Page 1: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

Panoura 18SPanoramic X-Ray Calibration and Alignment

Page 2: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

2ImageWorks Proprietary – Confidential Information

These instructions serve as supplemental assembly guidance of the Panoura. They are not meant to

replace the installation manual, but rather serve to complement it.

If there are any questions, please do not hesitate to contact us

914-592-6100

Summary

Page 3: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

3ImageWorks Proprietary – Confidential Information

1. Computer set up2. Tube head aging3. Panoramic alignment4. Panoramic sensor calibration5. Final Test Exposure

Agenda

Page 4: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

4ImageWorks Proprietary – Confidential Information

This procedure assumes PC was either provided or staged by ImageWorks

Start PC and open Task Manager (Ctl-Alt-Del and choose Task Manager)Verify that these processes are running:

1. P18_CM.exe2. P18_Launcher.exe3. P18S_Setup.exe4. Panoura18S.exe

1. Computer Set Up

Page 5: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

5ImageWorks Proprietary – Confidential Information

Confirm that key programs are running by looking for the following icons in the system tray (usually bottom right of screen):

1. Control Manager

2. Image Creator

1. Computer Set Up

Page 6: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

6ImageWorks Proprietary – Confidential Information

On the computer in the lower right hand corner of the screen, Double click the Control Manager icon (white tooth on light blue background).

Hold down the shift key and click on the setup button to open the CS service menus.

It will ask for password, which is p18p18

Select CS95 to perform Tube Head Aging Routing

2. Tube Head Aging

Page 7: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

7ImageWorks Proprietary – Confidential Information

Follow directions on screen. To initiate exposure, press hand switch and hold until beeping stops (x-ray cycle will be about 4 seconds).

Screen will show countdown during cool-down period and will say when it’s ready for next exposure(about 80 seconds).

Repeat this process 13 times. At the completion of the 13th cycle, the screen should show “Pass”

Click on “Main” tab to exit CS mode

2. Tube Head AgingNote:

If exposure switch is pressed before cooling cycle is complete, an

FL03 warning is displayed.

If switch is released before exposure is complete, an ER08

warning is displayed.

These error codes can be cleared by pressing

RESET on panel.

Page 8: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

8ImageWorks Proprietary – Confidential Information

Repeat this process 13 times. At the completion of the 13th cycle, the screen should show “Pass”

Click on “Main” tab to exit CS mode

2. Tube Head Aging

Page 9: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

9ImageWorks Proprietary – Confidential Information

3. Panoramic Mechanical Alignment

If inside cover on tube head is in place, remove this to reveal collimater assembly

Collimator

assembly

Page 10: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

10ImageWorks Proprietary – Confidential Information

For mechanical alignment and calibration, the FL07 exposure lock must be switched off. Do this by pressing and holding down the MODE and FIX buttons on the indicator panel simultaneously for about 2 seconds.Make sure sensor is installed in Pan position

3. Panoramic Mechanical Alignment

Open Ajat Dental Imaging Software

Select Tools Maintenance Mechanical Alignment

The mechanical alignment tools screen is displayed

Page 11: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

11ImageWorks Proprietary – Confidential Information

On the computer in the lower right hand corner of the screen, Double click the Control Manager icon (white tooth on light blue background).

Hold down the shift key and click on the setup button to open the CS service menus.

It will ask for password, which is p18p18

Select CS75 to set parameters for the scan.

3. Panoramic Mechanical Alignment

Page 12: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

12ImageWorks Proprietary – Confidential Information

Adjust parameters (kv, ma and time)click “Set” …

…to match those listed in the mechanical alignment tool

3. Panoramic Mechanical Alignment

Page 13: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

13ImageWorks Proprietary – Confidential Information

Place 20 mm aluminum block over the pan collimator slit

(a magnetized block does notrequire any tape)

3. Panoramic Mechanical Alignment

Page 14: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

14ImageWorks Proprietary – Confidential Information

Press handswitch to initiate x-ray. White column will appear on the screen with arrows above and below showing how far each end is out of alignment.

