quadtrap comparison 1

Upload: rmaytum

Post on 06-Apr-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/3/2019 QuadTrap Comparison 1

    1/105

  • 8/3/2019 QuadTrap Comparison 1

    2/105

    2

    Comparison of Quads and Traps

    Ion Traps Quadrupoles

    Mass Separation in time Mass Separation in space

    High sensitivity Full Scan Lower sensitivity Full Scan

    Lower sensitivity SIM and SRM High sensitivity SIM and SRM

    Offer multiple stages of MSn

    Parent and neutral loss scans

    Offers Only MS or MS/MS

  • 8/3/2019 QuadTrap Comparison 1

    3/105

    3

    LCQ Instrumentation

    ClassicClassic

    DecaDecaDuoDuo

  • 8/3/2019 QuadTrap Comparison 1

    4/105

    4

    Duo/Deca Comparisons

    LCQ DUO LCQ DECA 500um capillary

    1 square quadrupole & 1 octopole2 rotary pumps

    400um capillary

    2 octopolesOne rotary pump

    Deca: approximately 10 x better signal than Duo

  • 8/3/2019 QuadTrap Comparison 1

    5/105

    5

    Current LCQ Generation

  • 8/3/2019 QuadTrap Comparison 1

    6/105

    6

    Advantage/XP Comparisons

    450um Ion Transfer Tube Orthogonal Probes

    550um Ion Transfer Tube Orthogonal Probes

    XP: approximately 10 x better signal than Advantage

  • 8/3/2019 QuadTrap Comparison 1

    7/105

  • 8/3/2019 QuadTrap Comparison 1

    8/105

    8

    Mass Spectrometry Simplified

    G

    MS

    D

    enerate

    oveelect

    etect

    Ion productionIon production

    Ion opticsIon optics

    Mass filterMass filter

    Electron multiplierElectron multiplier

  • 8/3/2019 QuadTrap Comparison 1

    9/105

    9

    The Mass Spectrum

    Red

    Prism

    Light,all colors Green

    Blue

    100

    Ions,

    various massesMass

    Spectrometer200

    300

    T d d b i ll b d

  • 8/3/2019 QuadTrap Comparison 1

    10/105

    10

    Trace produced by summing all observedTrace produced by summing all observed

    masses in each scanmasses in each scan

    TotalTotal Ion Chromatogram or TICIon Chromatogram or TIC

  • 8/3/2019 QuadTrap Comparison 1

    11/105

    11

    Ionization vs. Fragmentation

    Ionization HardSoft

    No

    Fragments

    Fragments

    API EICI

  • 8/3/2019 QuadTrap Comparison 1

    12/105

    12

    EI and CI Mass Spectra of Ephedrine

    0

    50

    100

    0 20 40 60 80 100 120 140 160 180

    %

    RelativeIntensity

    m/z

    EI

    0

    50

    100

    0 20 40 60 80 100 120 140 160 180

    CI

    58

    148

    166

    MW = 165Th

    (M+H)+

  • 8/3/2019 QuadTrap Comparison 1

    13/105

    13

    Ionization Techniques

    ESI

    GCPBI

    TSPFAB

    Molecular

    W

    eight

    APCI

    200,000

    15,000

    1,000

    Non Polar Polar

  • 8/3/2019 QuadTrap Comparison 1

    14/105

    14

    What is API?

    Atmospheric Pressure Ionization

    Source Types:

    1. Electrospray Ionisation (ES) Solution phase process (forthe most part).

    2. APCI (Atmospheric Pressure Chemical Ionization) - Gas-phase process.

    Source Purpose:1. Ionize the analyte (APCI) or transport ion in solution to the

    gas phase.

    2. Desolvate sample flow for introduction into mass

    spectrometer.3. Baffle the first vacuum region of the MS from atmospheric

    pressure in the source.

    4. Pump away neutrals and opposite charged ions which would

    otherwise interfere with the analysis of the desired polarity.

  • 8/3/2019 QuadTrap Comparison 1

    15/105

    15

    LCQ Classic/Duo/Deca API Probes

    Electrospray Ionization (ESI)

    Peek insulator

    Atmospheric Pressure

    Chemical Interface (APCI)

  • 8/3/2019 QuadTrap Comparison 1

    16/105

    16

    LCQ Advantage/XP API Probes

    Atmospheric Pressure

    Chemical Interface (APCI)Electrospray Ionization (ESI)

    Orthogonal ESI & APCI probes

  • 8/3/2019 QuadTrap Comparison 1

    17/105

    17

    Chemistry Considerations

    ESI:

    Ions formed by solution chemistry

    Good for Thermally labile analytes

    Good for Polar analytes

    Good for Large Molecules (Proteins / Peptides)

    APCI:

    Ions formed by gas phase chemistry

    Good for Volatile / Thermally Stable

    Good for Non-polar analytes

    Good for Small Molecules (Steroids)

  • 8/3/2019 QuadTrap Comparison 1

    18/105

    18

    ESI Versatility Advantage/XP

    (Right: Picture represents

    a low flow position,

  • 8/3/2019 QuadTrap Comparison 1

    19/105

    19

    ElectrosprayBasic Layout

    Heated Capillary

    ESI Needle+/- 5 kV

    Taylor Cone

    Solvent evaporation and ion release

    ++

    +++ ++

    +

    ++

    +++ +

    ++

    +

    +

    +

    +++

    ++

    +

    +

    +

    ++

    +++ +

    ++

    ++

    +++ +

    ++

    ++

    ++

    ++

    +

    +

    +

    +

    +

    +

    ++

    ++

    ++

    +

    +

    +

    ++

    +

    + ++ +

    +++

    ++

    +

    ++

    +

    +++

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

    +

  • 8/3/2019 QuadTrap Comparison 1

    20/105

    20

    APCI Position on Advantage/XP

    In/Out MovableCorona Discharge Pin

    Total control for any

    flow rate (200ul/min -

    2000ul/min)

  • 8/3/2019 QuadTrap Comparison 1

    21/105

    21

    LC Flow Rates

    ESI:ESI:3 L/min - 1mL/minute

    Optimal Flow Rate: 200 L/minGenerally, higher flow rates require higher heatedcapillary temperatures and higher gas flow rates.

