time of flight mass spectrometry: review
TRANSCRIPT
www.wjpps.com Vol 4, Issue 07, 2015.
614
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
TIME OF FLIGHT MASS SPECTROMETRY: REVIEW
T. Sriveena*, A.Srividya, A.Ajitha and V.Uma Maheswara Rao
Department of Pharmaceutical Analysis and Quality Assurance, CMR College of
Pharmacy, Kandlakoya (v), Medchal road, Hyderabad – 501401, T.S, India.
ABSTRACT
Time-of-flight instruments form a well-established group of mass
spectrometers, with its popularity still increasing. Methods of gating
ion populations and the special requirements in the detection and
digitization of the signals in the TOF-MS are of special importance.
Time of flight mass spectrometry (TOF-MS) has been an attractive
choice of instrument for many years due to its potentially unlimited
m/z range, high-speed acquisition, accurate mass measurement
capability and sensitivity. In which an ion's mass-to-charge ratio is
determined via a time measurement. Ions are accelerated by an electric
field of known strength.
KEY WORDS: TOF-MS, mass to charge ratio, high speed.
INTRODUCTION OF MASS SPECTROMETRY
Mass spectrometry (MS) is an analytical chemistry technique that
helps identify the amount and type of chemicals present in a sample by measuring the mass-
to-charge ratio and abundance of gas-phase ions. A mass spectrum (plural spectra) is a plot of
the ion signal as a function of the mass-to-charge ratio. The spectra are used to determine the
elemental or isotopic signature of a sample, the masses of particles and of molecules, and to
elucidate the chemical structures of molecules, such as peptides and other chemical
compounds. Mass spectrometry works by ionizing chemical compounds to generate charged
molecules or molecule fragments and measuring their mass-to-charge ratios.
In a typical MS procedure, a sample, which may be solid, liquid, or gas, is ionized, for
example by bombarding it with electrons. This may cause some of the sample's molecules to
break into charged fragments. These ions are then separated according to their mass-to-
WWOORRLLDD JJOOUURRNNAALL OOFF PPHHAARRMMAACCYY AANNDD PPHHAARRMMAACCEEUUTTIICCAALL SSCCIIEENNCCEESS
SSJJIIFF IImmppaacctt FFaaccttoorr 55..221100
VVoolluummee 44,, IIssssuuee 0077,, 661144--662255.. RReevviieeww AArrttiiccllee IISSSSNN 2278 – 4357
Article Received on
10 May 2015,
Revised on 02 June 2015,
Accepted on 26 June 2015
*Correspondence for
Author
T. Sriveena
Department Of
Pharmaceutical Analysis
And Quality Assurance,
CMR College of
Pharmacy, Kandlakoya
(v), Medchal road,
Hyderabad – 501401, T.S,
India.
www.wjpps.com Vol 4, Issue 07, 2015.
615
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
charge ratio, typically by accelerating them and subjecting them to an electric or magnetic
field: ions of the same mass-to-charge ratio will undergo the same amount of deflection. The
ions are detected by a mechanism capable of detecting charged particles, such as an electron
multiplier. Results are displayed as spectra of the relative abundance of detected ions as a
function of the mass-to-charge ratio. The atoms or molecules in the sample can be identified
by correlating known masses to the identified masses or through a characteristic
fragmentation pattern.
TIME OF FLIGHT MASS SPECTROMETRY
Time-of-flight mass spectrometry (TOFMS) is a method of mass spectrometry in which
an ion's mass-to-charge ratio is determined via a time measurement. Ions are accelerated by
an electric field of known strength. This acceleration results in an ion having the
same kinetic energy as any other ion that has the same charge. The velocity of the ion
depends on the mass-to-charge ratio. The time that it subsequently takes for the particle to
reach a detector at a known distance is measured. This time will depend on the mass-to-
charge ratio of the particle (heavier particles reach lower speeds). From this time and the
known experimental parameters one can find the mass-to-charge ratio of the ion.
