hplc column troubleshooting - agilent€¦ · 4 soc 6 c 12-oc j.c. 268(1983) 1 j.c. 111(1975) 149...

64
HPLC Column Troubleshooting Page 1 Gulf Coast Conference 2014 John Palmer

Upload: others

Post on 05-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

HPLC Column Troubleshooting

Page 1

Gulf Coast Conference 2014

John Palmer

Page 2: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Page 2

Pump

Injector/Autosampler

Column

Detector

Data System/Integrator

HPLC System Components

Problems Can Be Related to All Components in the System

Page 3: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

There Can Be Many Causes Start Asking Questions That Lead to The Cause

•1st Did System Suitability or the Sample Fail?

•2nd Review Method for Compliance - Is The Procedure Being Followed Properly?

- Are Instrument Settings Correct?

•3rd Ask More Questions! - When Did the System Last Function Properly?

- Has Anything Been Changed?

•4th Review ALL parameters! - The Obvious Is Not Always the Cause

- Was There More Than One Change?

Page 3

Page 4: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Page 4

Categories of Column and System Problems

A. Pressure

B. Peak shape

C. Retention

Page 5: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Page 5

Column Observations Potential Problems

High pressure - Plugged frit

- Column contamination

- Plugged packing

Low Pressure - Leak

- Flow Incorrect

Pressure Issues

Page 6: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Determining the Cause and Correcting High Back Pressure

Page 6

• Check pressure with/without column

• Remove Guard

• Remove Column

• If Column pressure is high:

• Back flush column – Clear “dirty” frit surface • • Wash column – high molecular weight/adsorbed compounds – precipitate from sample or buffer

• Replace Column

Change frit – Clear plugged frit PREVENT THIS!

Page 7: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Peak Splitting Caused By Disrupted Sample Path

Page 7

Split or Double Peaks

Normal Double

Peaks

Tip: Similar Effect Can be Caused by Partially Plugged Frit

•Flow Path Disrupted by Void

•Sample Allowed to Follow Different Paths

Through Column

•Poorly Packed Bed Settles in Use

•High pH Dissolves Silica

Page 8: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Split Peaks from Injection Solvent Effects

September 28, 2014Month

##, 200X

Group/Presentation Title

Agilent Restricted

Page 8

Column: StableBond SB-C8, 4.6 x 150 mm, 5 mm Mobile Phase: 82% H2O : 18% ACN

Injection Volume: 30 mL Sample: 1. Caffeine 2. Salicylamide

A. Injection Solvent

100% Acetonitrile

B. Injection Solvent

Mobile Phase

Tip: Injecting in a solvent stronger than the mobile phase can cause peak shape

problems such as peak splitting or broadening

Trick: Keep Organic Concentration in Sample Solvent < Mobile Phase

0 10Time (min)

1

2

0 10Time (min)

1

2

Page 9: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Split Peaks from Column Contamination

Page 9

0 5 10 15

1

3

4

2

Time (min)0 5 10 15

1

3

4

2

Time (min)0 5 10 15

1

3

4

2

Time (min)

Column: StableBond SB-C8, 4.6 x 150 mm, 5 mm Mobile Phase: 60% 25 mM Na2HPO4, pH 3.0 : 40% MeOH Flow Rate: 1.0 mL/min

Temperature: 35°C Detection: UV 254 nm Sample: Filtered OTC Cold Medication: 1. Pseudoephedrine 2. APAP 3. Unknown 4. Chlorpheniramine

Injection 1 Injection 30 Injection 1

After Column Wash

with 100% ACN

Tip: Column washing eliminates the peak splitting, which resulted from a contaminant on the column

How could this be prevented? (Guard Column, SPE clean up of samples, Periodic column wash)

Page 10: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Peak Tailing - Column Contamination

Page 10

0.0 2.5 5.0

2

41

3

Time (min)

2

4

1

3

0.0 2.5 5.0

Time (min)

2

41

3

0.0 2.5 5.0

Time (min)

Column: StableBond SB-C8, 4.6 x 250 mm, 5mm Mobile Phase: 20% H2O : 80% MeOH Flow Rate: 1.0 mL/min

Temperature: R.T. Detection: UV 254 nm Sample: 1. Uracil 2. Phenol 3. 4-Chloronitrobenzene 4. Toluene

Plates TF

1. 7629 2.08

2. 12043 1.64

3. 13727 1.69

4 13355 1.32

Plates TF

1. 7906 1.43

2. 12443 1.21

3. 17999 1.19

4 17098 1.25

Plates TF

1. 7448 1.06

2. 12237 1.21

3. 15366 1.11

4 19067 1.17

QC test forward

direction QC test reverse direction

QC test after cleaning

100% IPA, 35°C

Tip: Quick Test to Determine if Column is Dirty or Damaged

Trick: Reverse Column and Run Sample –If Improved, Possible Cleaning Will Help

- No improvement - Column Damaged and Needs to be Replaced

Page 11: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Column Cleaning

Page 11

Use at least 25 mL of each solvent for long analytical columns

Flush with stronger solvents than

your mobile phase.

