Technical specification FLUXUS® F608**-A2 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 1 FLUXUS F608 supported by handle Measurement with transducers mounted with the portable Variofix VP Measurement with the flow transmitter fixed to the pipe with the QuickFix pipe mounting fixture FLUXUS F608 FLUXUS F608 Portable ultrasonic flow measurement of liquids in hazardous areas Portable instrument for non-invasive, quick ultrasonic flow measurement with clamp-on technology for all types of piping Features Precise bi-directional and highly dynamic flow measure- ment with the non-invasive clamp-on technology High precision at fast and slow flow rates, high tem- perature and zero point stability Portable, easy-to-use flow transmitter with 2 flow chan- nels, multiple inputs/outputs, an integrated data logger with a serial interface Extremely resistant carbon fiber housing Covered by ATEX/IECEx zone 2 certification Compact and very lightweight, allowing the measuring system to be easily carried as personal luggage, e.g. for offshore visits Water tight; resistant against oil, many liquids and dirt Li-Ion battery provides up to 25 hours of measurement operation Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results User-friendly design Transducers available for a wide range of inner pipe di- ameters and fluid temperatures Rugged transducers (ATEX/IECEx zone 1 and 2, resis- tant to rough environments, dust and humidity) Robust, water-tight (IP67) transport case with compre- hensive accessories HybridTrek automatically switches between transit time and NoiseTrek mode of measurement when high partic- ulate flows are encountered QuickFix for fast mounting of the flow transmitter in diffi- cult conditions Measurement is unaffected by fluid density, viscosity and solid content (max. 10 % of volume) Applications Designed for the following industries: Upstream (on- and offshore) Midstream and downstream (pipelines and refineries) Chemical industry Energy sector (e.g. HVAC, geothermal, power plants)

Upload: vanduong

Post on 17-Sep-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

Technical specification

FLUXUS® F608**-A2

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 1

FLUXUS F608 supported by handle

Measurement with transducers mounted with the portable Variofix VP

Measurement with the flow transmitter fixed to the pipe with the QuickFix pipe mounting fixture

FLUXUS F608

FLUXUS F608

Portable ultrasonic flow measurement of liquids in hazardous areas

Portable instrument for non-invasive, quick ultrasonic flow measurement with clamp-on technology for all types of piping

Features

• Precise bi-directional and highly dynamic flow measure-ment with the non-invasive clamp-on technology

• High precision at fast and slow flow rates, high tem-perature and zero point stability

• Portable, easy-to-use flow transmitter with 2 flow chan-nels, multiple inputs/outputs, an integrated data logger with a serial interface

• Extremely resistant carbon fiber housing

• Covered by ATEX/IECEx zone 2 certification

• Compact and very lightweight, allowing the measuring system to be easily carried as personal luggage, e.g. for offshore visits

• Water tight; resistant against oil, many liquids and dirt

• Li-Ion battery provides up to 25 hours of measurement operation

• Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results

• User-friendly design

• Transducers available for a wide range of inner pipe di-ameters and fluid temperatures

• Rugged transducers (ATEX/IECEx zone 1 and 2, resis-tant to rough environments, dust and humidity)

• Robust, water-tight (IP67) transport case with compre-hensive accessories

• HybridTrek automatically switches between transit time and NoiseTrek mode of measurement when high partic-ulate flows are encountered

• QuickFix for fast mounting of the flow transmitter in diffi-cult conditions

• Measurement is unaffected by fluid density, viscosity and solid content (max. 10 % of volume)

Applications

Designed for the following industries:

• Upstream (on- and offshore)

• Midstream and downstream (pipelines and refineries)

• Chemical industry

• Energy sector (e.g. HVAC, geothermal, power plants)

FLUXUS® F608**-A2 Technical specification

Table of contents

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-232

Function ........................................................................................................................................................... 3Measurement principle...................................................................................................................................... 3Calculation of volumetric flow rate .................................................................................................................... 3Number of sound paths..................................................................................................................................... 4Typical measurement setup .............................................................................................................................. 5

