quadtrap comparison 1
TRANSCRIPT
-
8/3/2019 QuadTrap Comparison 1
1/105
-
8/3/2019 QuadTrap Comparison 1
2/105
2
Comparison of Quads and Traps
Ion Traps Quadrupoles
Mass Separation in time Mass Separation in space
High sensitivity Full Scan Lower sensitivity Full Scan
Lower sensitivity SIM and SRM High sensitivity SIM and SRM
Offer multiple stages of MSn
Parent and neutral loss scans
Offers Only MS or MS/MS
-
8/3/2019 QuadTrap Comparison 1
3/105
3
LCQ Instrumentation
ClassicClassic
DecaDecaDuoDuo
-
8/3/2019 QuadTrap Comparison 1
4/105
4
Duo/Deca Comparisons
LCQ DUO LCQ DECA 500um capillary
1 square quadrupole & 1 octopole2 rotary pumps
400um capillary
2 octopolesOne rotary pump
Deca: approximately 10 x better signal than Duo
-
8/3/2019 QuadTrap Comparison 1
5/105
5
Current LCQ Generation
-
8/3/2019 QuadTrap Comparison 1
6/105
6
Advantage/XP Comparisons
450um Ion Transfer Tube Orthogonal Probes
550um Ion Transfer Tube Orthogonal Probes
XP: approximately 10 x better signal than Advantage
-
8/3/2019 QuadTrap Comparison 1
7/105
-
8/3/2019 QuadTrap Comparison 1
8/105
8
Mass Spectrometry Simplified
G
MS
D
enerate
oveelect
etect
Ion productionIon production
Ion opticsIon optics
Mass filterMass filter
Electron multiplierElectron multiplier
-
8/3/2019 QuadTrap Comparison 1
9/105
9
The Mass Spectrum
Red
Prism
Light,all colors Green
Blue
100
Ions,
various massesMass
Spectrometer200
300
T d d b i ll b d
-
8/3/2019 QuadTrap Comparison 1
10/105
10
Trace produced by summing all observedTrace produced by summing all observed
masses in each scanmasses in each scan
TotalTotal Ion Chromatogram or TICIon Chromatogram or TIC
-
8/3/2019 QuadTrap Comparison 1
11/105
11
Ionization vs. Fragmentation
Ionization HardSoft
No
Fragments
Fragments
API EICI
-
8/3/2019 QuadTrap Comparison 1
12/105
12
EI and CI Mass Spectra of Ephedrine
0
50
100
0 20 40 60 80 100 120 140 160 180
%
RelativeIntensity
m/z
EI
0
50
100
0 20 40 60 80 100 120 140 160 180
CI
58
148
166
MW = 165Th
(M+H)+
-
8/3/2019 QuadTrap Comparison 1
13/105
13
Ionization Techniques
ESI
GCPBI
TSPFAB
Molecular
W
eight
APCI
200,000
15,000
1,000
Non Polar Polar
-
8/3/2019 QuadTrap Comparison 1
14/105
14
What is API?
Atmospheric Pressure Ionization
Source Types:
1. Electrospray Ionisation (ES) Solution phase process (forthe most part).
2. APCI (Atmospheric Pressure Chemical Ionization) - Gas-phase process.
Source Purpose:1. Ionize the analyte (APCI) or transport ion in solution to the
gas phase.
2. Desolvate sample flow for introduction into mass
spectrometer.3. Baffle the first vacuum region of the MS from atmospheric
pressure in the source.
4. Pump away neutrals and opposite charged ions which would
otherwise interfere with the analysis of the desired polarity.
-
8/3/2019 QuadTrap Comparison 1
15/105
15
LCQ Classic/Duo/Deca API Probes
Electrospray Ionization (ESI)
Peek insulator
Atmospheric Pressure
Chemical Interface (APCI)
-
8/3/2019 QuadTrap Comparison 1
16/105
16
LCQ Advantage/XP API Probes
Atmospheric Pressure
Chemical Interface (APCI)Electrospray Ionization (ESI)
Orthogonal ESI & APCI probes
-
8/3/2019 QuadTrap Comparison 1
17/105
17
Chemistry Considerations
ESI:
Ions formed by solution chemistry
Good for Thermally labile analytes
Good for Polar analytes
Good for Large Molecules (Proteins / Peptides)
APCI:
Ions formed by gas phase chemistry
Good for Volatile / Thermally Stable
Good for Non-polar analytes
Good for Small Molecules (Steroids)
-
8/3/2019 QuadTrap Comparison 1
18/105
18
ESI Versatility Advantage/XP
(Right: Picture represents
a low flow position,
-
8/3/2019 QuadTrap Comparison 1
19/105
19
ElectrosprayBasic Layout
Heated Capillary
ESI Needle+/- 5 kV
Taylor Cone
Solvent evaporation and ion release
++
+++ ++
+
++
+++ +
++
+
+
+
+++
++
+
+
+
++
+++ +
++
++
+++ +
++
++
++
++
+
+
+
+
+
+
++
++
++
+
+
+
++
+
+ ++ +
+++
++
+
++
+
+++
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
+
-
8/3/2019 QuadTrap Comparison 1
20/105
20
APCI Position on Advantage/XP
In/Out MovableCorona Discharge Pin
Total control for any
flow rate (200ul/min -
2000ul/min)
-
8/3/2019 QuadTrap Comparison 1
21/105
21
LC Flow Rates
ESI:ESI:3 L/min - 1mL/minute
Optimal Flow Rate: 200 L/minGenerally, higher flow rates require higher heatedcapillary temperatures and higher gas flow rates.