Reversed-Phase Solvent Choices in Order of Increasing Strength

• Mobile phase without buffer salts

• 100% Methanol

• 100% Acetonitrile

• 75% Acetonitrile:25%

Isopropanol

• 100% Isopropanol

• 100% Methylene Chloride*

• 100% Hexane*

Must Reverse

to

Re-Equilibrate

This Is Time Consuming

Often Performed Offline

Page 12: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Changing a Frit May Not Be a Good Idea May not be possible with new generation columns

May damage high performance columns

Page 12

Column

Inlet Frit

Compression

Ferrule

Column Body

Female End Fitting

Male End Fitting

Wear gloves

Do not allow bed to dry

Do not touch the column -

body heat will extrude packing

Do not overtighten

Tip: Prevention is a Much Better Idea!

Page 13: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

The Trick: Prevention Techniques - A Better Choice!

Page 13

• Filter samples

• Filter buffered mobile phases

• Use column protection

- In-line filters

- Guard columns

• Sample clean-up (i.e. SPE)

• Appropriate column flushing

Easy

Not As Easy

Page 14: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Inexpensive Filters Prevent Column Frit Plugging

Page 14

Regenerated Cellulose (RC) Recommended •Universal hydrophilic membrane, compatible

with most solvents - aqueous and organic

•High purity, extremely low extractables and

binding

•More Uniform Surface

•Different than Other Cellulose Filters!!

In-line Filters Easy to Use and replace

Frits Available in 0.2,0.5 and 2.0µ Porosity

Much Less expensive than a Column

Easier and Faster to Replace than a Column Frit

Page 15: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Mini-UniPrep Filtering Vials

•Pre-assembled filtration devices for removing particulate matter from samples.

•Single disposable unit can replace the combination of syringe filters, syringes, auto-sampler vials, transfer containers, septa and caps.

•Quick, environmentally conservative way to filter samples prior to HPLC analysis.

Page 15 Manufactured by Whatman, a division of GE Healthcare

Page 16: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Separation Conditions and the Effect on Retention*

16

Flow Rate +/- 1% +/- 1% Tr

Temp +/- 1 deg C +/- 1 to 2% Tr

%Organic +/- 1% +/- 5 to 10% Tr

pH +/- 0.01% +/- 0 to 1% Tr

*excerpted from “Troubleshooting HPLC Systems”, J. W. Dolan and L. R. Snyder, p 442.

Page 17: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Mobile Phase: Aqueous Component Experimental Variables That Impact Resolution

Column

Mobile Phase

• Aqueous Component - Importance of Buffers

- Considerations for Buffer Selection

- Buffer pH

- Buffer Concentration

• Organic Component

Sample

Instrument

Gradient Separations

September 28, 2014

CI0126C

17

Your opportunity to improve

robustness and ruggedness

Page 18: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Mobile Phase pH and pH Buffers Why Are These So Important in HPLC?

•pH Effects Ionization

- Silica Surface of Column

- Sample Components of Interest

• Buffers

- Resist Changes in pH and Maintain Retention

- Improve Peak Shape for Ionizable Compounds

• Effects Column Life

- Low pH strips Bonded Phase

- High pH Dissolves Silica

Page 18

Page 19: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Why Worry About pH? pH, pKa and Weak Bases

Page 19

At pH 9 – the sample exists as protonated and unprotonated diphenhydramine

in a ratio of 1:1. Peak shape can be poor.

At pH 10 – 91% of the sample exists as unprotonated diphenhydramine.

At pH 8 – 91% of the sample exists as protonated diphenhydramine.

Ka = [R3N][H+]

[R3NH+]

Ka = 1 x 10-9

pKa = 9

R3NH+ R3N + H+

CHOCH CH N2 2

CH3

CH3

+

HCH

3

CH3

CHOCH CH N2 2 + H+

Page 20: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Why Worry About pH? pH, pKa and Weak Acids

Page 20

At pH 4.2 – the sample exists as benzoic acid and the benzoate ion in a ratio

of 1:1. Peak shape can be poor

At pH 5.2 – 91% of the sample exists as the benzoate ion. RP retention decreases.

At pH 3.2 – 91% of the sample exists as benzoic acid. RP retention increases.

RCOOH RCOO- + H+

Ka = 6.4 x 10-5

pKa = 4.2

Ka = [RCOO-][H+]

[RCOOH]

+ H+

COOH COO _

Page 21: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

pH vs. Selectivity for Acids and Bases

Page 21

5

SCD

+

+

2 +

1

1.5

1.0

0.5

0.0

-0.5

log

3 4 5 6 7 8 pH

A B C

2

4

5

3

3 5 7 9

ELUENT pH

40

30

20

10

10

Ret

entio

n

+++

3

17,12 - OC

UDC

SOC4

6

C12-OC

J.C. 268(1983) 1

J.C. 111(1975) 149

Column: Nucleosil-C18

Mobile Phase: 45% ACN/55% phosphate buffer

Sample: Bile Acids

Column: mBondapak-C18

Mobile Phase: 60% 25 mM phosphate buffer

40% Methanol

1. Salicylic acid

2. Phenobarbital

3. Phenacetin

4. Nicotine

5. Methamphetamine

• Retention and selectivity can change dramatically when pH is changed.

Page 22: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Why Use Buffered Mobile Phases?

September 28, 2014

CI0126C

22

A = pH 7.0

water A = pH 7.0, 25 mM phosphate

buffer

Column: ZORBAX Rapid Resolution Eclipse XDB-C8, 4.6 x 75 mm, 3.5 µm Mobile Phase: 44% A : 56% methanol

Flow Rate: 1.0 mL/min Temperature: 25°C Detection: UV 250 nm

Sample: 1. ketoprofen 2. ethyl paraben 3. hydrocortisone 4. fenoprofen 5. propyl paraben 6. propranolol 7. ibuprofen

• Buffered mobile phases enhance retention, resolution, and peak shape.