    APCI:APCI:200 L/min - 2mL/minute.Optimal Flow Rate: 500 L/minGenerally, higher flow rates require more sheath andauxiliary gas, but do not require higher heatedcapillary temperatures.

  • 8/3/2019 QuadTrap Comparison 1

    22/105

    22

    Theoretical Increase in Response

    DConc max =2

    Column 1D

    2Column 2

    Col. Diameter mm 1.0

    0.05

    21

    0.5

    2.3

    3.04.6

    0.2

    5

    2.0 Capillary

    Flow Rate ml/min 1 < 10 l / min1Theoretical

    Increase

    LC Additi

  • 8/3/2019 QuadTrap Comparison 1

    23/105

    23

    LC Additives

    Acids

    Do not use inorganic acids (may cause source corrosion)

    Formic and acetic acid are recommended

    Bases

    Do not use alkali metal bases (may cause source corrosion)

    Ammonium hydroxide is recommended

    Surfactants (surface active agents)

    Detergents and other surface active agents may suppressionization

    Trifluoroacetic Acid (TFA)

    May enhance chromatographic resolution, but causes ion

    suppression in both negative and positive ion mode

    Isopropyl Alcohol

    May Enhance Negative Ion Formation

    B ff ( H)

  • 8/3/2019 QuadTrap Comparison 1

    24/105

    24

    Buffers (pH)

    Avoid using non-volatile HPLC additives such as:

    Alkali Metal Phosphates

    Borates

    Citrates

    Keep Buffer concentrations below 20 mM using volatilesalts such as ammonium acetate.

    When using buffers, more frequent cleaning of the heatedcapillary and API stack will be necessary

    LC/MS Additi d B ff S

  • 8/3/2019 QuadTrap Comparison 1

    25/105

    25

    LC/MS Additives and Buffers Summary

    Acetic Acid

    Formic Acid

    Ammonium Hydroxide

    Ammonia Solutions

    Trichloroacetic Acid (< 0.1% v/v)

    Trifluoroacetic Acid (< 0.1% v/v)

    Isopropyl Alcohol

    (10% of organic phase)

    Ammonium Acetate

    Ammonium Formate

    Proton Donors

    Proton Acceptors

    Chromatographic

    Separation

    Negative ion

    formation

    Buffers

    C LC/MS S l t

  • 8/3/2019 QuadTrap Comparison 1

    26/105

    26

    Common LC/MS Solvents

    Methanol

    AcetonitrileWater

    IsopropanolDichloromethane

    Chloroform

    Hexane

    Effects of Solvents and Additives on ESI

  • 8/3/2019 QuadTrap Comparison 1

    27/105

    27

    Effects of Solvents and Additives on ESI

    50/50 MeOH/H2O

    50/50 ACN/H2O

    100 H2O

    100 MeOH

    100 ACN

    50/50 MeOH/H2O 1% Acetic

    50/50 MeOH/H2O 0.1% Formic

    50/50 ACN/H2O 1% Acetic

    50/50 ACN/H2O 0.1% Formic

    50/50 MeOH/H2O 5mM NH4OAc

    50/50 MeOH/H2O 10mM NH4OAc

    50/50 MeOH/H2O 0.1% TFA

    50/50 MeOH/H2O 0.05% TFA

    50/50 MeOH/H2O 0.02% TFA50/50 ACN/H2O 0.1% TFA

    50/50 ACN/H2O 0.05% TFA

    50/50 ACN/H2O 0.02% TFA

    50/50 MeOH/H2O 0.1% NH4OH

    50/50 ACN/H2O 0.1% NH4OH

    Solve

    ntSystem

    Counts (protonated ion species)

    0 100000 200000 300000 400000 500000

    Tyr-Gly-Gly-Phe-Leu

    Leucine Enkephalin

    API Stack

  • 8/3/2019 QuadTrap Comparison 1

    28/105

    28

    API Stack

    LCQClassic, LCQDUO,

    LCQDECALCQ DECAXP,LCQAdvantage

    Ion Transfer Tube and Removal Tool

  • 8/3/2019 QuadTrap Comparison 1

    29/105

    29

    Ion Transfer Tube and Removal Tool

    Removal

    tool

    Ion transfer tube

    Heated capillary

    Vent Prevent Mechanism

  • 8/3/2019 QuadTrap Comparison 1

    30/105

    30

    Vent Prevent Mechanism

    Heated Tube

    in-situ

    Heated Tube

    removed

    Tungsten Vent

    Prevent

    Ion Optics

  • 8/3/2019 QuadTrap Comparison 1

    31/105

    31

    Ion Optics

    First

    multipole

    Lens

    Intermultipole

    Lens

    Second

    multipoleLens

    Octapole

    Mount

    Vacuum

    Baffle

    Analyzer

    Mount

    IONS

    IN

    IONS

    OUT

    Multipole Potential Wells

  • 8/3/2019 QuadTrap Comparison 1

    32/105

    32

    Multipole Potential Wells

    Octapole

    Mass RangeTransEfficiency

    Round

    Quadrupole

    Square

    Quadrupole

    Mass Analyzer (Ion Trap)