APCI:APCI:200 L/min - 2mL/minute.Optimal Flow Rate: 500 L/minGenerally, higher flow rates require more sheath andauxiliary gas, but do not require higher heatedcapillary temperatures.
-
8/3/2019 QuadTrap Comparison 1
22/105
22
Theoretical Increase in Response
DConc max =2
Column 1D
2Column 2
Col. Diameter mm 1.0
0.05
21
0.5
2.3
3.04.6
0.2
5
2.0 Capillary
Flow Rate ml/min 1 < 10 l / min1Theoretical
Increase
LC Additi
-
8/3/2019 QuadTrap Comparison 1
23/105
23
LC Additives
Acids
Do not use inorganic acids (may cause source corrosion)
Formic and acetic acid are recommended
Bases
Do not use alkali metal bases (may cause source corrosion)
Ammonium hydroxide is recommended
Surfactants (surface active agents)
Detergents and other surface active agents may suppressionization
Trifluoroacetic Acid (TFA)
May enhance chromatographic resolution, but causes ion
suppression in both negative and positive ion mode
Isopropyl Alcohol
May Enhance Negative Ion Formation
B ff ( H)
-
8/3/2019 QuadTrap Comparison 1
24/105
24
Buffers (pH)
Avoid using non-volatile HPLC additives such as:
Alkali Metal Phosphates
Borates
Citrates
Keep Buffer concentrations below 20 mM using volatilesalts such as ammonium acetate.
When using buffers, more frequent cleaning of the heatedcapillary and API stack will be necessary
LC/MS Additi d B ff S
-
8/3/2019 QuadTrap Comparison 1
25/105
25
LC/MS Additives and Buffers Summary
Acetic Acid
Formic Acid
Ammonium Hydroxide
Ammonia Solutions
Trichloroacetic Acid (< 0.1% v/v)
Trifluoroacetic Acid (< 0.1% v/v)
Isopropyl Alcohol
(10% of organic phase)
Ammonium Acetate
Ammonium Formate
Proton Donors
Proton Acceptors
Chromatographic
Separation
Negative ion
formation
Buffers
C LC/MS S l t
-
8/3/2019 QuadTrap Comparison 1
26/105
26
Common LC/MS Solvents
Methanol
AcetonitrileWater
IsopropanolDichloromethane
Chloroform
Hexane
Effects of Solvents and Additives on ESI
-
8/3/2019 QuadTrap Comparison 1
27/105
27
Effects of Solvents and Additives on ESI
50/50 MeOH/H2O
50/50 ACN/H2O
100 H2O
100 MeOH
100 ACN
50/50 MeOH/H2O 1% Acetic
50/50 MeOH/H2O 0.1% Formic
50/50 ACN/H2O 1% Acetic
50/50 ACN/H2O 0.1% Formic
50/50 MeOH/H2O 5mM NH4OAc
50/50 MeOH/H2O 10mM NH4OAc
50/50 MeOH/H2O 0.1% TFA
50/50 MeOH/H2O 0.05% TFA
50/50 MeOH/H2O 0.02% TFA50/50 ACN/H2O 0.1% TFA
50/50 ACN/H2O 0.05% TFA
50/50 ACN/H2O 0.02% TFA
50/50 MeOH/H2O 0.1% NH4OH
50/50 ACN/H2O 0.1% NH4OH
Solve
ntSystem
Counts (protonated ion species)
0 100000 200000 300000 400000 500000
Tyr-Gly-Gly-Phe-Leu
Leucine Enkephalin
API Stack
-
8/3/2019 QuadTrap Comparison 1
28/105
28
API Stack
LCQClassic, LCQDUO,
LCQDECALCQ DECAXP,LCQAdvantage
Ion Transfer Tube and Removal Tool
-
8/3/2019 QuadTrap Comparison 1
29/105
29
Ion Transfer Tube and Removal Tool
Removal
tool
Ion transfer tube
Heated capillary
Vent Prevent Mechanism
-
8/3/2019 QuadTrap Comparison 1
30/105
30
Vent Prevent Mechanism
Heated Tube
in-situ
Heated Tube
removed
Tungsten Vent
Prevent
Ion Optics
-
8/3/2019 QuadTrap Comparison 1
31/105
31
Ion Optics
First
multipole
Lens
Intermultipole
Lens
Second
multipoleLens
Octapole
Mount
Vacuum
Baffle
Analyzer
Mount
IONS
IN
IONS
OUT
Multipole Potential Wells
-
8/3/2019 QuadTrap Comparison 1
32/105
32
Multipole Potential Wells
Octapole
Mass RangeTransEfficiency
Round
Quadrupole
Square
Quadrupole
Mass Analyzer (Ion Trap)
-
8/3/2019 QuadTrap Comparison 1
33/105
33
Mass Analyzer (Ion Trap)
Vacuum System
-
8/3/2019 QuadTrap Comparison 1
34/105
34
Vacuum System
Every mass analyzer must operate under vacuum inEvery mass analyzer must operate under vacuum inorder to minimize both ion/molecule andorder to minimize both ion/molecule and
molecule/ molecule collisions.molecule/ molecule collisions.
