DELTA
Venlighedsvej 4
2970 Hørsholm
Denmark
Tel. (+45) 72 19 40 00
Fax (+45) 72 19 40 01
www.delta.dk
This report is issued under the rules of DANAK (Danish Accredi tation) and ILAC (International Laboratory Accreditation Cooperation) including i ts MRA (Mutual Recognition Arrangement) . Further information can be found at www.danak.dk and www.i lac.org. The report must not be reproduced, except in ful l , without the wri tten approval of DELTA.
DELTA Test and Examination Report
Version m
TEST Reg. no. 19
Type examination and testing of non-automatic weighing indicator MV2 according to OIML R76:2006 Performed for Cardinal Scale Manufacturing Company DANAK-1913746
Project no.: T205886, Revision 1
Page 1 of 67
including 2 annexes
29 January 2014
DANAK-1913746 DELTA T205886, Rev.1 Page 2 of 67
JHJ/jes
Title
Type examination and testing of non-automatic weighing
indicator MV2 according to OIML R76:2006
Test object Non-automatic weighing indicator MV2
Report no. DANAK-1913746
Project no. T205886, Revision 1
Test and Examination period
19 August 2013 to 15 January 2014
Client Cardinal Scale Manufacturing Company
203 East Daugherty Street,
Webb City, Missouri 64870
USA
Tel.: +1 417 673 4631
Contact person Stephen Langford
E-mail: [email protected]
Manufacturer Cardinal Scale Manufacturing Company
Specifications OIML R76:2006
EN 45501:1992/AC:1993
Welmec Guide 2.1:2001
Results The tested equipment fulfilled the requirements
Test personnel Claus Momme Thomsen (CMT),
Michael Nielsen (MI,
Jan Bjørnø Nielsen (JBN),
Jens Hovgård (JHJ)
Date 29 January 2014
Responsible
Jens Hovgård
Specialist, M.Sc.E.E.
CTE-Certification
DELTA
EC Notified Body No. 0199
DANAK-1913746 DELTA T205886, Rev.1 Page 3 of 67
JHJ/jes
Table of contents Page
1. Résumé and conclusion 4
Annex 1 OIML R 76:2006 checklist for MV2 5
Annex 2 OIML R 76:2006 test report for MV2 20
DANAK-1913746 DELTA T205886, Rev.1 Page 4 of 67
JHJ/jes
1. Résumé and conclusion
The report is about type examination and testing of the non-automatic medical weighing indicator
MV2 manufactured by Cardinal Scale Manufacturing Company, USA.
The MV2 is an electronic indicator with embedded software intended to be used as a module for
non-automatic medical scales. It can be interfaced to a height rod for height measurement and
BMI calculation, but these values are not to be regarded as legal values.
The examined software version is 1-01-03.
The test was performed according to OIML R76:2006, EN 45501:1992/AC:1993 and Welmec
Guide 2.1:2001.
The test items fulfilled OIML R76:2006 – and by that also EN 45501:1992/AC:1993 - for non-
automatic weighing instruments, as described in Annex 1 and 2.
A brief summary of the specifications for the MV2 indicator is as follows:
Accuracy class III and IIII
Weighing range: Single interval
Maximum number of verification
scale intervals (n): 6000 (class III), 1000 (class IIII)
Maximum capacity (Max): n × e
Maximum tare effect: -Max
Fractional factor: pi = 0.5
Minimum verification scale interval(e): 0.83 µV
Excitation voltage: 5 VDC
Circuit for remote sense: No
Minimum load cell input impedance: 87.5 Ohm
Maximum load cell input impedance: 1050 Ohm
Temperature range. +5 °C to +40 °C
Power supply: 100-240 VAC, 50-60 Hz
DANAK-1913746 DELTA T205886, Rev.1 Page 6 of 67
JHJ/jes
17 CHECKLIST
Application N°: T205886
Pattern designation: MV2 non-automatic weighing indicator
17.1 All types of weighing instruments except non-self-indicating instruments (6.1-6.9, R 76-1)
Requirement Testing pro-
cedures
PA
SS
ED
FA
ILE
D
Remarks
Descriptive markings
7.1.1 A.3 Compulsory in all cases:
manufacturer's mark or name X
accuracy class X
(+3.3.1) maximum capacity, Max, Max1, Max2,... X
minimum capacity, Min X
(+3.3.1) verification scale interval, e, e1, e2,... X
7.1.2 A.3 Compulsory if applicable:
name or mark of manufacturer’s agent /
serial number X
identification marks on separate but associated units /
type approval mark X
scale interval d (d <e) /
software identification (if applicable) / Displayed at power up
maximum tare effect T (subtractive tare only if T Max) /
maximum safe load, Lim (if Lim > Max + T) /
special temperature limits X
counting ratio /
ratio between weight platform and load platform /
range of plus/minus indication /
7.1.3 A.3 Additional markings:
not to be used for direct sales to the public /
to be used exclusively for: /
the stamp does not guarantee …/guarantees only ... /
to be used only as follows: /
3.2 special applications clearly marked (weighings ranges in
classes I and II or II and III)
/
4.15 near display “not to be used for direct sales to the public” (for
instruments similar to those used for direct sales to the pub-
lic)
/
7.1.4 A.3 Presentation of markings:
indelible X
easily readable X
grouped together in a clearly visible place X
Max, Min, e and d (if d e) on or near display permanently
shown in a clearly visible position
X
possible to seal and apply a control mark/removal will result
in destruction
X
Markings B and G X
7.1.4 and
7.1.1 B,
7.1.2 G
additional information shown alternatively on a plate or dis-
played by a software solution either permanently or accessed
by a simple manual command
/
7.1.5.1 A.3 Instruments with several load receptors and load measuring devices:
identification mark, Max, Min and e of each load receptor on
relating load measuring device (Lim and T = + if applicable)
/
DANAK-1913746 DELTA T205886, Rev.1 Page 7 of 67
JHJ/jes
7.1.5.2 A.3 Separately-built main parts:
identification mark repeated in descriptive markings /
4.1.1.3 Suitability for verification:
identification of devices which have been subject to separate
type examination
/
Verification marks and sealing
7.2 A.3 Verification mark:
cannot be removed X
easy application X
visibility without the instrument to be moved when it is in
service
X
7.2.2 Verification mark support or space:
which ensures conservation of the mark X
for stamp, stamping area 150 mm2 /
for self-adhesive type, ø 15 mm X
4.1.2.4 A.3 Securing of components and pre-set controls:
location X
form X
4.1.2.4 Securing with software means
4.1.2.4 a legal status of the instrument recognizable /
evidence of any intervention /
4.1.2.4 b protection against changes of parameters and the reference
numbers
/
4.1.2.4 c facilities for affixing the reference number /
4.1.2.5 Span adjustment device (automatic or semi-automatic): Existent Non-existent X
external influence impossible after securing
4.1.2.6 Gravity compensation: Existent X Non-existent
external influence on or access to impossible after securing X
Documentation
8.2.1 A.1 Technical information and data:
8.2.1.1, characteristics of the instrument X
3.10.2 specifications of modules X
3.10.2.1 fractions pi (modules tested separately) X 0.5
3.10.4 specifications of families /
specifications of components X
8.2.1.2 Applicable descriptive documents (acc. to No 1-11) X
5.3.6.1 A.1 specific declaration of the manufacturer X
3.9.1.1 limiting value of tilting defined by the manufacturer X Level indicator
8.2.2 A.2 Examination of:
documents X
functions (spotchecks) X
test reports from other authorities /
Indicating device
4.2.1 Reading:
reliable, easy and unambiguous X
overall inaccuracy 0.2 e (analogue indication) /
size, shape and clarity X
by simple juxtaposition X
4.2.2.1 A.3 Units of:
mass X
price /
DANAK-1913746 DELTA T205886, Rev.1 Page 8 of 67
JHJ/jes
4.2.2.1 Form of indication:
for one indication, one unit of mass X
scale interval in the form (1, 2 or 5) x 10k X
same scale interval for all indicating devices, printing devices
and tare weighing devices
X
4.2.2.2 Form of digital indication:
at least one figure at right X
Decimal sign:
shall maintain its position (scale interval changed automati-
cally)
X
separate at least one figure to the left and all to the right X
on one line with the bottom of the figures X
Zero:
only one non-significant zero to the right X
for values with decimal sign, non-significant zero only in
third position
X
4.2.3 Limits:
preventing of indication above Max + 9 e X
preventing of indication below zero unless a tare device is in
operation (–20d is accepted)
X
4.2.4 “Approximate” displaying device: Existent Non-existent X
scale interval > Max/100 without being smaller than 20 e
4.2.5 Semi-self indicating instruments:
extension of self-indication range self-indication capacity /
Analogue indication:
4.3.1 thickness and length of scale marks /
4.3.2 scale spacing /
4.3.3 limit of movement below zero and above capacity of self-
indication
/
4.3.4 damping of oscillations of displaying component /
4.4.1 Changing of digital indication:
after change in load, previous indication not longer than 1 s X Unless HOLD is active
4.4.2 Stable equilibrium of digital indication:
printed or stored weight values do not deviate more than 1 e
from the final weight value
X
zero or tare operations are within their accuracy requirements X
No printing, data storage, zero-setting, or taring during con-
tinuous or temporary disturbance of equilibrium
X
4.4.3 Extended digital indication: Existent Non-existent X
not allowed when there is a differentiated scale division
displaying a smaller scale interval only during pressing a key
at most, 5 s after manual command
prevention of printing while the device is in operation