  • 8/3/2019 QuadTrap Comparison 1

    33/105

    33

    Mass Analyzer (Ion Trap)

    Vacuum System

  • 8/3/2019 QuadTrap Comparison 1

    34/105

    34

    Vacuum System

    Every mass analyzer must operate under vacuum inEvery mass analyzer must operate under vacuum inorder to minimize both ion/molecule andorder to minimize both ion/molecule and

    molecule/ molecule collisions.molecule/ molecule collisions.

    At atmospheric pressure, the mean free path of a typicalAt atmospheric pressure, the mean free path of a typicalion is only ca. 52 nm and at 1ion is only ca. 52 nm and at 1 mTorrmTorr, it is 40 m., it is 40 m.

    Without vacuum, the ions produced in the source wontWithout vacuum, the ions produced in the source wontmake it to the detector.make it to the detector.

    The LCQ vacuum is maintained by a both rotary andThe LCQ vacuum is maintained by a both rotary andturbomolecularturbomolecularpumpspumps

    Ion Optics (Operating Pressures)

  • 8/3/2019 QuadTrap Comparison 1

    35/105

    35

    Ion Optics (Operating Pressures)

    1.3 torr760 torr 1.7x10-3 torr 2.0 x10-5 torr (1.0x10-5 torr He)

    3.5x10-3 torr He

    220 L/sec60 m

    3

    /hr 100 L/sec

    Steps to Ion Trap Scan Functions

  • 8/3/2019 QuadTrap Comparison 1

    36/105

    36

    Steps to Ion Trap Scan Functions

    Trapping- all scans

    Isolation- SIM and MSn

    Excitation- MSn

    Ejection- all scans

    Helium as a Damping Gas

  • 8/3/2019 QuadTrap Comparison 1

    37/105

    37

    Helium as a Damping Gas

    +

    +

    +

    +Without HeliumWithout Helium

    He

    collision HeHe

    He

    He

    +

    +

    +

    +

    With HeliumWith Helium

    Discovery of the Effects of Helium

  • 8/3/2019 QuadTrap Comparison 1

    38/105

    38

    y

    Helium as a Damping Gas

  • 8/3/2019 QuadTrap Comparison 1

    39/105

    39

    p g

    Ion Trap ResolutionEffect of Damping Gas

  • 8/3/2019 QuadTrap Comparison 1

    40/105

    40

    p p g

    Traps injected ions by removing kinetic energy

    Damps ion motion to center of trap

    Result...Increase in resolution and sensitivity

    Helium Effect

  • 8/3/2019 QuadTrap Comparison 1

    41/105

    41

    S#:1 RT:0.00 AV:1 SM:7G NL:2.50E7T:+ p Full ms

    514 516 518 520 522 524 526 528

    m/z

    0

    20

    40

    60

    80

    100

    RelativeAb

    undance

    524.3

    525.3

    S#:23-32 RT:0.71-1.00AV:10 SM:7G NL:5.61E7T:+ p Full ms

    500 1000 1500 2000

    m/z

    0

    20

    40

    60

    80

    100

    RelativeAb

    undance

    1522.04

    1621.971322.06

    1721.89

    1222.141821.95524.26

    1122.21 1921.88195.15

    1022.09

    S#:1 RT:0.02 AV:1 SM:7G NL:9.70E6T:+ p Full ms

    514 516 518 520 522 524 526 528m/z

    0

    20

    40

    60

    80

    100

    RelativeAbun

    dance

    522.6523.0

    521.8

    521.2 523.9

    520.7

    S#:23-32 RT:0.39-0.54AV:10 SM:7G NL:2.80E7T:+ p Full ms

    500 1000 1500 2000m/z

    0

    20

    40

    60

    80

    100

    RelativeAbun

    dance

    1620.791520.26

    1720.441320.95

    1220.75523.01 1919.96

    1120.90

    192.17

    Helium shut off and not flowing into trapHelium shut off and not flowing into trap

    S#:1 RT:0.00 AV:1 SM:7G NL:2.50E7T:+ p Full ms

    514 516 518 520 522 524 526 528

    m/z

    0

    20

    40

    60

    80

    100

    RelativeAb

    undance

    524.3

    525.3

    S#:23-32 RT:0.71-1.00AV:10 SM:7G NL:5.61E7T:+ p Full ms

    500 1000 1500 2000

    m/z

    0

    20

    40

    60

    80

    100

    RelativeAb

    undance

    1522.04

    1621.971322.06

    1721.89

    1222.141821.95524.26

    1122.21 1921.88195.15

    1022.09

    S#:1 RT:0.02 AV:1 SM:7G NL:9.70E6T:+ p Full ms

    514 516 518 520 522 524 526 528m/z

    0

    20

    40

    60

    80

    100

    RelativeAbun

    dance

    522.6523.0

    521.8

    521.2 523.9

    520.7

    S#:23-32 RT:0.39-0.54AV:10 SM:7G NL:2.80E7T:+ p Full ms

    500 1000 1500 2000m/z

    0

    20

    40

    60

    80

    100

    RelativeAbun

    dance

    1620.791520.26

    1720.441320.95

    1220.75523.01 1919.96

    1120.90

    192.17

    Helium shut off and not flowing into trapHelium shut off and not flowing into trap

