Machine-specific QC in Aarhus using EPIDMai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen, Jørgen B.B. Pedersen, Maria F. JensenAarhus University HospitalDepartment of Medical Physics
![Page 2: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/2.jpg)
Rutine QC
2 ▪ www.regionmidtjylland.dk
IMRT VMAT
Machine specific Sweeping gap
Picket fence (PF)
Sweeping gap
Picket fence (PF)
Dose rate versus gantry speed (DRGS)
Dose rate versus MLC speed (DRMLC)
Patient specific PDI
(Delta4)
(PTW729 array)
PDI
(Delta4)
![Page 3: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/3.jpg)
Picket fence, PF
For every MLC and gab…Position:
Relative rotation of EPID and collimator
Intensity (”width”):
3 ▪ www.regionmidtjylland.dk
ThresholdyxI
ThresholdyxIMLC
MLC
MLC
MLC
y
yxIy
gapxp
),(
),(
),(
),(
ThresholdyxI
MLCMLCsum
MLC
yxIgapxI),(
),(),(
![Page 4: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/4.jpg)
PF: ex. ”Eye test”
![Page 5: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/5.jpg)
PF: ex. Deviation in position, tolerance ±0.3mm
5 ▪ www.regionmidtjylland.dk
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0 10 20 30 40 50 60
MLC leaf pair number
De
viat
ion
in p
osi
tion
(m
m)
![Page 6: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/6.jpg)
PF: ex. Deviation in intensity, tolerance ±300
-300
-200
-100
0
100
200
300
400
0 10 20 30 40 50 60
MLC leaf pair number
Dev
iatio
n in
inte
nsity
![Page 7: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/7.jpg)
PF: ex. ”Eye test version 2”
![Page 8: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/8.jpg)
PF: In total ~30 test on each accelerator, one every month
Deviations
012345678
1 2 3 4 5 6 7 8 9 10
Accelerator
Cou
nts
Position Intensity Relative rotation
![Page 9: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/9.jpg)
Dose rate vs. gantry speed, DRGS
9 ▪ www.regionmidtjylland.dk
%100),(),(
),( yxIyxI
yxIMLCOpen
MLCDRGSMLCcorr
1y),(xI
y),(xI )xDeviation(
],1[MLCcorr
MLCcorr
jMLC mj
j
![Page 10: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/10.jpg)
DRGS: ex. typical deviation
-0.2
0
0.2
0.4
0.6
0.8
1
1.2
0 100 200 300 400 500 600 700
y (pixels)
Nor
mal
ised
sca
le
DRGS
Open field
![Page 11: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/11.jpg)
DRGS: ex. cable issue
11 ▪ www.regionmidtjylland.dk
![Page 12: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/12.jpg)
DRGS: In total ~30 test on each accelerator, one every month
Deviation in DRGS test
0
2
4
6
8
10
12
14
1 2 3 4 5 6 7 8 9 10
Accelerator
Cou
nts
![Page 13: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/13.jpg)
Dose rate vs. MLC speed, DRMLC
13 ▪ www.regionmidtjylland.dk
%100),(),(
),( yxIyxI
yxIMLCOpen
MLCDRMLCMLCcorr
1y),(xI
y),(xI )xDeviation(
],1[MLCcorr
MLCcorr
jMLC mj
j
![Page 14: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/14.jpg)
-0.2
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
0 100 200 300 400 500 600 700
Pixel
No
rm
Leaf 30 Leaf 30 baggrund
DRMLC: ex. segment 4 DR120, 0.4cm/s
14 ▪ www.regionmidtjylland.dk
![Page 15: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/15.jpg)
-0.2
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
0 100 200 300 400 500 600 700
Pixel
No
rm
Leaf 30 Leaf 30 baggrund
DRMLC: ex. segment 2 DR600, 2.4cm/s
![Page 16: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/16.jpg)
-0.2
0
0.2
0.4
0.6
0.8
1
1.2
1.4
0 100 200 300 400 500 600 700
Pixel
No
rm
Leaf 30 Leaf 30 baggrund
DRMLC: ex. segment 4 DR120, 0.4cm/s segment 1 DR480, 1.6cm/s
![Page 17: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/17.jpg)
DRMLC: In total ~30 test on each accelerator, one every month
Deviations in DRMLC
0
5
10
15
20
25
30
35
1 2 3 4 5 6 7 8 9 10
Accelerator
Co
un
ts
![Page 18: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/18.jpg)
Time consumption
Measurements: ~10 min Analysis (home-made software): ~1 min Total: ~11 min
18 ▪ www.regionmidtjylland.dk
![Page 19: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/19.jpg)
Effect of chain adjustments on the DRGS and DRMLC tests
Mehtod 3 DRGS and 3 DRMLC test with a loose chain 3 DRGS and 3 DRMLC test with a thight chain
Result, comparing loose and thight chain DRGS: Standard deviation was unchanged DRMLC: Std 0.6 (loose) -> 0.2 (tight)
19 ▪ www.regionmidtjylland.dk
![Page 20: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/20.jpg)
-0.2
0
0.2
0.4
0.6
0.8
1
1.2
1.4
0 100 200 300 400 500 600 700
Pixel
No
rm
Leaf 30 Leaf 30 baggrund
Loose chain: unstable characteristic
![Page 21: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/21.jpg)
-0.2
0
0.2
0.4
0.6
0.8
1
1.2
1.4
0 100 200 300 400 500 600 700
Pixel
No
rm
Leaf 30 Leaf 30 baggrund
Tight chain: Stable characteristic
![Page 22: Www.regionmidtjylland.dk Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,](https://reader035.vdocuments.us/reader035/viewer/2022062516/56649d6f5503460f94a50765/html5/thumbnails/22.jpg)
Discussion
Deviations in PF are covered in next, next session
Deviations in DRGS and DRMLC Chain adjustments have an effect on DRGS and
DRMLC test, but it does not solve everything Cable issues Beam adjustments?
For more information on the calculations and tolerances: Med. Phys. 38, 1425 (2011);
22 ▪ www.regionmidtjylland.dk