4.4.4 Digital indications other than primary indications: Existent Non-existent X
additional indications do not lead to any ambiguity to primary
indications
quantities identified by units, symbols, signs or designations
thereof
weight values (not weighed) shall be clearly identified or
display only temporarily on manual command and
shall not be printed
the inoperative weighing mode is clear and unambiguously
recognisable
4.4.5 Digital printing: Existent Non-existent X
clear and permanent
figures 2 mm high
name or symbol of units to the right of the value
above column of values
printing impossible when equilibrium not stable
DANAK-1913746 DELTA T205886, Rev.1 Page 9 of 67
JHJ/jes
4.4.6 Memory storage: Existent Non-existent X
storage, transfer, totalizing, etc. inhibited when equilibrium
not stable
3.4.1 Auxiliary indicating device (Classes I and II only;
not allowed on multi-interval instruments) Existent Non-existent X
If existent, type: rider interpolation complementary differentiated scale division
only to the right of decimal sign
3.4.2 d < e 10 d, e = 10 k kg or e = 1 mg for class I with d < 1 mg
Differences between results
3.6.3 Differences:
between multiple indications: mpe /
between digital indications and printout: zero /
3.6.4 between two results: mpe for same load when method of
balancing changed (semi-self-indicating)
/
3.9.1.1 Tilting of instrument of class , or
a marking on the level indicator shows the limiting value of
tilting
/
level indicator fixed firmly in a place clearly visible to the
user
/
an automatic tilt sensor releases a display switch-off or other
appropriate alarm signal
/
and inhibits the printout and data transmission /
Zero-setting, -tracking and –indicating Existent Non-existent
Initial zero-setting X
Automatic zero-setting X
Semi-automatic zero-setting X
Non automatic zero-setting X
Zero-tracking X
Zero-indicating X
4.5.1 Effect shall not alter Max X
A.4.2.1 Overall effect of: zero-setting X ±2%
zero-tracking X
initial zero-setting X ≤ ±10%
4.5.2 A.4.2.3 Accuracy:
deviation 0.25 e X
deviation 0.5 d (auxiliary indicating device) /
4.5.3 Multiple range: Existent Non-existent X
effective for greater weighing range (if switching when load-
ed possible)
4.5.4 Control of zero-setting:
separate from that of tare weighing device X
Semi-automatic zero-setting: functions only
in stable equilibrium and X
if it cancels any previous tare operation X
DANAK-1913746 DELTA T205886, Rev.1 Page 10 of 67
JHJ/jes
4.5.5 A.4.2.2 Zero-indicating device (digital indication):
shows deviation 0.25 e X
not mandatory if auxiliary indicating device or rate of zero-
tracking 0.25 d/s
/
4.5.6 Automatic zero-setting:
operates only when equilibrium stable and /
indication has remained stable below zero at least 5 seconds /
4.5.7 Zero-tracking:
operates only when indication at zero or X
at negative net value equivalent to gross zero and X
equilibrium stable X
corrections 0.5 d/s X
when operates after tare, the overall effect may be 4 % of
Max
X ±2%
Tare devices Existent . Non-existent
Tare weighing X
Tare balancing X
Combined zero-setting and tare balancing X
Tare indicating X
Type : Subtractive X Additive
4.6.1 applicable requirements from 4.1 through 4.4 are fulfilled X
4.6.2 Tare weighing device:
dT = d X
4.6.3 A.4.6.2 Accuracy:
± 0.25 e (electronic instruments and instruments with ana-
logue indication), e = e1 for multi-interval
X
better than 0.5 d (mechanical instruments with digital indi-
cation
/
4.6.4 Operating range:
prevention of operation at its zero effect X
or below its zero effect X
prevention of operation above its maximum indicated X
4.6.5 Visibility of operation:
operation indicated X
net with sign "NET", "Net", "net" or complete word (digital
indication)
X
NET disappears if gross displayed temporarily X
tare value or letter "T" (mechanical additive tare device) /
4.6.6 Subtracting tare:
prevention of use above Max or indication that capacity is
reached
X
4.6.7 Multiple range:
operation effective in greater weighing ranges if switching
when loaded possible
/
tare values are rounded to the scale interval of the actual
weighing range which is in operation
/
4.6.8 Semi-automatic or automatic tare:
operation only in stable equilibrium X
DANAK-1913746 DELTA T205886, Rev.1 Page 11 of 67
JHJ/jes
4.6.9 Combined zero/tare:
accuracy (4.5.2) /
zero indicating device (4.5.5) /
zero-tracking (4.5.7) /
4.6.10 Consecutive tare operations:
indicated or printed tare weight values clearly designated (if
tare devices operative at the same time)
/
4.6.11 Printing net or gross:
without designation /
designation: by G or B (gross) /
by N (only net printed) /
designation of net and tare by N and T (if net printed with
gross and/or tare)
/
instead of G, B, N and T, complete words /
printing separately net and tare with identification (deter-
mined by different tare devices)
/
Preset tare Existent X Non-existent
4.7.1 dT = d or automatically rounded to d X
transferred from one range to another one with larger ei, shall
be rounded to the latter (multiple range)
/
tare value Max1 for the same net weight value (multi-
interval) and calculated net value rounded to the scale inter-
val for the same net weight value
/
4.7.2 4.6.10 applies X No consecutive tare
cannot be modified/cancelled if tare operated after the preset
tare is still in use
/
operates automatically if clearly identified with load /
4.7.3 4.6.5 applies X
possibility to indicate preset tare X
if calculated net printed then preset tare value is printed as
well
X
4.6.11 applies X
designation of preset tare by PT or complete word X
Locking devices Existent Non-existent X
4.8.1 Positions:
only two stable positions
weighing only in ‘weigh” position
4.8.2 positions clearly shown
Multiple ranges Existent Non-existent X
4.10 Selection of weighing ranges:
range in operation clearly indicated
selection from smaller to greater range possible at any load
(manual)
selection from smaller to the following greater range (auto-
matic) possible only for load Maxi of smaller range
selection from a greater to a smaller range (manually) or to
the smallest range (automatically) only
- at no load when zero or negative net value is indicated
- tare is cancelled automatically
- zero is set to ± 0.25 e1 automatically
DANAK-1913746 DELTA T205886, Rev.1 Page 12 of 67
JHJ/jes
Selection between load receptors, transmitting and measuring devices Existent Non-existent X
4.11, 4.11.1
4.11.2
4.11.3
4.11.4
compensation for unequal no-load effect
zero-setting without ambiguity and in accordance with 4.5
weighing impossible while selection
combinations easy identifiable
4.12 “Plus and “minus” comparator instruments
4.12.1 Distinction of zones:
"+" and "-" signs (analogue indication) /
by inscription (digital indication) /
4.12.2 Scale:
with at least one scale division d = e on either side of zero
and
/
value of d = e shown at either end /
Mechanical counting instruments with unit weigh receptor
4.17.1 Scale:
with at least one scale division d = e on either side of zero
and
/
value of d = e shown on the scale /
4.17.2 Counting ratio:
shown clearly above each counting platform or /
each counting scale mark /
4.20 Modes of operation: Non-existent
clearly identification of mode which is actually in operation /
manual switching back to weighing mode in any mode and at
any time possible
/
automatic selection of mode only within a weighing sequence /
automatic switching back to the weighing mode at the end of
the weighing sequence
/
zero indication after returning from switch-off condition X
automatic check of zero position before returning from
switch-off condition
X
DANAK-1913746 DELTA T205886, Rev.1 Page 13 of 67
JHJ/jes
17.2 Instruments for direct sales to the public and price computing and labelling instruments
Requirement Testing
procedures
PA
SS
ED
FA
ILE
D
Remarks
Miscellaneous checkings (direct sales to the public)
4.5.4 Combined semi-automatic zero-setting device and semi-automatic tare-balancing
device operated by the same key:
not allowed /
4.8.1 "Preweigh" position:
not allowed /
4.13.10 Counting ratio:
1/10 or 1/100 (mechanical counting instrument) /
4.13.5 Impossibility of weighing during:
locking operation /
adding or subtracting weights /
4.13.7 Auxiliary and extended indicating device:
not allowed /
4.13.9 When significant fault has been detected (electronic instruments):
visible or audible alarm provided for customer and (1) /
data transmission prevented (1) /
until user takes action or cause disappears /
Indication device (direct sales to the public)
4.13.1, 4.13.6 Primary indications (4.14.1) to both vendor and customer:
2 display sets, one vendor- and one customer display: Yes No
One display set for vendor and customer Yes No
weight /
information about correct zero position /
tare operation /
preset tare operation /
height of numerical figures displayed to the customer
≥ 9.5 mm
/
Instruments to be used with weights:
value of weights possible to distinguish /
Zero-setting device (direct sales to the public)
4.13.2 Non-automatic zero-setting:
only allowed when operated with a tool /
(1) Checked by verifying the compliance with documents or by simulating faults; this check does not duplicate the disturbance
tests 12.1 through 12.7.
