Wireless Electrocardiogram Monitor
BioNB 440
Sean Angeles, John Martin Lee and Nick Liu
[email protected],[email protected], [email protected]
12/08/03
I. Introduction
The electrocardiogram (ECG or EKG) is a noninvasive test used to measure the electrical
activity of the heart. An ECG can be used to measure the rate and regularity of heartbeats, the
position of the chambers, the presence of any damage to the heart and the effects of drugs anddevices used to regulate the heart. This procedure is very useful for monitoring people with heart
disease or to provide diagnosis when someone has chest pains or palpitations.
Leads are placed on the body in several pre-determined locations, usually the extremities or the
front of the chest, to provide information about heart conditions. For our final project, weimplemented a wireless electrocardiogram monitor.
II. High-Level Design
The following figure describes the overall high-level design for our ECG monitor:
Figure 1: High-Level Block Diagram
mailto:[email protected]:[email protected]:[email protected]:[email protected]://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/High-Level.JPGmailto:[email protected]:[email protected]:[email protected] -
8/2/2019 Wireless Electrocardiogram Monitor
2/14
Three leads are placed on the subject --usually one on each side of the chest and on the lower
abdomen. This signal is sent to the amplifier where it is amplified by a factor of one thousand.
The signal is then sent to the voltage to frequency converter (VFC), which converts the signal toa frequency so that it can be transmitted. Since we desired the amplifier, VFC and transmitter to
operate using only a single 9 V battery, a separate source splitter circuit was used to provide the
proper voltage to each of the components.
Once the signal is received using the radio receiver, a voltage summer is used to add an offsetvoltage of approximately 800 mV to the signal in order to make the signal entirely positive. This
signal is then amplified by a factor of three so that its maximum value exceeds the threshold of
the required voltage for the frequency to voltage converter (FVC). After the signal is passedthrough the FVC, the output signal is displayed on an oscilloscope.
III. Hardware Design
Transmitter Circuitry:
The following schematic shows the amplifier section of the circuit:
Figure 2: Amplifier Circuit
With the reference lead of the the subject placed to ground, each of the input chest leads is sentto an input of the INA121 instrumentation amplifier. Using a 4.7k resistor, a gain of 11.7 results
from this stage. Following the instrumentation amp, the signal is passed through two 10 uF
capacitors, placed back to back. The capacitors are used to prevent baseline drift in the ECG
signal. Putting two directional capacitors back-to-back forms a bi-directional capacitor. A timeconstant of 0.5 second was chosen to approximate the frequency of a standard ECG signal (a
resistor of 100k can be connected to ground after the capacitor in order to make a time constant
of 0.5 sec, but we found that this resistive element is unnecessary). This section is followed bytwo inverting amplifiers each with a gain of ten. The total gain of this part of the circuit is
approximately equal to one thousand.
-
8/2/2019 Wireless Electrocardiogram Monitor
3/14
The following schematic, obtained from the LM231 data sheet, shows the VFC circuit used in
our monitor:
Equation used to calculated the frequency output.
V_logic is connected to the power supply Vs and all ground are connected to the negative
terminal of the battery. Fout is a square wave of varying frequency with a maximum amplitudeof Vs. A voltage divider is needed (200k and 10k variable resistor connected in series) is needed
at pin 3 to scale the voltage down to around 20 -50 mV.
-
8/2/2019 Wireless Electrocardiogram Monitor
4/14
The output of the voltage divider is connected to the input of the transmitter shown in the circuit
diagram below.
Please note: We added everything on this schematics except for the 22k and the offset
adjust. We neglected those two elements completely.
The following schematic describes the the source-splitting and transmitter section of our project:
The source-splitting amplifier allows three different potential reference: + 4.5V, -4.5V and
Ground. Since the BA1404 can only have +3V as its power supply, we used two diodes to createa total drop of 1.4V and this allows the transmitter to function properly. Another advantage to
this setup is that every elements on this circuit can be powered off a signal 9V battery. Although
not noted on this schematics, one should know that the input to the transmitter should be on theorder of mV (5 - 50mV). Implementing a variable voltage divider to the input is very important.The nice thing about this setup is that one does not need an DC offset circuitry to adjust ECG
signal from the output of the amplifier. Since our VFC is powered between -4.5V and +4.5V, it
has a 4.5V offset already. If the ECG signal is centered at 0V with a swing from -0.5 to +0.5V,then the VFC sees it as 3.5V to 4.5V swing. It is important to know that VFC cannot have
negative voltage as its input. Furthermore, making an inductor at the tunable FM transmitter
-
8/2/2019 Wireless Electrocardiogram Monitor
5/14
range is a painstaking process. We found that by turning a wire 4 times around the pen allows the
signal to be transmitted at 90 MHz.