    Helium flowing into trap

    Ion Trap Stability Diagram

  • 8/3/2019 QuadTrap Comparison 1

    42/105

    42

    y

    The region shaded blue indi-cates a (DC) and q (RF)

    values which provide stable

    trajectories in the r-directionThe region shaded yellow

    indicates the z-stable a and q

    combinations

    The green area where the r-

    and z-stable regions overlap

    indicates the a and q combi-nations under which ions will

    be stable in the trap

    Stability Diagram for Commercial Traps

  • 8/3/2019 QuadTrap Comparison 1

    43/105

    43

    q kV

    m ez =

    ( / )

    LCQ Scan-Out (Ejection) Rates

  • 8/3/2019 QuadTrap Comparison 1

    44/105

    44

    Normal Scan (5500 amu/sec)

    Common full, SIM, or MSn (SRM and CRM)scanning

    Resolution (FWHM) = 0.50, Mass Accuracy = 0.05

    Zoom Scan (280 amu/sec)

    Increases resolution and mass accuracy across anarrow range (allows charge state determination)

    Resolution (FWHM) = 0.15, Mass Accuracy = 0.02

    Turbo Scan (55,000 amu/sec)

    Decreases total scan time of a full scan, thus

    increasing number of scans across achromatographic peak

    Resolution (FWHM) = 3.0, Mass Accuracy = 0.5Used for better quantitation due to an increase

    of scans across a chromatographic peak

    What is AGC and Why Is it Important?

  • 8/3/2019 QuadTrap Comparison 1

    45/105

    45

    Camera AE

    Too much light degrades the image

    stored on film, causing a loss of

    color and image resolution.

    Too little light results in dark

    picture with no fine details visible.

    Cameras with high quality light

    meters and AE controls produce

    high quality pictures over a wide

    dynamic range of lighting

    conditions.

    LCQ Series AGC

    Controls amount of ions (light)

    entering the ion trap (film)

    Too many ions degrade the

    spectral quality in the trap,

    causing loss in mass resolution

    and mass assignment. Too fewions result in poor sensitivity to

    low level or minor components.

    AGC ensures excellent quality

    MS, SIM and MS/MS spectra,

    as well as excellent sensitivity

    over a wide dynamic range.

    Automatic Gain Control (AGC)

  • 8/3/2019 QuadTrap Comparison 1

    46/105

    46

    Prescan before the analytical scan

    - Measures the # of ions in the trap for a

    pre-defined time (10 ms)

    -Allows software to determine optimum

    ion injection time

    No AGC Spectrum of Ultramark 1621, Caffeine,MRFA Calibration Mixture space charging

  • 8/3/2019 QuadTrap Comparison 1

    47/105

    47

    MRFA Calibration Mixture space charging

    Spectrum of Ultramark 1621, Caffeine, MRFACalibration Mixture with AGC

  • 8/3/2019 QuadTrap Comparison 1

    48/105

    48

    Calibration Mixture with AGC

    AGC (Ion Population Control)

  • 8/3/2019 QuadTrap Comparison 1

    49/105

    49

    ~ 300 Ions ~ 1500 Ions ~ 3000 Ions ~ 6000 Ions

    530m/z

    0

    20

    40

    60

    80

    100

    Relative

    Abundance

    524.3

    525.3

    526.3

    530m/z

    0

    20

    40

    60

    80

    100

    Relative

    Abundance

    524.4

    525.4

    526.3

    527.5

    530m/z

    0

    20

    40

    60

    80

    100

    Relative

    Abundance

    524.5

    525.5

    526.5

    527.5

    530m/z

    0

    20

    40

    60

    80

    100

    Relative

    Abundance

    524.8

    525.7

    526.7

    522 522 522 522

    Poor ResolutionGood Resolution

    Calculation of Ion Time

  • 8/3/2019 QuadTrap Comparison 1

    50/105

    50

    Constant During Prescan

    AGC Prescan Signal =

    Multiplier Gain x Prescan TimeNumber of Ions x

    (3x105 counts) (10 ms)

    Calculated Ion Time = Target Value

    AGC Prescan Signal(how long the gate lens is open)

    Triplicate Injection of 5 nmol of MRFA (AGC ON)

  • 8/3/2019 QuadTrap Comparison 1

    51/105

    51

    Scan Number

    UnscaledTIC(counts

    )

    0.0E+00

    2.0E+074.0E+07

    6.0E+07

    8.0E+071.0E+08

    1.2E+08

    1.4E+08

    1.6E+081.8E+08

    2.0E+08

    0 100 200 300 400 500 600

    Unscaled TIC

    0

    10

    20

    30

    40

    50

    60

    0 100 200 300 400 500 600Scan Number

    InjectionTime(ms)

    Injection Time

    Scaled TIC

    0.0E+00

    1.0E+08

    2.0E+08

    3.0E+08

    4.0E+08

    5.0E+08

    ScaledTIC(counts)

    8.0E+08

    7.0E+08

    6.0E+08

    0 100 200 400 500 600300

    Scan Number

    Isolation of Ions

  • 8/3/2019 QuadTrap Comparison 1

    52/105

    52

    Ion we wish to isolateIon we wish to isolate

    7

    0.908qz

    0.0

    Ions at different qz values oscillate at

    different frequencies (

    o)

    22

    qzo

    Isolation Waveforms

  • 8/3/2019 QuadTrap Comparison 1

    53/105

    53

    ~ m/z 200

    q axis .908

    500 Hz

    16 msec

    q axis.908

    Why MS/MS or MSn ?