At atmospheric pressure, the mean free path of a typicalAt atmospheric pressure, the mean free path of a typicalion is only ca. 52 nm and at 1ion is only ca. 52 nm and at 1 mTorrmTorr, it is 40 m., it is 40 m.
Without vacuum, the ions produced in the source wontWithout vacuum, the ions produced in the source wontmake it to the detector.make it to the detector.
The LCQ vacuum is maintained by a both rotary andThe LCQ vacuum is maintained by a both rotary andturbomolecularturbomolecularpumpspumps
Ion Optics (Operating Pressures)
-
8/3/2019 QuadTrap Comparison 1
35/105
35
Ion Optics (Operating Pressures)
1.3 torr760 torr 1.7x10-3 torr 2.0 x10-5 torr (1.0x10-5 torr He)
3.5x10-3 torr He
220 L/sec60 m
3
/hr 100 L/sec
Steps to Ion Trap Scan Functions
-
8/3/2019 QuadTrap Comparison 1
36/105
36
Steps to Ion Trap Scan Functions
Trapping- all scans
Isolation- SIM and MSn
Excitation- MSn
Ejection- all scans
Helium as a Damping Gas
-
8/3/2019 QuadTrap Comparison 1
37/105
37
Helium as a Damping Gas
+
+
+
+Without HeliumWithout Helium
He
collision HeHe
He
He
+
+
+
+
With HeliumWith Helium
Discovery of the Effects of Helium
-
8/3/2019 QuadTrap Comparison 1
38/105
38
y
Helium as a Damping Gas
-
8/3/2019 QuadTrap Comparison 1
39/105
39
p g
Ion Trap ResolutionEffect of Damping Gas
-
8/3/2019 QuadTrap Comparison 1
40/105
40
p p g
Traps injected ions by removing kinetic energy
Damps ion motion to center of trap
Result...Increase in resolution and sensitivity
Helium Effect
-
8/3/2019 QuadTrap Comparison 1
41/105
41
S#:1 RT:0.00 AV:1 SM:7G NL:2.50E7T:+ p Full ms
514 516 518 520 522 524 526 528
m/z
0
20
40
60
80
100
RelativeAb
undance
524.3
525.3
S#:23-32 RT:0.71-1.00AV:10 SM:7G NL:5.61E7T:+ p Full ms
500 1000 1500 2000
m/z
0
20
40
60
80
100
RelativeAb
undance
1522.04
1621.971322.06
1721.89
1222.141821.95524.26
1122.21 1921.88195.15
1022.09
S#:1 RT:0.02 AV:1 SM:7G NL:9.70E6T:+ p Full ms
514 516 518 520 522 524 526 528m/z
0
20
40
60
80
100
RelativeAbun
dance
522.6523.0
521.8
521.2 523.9
520.7
S#:23-32 RT:0.39-0.54AV:10 SM:7G NL:2.80E7T:+ p Full ms
500 1000 1500 2000m/z
0
20
40
60
80
100
RelativeAbun
dance
1620.791520.26
1720.441320.95
1220.75523.01 1919.96
1120.90
192.17
Helium shut off and not flowing into trapHelium shut off and not flowing into trap
S#:1 RT:0.00 AV:1 SM:7G NL:2.50E7T:+ p Full ms
514 516 518 520 522 524 526 528
m/z
0
20
40
60
80
100
RelativeAb
undance
524.3
525.3
S#:23-32 RT:0.71-1.00AV:10 SM:7G NL:5.61E7T:+ p Full ms
500 1000 1500 2000
m/z
0
20
40
60
80
100
RelativeAb
undance
1522.04
1621.971322.06
1721.89
1222.141821.95524.26
1122.21 1921.88195.15
1022.09
S#:1 RT:0.02 AV:1 SM:7G NL:9.70E6T:+ p Full ms
514 516 518 520 522 524 526 528m/z
0
20
40
60
80
100
RelativeAbun
dance
522.6523.0
521.8
521.2 523.9
520.7
S#:23-32 RT:0.39-0.54AV:10 SM:7G NL:2.80E7T:+ p Full ms
500 1000 1500 2000m/z
0
20
40
60
80
100
RelativeAbun
dance
1620.791520.26
1720.441320.95
1220.75523.01 1919.96
1120.90
192.17
Helium shut off and not flowing into trapHelium shut off and not flowing into trap
Helium flowing into trap
Ion Trap Stability Diagram
-
8/3/2019 QuadTrap Comparison 1
42/105
42
y
The region shaded blue indi-cates a (DC) and q (RF)
values which provide stable
trajectories in the r-directionThe region shaded yellow
indicates the z-stable a and q
combinations
The green area where the r-
and z-stable regions overlap
indicates the a and q combi-nations under which ions will
be stable in the trap
Stability Diagram for Commercial Traps
-
