MasterScreen IOS MasterScreen PFT MasterScreen Body Powerful, modular software and hardware design • Adhering to the latest 2005 ATS/ERS standardization guidelines • Suitable for physician offices, specialty practices, hospitals, and for research purposes. • MasterScreen is centered around the unique SentrySuite Software Package focusing on efficiency and productivity. MasterScreen/SentrySuite ® Technical Specifications “Rely on our JAEGER pneumotach for accurate and reliable measurements with a complete hygiene solution for patient safety”

Upload: vannga

Post on 15-Jul-2018

290 views

Category:

Documents


0 download

TRANSCRIPT

MasterScreen IOS MasterScreen PFT MasterScreen Body

• Powerful, modular software and hardware design

• Adheringtothelatest2005ATS/ERSstandardizationguidelines

• Suitableforphysicianoffices,specialtypractices,hospitals,andforresearchpurposes.

• MasterScreeniscenteredaroundtheuniqueSentrySuiteSoftwarePackagefocusingonefficiencyandproductivity.

MasterScreen/SentrySuite® Technical Specifications

“Rely on our JAEGER™ pneumotach for accurate and reliable measurements with a complete hygiene solution for patient safety”

MasterScreen Pneumo MasterScreen Pneumo with Rocc MasterScreen IOS MasterScreen Diffusion (RT)

PCbasedspirometrysystemforthemeasure-ment, recording and assessment of slow and forcedspirometry,aswellasmaximumvol-untaryventilation(MVV)andPre/Posthan-dling.

Thesystemisequippedwithapneumotachforexactmeasurement resultsandacom-pletehygienicconceptforpatientsafety.

PCbasedspirometrysystemforthemeasure-ment, recording and assessment of slow and forcedspirometry,maximumvoluntaryven-tilation(MVV)andairwayresistance(ROcc).

Thesystemisequippedwithapneumotachand shutter forexactmeasurement resultsandacompletehygienicconceptforpatientsafety.

Measuring parameters:Forced Spirometry (FVC):FVC,FEV1,FEF25,FEF50,FEF75,PEF,PIF,FEV1%VC,...Slow Spirometry (SVC):ERV, IRV, IC,VT,VCIN,VCEX,VCMAXMaximum voluntary ventilation (MVV):MVV,BF,VT, ...

Delivery:Hardware:•Digitalhandle•JAEGER™pneumotach(heatedorunheated)

•Accessories (mouthpieces,MicroGard™ II filter,noseclips, ...)

•Calibrationpump(option)•PowerUnit (forheatedpneumotachonly)

•DesktoporNotebookcomputer (option)•Color inkjetprinter (option)•Trolley (console)withsupportarm(option)

Software:•Windows®7SP1,64bit•SentrySuiteV2.15•MeasuringprogramSpirometry•APSproBronchialChallenge(option)

Measuring parameters:identical to the MasterScreen Pneumoplus:Airway Resistance:ROcc,GOcc,Palv, ...

Delivery:Hardware:identical to the MasterScreen Pneumoplus:•Shutterwithpowerunit

Software:identical to the MasterScreen Pneumoplus:•MeasuringprogramROcc•MeasuringprogramP0.1(option)•MeasuringprogramMIP/MEP(option)

Measuring parameters:identical to the MasterScreen Pneumoplus:Impulse Oscillometry:R5Hz,X5Hz,Fres.,AX,D5-20%,VT, ...

Complete system for recording spirometryandpulmonarydiffusingcapacityonthebasisof CO diffusion with simultaneous determi-nationofalveolar volumeandotherabso-lute pulmonary volumes. The continuous,high speedgas analysis ofCO/CH4allowsthemeasurementofSBDiffusioninRealtimeandadditionallywiththeIntrabreathmethodwithnon-breathholdmanoeuvre.

Measuring parameters:identical to the MasterScreen Pneumo plus:Diffusion SB Realtime:DLCOSB,DLCOSBc,VASB,KCO,KCOc...Diffusion SB Intrabreath:DLCOIB,DLCOIBc,VAIB,KCOIB,KCOIBc...