DANAK-1913746 DELTA T205886, Rev.1 Page 14 of 67
JHJ/jes
Tare device (direct sales to the public)
4.13.3 not allowed on mechanical instrument with weights receptor /
on instruments with one platform public can see whether
- tare is in use
/
- tare setting is altered /
only one tare shall be in operation at any given time /
while tare or preset tare is in operation recalling of gross val-
ues is prohibited
/
4.13.3.1 Non-automatic tare:
displacement of 5 mm at most e /
4.13.3.2 Semi-automatic tare:
reduction of value of tare not permitted and /
cancelling of tare effect only if no load on the receptor /
One of the following condition fulfilled:
tare value indicated permanently in a separate display /
indicated with sign "-" when no load on the receptor /
tare effect cancelled automatically when unloading after net
weighing
/
4.13.3.3 Automatic tare:
not allowed /
4.13.4 Preset tare:
indicated on separate display clearly differentiated from
weight display
/
reduction of tare value not permitted and /
cancelling of tare effect only if no load on the receptor /
impossible to operate if tare device in operation /
cancelled at the same time as PLU if associated with PLU /
4.13.11 Self-service instruments: with one set of scales or displays
two sets of scales or displays
instrument has two sets of scales or displays /
Primary indications shall include the product designation if a
ticket is printed
/
Price computing instruments and price scales
(direct sales to the public)
4.14 Requirements of 4.13 for direct sales to the public are met /
4.14.1 Supplementary primary indications (4.13.6)
unit price /
price to pay /
if applicable number, unit price and price to pay for non-
weighed articles, price totals
/
4.14.2
4.2
4.3.1-4.3.3
Price scales:
4.2 and 4.3.1 through 4.3.3 /
error of price scale W • U - P e • U /
4.14.3 Price computing:
multiplication of indicated weight and unit price as indicated /
rounding to the nearest interval of price to pay /
unit price: Price/100 g or price/kg /
Indications of weights, unit price and price to pay visible:
while load on load receptor and for at least 1 s after stable
weight indication or after any introduction of unit price
/
freezing for 3 s after removing load and not possible to
introduce or change unit price (if indication has been stable
before and would otherwise be zero)
/
printing weight, unit price and price to pay /
DANAK-1913746 DELTA T205886, Rev.1 Page 15 of 67
JHJ/jes
Stored in memory:
before printing /
same data not to be printed twice for customer /
4.14.4 Additional functions for trade and management:
all transactions are printed for customer /
they shall not lead to confusion /
4.14.4.1 Prices-to-pay (positive or negative) of non-weighed articles:
weight indication zero or /
weighing mode inoperative /
prices shall be shown on price-to-pay display /
Prices for more than one equal articles:
number of articles shown on weight display without being
taken for a weight
/
price for one article shown on unit price display /
supplementary display for number of articles and/or article
prices
/
4.14.4.2 Totalization of transactions on one or several tickets:
price total indicated on price-to-pay display and /
printed accompanied by a special word or symbol and /
reference to commodities whose prices are totalized if a sepa-
rate ticket is issued for total
/
all prices-to-pay shall be printed and price total shall be the
algebraic sum of these printed prices
/
Totalization of transactions from linked instruments:
price-to-pay scale intervals of all connected instruments iden-
tical
/
4.14.4.3 Instrument used by several vendors or to serve more than one customer at the same
time:
connection between transactions and vendor or customer
identified
/
4.14.4.4 Cancelling previous transactions:
transaction is already printed: the price-to-pay cancelled shall
be printed with comment
/
transaction not yet printed and displayed to customer: trans-
action clearly differentiated from normal transactions
/
4.14.4.5 Printing additional information:
clearly correlated to transaction and /
does not interfere with assignment of weight value to unit
symbol
/
Price labelling instruments
4.16 Requirements 4.13.8, 4.14.3 (paragraphs 1 and 5), 4.14.4.1
(paragraph 1) and 4.14.4.5 are met
/
Display:
for weight /
possibility to verify values of unit price and preset tare during
the use of the instrument
/
Printing:
prevention of printing below Min /
labels with fixed values of weight, unit price and price-to-pay
allowed provided weighing mode is inoperative
/
DANAK-1913746 DELTA T205886, Rev.1 Page 16 of 67
JHJ/jes
4.18.1 Mobile instruments used outside Existent Non-existent X
means to indicate that the limiting value of tilting has been
exceeded and to inhibit printout and data transmission
automatic zero-setting or tare balancing operation after each
moving of the vehicle
indication when instrument is not in the weighing window
equipped with an appropriate protection system if the load
measuring device is sensitive to moving or driving influences
prevention of wrong weighing results if the cardanic suspen-
sion system or load receptor comes into contact with the sur-
rounding frame construction
4.18.2 Other mobile instruments not to be used outside
with a levelling device and a level indicator
the levelling device shall be operated easily without tools /
appropriate inscription pointing the user to the necessity of
levelling after each movement
/
DANAK-1913746 DELTA T205886, Rev.1 Page 17 of 67
JHJ/jes
17.3 Electronic weighing instruments
Requirement Testing
procedures
PA
SS
ED
FA
ILE
D
Remarks
Disturbances
5.1.1 indication of significant faults in the display does not lead to
confusion with other messages
X
5.2 Acting upon significant faults in case 5.1.1, b):
instrument made automatically inoperative (1), or /
visual or audible indication until user takes action or fault
disappears (1)
X
Display check
5.3.1 Upon switch-on:
signs of indication are active and non-active long enough to
be checked by operator
X
External equipment
5.3.6 Interfaces (mechanical, electrical, logical) do not allow:
- functions and measurement data to be inadmissibly influ-
enced by peripheral devices, or other connected instruments,
or disturbances
X
5.3.6.1 - displaying data which could be mistaken for a weighing
result
X
- falsifying weighing results (displayed, processed, stored) X
- changing adjustment factor or adjusting the instrument (ex-
cept authorized cases)
X
- falsifying displayed primary indications (direct sales) /
5.3.6.2 interfaces that do not fulfil 5.3.6.1 can be secured /
5.3.6.3 interfaces transmit data so that peripheral device can meet
requirements
X
5.3.6 Metrologically relevant functions performed or initiated
through the interface meet relevant requirements of No 4,
R76-1
/
(1) Checked by verifying the compliance with documents or by simulating faults; this check does not duplicate the
disturbance tests 12.1 through 12.7.
DANAK-1913746 DELTA T205886, Rev.1 Page 18 of 67
JHJ/jes
17.4 Software-controlled digital devices and instruments
Requirement Testing
procedures
PA
SS
ED
FA
ILE
D
Remarks
Devices with embedded software Existent X Non-existent
5.5.1 G.1 Declaration of the manufacturer that the software
- is used in a fixed hardware and software environment, and
X
- cannot be modified or uploaded by any means after secur-
ing/ verification
X
The software documentation contains:
- description of the legally relevant functions
X
- description of the securing means (evidence of an interven-
tion)
X
- software identification X REV 1-01-03
- description how to check the actual software identification X Shown at power up
The software identification is
- clearly assigned to the legally relevant software and func-
tions
X No software separation
- provided by the instrument as documented X
Personal computers, instruments with PC components, and other instruments, devices, modules,
and elements with programmable or loadable legally relevant software Existent Non-existent X
5.5.2.2 d G.2.1 The legally relevant software is
- documented with all relevant information
5.5.2.2 a - protected against accidental or intentional changes
5.5.2.2 a Evidence of intervention is available until the next verifica-
tion / inspection
5.5.2.2 G.2.2.1 Operation system / programs not accessible for the user
description of all commands via keys or interfaces
declaration of completeness of commands
5.5.2.2 G.2.2.2 Operating system / programs accessible for the user
checksum or signature generated over the machine code of
the legally relevant software
legally relevant software cannot be started if the code is falsi-
fied
G.2.2.3 In addition to the cases G.2.2.1 or G.2.2.2
device-specific parameters sufficiently protected
audit trail for the protection of the parameters and description
some practical spot checks performed
5.5.2.2 b G.2.3 Software interfaces
If there is associated software providing other than measuring
functions, the legally relevant software part
- is separated from associated software
- identified
- cannot be influenced by the associated software
program modules of legally relevant software are defined and
separated from the modules of associated software by a de-
fined protective software interface
protective software interface itself is part of the legally rele-
vant software
description and definition of functions of the legally relevant
software that can be released via the protective software in-
terface
description and definition of parameters that may be ex-
changed via the protective software interface
DANAK-1913746 DELTA T205886, Rev.1 Page 19 of 67
JHJ/jes
Requirement Testing
procedures
PA
SS
ED
FA
ILE
D
Remarks
description of the functions and parameters conclusive and
complete
each documented function and parameter does not contradict
to the requirements of this Recommendation
appropriate instructions for the application programmer con-
cerning the protectiveness of the software interface
5.5.2.2 c G.2.4 Software identification
The legally relevant software is identified by a software iden-
tification
The software identification
- covers all program modules of the legally relevant software
and the type-specific parameters at runtime of the instrument
- is easily provided by the instrument
- can be compared with the reference identification fixed at
type approval
Spot checks whether the checksums (signatures) are generat-
ed and work as documented
There exists an effective audit trail
Data storage devices (DSD) Existent Non-existent X
5.5.3 G.3.1 DSD realised with embedded software (examine software acc. to G.1) Yes No
DSD realised with programmable/loadable software
(examine software acc. to G.1) Yes No
documentation with all relevant information
5.5.3.1 G.3.2 sufficient storage capacity for the intended purpose
data are stored and given back correctly
sufficient description of measures to prevent data loss
5.5.3.2 G.3.3 storage of all relevant information necessary to reconstruct an
earlier weighing, i.e. gross, net, tare values, decimal signs,
units, identifications of the data set, instrument number, load
receptor, (if applicable), checksum / signature of the data set
stored.