8/3/2019 QuadTrap Comparison 1
43/105
43
q kV
m ez =
( / )
LCQ Scan-Out (Ejection) Rates
-
8/3/2019 QuadTrap Comparison 1
44/105
44
Normal Scan (5500 amu/sec)
Common full, SIM, or MSn (SRM and CRM)scanning
Resolution (FWHM) = 0.50, Mass Accuracy = 0.05
Zoom Scan (280 amu/sec)
Increases resolution and mass accuracy across anarrow range (allows charge state determination)
Resolution (FWHM) = 0.15, Mass Accuracy = 0.02
Turbo Scan (55,000 amu/sec)
Decreases total scan time of a full scan, thus
increasing number of scans across achromatographic peak
Resolution (FWHM) = 3.0, Mass Accuracy = 0.5Used for better quantitation due to an increase
of scans across a chromatographic peak
What is AGC and Why Is it Important?
-
8/3/2019 QuadTrap Comparison 1
45/105
45
Camera AE
Too much light degrades the image
stored on film, causing a loss of
color and image resolution.
Too little light results in dark
picture with no fine details visible.
Cameras with high quality light
meters and AE controls produce
high quality pictures over a wide
dynamic range of lighting
conditions.
LCQ Series AGC
Controls amount of ions (light)
entering the ion trap (film)
Too many ions degrade the
spectral quality in the trap,
causing loss in mass resolution
and mass assignment. Too fewions result in poor sensitivity to
low level or minor components.
AGC ensures excellent quality
MS, SIM and MS/MS spectra,
as well as excellent sensitivity
over a wide dynamic range.
Automatic Gain Control (AGC)
-
8/3/2019 QuadTrap Comparison 1
46/105
46
Prescan before the analytical scan
- Measures the # of ions in the trap for a
pre-defined time (10 ms)
-Allows software to determine optimum
ion injection time
No AGC Spectrum of Ultramark 1621, Caffeine,MRFA Calibration Mixture space charging
-
8/3/2019 QuadTrap Comparison 1
47/105
47
MRFA Calibration Mixture space charging
Spectrum of Ultramark 1621, Caffeine, MRFACalibration Mixture with AGC
-
8/3/2019 QuadTrap Comparison 1
48/105
48
Calibration Mixture with AGC
AGC (Ion Population Control)
-
8/3/2019 QuadTrap Comparison 1
49/105
49
~ 300 Ions ~ 1500 Ions ~ 3000 Ions ~ 6000 Ions
530m/z
0
20
40
60
80
100
Relative
Abundance
524.3
525.3
526.3
530m/z
0
20
40
60
80
100
Relative
Abundance
524.4
525.4
526.3
527.5
530m/z
0
20
40
60
80
100
Relative
Abundance
524.5
525.5
526.5
527.5
530m/z
0
20
40
60
80
100
Relative
Abundance
524.8
525.7
526.7
522 522 522 522
Poor ResolutionGood Resolution
Calculation of Ion Time
-
8/3/2019 QuadTrap Comparison 1
50/105
50
Constant During Prescan
AGC Prescan Signal =
Multiplier Gain x Prescan TimeNumber of Ions x
(3x105 counts) (10 ms)
Calculated Ion Time = Target Value
AGC Prescan Signal(how long the gate lens is open)
Triplicate Injection of 5 nmol of MRFA (AGC ON)
-
8/3/2019 QuadTrap Comparison 1
51/105
51
Scan Number
UnscaledTIC(counts
)
0.0E+00
2.0E+074.0E+07
6.0E+07
8.0E+071.0E+08
1.2E+08
1.4E+08
1.6E+081.8E+08
2.0E+08
0 100 200 300 400 500 600
Unscaled TIC
0
10
20
30
40
50
60
0 100 200 300 400 500 600Scan Number
InjectionTime(ms)
Injection Time
Scaled TIC
0.0E+00
1.0E+08
2.0E+08
3.0E+08
4.0E+08
5.0E+08
ScaledTIC(counts)
8.0E+08
7.0E+08
6.0E+08
0 100 200 400 500 600300
Scan Number
Isolation of Ions
-
8/3/2019 QuadTrap Comparison 1
52/105
52
Ion we wish to isolateIon we wish to isolate
7
0.908qz
0.0
Ions at different qz values oscillate at
different frequencies (
o)
22
qzo
Isolation Waveforms
-
8/3/2019 QuadTrap Comparison 1
53/105
53
~ m/z 200
q axis .908
500 Hz
16 msec
q axis.908
Why MS/MS or MSn ?