Flow transmitter .............................................................................................................................................. 6Technical data................................................................................................................................................... 6Dimensions ....................................................................................................................................................... 8Standard scope of supply.................................................................................................................................. 8Adapters (optional)............................................................................................................................................ 9

Transducers................................................................................................................................................... 11Transducer selection....................................................................................................................................... 11Transducer order code.................................................................................................................................... 12Technical data................................................................................................................................................. 13

Transducer mounting fixture ....................................................................................................................... 19

Coupling materials for transducers ............................................................................................................ 20

Connection systems ..................................................................................................................................... 21Transducer cable ............................................................................................................................................ 21

Clamp-on temperature probe (optional) ..................................................................................................... 22

Wall thickness measurement (optional) ..................................................................................................... 24

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 3

Technical specification FLUXUS® F608**-A2

Function

Measurement principle

Transit time difference principle

In order to measure the flow of a fluid in a pipe, ultrasonic signals are used, employing the transit time differ-ence principle. Ultrasonic signals are emitted by a transducer installed on the pipe and received by a secondtransducer. These signals are emitted alternately in the flow direction and against it.

As the fluid in which the signals propagate is flowing, the transit time of the ultrasonic signals in the flow direc-tion is shorter than against the flow direction.

The transit time difference, ∆t, is measured and allows the flowmeter to determine the average flow velocityalong the propagation path of the ultrasonic signals. A flow profile correction is then performed in order to ob-tain the area averaged flow velocity, which is proportional to the volumetric flow rate.

Two integrated microprocessors control the entire measuring process. This allows the flowmeter to removedisturbance signals, and to check each received ultrasonic wave for its validity which reduces noise.

HybridTrek

If the gaseous or solid content in the fluid increases occasionally during measurement, a measurement withthe transit time difference principle is no longer possible. NoiseTrek mode will then be selected by the flowme-ter. This measurement method allows the flowmeter to achieve a stable measurement even with high gas-eous or solid content.

The transmitter can switch automatically between transit time and NoiseTrek mode without any changes tothe measurement setup.

Calculation of volumetric flow rate

= kRe . A . ka . ∆t/(2 . tfl)

where

Path of the ultrasonic signal Transit time difference ∆t

- volumetric flow ratekRe - fluid mechanics calibration factorA - cross-sectional pipe areaka - acoustical calibration factor∆t - transit time differencetfl - transit time in the fluid

4 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Number of sound paths

The number of sound paths is the number of transits of the ultrasonic signal through the fluid in the pipe. De-pending on the number of sound paths, the following methods of installation exist:

• reflection arrangementThe number of sound paths is even. Both of the transducers are mounted on the same side of the pipe. Correct positioning of the transducers is easier.

• diagonal arrangementThe number of sound paths is odd. Both of the transducers are mounted on opposite sides of the pipe. In the case of a high signal attenuation by the fluid, pipe and coatings, diagonal arrangement with 1 sound path will be used.

The preferred method of installation depends on the application. While increasing the number of sound pathsincreases the accuracy of the measurement, signal attenuation increases as well. The optimum number ofsound paths for the parameters of the application will be determined automatically by the transmitter.

As the transducers can be mounted with the transducer mounting fixture in reflection arrangement or diagonalarrangement, the number of sound paths can be adjusted optimally for the application..

a - transducer distance

Reflection arrangement, number of sound paths: 2

Diagonal arrangement, number of sound paths: 3

Diagonal arrangement, number of sound paths: 1 Diagonal arrangement, number of sound paths: 1,negative transducer distance

a

a

a > 0 a < 0

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 5

Technical specification FLUXUS® F608**-A2

Typical measurement setup

zone 2

zone 2/1

!

" # " $ %

transducers

outputs

zone 2

outside of explosive atmosphere

temperature probes

output adapter

(power adapter)

transmitter

!

