    EPOCH LT Training Presentation

    Introduction and Basic Operation

  • 8/10/2019 Epoch Lt Training 2-06

    2/58

    2

    Panametrics-NDT Flaw Detector Line

    EPOCH LT

    EPOCH 4B

    EPOCH 4PLUS

    EPOCH 4

  • 8/10/2019 Epoch Lt Training 2-06

    3/58

    3

    Instrument Summary and Key Features

    EPOCH 4PLUS Ultrasonic flexibility of the EPOCH 4 with the addition of a color liquid crystal display.Several unique software features include Dynamic Waveform Color Change, Backwall

    Echo Attenuation, Spotweld Assistant, special RF detection, and high-speed USB

    communication protocol.

    EPOCH 4 High end flaw detector with a large array of specialized software options includingencoded B-Scan. Unprecedented inspection flexibility with dual pulsers and user

    customizable filters open the door to many new applications that previous generation

    flaw detectors cannot tackle.

    EPOCH 4B Basic version of the EPOCH 4 Flaw Detector. Offers standard dual pulsers, optionalnarrow band filtering and a choice between LCD and ELD screen types. Similar option

    package to EPOCH LT.

    EPOCH LT Basic instrument with a standard negative spike pulser, broadband 20MHzreceiver, high performance LCD, high-speed USB communication port, and

    standard DAC/TVG. The lightest and smallest instrument available. Options

    include Adv. DAC/TVG, DGS/AVG, API 5UE, Low PRF, AWS D1.1/D1.5, Tunable

    Square Wave Pulser.

    Panametrics-NDT Flaw Detector Line

  • 8/10/2019 Epoch Lt Training 2-06

    4/58

    4

    Product Overview

    The EPOCH LT was launched at theNDT World Conference in September2004.

    The EPOCH LT is Olympus NDTs basicflaw detector platform providing

    customers with class-leading size andweight as well as excellent ultrasonicperformance.

    The EPOCH LT is designed to handlethe most common UT Flaw Detector

    applications with standard basic DACand TVG as well as common softwareoptions. It is fully compatible with theEPOCH 4 series instruments as well.

  • 8/10/2019 Epoch Lt Training 2-06

    5/58

    5

    Key Features

    Size and Weight Only 2.35 lbs. (1.0kg)

    Only 1.35 in. (38mm) thick

    Provides truly handheld performance

    Large Bandwidth Broadband Receiver (20 MHz at -3dB) Allows use of high frequency transducers for

    thinner materials

    Large Memory Capacity (100 W / 2,000 T)

    Option to expand this capacity to

    (500 W / 5,000 T)

    Auto-Calibration Feature

    Greatly speeds and simplifies the calibration process

  • 8/10/2019 Epoch Lt Training 2-06

    6/58

    6

    Key Features

    Nickel Metal Hydride (NiMH) Battery

    High power to weight ratio

    Internal pack design

    Time of Flight Measurement Readout

    For sec readout

    USB Communication Port

    Communication with GageView Pro Interface Program

    Real-Time Clock Saved data is automatically tagged with date/time

  • 8/10/2019 Epoch Lt Training 2-06

    7/58

    7

    Key Features

    Thickness Measurements are Taken from Raw Data

    True time of flight instrument

    Resolution NOT tied to screen range

    5 Function Soft Keys Direct access to preset parameter values

    Gasketed Case

    Seals out the environment

  • 8/10/2019 Epoch Lt Training 2-06

    8/58

    8

    Expanded Memory Capacity

    Echo to Echo Measurement

    Extended Range

    Standard range is 200in. (steel velocity)

    Expand to 400in. (steel velocity)

    Expanded Memory

    Capacity

    Standard Memory

    Capacity

    Thickness Readings 10,000 2,000

    Waveforms 500 100

    Basic Options

  • 8/10/2019 Epoch Lt Training 2-06

    9/58

    9

    EPOCH Design Philosophy

    EPOCH Flaw Detectors have been designed with the operator in mind. Theuser interface has been designed to provide Direct Access to settings.

    Examples of Direct Access Philosophy

    Split Screen ViewAllows the operator to view all instrument settings on onescreen.

    KeypadFunctionally grouped keypad allows access to all major instrumentfunctions with one or two key presses.

    Function (F) KeysUser can jump directly to various instrument settings rather thanmanual adjustment. Fully customizable values with standard Editable Parametersfeature.

    Single Hand OperationAll instrument functions accessible with ENTER key andadjustable with slewing keys.