-
8/3/2019 QuadTrap Comparison 1
54/105
54
Signal to Noise Improvement
0 1 2 3 4 5 6
Stages of Analysis
Intensity Signal
Noise
S/N
MS/MS Parameters
-
8/3/2019 QuadTrap Comparison 1
55/105
55
Precursor ion m/z
Excitation voltage
Excitation qz
1.0 2.0 3.0
Excitation Voltage (V)
Intensity
Precursor ion
(Product ions)
0.0 4.0
77001-1285
970219
Resonant Excitation qz Value
-
8/3/2019 QuadTrap Comparison 1
56/105
56
Fragmentation
Energy
0.908
qz
0.00.2250.225
0.908
qz
0.0
0.908
qz
0.0
0.300.30
0.450.45
Fragment ions
not trapped
Product Ion
m/z Range
1/41/4
1/31/3
1/21/2
Ion Trap Scan Functions
-
8/3/2019 QuadTrap Comparison 1
57/105
57
1. Collect
2. Isolate
3. Fragment
4. Eject
Summary (Ion Trap Functions)
-
8/3/2019 QuadTrap Comparison 1
58/105
58
1)1)CollectionCollection
2)2)IsolationIsolation
3)3)ExcitationExcitation
4)4)Ejection
For Scans: All
By: Ring Electrode
Method: Alternating RF
frequency (760 kHz) at a set
amplitude along with Hedampening gas traps and
cools the ions to the center of
the trap.Ejection
Summary (Ion Trap Functions)
-
8/3/2019 QuadTrap Comparison 1
59/105
59
1)1)TrappingTrapping
2)2)IsolationIsolation
3)3)ExcitationExcitation
4)4)Ejection
For Scans: SIM, MSn
By: Endcap Electrodes
Method: a) Tailored waveform
applied to all ions in the trap
except ion of interest
b) Thus, only ions of interest
remain in the trap.
Ejection
Summary (Ion Trap Functions)
-
8/3/2019 QuadTrap Comparison 1
60/105
60
1)1)TrappingTrapping
2)2)IsolationIsolation
3)3)ExcitationExcitation
4)4)Ejection
For Scans: MSn
By: Endcap Electrodes
Method: a) Cool ion of
interest back to set q value
(default = 0.25). b) Apply
custom RF waveform inresonance with the set q
value, activation time
(default = 30 msec), andoptimized activation
amplitude.
Ejection
Summary (Ion Trap Functions)
-
8/3/2019 QuadTrap Comparison 1
61/105
61
1)1)TrappingTrapping
2)2)IsolationIsolation
3)3)ExcitationExcitation
4)4)Ejection
For Scans: All
By: Ring Electrode
Method: Ramp ring RFpower to increase the q values
of all ions in desired scan
range, low mass to high mass.(i.e. Mass Selective Scanning)
Also, ramp the RF amplitude
on the endcap electrodes toconsolidate the ions to a
group (Resonance Ejection)
Ejection
Tune Page
convection gauge
-
8/3/2019 QuadTrap Comparison 1
62/105
62
Ion gauge
Pressure
-
8/3/2019 QuadTrap Comparison 1
63/105
63
Caffeine stock: 1 mg/ml in methanol
MFRA stock: Dissolve 3.0 mg MRFA in 1 ml 50:50 methanol:water
Ultramark stock: Measure 10 l of Ultramark 1621, and dissolve it in10 ml acetonitrile
ESI calibration solutionESI calibration solution
Into a clean vial pipette 100 l of caffeine stock, 5 l of MRFA stock and2.5 ml of Ultramark stock. Add 50 l of glacial acetic acid and 2.34 ml50:50 methanol:water
Ion Trap Animation
-
8/3/2019 QuadTrap Comparison 1
64/105
64
The Eighth Generation Triple Quad
-
8/3/2019 QuadTrap Comparison 1
65/105
65
TSQ 15,
TSQ 45,
TSQ 46,TSQ 70,
TSQ 700,TSQ 7000,
TSQ,TSQ Quantum
Smaller because
-
8/3/2019 QuadTrap Comparison 1
66/105
66
25 cm quad
25 cm quad
25 cm quad
25 cm quad
8degrees
90 degrees
TSQ 7000
Quantum
90 Degree
Square
quad
collisioncell
HyperQuadsTMHyperbolic Quadrupoles
-
8/3/2019 QuadTrap Comparison 1
67/105
67
Forms Pure Quadrupolar Fields
Reduces Fringing Effects
Significantly Improves Resolution Improves Transmission
Improves Peak Shapes
TSQ 7000
1993 to 2000
r0 = 4 mmL = 250 mm
TSQ Quantum
2001 to
r0 = 6 mm
L = 250 mm
Quadrupole Mass Analyzer
-
8/3/2019 QuadTrap Comparison 1
68/105
68
+
+
++
The ion is transmitted along the quadrupole in a
stable trajectory Rf field. The ion does not have astable trajectory and is ejected from the quadrupole.
