question: Current approved devices require a CIED clinic staff to test and reprogram devices pre- and post-MR scan.

A) True

B) False

CIED clinic staff ensures device testing is within appropriate ranges for MR-scan (lead impedance, sensing and capture) and ensures MRI programming for the device during scan. Post scan testing to ensure device functionality and restores permanent programming.

Page 47: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

47

Document identifier

Clicker Question: Next generation MR devices do not required a CIED clinic staff to test and reprogram devices pre- and post-MR scan?

A)True

B) False

Note: Next generation of MR compatible devices may include devices that auto- detect MRI environment and reprogram themselves or have an MRI activator the radiologist may use to temporarily reprogram the device pre- and post- MR scan.

Page 48: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

48

Document identifier

Clicker Question: Next generation MR devices do not required a CIED clinic staff to test and reprogram devices pre- and post-MR scan?

A)True

B) False

Regardless of MRI programming of the device, an full interrogation and evaluation of the CIED system is required pre- and post-MRI as per CHRS guidelines.

Page 49: Winter Arrhythmia School Blue Mountain€¦ · 8 Document identifier BIOTRONIK ProMRI ®-Developing the industry’smost complete MR-conditional CRM portfolio 2003 2008 2009 2010

49

Document identifier

Thank you for your participation!