0 1 2 3 4 5Time (min)

1

2

3

4

6

7

5

0 1 2 3 4

1, 4, 6, 7

2

3

5

Time (min)

Page 23: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Considerations For Buffer Selection

Buffer Type

• Inorganic vs. organic buffers—choice can affect resolution, column lifetime and MS compatibility

Buffer pH

• Select buffer based on desired pH and optimum buffer pH range.

• Measure pH of buffer solution before mixing with organic modifier.

• Compare resolution at desired pH ± 0.1–0.2 pH units. Buffer Concentration and Ionic Strength

• Start at 20 – 25 mM.

• Prepare buffer according to accepted procedures.

• Avoid overshoot and readjustment when setting pH.

• Compare resolution at desired buffer concentration ± 5–10 mM.

September 28, 2014

CI0126C

23

Page 24: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Changes in Buffer Concentration Can Affect Retention, Peak Width and Peak Shape

September 28, 2014

CI0126C

24

USP Tf

(5%)

1. 1.62

2. 1.65

3. 1.63

4. 1.77

5. 1.83

6. 1.12

0.0 0.02.5 2.55.0 5.0

6

6

5

54

4 3

3

2

2

1

1

7.5

Time (min) Time (min)

10 mM Phosphate 25 mM Phosphate

Column: ZORBAX Eclipse XDB-C8, 4.6 x 150 mm, 5 µm

Mobile Phase: 40% phosphate buffer (pH 7.0) : 60% ACN Flow Rate: 1.5 mL/min Temperature:

40°C

Sample: Tricyclic Antidepressants, 1. Desipramine 2. Nortriptyline 3. Doxepin 4. Imipramine 5. Amitriptyline 6.

Trimipramine

USP Tf

(5%)

1. 1.41

2. 1.50

3. 1.33

4. 1.39

5. 1.36

6. 1.00

Page 25: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Effect of pH on Peak Shape at or Near the Sample pKa

Page 25

0 1 2 3 4 5 6 7 8 9 10 0 1 2 3 4 5 6 7 8 9 10

Time (min) Time (min)

Column: ZORBAX SB-C8 4.6 x 150 mm, 5 mm Mobile Phase: 40% 5 mM KH2PO4: 60% ACN

Flow Rate: 1.0 mL/min. Temperature: RT

pH 4.4 pH 3.0

Ibuprofen

pKa = 4.4

CH3CHCOOH

CH2CH(CH3)2

• Inconsistent and tailing peaks may occur when operating close to an analyte

pKa and should be avoided.

Page 26: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Lot-to-Lot Selectivity Change Related to pH Choice

September 28, 2014Month

##, 200X

Group/Presentation Title

Agilent Restricted

Page 26

0 2 4 6 8 10 12 14 16 18

Time (min)

2-Base

3

4-Base

1

0 2 4 6 8 10 12 14 16 18

Time (min)

2

3

4

1

• pH 4.5 shows selectivity change from lot-to-lot for basic compounds

• pH 3.0 shows no selectivity change from lot-to-lot

• Indication of poorly controlled ionization

pH 4.5 - Lot 1 pH 3.0 - Lot 1

pH 4.5 - Lot 2 pH 3.0 - Lot 2

0 2 4 6 8 10 12 14 16 18

Time (min)

2-Base

3

4-Base

1

0 2 4 6 8 10 12 14 16 18

Time (min)

2

3

4

1

Page 27: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Test for pH Robustness

September 28, 2014

CI0126C

27

1 2

3

4

6 7

5

0 1 2 3 4 5 6

Time (min)

0 1 2 3 4 5 6

1 2

3

4

6

7

5

Time (min)

pH 7.00

pH 7.25

Column: ZORBAX Rapid Resolution Eclipse XDB-C8, 4.6 x 75 mm, 3.5 µm

Mobile Phase: 44% 25 mM phosphate, pH 7.00 : 56% methanol Flow Rate: 1.0 mL/min Temperature: 25°C

Detection: UV 250 nm

Sample: 1. ketoprofen 2. ethyl paraben 3. Hydrocortisone 4. fenoprofen 5. propyl paraben 6. Propranolol 7.

ibuprofen

• The resolution of ionizable compounds can change

markedly with pH changes—even as small as 0.05–

0.25 pH units.

Page 28: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Don’t Forget - Match Column to pH of Mobile Phase for Maximum Column Lifetime low pH and high temperature (pH 0.8, 90°C)

Page 28 000023P1.PPT

Purge Solvent:

50% methanol/water with

1.0% TFA

Solute: Toluene

Kirkland, J.J. and J.W. Henderson, Journal of Chromatographic Science, 32 (1994) 473-480.

