7620 Executive Drive | Eden Prairie, MN 55344-3677 | USA Phone: 952-934-2100 | Fax: 952-934-2737 | Email: [email protected] | www.svta.com Engines for Thin Film Innovation In-Situ 4000 Design Philosophy SVT Associates has recognized these difficulties and has, in conjunction with our own MBE processing lab and UHV system and components expertise, designed a new pyrometer system which solves many of the problems preventing pyrometry from becoming an indispensable tool for both MBE and MOCVD processes. This system combines traditional pyrometry with specular reflectometry to provide a single instrument to monitor both substrate temperature and film thickness in real time. The schematic diagram above shows the layout of the In-Situ 4000 system. The integrated optical head contains the reflectometer illumination source, computer communications interface, and a video camera for ease of aiming and alignment with the substrate. SOLVING SUBSTRATE TRANSPARENCY The problem of low emissivity and substrate transparency at infrared wavelengths is addressed by careful selection of the pyrometry wavelength. The In-Situ 4000 performs single wavelength pyrometry at 950 nm which is short enough to ensure that the substrate is opaque, and long enough to ensure a measurable optical radiation. Ensuring substrate opacity is a key element in addressing several of the previously listed difficulties. An opaque substrate prevents radiation from a filament heater from being transmitted though to the pyrometer. It also prevents other radiation sources in the chamber (effusion cells, ion gauge filaments, etc.) from scattering their light from the rough back side of the substrate into the pyrometer. Since optical absorption and radiation are closely coupled, substrate opacity ensures that the emissivity of the substrate is high enough to provide a measurable signal. Typical substrates such as silicon, GaAs and InP are sufficiently opaque at 950 nm for substrate temperatures typical of epitaxial growth. In-Situ 4000 White Paper Solving the problems of pyrometry and thickness measurement during MBE and MOCVD. S T SVT Associates, Inc. Background Pyrometry is a well established measurement technique and is successfully used in many applications. It has one main feature which makes it attractive for substrate temperature measurement during MBE or MOCVD: non- contact, vacuum compatible temperature sensing. However as those who have used pyrometry for semiconductor applications know, this technique is fraught with problems which greatly limit its usefulness. Some of those problems are listed in the table below. SUBSTRATE TRANSPARENCY Semiconductors by nature are partially transparent for wavelengths below the band gap which are usually in the infrared portion of the spectrum. This transparency causes two severe limitations in pyrometric measurements: low substrate emissivity and optical interference from other hot elements in the chamber. BANDGAP SHIFTS. The band gap of the semiconductor substrate is a strong function of temperature and so the basic assumption of constant emissivity made by standard pyrometers is not valid. VIEWPORT COATING Standard pyrometry computes temperature from the intensity of emitted radiation. If the viewport window through which the pyrometer measures the radiation becomes coated over time, the pyrometer temperature calibration will drift accordingly. OBSCURATION ERRORS This problem is similar to viewport coating in that there are errors introduced into the collection of radiated intensity by misalignment of the optics to the viewport window or other obscuration problems. PICKUP OF STRAY LIGHT Any hot surface in the chamber can radiate light. If that light makes its way into the pyrometer, the pyrometer will not be able to distinguish between “black body” radiation of the substrate itself and that of the stray light. Thus the stray light introduces an error into the measurement usually causing a higher than actual reading. INTERFERENCE EFFECTS When thin films having different index of refraction from the substrate are deposited, the resulting interference effects cause the radiated light to rise and fall with the film thickness even if the actual temperature is constant. This causes temperature measurement errors during film growth and can cause actual temperature fluctuations due to non-constant radiative cooling of the substrate. FILM ABSORPTION EFFECTS If a narrow band semiconductor film (e.g. InAs) is deposited on a wider band substrate (e.g. GaAs) there is a strong change in the optical absorption and emission characteristics of the film/substrate system and often there is a corresponding real change in the temperature. As a result of these problems, pyrometry has experienced only limited success in solving the critically important problem of accurate and repeatable substrate temperature measurement. Video Display Computer Optical Head Vacuum Chamber Wafer Heater

Upload: others

Post on 13-Jun-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 4000 WhitePaper Solvingtheproblemsofpyrometryandthickness T … Paper.pdf · 7620 Executive Drive | Eden Prairie, MN 55344-3677 | USA Phone: 952-934-2100 | Fax: 952-934-2737 | Email:

7620 Executive Drive | Eden Prairie, MN 55344-3677 | USA Phone: 952-934-2100 | Fax: 952-934-2737 | Email: [email protected] | www.svta.com Engines for Thin Film Innovation

In-Situ 4000 Design PhilosophySVT Associates has recognized these difficulties and has, in conjunction with our own MBE processing lab and UHV system and components expertise, designed a new pyrometer system which solves many of the problems preventing pyrometry from becoming an indispensable tool for both MBE and MOCVD processes. This system combines traditional pyrometry with specular reflectometry to provide a single instrument to monitor both substrate temperature and film thickness in real time.