How does the Quadrupole work ?
-
8/3/2019 QuadTrap Comparison 1
69/105
69
The quadrupole consists of four parallel rods. The opposing rodshave the same polarity while adjacent rods have opposite polarity.
-ve
+ve
Each rod is applied with a DC and
an RF voltage. Ions are scanned byvarying the DC/Rf quadrupole
voltages.
Only ions with the selected massto charge ratio will have the
correct oscillatory pathway in the
Rf field.
Effect of Peak Width On Transmission
-
8/3/2019 QuadTrap Comparison 1
70/105
70
0
20
40
60
80
100
%
T
r
a
n
s
m
i
s
s
i
o
n
2 1.5 0.7 0.5 0.2 0.1
Peak Width FWHM
HYPERQUAD ROUND RODS
Effect of Peak Width on Resolution
-
8/3/2019 QuadTrap Comparison 1
71/105
71
0
2000
4000
6000
8000
10000
12000
0 200 400 600 800 1000 1200
m/z
R
esolution
Quad 0.7 FWHM
Quad 0.1 FWHM
Sector
Quantum, API 4000, Ultima
at 0.7 FWHM at m/z 1000
R = 1428
R is relatively flat across m/z
Quantum operating at
0.1 FWHM at m/z 1000
R = 10,000
The Power of Resolution
-
8/3/2019 QuadTrap Comparison 1
72/105
72
Separation of ions with samenominal m/z value
Unequivocal determination ofcharge state (ESI)
High resolution precursor ionselection for MS/MS
High resolution product ion for
charge state determination
Effect of changing resolution on peak shape
-
8/3/2019 QuadTrap Comparison 1
73/105
73
1.0 FWHM 0.7 FWHM 0.5 FWHM
0.3 FWHM0.2 FWHM 0.1 FWHM
Resolution vs. Intensity
-
8/3/2019 QuadTrap Comparison 1
74/105
74
1.0 FWHM 0.7 FWHM 0.5 FWHM
2.5e6 2.1e61.9e6
Resolution vs. Intensity
-
8/3/2019 QuadTrap Comparison 1
75/105
75
0.3 FWHM 0.2 FWHM 0.1 FWHM
0.8 e61.5e6 1.4e6
Quadrupole Mass Analyzer
-
8/3/2019 QuadTrap Comparison 1
76/105
76
If one MS scan between m/z 100 and 500is completed in one second, then eachm/z will be allowed to pass for 2.5 ms.
1000 ms== 2.5 ms/amu
400 amu
+
++
+
+
To
detector
+
Quadrupole Mass Analyzer
-
8/3/2019 QuadTrap Comparison 1
77/105
77
And, for the same peak, for example, thequadrupole performs 5 complete scansfrom 100 500 Da each taking 1 sec.
5 sec
100 500 Da
100 500 Da
100 500 Da
100 500 Da 100 500 Da
Ion Trap Pre-Scan
-
8/3/2019 QuadTrap Comparison 1
78/105
78
The length of time that the trap staysopen to collect ions is determined by a
pre-scan which measures total ion current(prevents space charging, so no ghostpeaks)
5 sec
Pre-Scan
Ion Trap Pre-Scan Contd
-
8/3/2019 QuadTrap Comparison 1
79/105
79 5 sec
Across a peak 5 sec. wide the trap mightfill and empty 5 times. So, a group of ionsare collected ca. every 1 sec each group
is then ejected to the detector, smallerions first. 1 sec
1 sec
1 sec
1 sec
1 sec
Comparison of Quads and Traps
-
8/3/2019 QuadTrap Comparison 1
80/105
80
But for a trap, each m/z (andall m/z at the same time) is/are
collected for ca. 750 ms (taking
pre-scan and interscan timesinto account) and then scanned
to the detector.
For the quadrupole,
each m/z is scanned
(sequentially)
to the detector for 2.5
ms.
+
+
++
Comparison of Quads and Traps
-
8/3/2019 QuadTrap Comparison 1
81/105
81
A trap will fill similar tofilling a glass with water. All
ions entering the trap will be
collected until the trap fillswith a pre-defined amount of
ions (AGC target value).
At any one particularinstant, a quad will only
scan/look for only one
m/z. All other m/z willbe ignored. In this
example, each m/z is
scanned for 2.5 ms.
+
+
++
Comparison of Quads and Traps
-
8/3/2019 QuadTrap Comparison 1
82/105
82
So, for full scan MS, a trap will givebetter sensitivity because there aremore ions representing each m/zarriving at the detector for each scan.
+
++
+
How Much More???
-
8/3/2019 QuadTrap Comparison 1
83/105
83
Accounting for pre-scan/interscantiming, the trap produces ca. 300 times(750 ms/2.5 ms) for the collection ofeach m/z compared to a quadrupole (i.e
2 orders of magnitude).