Page 29: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Don’t Forget - Match Column to pH of Mobile Phase for Maximum Column Lifetime High pH and Room Temperature (pH 11 RT)

Page 29

Tip: Use Columns Designed for chosen pH

Mobile Phase: 50%ACN: 50% Water : 0.2% TEA

(~ pH 11)

After 30 injections

Initial

Page 30: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Mobile Phase: Organic Component

Column

Mobile Phase • Aqueous Component

• Organic Component

- % Organic Modifier

Sample

Instrument

Gradient Separations

September 28, 2014

CI0126C

30

Page 31: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Even a Small Change in % Organic Modifier Can Change Resolution

September 28, 2014

CI0126C

31

12

34

5

6

0 1 2 3 4

Time (min)

0 1 2 3 4

12

34

5

6

Time (min)

1 2

34

5

6

0 1 2 3Time (min)

4

• Verify that resolution doesn’t change

significantly around desired

conditions (for example, %B ± 1–

2%).

59% MeOH

57.5% MeOH

56% MeOH

Column: ZORBAX Rapid Resolution Eclipse

XDB-C8

4.6 x 75 mm, 3.5 µm

Mobile Phase: A: 25 mM phosphate, pH

7.00 (10 mM TEA)

B: methanol (10 mM TEA)

Flow Rate: 1.0 mL/min

Temperature: 25°C controlled

Injection: 5 mL

Detection: 275 nm

Sample: 1. ketoprofen 2. ethyl paraben 3.

hydrocortisone

4. fenoprofen 5. propyl paraben 6.

propranolol

Rs (5,6) =

2.5

Rs (5,6) =

1.7

Rs (5,6) =

1.2

Page 32: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

The Sample Experimental Variables That Impact Resolution

Column

Mobile Phase

Sample

• Injection volume

• Sample solvent

strength

Instrument

Gradient Separations

September 28, 2014

CI0126C

32

Page 33: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Injection Volume and Sample Solvent Strength

Injection Volume

• Lack of ruggedness typically seen

– when Vinj is increased to improve signal-to-noise (S/N) ratio, or,

– when column size is decreased.

• Use minimum Vinj for required repeatability and limit of detection.

• Compare resolution, peak shape and repeatability at 0.2X, 1X and 2–5X Vinj.

Sample Solvent Strength

• Match % organic modifier in mobile phase (or weaker).

• If stronger sample solvent needed (solubility, stability), keep Vinj to

minimum.

• Compare resolution, peak shape and width at desired solvent strength ±50% relative.

September 28, 2014

CI0126C

33

Page 34: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Test For Injection Volume Robustness

September 28, 2014

CI0126C

34

0 1 2 3 4 5

0 1 2 3 4 5

1 2

3

4 5

T im e (m in)

6 7

1 2

3

45

6

7

1

2

3

45

6

7

0 1 2 3 4 5

1 µL

5 µL

Column: ZORBAX Rapid Resolution Eclipse XDB-C8 4.6 x 75 mm, 3.5 µm

Mobile Phase: 44% 25 mM phosphate,

pH 7.00

56% methanol

Flow Rate: 1.0 mL/min

Temperature: 25°C

Detection: UV 250 nm

Sample: 1. ketoprofen 2. ethyl paraben 3. hydrocortisone 4. fenoprofen 5. propyl paraben 6. propranolol 7. ibuprofen

Rs(6,7) =

2.5

Rs(6,7) =

2.1

Rs(6,7) =

1.6

10 µL

• Varying injection volume can sometimes reveal lack of robustness for resolution and peak shape.

Page 35: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Peak Tailing, Broadening and Loss of Efficiency

Page 35

May be caused by:

• Column “secondary

interactions”

• Column contamination

• Column aging

• Column loading

• Extra-column effects

Page 36: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Peak Shape: Tailing Peaks

Page 36

Normal Tailing

Normal Tailing

Symmetry > 1.2

All Peaks Tail:

Extra-Column Effects.

Build up of Contamination on Column

Inlet.

Bad Column.

Causes

Some Peaks Tail:

Secondary - Retention Effects.

Residual Silanol Interactions.

Small Peak Eluting on Tail of Larger Peak.

Page 37: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Peak Tailing

Identifying Column “Secondary Interactions”

Page 37

Tip: Mobile phase modifier (TEA) competes with Sample for surface ion exchange

sites at mid-range pH values

Column: Alkyl-C8, 4.6 x 150 mm, 5mm Mobile Phase: 85% 25 mM Na2HPO4 pH 7.0 : 15% ACN Flow Rate: 1.0 mL/min

Temperature: 35°C Sample: 1. Phenylpropanolamine 2. Ephedrine 3. Amphetamine 4. Methamphetamine 5. Phenteramine

No TEA USP TF (5%)

1. 1.29

2. 1.91

3. 1.63

4. 2.35

5. 1.57

10 mM TEA USP TF (5%)

1. 1.19

2. 1.18

3. 1.20

4. 1.26

5. 1.14

TIme (min) Time (min)

0.0 2.5 5.0

54

32

1

5

4

3

2

1

0.0 2.5 5.0

Page 38: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Peak Tailing

Low pH Minimizes “Secondary Interactions” for Amines

Page 38

Column: Alkyl-C8, 4.6 x 150 mm, 5mm Mobile Phase: 85% 25 mM Na2HPO4 : 15% ACN Flow Rate: 1.0 mL/min

Temperature: 35°C Sample: 1. Phenylpropanolamine 2. Ephedrine 3. Amphetamine 4. Methamphetamine 5. Phenteramine

pH 3.0

USP TF (5%)

4. 1.33

pH 7.0

USP TF (5%)

4. 2.35

Tip: Reducing mobile phase pH reduces interactions with silanols and peak tailing.