  • 8/3/2019 QuadTrap Comparison 1

    54/105

    54

    Signal to Noise Improvement

    0 1 2 3 4 5 6

    Stages of Analysis

    Intensity Signal

    Noise

    S/N

    MS/MS Parameters

  • 8/3/2019 QuadTrap Comparison 1

    55/105

    55

    Precursor ion m/z

    Excitation voltage

    Excitation qz

    1.0 2.0 3.0

    Excitation Voltage (V)

    Intensity

    Precursor ion

    (Product ions)

    0.0 4.0

    77001-1285

    970219

    Resonant Excitation qz Value

  • 8/3/2019 QuadTrap Comparison 1

    56/105

    56

    Fragmentation

    Energy

    0.908

    qz

    0.00.2250.225

    0.908

    qz

    0.0

    0.908

    qz

    0.0

    0.300.30

    0.450.45

    Fragment ions

    not trapped

    Product Ion

    m/z Range

    1/41/4

    1/31/3

    1/21/2

    Ion Trap Scan Functions

  • 8/3/2019 QuadTrap Comparison 1

    57/105

    57

    1. Collect

    2. Isolate

    3. Fragment

    4. Eject

    Summary (Ion Trap Functions)

  • 8/3/2019 QuadTrap Comparison 1

    58/105

    58

    1)1)CollectionCollection

    2)2)IsolationIsolation

    3)3)ExcitationExcitation

    4)4)Ejection

    For Scans: All

    By: Ring Electrode

    Method: Alternating RF

    frequency (760 kHz) at a set

    amplitude along with Hedampening gas traps and

    cools the ions to the center of

    the trap.Ejection

    Summary (Ion Trap Functions)

  • 8/3/2019 QuadTrap Comparison 1

    59/105

    59

    1)1)TrappingTrapping

    2)2)IsolationIsolation

    3)3)ExcitationExcitation

    4)4)Ejection

    For Scans: SIM, MSn

    By: Endcap Electrodes

    Method: a) Tailored waveform

    applied to all ions in the trap

    except ion of interest

    b) Thus, only ions of interest

    remain in the trap.

    Ejection

    Summary (Ion Trap Functions)

  • 8/3/2019 QuadTrap Comparison 1

    60/105

    60

    1)1)TrappingTrapping

    2)2)IsolationIsolation

    3)3)ExcitationExcitation

    4)4)Ejection

    For Scans: MSn

    By: Endcap Electrodes

    Method: a) Cool ion of

    interest back to set q value

    (default = 0.25). b) Apply

    custom RF waveform inresonance with the set q

    value, activation time

    (default = 30 msec), andoptimized activation

    amplitude.

    Ejection

    Summary (Ion Trap Functions)

  • 8/3/2019 QuadTrap Comparison 1

    61/105

    61

    1)1)TrappingTrapping

    2)2)IsolationIsolation

    3)3)ExcitationExcitation

    4)4)Ejection

    For Scans: All

    By: Ring Electrode

    Method: Ramp ring RFpower to increase the q values

    of all ions in desired scan

    range, low mass to high mass.(i.e. Mass Selective Scanning)

    Also, ramp the RF amplitude

    on the endcap electrodes toconsolidate the ions to a

    group (Resonance Ejection)

    Ejection

    Tune Page

    convection gauge

  • 8/3/2019 QuadTrap Comparison 1

    62/105

    62

    Ion gauge

    Pressure

  • 8/3/2019 QuadTrap Comparison 1

    63/105

    63

    Caffeine stock: 1 mg/ml in methanol

    MFRA stock: Dissolve 3.0 mg MRFA in 1 ml 50:50 methanol:water

    Ultramark stock: Measure 10 l of Ultramark 1621, and dissolve it in10 ml acetonitrile

    ESI calibration solutionESI calibration solution

    Into a clean vial pipette 100 l of caffeine stock, 5 l of MRFA stock and2.5 ml of Ultramark stock. Add 50 l of glacial acetic acid and 2.34 ml50:50 methanol:water

    Ion Trap Animation

  • 8/3/2019 QuadTrap Comparison 1

    64/105

    64

    The Eighth Generation Triple Quad

  • 8/3/2019 QuadTrap Comparison 1

    65/105

    65

    TSQ 15,

    TSQ 45,

    TSQ 46,TSQ 70,

    TSQ 700,TSQ 7000,

    TSQ,TSQ Quantum

    Smaller because

  • 8/3/2019 QuadTrap Comparison 1

    66/105

    66

    25 cm quad

    25 cm quad

    25 cm quad

    25 cm quad

    8degrees

    90 degrees

    TSQ 7000

    Quantum

    90 Degree

    Square

    quad

    collisioncell

    HyperQuadsTMHyperbolic Quadrupoles

  • 8/3/2019 QuadTrap Comparison 1

    67/105

    67

    Forms Pure Quadrupolar Fields

    Reduces Fringing Effects

    Significantly Improves Resolution Improves Transmission

    Improves Peak Shapes

    TSQ 7000

    1993 to 2000

    r0 = 4 mmL = 250 mm

    TSQ Quantum

    2001 to

    r0 = 6 mm

    L = 250 mm

    Quadrupole Mass Analyzer

  • 8/3/2019 QuadTrap Comparison 1

    68/105

    68

    +

    +

    ++

    The ion is transmitted along the quadrupole in a

    stable trajectory Rf field. The ion does not have astable trajectory and is ejected from the quadrupole.