The schematic diagram above shows the layout of the In-Situ 4000 system. The integrated optical head contains the reflectometer illumination source, computer communications interface, and a video camera for ease of aiming and alignment with the substrate.

SolvINg SUbStrAtE trANSPArENcy The problem of low emissivity and substrate transparency at infrared wavelengths is addressed by careful selection of the pyrometry wavelength. The In-Situ 4000 performs single wavelength pyrometry at 950 nm which is short enough to ensure that the substrate is opaque, and long enough to ensure a measurable optical radiation. Ensuring substrate opacity is a key element in addressing several of the previously listed difficulties. An opaque substrate prevents radiation from a filament heater from being transmitted though to the pyrometer. It also prevents other radiation sources in the chamber (effusion cells, ion gauge filaments, etc.) from scattering their light from the rough back side of the substrate into the pyrometer. Since optical absorption and radiation are closely coupled, substrate opacity ensures that the emissivity of the substrate is high enough to provide a measurable signal. Typical substrates such as silicon, GaAs and InP are sufficiently opaque at 950 nm for substrate temperatures typical of epitaxial growth.

In-Situ 4000 White PaperSolving the problems of pyrometry and thickness

measurement during MbE and MocvD.

S TSVT Associates, Inc.

backgroundPyrometry is a well established measurement technique and is successfully used in many applications. It has one main feature which makes it attractive for substrate temperature measurement during MBE or MOCVD: non-contact, vacuum compatible temperature sensing. However as those who have used pyrometry for semiconductor applications know, this technique is fraught with problems which greatly limit its usefulness. Some of those problems are listed in the table below.

SUbStrAtE trANSPArENcy Semiconductors by nature are partially transparent for wavelengths below the band gap which are usually in the infrared portion of the spectrum. This transparency causes two severe limitations in pyrometric measurements: low substrate emissivity and optical interference from other hot elements in the chamber.

bANDgAP ShIFtS. The band gap of the semiconductor substrate is a strong function of temperature and so the basic assumption of constant emissivity made by standard pyrometers is not valid.

vIEWPort coAtINg Standard pyrometry computes temperature from the intensity of emitted radiation. If the viewport window through which the pyrometer measures the radiation becomes coated over time, the pyrometer temperature calibration will drift accordingly.

obScUrAtIoN ErrorS This problem is similar to viewport coating in that there are errors introduced into the collection of radiated intensity by misalignment of the optics to the viewport window or other obscuration problems.

PIckUP oF StrAy lIght Any hot surface in the chamber can radiate light. If that light makes its way into the pyrometer, the pyrometer will not be able to distinguish between “black body” radiation of the substrate itself and that of the stray light. Thus the stray light introduces an error into the measurement usually causing a higher than actual reading.

INtErFErENcE EFFEctS When thin films having different index of refraction from the substrate are deposited, the resulting interference effects cause the radiated light to rise and fall with the film thickness even if the actual temperature is constant. This causes temperature measurement errors during film growth and can cause actual temperature fluctuations due to non-constant radiative cooling of the substrate.

FIlM AbSorPtIoN EFFEctS If a narrow band semiconductor film (e.g. InAs) is deposited on a wider band substrate (e.g. GaAs) there is a strong change in the optical absorption and emission characteristics of the film/substrate system and often there is a corresponding real change in the temperature.

As a result of these problems, pyrometry has experienced only limited success in solving the critically important problem of accurate and repeatable substrate temperature measurement. SVT Associates, Inc. 7620 Executive Drive, Eden Prairie, MN 55344, USA 952-934-2100 FAX 952-934-2737 www.svta.com

In-Situ 4000 White PaperSolving the problems of pyrometry and thicknessmeasurement during MBE and MOCVD.

BackgroundPyrometry is a well established measurement technique and issuccessfully used in many applications. It has one mainfeature which makes it attractive for substrate temperaturemeasurement during MBE or MOCVD: non-contact, vacuumcompatible temperature sensing. However as those who haveused pyrometry for semiconductor applications know, thistechnique is fraught with problems which greatly limit itsusefulness. Some of those problems are listed in the tablebelow.