  • 8/10/2019 Epoch Lt Training 2-06

    10/58

    10

    Physical Features

    Liquid Crystal Display

    High Speed USB Port

    AC Adaptor Input LEMO 00 Transducer

    Connectors

    LM-BF adaptors are

    available for BNC Cables

    Direct Access Function Keys

    Keypad (English or

    International Symbol)

    EPOCH LT shown with

    English Keypad and option

    rubber boot.

    LEMO 00 transducer

    connectors are the only type

    available due to connector

    size restrictions.

  • 8/10/2019 Epoch Lt Training 2-06

    11/58

    11

    Display

    Liquid Crystal Display (LCD) 1/4 VGA (320 W x 240 H Pixels)

    Superior Outdoor Visibility

    Contrast Adjustment via Keypad

    Backlight Adjustment via Keypad

  • 8/10/2019 Epoch Lt Training 2-06

    12/58

    12

    Keypad (English Version)

  • 8/10/2019 Epoch Lt Training 2-06

    13/58

    13

    Outputs

    USB Communication Port

    AC Adaptor Input

    LEMO Transducer Posts

  • 8/10/2019 Epoch Lt Training 2-06

    14/58

    14

    Battery Charging

    OPTION #1 - Charge Battery Inside Unit Plug charger into unit and AC Main Line.

    Full charge will be achieved in approximately 2-3 hours. Battery will also charge

    when unit is powered on, although at a slower rate.

    OPTION #2 - Charge Battery Externally.

    Optional External Battery Charger

    Panametrics Part Number: EPLT-EC

    Full charge will be achieved in approximately 2-3 hours

  • 8/10/2019 Epoch Lt Training 2-06

    15/58

    15

    Option Security Key

    Security Key allows installation of instrument software options

    Options can be programmed remotely - just contact Waltham

    All EPOCH LT Instruments have their security key soldered onto the main

    board. The keys cannot be removed as in older EPOCH 4 series instruments.

  • 8/10/2019 Epoch Lt Training 2-06

    16/58

    16

    EPOCH LTOPERATION

  • 8/10/2019 Epoch Lt Training 2-06

    17/58

    17

    The [ON/OFF] key is used to

    turn the EPOCH LT on and off.

    ON

    OFF

    EPOCH LT will go through a Self Teston Power up. This assures that the

    instrument is functioning properly.

    Power Up

  • 8/10/2019 Epoch Lt Training 2-06

    18/58

    18

    Liquid Crystal Display (LCD)Backlight On/Off

    Contrast Adjustment

    Press: then to adjust the contrast.2NDFCONTRAST

    CALIBRATION

    Contrast Adjustment

  • 8/10/2019 Epoch Lt Training 2-06

    19/58

    19

    Directory

    Filename

    Gain/Range

    Parameter Box

    Flaw Information

    (Always displayed)

    Split Screen View

    A-Scan and Setup Parameters can be Viewed Simultaneously

    Setup

    Parameters

    To Toggle Between Screen Views.

    Battery Level Indicator

    (Always displayed)Function Key Presets

    Split Screen

    A-Scan withflags

    PRESS: DISPLAY

    PRINT

  • 8/10/2019 Epoch Lt Training 2-06

    20/58

    20

    Full Screen View Provides a Large View of the A-Scan for Increased Resolution

    Directory

    Filename

    Gain/Range

    Parameter Box

    Flaw Information

    (Always displayed) Split ScreenA-Scan withflags

    Battery Level Indicator

    (Always displayed)

    Function Key Presets

  • 8/10/2019 Epoch Lt Training 2-06

    21/58

    21

    100%~ 8 Hours Backlight Off

    ~ 6 Hours Backlight On

    Battery Level Indicator

    50%~ 4 Hours Backlight Off

    ~ 3 Hours Backlight On

  • 8/10/2019 Epoch Lt Training 2-06

    22/58

    22

    This menu is used to:- Access Software Features / Options (DAC, TVG, etc.)

    - Adjust Operating Parameters (Language, Units, PRF, etc.)

    Press: to access menu.

    Use the Slewing keys

    to move highlight bar.

    Option Menu

    Press: to access a function.

    Press: again to exit menu.