System for performing oscillometric diag-nosticsindependentofpatientcooperation.TheIOSdoesnotonlyallowforaquickandreliabledeterminationofpulmonaryobstruc-tionsbutfurtherdifferentiatesbetweencen-tralandperipheralcomponents.

Delivery*:Hardware:identical to the MasterScreen Pneumoplus:• IOSheadwithspeaker•Y-adapter•Powerunitorsystemboxwith isolation

transformer•Referenceresistance

Software:identical to the MasterScreen Pneumoplus:•MeasuringprogramIOS

* The IOS isavailableasan independentsystemorasanaddition toaMasterScreenSystem.

Delivery:Hardware: •Trolleywithsupportarm•JAEGER™ Pneumotach•Accessories (mouthpieces,MicroGard™ II filter,noseclips,…)•FastCO/CH4Diffusiongasanalyzer•Systembox incl. isolationtransformer•Calibrationpump(option)•Desktopcomputer/TFTmonitor•Color inkjetprinter

Software:identical to the MasterScreen Pneumo plus:•Measuringprogram -DiffusionSBRealtime -DiffusionSB Intrabreath

MasterScreen Body MasterScreen Body/Diff MS Body XXL-size two doors cabine

Whole body plethysmograph capable ofcomprehensive pulmonary function test-ingequippedwith spaciousboxaswell aseasy-to-exchange,heatedpneumotachandaquickshutter.Maximizepatientcomfortbyheight-adjustable seat and easy-to-handlesupportarm.

IdenticalwithMasterScreenBodybutwith integrated fast gas analyzer for measur-ingdiffusioncapacityvia thesingle-breathmethodinRealtimeandadditionallywiththeIntrabreath method with non-breath holdmanoeuvre.

IdenticalwithMasterScreenBodyorBody/DiffbutwithaXXL-size twodoors cabinewith increased leg room and access for wheelchair.

Measuring parameters:identical to the MasterScreen Pneumoplus:Bodyplethysmography:FRCpleth,ERV,RV,TLC,VC,...sReff,Reff,sRtot,Rtot, ...

Delivery:Hardware:•Trolley•JAEGER™ Pneumotach•Shutter•Accessories (mouthpieces,MicroGard™ II filter,noseclips, ...)

•Traditionalcomfort-sizedbodybox•Systembox incl. isolationtransformer•Calibrationpump(option)•Desktopcomputer/TFTmonitor•Color inkjetprinterOptions:•XXL-sizetwodoorscabinwith increased

leg room and access for wheelchair•APSProvocationsystem(insidebox)• IOS ImpulseOscillometry•MasterScreenPFT

Software:identical to the MasterScreen Pneumoplus:• Measuring program - Bodyplethysmography- ROcc,P0.1,MIP/MEP(options)

Measuring parameters:identical to the MasterScreen Pneumo plus:Bodyplethysmography:FRCpleth,ERV,RV,TLC,VC,...SReff,Reff,SRtot,Rtot, ...Diffusion SB:DLCOSB,DLCOSBc,VASB,KCO,KCOc...

Delivery:Hardware:•Trolley•JAEGER™ Pneumotach•Shutter•Accessories (mouthpieces,MicroGard™ II filter,noseclips, ...)

•Traditionalcomfort-sizedbodybox• Integratedgasanalyzer•DemandValve•Systembox incl. isolationtransformer•Calibrationpump(option)•Desktopcomputer/TFTmonitor•Color inkjetprinterOptions:•APSProvocationsystem(insidebox)• IOS ImpulseOscillometry•XXL-sizetwodoorscabinSoftware:identical to the MasterScreen Pneumo plus:• Measuring program- Bodyplethysmography-DiffusionSBRealtime/Intrabreath-ROcc,P0.1,MIP/MEP(options)

Measuring parameters:identical to the MasterScreen Body

Delivery:Hardware:•Trolley•JAEGER™ Pneumotach•Shutter•Accessories (mouthpieces,MicroGard™ II filter,noseclips,…)

•XXL-sizetwodoorsboxwith increasedleg room and access for wheelchair

•Rampforeasywheelchairaccess•Removableheightadjustablechair•Systembox incl. isolationtransformer•Calibrationpump(option)•Desktopcomputer/TFTmonitor•Color inkjetprinterOptions:•APSProvocationsystem(insidebox)• IOS ImpulseOscillometry•MasterScreenPFT

Software:identical to the MasterScreen Body

MasterScreen PFT

Measuring parameters:identical to the MasterScreen Pneumo plus:FRC Rebreathing:FRC-He,ERV,RV,TLC,VC,...Diffusion SB Realtime:DLCOSB,DLCOSBc,VASB,KCO,KCOc...Diffusion SB Intrabreath:DLCOIB,DLCOIBc,VAIB,KCOIB,KCOIBc...