Substrate transparencySemiconductors by nature are partially transparent forwavelengths below the band gap which are usually in theinfrared portion of the spectrum. This transparency causes twosevere limitations in pyrometric measurements: low substrateemissivity and optical interference from other hot elements inthe chamber.

Bandgap shifts.The band gap of the semiconductor substrate is a strongfunction of temperature and so the basic assumption ofconstant emissivity made by standard pyrometers is not valid.

Viewport coatingStandard pyrometry computes temperature from the intensityof emitted radiation. If the viewport window through which thepyrometer measures the radiation becomes coated over time,the pyrometer temperature calibration will drift accordingly.

Obscuration errorsThis problem is similar to viewport coating in that there areerrors introduced into the collection of radiated intensity bymisalignment of the optics to the viewport window or otherobscuration problems.

Pickup of stray lightAny hot surface in the chamber can radiate light. If that lightmakes its way into the pyrometer, the pyrometer will not beable to distinguish between “black body” radiation of thesubstrate itself and that of the stray light. Thus the stray lightintroduces an error into the measurement usually causing ahigher than actual reading.

Interference effectsWhen thin films having different index of refraction from thesubstrate are deposited, the resulting interference effectscause the radiated light to rise and fall with the film thicknesseven if the actual temperature is constant. This causestemperature measurement errors during film growth and cancause actual temperature fluctuations due to non-constantradiative cooling of the substrate.

Film absorption effectsIf a narrow band semiconductor film (e.g. InAs) is deposited ona wider band substrate (e.g. GaAs) there is a strong change inthe optical absorption and emission characteristics of thefilm/substrate system and often there is a corresponding realchange in the temperature.

As a result of these problems, pyrometry has experienced onlylimited success in solving the critically important problem ofaccurate and repeatable substrate temperature measurement.

In-Situ 4000 Design PhilosophySVT Associates has recognized these difficulties and has, inconjunction with our own MBE processing lab and UHV systemand components expertise, designed anew pyrometer system which solvesmany of the problems preventingpyrometry from becoming anindispensable tool for both MBE andMOCVD processes. This systemcombines traditional pyrometry withspecular reflectometry to provide asingle instrument to monitor both substrate temperature andfilm thickness in real time.

VideoDisplay

Computer

OpticalHead

VacuumChamber

Wafer

Heater

The schematic diagram above shows the layout of the In-Situ4000 system. The integrated optical head contains thereflectometer illumination source, computer communicationsinterface, and a video camera for ease of aiming and alignmentwith the substrate.

Solving Substrate TransparencyThe problem of low emissivity and substrate transparency atinfrared wavelengths is addressed by careful selection of thepyrometry wavelength. The In-Situ 4000 performs singlewavelength pyrometry at 950 nm which is short enough toensure that the substrate is opaque, and long enough toensure a measurable optical radiation. Ensuring substrateopacity is a key element in addressing several of the previouslylisted difficulties. An opaque substrate prevents radiation froma filament heater from being transmitted though to thepyrometer. It also prevents other radiation sources in thechamber (effusion cells, ion gauge filaments, etc.) fromscattering their light from the rough back side of the substrateinto the pyrometer. Since optical absorption and radiation areclosely coupled, substrate opacity ensures that the emissivity

Page 2: 4000 WhitePaper Solvingtheproblemsofpyrometryandthickness T … Paper.pdf · 7620 Executive Drive | Eden Prairie, MN 55344-3677 | USA Phone: 952-934-2100 | Fax: 952-934-2737 | Email:

©2011 Svt Associates, Inc. All rights reserved.

The figure above shows the temperature error which can occur when depositing multilayer thin films. The graph shows the measured substrate temperature during the MBE growth of GaN on sapphire. The lower curve shows the temperature as measured using traditional pyrometry while the upper curve shows the temperature after compensation by the reflectometry system.

These new features of the In-Situ 4000 Process Monitor now allow accurate pyrometry to be performed in an MBE or MOCVD system.