" # " $ %

zone 2 zone 1

outputs

outside of explosive atmosphere

output adapter

(power adapter)

transducers

transmitter

6 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Flow transmitter

Technical data

FLUXUS F608**-A2design portable, zone 2

measurementmeasurement principle transit time difference correlation principle,

automatic NoiseTrek selection for measurements with high gaseous or solid contentflow velocity 0.01...25 m/srepeatability 0.15 % of reading ±0.01 m/sfluid all acoustically conductive liquids with < 10 % gaseous or solid content in volume (transit time difference

principle)temperature compensation corresponding to the recommendations in ANSI/ASME MFC-5.1-2011

accuracy1

with standard calibration ±1.6 % of reading ±0.01 m/swith advanced calibration (optional)

±1.2 % of reading ±0.01 m/s

with field calibration2 ±0.5 % of reading ±0.01 m/sflow transmitterpower supply 100...230 V/50...60 Hz (power supply unit, outside of explosive atmosphere)

10.5...15 V DC (socket at transmitter, with power adapter (optional))integrated battery

integrated battery Li-Ion, 7.2 V/6.2 Ah- operating time > 14 h (without outputs, inputs and backlight)

> 25 h (1 measuring channel, ambient temperature > 10 °C, without outputs, inputs and backlight)power consumption < 6 W (with outputs, inputs and backlight)number of flow measuring channels

2

damping 0...100 s, adjustablemeasuring cycle (1 channel) 100...1000 Hzresponse time 1 s (1 channel), option: 70 mshousing material PA, TPS, PC, Polyester, stainless steeldegree of protection accord-ing to IEC/EN 60529

IP65

dimensions see dimensional drawingweight 2.2 kgfixation QuickFix pipe mounting fixtureambient temperature -10...+60 °Cdisplay 2 x 16 characters, dot matrix, backlightmenu language English, German, French, Dutch, Spanishexplosion protection

ATEX/IECEx

categoryEPLzone

gas: 3G dust: 2D Gc Db 2 21

marking

Ex nA nC [ic] IIC (T6)T4 GcEx tb IIIC T 100 °C DbTa -10...+(50)60 °C

certification ATEX IBExU10ATEX1067certification IECEx IECEx IBE 12.0006type of protection gas: non sparking

dust: protection by enclosuretemperature inputs: intrinsic safety

intrinsic safety parameters Um = 16 V DC

intrinsically safe inputs: Uo = 22 V, Io = 6 mA, Po = 33 mW, Co = 450 nF, Lo = 10 mH,Ci = 1.8 nF, Li = 10 μH

1 for transit time difference principle, reference conditions and v > 0.15 m/s2 reference uncertainty < 0.2 %

II3GII2D

0637

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 7

Technical specification FLUXUS® F608**-A2

measuring functionsphysical quantities volumetric flow rate, mass flow rate, flow velocity,

heat flow (if temperature inputs are installed)totalizer volume, mass, optional: heat quantitycalculation functions average, difference, sumdiagnostic functions sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit timescommunication interfacesdiagnostic interfaces - RS232

- USB (with adapter)serial data kitsoftware - FluxDiagReader: download of measured values and parameters, graphical presentation

- FluxDiag (optional): download of measurement data, graphical presentation, report generation

- FluxSubstanceLoader: upload of fluid data setsсable RS232adapter RS232 - USBdata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity > 100 000 measured valuestransport casedimensions 500 x 400 x 190 mmoutputs

The outputs are galvanically isolated from the transmitter.number- analog outputs max. 4

0, 2 or 4 active current outputs or passive current outputs or frequency outputs or2 active current outputs and 2 passive current outputs or2 active current outputs and 2 frequency outputs or2 passive current outputs and 2 frequency outputs

- binary outputs max. 4accessories output adapter (necessary, option)

current outputrange 0/4...20 mAaccuracy 0.1 % of reading ±15 μAactive output Rext < 200 Ωpassive output Uext = 4...9 V, depending on Rext

Rext < 200 Ωfrequency output

range 0...5 kHzopen collector 24 V/4 mA

binary outputoptorelay 26 V/100 mAbinary output as alarm output- functions limit, change of flow direction or errorbinary output as pulse output mainly for totalizing- pulse value 0.01...1000 units- pulse width 1...1000 msinputs

The inputs are galvanically isolated from the transmitter.number max. 4accessories input adapter (if number of inputs > 2)

temperature input (intrinsic safety)type Pt100/Pt1000connection 4-wirerange -150...+560 °Cresolution 0.01 Kaccuracy ±0.01 % of reading ±0.03 K