5.5.3.3 G.3.4 protection of the stored legally relevant data against acci-
dental or intentional changes
protection of the stored legally relevant data at least with a
parity check during transmission to the storage device
protection of the stored legally relevant data at least with a
parity check of a storage device with embedded software
(5.5.1)
protection of the stored legally relevant data by an adequate
checksum or of a storage device with programmable or
loadable software (5.5.2)
5.5.3.4 G.3.5 identification and indication of the stored legally relevant
data with an identification number
record of the identification number on the official transaction
medium, i.e. on the print-out
5.5.3.5 G.3.6 automatic storage of the legally relevant data
5.5.3.6 G.3.7 a device subject to legal control prints or displays the stored
legally relevant data for verifying
Annex for report R76 - Report page 21 of 67 DANAK-1913746
OIML R 76 TEST REPORT
ORGANISATION INTERNATIONALE DE MÉTROLOGIE LÉGALE
INTERNATIONAL RECOMMENDATION
OIML R 76WELMEC 2.1
Non-automatic weighing instruments
Testing of module
Test item
MV2
manufactured byCardinal Scale Manufacturing Company
Task: T205886
DANAK-1913746
15 January 2014
DELTA Danish Electronics, Light & Acoustics
Jens Hovgård, DELTA Weighing Laboratory19 August to 18 November 2013
Authorized user of DTPR76: DELTA
Evaluator / Responsible:
Report No:
Date of Issue:
Test laboratory:
Issued by DELTA Report DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 22 of 67 DANAK-1913746
CONTENTS
PAGE
Summary of results …………………………………………………………………………… 23
Information concerning the instrument …………………………………………………… 24
Test specification …………………………………………………………………………… 26
Setup of load cell simulator, calibration and traceability ……………………………... 27
Test equipment …………………………………………………………………………….. 28
Estimation of measurement uncertainty …………………………………………………. 29
Photographs ……………………………………………………………………………………. 30
Initial Intrinsic error test ………………………………………………………………… 31
Temperature testsInitial test …………………………………………………………………………. 32
High temperature test …………………………………………………………… 33
Low temperature test …………………………………………………………… 34
Final temperature test ………………………………………………………… 35
Temperature effect on no-load indication ……………………………………………… 36
Repeatability ……………………..……………………………………………………… 37
TareTare analysis based on Intrinsic data …………………………………………… 38
Tare measurement No: 1 ………………………………………………………… 40
Tare measurement No: 2 ………………………………………………………… 41
Warm-up time ……………………..……………………………………………………… 42
Variations of voltage ……………………..……………………………………………… 44
Electrical disturbances, summary …………………………………………………… 45
Electrical disturbances, tests …………………………………………………. 46
Damp heat, steady stateTest before humidity ……………………..……………………………………… 56
Test during humidity ……………………..……………………………………… 57
Test after humidity ……………………..……………………………………… 58
Span stability Span measurement No: 1, Reference condition ……………………………… 59
Span measurement No: 2, After temperature test …………………………… 60
Span measurement No: 3, No change of condition …………………………… 61
Span measurement No: 4, After power interruption …………………………… 62
Span measurement No: 5, No change of condition …………………………… 63
Span measurement No: 6, After humidity test …………………………… 64
Span measurement No: 7, After power interruption …………………………… 65
Span measurement No: 8, No change of condition …………………………… 66
Result of span stability test ……………………………………………………………… 67
Tests:
Issued by DELTA Cont DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 23 of 67 DANAK-1913746
Summary of results
Application / Task: T205886
Applicant name: Cardinal Scale Manufacturing Company
Type: MV2
OIML: R76 accuracy Class III
R76
No
Inputimpe-dance
Report sheet
PASS FAIL Remarks
Checking of zero setting N/A -
Low X
High X
Low X
High X
Low X Conclusion for n = 6000
High X and 0,83 [µV] / n
Low X
High X
Low
3 Eccentricity -
4 Discrimination -
5 Repeatability Low Rep X
6.1 Zero return -
6.2 Creep -
7 -
8 Tilting -
9 Tare L/H/L Tare(-) X Based on a tare analysis and tare test
10 Warm-up time High W X
11 Voltage variations Low Var X
12.1 Short time power reductions High EMC X
a) Power supply lines High EMC X
b) I/O circuits and comms. Lines High EMC X
a) Direct application High EMC X Result: 0,48 [µV / VSI]
b) Indirect application High EMC X
a) Radiated 10 V/m test High EMC X
b) Conducted 10 V test High EMC X
a) Power supply lines 1 [kV] High EMC X
b) I/O lines - [kV] - N.A.
a) Power supply lines - N.A.
b) Other lines - N.A.
a) Initial test at ref. conditions High H(1) X
b) Test at high humidity and temp. High H(2) X
c) Final test at ref. conditions High H(3) X
14 Span stability Low SP(res) X
Low Span_T1 X
High Span_T2 X
-
-
Sense interruptions -
a) Test before at ref. conditions High -
b) Test after at ref. Conditons High -
16 Exam X
17 Check X
In all:
*) WELMEC 2.1, Annex 5.
PASSED
12.4
Checklist
Examination of the construction
Load cell interface: Maximum cable length to the junction box for load cells
Immunity to RF fields
15
Surge
*)
13
12.3Electrostatic discharges
*)
Endurance
Examinations
Low Z X
T(4)
-
Result: 0,04 [µV / VSI]
Tests
Test at high temperature
Test at low temperature
Final test at ref. temp.
5 °C
40 °C
Calculation of intrinsic error
20 °C
12.2Electrical bursts
1
2 Temperature effect on no-load indication
20 °C
Weighing performance
Initial test at ref. temp.
Made: 2013-8-19Low XIntrinsic
Temperature effect on span
Damp heat, steady state
T(1)
T(2)
T(3)
Powered from vehicle battery
Stability of equilibrium
Issued by DELTA Sum DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 24 of 67 DANAK-1913746
Information concerning the instrumentInformation submitted by the manufacturer.
Applicant name:address:
Contact person:e-mail:
Telephone:
Manufacturer name:address:
Type:
Test item device:
S/N:Type subject to EMC tests:S/N for test item subject to EMC tests:Separate load receptor, if submitted Type: Capacity [kg]: S/N:
Load cell(s), if submitted: Type: Manuf.: Number: S/N:
Load cell class: Capacity [kg]: Cable[m]:
Rated output: [mV/V] umin = [µV] per VSI
Accuracy class: OIML R76: III
Optional for AWI, if applied: AWI class: 0
"R61: Specified Ref. Class" (Not used)
Weighing range(s) specified: Single-interval
Number (i) of Intervals / ranges specified: 1
Change-over point for multi-range, if available: Maxi + 9 ei
Unit marked on test item: [kg]
Internal resolution of measurements:
Number of Verification Scale Intervals (ni): 10000 ni = Maxi / ei
Maximum capacity of partial ranges (Maxi): 200 [kg] Min: [kg]
Verification Scale Interval (ei): 0,02 [kg] e: [kg]
Fractional factor (pi): 0,5
Excitation voltage: 5 [Vdc]Minimum input-voltage per VSI (∆umin): 0,50 [µV] Later changed to 0,83 [µV]Minimum dead load (Dmin): 0 [mV]
Maximum input range: 15 [mV]Rated maximum input-impedance for load cells: 1050 [ohm]Rated minimum input-impedance for load cells: 87,5 [ohm]
Sense (6-wires) or no-sense (4-wires): Sense
Maximum resistance of cable to JB for load cells: each wire [ohm]
Zero-setting devices Range [% of Max] Tare devicesNonautomatic: FALSE Tare balancing: TRUE
Semi-automatic: TRUE -2 +2 Tare Weighing: FALSEAutomatic zero-setting: FALSE Preset tare device: FALSE
Initial zero-setting: TRUE Subtractive tare: TRUEZero-tracking (+/- 0.25e): TRUE -2 +2 Additive tare: FALSE
Combined zero/tare device: FALSE Max. Tare effect: 100,0 [% of Max]
Operating temperature rangeMin: 5Max: 40
Power voltage supply [Vac]Un: 50/60 [Hz] External supply
Umin: 100 with 12 VDC outputUmax: 240
Battery supply [Vdc]Un: 9
Umin:Umax:
Load cell linearizing, if available:
+1 417 673 4631 ext. 231
Cardinal Scale Manufacturing Company203 East Daugherty, P.O. Box 15, Webb City, MO 64870, USA
E13613-0307
MV2
0,4
20
Cardinal Scale Manufacturing Company203 East Daugherty, P.O. Box 15, Webb City, MO 64870, USA
Brian [email protected]
Indicator
Issued by DELTA Info(I) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 25 of 67 DANAK-1913746
Printer subject to test: Non present, but connectable
No. of interfaces: 3
Specification of interfaces available
Interface No. 1 (applied for recording of measurements during tests) xformat / standard: cable length: < 3protective or not: shield/no shield: shielded
Interface No. 2. xformat / standard: cable length: 1,8protective or not: shield/no shield: shielded
Interface No. 3.format / standard: cable length: 1,8protective or not: shield/no shield: shielded
Peripheral equipment:
Remarks:
Digital Hight Rod I/F
Connections during electrical disturbance tests:
Protective
Connecting cables
ProtectiveRS-232C
USBProtective
Issued by DELTA Info(I) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 26 of 67 DANAK-1913746
Test specificationOIML: R76WELMEC: Guide 2.1
Application / Task: T205886Applicant name:Manufacturer name: Cardinal Scale Manufacturing Company
Type MV2Test item device: Indicator
S/N: E13613-0307Type subject to EMC tests:S/N for test item subject to EMC tests:
Maximum permissible errors apply to: Initial verification
Accuracy class: III
Unit of measurement: [count]
Weighing range(s) subject to test: Single interval
Number of intervals / ranges: 1
Change-over point for multi-interval, if available: Maxi +: ei Not Applicable
Number of Verification Scale Intervals (ni): 6000Maximum indication during test (Maxi): 65000 [count]Maximum capacity (Max): 6001,85 [ei]
Indication for one (e) during test: 10,83 [count]
Min: 216,6 [count]
Resolution during test: 1 [count]
Fractional factor (pi): 0,5
Initial zero-setting range 20 [%]
Excitation voltage: 5 [Vdc]
Minimum input-voltage per VSI (∆umin): 0,83 [µV]
Minimum dead load (Dmin): 0 [mV]
Dead load used for the tests (zero offset): 0 [mV] Proposal to begin with
Minimum number of load cells (each 350 ohm): 1 Impedance: 350 [ohm]
Maximum number of load cells (each 350 ohm): 4 Impedance: 87,5 [ohm]Sense (6-wires) or no-sense (4-wires): No-sense
Operating temperature range: Min / Max = 5 °C / 40 [°C]
Power voltage supply: Un Umin Umax
External supply AC supply: 100 240 [Vac] 50/60 [Hz]Battery supply: 9 [Vdc]
Submitted Documentation.