Time (min)

0.0 2.5 5.0

54

32

1

5

4

32

1

Time (min)

0.0 2.5 5.0

Page 39: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Peak Tailing

High pH Eliminates “Secondary Interactions” for Amines

Peak Shape and Retention of this sample of basic compounds improves at high pH where column has high IEX activity. Why?

Page 39

0 5

1

2,3

4

5

Time (min)

7

6

0 5 10

1

2

3

4

Time (min)

6

5

7

Column: ZORBAX Extend-C18, 4.6 x 150 mm, 5 m m Mobile Phase: See Below Flow Rate: 1.0 mL/min Temperature: RT

Detection: UV 254 nm

Sample: 1. Maleate 2. Scopolamine 3. Pseudoephedrine 4. Doxylamine 5. Chlorpheniramine 6. Triprolidine 7. Diphenhydramine

pH 7

30% 20 mM Na2HPO4

70% MeOH

pH 11

30% 20 mM TEA

70% MeOH

tR = 8.5 tR = 11.4

Page 40: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

40

Peak Shape: Broad Peaks

All Peaks Broadened:

• Loss of Column Efficiency.

• Column Void.

• Large Injection Volume.

Some Peaks Broadened:

• Late Elution from Previous Sample

(Ghost Peak).

- High Molecular Weight.

- Sample - Protein or Polymer.

Page 40

Page 41: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Unknown “Phantom” Peaks

Page 41

Time (min)

3

1

2

0 5 10 15

Column: Extend-C18, 4.6 x 150 mm, 5 mm Mobile Phase: 40% 10 mM TEA, pH 11 : 60% MeOH Flow Rate: 1.0 mL/min

Temperature: R.T. Detection: UV 254 Sample: 1. Maleate 2. Pseudoephedrine 3. Chlorpheniramine

Plates

1. 5922

2. 9879

3. 779

Tip: The extremely low plates for moderately retained peaks are an indication of a

very late eluting peak from a preceding run.

Time (min)

1

0 5 10

Sample 1: Chlorpheniramine maleate

Peak 1: maleate

Sample 2 : Chlorpheniramine

maleate

and Pseudoephedrine

Peak 1: maleate

Peak 2: pseudoephedrine

Peak 3: chlorpheniramine (from 1st

injection)

“Phantom” peak from

first injection

Page 42: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Changes in Retention Can Be Chemical or Physical

Page 42

May be caused by:

• Column aging

• Column contamination

• Insufficient equilibration

• Poor column/mobile phase combination

• Change in mobile phase

• Change in flow rate

• Different Gradient Delay Volumes

Page 43: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Column Aging/Equilibration Causes Retention/Selectivity Changes

September 28, 2014Month

##, 200X

Group/Presentation Title

Agilent Restricted

Page 43

Column 1 - After Cleaning

with 1% H3PO4

/Equilibration

• The primary analyte was sensitive to mobile phase aging/

conditioning of the column

• The peak shape was a secondary issue (metal chelating

compound) resolved by “de-activating” the active metal

contamination

Column 1 - Next Day Column 1 - Initial

0 3 5 9 12 15Time (min)

2

1

0 3 5 9 12 15Time (min)

2

1

Page 44: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Column Temperature Adequate Temperature Control is Essential

Laboratory temperatures can vary by ±5°C or more. Column temperature changes affect resolution and repeatability. Useful tool for changing selectivity, retention and efficiency when developing separations. Important parameter to control during method development and validation. Compare resolution, peak width and peak shape at desired temperature ±5°C.

September 28, 2014

CI0126C

44

Page 45: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Small Temperature Changes Can Cause Dramatic Changes in Resolution

September 28, 2014

CI0126C

45

12

34 6

5

0 0.5 1 1.5 2 2.5 3 3.5

Time (min)

0 0.5 1 1.5 2 2.5 3 3.5

12

34 6

5

Time (min)

0 0.5 1 1.5 2 2.5 3 3.5

1 2

34 6

5

Time (min)

Column: ZORBAX Rapid Resolution Eclipse XDB-C8 4.6 x 75 mm, 3.5 µm

Mobile Phase: Isocratic, 28%B : 72%A

A: 5/95 methanol/pH 7.00 buffer 25 mM, 10 mM TEA

B: 80/20 methanol/pH 7.00 buffer 25 mM, 10 mM TEA

Flow Rate: 1.0 mL/min.

Temperature: See Figure

Injection: 5 mL

Detection: 275 nm

Sample: 1. ketoprofen 2. ethyl paraben 3. hydrocortisone 4. fenoprofen 5. propyl paraben 6. propranolol

20°C

25°C

30°C

• Column temperature

control will produce the

most consistent

results.

Page 46: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

CASE STUDIES

Page 46

Page 47: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Example: Change in Retention/Selectivity

September 28, 2014Month

##, 200X

Group/Presentation Title

Agilent Restricted

Page 47

Unintended Mobile Phase Variation

Tip: The Source of the Problem is Often Not the Obvious Change

“I have experimented with our mobile phase, opening new bottles of all mobile phase components. When I use all fresh ingredients, the problem ceases to exist, and I have narrowed the problem to either a bad bottle of TEA or phosphoric acid. Our

problem has been solved.”