  • 8/10/2019 Epoch Lt Training 2-06

    23/58

  • 8/10/2019 Epoch Lt Training 2-06

    24/58

    24

    Setup Menu

    Audible Tone Upon Keypress

    Turns A-Scan Grid ON/OFF

    Keypad and Calibration Lock

    Filled Frozen A-Scan when

    Filled Live A-Scan Waveform

  • 8/10/2019 Epoch Lt Training 2-06

    25/58

    25

    Real Time Clock

  • 8/10/2019 Epoch Lt Training 2-06

    26/58

    26

    Advanced User Selection

    The User Interface can be customized for Advanced Users

    Preset or Customizable Narrow Band Filters

    Zero Offset Preset Values

    Measure horizontal distance values from the front of an angle beam wedge rather than the BeamIndex Point

    Save MIN THICKNESS instead of CURRENT THICKNESS when the SAVE THICK key is

    pressed.

  • 8/10/2019 Epoch Lt Training 2-06

    27/58

    27

    EPOCH LT

    Bas ic Operat ing Parameters

  • 8/10/2019 Epoch Lt Training 2-06

    28/58

    28

    Basic Operation

    Access each parameter through a direct or

    2nd function key press.

    Press to toggle parameters.

    Adjust each parameter using:

    The Function Keys for preset values

    the Slewing Keys for fine tuning

    F1 F2 F3 F4 F5

  • 8/10/2019 Epoch Lt Training 2-06

    29/58

    29

    Press: or to fine tune the value.

    Press: , , , , or to select a preset value.

    Function Key Presets For quick access to preset values for each parameter

    Function Keys

    Press:

    Preset Values

    ANGLE

    THICKNESS

    F1 F2 F3 F4 F5

    Press: then in the Full Screen view to view the horizontal

    graticule division numbers instead of the function key parameter presets.

    2NDF ZERO

    OFFSET

    # DIV

    Example:

  • 8/10/2019 Epoch Lt Training 2-06

    30/58

    30

    Basic Operating Parameters

    GAINGAINREF

    FREEZE

    2NDFGAIN

    REF

    REFERENCE GAIN

    FREEZE/UNFREEZE SCREEN

    SAVE THICKNESS / AMPLITUDE / TOF DATA ONLY

    SAVE ENTIRE WAVEFORM AND SETUP PARAMETERS

  • 8/10/2019 Epoch Lt Training 2-06

    31/58

    31

    GATE 1 POSITION

    2NDF

    CONTRAST

    CALIBRATION

    GATE

    1

    ALARM 1

    GATE

    2

    ALARM 2

    GATE 1 ALARM

    GATE 2 POSITION

    GATE 2 ALARM

    GATE

    1

    ALARM 1

    2NDF GATE

    2

    ALARM 2

    AUTO-CALIBRATION INITIATION

    CONTRAST / BRIGHTNESS CONTROL2NDF

    CONTRAST

    CALIBRATION

    Basic Operating Parameters

  • 8/10/2019 Epoch Lt Training 2-06

    32/58

    32

    ZERO OFFSET

    2NDF ZEROOFFSET

    # DIV

    RANGE

    ZOOM

    VEL

    REJECT

    CHANGES X-AXIS IN FULL SCREEN TO

    DISPLAY GRATULE MARKINGS (0-10)

    SCREEN RANGE

    ZOOM / UNZOOMS SCREEN2NDF

    VELOCITY

    REJECT2NDF

    ZERO

    OFFSET

    # DIV

    RANGE

    ZOOM

    VEL

    REJECT

    Basic Operating Parameters

  • 8/10/2019 Epoch Lt Training 2-06

    33/58

    33

    REFRACTED ANGLE

    OPTION

    B-SCAN

    2NDF ANGLE

    THICKNESS

    MATERIAL THICKNESS

    OPENS EPOCH 4 OPTION MENU

    INITIALIZES THE B-SCAN OPTION2NDF

    ACCESSES THE PULSER FUNCTIONS

    ACCESSES THE PEAK MEM FEATURE2NDF

    ANGLE

    THICKNESS

    OPTION

    B-SCAN

    Basic Operating Parameters

  • 8/10/2019 Epoch Lt Training 2-06

    34/58

    34

    TOGGLES BETWEEN SPLIT / FULL SCREEN VIEW

    2NDF

    PRINT

    DISPLAY

    DEPTH

    %AMP

    ECHO-ECHO

    NOT ACTIVE IN EPOCH LT

    SELECTION FOR: -EDGE / PEAK DEPTH- AMPLITUDE

    - TOGGLE SOUNDPATH READINGS IN

    ANGLE BEAM MODE

    TURNS ON / OFF ECHO-ECHO FEATURE2NDF

    PRINT

    DISPLAY

    DEPTH

    %AMP

    ECHO-ECHO

    Basic Operating Parameters

  • 8/10/2019 Epoch Lt Training 2-06

    35/58

    35

    EPOCH LT

    Auto-Cal ibrat ion

    Straigh t Beam Example

  • 8/10/2019 Epoch Lt Training 2-06

    36/58

    36

    Transducer 5 Step Steel Test Block

    A109S-RM 2214E

    (5.0 MHz, 0.5 in) (0.100 to 0.500)