Manuals Owner's Manual
Drawings Electronic schematicsPCB layouts
Other
Remarks:
Cardinal Scale Manufacturing Company
Issued by DELTA Spec(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 27 of 67 DANAK-1913746
Setup of load cell simulator Ref.: OIML R76 and WELMEC 2.1
Before the formal start of a type test the A/D device and the load cell simulator shall be calibrated together.The A/D device should be calibrated with minimum input impedance and the maximum number of verification scale intervals and the minimum input-voltage per verification scale interval.
A procedure for finding the best fit of the setup based on the manufacturer's specifications is given below.
Test item:
Manufacturer:Type and serial no.:Number of intervals / ranges: 1
Maximum indication during test (Maxi): 65000 [count]
One verification scale interval (VSI equals e): 10,83 [count]Number of Verification Scale Intervals (n): 6000Resolution during test: 1 [count]Excitation voltage: 5 [Vdc]Minimum input-voltage per VSI (∆umin): 0,83 [µV]Minimum dead load (Dmin): 0 [mV] = 0 [mV/V]
Span scale range (n * ∆umin): 4,98 [mV] = 0,996 [mV/V]Scale range (span + Dmin): 4,98 [mV] = 0,996 [mV/V]Fractional factor (pi): 0,5
Maximum permissible error at zero (Mpe * p): 0,25 *e 2,7075 [count]Minimum number of load cells (each 350 ohm): 1Maximum number of load cells (each 350 ohm): 4 87,5 [ohm]Sense (6-wires) or no-sense (4-wires): No-sense
Load cell simulator
Load cell simulator applied: Type: HBM, K3608 S/N:Specified setup:Length of cable between indicator and simulator: < 3 [m]Lowest value of the scale (mV/V): 0 [mV/V] 0 [mV]Highest value of the scale (mV/V): 0,996 [mV/V] 4,98 [mV]Span range (mV/V): 0,996 [mV/V] 4,98 [mV]
Simulator values used as Load values: Initial Intrinsic values
Setup for best fit
Selected calibration range: 1 [mV/V]Selected lowest value of the scale (mV/V): 0 [mV/V] 0 [%] of the rangeSelected highest value of the scale (mV/V): 1,00 [mV/V] 100 [%] of the rangeSelected span test range (mV/V) : 1,00 [mV/V] 5 [mV]Re-calculated ∆umin (best fit): 0,83 [µV]
Simulator calibration values Simulator LoadRange 0,5 [mV/V] 1 [mV/V] 2 [mV/V] 5 [mV/V] 10 [mV/V] Linear
[%] [mV/V] [mV/V] [mV/V] [mV/V] [mV/V] [%] [mV/V] [count] [mV/V] [count] [count] [mV/V] [count] Selected
0 0,000000 0,000000 0 0,00 0 0,000000 0 0 0,000000 010 0,099998 0,499989 10 0,10 6500 0,099998 6500,1 6500 0,100000 650020 0,199995 0,999981 20 0,20 13000 0,199995 13000,1 13000 0,200000 1300030 0,299991 1,499959 30 0,30 19500 0,299991 19500 19500 0,300000 1950040 0,399988 1,999940 40 0,40 26000 0,399988 26000 26000 0,400000 2600050 0,499985 2,499916 50 0,50 32500 0,499985 32500 32500 0,500000 3250060 0,599986 2,999889 60 0,60 39000 0,599986 39000,2 39000 0,600000 3900070 0,699976 3,499883 70 0,70 45500 0,699976 45499,7 45500 0,700000 4550080 0,799972 3,999862 80 0,80 52000 0,799972 51999,7 52000 0,800000 5200090 0,899971 4,499850 90 0,90 58500 0,899971 58499,8 58500 0,900000 58500
100 0,999971 4,999842 100 1,00 65000 0,999971 65000 65000 1,000000 65000
Remarks:
Indicator
MV2 S/N: E13613-0307Cardinal Scale Manufacturing Company
90220004
Range LoadInitial intrinsicCalibrated
Issued by DELTA Sim_Set DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 28 of 67 DANAK-1913746
Test equipment Ref.: OIML R76 and WELMEC 2.1
T205886
Test loads Weights not applied
Accuracy class: M1Other information:
Simulator
Type and serial number: HBM, K3608, S/N: 90220004
Traceability: to DKD
Laboratory for the performance tests
Facilities for influence factor testing
Equipment applied for the performance tests
Facilities for electrical disturbance testing
See page EMC(T)
Photograph of test equipment and setup
See page Photographs
Loading using Weights The table below is not used
Accuracy class: M1
Not applied: 0
Not applied:Load Accuracy Intervals in use: 1
No. N/A N/A n ni %
0 0 OIML R 111OIML R 111OIML R 111OIML R 111OIML R 111OIML R 111OIML R 111OIML R 111OIML R 111OIML R 111OIML R 111
Maximum value: 0
Hygro-/termometer Elpro-buchs Ecolog TH-2 40172Multimeter Fluke 27 23778
Jens Hovgård, DELTA Weighing Laboratory Application / Task:
Evaluator:
Test laboratory conditions:
Weight Sets No. 40102, 40104, 40107, 40101
Air conditioned 20 °C, 50 % RH
Load used
Test laboratory location:
Humidity chamber:
NameVariotransformer
2A12C
Type No.Manufacture
Temperature chamber:
Lübcke
VKF875-2
2A12
R54-260B 23733
Issued by DELTA Equip DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 29 of 67 DANAK-1913746
Estimation of measurement uncertainty
See page Equip
Measuring chain for determination of errors of measurements
Standard bridge calibrator BN100 with traceability to DKD in Germany
High precision measuring instrument DMP39 with traceability to BN100
Load cell simulator K3607 with traceability to DMP39
Uncertainty of measurements
Max contribution
a BN100: 0,001 [%]
b DMP39 & BN100 (repeatability error): 0,0002 [%]
c K3607 & DMP39 (repeatability error): 0,0001 [%]
d K3607 (temperature dependence): 0,001 [% of span/°C]
e Temp. measurements at influence tests: ± 0,2 [°C]
f Temp. measurements in test lab.: ± 0,2 [°C]
g LIS 1 for cable length testing: ± 2 [% of the nominal value]
h Variation of test lab. temp. in the test period: 1,3 [°C]
Confidence factor included in b and c: 2,5
Intrinsic curve : uintrinsic = 0,0002 [%] √ ( ( c )2 + ( b )2 )
Rest of the tests : u = 0,0017 [%] √ ( ( ( h + 2 * f ) * d )2 + ( c )2 + ( b )2 )
∆umin : uumin = 0,0008 [µV / VSI] √ ( ( u )2 + ( ( e + e ) / ∆temp )2 )
Cable length : ucable = - [m] √ ( ( u )2 + ( g )2 )
Everything considered the accuracy and stability of the measurements are sufficient good
to comply with R76 paragraph 3.7.1 and A.4.4.4.
The evidence is given by the Intrinsic error curve (see Intrinsic).