Column 1 Column 2 - Fresh

mobile phase Column 2

0 4 6Time (min)

0 2 3 4 5 6 7

Time (min)

0 4 6Time (min)

Page 48: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Case History: Problem – Selectivity Does Not Appear the Same from Column-to-Column

Details:

• 3 Columns with the Same Bonded Phase were used

• They were the same dimensions, just different particle sizes (and therefore

different lots of material)

• They were tested on the same day on the same instrument and with the

same mobile phase

• Isocratic elution; binary (two channel) pump generated mobile phase

Problem:

• The selectivity was different on each of the columns

48

Page 49: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

49

Selectivity between particle sizes of Eclipse Plus C18, 4.6 x 50 mm, using same test solutes

John W. Henderson JrGroup/Presentation

Title

Agilent Restricted Month ##, 200X

min 0 1 2 3 4 5

mAU

-5

0

5

10

15

20

25

DAD1 B, Sig=280,16 Ref=360,20 (F:\EPC18SELECTIVITY\ANALINE000015.D)

0.2

93

0.3

60

0.7

12

0.9

61

1.2

29

1.6

35

3.9

07

min 0 1 2 3 4 5

mAU

-5

0

5

10

15

20

25

DAD1 B, Sig=280,16 Ref=360,20 (F:\EPC18SELECTIVITY\3P5MICRON200001.D)

0.0

17

0.3

00

0.3

69

0.7

25

0.9

96

1.2

90

1.7

22

4.1

98

min 0 1 2 3 4 5

mAU

-5

0

5

10

15

20

25

DAD1 B, Sig=280,16 Ref=360,20 (F:\EPC18SELECTIVITY\ANALINE000019.D)

0.2

99

0.7

84

1.0

71

1.3

89

1.9

00

4.8

97

2 3

5 1 4

6

5 um, UXE01033, NEP0652003

3.5 um, UXF01309, B08111

1.8 um, UXG03882, B08021

m.p.: A: water, B: acetonitrile (60:40 A:B)

Flow : 1.5 mL/min

Temp: 25 C

Detection: UV 280nm,16, ref=360,20

Flow cell: 6mm, 5uL

Data rate: 0.2s

Inj. Vol 2 uL

N=3800

N=7100

N=9900

tr k alpha

uracil 0.3 0.00

aniline 0.78 1.60 #DIV/0!

o-toluidine 1.07 2.57 1.60

4-chloroaniline 1.39 3.63 1.42

2-naphthylamine 1.9 5.33 1.47

3,3'-dichlorobenzidine 4.9 15.33 2.88

tr k alpha

uracil 0.3 0.00

aniline 0.73 1.43 #DIV/0!

toluidine 1 2.33 1.63

chloroaniline 1.29 3.30 1.41

naphthylamine 1.72 4.73 1.43

dichlorobenzidine 4.2 13.00 2.75

tr k alpha

uracil 0.29 0.00

aniline 0.71 1.45 #DIV/0!

toluidine 0.96 2.31 1.60

chloroaniline 1.23 3.24 1.40

naphthylamine 1.63 4.62 1.43

dichlorobenzidine 3.91 12.48 2.70

Page 50: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

What Questions Should We Ask and What Should We Try to diagnose problem?

Compare more columns – see if they are different

Try another LC

Premix the mobile phase

50

Page 51: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

51

One channel premixed mobile phase shows similar α whereas previous two channels doesn’t show similar α!

John W. Henderson JrGroup/Presentation

Title

Agilent Restricted Month ##, 200X

0 1 2 3 4

mAU

-5

0

5

10

15

20

DAD1 B, Sig=280,16 Ref=360,20 (E:\ECLIPSE SELECTIVIY INVESTIGATION\ONECHANNEL00002.D)

0.2

49

0

.28

8

0.3

52

0.6

96

0.9

46

1.2

14

1.6

15

3.8

79

DAD1 B, Sig=280,16 Ref=360,20 (E:\ECLIPSE SELECTIVIY INVESTIGATION\ONECHANNEL00003.D)

0.2

88

0.3

53

0.6

95

0.9

44

1.2

13

1.6

12

3.8

73

5 um, UXE01033, NEP0652003

min 0 1 2 3 4 5 6

mAU

-5

0

5

10

15

20

25

30

DAD1 B, Sig=280,16 Ref=360,20 (E:\ECLIPSE SELECTIVIY INVESTIGATION\ONECHANNEL00005.D)

0.2

90

0.3

51

0.4

17

0.7

14

0.8

48

0.9

71

1.2

41

1.6

68

4.0

73

DAD1 B, Sig=280,16 Ref=360,20 (E:\ECLIPSE SELECTIVIY INVESTIGATION\ONECHANNEL00006.D)

0.2

87

0.3

48

0.7

13

0.9

70

1.2

41

1.6

66

4.0

72

1.8 um, UXG03882, B08021

one channel tr k alpha

uracil 0.29 0.00

aniline 0.69 1.38 #DIV/0!

o-toluidine 0.94 2.24 1.63

4-chloroaniline 1.21 3.17 1.42

2-naphthylamine 1.61 4.55 1.43

3,3'-dichlorobenzidine 3.87 12.34 2.71

one channel tr k alpha

uracil 0.29 0.00

aniline 0.71 1.45 #DIV/0!

o-toluidine 0.97 2.34 1.62

4-chloroaniline 1.24 3.28 1.40

2-naphthylamine 1.67 4.76 1.45

3,3'-dichlorobenzidine 4.07 13.03 2.74

m.p.: A: water, acetonitrile (60:40 v/v)

