panoura 18s with ceph operator training quick guide

28
Panoura 18S Panoramic X-Ray with Ceph Operation Training Quick Guide

Upload: don-vibbert

Post on 20-Jan-2017

39 views

Category:

Healthcare


0 download

TRANSCRIPT

Page 1: Panoura 18S with Ceph Operator Training Quick Guide

Panoura 18SPanoramic X-Ray with Ceph Operation Training

Quick Guide

Page 2: Panoura 18S with Ceph Operator Training Quick Guide

2ImageWorks Proprietary – Confidential Information

These instructions serve as supplemental guidance on the Panoura 18S. They are not meant to replace

the installation manual, but rather serve to complement it.

If there are any questions, please do not hesitate to contact us

914-592-6100

Summary

Page 3: Panoura 18S with Ceph Operator Training Quick Guide

3ImageWorks Proprietary – Confidential Information

1. Unit Overview

2. Powering on and Emergency Stop

3. Pan – Prep Unit for Patient

4. Pan – Position Patient

5. Pan – Initiate Scan

6. Ceph - Prep Unit for Patient

7. Ceph - Position Patient

8. Ceph – Initiate Scan

9. Changing Modes

Contents

Page 4: Panoura 18S with Ceph Operator Training Quick Guide

4ImageWorks Proprietary – Confidential Information

Forehead and temple supports

Bite Stick

Chin rest

Hand Holds

X-ray head

Image sensor

Console

1. Unit Overview

Page 5: Panoura 18S with Ceph Operator Training Quick Guide

5ImageWorks Proprietary – Confidential Information

Console

1. Unit Overview

Handswitch

Page 6: Panoura 18S with Ceph Operator Training Quick Guide

6ImageWorks Proprietary – Confidential Information

2. Power Switch

Power Switch

Page 7: Panoura 18S with Ceph Operator Training Quick Guide

7ImageWorks Proprietary – Confidential Information

Before bringing patient to machine…

Confirm that unit is in Pan mode and sensor is in pan position.In imaging software, make sure you are ready to acquire a panoramic x-ray with the correct device selected.Initiate image acquisition process in imaging software.Dialogue box appears.

Click on the “NEW” button to capture a new X-Ray scan.

3. Pan - Prep Unit for Patient

Page 8: Panoura 18S with Ceph Operator Training Quick Guide

8ImageWorks Proprietary – Confidential Information

Ensure that the “Radiation Mode Setup” dialog shown below is displayed.

Click on “Panorama” Tab.

Click on Auto Panoramic.

Press Ok.

Ready for exposure screen will appear.

3. Pan - Prep Unit for Patient

Page 9: Panoura 18S with Ceph Operator Training Quick Guide

9ImageWorks Proprietary – Confidential Information

3. Pan - Prep Unit for Patient

Open the head supports before the patient enters the machine

Move head rest to an approximate position of patient height by pressing Up/Down switch on the console.

Have patient step into machine and rest chin on chinrest.

Recommend patient cross arms and hold stabilizing bars. This drops the shoulders.

Have patient step both feet forward about a foot from the post (holding on to stabilizing bars for support).

Bite bitestick with upper and lower teeth in grooves (place sleeve over bitestick).

Page 10: Panoura 18S with Ceph Operator Training Quick Guide

10ImageWorks Proprietary – Confidential Information

4. Pan - Patient PositioningHave the patient remove jewelry, eyeglasses, etc and place the apron on the patient.

Press Beam button. Unit will reposition and laser lights will come on.

Close head supports around patient’s head

Utilize the mirror to position the patient. There are 3 beams that must be aligned:

1) Medial/midsagittal ; 2) Frankfurt plane/Oculoauricular; 3) Anteroposterior Plan/Canine tooth.

Page 11: Panoura 18S with Ceph Operator Training Quick Guide

11ImageWorks Proprietary – Confidential Information

4. Pan - Patient Positioning

1. Medial/midsagittal line should run through the middle point between the left and right incisors and the nose ridge

Page 12: Panoura 18S with Ceph Operator Training Quick Guide

12ImageWorks Proprietary – Confidential Information

4. Pan - Patient Positioning

2. Frankfurt plane should run from base of eye orbitals to lower portion of ear opening.

If chin too high – flat occlusal plane. Pan will look like a frown

If chin too low – pan will look like smile

Page 13: Panoura 18S with Ceph Operator Training Quick Guide

13ImageWorks Proprietary – Confidential Information

4. Pan - Patient Positioning

3) Anteroposterior Plan/Canine tooth line should align with mandibular third tooth (patient’s left canine tooth). To assist in showing the tooth, patient can be asked to smile. However, have patient close lips before scan.

Page 14: Panoura 18S with Ceph Operator Training Quick Guide

14ImageWorks Proprietary – Confidential Information

Press RESET on the console Turn beams off (if still on, BEAM button can be pressed)

Confirm ready for exposure screen still appears.If possible the operator should stand where they can observe both the patient in the machine and the computer screen.

5. Pan - Initiate Scan

Note: Unit will not scan when laser beams are on. Either wait until beam go off or press BEAM button to turn them

off.

Page 15: Panoura 18S with Ceph Operator Training Quick Guide

15ImageWorks Proprietary – Confidential Information

Confirm Ready LED is solid green (not blinking)

Instruct the patient to swallow, place the tongue on the roof of the mouth, and hold still.