Today's time-of-flight mass spectrometers offer excellent sensitivity and high mass range as
well as high-speed analysis. This has made TOFMS an essential instrument for biological
analysis applications - typically with MALDI and ESI ionization-particularly with the
development of high-resolution and hybrid instruments (for example Q-TOF and TOF-TOF
configurations). The TOFMS suitability for the analysis of fast transient signals has seen it
applied to studying fast gas-phase chemical reactions and to hyphenated flow or injection
analysis techniques such as ICP-TOF and GC-TOF. In addition time-of-flight mass
spectrometry is the dominant instrument for static SIMS.
Fig:1. Time of flight mass spectrometry
www.wjpps.com Vol 4, Issue 07, 2015.
616
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
A time-of-flight mass spectrometer uses the differences in transit time through a drift region
to separate ions of different masses. It operates in a pulsed mode so ions must be produced or
extracted in pulses. An electric field accelerates all ions into a field-free drift region with a
kinetic energy of qV, where q is the ion charge and V is the applied voltage. Since the ion
kinetic energy is 0.5mv2, lighter ions have a higher velocity than heavier ions and reach the
detector at the end of the drift region sooner.
Fig:2.Source of Time of flight mass spectrometry
In the source of a TOF analyzer, a packet of ions is formed by a very fast ionization pulse.
These ions are accelerated into the flight tube by an electric field (typically 2-25 kV) applied
between the sample plate and the charged grid. Since all the ions are accelerated across the
same distance by the same force, they have the same kinetic energy. Because velocity (v) is
dependent upon the kinetic energy (E ) and mass (m) lighter ions will travel faster.
K.E = mv2 ......... (1)
E is determined by the acceleration voltage of the instrument (V) and the charge of the ion (e
kinetic × z). Equation 2 rearranges to give the velocity of an ion (v) as a function of
acceleration voltage and m/z value.
v =√ ...........(2)
After the ions accelerate, they enter a 1 to 2 meter flight tube. The ions drift through this
field free region at the velocity reached during acceleration. At the end of the flight tube they
strike a detector. The time delay (t) from the formation of the ions to the time they reach the
detector dependents upon the length of the drift region (L), the mass to charge ratio of the
ion, and the acceleration voltage in the source
t = / ...........(3)
www.wjpps.com Vol 4, Issue 07, 2015.
617
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
From the above Equation 3 shows that low m/z ions will reach the detector first. The mass
spectrum is obtained by measuring the detector signal as a function of time for each pulse of
ions produced in the source region. Because all the ions are detected, TOF instruments have
very high transmission efficiency which increases the S/N level .
Time-of-flight mass spectrometry equation:
TF = L/v =L m/2zeV
Measurement of ion masses by Time-of-Flight Mass Spectrometry (TOF MS)
Time-of-flight mass spectrometers measure the time a sample molecule requires to fly a
known distance. Charged molecules of the sample are accelerated in an electric field with a
known energy.
The speed a molecule gains is then:
From this equation the mass can be calculated, by measuring the time such a molecule
requires to fly a known distance. Heavy ions will take longer than light ones.
Fig:3.Principle of time of flight mass spectrometry
This picture shows the working principle of a time of flight mass spectrometer. To allow the
ions to fly through the flight path without hitting anything else, all the air molecules have
been pumped out to create an ultra high vacuum. In the vacuum in the MS-200 chamber an
ion can fly on average 600 meters before it will hit an air molecule.
At the end of the ion flight path is a detector which produces electrical pulses as the ions
impact the detector. The time measurement is done by the timing electronics, which applies a
pulse of voltage to accelerate the ions and measures the time between this pulse, and the
electrical impulse from the detector.
www.wjpps.com Vol 4, Issue 07, 2015.
