panoura 18s pan-only calibration and alignment quick guide
TRANSCRIPT
Panoura 18SPanoramic X-Ray Calibration and Alignment
05/0
1/20
23
2ImageWorks Proprietary – Confidential Information
These instructions serve as supplemental assembly guidance of the Panoura. They are not meant to
replace the installation manual, but rather serve to complement it.
If there are any questions, please do not hesitate to contact us
914-592-6100
Summary
05/0
1/20
23
3ImageWorks Proprietary – Confidential Information
1. Computer set up2. Tube head aging3. Panoramic alignment4. Panoramic sensor calibration5. Final Test Exposure
Agenda
05/0
1/20
23
4ImageWorks Proprietary – Confidential Information
This procedure assumes PC was either provided or staged by ImageWorks
Start PC and open Task Manager (Ctl-Alt-Del and choose Task Manager)Verify that these processes are running:
1. P18_CM.exe2. P18_Launcher.exe3. P18S_Setup.exe4. Panoura18S.exe
1. Computer Set Up
05/0
1/20
23
5ImageWorks Proprietary – Confidential Information
Confirm that key programs are running by looking for the following icons in the system tray (usually bottom right of screen):
1. Control Manager
2. Image Creator
1. Computer Set Up
05/0
1/20
23
6ImageWorks Proprietary – Confidential Information
On the computer in the lower right hand corner of the screen, Double click the Control Manager icon (white tooth on light blue background).
Hold down the shift key and click on the setup button to open the CS service menus.
It will ask for password, which is p18p18
Select CS95 to perform Tube Head Aging Routing
2. Tube Head Aging
05/0
1/20
23
7ImageWorks Proprietary – Confidential Information
Follow directions on screen. To initiate exposure, press hand switch and hold until beeping stops (x-ray cycle will be about 4 seconds).
Screen will show countdown during cool-down period and will say when it’s ready for next exposure(about 80 seconds).
Repeat this process 13 times. At the completion of the 13th cycle, the screen should show “Pass”
Click on “Main” tab to exit CS mode
2. Tube Head AgingNote:
If exposure switch is pressed before cooling cycle is complete, an
FL03 warning is displayed.
If switch is released before exposure is complete, an ER08
warning is displayed.
These error codes can be cleared by pressing
RESET on panel.
05/0
1/20
23
8ImageWorks Proprietary – Confidential Information
Repeat this process 13 times. At the completion of the 13th cycle, the screen should show “Pass”
Click on “Main” tab to exit CS mode
2. Tube Head Aging
05/0
1/20
23
9ImageWorks Proprietary – Confidential Information
3. Panoramic Mechanical Alignment
If inside cover on tube head is in place, remove this to reveal collimater assembly
Collimator
assembly
05/0
1/20
23
10ImageWorks Proprietary – Confidential Information
For mechanical alignment and calibration, the FL07 exposure lock must be switched off. Do this by pressing and holding down the MODE and FIX buttons on the indicator panel simultaneously for about 2 seconds.Make sure sensor is installed in Pan position
3. Panoramic Mechanical Alignment
Open Ajat Dental Imaging Software
Select Tools Maintenance Mechanical Alignment
The mechanical alignment tools screen is displayed
05/0
1/20
23
11ImageWorks Proprietary – Confidential Information
On the computer in the lower right hand corner of the screen, Double click the Control Manager icon (white tooth on light blue background).
Hold down the shift key and click on the setup button to open the CS service menus.
It will ask for password, which is p18p18
Select CS75 to set parameters for the scan.
3. Panoramic Mechanical Alignment
05/0
1/20
23
12ImageWorks Proprietary – Confidential Information
Adjust parameters (kv, ma and time)click “Set” …
…to match those listed in the mechanical alignment tool
3. Panoramic Mechanical Alignment
05/0
1/20
23
13ImageWorks Proprietary – Confidential Information
Place 20 mm aluminum block over the pan collimator slit
(a magnetized block does notrequire any tape)
3. Panoramic Mechanical Alignment
05/0
1/20
23
14ImageWorks Proprietary – Confidential Information
Press handswitch to initiate x-ray. White column will appear on the screen with arrows above and below showing how far each end is out of alignment.