Delivery:Hardware:•Trolleywithsupportarm•JAEGER™ Pneumotach•Accessories (mouthpieces,MicroGard™ II filter,noseclips,…)

•PFTGasanalyzer•Systembox incl. isolationtransformer•Calibrationpump(option)•Desktopcomputer/TFTmonitor•Color inkjetprinter

Software:identical to the MasterScreen Pneumo plus:• Measuring program - DiffusionSBRealtime- DiffusionSB Intrabreath- FRCRebreathing

Practice-orientedpulmonaryfunctiontestingstation for determining static and dynamic lungvolumesviatheclassicHe-Rebreathingmethod and thediffusion capacity via thesingle-breath method. MasterScreen PFTcanbecombinedwithMasterScreenBody,APS Provocation system and IOS ImpulseOscillometry.

JAEGER™ Pneumotach

MasterScreen Body

Bodyplethysmograph - MasterScreen Body

Boxsignal Transducertype piezoresistive/calibratedinshiftvolumeTransducer range 0to±800mLat1000hPaambientpressureAccuracy ±2%Resolution 0.4mLCompensationchamber ReductionofambientpressurevariationsCalibrationvolume 50mL,motor-drivencalibrationpump

Standardbox Volume 830LSize 88x80x185cm(35”x32”x73”)(WxDxH)Weight 120kg(265lbs)

Wheelchairbox Volume 1350LSize 133x100x183cm(52”x39”x72”)(WxDxH)Weight 165kg(364lbs)

Chair Maximalload 150kg(331lbs)Material Box Aluminum/acrylicglassconstruction

Flow-Volume/pressure measurements

Flow

Type PneumotachRange 0to±20L/sAccuracy 0.2to12L/s:±2%or0.2L/s(whicheverisgreater)Resolution 10mL/sResistance <0.05kPa/(L/s)(0.5cmH2O/(L/s))at10L/sCMRR 60dBat50Hz

Volume Type Software IntegrationRange ±20LAccuracy 0.5to8L:±3%or±0.05L(whicheverisgreater)Resolution 1mL

MouthpressureMasterScreenPneumo/PFT/Body/ Body-Diff

Type PiezoResistiveRange ±20kPa(±150mmHg)Accuracy ±2%Resolution 0.01kPa(0.075mmHg)CMRR 60dBat50Hz

MouthpressureMasterScreen IOS

Type piezoresistiveRange ±2kPa(±15mmHg)Accuracy ±2%Resolution 0.003kPa(0.023mmHg)

MasterScreen Body/Diffusion (RT)

Gas Analyzer MasterScreen Body/Diff (RT), MasterScreen Diff (RT):

MultigasAnalyzer(CO/CH4/C2H2): Type Infrared

Range 0to0.33%

Accuracy ±0.003%absoluteor±2%relative(whicheverisgreater)

Resolution 0.0005%

T10-90Sensor <350ms

MasterScreen PFT

Single-Breath Diffusion Realtime

FRC Rebreathing (He)

Gas Analyzer - MasterScreen PFT

HeliumAnalyzer(He): Principle Principleofthermalconductivity

Range 0to10%He

Accuracy ±0.05%Heabsoluteor±2%relative(whicheverisgreater)

Resolution 0.005%He

T10-90Sensor <200ms

CarbonMonoxideAnalyzer(CO): Type Electrochemicalcell

Range 0to0.4%CO

Accuracy ±0.003%COabsoluteor±1%relative(whicheverisgreater)