    EXAMPLE:

  • 8/10/2019 Epoch Lt Training 2-06

    37/58

    37

    CalibrationInitial Steps

    GAIN 10 dB

    ZERO OFFSET 0.000 us

    VELOCITY 2320 in /us (L-Wave Steel)

    RANGE 1in Full Screen

    ADJUST THE FOLLOWING PARAMETERS:GAIN

    REF

    ZERO

    OFFSET

    # DIV

    RANGE

    ZOOM

    VEL

    REJECT

  • 8/10/2019 Epoch Lt Training 2-06

    38/58

    38

    Rectification:Fullw ave, RF, Halfwave +, Halfwave -

    Pulser Energy:Low (100V), Med (200V), Hig h (300V), Max (400V)

    Pulser Damping:50, 63, 150, 400 ohms

    Pulser Mode:Pulse Echo, Dual, Through-Transmissio n

    ADJUST PULSER SETTINGS

    Multiple presses of the key toggles through pulser parameters.

    Pressing then will go directly to the filter selections2ND F

    CalibrationInitial Steps

  • 8/10/2019 Epoch Lt Training 2-06

    39/58

    39

    Pulser Type Selection

    The EPOCH LT has a standard negative spike excitation pulser and an

    OPTIONALtunable square wave pulser.

    Negative Spike Excitation Pulser:

    Easy to use

    No additional parameters needed for adjustment

    Tunable Square Wave Pulser

    Pulser is tuned to a frequency corresponding to the transducer

    Provides better signal to noise ratio for most applications

    Note: If Square Wave Pulser is selected, user must enter a frequency.

    CalibrationInitial Steps

  • 8/10/2019 Epoch Lt Training 2-06

    40/58

    40

    Resolution of Measurements

    Thickness and Time-of-Flight (msec) measurements are taken from theinstruments raw data.

    Measurement resolution is NOT tied to screen range.

    DEMONSTRATION!

  • 8/10/2019 Epoch Lt Training 2-06

    41/58

    41

    STEP #1:

    Couple the transducer to the THIN

    reference sample.(0.200 step)

    Position Gate 1 over the FIRST

    backwall echo.

    Adjust the Gain so the signal

    amplitude is ~ 80%

    Straight Beam Calibration

  • 8/10/2019 Epoch Lt Training 2-06

    42/58

    42

    Press:

    Wait for a steady reading, then press

    Uncouple the transducer.

    Enter the known thickness of the THIN reference sample.

    CONTRAST

    CALIBRATION

    ZERO

    OFFSET

    # DIV

    Straight Beam Calibration

  • 8/10/2019 Epoch Lt Training 2-06

    43/58

    43

    STEP #2:

    Press

    Couple the transducer to the THICK

    reference sample.(0.500 step)

    Position Gate 1 over the backwall echo.

    CONTRAST

    CALIBRATION

    Straight Beam Calibration

  • 8/10/2019 Epoch Lt Training 2-06

    44/58

    44

    Wait for a steady reading, then press

    Uncouple the transducer.

    Enter the known thickness of the THICK reference sample.

    Press to complete the calibration.

    VEL

    REJECT

    Straight Beam Calibration

  • 8/10/2019 Epoch Lt Training 2-06

    45/58

    45

    EPOCH LTAuto-Cal ibrat ion

    Ang le Beam Example

  • 8/10/2019 Epoch Lt Training 2-06

    46/58

    46

    Transducer 45oWedge IIW Type I Test Block

    A430S-SB ABWS-6-45 1018 Steel TB7541-1(2.25 MHz, 0.625 in)

    EXAMPLE:

  • 8/10/2019 Epoch Lt Training 2-06

    47/58

  • 8/10/2019 Epoch Lt Training 2-06

    48/58

    48

    Locating the Beam Index Point (B.I.P.)

    1. Align the wedge with the zero mark on the

    IIW Type I Block.

    2. Move the probe forward and backward untilthe echo amplitude from the 4 arc is peaked(The Peak Memory feature is helpful).