FLUXUS F608**-A2

8 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Dimensions

Standard scope of supply

in mm

F608 Standard F608 Energy F608 Double Energyapplication flow measurement of liquids

2 independent measuring channelstemperature-compensated calculation of mass flow rateintegrated heat flow computer for monitoring of energy flowssimultaneous monitoring of flow and energy flow

simultaneous monitoring of 2 energy flows, e.g. heating sys-tems, heat exchangers)

outputspassive current output 2 2 2binary output 2 2 2inputstemperature input - 2 4accessoriestransport case x x xpower supply unit, mains cable x x xbattery x x xpower adapter1 - - -output adapter1 - - -input adapter - - 2QuickFix pipe mounting fixture for transmitter

x x x

serial data kit x x xmeasuring tape x x xuser manual,safety instructions,Quick start guide

x x x

connector board at the upper side of the transmitter

1 if required, to be ordered separately

!

" # " $ %

22659

213

! $ ! "

& ' ( & '

! !

! $ ! "

& ' ( & '

! !

) ( & '

! $ ! "

& ' ( & '

! !

) ( & '

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 9

Technical specification FLUXUS® F608**-A2

Adapters (optional)

Technical data

output adapter power adaptertechnical type OA608A2 PA608A2connection voltage 10.5...15 V DCdimensions see dimensional drawingweight kg 0.26 0.26materialhousing polyestergasket siliconedegree of protection according to IEC/EN 60529

IP66

ambient temperaturemin. °C -20max. °C +90explosion protection

ATEX

zone 2marking

II3G Ex nA IIC T6 GcTa -10...+60 °C

type of protection non sparking

! $

& ' ( & '

! !

) ( & '

! "

output adapter

input adapter

transducersmeasuring channel A

RS232 outputsinputs

transducersmeasuring channel B

power supply unit/battery charging unit

power adapter

10 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Dimensions

Terminal assignment

output adapter power adapter

in mm

output adapter power adapter

110 56

75

thread: 3x M20 x 1.5

56

75

80

thread: M20 x 1.5

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 11

Technical specification FLUXUS® F608**-A2

Transducers

Transducer selection

transducer order code

FSG 400 500 4000 6500

FSK 100 200 2000 6500

FSM 50 100 1000 3400

FSP 25 50 400 600

FSQ 10 25 150 400

5 10 50 100 500 1000 5000inner pipe diameter [mm]

transducer order code

FSG 11

FSK 5

FSM 2

FSP 1

FSQ 0.6

5 10 15 20 25 30pipe wall thickness [mm]

recommended possible

12 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Transducer order code

1, 2 3 4 5, 6 7, 8 9...11 12, 13 no. of character

tra

nsdu

cer

tra

nsdu

cer

fre

quen

cy

- ambi

ent

tem

pera

ture

expl

osio

n pr

ote

ctio

n

conn

ectio

n sy

stem

- exte

nsio

n ca

ble

/ optio

n

description

FS set of ultrasonic flow transducers for liquids measurement, shear wave

G 0.2 MHz

K 0.5 MHz

M 1 MHz

P 2 MHz

Q 4 MHz

N normal temperature range

E extended temperature range (FSM, FSP, FSQ)

A1 ATEX zone 1/IECEx zone 1

A2 ATEX zone 2/IECEx zone 2

NL with Lemo connector

XXX cable length in m, for max. length of extension cable see page 21 (connector outside of ATEX zone 1/IECEx zone 1)

LC long transducer cable

example

FS M - N A2 NL - 010 shear wave transducer 1 MHz, normal temperature range, ATEX zone 2/IECEx zone 2, connection system NL with Lemo connec-tor and extension cable 10 m