Resolution 0.0002%CO

T10-90Sensor <40sec

OxygenAnalyzer(O2): Type Electrochemicalcell

Range 0to100%O2

Accuracy ±1.0%O2absolute

Resolution 0.05%O2

T10-90Sensor <12sec

Service life: at least2years

MultigasAnalyzer(CO/CH4/C2H2): Type Infrared

Range 0to0.33%

Accuracy ±0.003%absoluteor±2%relative(whicheverisgreater)

Resolution 0.0005%

T10-90Sensor <350ms

NitrogenMonoxideAnalyzer(NO): Type Electrochemicalcell

Range 0to100ppm

Accuracy 3ppmNOabsolute

Resolution 0.1ppmNO

T10-90Sensor <15sec

Pressure reducer

Diffusionpressurereducerfordemandvalve(single-breathmethod)

Inletpressure 200barOutlet pressure 7bar(fixed)Flow rate atleast500L/minFlow rate atmost700L/min2outlets G1/4internalscrewthread

O2pressurereducer Inletpressure 200barOutlet pressure 2.5bar(fixed)1outlet G1/4internalscrewthread

Othertestgaspressurereducer Inletpressure 200barOutlet pressure 3bar(fixed)1outlet G1/4internalscrewthread

Calibration pump

Volume: 1L/3LAccuracy: +/-12mL

APS pro

Compressor Flow 7L/minPressure 1.0bar

Nebulizer Name PhilipsRespironicsSidestream®

Type CompressedairnebulizerMean mass diameter 3.2μmOutputpower 240mg/min

MasterScreen IOS

Test signal

Type ImpulsesofalternatingdirectionPulseinterval 0.1-10sImpulselength 40msFrequencyrangeofsingleimpulse

0-100Hz

Maximalimpulsepressure 0.3kPa(3cmH2O)Powerspectrum Maximumat5Hzto-25dBat50Hz

Systemverification Referenceimpedance 0.2kPa/(L/s)(2cmH2O/(L/s))Accuracy <±2%

MicroGard™ II Bacterial/Viral Filter

MicroGard™IIB/IIC Application SinglepatientuseLatex Latex-freeproductlnspiratoryResistance <0.04kPa/(L/s)(0.4cmH2O/(L/s))at1L/SExpiratoryResistance <0.04kPa/(L/s)(0.4cmH2O/(L/s))at1L/SFilteringEfficiency >99.999%(NelsonTestLab)FilterVolume 55mL

MicroGard™ IIC Connecting Cone 30mmID.30mmOD.TaperedportsMicroGard™ IIB (Filterwithintegratedmouthpiece)

Connecting Cone 30mmID.Taperedport

MicroGard™ II Filter

APS pro Stand alone with nebulizer

MS Pneumo with Rocc and power supply

MS IOS with APS pro and Trolley (Console)

MS IOS Stand alone with Power Unit

Power supply

PCornotebook Mainsvoltage 100to240VAC;50to60Hz,max1.5APowerinput max.1.5A(dependingonPC/notebook)

MasterScreen Pneumo unheated(handlewithoutpowerunitandwithoutisolatingtransformer)

Mainsvoltage 5VDCPowerinput 1A

MasterScreen Pneumo heated(handlewithpowerunitandwithoutisolatingtransformer)

Mainsvoltage 100/115/125/230VAC,50to60HzPowerinput 20VA

Electricalsafety Protection class IMasterScreen Pneumo with isolating transformer

Mainsvoltage 115/125/230VAC,50to60HzPowerinput 900VAElectricalsafety Protection class I

MasterScreen IOS (Standalonewithpowerunitandwithouttrolley)

Mainsvoltage 100/115/125/230VAC,50to60HzPowerinput 20VAElectricalsafety Protection class I

MasterScreenwithtrolleyandsystembox e.g.MasterScreenPFT/MasterScreenBody/MasterScreenBody-Diff/MasterScreenIOSwithsystembox/APSprowithsystembox

Mainsvoltage 100/115/125/230VAC,50to60HzPowerinput 508VAElectricalsafety Protection class I

APSproStandalone Mainsvoltage 100/115/230VAC,50to60HzPowerinput 10to100VA(Int.)Electricalsafety Protection class I

