introduction to time-of-flight mass spectrometry and ... · introduction to time-of-flight mass...
TRANSCRIPT
Introduction to Time-of-Flight Mass Spectrometry and DeconvolutionSpectrometry and Deconvolution
&
Nicholas Hall – SpeakerMark Libardoni, Pete Stevens, Joe Binkley
Life Science & Chemical Analysis CentreSt. Joseph, Michigan, USA
e-Seminar Goals
• Overview of a Time-of-Flight mass spectrometer– Comparison to scanning instruments– LECO’s TOF MS Instruments
• Spectral reproducibility• Fast spectral acquisition rates • Increased linear dynamic range
• Deconvolution OverviewDeconvolution Overview• Experimental Studies for:
– Sampling Rate– Deconvolution Capability– Usage as a High-Speed GC and GCxGC detector
Time-of-Flight Mass Analyzerg yA Time-of-Flight (TOF) is a non-scanning mass analyzerwhich emits pulses of ions (or transients) from the source.These pulses are accelerated so that the ions have equal
LECO’s Pegasus TOFMSThese pulses are accelerated so that the ions have equalkinetic energy as they are directed into a flight tube. Sincekinetic energy is equal to 1/2 mv2, where m is the mass ofthe ion and v is the ion velocity, the lower the ion's mass, thegreater the velocity and shorter its flight time. The travel timefrom the ion source through the flight tube to the detector,measured in microseconds, can be transformed to the m/zvalue through the relationships described above. Becauseall ion masses are measured for each transient, TOF massspectrometers offer high sensitivity as well as rapidacquisition. A non-scanning instrument such as TOF MSoffers many advantages including fast acquisition ratesoffers many advantages including fast acquisition rates,exceptional dynamic range, spectral continuity, the ability toacquire full mass range spectra without sacrificing speed orsensitivity, and capability to perform quantitative analysesacross a wide dynamic range in the presence of complexmatrices.matrices.
Quadrupolep
The strength and frequency of the RF field determines whether or not an ion of a certain mass passes through
Molecules
Quadrupole Ion Analyzer
Successful Ion Path
Detector
whether or not an ion of a certain mass passes through the rods (and is counted by the detector) or smashes into a nearby surface. For example, in a 120 volt field at a radio frequency of 2 MHz, only ions of 16 Daltons (Da) will navigate through the rods and into the detector. Heavier or lighter ions do not survive the journey to the
Ion Path
Electron Ion Beam
The quadrupole mass spectrometer is made with 4 rods of hyperbolic surfaces A direct current field is
g j ydetector. In this manner, you can control the mass of the ions that the detector collects.
rods of hyperbolic surfaces. A direct current field is applied to 2 rods and a radio frequency (RF) field is applied to the other 2 rods. These rods generate an electric field through which the ions can move.
Benefits of Quads and TOFsBenefits of Quads and TOFs• Quadrupole • Time-of-Flight
– Inexpensive
Robust
– Excellent Qualitative and Quantitative results in single run
– Robust
– Good for Qualitative and Quantitative work
– No Spectral Bias (skewing)
– Excellent Detection LimitsQuantitative work
– Enhanced Detection Limits when operating in SIM
Excellent Detection Limits
– Very Fast acquisition rates (up to 500 spectra/s)p g ( p p )
Sampling Ratep gQuadrupoles Time-of-Flight
• Scanning Instrument at 10,000 u/sec (~20 Hz)
Duty Cycle Time
• Non-Scanning Instrument– Full Mass Range Acquisition
All Masses are Collected– Duty Cycle Time
• Increase Scan Rate by Limiting Mass Range
– All Masses are Collected– Acquisition Rates up to 500
spectra/s (500 Hz)Limiting Mass Range– (40 – 200 amu)
Sampling Rate is important for properly defined peaks and deconvolution
Proper Sampling Ratesp p g• Literature defines the proper sampling rate of a
chromatographic detector to have an acquisition speed g p q pcapable of delivering at least 10 points across a fully resolved peak to deliver accurate quantitative data
• Therefore if you have a:– 1 second wide peak, you will need a detector speed of 10 Hzp , y p 0– 0.5 second wide peak (500 ms), you will need 20 Hz– 0.2 second wide peak (200 ms), you will need 50 Hz
0 1 second wide peak (100 ms) you will need 100 Hz– 0.1 second wide peak (100 ms), you will need 100 Hz– 0.05 second wide peak (50 ms), you will need 200 Hz
LECO Separation SciencesDelivering the Right Results
P 4D d HT T TOF HTPegasus 4D and HT- Fastest GCMS Available- Market’s only GCxGC-TOFMS
TruTOF HT- 2nd Fastest GCMS Available- All the advantages of TOF at quad
prices
Proper Sampling Ratesp p g• Most samples are complex and do not contain fully resolved
chromatographic peaks – therefore some method of deconvolution i d dis needed.