    How does the Quadrupole work ?

  • 8/3/2019 QuadTrap Comparison 1

    69/105

    69

    The quadrupole consists of four parallel rods. The opposing rodshave the same polarity while adjacent rods have opposite polarity.

    -ve

    +ve

    Each rod is applied with a DC and

    an RF voltage. Ions are scanned byvarying the DC/Rf quadrupole

    voltages.

    Only ions with the selected massto charge ratio will have the

    correct oscillatory pathway in the

    Rf field.

    Effect of Peak Width On Transmission

  • 8/3/2019 QuadTrap Comparison 1

    70/105

    70

    0

    20

    40

    60

    80

    100

    %

    T

    r

    a

    n

    s

    m

    i

    s

    s

    i

    o

    n

    2 1.5 0.7 0.5 0.2 0.1

    Peak Width FWHM

    HYPERQUAD ROUND RODS

    Effect of Peak Width on Resolution

  • 8/3/2019 QuadTrap Comparison 1

    71/105

    71

    0

    2000

    4000

    6000

    8000

    10000

    12000

    0 200 400 600 800 1000 1200

    m/z

    R

    esolution

    Quad 0.7 FWHM

    Quad 0.1 FWHM

    Sector

    Quantum, API 4000, Ultima

    at 0.7 FWHM at m/z 1000

    R = 1428

    R is relatively flat across m/z

    Quantum operating at

    0.1 FWHM at m/z 1000

    R = 10,000

    The Power of Resolution

  • 8/3/2019 QuadTrap Comparison 1

    72/105

    72

    Separation of ions with samenominal m/z value

    Unequivocal determination ofcharge state (ESI)

    High resolution precursor ionselection for MS/MS

    High resolution product ion for

    charge state determination

    Effect of changing resolution on peak shape

  • 8/3/2019 QuadTrap Comparison 1

    73/105

    73

    1.0 FWHM 0.7 FWHM 0.5 FWHM

    0.3 FWHM0.2 FWHM 0.1 FWHM

    Resolution vs. Intensity

  • 8/3/2019 QuadTrap Comparison 1

    74/105

    74

    1.0 FWHM 0.7 FWHM 0.5 FWHM

    2.5e6 2.1e61.9e6

    Resolution vs. Intensity

  • 8/3/2019 QuadTrap Comparison 1

    75/105

    75

    0.3 FWHM 0.2 FWHM 0.1 FWHM

    0.8 e61.5e6 1.4e6

    Quadrupole Mass Analyzer

  • 8/3/2019 QuadTrap Comparison 1

    76/105

    76

    If one MS scan between m/z 100 and 500is completed in one second, then eachm/z will be allowed to pass for 2.5 ms.

    1000 ms== 2.5 ms/amu

    400 amu

    +

    ++

    +

    +

    To

    detector

    +

    Quadrupole Mass Analyzer

  • 8/3/2019 QuadTrap Comparison 1

    77/105

    77

    And, for the same peak, for example, thequadrupole performs 5 complete scansfrom 100 500 Da each taking 1 sec.

    5 sec

    100 500 Da

    100 500 Da

    100 500 Da

    100 500 Da 100 500 Da

    Ion Trap Pre-Scan

  • 8/3/2019 QuadTrap Comparison 1

    78/105

    78

    The length of time that the trap staysopen to collect ions is determined by a

    pre-scan which measures total ion current(prevents space charging, so no ghostpeaks)

    5 sec

    Pre-Scan

    Ion Trap Pre-Scan Contd

  • 8/3/2019 QuadTrap Comparison 1

    79/105

    79 5 sec

    Across a peak 5 sec. wide the trap mightfill and empty 5 times. So, a group of ionsare collected ca. every 1 sec each group

    is then ejected to the detector, smallerions first. 1 sec

    1 sec

    1 sec

    1 sec

    1 sec

    Comparison of Quads and Traps

  • 8/3/2019 QuadTrap Comparison 1

    80/105

    80

    But for a trap, each m/z (andall m/z at the same time) is/are

    collected for ca. 750 ms (taking

    pre-scan and interscan timesinto account) and then scanned

    to the detector.

    For the quadrupole,

    each m/z is scanned

    (sequentially)

    to the detector for 2.5

    ms.

    +

    +

    ++

    Comparison of Quads and Traps

  • 8/3/2019 QuadTrap Comparison 1

    81/105

    81

    A trap will fill similar tofilling a glass with water. All

    ions entering the trap will be

    collected until the trap fillswith a pre-defined amount of

    ions (AGC target value).

    At any one particularinstant, a quad will only

    scan/look for only one

    m/z. All other m/z willbe ignored. In this

    example, each m/z is

    scanned for 2.5 ms.

    +

    +

    ++

    Comparison of Quads and Traps

  • 8/3/2019 QuadTrap Comparison 1

    82/105

    82

    So, for full scan MS, a trap will givebetter sensitivity because there aremore ions representing each m/zarriving at the detector for each scan.

    +

    ++

    +

    How Much More???