Moistureprotection Completesystem IP20

Classification of applied parts

MasterScreen Body AppliedPart TypeBMasterScreenBody/Diff(RT) AppliedPart TypeBMasterScreenDiffusion(RT) AppliedPart TypeBMasterScreen PFT AppliedPart TypeBMasterScreenIOSwithsystembox AppliedPart TypeBMasterScreenIOSwithpowerunit AppliedPart TypeBFOptionIOS AppliedPart TypeBMasterScreen Pneumo AppliedPart TypeBFAPSproSAwithsystembox AppliedPart TypeBAPSproSAwithpowerunit/compressor AppliedPart TypeBFOptionAPS AppliedPart TypeB

Category according to MDD 93/42/EEC (2007)

Completesystem ActiveclassIIamedicalproduct

Operating mode

Completesystem(apartfromAPSprowithpowerunit)

Continuousoperation

APSpro(withpowerunitandcompressor)

Intermittentoperation

MasterScreen Diffusion (RT)

Forced Spirometry (FVC)

IOS

Bodyplethysmography

Ambient conditions

CompleteSystem(apartfromAPSpro) Altitude ≤2000mTemperature +10°Cto+34°C(+50°Fto93.2°F)Rel.humidity 15to95%RH,non-condensingAmbientpressure 700to1060hPa(525to795mmHg)

APSproSA(withpowerunit/compressor) Temperature +10°Cto+40°C(+50°Fto104°F)Rel.humidity 30to75%RH,non-condensingAmbientpressure 700to1060hPa(525to795mmHg)

Transport and storage conditions

MasterScreenPneumo/IOS/PFT/Body/Body-Diff,APSpro

Temperature -20°Cto+50°C(-4°Fto122°F)Rel.humidity 15to95%RH,non-condensingAmbientpressure 600to1200hPa(450to900mmHg)

SentrySuite® software

OS Windows®7Professional/Ultimate/Enterprise64bitSP1 andWindows®8.1Pro/Enterprise64bit Forupdates:Windows®7Professional32bitSP1

Otherpatches NoneDataplatform Microsoft®SQLServer®2008R2Edition StandardorEnterpriseeditionLanguages English,English(US),German,French,Italian,Spanish,Portuguese,

Danish,Swedish,Norwegian,Finnish,Polish,Russian.NETframework 3.5forMSSQLServer,4.5forSentrySuiteandSentry.NET

Instrument, workstation and small system server requirements

Computer Manufacturer Dell™

Processor MinimumDualCore™with2.0GHzormore RecommendedQuadCore™processorwith2.8GHzormore

RAM Minimum3GBfor32bitOS,4GBfor64bitOS Recommended4GBfor32bitOS,8GBormorefor64bitOS

Harddrive Minimum100GBHDD,recommended320GBSSD,20GBfreediscspace

Otherdrive DVD-ROM,recommendedDVD-RWNetwork Ifnetworking:100Mbit/sTCP/IP,recommended1000Mbit/sOS Windows®7Professional/Ultimate/Enterprise64bitSP1

andWindows®8.1Pro/Enterprise64bit Forupdates:Windows®7Professional32bitSP1

Interfaces MasterScreen Pneumo unheated

1freeserialRS232or1freeUSB2.0 1freePS/2or1freeUSB2.0

MasterScreen heated 1freeserialRS232or1freeUSB2.0MasterScreen IOS 1freeserialRS232or1freeUSB2.0MasterScreenPFT/Body/Diff 1freePCIeinterfaceAPSpro 1freeserialRS232or1freeUSB2.0Fingerprintreader 1freeUSB2.0;supportedwithWindows®732bitonly

Monitor Screen resolution Minimum1280x900,wideornormalscreenDisplay 32bitcolor

Power Computer/Monitor 100to240VAC,4.6ASmallsystemserver Networkedstations Maximum5stationsincludingserver

APS pro Bronchial Provocation

Review Program for Interpretation

Sentry.NET

Server system requirements

Computer Processor MinimumDualCore™processorwith2.0GHz RecommendedQuadCore™processorwith2.8GHzormore

RAM <100000patients:Minimum3GBforWindows®732bitOS;Minimum8GBfor64bitOS >100000patients:Minimum16GBfor64bitOS

Harddrive PerformancebasedRAIDlevelorSSD <100000patients:Minimum360GBfreediscspace >100000patients:Minimumnative100GBfreediscspace+ 200GBper100000patients;incl.databasebackups:minimum360GBfreediscspace+600GBper100000patients