618
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
The MS-200 mass spectrometer has an flight path length of around 0.6m. A molecule with a
mass 26µm (atomic mass units) requires 6 µsec (6 x 10-6
seconds) to fly through the flight
path. For a sufficient distinction between different masses the timing electronics is capable of
measurement with a resolution of 2 nsec (2 x 10-9
seconds). Every 20 µsec the analyte in the
ionization area is accelerated and the masses of the molecules are recorded. In an experiment
of 1 second, 50,000 analysis cycles are performed. Therefore the gathered spectra is a good
representation of the sample.
PRINCIPLE
Fig:4.Basic principle representation of time-of-flight mass spectrometry
1. In a time-of-flight mass spectrometer the ions are formed in a similar manner by electron
bombardment, and the resulting ions accelerated between electrically charged plates.
2. Again, the sample must be a gas or vaporized and is bombarded with an electron beam or
laser beam to knock off electrons to produce positive ions.
3. However, the method of separation due to different m/e (mass/charge) values is then
dependent on how long it takes the ion to travel in the drift region' i.e. the region NOT
under the influence of an accelerating electric field.
4. The ions are accelerated in the same way between positive to negative plates in an electric
field of fixed strength i.e. constant potential difference.
5. The smaller the mass of the ionized particle (ionized atom, fragment or whole molecule)
the shorter the time of flight in the drift region where no electric field operates.
6. This is because for a given accelerating potential difference, a lighter particle is
accelerated more to a higher speed than a heavier ion, so the 'time of flight' down the tube
is shorter.
7. Therefore the ions are distinguished by different flight times NOT by different masses
being brought into focus with a magnetic field BUT the separation by time of flight is still
determined by the m/e (m/z) value of the ion.
www.wjpps.com Vol 4, Issue 07, 2015.
619
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
t = Kinst√(m/q)
t = time of flight, m = mass of ion, q = charge onion,
Kinst = a proportionality constant based on the instrument settings and characteristics e.g. the
electric field strength, length of analyzing tube etc.
Therefore t is proportional to the square root of the mass of the ion for particles carrying the
same charge - the bigger the mass the longer the 'flight time'.
KE = qV
the kinetic energy imparted to the ion is given by its charge x the potential difference of the
accelerating electric field.
The acceleration, for a fixed electric field, results in an ion having the same kinetic
energy (KE) as any other ion of the same charge q but the velocity v of the ion depends on
the m/e (m/z) value.
v = d/t
where v = velocity of accelerated particle in the drift region,
d = length of tube in the drift region. (or t = d/v)
KE = 1/2mv
2,
so the bigger m, the smaller is v in the drift region and hence the basis of detecting ions of
different mass by different 'flight times'.
INSTRUMENTATION AND OPERATION
Fig:5.Instrumentation of time-of-flight mass spectrometry
The TOFMS have subsystems
1. Vacuum system
2. Optical system
3. Source/analyzer
4. Reflectron
5. Electronics/data system
www.wjpps.com Vol 4, Issue 07, 2015.
620
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
Vacuum System
The vacuum system consists of a 52-in.3 coffin-type chamber machined from a block of
aluminum with a clear Lexan top and multiple polymer O-ring sealed vacuum ports (ISO-KF)
machined into its walls. A Pfeiffer combination turbo molecular/drag pump (model TPD011)
is used to evacuate the chamber into the low 106 torr range. With a pumping speed of 10 L/s
and a mass of 2.5 kg, the Pfeiffer pump is the smallest one of its kind that is commercially
available. This type of combination pump can exhaust into the relatively low vacuum of a
diaphragm pump. A KnF-Neuberger diaphragm pump (Model N84.4ANDC) capable of
operating at 1.5 torr and 4.8 L/min is used in this application. The chamber pressure is
measured with a Pfeiffer wide-range vacuum gauge (combination Pirani/cold-cathode gauge).
Optical System
The optical system of a typical MALDI TOFMS is designed to deliver a series of short
ultraviolet (UV) laser pulses to the source region of the TOF analyzer. The sample to be
analyzed is co-deposited with a UV-absorbing substance (matrix) onto a probe and inserted
into the source of the analyzer at the focal point of the optical system. The laser energy is
absorbed by the matrix and transferred to the analyte. Both analyte and matrix molecules are
desorbed, ionized, and accelerated into the TOF analyzer. Standard optical components
(lenses and windows) are used in nearly all commercial and custom TOF mass spectrometers.