- - /

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 13

Technical specification FLUXUS® F608**-A2

Technical data

Shear wave transducers (zone 1)

technical type CDG1NW1 CLG1NW1 CDK1NW1 CLK1NW1order code FSG-NA1NL FSG-NA1NL/LC FSK-NA1NL FSK-NA1NL/LCtransducer frequency MHz 0.2 0.5

inner pipe diameter dmin. extended mm 400 100min. recommended mm 500 200max. recommended mm 4000 2000max. extended mm 6500 6500pipe wall thicknessmin. mm 11 5materialhousing PEEK with stainless steel cap and

transducer shoe 304 (1.4301)PEEK with stainless steel cap and transducer shoe 304 (1.4301)

contact surface PEEK PEEKdegree of protection according to IEC/EN 60529

IP65 IP66

transducer cabletype 1699 1699 1699 1699length m 5 9 5 9dimensionslength l mm 136.5 136.5width b mm 59 59height h mm 90.5 90.5dimensional drawing

ambient temperaturemin. °C -40 -40max. °C +130 +130temperature compen-sation

x x

explosion protection

ATEX/IECEx

categoryEPLzone

gas: 2G dust: 2DGb Db1 21

gas: 2G dust: 2DGb Db1 21

explosion protection temperature (pipe surface)min. °C -55 -55max. °C +180 +180marking

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling,

non sparkingdust: protection by enclosure

gas: powder filling,non sparkingdust: protection by enclosure

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637

14 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

technical type CDM2NW1 CLM2NW1 CDP2NW1 CLP2NW1 CDQ2NW1 CLQ2NW1order code FSM-NA1NL FSM-NA1NL/LC FSP-NA1NL FSP-NA1NL/LC FSQ-NA1NL FSQ-NA1NL/LCtransducer frequency MHz 1 2 4

inner pipe diameter dmin. extended mm 50 25 10min. recommended mm 100 50 25max. recommended mm 1000 400 150max. extended mm 3400 600 400pipe wall thicknessmin. mm 2 1 0.6materialhousing PEEK with stainless steel cap and

transducer shoe 304 (1.4301)PEEK with stainless steel cap and transducer shoe 304 (1.4301)

PEEK with stainless steel cap and transducer shoe 304 (1.4301)

contact surface PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP66 IP66 IP65

transducer cabletype 1699 1699 1699 1699 1699 1699length m 4 9 4 9 3 9dimensionslength l mm 84 84 70width b mm 40 40 30height h mm 59 59 47.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +130 +130 +130temperature compen-sation

x x x

explosion protection

ATEX/IECEx

categoryEPLzone

gas: 2G dust: 2DGb Db1 21

gas: 2G dust: 2DGb Db1 21

gas: 2G dust: 2DGb Db1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +180 +180 +180marking

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling,

non sparkingdust: protection by enclosure

gas: powder filling,non sparkingdust: protection by enclosure

gas: powder filling,non sparkingdust: protection by enclosure

l

hb

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 15

Technical specification FLUXUS® F608**-A2

Shear wave transducers (zone 1, extended temperature range)

technical type CDM2EW5 CLM2EW5 CDP2EW5 CLP2EW5 CDQ2EW5 CLQ2EW5order code FSM-EA1NL FSM-EA1NL/LC FSP-EA1NL FSP-EA1NL/LC FSQ-EA1NL FSQ-EA1NL/LCtransducer frequency MHz 1 2 4

inner pipe diameter dmin. extended mm 50 25 10min. recommended mm 100 50 25max. recommended mm 1000 400 150max. extended mm 3400 600 400pipe wall thicknessmin. mm 2 1 0.6materialhousing PI with stainless steel cap and

transducer shoe 304 (1.4301)PI with stainless steel cap and transducer shoe 304 (1.4301)

PI with stainless steel cap and transducer shoe 304 (1.4301)

contact surface PI PI PIdegree of protection according to IEC/EN 60529

IP66 IP66 IP56

transducer cabletype 6111 6111 6111 6111 6111 6111length m 4 9 4 9 3 9dimensionslength l mm 84 84 70width b mm 40 40 30height h mm 59 59 47.5dimensional drawing

ambient temperaturemin. °C -30 -30 -30max. °C +200 +200 +200temperature compen-sation

x x x

explosion protection

ATEX/IECEx

categoryEPLzone

gas: 2G dust: 2DGb Db1 21

gas: 2G dust: 2DGb Db1 21

gas: 2G dust: 2DGb Db1 21

explosion protection temperature (pipe surface)min. °C -45 -45 -45max. °C +225 +225 +225marking