Issued by DELTA Uncertain DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 30 of 67 DANAK-1913746
Remarks:
Photographs
Issued by DELTA Photo DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 31 of 67 DANAK-1913746
TEST: Initial intrinsic error (OIML R76: T.5.5.3, figure 2)WELMEC 2.1
Application / Task: T205886 Start MaxType designation: Temp.[°C]: 21Test item: Indicator Rel. H.[%]: 51Serial No.: E13613-0307 Time: 14:30Accuracy class: III Date:
Max: 6001,85 [ei] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]
HBM, K3608, S/N: 902200041 [mV/V], 0 [%] to 100 [%], 0,83 [µV] per VSI
Not in operation
No
4 load cell 87,5 [Ohm]
E = I + ½e - L - L Ec = E - E0
pi = 0,5
↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,000000 0 0 0 0 0 0 0 2,7 2,710 0,099998 6500,1 6500 6500 -0,1 -0,1 -0,1 -0,1 5,4 5,420 0,199995 13000,1 13000 13000 -0,1 -0,1 -0,1 -0,1 5,4 5,430 0,299991 19500 19500 19500 0 0 0 0 5,4 5,440 0,399988 26000 26000 26000 0 0 0 0 8,1 8,150 0,499985 32500 32500 32500 0 0 0 0 8,1 8,160 0,599986 39000,2 39000 39000 -0,2 -0,2 -0,2 -0,2 8,1 8,170 0,699976 45499,7 45500 45500 0,3 0,3 0,3 0,3 8,1 8,180 0,799972 51999,7 52000 52000 0,3 0,3 0,3 0,3 8,1 8,190 0,899971 58499,8 58500 58500 0,2 0,2 0,2 0,2 8,1 8,1100 0,999971 65000 65000 65000 0 0 0 0 8,1 8,1
Passed / Failed: PASSED
Observer:Remarks:
MV2
E EcCorrected error
Aut. Zero-setting and zero-tracking device is:
Initial zero-setting ( > 20% of Max ):
Error
Simulator applied:Simulator range:
(Calibrated values)
JBN
L I LLoad Indication Add. Load
Load cell number and input impedance:
MPE* pi
19-08-2013
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [c
ount
]
Load [count]
Initial intrinsic error curve
Up loading Down loading Mpe
Issued by DELTA Intrinsic DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 32 of 67 DANAK-1913746
TEST: Weighing Performance (OIML R76: T.5.5.3 (figure 2) & A.5.3.1)WELMEC 2.1 Begin Max End
Application / task: T205886 Room temp.[°C]: 21 21Type designation: Temp.[°C]: 21 21Test item: Indicator Rel. H.[%]: 52Serial No.: E13613-0307 Time: 14:30 15:00Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000 Temp. Test: 1 Ref. 1Resolution during test: 1 [count]
HBM, K3608, S/N: 90220004 Placed outside the climatic chamber.1 [mV/V], 0 [%] to 100 [%], 0,83 [µV] per VSI
Not in operation
4 load cell(s) 87,5 [Ohm]
2.5 [m] 4-wires
E = I + ½e - L - L Ec = E - E0 pi = 0,5
P
/
↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ F% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,000000 0 0 0 0 0 0 0 2,7 2,7 P10 0,100000 6500 6500 6500 0 0 0 0 5,4 5,4 P20 0,200000 13000 13000 13000 0 0 0 0 5,4 5,4 P30 0,300000 19500 19500 19500 0 0 0 0 5,4 5,4 P40 0,400000 26000 26000 26000 0 0 0 0 8,1 8,1 P50 0,500000 32500 32500 32500 0 0 0 0 8,1 8,1 P60 0,600000 39000 39000 39000 0 0 0 0 8,1 8,1 P70 0,700000 45500 45500 45500 0 0 0 0 8,1 8,1 P80 0,800000 52000 52000 52000 0 0 0 0 8,1 8,1 P90 0,900000 58500 58500 58500 0 0 0 0 8,1 8,1 P
100 1,000000 65000 65000 65000 0 0 0 0 8,1 8,1 PNew no-load indication: (if it has changed from the first recorded value)
Passed / Failed: PASSED
Additional measurements to cover the whole impedance range:1 350 [Ohm]
0 0,000000 0 0 0 0100 1,000000 65000 65236 236 236
Corrected for init span: 0 8,1 P
Observer:Remarks:
MPE
Load cell number and input impedance:
(Initial Intrinsic values)
JBN
L I E EcL * pi
Corrected errorErrorLoad Add. LoadIndication
Length of cable outside the climatic chamber:
Simulator range:Simulator applied:
19-08-2013
MV2
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve
Up loading Down loading Mpe
Issued by DELTA T(1) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 33 of 67 DANAK-1913746
TEST: Weighing Performance (OIML R76: T.5.5.3 (figure 2) & A.5.3.1)WELMEC 2.1 Begin Max End
Application / task: T205886 Room temp.[°C]: 20,6 20,6Type designation: Temp.[°C]: 40,1 40,1Test item: Indicator Rel. H.[%]: 49Serial No.: E13613-0307 Time: 09:47 10:00Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000 Temp. Test: 2 High TResolution during test: 1 [count]
HBM, K3608, S/N: 90220004 Placed outside the climatic chamber.1 [mV/V], 0 [%] to 100 [%], 0,83 [µV] per VSI
Not in operation
4 load cell(s) 87,5 [Ohm]
2.5 [m] 4-wires
E = I + ½e - L - L Ec = E - E0 pi = 0,5
P
/
↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ F% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,000000 0 -1 -1 -1 -1 0 0 2,7 2,7 P10 0,100000 6500 6497 6497 -3 -3 -2 -2 5,4 5,4 P20 0,200000 13000 12997 12997 -3 -3 -2 -2 5,4 5,4 P30 0,300000 19500 19497 19497 -3 -3 -2 -2 5,4 5,4 P40 0,400000 26000 25997 25997 -3 -3 -2 -2 8,1 8,1 P50 0,500000 32500 32497 32497 -3 -3 -2 -2 8,1 8,1 P60 0,600000 39000 38996 38996 -4 -4 -3 -3 8,1 8,1 P70 0,700000 45500 45496 45496 -4 -4 -3 -3 8,1 8,1 P80 0,800000 52000 51996 51996 -4 -4 -3 -3 8,1 8,1 P90 0,900000 58500 58496 58496 -4 -4 -3 -3 8,1 8,1 P
100 1,000000 65000 64996 64996 -4 -4 -3 -3 8,1 8,1 PNew no-load indication: -1 (if it has changed from the first recorded value)
Passed / Failed: PASSED
Additional measurements to cover the whole impedance range:1 350 [Ohm]
0 0,000000 0 -1 -1 0100 1,000000 65000 65232 232 233
Corrected for init span: -3 8,1 P
Observer:Remarks:
MV2
Corrected error
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Error
E Ec
Load
JBN
Load cell number and input impedance:
L I L
(Initial Intrinsic values)
20-02013
* pi
Indication Add. Load
Length of cable outside the climatic chamber:
Simulator applied:Simulator range:
MPE
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [c
ount
]
Load [count]
Error Curve
Up loading Down loading Mpe
Issued by DELTA T(2) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 34 of 67 DANAK-1913746
TEST: Weighing Performance (OIML R76: T.5.5.3 (figure 2) & A.5.3.1)WELMEC 2.1 Begin Max End
Application / task: T205886 Room temp.[°C]: 20,5 20,6Type designation: Temp.[°C]: 5,0 5,1Test item: Indicator Rel. H.[%]: 55Serial No.: E13613-0307 Time: 09:10 10:20Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000 Temp. Test: 3 Low TResolution during test: 1 [count]
HBM, K3608, S/N: 90220004 Placed outside the climatic chamber.1 [mV/V], 0 [%] to 100 [%], 0,83 [µV] per VSI
Not in operation
4 load cell(s) 87,5 [Ohm]
2.5 [m] 4-wires
E = I + ½e - L - L Ec = E - E0 pi = 0,5
P
/
↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ F% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,000000 0 0 1 0 1 0 1 2,7 2,7 P10 0,100000 6500 6501 6501 1 1 1 1 5,4 5,4 P20 0,200000 13000 13002 13002 2 2 2 2 5,4 5,4 P30 0,300000 19500 19503 19503 3 3 3 3 5,4 5,4 P40 0,400000 26000 26004 26004 4 4 4 4 8,1 8,1 P50 0,500000 32500 32504 32504 4 4 4 4 8,1 8,1 P60 0,600000 39000 39005 39005 5 5 5 5 8,1 8,1 P70 0,700000 45500 45506 45506 6 6 6 6 8,1 8,1 P80 0,800000 52000 52007 52007 7 7 7 7 8,1 8,1 P90 0,900000 58500 58507 58508 7 8 7 8 8,1 8,1 P
100 1,000000 65000 65008 65008 8 8 8 8 8,1 8,1 PNew no-load indication: (if it has changed from the first recorded value)
Passed / Failed: PASSED
Additional measurements to cover the whole impedance range:1 350 [Ohm]
0 0,000000 0 1 1 0100 1,000000 65000 65242 242 241
Corrected for init span: 5 8,1 P
Observer:Remarks:
MV2
Corrected error
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Error
E Ec
Load
JBN
Load cell number and input impedance:
L I L
(Initial Intrinsic values)
21-08-2013
* pi
Indication Add. Load
Length of cable outside the climatic chamber:
Simulator applied:Simulator range:
MPE
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [c
ount
]
Load [count]
Error Curve
Up loading Down loading Mpe
Issued by DELTA T(3) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 35 of 67 DANAK-1913746
TEST: Weighing Performance (OIML R76: T.5.5.3 (figure 2) & A.5.3.1)WELMEC 2.1 Begin Max End
Application / task: T205886 Room temp.[°C]: 20,4 20,4Type designation: Temp.[°C]: 20,7 20,7Test item: Indicator Rel. H.[%]: 52Serial No.: E13613-0307 Time: 15:35 15:45Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000 Temp. Test: 4 20 °CResolution during test: 1 [count]
HBM, K3608, S/N: 90220004 Placed outside the climatic chamber.1 [mV/V], 0 [%] to 100 [%], 0,83 [µV] per VSI
Not in operation
4 load cell(s) 87,5 [Ohm]
2.5 [m] 4-wires
E = I + ½e - L - L Ec = E - E0 pi = 0,5
P
/
↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ F% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,000000 0 0 0 0 0 0 0 2,7 2,7 P10 0,100000 6500 6500 6500 0 0 0 0 5,4 5,4 P20 0,200000 13000 13000 13000 0 0 0 0 5,4 5,4 P30 0,300000 19500 19500 19500 0 0 0 0 5,4 5,4 P40 0,400000 26000 26000 26000 0 0 0 0 8,1 8,1 P50 0,500000 32500 32500 32500 0 0 0 0 8,1 8,1 P60 0,600000 39000 39000 39000 0 0 0 0 8,1 8,1 P70 0,700000 45500 45500 45500 0 0 0 0 8,1 8,1 P80 0,800000 52000 52000 52000 0 0 0 0 8,1 8,1 P90 0,900000 58500 58500 58500 0 0 0 0 8,1 8,1 P
100 1,000000 65000 65000 65000 0 0 0 0 8,1 8,1 PNew no-load indication: 1 (if it has changed from the first recorded value)
Passed / Failed: PASSED
Additional measurements to cover the whole impedance range:1 350 [Ohm]
0 0,000000 0 1 1 0100 1,000000 65000 65237 237 236
Corrected for init span: 0 8,1 P
Observer:Remarks:
MV2
Corrected error
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Error
E Ec
Load
Jens Hovgård
Load cell number and input impedance:
L I L
(Initial Intrinsic values)
22-08-2013
* pi
Indication Add. Load
Length of cable outside the climatic chamber:
Simulator applied:Simulator range:
MPE
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve
Up loading Down loading Mpe
Issued by DELTA T(4) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 36 of 67 DANAK-1913746
TEST: Temperature Effect on No-load Indication (OIML R76: A.5.3.2) WELMEC 2.1
Application / task: : T205886Type designation: :Test item: : IndicatorSerial number : E13613-0307Observer : JBNAccuracy class: : IIIMax: : 65000 [count]Verification scale interval, e: : 10,83 [count]No. of Verification scale intervals, n: : 6000Resolution during test (smaller than e) : 1 [count]Requirement : 5,4 [count]Fraction (pi) : 0,5
Excitation load (load cell input impedance : 87,5 [ohm]Excitation voltage : 5 [Vdc]Measuring range per volt : 0 to 1,00 [mV/V]Measuring range : 0 to 5 [mV]Load cell simulator : Type: HBM, K3608 S/N: 90220004(simulator outside the climatic chamber) Range : 0 to 100 [%]
Setting : 1 [mV/V]: 0,83 [µV] Later changed to 0,83 [µV]
Sense (6-wires) or no-sense (4-wires) :: Not in operation
P=I + ½e - LSee Date Time Temp Zero Add. Temp. Zero- Require-
Work indication load diff. change mentsheet (I) (L) (P) (P) per 5 °C (pi * VSI)
2013 hh:mm [°C] [count] [count] [count] [count] [K] [count] [count]
T (1) 19-aug 14:30 21 0 020-02013 09:47 40,1 -1 -1
T (2)20-02013 09:47 40,1 -1 -1
21-aug 09:10 5 0 0T (3)
21-aug 09:10 5,1 0 022-aug 15:35 20,7 0 0
T (4)22-aug 15:35 20,7 1 1
T (5) 0 0
Equivalent minimum input-voltage per verification scale interval : Result: 0,04 [µV / VSI]
Equivalent minimum input-voltage per verification scale interval In'min = ( Max. Zero-change per 5°C / Requirement ) * VSI
P = difference of P for two consecutive tests at different temperatures.