Number on top and bottom represent distance each side is out of alignment. If number is less than 0,20 mm, it will be in green, which is within the tolerance for alignment. The objective is to adjust collimater position such that both top and bottom are green (< 0,20 mm).
If either number is red (greater than 0,20), then the collimater must be adjusted.
Note: these distances do NOT correlate with how far collimater is out of alignment
3. Panoramic Mechanical Alignment
05/0
1/20
23
15ImageWorks Proprietary – Confidential Information
The direction of the arrow represents the side to which the beam has moved too far when facing the sensor. Therefore, the beam must be moved in the opposite direction when facing the sensor.
In the example shown, the red arrows pointing to the left means that the beam is too far to the left (when facing the sensor), and must be moved to the right. Therefore, in this situation, the collimator slit must be moved to the left when facing the collimator.
Tip: The mechanical alignment process is most efficient if each adjustment of the collimator is done with deliberation and care. Small adjustments to collimator can have large impact on alignment, so it’s best to start with very small adjustments so that larger misalignments don’t result.
3. Panoramic Mechanical Alignment
05/0
1/20
23
16ImageWorks Proprietary – Confidential Information
To adjust collimator left or right, slightly loosen the two adjusting screws and gently move the slit in the desired direction.
Note: Do not overly loosen the screws as the slit will not stay in the desired position after it is moved.
Note: vertical or rotational adjustments may be required. Please refer to manual
3. Panoramic Mechanical Alignment
Adjusting screws
Panoramic Slit
05/0
1/20
23
17ImageWorks Proprietary – Confidential Information
Turn off FL07 by pressing and holding down the MODE and FIX buttons on the indicator panel simultaneously for about 2 seconds.
Repeat steps to open Control Manager • Double click the Control Manager icon• Hold down the shift key and click on the setup button to open the CS
service menus.• Enter password: p18p18• Select CS75 to set parameters for the scan.
Initiate scan.
If arrows are red, repeat collimator position adjustment steps
If arrows are green, tighten adjusting screws, and repeat steps for power cycling the unit, reopening Control Manager and scanning.Confirm that arrows remain green after tightening. If all is green, then click Close.
3. Panoramic Mechanical Alignment
05/0
1/20
23
18ImageWorks Proprietary – Confidential Information
4. Panoramic Sensor CalibrationRemove all adapters such as chin rest block from irradiation path of the x-ray
Open Ajat Dental Imaging Software
Select Tools Maintenance Calibration
Calibration screen is displayed. Click Calibrate
Click Yes
Note: unit needs to be
powered on for about 5 min to
perform calibration
05/0
1/20
23
19ImageWorks Proprietary – Confidential Information
4. Panoramic Sensor CalibrationThe calibration includes a series of 3 x-ray exposures.
The display details the parameters needed for each exposure. Set the indicator panel to reflect these parameters.
Note: when 20 mm Al is displayed, then the aluminum block must be attached over the panoramic slit. When “No Filter” is displayed, then remove the block
Repeat cycles updating parameters requested for each exposures
When the calibration is finished, press Ok
05/0
1/20
23
20ImageWorks Proprietary – Confidential Information
Remove chinrest block
Mount installation phantom
5. Final Test Exposure
05/0
1/20
23
21ImageWorks Proprietary – Confidential Information
Run exposure and evaluate image
Seven ball bearings should be checked to see that:1. Each ball appears circular2. Distance between balls is symmetrical to left and right of
center ball
5. Final Test Exposure
Note: Because the test scan does
not have patient anatomy, it is possible that lines or artifacts may be seen in other parts of the test image. This will not affect quality of actual dental
images taken.
05/0
1/20
23
22ImageWorks Proprietary – Confidential Information
These instructions serve as supplemental assembly guidance of the Panoura. They are not meant to
replace the installation manual, but rather serve to complement it.
If there are any questions, please do not hesitate to contact us
914-592-6100
Summary