Ex q IIC T6...T2 GbEx tb IIIA TX Db

Ex q IIC T6...T2 GbEx tb IIIA TX Db

Ex q IIC T6...T2 GbEx tb IIIA TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling,

non sparkingdust: protection by enclosure

gas: powder filling,non sparkingdust: protection by enclosure

gas: powder filling,non sparkingdust: protection by enclosure

l

hb

lh

bl

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

16 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Shear wave transducers (zone 2)

technical type CDG1NH1 CLG1NH1 CDK1NH1 CLK1NH1order code FSG-NA2NL FSG-NA2NL/LC FSK-NA2NL FSK-NA2NL/LCtransducer frequency MHz 0.2 0.2 0.5 0.5

inner pipe diameter dmin. extended mm 400 400 100 100min. recommended mm 500 500 200 200max. recommended mm 4000 4000 2000 2000max. extended mm 6500 6500 6500 6500pipe wall thicknessmin. mm 11 11 5 5materialhousing PEEK with stainless steel

cap and transducer shoe 304 (1.4301)

PEEK with stainless steel cap and transducer shoe 304 (1.4301)

PEEK with stainless steel cap and transducer shoe 304 (1.4301)

PEEK with stainless steel cap and transducer shoe 304 (1.4301)

contact surface PEEK PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP65 IP65 IP66 IP66

transducer cabletype 1699 1699 1699 1699length m 5 9 5 9dimensionslength l mm 136.5 136.5 136.5 136.5width b mm 59 59 59 59height h mm 90.5 90.5 90.5 90.5ambient temperaturemin. °C -40 -40 -40 -40max. °C +130 +130 +130 +130temperature compen-sation

x x x x

explosion protection

ATEX/IECEx

categoryEPLzone

gas: 3G dust: 2DGc Db2 21

gas: 3G dust: 2DGc Db2 21

gas: 3G dust: 2DGc Db2 21

gas: 3G dust: 2DGc Db2 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55 -55max. °C gas: +190, dust: +180 gas: +190, dust: +180 gas: +190, dust: +180 gas: +190, dust: +180marking

Ex nA IIC T6...T2 GcEx tb IIIC TX Db

Ex nA IIC T6...T2 GcEx tb IIIC TX Db

Ex nA IIC T6...T2 GcEx tb IIIC TX Db

Ex nA IIC T6...T2 GcEx tb IIIC TX Db

certification ATEX IBExU10ATEX1163 X IBExU10ATEX1163 X IBExU10ATEX1163 X IBExU10ATEX1163 Xcertification IECEx IECEx IBE 12.0005X IECEx IBE 12.0005X IECEx IBE 12.0005X IECEx IBE 12.0005Xtype of protection gas: non sparking

dust: protection by enclosure

gas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

II3GII2D

0637 II3GII2D

0637 II3GII2D

0637 II3GII2D

0637

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 17

Technical specification FLUXUS® F608**-A2

technical type CDM2NH1 CDP2NH1 CDQ2NH1order code FSM-NA2NL FSP-NA2NL FSQ-NA2NLtransducer frequency MHz 1 2 4

inner pipe diameter dmin. extended mm 50 25 10min. recommended mm 100 50 25max. recommended mm 1000 400 150max. extended mm 3400 600 400pipe wall thicknessmin. mm 2 1 0.6materialhousing PEEK with stainless steel

cap and transducer shoe 304 (1.4301)

PEEK with stainless steel cap and transducer shoe 304 (1.4301)

PEEK with stainless steel cap and transducer shoe 304 (1.4301)

contact surface PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP66 IP66 IP65

transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 84 84 70width b mm 40 40 30height h mm 59 59 47.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +130 +130 +130temperature compen-sation

x x x

explosion protection

ATEX/IECEx

categoryEPLzone

gas: 3G dust: 2DGc Db2 21

gas: 3G dust: 2DGc Db2 21

gas: 3G dust: 2DGc Db2 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C gas: +190, dust: +180 gas: +190, dust: +180 gas: +190, dust: +180marking