Temp. diff. = difference of temp for two consecutive tests at different temperatures.
Check if the zero-change per 5 °C is smaller than e (class II, III or IIII)
PASSED for 0,83 [µV / VSI] and also Passed for 0,04 [µV / VSI]
Observer: Jens HovgårdRemarks:
MV2
Passed/
Failed
-1 19,1 0,26 5,4 PASSED
PASSED
0 15,6 0 5,4 PASSED
-20,7 5,4
Aut. Zero-setting and zero-tracking device is
Minimum input-voltage per verification scale interval (∆umin)No-sense
Passed / Failed:
5,41 35,1 0,14
Issued by DELTA Z DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 37 of 67 DANAK-1913746
TEST: Repeatability (OIML R76: A.4.10)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 20,4 20,4Test item: Indicator Rel. H.[%]:Serial No.: E13613-0307 Time: 13:55Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]: 00:00Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
No load (L0): 0 [count] 0 [%] of the range
Load about half Max: (Calibrated values) Load about Max: (Calibrated values)
Load Add. Load Add.Load P E Load P E
L I0 I L = I - I0 L I0 I L = I - I0No. [count] [count] [count] [count] [count] [count] No. [count] [count] [count] [count] [count] [count]1 32500 0 32501 32501 1 1 65000 0 65002 65002 22 32500 0 32501 32501 1 2 65000 0 65002 65002 23 32500 0 32501 32501 1 3 65000 0 65002 65002 24 32500 0 32501 32501 1 4 65000 0 65002 65002 25 32500 0 32501 32501 1 5 65000 0 65002 65002 26 32500 0 32501 32501 1 6 65000 0 65002 65002 27 32500 0 32501 32501 1 7 65000 0 65002 65002 28 32500 0 32501 32501 1 8 65000 0 65002 65002 29 32500 0 32501 32501 1 9 65000 0 65002 65002 2
10 32500 0 32501 32501 1 10 65000 0 65002 65002 2
Emax: 1 [count] Emax: 2 [count]
Emin: 1 [count] Emin: 2 [count]
Emax - Emin: 0 [count] Emax - Emin: 0 [count]
pi = 0,5 Mpe: 8,1 [count] Mpe: 8,1 [count]
Passed / Failed: P Passed / Failed: P
Overall Passed / Failed: Passed
Observer:
Remarks:
02-09-2013
Jens Hovgård
Indication Indication
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Issued by DELTA Rep DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 38 of 67 DANAK-1913746
TEST: Tare "Weighing test" (OIML R76: A.4.6.1 & 3.5.3.3)Analysis based on Initial Intrinsic data.
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 21 21Test item: Indicator Rel. H.[%]: 51Serial No.: E13613-0307 Time: 14:30 15:00Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
Tare load applied: 6500,1 [count] Max. Tare: 100 %Indication of tare load: 6500 [count] Max. for analysis: 100 %
E = I + ½e - L - L Ec = E - E0 pi = 0,5
(Initial Intrinsic values) ↑ ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,000000
10 0,100000 0 0 0 0 0 0 0 2,7 2,720 0,200000 6500 6500 6500 0 0 0 0 5,4 5,430 0,300000 12999,9 13000 13000 0,1 0,1 0,1 0,1 5,4 5,440 0,400000 19499,9 19500 19500 0,1 0,1 0,1 0,1 5,4 5,450 0,500000 25999,9 26000 26000 0,1 0,1 0,1 0,1 8,1 8,160 0,600000 32500,1 32500 32500 -0,1 -0,1 -0,1 -0,1 8,1 8,170 0,700000 38999,6 39000 39000 0,4 0,4 0,4 0,4 8,1 8,180 0,800000 45499,6 45500 45500 0,4 0,4 0,4 0,4 8,1 8,190 0,900000 51999,7 52000 52000 0,3 0,3 0,3 0,3 8,1 8,1100 1,000000 58499,9 58500 58500 0,1 0,1 0,1 0,1 8,1 8,1
Passed / Failed: PASSEDRemarks:
Observer:
Add. Load
19-08-2013
MPE
* pi
Jens Hovgård
Ec
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
I L E
Error Corrected error
L
Load Indication of NET
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [c
ount
]
Load [count]
Error Curve for Tare = 6500,1 [count]
Up loading Down loading Mpe
Issued by DELTA Tare(A) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 39 of 67 DANAK-1913746
PASSED PASSED
PASSED PASSED
PASSED PASSED
PASSED PASSED
PASSED PASSEDPassed/Failed:
Passed/Failed:
Passed/Failed:
Passed/Failed:
Passed/Failed:
Passed/Failed:
Passed/Failed:
Passed/Failed:
Passed/Failed:
Passed/Failed:
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 6500,1 [count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 13000,1 [count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 19500 [count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 26000 [count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 32500 [count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 39000,2 [count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 45499,7 [count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 51999,7 [count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 58499,8 [count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve for Tare = 65000 [count]
Issued by DELTA Tare(A) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 40 of 67 DANAK-1913746
TEST: Tare "Weighing test" (OIML R76: A.4.6.1 & 3.5.3.3)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 20,3 20,4Test item: Indicator Rel. H.[%]:Serial No.: E13613-0307 Time: 13:40 13:48Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]: 1013,0Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004 The test is superfluousSimulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
1 load cell(s) 350 [Ohm]
Tare load applied: 19500 [count]Indication of tare load: 19501 [count]
E = I + ½e - L - L Ec = E - E0 pi = 0,5MPE P
* pi /(Initial Intrinsic values) ↑ ↓ ↑ ↓ ↑ ↓ ↑ F% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,00000010 0,10000020 0,20000030 0,300000 0 0 0 0 0 0 0 2,7 P40 0,400000 6500 6500 6500 0 0 0 0 5,4 P50 0,500000 13000 13000 13000 0 0 0 0 5,4 P60 0,600000 19500,2 19500 19500 -0,2 -0,2 -0,2 -0,2 5,4 P70 0,700000 25999,7 26000 26000 0,3 0,3 0,3 0,3 8,1 P80 0,800000 32499,7 32500 32501 0,3 1,3 0,3 1,3 8,1 P90 0,900000 38999,8 39001 39001 1,2 1,2 1,2 1,2 8,1 P100 1,000000 45500 45501 1 1 8,1 P
Passed / Failed: PASSEDObserver:Remarks:
Tare loading using Weights
Not applied: 7 Not applied: LoadNo. N/A N/A n
Tare: 19500 [count] 0 20 201 220 220
Max tare: 65000 [count] 2 500 5003 1000 10004 1500 15005 2000 20006 3000 3000
[count] 7 4000 4000[count][count]
Load used
Ec
Jens Hovgård
L I L E
02-09-2013
Indication of NET Add. Load Corrected error
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
ErrorLoad
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or
[co
un
t]
Load with tare 19500 [count]
Error Curve
Up loading Down loading Mpe
Issued by DELTA Tare(1) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 41 of 67 DANAK-1913746
TEST: Tare "Weighing test" (OIML R76: A.4.6.1 & 3.5.3.3)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 20,4 20,4Test item: Indicator Rel. H.[%]:Serial No.: E13613-0307 Time: 13:50 13:55Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]: 1013,0Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004 The test is superfluousSimulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
Tare load applied: 39000 [count]Indication of tare load: 39001 [count]
E = I + ½e - L - L Ec = E - E0 pi = 0,5MPE P
* pi /(Initial Intrinsic values) ↑ ↓ ↑ ↓ ↑ ↓ ↑ F% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,00000010 0,10000020 0,20000030 0,30000040 0,40000050 0,50000060 0,600000 0,2 0 -0,2 -0,2 2,7 P70 0,700000 6499,7 6501 6501 1,3 1,3 1,3 1,3 5,4 P80 0,800000 12999,7 13001 13001 1,3 1,3 1,3 1,3 5,4 P90 0,900000 19499,8 19501 19501 1,2 1,2 1,2 1,2 5,4 P100 1,000000 26000 26002 2 2 8,1 P
Passed / Failed: PASSEDObserver:Remarks:
Tare loading using Weights
Not applied: Not applied: LoadNo. N/A N/A n
Tare: 39000 [count]
Max tare: 65000 [count]
[count][count][count]
Load used
02-09-2013
Jens Hovgård
Error Corrected error
Ec
Aut. Zero-setting and zero-tracking device is:
E
Load cell number and input impedance:
Load Indication of NET Add. Load
L I L
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 10000 20000 30000 40000 50000 60000
Err
or [c
ount
]
Load with tare 39000 [count]
Error Curve
Up loading Down loading Mpe
Issued by DELTA Tare(2) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 42 of 67 DANAK-1913746
TEST: Warm-up time (OIML R76: A.5.2)WELMEC 2.1
Application / task: T205886 Begin EndType designation: Temp.[°C]: 20,1 20,3Test item: Indicator Rel. H.[%]:Serial No.: E13613-0307 Time: 13:00 13:38Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]: 1013Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
1 load cell(s) 350 [Ohm]
Duration of power off before the test: hours
Time from power on until the first measurement is readible: 0,1 minutes
E = I + ½e - L - L Ec = EL - E0 pi = 0,5
Corrected mpe = Perror MPE /
Lapse Real L0 L I0 IL E0 EL Ec * pi F
mm:ss hh:mm:ss [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 4 4
65000 65005 5 1 8,1 P0
650000
650000
650000
650000 0 0
65000 65004 4 4 8,1 P0
650000
650000
650000
650000
650000
650000
650000
650000
650000 0 0
65000 65003 3 3 8,1 P0
650000
650000
650000
650000
650000
650000
650000
650000
650000
650000
650000
650000
650000
650000 0 0
65000 65002 2 2 8,1 P(*) Counted from the moment an indication has first appeared.