  • 8/3/2019 QuadTrap Comparison 1

    83/105

    83

    Accounting for pre-scan/interscantiming, the trap produces ca. 300 times(750 ms/2.5 ms) for the collection ofeach m/z compared to a quadrupole (i.e

    2 orders of magnitude).

    +

    +++

    But

  • 8/3/2019 QuadTrap Comparison 1

    84/105

    84

    What if I wanted to pass (filter) only onem/z ion to the detector (i.e. SIM or SRM) then I could spend more time on that ion

    +

    + ++

    Comparison of Quads and Traps

  • 8/3/2019 QuadTrap Comparison 1

    85/105

    85

    Yes, on a quadrupole,

    interscan times are

    relatively short and sothe quadrupole remains

    fixed on that one ion a

    duty cycle of close to100%

    But a trap will still only

    collect ions in batches-

    and prescan/interscantimes afford a duty cycle of

    about 75%

    +

    +++

    What is a Duty Cycle?

  • 8/3/2019 QuadTrap Comparison 1

    86/105

    86

    Scan 1 ISD Scan 2

    Definition:-

    Time taken to acquire 1 scan and be ready to acquire

    the next one

    Interscan delay (ISD) is the time taken to return all

    system voltages to the start values and reach a stable

    stateThis is dead time and should be minimized

    Duty Cycle

    In Addition

    Whil h b i i i l

  • 8/3/2019 QuadTrap Comparison 1

    87/105

    87

    While the beam instrument is continuouslydetecting one particular m/z a trap builds acurve from an average over each collection

    time and the points are least frequent atthe most important region for quantitation

    (the take off).

    For Example

    SRM f 5 Al l ith LCQ D

  • 8/3/2019 QuadTrap Comparison 1

    88/105

    88

    SRM of 5 pg Alprazolam with LCQ Decagives a %RSD of 6.11 while

    SRM of 750 fg Alprazolam with TSQ 7000gives a %RSD of 1.87.

    Signal to noise ratio (S/N) are similar

    (20:1 and 28:1 respectively).

    100

    RT: 4.26

    SN: 20

    %RSD 6 11%RSD 6 11 SRM 5SRM 5 Al lAl l

    90

    80

    an

    70ce

  • 8/3/2019 QuadTrap Comparison 1

    89/105

    89

    0

    10

    20

    30

    40

    50

    Relative

    %RSD 6.11%RSD 6.11withwithLCQLCQDecaDeca

    SRM 5pgSRM 5pg AlprazolamAlprazolam60A

    bundan

    RT: 4.41

    SRM 750fgSRM 750fg AlprazolamAlprazolam with TSQwith TSQ

    SN: 28

    %RSD 1.87%RSD 1.87

    0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5Time (min)

    0

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    The Effect of Ion Trap Scan Speeds onQuantitative Performance

  • 8/3/2019 QuadTrap Comparison 1

    90/105

    90

    In general, faster scanning produces more points

    across a chromatographic peak, hence better

    precision and lower LOQs.

    In fact, for an ion-trap, scan speed refers only to

    the time taken to scan ions from the trap duringmass analysis.

    Scan speed does not refer to the total analyticalcycle. In an ion trap device, an MSn analytical

    scan comprises at least four events:

    The Effect of Ion Trap Scan Speeds onQuantitative Performance

  • 8/3/2019 QuadTrap Comparison 1

    91/105

    91

    1- AGC Pre-scan

    2- Ion Injection (usually the rate-determining step)

    3- Isolation and activation of the parent ion within

    the trap

    4- Scanning the ions out of the trap (mass analysis)

    MS Scan FunctionMass

    Analysis

    Ion

  • 8/3/2019 QuadTrap Comparison 1

    92/105

    92

    IonActivation

    Ion

    isolation

    Ion

    Injection

    Analytical ScanAGC Prescan

    Scan TerminologyScan Terminology

  • 8/3/2019 QuadTrap Comparison 1

    93/105

    93

    Prescan Mass Analysis Prescan Mass Analysis Prescan Mass Analysis

    1st Microscan 2nd Microscan 3rd Microscan Save Data

    Complete Scan Cycle

    The Effect of Ion Trap Scan Speeds onQuantitative Performance

  • 8/3/2019 QuadTrap Comparison 1

    94/105

    94

    The injection time constitutes the majority of the

    total scan time.

    For good quantitative reproducibility, it is

    necessary to take enough points to precisely

    determine the chromatographic peak particularly

    at the take-off point.

    Increasing the scan speed in the trap does not

    significantly increase the number of data pointstaken across the peak.