    3. Once the signal is peaked, mark the side ofthe wedge directly over the zero mark. This isthe Beam Index Point.

  • 8/10/2019 Epoch Lt Training 2-06

    49/58

    49

    NOTE: The refracted angle of the wedge should have already been entered intothe EPOCH LT at the beginning of the calibration. However, this angle must be

    verified due to material variations or wear on the wedge.

    Verify the Refracted Angle (Beta)

    1. Position the probe over the appropriate angle mark on the IIW block (matched

    to the angle on the wedge).

    2. Move the probe forward and backward to peak up the echo from the large

    circular hole in the block. (Again, the Peak Memory feature simplifies this

    process.)

  • 8/10/2019 Epoch Lt Training 2-06

    50/58

    50

    3. Once the signal is peaked, hold the probe stationary. Note the degree mark on the

    block that lines up with the Beam Index Point (B.I.P. was found in the previous step).

    This is the actual refracted angle (Beta) for the particular transducer and probe.

    4. If the Beta value differs from the one initially entered, enter a new one by

    pressing and then fine tuning the value using the or keys.ANGLE

    Verify the Refracted Angle (Beta)

  • 8/10/2019 Epoch Lt Training 2-06

    51/58

    51

    Calibrate for Distance

    1. Couple the probe to the IIW Type 1 Reference block so that the B.I.P. is directly

    over the 0 mark.2. Set the initial Range to 10 inches.

    3. Adjust the [ZERO OFFSET] so that both the 4 and 9echoes are visible on the

    screen. The [GAIN] may have to be increased to view the 9 echo.

    NOTE: An IIW Type 2 Reference Block will produce echoes at the 2

    and 4 screen divisions.

  • 8/10/2019 Epoch Lt Training 2-06

    52/58

    52

    Position Gate 1 over the 4 echo.

    Press .

    Wait for a steady reading, then press .

    Use the keypad to enter a thickness of 4.000 inches.

    CONTRAST

    CALIBRATION

    ZERO

    OFFSET

    Calibrate for Distance

  • 8/10/2019 Epoch Lt Training 2-06

    53/58

    53

    Position Gate 1 over the 9 echo.

    Press

    Wait for a steady reading, then

    press .

    Use the keypad to enter a thickness

    of 9.000 inches.

    Press to complete the calibration.

    VEL

    REJECT

    CONTRAST

    CALIBRATION

    Calibrate for Distance

  • 8/10/2019 Epoch Lt Training 2-06

    54/58

    54

    1. Couple the probe to the IIW Type 1 Reference Block so that the transducer is

    aimed at the 0.060 side drilled hole.

    2. Move the probe forward and backward to peak up the echo (The Peak

    Memory feature aids in this process)

    3. Use the [GAIN] key to bring the gain to a desired level.

    4. Lock in a reference gain and enable scanning gain by

    pressing then .

    Calibrate for Sensitivity

    0.060 sidedrilled hole

    GAIN

    REF

    2NDF

  • 8/10/2019 Epoch Lt Training 2-06

    55/58

    55

    Toggle Feature

    When utilizing angle beam transducers during inspections, the user can select whichsound path reading is displayed in large text.

    Press: then

    DEMONSTRATION!

    DEPTH

    %AMP

    ECHO-ECHO

    F4

  • 8/10/2019 Epoch Lt Training 2-06

    56/58

    56

    EPOCH LT

    DATALOGGER

  • 8/10/2019 Epoch Lt Training 2-06

    57/58

    57

    Datalogger Specifications

    Storage of 100 Waveforms/2,000 Thickness Readings

    Option to increase to 500 Waveforms/10,000 Thickness Readings

    Press 2ndF to enter/exit the memory screen.

    File Content Window

    File Directory Window

    Datalogger Capacity Remaining

  • 8/10/2019 Epoch Lt Training 2-06

    58/58

    Software Options

    DISTANCE AMPLITUDE CORRECTION (DAC) NOW STANDARD TIME VARIED GAIN (TVG) NOW STANDARD

    ADVANCED DAC/TVG

    ONBOARD DGS/AVG

    API 5UE

    SPOTWELD

    INTERFACE GATE

    FLOATING GATE

    CURVED SURFACE CORRECTION (CSC)

    PULSE REPETITION FREQUENCY (PRF)

    Low PRF (30 Hz)

    High PRF (1 KHz)

    B-SCAN

    WAVE ANALYSIS

    AWS D1.1

    EXPANDED MEMORY

    GAGEV IEW PRO PC INTERFACE PROGRAM

    *Refer to separate presentation

    for details on operationof software options