In contrast, the APL Suitcase TOF uses fiber-optics to deliver the energy from a pulsed
nitrogen laser featuring a small, sealed plasma cartridge (LSI Model VSL-337, 140 J per
pulse, 5-ns pulse width, 337-nm wavelength, and 1- to 20-Hz pulse rate). This pulsed
nitrogen laser is also the smallest and lightest commercially available one that will work in
this application.
Fiber-optic components simplify the optical mounts and increase the overall ruggedness of
the instrument. The light is coupled into a 200um core UV, transmitted to a U-bracket (Oz
Optics Model UB-01), collimated and launched across the air gap of the U-bracket through a
linear-wedge neutral-density filter, and then coupled back into a second 200 m core UV fiber.
The linear-wedge 0–2 optical-density filter is used to attenuate the laser energy delivered to
the sample. A small lens pigtailed to the end of the second fiber focuses the light through a
vacuum window and onto the center of the sample probe. A simple sliding gimbal mount is
used to align the beam and set the spot size at the sample. Finally, a fast photodiode mounted
near the exit aperture of the laser detects scattered light to be used as the start trigger for the
TOF measurement.
www.wjpps.com Vol 4, Issue 07, 2015.
621
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
Source/Analyzer
Analyte and matrix ions formed in the electric field of the source region are all accelerated to
the same kinetic energy. A flat sample plate held at a high voltage (≈10 kV) separated by a
small distance from an extraction grid at ground potential typically defines a source. After
exiting the source, the ions drift in a field-free region until they strike a detector. Because all
the ions have the same kinetic energy, a more massive or heavier ion will have lower velocity
than a less massive or lighter ion. The velocity v and TOF t of the ions are related to their
mass m by the following equations:
v = [ 2zeEs/m]1/2,
t = [ m/2zeEs]1/2 D,
where z = the number of charges (e) on the ion,
E = electric field in the source region of width s,
D = the length of the drift region.
In a simple linear analyzer, the initial energy, temporal, and spatial distributions of ions in the
source lead to peak spreading at the detector. One method of correcting the initial energy
distribution and decreasing the peak width is to use a “reflectron,” a retarding electric field
that reflects the ions back along their original path to strike a detector placed at the ions’
spatial focus. Ions of the same mass but with slightly more energy enter the reflectron first,
penetrate slightly deeper, and exit later than less energetic ions. As they approach the
detector, the more energetic ions catch up with slower ones and come into a spatial focus at
the detector. The reflectron increases signal resolution, defined as (t/2Δt), by correcting for
the initial energy spread (decreasing Δt and increasing the effective drift length of the
analyzer (increasing t). An ion detector can be placed to the side of the original ion path or a
detector with a center hole can be placed coaxially with the original ion path so that the ions
pass through the center of the detector, are reflected by the ion mirror, and then are
accelerated back toward the active face of the detector. A reflectron with a linearly increasing
potential (constant electric field) corrects ion energy spreads to first order. Deviating from the
linear potentials typically used in commercial instruments to a slightly curved potential can
create higher-order corrections.7,8 APL’s Suitcase TOF analyzer is a miniature, nonlinear
reflectron TOF with a very efficient grid less focusing source and unique detectors that
produce very clean signals.
www.wjpps.com Vol 4, Issue 07, 2015.