Ex nA IIC T6...T2 GcEx tb IIIC TX Db

Ex nA IIC T6...T2 GcEx tb IIIC TX Db

Ex nA IIC T6...T2 GcEx tb IIIC TX Db

certification ATEX IBExU10ATEX1163 X IBExU10ATEX1163 X IBExU10ATEX1163 Xcertification IECEx IECEx IBE 12.0005X IECEx IBE 12.0005X IECEx IBE 12.0005Xtype of protection gas: non sparking

dust: protection by enclosure

gas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

l

hb

l

hb

l

hb

II3GII2D

0637 II3GII2D

0637 II3GII2D

0637

18 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Shear wave transducers (zone 2, extended temperature range)

technical type CDM2EH5 CDP2EH5 CDQ2EH5order code FSM-EA2NL FSP-EA2NL FSQ-EA2NLtransducer frequency MHz 1 2 4

inner pipe diameter dmin. extended mm 50 25 10min. recommended mm 100 50 25max. recommended mm 1000 400 150max. extended mm 3400 600 400pipe wall thicknessmin. mm 2 1 0.6materialhousing PI with stainless steel cap

and transducer shoe 304 (1.4301)

PI with stainless steel cap and transducer shoe 304 (1.4301)

PI with stainless steel cap and transducer shoe 304 (1.4301)

contact surface PI PI PIdegree of protection according to IEC/EN 60529

IP66 IP66 IP56

transducer cabletype 6111 6111 6111length m 4 4 3dimensionslength l mm 84 84 70width b mm 40 40 30height h mm 59 59 47.5dimensional drawing

ambient temperaturemin. °C -30 -30 -30max. °C +200 +200 +200temperature compen-sation

x x x

explosion protection

ATEX/IECEx

categoryEPLzone

gas: 3G dust: 2DGc Db2 21

gas: 3G dust: 2DGc Db2 21

gas: 3G dust: 2DGc Db2 21

explosion protection temperature (pipe surface)min. °C -45 -45 -45max. °C gas: +235, dust: +225 gas: +235, dust: +225 gas: +235, dust: +225marking

Ex nA IIC T6...T2 GcEx tb IIIA TX Db

Ex nA IIC T6...T2 GcEx tb IIIA TX Db

Ex nA IIC T6...T2 GcEx tb IIIA TX Db

certification ATEX IBExU10ATEX1163 X IBExU10ATEX1163 X IBExU10ATEX1163 Xcertification IECEx IECEx IBE 12.0005X IECEx IBE 12.0005X IECEx IBE 12.0005Xtype of protection gas: non sparking

dust: protection by enclosure

gas: non sparkingdust: protection by enclosure

gas: non sparkingdust: protection by enclosure

l

hb

l

hb

lh

b

II3GII2D

0637 II3GII2D

0637 II3GII2D

0637

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 19

Technical specification FLUXUS® F608**-A2

Transducer mounting fixture

Order code

1, 2 3 4 5 6 7...9 no. of character

tran

sduc

er

mou

ntin

g fix

ture

tran

sduc

er

- mea

sure

men

t ar

rang

emen

t

size

- fixat

ion

oute

r pi

pe

diam

e-

ter

description

VP portable Variofix

A all transducers

D reflection arrangement or diagonal arrangement

R reflection arrangement

M medium

C chains

N without fixation

055 10...550 mm

example

VP A - D M - C 055 portable Variofix and chains

- -

portable Variofix VP and chains

material: stainless steel 304 (1.4301), 301 (1.4310), 303 (1.4305)

dimensions:414 x 94 x 76 mm

chain length: 2 m

20 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Coupling materials for transducers

Technical data

normal temperature range(4th character of transducer order code = N)

extended temperature range(4th character of transducer order code = E)

< 100 °C < 170 °C < 150 °C < 200 °C

coupling compound type N coupling compound type E coupling compound type E coupling compound type E or H

type ambient temperature material°C

coupling compound type N

-30...+130 mineral grease paste

coupling compound type E

-30...+200 silicone paste

coupling compound type H

-30...+250 fluoropolymer paste

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 21

Technical specification FLUXUS® F608**-A2

Connection systems

Transducer cable

Technical data

connection system NL

transducer frequency(3d character of transducer

order code)