Receiver Circuitry:
The following schematic shows the voltage summer and amplifier section of the project:
This section consists os a summing amplifier with a gain of one cascaded with an invertingamplifier with a gain of approximately two. The summer takes the input obtained from thereceiver (a FM radio tuned at approximately 90 MHz) and adds it to a constant voltage
obtained using a simple voltage divider. This signal is then sent to an inverting amplifier which
provides a gain of two and inverts the signal after it has been inverted by the summing amplifier.The reason we amplified the signal is due to the fact that FVC (LM231) needs at least a 2V peak-
to-peak amplitude. The signal coming from the radio receiver has a peak-to-peak amplitude
around 500 mV. Increasing the volume will normally increase the signal amplitude but it will
also decrease the signal-to-noise (SNR) ratio. The amplifier we designed at the receiving endincrease the signal to at least 5V of peak-to-peak amplitude, which is sufficient for FVC
conversion.
The following figure shows the FVC circuit, also obtained from the LM231 data sheet:
-
8/2/2019 Wireless Electrocardiogram Monitor
6/14
Equation used to calculated the voltage output due to the frequency response
Please note:
IV. Results
One of the most difficult parts about this project is setting up the VFC (Voltage-to-Frequency)
and FVC (Frequency-to-Voltage). Through many experimentations, we found out that the VFCcan not convert a signal varying more than 16 Hz into pulse train. For example, when we fed in a
100 Hz sine wave, we were getting a constant square as the output (thus a constant DC voltage).
This could be a potential problem for ECG transmission since the QRS peak can occur as fast as
20 to 50 Hz. However, when we fed in square waves of varying frequency into the FVC, wecould get a varying DC voltage as expected. This is probably the reason why we could not
receive a nice-looking ECG waveform on the receiving end.Also, we believe that the signal wasattenuated during the transmission process. We had a difficult time receiving a nice looking
square wave from the FM radio receive. However, we were able to fix that problem by
increasing the volume on the radio to create better rising and falling edges for the FVC.Finally,we noticed that our transmission range is about 10 feet, which is not very useful for a wireless
ECG. The pictures below demonstrate our final result:
-
8/2/2019 Wireless Electrocardiogram Monitor
7/14
Transmitter Section (this includes, pre-amp, source-splitting, FM transmitter and VFC).
http://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/114-1431_IMG.JPG -
8/2/2019 Wireless Electrocardiogram Monitor
8/14
A closer view of VFC.
http://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/114-1443_IMG.JPG -
8/2/2019 Wireless Electrocardiogram Monitor
9/14
A different angle of transmitter circuitry.
http://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/114-1441_IMG.JPG -
8/2/2019 Wireless Electrocardiogram Monitor
10/14
The receiver circuit.
http://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/114-1436_IMG.JPG -
8/2/2019 Wireless Electrocardiogram Monitor
11/14
FM Radio Receiver and the output of the ECG waveform. The picture shown is on a 250ms
division. The QRS-peak and the T-wave are extremely visible. (we are very pleased with theresult. For awhile, we were worried that we might not be able to detect the QRS peak).
http://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/114-1456_IMG.JPG -
8/2/2019 Wireless Electrocardiogram Monitor
12/14
Another view
http://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/ECG1.jpg -
8/2/2019 Wireless Electrocardiogram Monitor
13/14
Overall setup
-
8/2/2019 Wireless Electrocardiogram Monitor
14/14
Link to a video that demonstrate its operation
V. Conclusions
Overall, our wireless ECG monitor was able to detect and transmit the basic elements of a ECG
waveform, such as the QRS-complex and the T-wave. The two major problems of this design arethe transmitter and the voltage-to-frequency conversion. To solve the first problem, one could
employ a multi-channel digital transmitter, since digital transmitter are often more robust, easy to
manipulate, and transmit at a much higher frequency (less interruption). Furthermore,transmitting bits are much more reliable than analog voltage data. On the other hand, there are
several ways to go about solving the second problem; we could use a one-bit digital-to-analog
converter or a sigma-delta converter instead of a VFC. Oftentimes, a sigma-delta converter can
convert analog input into digital data at a much higher rate. Finally, if the VFC is somehowdesired, one might consider matching the input capacitance (CIn) versus the output load
capacitance (CL). One could attempt to match these two capacitors in order to increase theconversion rate. In conclusion, given the amount of time we had, this was a very intense butrewarding project.
VI. References
LM231.pdf (VFC and FVC)
LM158.pdf (Operational Amplifier)
INA121.pdf (Instrumentation Amplifier)
http://www.medicinenet.com/Electrocardiogram_ECG_or_EKG/article.htm
http://www.nlm.nih.gov/medlineplus/ency/article/003868.htm
http://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/demo.AVIhttp://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/LM231.pdfhttp://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/LM158.pdfhttp://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/ina121.pdfhttp://www.medicinenet.com/Electrocardiogram_ECG_or_EKG/article.htmhttp://www.nlm.nih.gov/medlineplus/ency/article/003868.htmhttp://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/demo.AVIhttp://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/demo.AVIhttp://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/LM231.pdfhttp://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/LM158.pdfhttp://courses.cit.cornell.edu/bionb440/FinalProjects/f2003/nwl2/Final%20Webpage/ina121.pdfhttp://www.medicinenet.com/Electrocardiogram_ECG_or_EKG/article.htmhttp://www.nlm.nih.gov/medlineplus/ency/article/003868.htm