In addition to accurate pyrometry, the reflectometer measures specular reflectance at two wavelengths: 950 nm and 470 nm. Reflectometry is obtained in real-time during deposition so that films which have differing index of refraction from the substrate produce reflectance oscillations in time. These reflectance oscillations can be analyzed in real-time to provide both film thickness and film index of refraction. The In-Situ 4000 has powerful analysis algorithms which can provide accurate growth rate, thickness, and index information. This valuable information can even be fed back to the growth system controller to achieve layer thickness precision improvements over traditional open-loop timed deposition control. The best growth rate and index of refraction measurements are obtained with films of at least l/4 in thickness where l is the reflectometer measurement wavelength (i.e. 950 nm or 470 nm).

The reflectometer uses two LEDs housed in the optical head to illuminate the wafer through the vacuum viewport window and detects the light with the same optical system as the pyrometry. Thus the system need only use a single, normal incidence viewport which is provided in many vacuum systems. The viewport must be large enough to allow room for both the illumination light to enter and the reflected and pyrometric radiated light to exit the chamber. The In-Situ 4000 requires a 2.75 inch Conflat viewport or larger to accomplish this measurement. Another requirement is that the viewport window be located at normal incidence to the substrate so that the illumination beam be specularly reflected back to the instrument. This places some limits on the substrate holders which need to maintain normal incidence alignment during use and especially during wafer rotation. The In-Situ 4000 system has special software features which can allow the measurement to function well even if the substrate tilts strongly during wafer rotation.

The In-Situ 4000 Process Monitor provides the MBE or MOCVD user with a complete tool for obtaining in real-time the two most important process parameters in film growth: temperature and thickness. The system uses only a single viewport window at normal incidence and is specifically designed for MBE and MOCVD processes.

S TSVT Associates, Inc.

SolvINg WINDoW coAtINg ProblEMSThe problem of viewport window coating and optical alignment errors is addressed in the In-Situ 4000 through the use of “two color” or “ratio” pyrometry which measures optical radiation at two wavelengths and takes the ratio of the measured intensities. If the pyrometer is well designed and the film coating the viewport is spectrally neutral, then the temperature may still be calculated since these errors affect both channels equally. Since many of the viewport coating problems deal with “metallic” type films (e.g. Ga, In, As), these films attenuate the light equally between the two wavelengths, the ratio calculation is still valid. If the viewport becomes coated with a dielectric film causing interference effects, then the ratio measurement becomes imbalanced and may not be valid. The intensity ratio vs. substrate temperature will differ for varying substrate materials, so the In-Situ 4000 system allows the flexibility of a look-up table to translate from intensity ratio to reported temperature. This look-up table can be changed for different substrates, or updated to compensate for any calibration errors or instrument drift. The ratio pyrometry feature of the In-Situ 4000 provides the user with a repeatable day-to-day calibration which is independent of viewport coating effects.

SolvINg vAryINg EMISSIvIty The problem of unknown or shifting emissivity is solved using information from the accompanying reflectometer system. Under the conditions of an opaque substrate, a perfectly specular substrate front surface, and perfect uniformity of the deposited films, the emissivity may be measured via reflectometry at the same wavelength as the pyrometry. The relation linking emissivity, ɛ, and reflectometry, R, under these conditions is: ɛ = 1 - R. The In-Situ 4000 provides a 950 nm reflectometer which is matched to the 950 nm pyrometry system thus allowing this “emissivity compensation” to provide an emissivity independent temperature measurement. Thus, if the film being deposited has a index of refraction as the substrate producing interference effects, the reflectometry will observe these changes and the software will correct the error in the temperature calculation. The success of this technique depends upon how closely the materials system conform to the assumptions listed above, namely substrate opacity, surface specularity (smoothness), and film uniformity. If the substrate is not sufficiently opaque or the surface is rough and scatters significant amount of light, the ɛ = 1 - R relation is no longer true. If the deposited film is not uniform, then the reflectance measurement (made at a single point on the substrate) may not be well correlated to the pyrometry measurement (made over a large area of the substrate). In-Situ 4000 White Paper Page 2

SVT Associates, Inc. 7620 Executive Drive, Eden Prairie, MN 55344, USA 952-934-2100 FAX 952-934-2737 www.svta.com

of the substrate is high enough to provide a measurable signal.Typical substrates such as silicon, GaAs and InP aresufficiently opaque at 950 nm for substrate temperaturestypical of epitaxial growth.