    Quantitation m/z 500 scanning 135 510

    Pre-scan 0 msec

    Isol /Activ / Download 80

    Pre-scan 0 msec

    Isol /Activ / Download 80

    Scan speeds in ion traps

  • 8/3/2019 QuadTrap Comparison 1

    95/105

    95

    Pre-scan 60 msec

    Isol/Activ/ Download 80

    375 amu @ 5500 amu/sec 70

    Injection time 500

    Total scan time 710

    Scans / 10 sec wide peak 14

    Pre-scan 60 msec

    Isol/Activ/ Download 80

    375 amu @ 5500 amu/sec 70

    Injection time 500

    Total scan time 710

    Scans / 10 sec wide peak 14

    Quantitation m/z 500 scanning 135-510Pre-scan 60 msec

    Isol /Activ / Download 80375 amu @ 13,000 amu/sec 30

    Injection time 500

    Total scan time 670

    Scans / 10 sec wide peak 15

    Pre-scan 60 msec

    Isol /Activ / Download 80

    375 amu @ 13,000 amu/sec 30

    Injection time 500

    Total scan time 670

    Scans / 10 sec wide peak 15

    Isol /Activ / Download 80

    375 amu @ 13,000 amu/sec 30

    Injection time 500

    Total scan time 610

    Scans / 10 sec wide peak 16

    Isol /Activ / Download 80

    375 amu @ 13,000 amu/sec 30

    Injection time 500Total scan time 610

    Scans / 10 sec wide peak 16

    The Effect of Ion Trap Scan Speeds onQuantitative Performance

  • 8/3/2019 QuadTrap Comparison 1

    96/105

    96

    The only significant way to increase the sampling

    rate across the peak is to reduce the injectiontime which can be achieved in two ways:

    1- Set a lower max injection time which reduces the

    number of ions in the trap, hence sensitivity

    2- Increase the efficiency of the source and lenses

    to improve the transmission of ions; I.e filling thetrap to the same level in a shorter period of time.

    Data Dependant Acquisition of MSn Spectra

    Data Dependant Acquisition: Intelligent decision-making software that

    selects precursor ion for MSn experiments based on user define criteria

  • 8/3/2019 QuadTrap Comparison 1

    97/105

    97

    selects precursor ion for MSn experiments based on user-define criteria.

    Critical for metabolite screening experiments.

    Scan event 1

    Full-Scan MSScan event 2

    Full-Scan MS2

    Software selects most intenseion from scan event 1 as

    precursor ion for ms2

    experiment in scan event 2,

    provide that its intensity is

    above a user selected threshold

    m/zm/z

    Dynamic Exclusion-- MS and MS/MS of Co-Eluters

    MS

    MS/MS

  • 8/3/2019 QuadTrap Comparison 1

    98/105

    98

    407

    407MS

    Time

    452

    452

    365

    206

    377255

    MS/MSm/z

    377

    231

    Threshold

    MS MS/MS

    MS/MS

    MS

    MS/MS

    MS

    MS MS

    MS

    m/z

    m/z

    m/z

    Comparison of Quads and Traps

    Major Strengths of Triple Quads

  • 8/3/2019 QuadTrap Comparison 1

    99/105

    99

    Major Strengths of Triple Quads

    SRM Sensitivity

    Neutral Loss Scan Mode Parent (Precursor) Scan Mode

    Major Strengths of the LCQ Deca XP Plus

    MSn Scan Mode

    Full Scan MS/MS Sensitivity

    Consecutive Reaction Monitoring (CRM)

    What are neutral loss scans ?

    Both Q1 and Q3 are scanned together

  • 8/3/2019 QuadTrap Comparison 1

    100/105

    100

    Both Q1 and Q3 are scanned together

    Q3 is offset by the neutral loss under

    investigation The precursor ions collide with Argon gas inQ2 to create fragment ions

    Only those compounds which give a fragmenthaving that specific loss are detected

    Since both Q1 and Q3 are scanning, neutralloss scan mode is slower than any other mode

    Neutral loss scans

    Neutral loss scans are used for screening experiments where agroup of compounds all give the same loss

  • 8/3/2019 QuadTrap Comparison 1

    101/105

    101

    g pgroup of compounds all give the same loss

    NN2H

    N2H

    N

    N

    N2H

    N2H - m/z 84

    N

    N

    N2H

    N 2H

    N

    N 2H

    N2H

    - m/z 84

    NN

    N - m/z 84

    N

    NOH

    N

    OH

    - m/z 84

    What are precursor ion scans ?

  • 8/3/2019 QuadTrap Comparison 1

    102/105

    102

    Precursor ion scans also known as parent ion scans

    Q1 is scanned

    Q3 is set to allow only a fragment ion of one m/z topass; (Q3 fixed)

    ions collide with Argon gas in Q2 to create fragment or

    product ions

    Only those compounds which give that specificfragment ion are detected

    Precursor ion scans

    Precursor ion scans are used for screening experiments where agroup of compounds all give the same fragment ion

  • 8/3/2019 QuadTrap Comparison 1

    103/105

    103

    g p p g g

    N

    N

    N

    N

    NOH

    N

    N

    N

    N

    N2H

    N 2H

    N

    N

    N2H

    N2Hm/z

    84

    m/z 162m/z 192

    m/z 192

    m/z 238

    m/z 268

    Comparison of MSn in a Triple Quad versesan Ion Trap Instrument.

  • 8/3/2019 QuadTrap Comparison 1

    104/105

    104

    Triple Quad(nonresonant excitation): Acceleration voltage applied

    equally to all masses. Get a mix of ms2

    , ms3

    msn

    products.

    Ion Trap(resonant excitation): Excitation energy is in resonance with

    only one mass at a time. Fragments, once formed, can not be furtherexcited unless they are purposely selected for next stage of MS.

    Allows one to take apart a molecule in a controlled, step-wise fashion.

    Comparison of Quads and Traps

    Ion Traps Quadrupoles

  • 8/3/2019 QuadTrap Comparison 1

    105/105

    105

    Mass Separation in time Mass Separation in space

    High sensitivity Full Scan Lower sensitivity Full Scan

    Lower sensitivity SIM and SRM High sensitivity SIM and SRM

    Offer multiple stages of MSn

    Parent and neutral loss scans

    Offers Only MS or MS/MS