+
+++
But
-
8/3/2019 QuadTrap Comparison 1
84/105
84
What if I wanted to pass (filter) only onem/z ion to the detector (i.e. SIM or SRM) then I could spend more time on that ion
+
+ ++
Comparison of Quads and Traps
-
8/3/2019 QuadTrap Comparison 1
85/105
85
Yes, on a quadrupole,
interscan times are
relatively short and sothe quadrupole remains
fixed on that one ion a
duty cycle of close to100%
But a trap will still only
collect ions in batches-
and prescan/interscantimes afford a duty cycle of
about 75%
+
+++
What is a Duty Cycle?
-
8/3/2019 QuadTrap Comparison 1
86/105
86
Scan 1 ISD Scan 2
Definition:-
Time taken to acquire 1 scan and be ready to acquire
the next one
Interscan delay (ISD) is the time taken to return all
system voltages to the start values and reach a stable
stateThis is dead time and should be minimized
Duty Cycle
In Addition
Whil h b i i i l
-
8/3/2019 QuadTrap Comparison 1
87/105
87
While the beam instrument is continuouslydetecting one particular m/z a trap builds acurve from an average over each collection
time and the points are least frequent atthe most important region for quantitation
(the take off).
For Example
SRM f 5 Al l ith LCQ D
-
8/3/2019 QuadTrap Comparison 1
88/105
88
SRM of 5 pg Alprazolam with LCQ Decagives a %RSD of 6.11 while
SRM of 750 fg Alprazolam with TSQ 7000gives a %RSD of 1.87.
Signal to noise ratio (S/N) are similar
(20:1 and 28:1 respectively).
100
RT: 4.26
SN: 20
%RSD 6 11%RSD 6 11 SRM 5SRM 5 Al lAl l
90
80
an
70ce
-
8/3/2019 QuadTrap Comparison 1
89/105
89
0
10
20
30
40
50
Relative
%RSD 6.11%RSD 6.11withwithLCQLCQDecaDeca
SRM 5pgSRM 5pg AlprazolamAlprazolam60A
bundan
RT: 4.41
SRM 750fgSRM 750fg AlprazolamAlprazolam with TSQwith TSQ
SN: 28
%RSD 1.87%RSD 1.87
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5Time (min)
0
10
20
30
40
50
60
70
80
90
100
The Effect of Ion Trap Scan Speeds onQuantitative Performance
-
8/3/2019 QuadTrap Comparison 1
90/105
90
In general, faster scanning produces more points
across a chromatographic peak, hence better
precision and lower LOQs.
In fact, for an ion-trap, scan speed refers only to
the time taken to scan ions from the trap duringmass analysis.
Scan speed does not refer to the total analyticalcycle. In an ion trap device, an MSn analytical
scan comprises at least four events:
The Effect of Ion Trap Scan Speeds onQuantitative Performance
-
8/3/2019 QuadTrap Comparison 1
91/105
91
1- AGC Pre-scan
2- Ion Injection (usually the rate-determining step)
3- Isolation and activation of the parent ion within
the trap
4- Scanning the ions out of the trap (mass analysis)
MS Scan FunctionMass
Analysis
Ion
-
8/3/2019 QuadTrap Comparison 1
92/105
92
IonActivation
Ion
isolation
Ion
Injection
Analytical ScanAGC Prescan
Scan TerminologyScan Terminology
-
8/3/2019 QuadTrap Comparison 1
93/105
93
Prescan Mass Analysis Prescan Mass Analysis Prescan Mass Analysis
1st Microscan 2nd Microscan 3rd Microscan Save Data
Complete Scan Cycle
The Effect of Ion Trap Scan Speeds onQuantitative Performance
-
8/3/2019 QuadTrap Comparison 1
94/105
94
The injection time constitutes the majority of the
total scan time.
For good quantitative reproducibility, it is
necessary to take enough points to precisely
determine the chromatographic peak particularly
at the take-off point.
Increasing the scan speed in the trap does not
significantly increase the number of data pointstaken across the peak.