Flow : 1.5 mL/min

Temp: 25 C

Detection: UV 280nm,16, ref=360,20

Flow cell: 6mm, 5uL

Data rate: 0.2s

Inj. Vol 2 uL

46 bar

246 bar

Page 52: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

52

Repeat experiment of two channel isocratic mobile phase confirms different selectivity probably due to wrong composition of mobile phase being delivered

John W. Henderson JrGroup/Presentation

Title

Agilent Restricted Month ##, 200X

min 0 1 2 3 4 5 6

mAU

-5

0

5

10

15

20

DAD1 B, Sig=280,16 Ref=360,20 (E:\ECLIPSE SELECTIVIY INVESTIGATION\TWOCHANNEL00005.D) DAD1 B, Sig=280,16 Ref=360,20 (E:\ECLIPSE SELECTIVIY INVESTIGATION\TWOCHANNEL00006.D)

min 0 1 2 3 4 5 6

mAU

-5

0

5

10

15

20

25

DAD1 B, Sig=280,16 Ref=360,20 (E:\ECLIPSE SELECTIVIY INVESTIGATION\TWOCHANNEL00002.D)

0.3

00

0.3

68

0

.38

7

0.7

81

1.0

75

1.4

01

1.9

18

4.9

91

DAD1 B, Sig=280,16 Ref=360,20 (E:\ECLIPSE SELECTIVIY INVESTIGATION\TWOCHANNEL00003.D)

0.3

00

0.3

67

0

.38

7

0.7

82

1.0

74

1.4

00

1.9

16

4.9

85

5 um, UXE01033, NEP0652003

1.8 um, UXG03882, B08021

45 bar

238bar

m.p.: A: water, B: acetonitrile (60:40 A:B)

Flow : 1.5 mL/min

Temp: 25 C

Detection: UV 280nm,16, ref=360,20

Flow cell: 6mm, 5uL

Data rate: 0.2s

Inj. Vol 2 uL

two channel (2nd exp) tr k alpha

uracil 0.29 0.00

aniline 0.72 1.48 #DIV/0!

o-toluidine 0.98 2.38 1.60

4-chloroaniline 1.25 3.31 1.39

2-naphthylamine 1.67 4.76 1.44

3,3'-dichlorobenzidine 4.03 12.90 2.71

two channel (2nd exp) tr k alpha

uracil 0.3 0.00

aniline 0.78 1.60 #DIV/0!

o-toluidine 1.07 2.57 1.60

4-chloroaniline 1.4 3.67 1.43

2-naphthylamine 1.92 5.40 1.47

3,3'-dichlorobenzidine 4.98 15.60 2.89

Page 53: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Typical Case History: Lack of Efficiency on Poroshell 300

53

Agilent Publication Customer Results

Page 54: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Looked Bad – What Did We Do?

1. Customer needed to understand that the column used – 2.1

x 75mm, Poroshell 300 column---is a low volume column, and

therefore, low column volume guidelines should be followed

2. What about the instrument and sample ?

3. Turns out that instrument was quaternary pump 1100, not set

up appropriately for low volume use (difficult to optimize

completely)

4. Customer provided with details on how to achieve a better

low volume set-up, based on 1100-RRHT Kit documents

5. Customer now achieving acceptable results

54

Page 55: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Conclusions

Page 55

• High pressure (prevention better than the cure)

• Undesirable peak shape

• Changes in retention/selectivity

Often these problems are not associated with the column and

may be caused by instrument and chemistry issues.

•pH of mobile Phase

•Instrument Connections

•Detector Settings

•Metal Contamination

Start With the Correct Questions

•Find the Answers

•The Answers will Lead to Solutions

HPLC column problems are evident as

Page 56: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

APPENDIX

Page 56

Page 57: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Important Buffer Systems

September 28, 2014

CI0126C

57

Buffer Selection

Buffer Selection

Buffer pKa pH Range UV Cutoff (A > 0.5)

Trifluoroacetic acid <<2 (0.5) 1.5-2.5 210 nm (0.1%)

KH2PO4/phosphoric acid 2.12 1.1-3.1 <200 nm (0.1%)

tri-K-Citrate/hydrochloric acid 1 3.06 2.1-4.1 230 nm (10 mM)

Potassium formate/formic acid 3.8 2.8-4.8 210 nm (10 mM)

tri-K-Citrate /hydrochloric acid 2 4.7 3.7-5.7 230 nm (10 mM)

Potassium acetate/acetic acid 4.8 3.8-5.8 210 nm (10 mM)

tri-K-Citrate /hydrochloric acid 3 5.4 4.4-6.4 230 nm (10 mM)

Ammonium formate 3.8

9.2

2.8-4.8

8.2-10.2 (50 mM)

Bis-tris propane·HCl/Bis-tris propane 6.8 5.8-7.8 215 nm (10 mM)

Ammonium acetate 4.8

9.2

3.8-5.8

8.2-10.2 (50 mM)

KH2PO4/K2HPO4 7.21 6.2-8.2 <200 nm (0.1%)

Tris·HCl/Tris 8.3 7.3-9.3 205 nm (10 mM)

Bis-tris propane·HCl/Bis-tris propane 9.0 8.0-10.0 225 nm (10 mM)

Ammonium hydroxide/ammonia 9.2 8.2-10.2 200 nm (10 mM)

Borate (H3BO3/Na2B4O7·10 H2O) 9.24 8.2-10.2

Glycine·HCl/glycine 9.8 8.8-10.8

1-methylpiperidine·HCl/1-methylpiperidine 10.1 9.1-11.1 215 nm (10 mM)

Diethylamine·HCl/diethylamine 10.5 9.5-11.5

Triethylamine·HCl/triethylamine 11.0 10.0-12.0 <200 nm (10 mM)

Pyrollidine·HCl/pyrollidine 11.3 10.3-12.3

Adapted from Practical HPLC Method Development, 2nd Edition, Snyder, L.R., Kirkland, J.J. and

Glajch, J.L., page 299.