Number on top and bottom represent distance each side is out of alignment. If number is less than 0,20 mm, it will be in green, which is within the tolerance for alignment. The objective is to adjust collimater position such that both top and bottom are green (< 0,20 mm).

If either number is red (greater than 0,20), then the collimater must be adjusted.

Note: these distances do NOT correlate with how far collimater is out of alignment

3. Panoramic Mechanical Alignment

Page 15: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

15ImageWorks Proprietary – Confidential Information

The direction of the arrow represents the side to which the beam has moved too far when facing the sensor. Therefore, the beam must be moved in the opposite direction when facing the sensor.

In the example shown, the red arrows pointing to the left means that the beam is too far to the left (when facing the sensor), and must be moved to the right. Therefore, in this situation, the collimator slit must be moved to the left when facing the collimator.

Tip: The mechanical alignment process is most efficient if each adjustment of the collimator is done with deliberation and care. Small adjustments to collimator can have large impact on alignment, so it’s best to start with very small adjustments so that larger misalignments don’t result.

3. Panoramic Mechanical Alignment

Page 16: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

16ImageWorks Proprietary – Confidential Information

To adjust collimator left or right, slightly loosen the two adjusting screws and gently move the slit in the desired direction.

Note: Do not overly loosen the screws as the slit will not stay in the desired position after it is moved.

Note: vertical or rotational adjustments may be required. Please refer to manual

3. Panoramic Mechanical Alignment

Adjusting screws

Panoramic Slit

Page 17: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

17ImageWorks Proprietary – Confidential Information

Turn off FL07 by pressing and holding down the MODE and FIX buttons on the indicator panel simultaneously for about 2 seconds.

Repeat steps to open Control Manager • Double click the Control Manager icon• Hold down the shift key and click on the setup button to open the CS

service menus.• Enter password: p18p18• Select CS75 to set parameters for the scan.

Initiate scan.

If arrows are red, repeat collimator position adjustment steps

If arrows are green, tighten adjusting screws, and repeat steps for power cycling the unit, reopening Control Manager and scanning.Confirm that arrows remain green after tightening. If all is green, then click Close.

3. Panoramic Mechanical Alignment

Page 18: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

18ImageWorks Proprietary – Confidential Information

4. Panoramic Sensor CalibrationRemove all adapters such as chin rest block from irradiation path of the x-ray

Open Ajat Dental Imaging Software

Select Tools Maintenance Calibration

Calibration screen is displayed. Click Calibrate

Click Yes

Note: unit needs to be

powered on for about 5 min to

perform calibration

Page 19: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

19ImageWorks Proprietary – Confidential Information

4. Panoramic Sensor CalibrationThe calibration includes a series of 3 x-ray exposures.

The display details the parameters needed for each exposure. Set the indicator panel to reflect these parameters.

Note: when 20 mm Al is displayed, then the aluminum block must be attached over the panoramic slit. When “No Filter” is displayed, then remove the block

Repeat cycles updating parameters requested for each exposures

When the calibration is finished, press Ok

Page 20: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

20ImageWorks Proprietary – Confidential Information

Remove chinrest block

Mount installation phantom

5. Final Test Exposure

Page 21: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

21ImageWorks Proprietary – Confidential Information

Run exposure and evaluate image

Seven ball bearings should be checked to see that:1. Each ball appears circular2. Distance between balls is symmetrical to left and right of

center ball

5. Final Test Exposure

Note: Because the test scan does

not have patient anatomy, it is possible that lines or artifacts may be seen in other parts of the test image. This will not affect quality of actual dental

images taken.

Page 22: Panoura 18S Pan-Only Calibration and Alignment Quick Guide

05/0

1/20

23

22ImageWorks Proprietary – Confidential Information

These instructions serve as supplemental assembly guidance of the Panoura. They are not meant to

replace the installation manual, but rather serve to complement it.

If there are any questions, please do not hesitate to contact us

914-592-6100

Summary