Press and hold switch during entirety of scan.

5. Pan - Initiate Scan

Note: If switch is released before exposure is complete, an ER08

warning is displayed. These error codes can be cleared by pressing RESET on panel.

Page 16: Panoura 18S with Ceph Operator Training Quick Guide

16ImageWorks Proprietary – Confidential Information

As scan is taken, preview screen will appear.

When preview screen shows completed image, system will process and contruct the image.

During this time, patient can step out of machine.

5. Pan - Initiate Scan

Page 17: Panoura 18S with Ceph Operator Training Quick Guide

17ImageWorks Proprietary – Confidential Information

Acquired image appears in window.

Click OK to transfer the image to the imaging software where it is permanently saved.

5. Pan - Initiate Scan

Page 18: Panoura 18S with Ceph Operator Training Quick Guide

18ImageWorks Proprietary – Confidential Information

Before bringing patient to machine…

Confirm that unit is in Ceph mode and sensor is in ceph position.In imaging software, make sure you are ready to acquire a ceph x-ray with the correct device selected.Initiate image acquisition process in imaging software.Dialogue box appears.

Then click on the “NEW” button to capture a new X-Ray scan.

6. Ceph - Prep Unit for Patient

Page 19: Panoura 18S with Ceph Operator Training Quick Guide

19ImageWorks Proprietary – Confidential Information

Ensure that the “Radiation Mode Setup” dialog shown below is displayed.

Click on (or verify set to) “Cephalo” Tab.

Click on desired sub mode (LA or PA).

Press Ok.

Ready for exposure screen will appear.

6. Ceph - Prep Unit for Patient

Page 20: Panoura 18S with Ceph Operator Training Quick Guide

20ImageWorks Proprietary – Confidential Information

6. Ceph - Prep Unit for Patient

Move the ear rod to an approximate position of patient height by pressing Up/Down switch on the console.

Rotate head piece depending on PA or LA.

Ear rods should match the heights of the patient’s ear canal (porus acusticus).

Spread the ear rods.

Page 21: Panoura 18S with Ceph Operator Training Quick Guide

21ImageWorks Proprietary – Confidential Information

7. Ceph - Patient Positioning

Have the patient remove jewelry, glasses, etc. and place the apron on the patient.If taking posteroanterior (PA), remove nose support.Insert ear rod on one side and close the ear rod.Gently insert the ear rod on the other side. Make sure the patient’s Frankfurt plane is positioned horizontally.If taking lateral (LA), then adjust the height of the nose support at the center between the patient’s eyebrows and gently place it to fix.

Page 22: Panoura 18S with Ceph Operator Training Quick Guide

22ImageWorks Proprietary – Confidential Information

Press RESET on the console

Confirm ready for exposure screen still appears.

If possible the operator should stand where they can observe both the patient in the machine and the computer screen.

7. Ceph - Initiate Scan

Note: Unit will not scan when laser beams are on. Either wait until beam go off or press BEAM button to turn them

off.

Page 23: Panoura 18S with Ceph Operator Training Quick Guide

23ImageWorks Proprietary – Confidential Information

Confirm Ready LED is solid green (not blinking)

Instruct the patient to swallow, place the tongue on the roof of the mouth, and hold still.

Press and hold switch during entirety of scan.

8. Ceph - Initiate Scan

Note: If switch is released before exposure is complete, an ER08

warning is displayed. These error codes can be cleared by pressing RESET on panel.

Page 24: Panoura 18S with Ceph Operator Training Quick Guide

24ImageWorks Proprietary – Confidential Information

As scan is taken, preview screen will appear.

When preview screen shows completed image, system will process and contruct the image.

During this time, patient can step out of machine.

8. Ceph - Initiate Scan

Page 25: Panoura 18S with Ceph Operator Training Quick Guide

25ImageWorks Proprietary – Confidential Information

Acquired image appears in window.

Open the ear rods and guide the patient out of the machine.

Click OK to transfer the image to the imaging software.

8. Ceph - Initiate Scan

Note: Default setting

displays head facing to the right.

Page 26: Panoura 18S with Ceph Operator Training Quick Guide

26ImageWorks Proprietary – Confidential Information

To change machine from Pan to Ceph or from Ceph to Pan, the sensor must be physically moved, and the unit switched to the new mode.

Warning: Sensor system is very robust and the fastening system has multiple support levels to assure

it is secure. However, the sensor can be damaged if dropped, so please use care in moving the sensor.

Hold sensor by handleTurn know to “Unlock” positionPull sensor off of railsMove sensor to new position, and slide onto railsTurn knob to lock into position

9. Changing Mode

Page 27: Panoura 18S with Ceph Operator Training Quick Guide

27ImageWorks Proprietary – Confidential Information

Press MODE on console.

Toggle to the desired mode (Pan, Ceph LA, Ceph PA).

Press RESET.

Sensor will turn on and system is ready to use in the new mode.

9. Changing Mode

Page 28: Panoura 18S with Ceph Operator Training Quick Guide

28ImageWorks Proprietary – Confidential Information

These instructions serve as supplemental guidance of the Panoura. They are not meant to replace the

manual, but rather serve to complement it.

If there are any questions, please do not hesitate to contact us

914-592-6100

Summary