F, G, H, K M, P Q S

NL

x y l x y l x y l x y lcable length m 2 3 ≤ 10 2 2 ≤ 10 2 1 ≤ 10 1 1 ≤ 10

cable length (option LC) m 2 7 ≤ 10 7 2 ≤ 10 8 1 ≤ 10 - - -

x, y - transducer cable lengthl - max. length of extension cable

transducer cable extension cabletype 1699 6111 1750standard length m see table above 5

10max. length m - see table aboveambient temperature °C -55...+200 -100...+225 < 80cable jacketmaterial PTFE PFA PEouter diameter mm 2.9 2.7 6thickness mm 0.3 0.5 0.5colour brown white blackshield x x xsheathmaterial stainless steel 304 (1.4301) stainless steel 304 (1.4301) stainless steel 304 (1.4301)outer diameter mm 8 8 9

tra

nsm

itter

l x y

22 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Clamp-on temperature probe (optional)

Technical data

Connection

Temperature probe

technical type PT12N PT12Fdesign short response timetype Pt100 Pt100connection 4-wire 4-wiremeasuring range °C -30...+250 -50...+250

accuracy T ±(0.15 °C + 2 . 10-3 . |T [°C]|)class A

±(0.15 °C + 2 . 10-3 . |T [°C]|)class A

accuracy ∆T(2x Pt matched accor-ding to EN 1434-1)

≤ 0.1 K (3 K < ∆T < 6 K), more corresponding to EN 1434-1

≤ 0.1 K (3 K < ∆T < 6 K), more corresponding to EN 1434-1

response time s 50 8housing aluminum PEEK, stainless steel 304 (1.4301), copperdegree of protection according to IEC/EN 60529

IP66 IP66

weight (without con-nector)

kg 0.25 0.32

fixation clamp-on clamp-onaccessoriesthermal conductivity paste 200 °C

x x

thermal conductivity foil 250 °C

x x

plastic protection plate, insulation foam

- x

dimensionslength l mm 15 14width b mm 15 30height h mm 20 27

PT12N

PT12F

extension cableh

l

h

l

red/blue red white/blue white

TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23 23

Technical specification FLUXUS® F608**-A2

Connector

Cable

pin cable of temperature probe extension cable1 white/blue blue2 red/blue grey3, 4, 5 not connected6 red red7 white white8 not connected

cable of temperature probe extension cabletype 4 x 0.25 mm² black LIYCY 8 x 0.14 mm² greystandard length m 3 5/10/25max. length m - 100cable jacket PTFE PVC

24 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

Wall thickness measurement (optional)The pipe wall thickness is an important pipe parameter which has to be determined exactly for a good mea-surement. However, the pipe wall thickness often is unknown.

The wall thickness probe can be connected to the transmitter instead of the flow transducers and the wallthickness measurement mode is activated automatically.

Acoustic coupling compound is applied to the wall thickness probe which then is placed firmly on the pipe.The wall thickness is displayed and can be stored directly in the transmitter.

Technical data

Cable

technical type DWR1NZ7

measuring range1 mm 1...250resolution mm 0.01accuracy 1 % ± 0.1 mmfluid temperature °C -20...+200,

short-time peak max. 500explosion protection -сabletype 2616length m 1.5

1 The measuring range depends on the attenuation of the ultrasonic signal in thepipe. For strongly attenuating plastics (e.g. PFA, PTFE, PP) the measuring range is smaller.

type 2616ambient temperature °C <200cable jacketmaterial FEPouter diameter mm 5.1colour blackshield x

DWR1NZ7

25 TSFLUXUS_F608xx-A2V2-1-3EN_Leu, 2017-05-23

FLUXUS® F608**-A2 Technical specification

FLEXIM GmbH

Wolfener Str. 36

12681 Berlin

Germany

Tel.: +49 (30) 93 66 76 60

Fax: +49 (30) 93 66 76 80

internet: www.flexim.com

e-mail: [email protected]

Subject to change without notification. Errors excepted.

FLUXUS® is a registered trademark of FLEXIM GmbH.