Report on SiPM Tests
•SiPM as a alternative photo detector to replace PMT.•Qauntify basic characteristics•Measure Energy, Timing resolution•Develop simulation code to confirm measurement.
Test Setup
Two SiPM sitting on top of tables.Hamamatsu SiPM.1x1mm2 area( fill factor : 30%).1600 pixels.
Bias Voltage : 71VNa22 for gamma sourceLYSO : 1x1x10mm3Lecroy 612A amplifier( 10X gain)
Gain vs Bias V( 25um pixel SiPM)
• Consistent with using the fast scope.• Factor ~2 increases in 70.1 and 71.9V.• Similar tendency as in the Hamamatsu Spec.(right figure)
Dark count rate(25um pixel SiPM)
Noise due thermal excitation.
Two discriminators:Threshold : 0.5 and 1.5 p.e.
Vary Bias 70.1 ~ 71.9V.
Read-out count rate.Lecroy 2551 100Mhz scaler.
Single p.e. rate is dominant.
Plans
•Measurement with different type of SiPM and crystal.•Connect signal to Multi-threshold board. •Refine setup to improve results.•Validate simulation code by experimental data.
Geant4 Simulation
•To confirm experimental results.•Data for validation. SiPM Test R9800 PMT+ LSO Crystal•Basis for future detector design study.
Simulation Setup
Pair of LSO + R9800PMTLSO( 6.25x6.25x25mm3)Wrapped with teflon tape.5cm distance between modules.
Borosilicate GlassMetal photocaothode inside Glass
Geant4 with Optical Physics Processes.(Light refraction, reflection, absorption…)Light Yield : 30,000/MeV (LSO)
# of photons detected
Experimental data using Fast oscilloscopeIntegrated charge under pulse shape.
Simulation results.# of photon detected at photocathode.PMT Q.E included.
Timing simulation
Top left: Photo arrival time at photocathode from generation.Top right : Convolution with gaussian for each photon.Bottoms : Rising edges for two modules A and B( arbitrary unit)
Timing simulation
T0 was extracted from linear fit.Time difference between two T0.
~250 ps in FWHM.c.f) ~300ps with real data