Otherdrive DVD-ROM,recommendedDVD-RWNetwork 100Mbit/sTCP/IP,recommended1000Mbit/s

OS Windows®7Professional/Ultimate/Enterprise64bitSP1 andWindows®8.1Pro/Enterprise64bit Windows®Server2008R2or2012 Forupdates:Windows®7Professional32bitSP1 Forupdates:Windows®Server2003SP2

Monitor Screen resolution Minimum1024x768,recommended1280x1024Power Server/Monitor 100to240VAC,4.6A

Network hardware

Ethernet Speed Minimum100Mbit/s,recommended1000Mbit/s

Sentry.NET server requirements

Computer Processor MinimumDualCore™processorwith2.0GHzormoreRecommendedQuadCore™processorwith2.8GHzormore

RAM Minimum3GBfor32bitOS,4GBfor64bitOS Recommended4GBfor32bitOS,8GBormorefor64bitOS

Harddrive Minimum100GBHDD,recommended320GBSSD,20GBfreediscspacetosaveatleast10000PFTmeasurements

Otherdrive DVD-ROM,recommendedDVD-RWNetwork 100Mbit/sTCP/IP,recommended1000Mbit/sOS -Windows®7Professional/Ultimate/Enterprise64bitSP1withIIS

7.5(withrestrictednumberofusers) -Windows®8.1Pro/Enterprise64bitwithIIS8.0(withrestricted numberofusers) -Windows®Server®2008R2withIIS7.5or2012withIIS8 -Forupdates:Windows®7Professional32bitSP1withIIS7.5 (withrestrictednumberofusers) -Forupdates:Windows®Server®2003SP2withIIS6

Monitor Screen resolution Minimum1280x900,wideornormalscreenDisplay 32bitcolor

Power Computer/Monitor 100to240VAC,4.6A

Sentry.NET clients requirements

Browser InternetExplorer9forWindows®7SP1andInternetExplorer10forWindows®8.1

Silverlightversion Microsoft®Silverlight5.1

Features at a GlanceMasterScreen Products with SentrySuite® Software

xStandard sStandard,dependingontheexistingmeasuringsytemoOption -notavailable

MS Pneumo MS IOSMS PFT (FRC/

Diff)MS Diff (RT) MS Body

MS Body/Diff (RT)

Report Station

Measurement ProgramsForcedSpirometry(FVC)/SlowSpirometry(SVC)/MVV x x x x x x sSpirometryAnimationPrograms x x x x x x -AirwaysResistancebyROcclusion o o o o o o sImpulseOscillometryIOS o x o o o o sFRCRebreathing(StaticLungVolumes) - - x - o - sSingle-BreathDiffusionRealtime - - x x o x sSingle-BreathDiffusionIntrabreath - - x x o x sNon-invasiveCardiacOutput(C2H2)

* - - o - o - sBodyplethysmography o - o - x x sBronchialChallengewithAPSpro o o o o o o sP0.1 o o o o o o sMIP/MEP o o o o o o s

Calibration ProgramsFlow/VolumeCalibration x x x x x x -GasCalibration - - x x - x -BodyBoxCalibration - - o - x x -

*ThisoptionisnotavailableintheUSA

MS Pneumo MS IOSMS PFT (FRC/

Diff)MS Diff (RT) MS Body

MS Body Diff (RT)

Report Station

Organization ProgramsSentrySuiteDesktop(Settings,Export/Import) x x x x x x xSentrySuiteReview x x x x x x xPatientData x x x x x x xReportOutput(Print,View,QuickView,Save,PDF) x x x x x x xMultiformatOutput(JPG,TIFF,RTF) x x x x x x xAutoInterpretation x x x x x x xATS/ERSQualityCheck x x x x x x xComments/PhysicianInterpretation x x x x x x xReportDesignerforgeneratingcustomizedreports x x x x x x xPredictedValuesandReferenceSetsDesign andEditor

x x x x x x x

OfflineDataInput x x x x x x xSeS Quality Management x x x x x x -Logfileviewer x x x x x x xBackup/Restore x x x x x x xInterConnectivityManagerforinterfacewithJLAB/Vmax/SPCSplatforms