Quantitation m/z 500 scanning 135 510
Pre-scan 0 msec
Isol /Activ / Download 80
Pre-scan 0 msec
Isol /Activ / Download 80
Scan speeds in ion traps
-
8/3/2019 QuadTrap Comparison 1
95/105
95
Pre-scan 60 msec
Isol/Activ/ Download 80
375 amu @ 5500 amu/sec 70
Injection time 500
Total scan time 710
Scans / 10 sec wide peak 14
Pre-scan 60 msec
Isol/Activ/ Download 80
375 amu @ 5500 amu/sec 70
Injection time 500
Total scan time 710
Scans / 10 sec wide peak 14
Quantitation m/z 500 scanning 135-510Pre-scan 60 msec
Isol /Activ / Download 80375 amu @ 13,000 amu/sec 30
Injection time 500
Total scan time 670
Scans / 10 sec wide peak 15
Pre-scan 60 msec
Isol /Activ / Download 80
375 amu @ 13,000 amu/sec 30
Injection time 500
Total scan time 670
Scans / 10 sec wide peak 15
Isol /Activ / Download 80
375 amu @ 13,000 amu/sec 30
Injection time 500
Total scan time 610
Scans / 10 sec wide peak 16
Isol /Activ / Download 80
375 amu @ 13,000 amu/sec 30
Injection time 500Total scan time 610
Scans / 10 sec wide peak 16
The Effect of Ion Trap Scan Speeds onQuantitative Performance
-
8/3/2019 QuadTrap Comparison 1
96/105
96
The only significant way to increase the sampling
rate across the peak is to reduce the injectiontime which can be achieved in two ways:
1- Set a lower max injection time which reduces the
number of ions in the trap, hence sensitivity
2- Increase the efficiency of the source and lenses
to improve the transmission of ions; I.e filling thetrap to the same level in a shorter period of time.
Data Dependant Acquisition of MSn Spectra
Data Dependant Acquisition: Intelligent decision-making software that
selects precursor ion for MSn experiments based on user define criteria
-
8/3/2019 QuadTrap Comparison 1
97/105
97
selects precursor ion for MSn experiments based on user-define criteria.
Critical for metabolite screening experiments.
Scan event 1
Full-Scan MSScan event 2
Full-Scan MS2
Software selects most intenseion from scan event 1 as
precursor ion for ms2
experiment in scan event 2,
provide that its intensity is
above a user selected threshold
m/zm/z
Dynamic Exclusion-- MS and MS/MS of Co-Eluters
MS
MS/MS
-
8/3/2019 QuadTrap Comparison 1
98/105
98
407
407MS
Time
452
452
365
206
377255
MS/MSm/z
377
231
Threshold
MS MS/MS
MS/MS
MS
MS/MS
MS
MS MS
MS
m/z
m/z
m/z
Comparison of Quads and Traps
Major Strengths of Triple Quads
-
8/3/2019 QuadTrap Comparison 1
99/105
99
Major Strengths of Triple Quads
SRM Sensitivity
Neutral Loss Scan Mode Parent (Precursor) Scan Mode
Major Strengths of the LCQ Deca XP Plus
MSn Scan Mode
Full Scan MS/MS Sensitivity
Consecutive Reaction Monitoring (CRM)
What are neutral loss scans ?
Both Q1 and Q3 are scanned together
-
8/3/2019 QuadTrap Comparison 1
100/105
100
Both Q1 and Q3 are scanned together
Q3 is offset by the neutral loss under
investigation The precursor ions collide with Argon gas inQ2 to create fragment ions
Only those compounds which give a fragmenthaving that specific loss are detected
Since both Q1 and Q3 are scanning, neutralloss scan mode is slower than any other mode
Neutral loss scans
Neutral loss scans are used for screening experiments where agroup of compounds all give the same loss
-
8/3/2019 QuadTrap Comparison 1
101/105
101
g pgroup of compounds all give the same loss
NN2H
N2H
N
N
N2H
N2H - m/z 84
N
N
N2H
N 2H
N
N 2H
N2H
- m/z 84
NN
N - m/z 84
N
NOH
N
OH
- m/z 84
What are precursor ion scans ?
-
8/3/2019 QuadTrap Comparison 1
102/105
102
Precursor ion scans also known as parent ion scans
Q1 is scanned
Q3 is set to allow only a fragment ion of one m/z topass; (Q3 fixed)
ions collide with Argon gas in Q2 to create fragment or
product ions
Only those compounds which give that specificfragment ion are detected
Precursor ion scans
Precursor ion scans are used for screening experiments where agroup of compounds all give the same fragment ion
-
8/3/2019 QuadTrap Comparison 1
103/105
103
g p p g g
N
N
N
N
NOH
N
N
N
N
N2H
N 2H
N
N
N2H
N2Hm/z
84
m/z 162m/z 192
m/z 192
m/z 238
m/z 268
Comparison of MSn in a Triple Quad versesan Ion Trap Instrument.
-
8/3/2019 QuadTrap Comparison 1
104/105
104
Triple Quad(nonresonant excitation): Acceleration voltage applied
equally to all masses. Get a mix of ms2
, ms3
msn
products.
Ion Trap(resonant excitation): Excitation energy is in resonance with
only one mass at a time. Fragments, once formed, can not be furtherexcited unless they are purposely selected for next stage of MS.
Allows one to take apart a molecule in a controlled, step-wise fashion.
Comparison of Quads and Traps
Ion Traps Quadrupoles
-
8/3/2019 QuadTrap Comparison 1
105/105
105
Mass Separation in time Mass Separation in space
High sensitivity Full Scan Lower sensitivity Full Scan
Lower sensitivity SIM and SRM High sensitivity SIM and SRM
Offer multiple stages of MSn
Parent and neutral loss scans
Offers Only MS or MS/MS