1 virgo commissioning status wg1 meeting potsdam, 21 st july 2006
TRANSCRIPT
1
Virgo Commissioning Status
WG1 meeting Potsdam, 21st July 2006
2
Full ITF re-lock
Locks not longer than a few minutes
Mid of March: 7 Watts entering the ITF
Recycling Cavity Power (1 day)
Recycling gain lower than expected
Recycling gain = 25
(instead of 45)
3
Other problems observed
Clipping of the beam discovered at the level of the output telscope
Suspended detection bench
4
Other problems observed
Reflected beam shape very bad
Matching of the beam with the cavities only 88%
Clipping of the beam discovered at the level of the output telscope
ITF locked
5
Tring to fix problems
Centering of the beam on the mirrors
Improvement of the matching
Clipping disappeared
Matching improved (88%96%)
6
Recycling gain from 25 to 40
1 month
Interferometer powers
280 W on the BS
10 times more power than C7 (25 W)
7
Re-lock of the ITF End of May
Lock still not stable
Oscillations around 30-50 Hz
Present “everywhere”
Not clearly connected with any longitudinal loop oscillation
Quite often they caused the unlock of the ITF
8
Locks long enough to see thermal effects
Sideband powersee Julien’s talk
9
dramatic change of the locking parameters in the firts minutes of lock acquisition
quite challenging to keep the ITF locked
Locks long enough to see thermal effects
10
Coupling with alignment fluctuations
0.4 rad
11
locking parameters very sensitive to alignment fluctuations (Locking/Alignment shifts)
small window in which the locking parameters make the ITF more stable
Coupling with alignment fluctuations
12
Useful tools On-line monitor: ugf of the longitudinal loops
P\Q of the error signals
Automatic gain adjustment (Differential ARM loop)
13
Useful tools
Scanning Fabry-Perot on the dark fringe beam
Sidebands always unbalanced, instabilities clearly correlated with one sideband vanishing
14
Easier to identify possible sources of instabilities
Help in tuning locking parameters
Improvements
15
Results
1- Lock acquisition reliable
W
16
Once the lock was stable enough..
further improvement of the locking stability
ITF conditions more repeatable * new alignment procedure tested and implemented
relaxed constraints on the locking parameters
… it was possible to close the automatic alignment:
17
Results
2- Typical locking periods of hours
Automatic alignment ON: 10 loops closed
18
Results
3- Start of low noise operations as in C7
Dark fringe controlled with B1 (OMC on resonance)
Re-allocation to the marionette + low noise coil drivers
More aggressive filters in the longitudinal loops
19
Sensitivity
BS length control noise
PR length control noise (direct coupling)
Arm mirror actuator noise
BS actuator noise
B1 shot noise
B1 electronic noise
Oscillator phase noise
Laser frequency noise ?
19
Presented by Romain at the last
collaboration meeting
Yes!
Noise Budget
Sensitivity
B5_ACp (freq. servo error signal)
20
PRCL sensing noise reduction
Further noise reduction by
switching on B2
Current sensitivity
B2_3f readout noise (4 mW)
B2_3f readout noise with 100 mW
Increasing of the light impinging the photodiode (just done)
Expectations, no direct measurement yet
21
Noise Reduction: near future
Length control noise: Improvement of MICH-PRCL longitudinal loop decoupling
Optimization of the automatic alignment loops (see Maddalena’s talk)
Frequency noise reduction
22
Noise Reduction: already planned actions
Vibration isolation of external detection bench DONE
Actuators noise: plans for new coil drivers
Acoustic noise: plans ready for installation of acoustic enclosure in the laser
Diffused light mitigation: optimization of the benches optical set-up started, need few iterations
23
Conclusions
It seems that we can survive to the thermal effects, but studies are in progress
Short term plans: complete automatic alignment, increase locking robustness (difficult to recover a stable lock after changes in the system, see last week)
Noise hunting just started
24
Matching and beam centering
Spherical telescope
Parabolic telescope
Iterative procedure:
Beam-mirror centering
Astigmatism removal
Beam centering (clipping removal) and matching imply to act on injection bench telescopes
25
Centering techniques
methods:
Image analysis using camera local control (only for beam splitter)
Angle to length coupling
26
What has been intentionally changed?
27
Scan of the B2_3f demdoulation phase(automatic alignment closed, thermal drift over)
28
Laser frequency noise ?
Dark fringe signal coherent with B5_ACp between 240 Hz and 6 kHz
B5_ACp = error signal of the laser frequency stabilization loop (SSFS)
Also coherent with the angular error signals of IMC automatic alignment
B5_ACp superimposed on Dark Fringe Signal using the 444 and 1111 Hz lines
See logbook entry 12603 by M. Evans
Sensitivity
B5_ACp
Conclusion : B1_ACp and B5_ACp see the same noise (presumably laser frequency noise)
29
Global view for 2006
Main task Duration (weeks) In parallel/remarks
Completion of recycled interferometer commissioning
8 High frequency noise hunting robustness increase locking loops
Noise hunting: already planned operations (actuators noise reduction, acoustic enclosure, feet detection lab)
6 High frequency noise huntung
Noise hunting: control noises “first reduction” (angular and longitudinal)
6 High and intermediated frequency noise hunting
Noise hunting: scattered light reduction
4 High and intermediated frequency noise hunting
Total: 6 months (end of 2006)
Start data taking during long week-end as soon as the interferometer is stable and sensitivity better than C7 (september?)
Priority to the high-intermediate frequency range
Goal: factor 10 in the inspiral range
30
Plans for 2007
Science data taking
Shutdown with upgrades to be defined:
Eddy current removal
Thermal compensation implementation
Acoustic mitigation 2nd generation?
Optical table re-shuffling and diffused light mitigation 2nd generation ?
Mode-cleaner mirror replacement ?
Re-commissioning
Noise hunting 2nd phase