x x x x x x x

MedGraphicsandnSpireImportFunction (consultIFUforcompatibleSWversions)

x x x x x x x

GDTconnection(Germanstandard) x x x x x x xQuestionnaireDesigner o o o o o o oSeSQRemoteTabletQuestionnaire o o o o o o oSecurityandUserAdministration o o o o o o oSentryConnectforHISintegrationviaHL7 o o o o o o oSpirometerDataTransfer(FlowScreen®,SpiroPro®,MicroLoop,MicroLab)

x x x x x x x

SeSSQLDatabaseInterfaceQuery o o o o o o oPatientDataManagement o o o o o o oNetworking o o o o o o xSentry.NETWebApplication o o o o o o o

xStandard sStandard,dependingontheexistingmeasuringsytemoOption -notavailable

Features at a GlanceMasterScreen Products with SentrySuite® Software

Dimensions and Required SpaceMasterScreen Products

0

30

60

90

[cm]120

0306090 [cm]MasterScreen IOS with swivel-mounted tableRequiredSpace:90x120cm(WxD)

MasterScreen PFT/Diffusion with swivel-mounted tableRequiredSpace:90x120cm(WxD)

MasterScreen PFT + Body with swivel-mounted tableRequiredSpace:210x150cm(WxD)

MasterScreen Body with swivel-mounted tableRequiredSpace:180x150cm(WxD)

MasterScreen Body XXL with swivel-mounted tableRequiredSpace:420x170cm(WxD)

MasterScreen Body with swivel-mounted table with legRequiredSpace:230x150cm(WxD)

60cm

77cm12

8cm 185cm

80-147cm

MasterScreen PFT + Body, Front view

Dimensions:

Trolley60x66cm(WxD)

Table*70x50cm(WxD)

Trolley + Table*70x98cm(WXD)

Box Standard88x80x185cm(WxDxH)

Box XXL133x100x183cm(WxDxH)

Ramp63x73cm(WxD)

0

30

60

90

120

[cm]150

0306090120150180210[cm]240 0

30

60

90

120

[cm]150

0306090120150180210 [cm]

0

30

60

90

[cm]120

0306090 [cm]0

30

60

90

[cm]120

0306090 [cm]

0

30

60

90

120

[cm]150

0306090120150180 [cm]

0

30

60

90

120

150

[cm]180

0306090120150180210240270300330360390[cm]420

Ramp Space for wheelchairand support staff

FortheXXLBox,aminimumdoorwaywidthof100cm(40”)isrequired.

CareFusion Germany 234 GmbH Leibnizstrasse 7 97204 Hoechberg Germany

+49 931 4972-0 tel +49 931 4972-423 fax

www.carefusion.de

CareFusion 22745 Savi Ranch Parkway Yorba Linda, CA 92887

800.231.2466 toll-free 714.283.2228 tel 714.283.8493 fax

CAUTION -U.S.FederalLawrestrictsthisdevicetosalebyorontheorderofaphysician.

CareFusion Yorba Linda, CA

carefusion.co.uk

U.K Sales:CareFusion UK 236 LtdThe Crescent, Jays CloseBasingstoke, RG22 4BS, UK

+44 (0) 1256 388599 tel+44 (0) 1256 330860 fax

www.carefusion.co.uk

© 2015 CareFusion Corporation or one of its affiliates. Microsoft, Windows and Server are either registered trademarks or trademarks of Microsoft Corporation in the United States and/or other countries. Dell is a trademark of Dell Computer Corporation in the United States and/or other countries. Core is a trademark of Intel Corporation in the U.S. and/or other countries. Sidestream is a registered trademark of Respironics UK Ltd.All rights reserved. CareFusion, CareFusion logo, SentrySuite, JAEGER, MicroGard, FlowScreen and SpiroPro are trademarks or registered trademarks of CareFusion Corporation or one of its affiliates. All trademarks are property of their respective owners. RC1909 (0810/Qty) L3339, Version 09.00

Pleasenote:notallproducts,servicesorfeaturesofproductsandservicesmaybeavailable inyour localarea.Pleasecheckwithyour localCareFusionrepresentative.