622
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
Fine-mesh metallic grids are employed in most mass spectrometers to establish boundaries
between regions of differing electric fields. For example, the high electric field of the source
region is typically separated from the field-free drift region by a grid. In our Suitcase TOF,
we have eliminated all but one grid, which is located a few millimeters from the detector. The
grid less focusing source uses a very small three-element electrostatic lens to extract the ions
with a high field (≈5 kV/mm) and then slow them down and focus them for maximum
transmission through the center of the detector and into the reflectron. Typical reflectrons are
built by stacking many thin metal elements onto insulating rods and applying the reflector
potentials with a resistor network. We have simplified and ruggedized the reflectron by
defining the electrode structure on a flexible circuit board material, rolling it around a
mandrel, and encasing it in a fiber glass cylinder. A surface-mount resistor network is
soldered to pads on the end of the flexible circuit board. Finally, the analyzer uses a custom
micro channel plate (MCP) detector assembly with a novel anode structure that drastically
reduces the signal ringing that is often associated with coaxial detectors. A small polished pin
replaces the standard disk-shaped anode to collect the pulse of electrons generated by the
stack of MCPs. A reduction in the stray capacitance significantly improves the impedance
match between the anode and the 50-Ω transmission line, resulting in much less ambient
electrical noise coupling into the line and virtually no signal ringing.
Reflectron TOF
Fig:6.Representation of Reflectron TOFMS
The kinetic energy distribution in the direction of ion flight can be corrected by using a
reflectron. The reflectron uses a constant electrostatic field to reflect the ion beam toward the
detector. The more energetic ions penetrate deeper into the reflectron, and take a slightly
longer path to the detector. Less energetic ions of the same mass-to-charge ratio penetrate a
shorter distance into the reflectron and, correspondingly, take a shorter path to the detector.
www.wjpps.com Vol 4, Issue 07, 2015.
623
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
The flat surface of the ion detector (typically a micro channel plate, MCP) is placed at the
point where ions with different energies reflected by the reflectron hit a surface of the
detector at the same time counted with respect to the onset of the extraction pulse in the ion
source. A point of simultaneous arrival of ions of the same mass and charge but with different
energies is often referred as time-of-flight focus. An additional advantage to the re-TOF
arrangement is that twice the flight path is achieved in a given length of instrument.
Electronics/Data System
The current version of the Suitcase TOF uses a mix of commercial and custom electronics to
control the instrument and collect the data. Five high-voltage power supply modules are used
to manually control the source, reflectron, and detector voltages. A full-size LeCroy
oscilloscope (Model 9354, 500-MHz analog bandwidth, 500 mega samples/s) digitizes the
TOF signal and averages between 10 and 100 individual shots to produce one TOF mass
spectrum. A very small laptop computer (Panasonic Tough Book) running a National
Instruments Lab View program downloads the data from the oscilloscope and then processes,
formats, stores, and presents the mass spectra. Efforts at greatly reducing the size, weight,
power, and packaging of the electronics are being addressed in the next prototype.
BENEFITS
1. Fastest MS analyzer.
2. Well suited for pulsed ionization methods (method of choice for majority of MALDI
mass spectrometer systems).
3. High ion transmission.
4. MS/MS information from post-source decay.
5. Highest practical mass range of all MS analyzers
ADVANTAGES
1. Ii is Simplicity and Ruggedness.
2. Accessibility of the ion source and Unlimited mass range
3. Major advantages of the TOF technique over quadrapole and sector type analyzers are the
extremely high transmission, the parallel detection of all masses and the unlimited mass
range.
4. The 3-D ion trap time-of-flight mass spectrometer(LCMS-IT-TOF) is used to analyze
mass of components eluting from the HPLC column, and for structural elucidation.
www.wjpps.com Vol 4, Issue 07, 2015.
624
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
5. Time-of-flight (TOF) has quickly established itself as the preferred type of mass analyzer
for the characterization of synthetic macromolecules. TOF combines a high sensitivity
with a broad mass range and a high spectral resolution and accuracy.
DISADVANTAGES
1. Sensitivity
2. Resolution
LIMITATIONS
1. Requires pulsed ionization method or ion beam switching (duty cycle is a factor) .
2. Fast digitizers used in TOF can have limited dynamic range .
3. Limited precursor-ion selectivity for most MS/MS experiments.
APPLICATIONS
1. Very fast GC/MS systems
2. Matrix-assisted laser desorption ionization (MALDI) is a pulsed ionization technique that
is readily compatible with TOF MS.