Solving Window Coating ProblemsThe problem of viewport window coating and optical alignmenterrors is addressed in the In-Situ 4000 through the use of “two-color” or “ratio” pyrometry which measures optical radiation attwo wavelengths and takes the ratio of the measuredintensities. If the pyrometer is well designed and the filmcoating the viewport is spectrally neutral, then the temperaturemay still be calculated since these errors affect both channelsequally. Since many of the viewport coating problems dealwith “metallic” type films (e.g. Ga, In, As), these films attenuatethe light equally between the two wavelengths, the ratiocalculation is still valid. If the viewport becomes coated with adielectric film causing interference effects, then the ratiomeasurement becomes imbalanced and may not be valid.The intensity ratio vs. substrate temperature will differ forvarying substrate materials, so the In-Situ 4000 system allowsthe flexibility of a look-up table to translate from intensity ratioto reported temperature. This look-up table can be changedfor different substrates, or updated to compensate for anycalibration errors or instrument drift. The ratio pyrometryfeature of the In-Situ 4000 provides the user with a repeatableday-to-day calibration which is independent of viewport coatingeffects.

Solving Varying EmissivityThe problem of unknown or shifting emissivity is solved usinginformation from the accompanying reflectometer system.Under the conditions of an opaque substrate, a perfectlyspecular substrate front surface, and perfect uniformity of thedeposited films, the emissivity may be measured viareflectometry at the same wavelength as the pyrometry. Therelation linking emissivity, ε, and reflectometry, R, under theseconditions is: ε = 1 - R. The In-Situ 4000 provides a 950 nmreflectometer which is matched to the 950 nm pyrometrysystem thus allowing this “emissivity compensation” to providean emissivity independent temperature measurement. Thus, ifthe film being deposited has a index of refraction as thesubstrate producing interference effects, the reflectometry willobserve these changes and the software will correct the errorin the temperature calculation. The success of this techniquedepends upon how closely the materials system conform to theassumptions listed above, namely substrate opacity, surfacespecularity (smoothness), and film uniformity. If the substrateis not sufficiently opaque or the surface is rough and scatterssignificant amount of light, the ε = 1 - R relation is no longertrue. If the deposited film is not uniform, then the reflectancemeasurement (made at a single point on the substrate) maynot be well correlated to the pyrometry measurement (madeover a large area of the substrate).

The figure below shows the temperature error which can occurwhen depositing multilayer thin films. The graph shows themeasured substrate temperature during the MBE growth ofGaN on sapphire. The lower curve shows the temperature asmeasured using traditional pyrometry while the upper curveshows the temperature after compensation by thereflectometry system.

700

750

800

850

0 2000 4000 6000 8000 10000Time (s)

Sub

stra

teTe

mpe

ratu

re (°

C)

CompensatedTemperature

UncompensatedTemperatureGrowth of GaN on sapphire

These new features of the In-Situ 4000 Process Monitor nowallow accurate pyrometry to be performed in an MBE orMOCVD system.

In addition to accurate pyrometry, the reflectometer measuresspecular reflectance at two wavelengths: 950 nm and 470 nm.Reflectometry is obtained in real-time during deposition so thatfilms which have differing index of refraction from the substrateproduce reflectance oscillations in time. These reflectanceoscillations can be analyzed in real-time to provide both filmthickness and film index of refraction. The In-Situ 4000 haspowerful analysis algorithms which can provide accurategrowth rate, thickness, and index information. This valuableinformation can even be fed back to the growth systemcontroller to achieve layer thickness precision improvementsover traditional open-loop timed deposition control. The bestgrowth rate and index of refraction measurements are obtainedwith films of at least λ/4 in thickness where λ is thereflectometer measurement wavelength (i.e. 950 nm or470 nm).

The reflectometer uses two LEDs housed in the optical head toilluminate the wafer through the vacuum viewport window anddetects the light with the same optical system as thepyrometry. Thus the system need only use a single, normalincidence viewport which is provided in many vacuum systems.The viewport must be large enough to allow room for both theillumination light to enter and the reflected and pyrometricradiated light to exit the chamber. The In-Situ 4000 requires a2.75 inch Conflat viewport or larger to accomplish thismeasurement. Another requirement is that the viewportwindow be located at normal incidence to the substrate so thatthe illumination beam be specularly reflected back to theinstrument. This places some limits on the substrate holderswhich need to maintain normal incidence alignment during useand especially during wafer rotation. The In-Situ 4000 systemhas special software features which can allow themeasurement to function well even if the substrate tilts stronglyduring wafer rotation.

The In-Situ 4000 Process Monitor provides the MBE orMOCVD user with a complete tool for obtaining in real-time thetwo most important process parameters in film growth:temperature and thickness. The system uses only a singleviewport window at normal incidence and is specificallydesigned for MBE and MOCVD processes.