ongoing standardisation of testing for blood-borne pathogens through an internet-based application
DESCRIPTION
Ongoing Standardisation of Testing for Blood-Borne Pathogens through an Internet-based Application. Elizabeth M. Dax for the staff of the National Serology Reference Laboratory. NRL’s Activities. Evaluations of Test Kits. Performance Monitoring Methods: QC - PowerPoint PPT PresentationTRANSCRIPT
Ongoing Standardisation of Testing for Blood-Borne Pathogens through an Internet-based ApplicationElizabeth M. Dax for the staff of the National Serology Reference Laboratory
NRL’s Activities
Evaluations of Test Kits
Quality Management
Performance Monitoring Methods: QC
Monitoring Specificity
Batch
Delivers reports and programsto generate quality in laboratories
EQAS
Reference and QA sample testing
Testing strategy development*Education
I-Collaborations
Sample characterisation
Panel Distribution Panel Distribution Panel Distribution
Sourcing samples Sourcing samples Sourcing samples
Sample selection
Sample manufacture
Results collection
Results analysis
Reports
KEYAcroMetrix
HealthMetrx
NRL
QC EQASDigital PT
EDCNet
Internet-based QC monitoring system
Laboratories submit data into ‘live’ datagraphically displays data in chartsinstantaneous and confidentialInter- and -intralaboratory variationmonitors performance of instruments, reagents and operators
Facilitating a QC programme in real time
NRL application (https://www.nrlqa.net).
* 2 Plasma samples
Data entered in 2005 for Selected Assays
0
5000
10000
15000
20000
PanBio EBVVCA IgM
ELISA
Murex anti-HCV (version
4.0)
RocheCOBASAmplicor
Chlamydia
Chiron TMAAssays
AbbottAxSYMHBsAg
Version 2MEIA
AbbottPRISMHBsAgChLIA
Assay type
No.
of d
ata
entr
ies
01020304050607080
Number of entries Number of laboratories Number of batches of reagent
Monitoring Precision & AccuracyS/
Co
02468
101214
Mean
Mean+2SD
Mean-2SD
A B C D E F G H I J
Performance of QC sample hivspy004
06/2002 - 10/2003 (7 Laboratories)Batch N mean SD -2SD +2SD Min Max CV
011701 230 17.75 2.25 13.25 22.25 7.65 29.84 12.68
203159 793 17.34 2.66 12.02 22.66 0.05 23.34 15.34
207123 1288 17.89 2.56 12.77 23.01 0.13 24.24 14.31
208146 1825 17.84 2.55 12.74 22.94 0.39 28.00 14.29
210096 16 18.60 3.67 11.26 25.94 7.29 22.85 19.73
210509 933 18.10 3.08 11.94 24.26 3.26 26.15 17.02
TOTAL 5085 17.82 2.67 12.48 23.16 0.05 29.84 14.98
Precision - CV (%)0 5 10 15 20 25 30
Acc
urac
y - B
ias
(%)
-30
-20
-10
0
10
20
30
instruments testing HCV instruments testing HBsAg instruments testing HIV instruments testing HTLV
Accuracy and Precision of the Abbott PRISM
Bias - difference between mean ratio for individual laboratory and mean ratio for all laboratories, expressed as % of mean ratio for all laboratories
Variation by Laboratory
1 2 3 4 5 6 7 Laboratory
25201510 5 0
S/C
o
Variation by Batch
Ongoing monitoring of Changes in Performance