3. Atom probe tomography also takes advantage of TOF mass spectrometry.
4. Photoelectron photo ion coincidence spectroscopy uses soft photo ionization for ion
internal energy selection and TOF mass spectrometry for mass analysis.
5. Common applications of TOF spectroscopy within pharmaceutical development include
metabolite profiling, surface analysis of organic and inorganic molecules, and biomarker
discovery
6. Matrix-assisted laser desorption/ionization (MALDI) TOF spectroscopy is efficient and
highly sensitive for analyzing polymeric compounds such as protein and polysaccharides.
7. Purchasing considerations for pharmaceutical TOF equipment include the molecular
stability of the compounds to be tested, requirements for instrument size, possible
instrument compatibility with quantification software, and the laboratory.
CONCLUSION
Advances in the technologies of ionization, ion detection as well as high speed signal and
data processing make TOF-MS and attractive alternative to conventional mass analyzers in
specific applications. Developments in ion optics and electronics have substantially removed
the limitations of early TOF instruments. In this many important instrumental and
performance of TOF instruments have been discussed. There is considerable scope for further
www.wjpps.com Vol 4, Issue 07, 2015.
625
Sriveena et al. World Journal of Pharmacy and Pharmaceutical Sciences
development of the facilitating technologies to allow the speed of TOF to be exploited while
maintaining a dynamic range commensurate with scanning MS instruments.
REFERENCES
1. Bhowmick.A. Development of a New High Resolution Reflectron Time-of-Flight Mass
Spectrometer. Bhabha Atomic Research Centre, 2005; 64-67.
2. Chemushevich. Duty cycle improvement for a quadrapole-time-of-flight mass
spectrometer and its use for precursor ion scan. Eur. J. Mass Spectrom.2000; 6: 1-6.
3. Cherokee S. Ion Trap/Ion Mobility/Quadrupole/Time-of-Flight Mass Spectrometry for
Peptide Mixture Analysis. A n a l . C h e m .2001; 73: 177-184.
4. Christian Weickhardt. Time-of-flight mass spectrometry: State-of the-art in chemical
analysis and molecular science. 1996; 15(3): 139-162.
5. Ecelberger . Scott A. Suitcase TOF: A Man-Portable Time-of-Flight Mass Spectrometer.
JOHNS HOPKINS APL TECHNICAL DIGEST. 2014; 25: 13-18.
6. Michael Guilhaus. Principles and Instrumentation in Time-of-flight Mass Spectrometry.
JOURNAL OF MASS SPECTROMETRY. 1995; 30: 1519-1532.
7. Per Hakansson. An Introduction to the Time-of-Flight Technique. Brazilian Journal of
Physics. 1999; 29(3): 422-427.
8. Scott E. Van Bramer. An Introduction to Mass Spectrometry. 1998; 19-20.
9. Timothy J. Cornish. Miniature time-of-flight mass spectrometer using a flexible
circuitboard reflector. Rapid Commun. Mass Spectrom. 2000; 14: 2408-2411.
10. WILEY W.C. Time of flight mass spectrometry with improved resolution. REVIEW OF
SCINTIFIC INSTRUMENTS. 1955; 26: 1150-1155.
11. 11.Basic principle time of flight mass spectrometry diagram - Available at
http//www.docbrown.info
12. Mass spectrometry introduction - Available at https//en.wikipedia.org
13. Structure of time of flight mass spectrometry - Available at
http//www.tissuegroup.chem.vt.edu
14. Time of flight mass spectrometry introduction-Available at
http//www.kore.co.uk/tutorial.htm
15. Time of flight mass analyzer - Available at http// img kid.com
16. Time of flight mass spectrometry applications - Available at www. american
pharmaceutical review
17. Time of flight mass spectrometry Advantages - Available at www.ssi.shimadzu.com