• LECO has published results showing that 18-20 data points are necessary for accurate quantitative deconvolution of overlapped peaks.
Spectral Bias (Skewing)
500
750
1000
Abu
nd.
4000
6000
8000
ensi
ty
500
750
1000 A
bund
.GC PeakTOFMS Scanning MS
60 80 100 120 1400
250
500
Rel
.
m/z
42.5 43.0 43.50
2000
4000
Inte
Time (sec)60 80 100 120 140
0
250Rel
.
m/z
1000
250
500
750
1000
Rel
. Abu
nd.
2000
4000
6000
8000
Inte
nsity
250
500
750
1000
Rel
. Abu
nd.
60 80 100 120 1400
m/z
750
1000
nd.
42.5 43.0 43.50
Time (sec)
6000
8000
60 80 100 120 1400
m/z
750
1000
nd.
60 80 100 120 1400
250
500
50
Rel
. Abu
n
/
42.5 43.0 43.50
2000
4000
6000
Inte
nsity
Time (sec)60 80 100 120 140
0
250
500
Rel
. Abu
m/z m/z( )m/z
Proper Sampling Ratep p g
Metolachlor 238Malathion 158
40 spectra/s
9 peaks locatedMalathion 158Fenthion 278Chlorpyrifos 199Parathion 291
9 peaks located
DCPA 301Cyanazine 68Isodrin 193T i hl t 269Trichloronat 269
Proper Sampling Ratep p g10 spectra/s
6 peaks located6 peaks located
Proper Sampling Ratep p g5 spectra/s
5 peaks located5 peaks located
Proper Sampling Ratep p g2 spectra/s
0 peaks located0 peaks located
40 spectra/s9 peaks located
10 spectra/s6 peaks located
2 spectra/s0 peaks located
True Signal True Signal DeconvolutionDeconvolutionA correlated set of ionsA correlated set of ions reaches the detector
erC
alip
True Signal Deconvolution
erC
alip
e
True Signal Deconvolution
erC
alip
e
T Si l D l tiA new correlated set of i h th d t t
True Signal Deconvolution
ions reaches the detector
erC
alip
e
T Si l D l ti
The first pack of correlated ions have apexed
True Signal Deconvolution
p pChromaTOF lays down a peak marker
erC
alip
e
T Si l D l tiTrue Signal Deconvolution
erC
alip
e
T Si l D l ti
The second pack of correlated ions have apexed
True Signal Deconvolution
p pChromaTOF lays down another peak marker
erC
alip
e
T Si l D l tiTrue Signal Deconvolution
erC
alip
e
T Si l D l tiTrue Signal Deconvolution
erC
alip
e
CoelutionCoelution of 3 Pesticidesof 3 Pesticides
Fenthion 278Chlorpyrifos 199Parathion 291
< 550 ms >
Spectral Spectral CoelutionCoelution of Pesticidesof Pesticidespp
* *FenthionChlorpyrifosParathion*
*
* ** ***
*
**
**
Spectral Spectral DeconvolutionDeconvolution of Pesticidesof Pesticides
Deconvoluted SpectraDeconvoluted Spectraof Chlorpyrifos
NIST Library Spectraof Chlorpyrifos
For More InformationFor More Information
Contact LECO at:
World Headquarters/United StatesIn United States: 800-292-6141 or 269-985-5496
Outside U.S.A.: 269-983-5531
Email: info@leco comEmail: [email protected]