Page 58: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Separation Ruggedness Buffer Preparation

1. Dissolve salt in organic-free water in 1- or 2-L beaker. Use appropriate volume to leave room for pH

adjustment solution. Equilibrate solution to room temperature for maximum accuracy.

2. Calibrate pH meter. Use 2-level calibration and bracket desired pH. Use appropriate audit solution to

monitor statistical control (for example, potassium hydrogen tartrate, saturated solution, pH = 3.56).

3. Adjust salt solution to desired pH. Minimize amount of time electrode spends in buffer solution

(contamination). Avoid overshoot and readjustment (ionic strength differences can arise).

4. Transfer pH-adjusted buffer solution quantitatively to volumetric flask, dilute to volume, and mix.

5. Filter through 0.45 µm filter. Discard first 50 – 100 mL filtrate. Rinse solvent reservoir with small

volume of filtrate and discard. Fill reservoir with remaining filtrate or prepare premix with organic

modifier.

– Agilent Solvent Filtration Kit, 250-mL reservoir, 1000-mL flask, p/n 3150-0577

– Nylon filter membranes, 47 mm, 0.45 µm pore size, p/n 9301-0895 (not for proteins!)

September 28, 2014

CI0126C

58

Page 59: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Using Buffers Successfully Initial Column and System Equilibration

In an appropriate vessel, test highest % organic/buffer ratio to verify

that buffer will not precipitate. With stirring, add organic to buffer

first, not vice versa.

Equilibrate column with, in order:

• 100% organic modifier (if brand new)

• mobile phase minus buffer

• buffered mobile phase containing highest % organic modifier (gradient high end)

• buffered mobile phase containing lowest % organic modifier (gradient low end).

Inject standard or sample several times until RTs stable, or for

gradient methods, precede former with 1 or 2 blank gradients.

September 28, 2014

CI0126C

59

Page 60: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Using Buffers Successfully Shutdown State and Instrument Flushing

Next day use—using same buffers

• Pump mobile phase very slowly (for example, 0.01 – 0.1mL/min). When flushing column or for longer term column storage • Flush with 20/80 organic/water, then 80/20 organic/water or 100%

organic.

Instrument flushing

• Replace column with capillary tubing. Leave disconnected from detector.

• Flush pumps with water, then connect capillary tubing to detector. • Inject water 2-3 times at maximum injection volume setting. • Flush all pumps with 100% organic for long term storage.

September 28, 2014

CI0126C

60

Page 61: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Importance of pH and Buffers A Practical Example

•Why the Sample Dictates Use

•What Happens When Buffer Used Effectively

•What Happens When Buffer Ignored or Used Improperly

Page 61

Page 62: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

Importance of pH and Buffers - A Practical Example Optimized Isocratic Conditions for Cardiac Drugs

Page 62

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14

Time (min)

5

4

32

1

0 1 2 3 4 5 6Time (min)

5

4

32

1Column: StableBond SB-C18, 4.6 x 150 mm, 5 mm Mobile Phase: 45% 25 mM NaH2PO4, pH 3.0 55% MeOH Flow Rate: 2.0 mL/min.

Temperature: 35°C

Detection: UV 254 nm

Sample: Cardiac Drugs 1. Diltiazem 2. Dipyridamole 3. Nifedipine 4. Lidoflazine

5. Flunarizine

Page 63: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

I Don’t Have Time to Make Buffers or Adjust pH …

Page 63

0 5 10 15 20 25

Time (min)

Column: StableBond SB-C18 4.6 x 150 mm, 5 mm

Mobile Phase: A: 20% H2O B: 80% MeOH

Flow Rate: 1.0 mL/min.

Temperature: 35°C

UV Detection: 254 nm

Sample: Cardiac Drugs

• Buffers are critical to good retention and peak shape in many separations.

Even at very high % MeOH Most Components

Strongly Retained with Poor peak Shape Due to

IEX at Surface

Page 64: HPLC Column Troubleshooting - Agilent€¦ · 4 SOC 6 C 12-OC J.C. 268(1983) 1 J.C. 111(1975) 149 Column: Nucleosil-C18 Mobile Phase: 45% ACN/55% phosphate buffer Sample: Bile Acids

What If You Work Outside the Buffer Range?

Page 64

0 5 10 15 20 25

5

1

2

34

Time (min)

Columns: StableBond SB-C18

4.6 x 150 mm, 5 mm

Mobile Phase:A: 30% 25 mM NaH2PO4, pH 4.8 unbuffered B: 70% MeOH Flow Rate: 1.0 mL/min.

Temperature: 35°C

UV Detection: 254 nm

Sample: Cardiac Drugs 1. Diltiazem 2. Dipyridamole 3. Nifedipine 4. Lidoflazine 5. Flunarizine

Unsuitable Peak Shape