13:28:00
13:29:00
13:30:00
30:00
13:31:00
13:33:00
13:34:00
13:27:00
13:32:00
13:18:00
13:24:00
13:25:00
13:26:00
13:23:00
13:19:00
13:20:00
02-09-2013
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Error
13:14:00
13:07:00
13:08:00
13:09:00
13:10:00
13:11:00
13:12:00
13:15:00
E
13:16:00
13:17:00
MV2
13:13:00
13:21:00
13:22:00
Time (*)
5:00
0:00
Add. LoadL
IndicationI
Load
15:00
13:35:00
13:36:00
13:37:00
Issued by DELTA W DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 43 of 67 DANAK-1913746
Passed / Failed: Passedmpe P / F
Initial zero-setting error: E0Init 4 2,7 (F) No requirement,
Maximum value of error unloaded: E0 4 2,7 (F) for information only.
Maximum span error: EL - E0 4 8,1 P
Observer:Remarks:
Jens Hovgård
[count]
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 5 10 15 20 25 30 35
Err
or
vari
atio
n o
f E
L -
Eo
[co
un
t]
Time [minutes]
Warm-up time
mpe Ec Ec stab.
-4
-3
-2
-1
0
1
2
3
4
5
0 5 10 15 20 25 30
Err
or
vari
atio
n
of
Eo
[co
un
t]
Time [minutes]
Warm-up time
Eo Limit Eo
Issued by DELTA W DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 44 of 67 DANAK-1913746
TEST: Voltage variations (OIML R76: A.5.4 & 3.9.3)WELMEC 2.1
Application / task: T205886 Begin EndType designation: Temp.[°C]: 20,1Test item: Indicator Rel. H.[%]: 52Serial No.: E13613-0307 Time: 12:23Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
AC Power voltage supply: Un: [Vac]
Marked AC voltage range: Umin - Umax: 100 - 240 [Vac]
E = I + ½e - L - L Ec = E - E0 pi = 1
MPE PU /
L0 L I0 IL E0 E F
[Vac] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 1 1 0 5,4 P
65000 65004 4 16,2 P0 0 0 -1 5,4 P
65000 65003 3 16,2 P0 0 0 -1 5,4 P
65000 65003 3 16,2 P0 0 0 -1 5,4 P
65000 65003 3 16,2 PMaximum value: 3
Passed / Failed: Passed
Remarks:
DC Battery supply: Un: 9 [Vdc] Time: 13:20
Marked DC voltage range: Umin - Umax: - [Vdc] Date:
Lowest limit of voltage for indication: 5,98 [Vdc]
E = I + ½e - L - L Ec = E - E0 pi = 1
MPE PU /
L0 L I0 IL E0 E F
[Vdc] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0 0 5,4 P
65000 65001 1 16,2 P0 0 0 5,4 P
65000 65001 1 16,2 P0 0 0 5,4 P
65000 65001 1 16,2 P0 0 0 5,4 P
65000 65001 1 16,2 P0
65000Maximum value: 1
Passed / Failed: Passed
Remarks:
Observer:
Overall Passed/Failed: Passed
1
0
01
1
0
0
9
11,7
1
2
9
5,98
Measured on battery connection terminals
1
2
3
170
85
264
03-09-2013
Corrected errorErrorLoad Indication Add. Load
MV2
LI E Ec
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Battery supply (6 x 1,.5V rechargeable batteries)
Jens Hovgård
[count]
Corrected errorE
Error(Initial Intrinsic values) I
3
[count]
3
3
3
4 170
03-09-2013
Load Add. LoadIndicationL
4
Ec
5
3
1
Issued by DELTA Var DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 45 of 67 DANAK-1913746
TEST: Electrical Disturbances (OIML R76: B.3)WELMEC 2.1
Application / task: T205886
Type designation: MV2Test item: IndicatorSerial No.: E13613-0307Accuracy class: IIIMax: 65000 [count]Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count] Simulator applied:
Test load: See the respective tests.Input-voltage per VSI (∆umin): 0,83 [µV] Later changed to 0,83 [µV]
Note: See "Set(T)" page for the configuration of the test setup.
1 load cell(s) corresponding to 350 [Ohm]
Summary of resultsSection Test P/F
[count] [µV] [count] [µV]B.3.1 Short time power reductions 1 0,0766 16,2 1,242 PB.3.2 Electrical bursts
a) Power supply lines: 1 [kV] 1 0,0766 16,2 1,242 Pb) I/O circuits and data lines: 0,5 [kV] 1 0,0766 16,2 1,242 P
B.3.3 Electrostatic discharges(B.3.4) a) Direct application: 0 0 16,2 1,242 P
b) Indirect application: 0 0 16,2 1,242 PB.3.4 Immunity to electromagnetic fields (emc)(B.3.5) a) Radiated Field strength: 10 [V / m] 7 0,5365 16,2 1,242 P(B.3.6) b) Conducted Level 10 [V ] 9 0,6898 16,2 1,242 P(B.3.3) Surge
a) Power supply lines: 1 [kV] 0 0 16,2 1,242 Pb) I/O circuits and data lines: - [kV] 16,2 1,242
(B.3.7) Electrical transients, road vehiclesa) Conduction along supply lines: 24 [V battery] 16,2 1,242b) Capacitive and inductive couplin 12 [V battery] 16,2 1,242
Overall: 9 0,6898 16,2 1,242 P
Passed / Failed: Passed for 0,83 [µV / VSI] and also Passed for 0,48 [µV / VSI]
Test location:Test laboratory:EMC report no.:Date of EMC report:Responsible person:Remarks:
Load cell number and input impedance:
RemarksMax. Error MPE
Issued by DELTA EMC DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 46 of 67 DANAK-1913746
TEST: Short time power reductions (OIML R76:2006 B.3.1)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 23Test item: Indicator Rel. H.[%]:Serial No.: E13613-0307 Time: 13:16Accuracy class: III Date:Max: 65000 [count]Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Load cell applied
Input-voltage per VSI (∆umin): 0,83 [µV]
Note: See "Set(T)" page for the configuration of the test setup.
Load cell number and input impedance: 1 load cell(s) 350 [Ohm]
Marked nominal voltage Un or voltage range: 100 - 240 [V]
Load: 40 [g]
No disturbance Disturbance ResultIndication Amplitude Duration Number of Repetition Indication Error Remarks
I1 % of Un cycles disturbances interval I2 I2-I1[count] [s] [count] [count]39545 0 0,5 10 10 39544 -139544 50 1 10 10 39544 039544 0 1 10 10 39544 039544 40 10 10 10 39544 039545 70 25 10 10 39544 -139544 80 250 10 10 39545 139545 0 250 10 10 39545 0 normal start-up
/Max. Error/: 1Passed / Failed: Passed
Tested in accordance with the method specified in OIML R76
Observer: CMTRemarks:
04-09-2013
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 47 of 67 DANAK-1913746
TEST: Electrical bursts (OIML R76:2006 B.3.2)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 25Test item: Indicator Rel. H.[%]: 49Serial No.: E13613-0307 Time: 15:37Accuracy class: III Date:Max: 65000 [count] Bar.pres.: [hPa]Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Load cell applied
Input-voltage per VSI (∆umin): 0,83 [µV]
Note: See "Set(T)" page for the configuration of the test setup.
Load cell number and input impedance: 1 load cell(s) 350 [Ohm]Load: 40 [g]
a) Power supply lines
Power supply lines: test voltage : 1 kV, duration of the test 1 minute at each polarity
No disturbance Disturbance Polarity ResultIndication phase neutral protec.earth Indication Error Remarks
I1 I2 I2-I1[count] ground ground ground [count] [count]39607 X pos 39607 039607 X neg 39607 039607 X pos 39607 039607 X neg 39607 039607 X pos 39607 039607 X neg 39606 -1
Max. Error: 1Passed / Failed: Passed
Tested in accordance with the method specified in OIML R76
Observer: CMTRemarks:
b) I/O circuits and communication lines.
I/O signals, data and control lines: test voltage : 0,5 kV, duration of the test 1 minute at each polarity
No disturbance Cable/Interface Polarity ResultIndication Indication Error
I1 I2 I2-I1[count] [count] [count]39602 Connection to load cell pos 39602 039602 < 3 m, shielded neg 39601 -139601 RS-232C cable pos 39601 039601 < 3 neg 39601 039601 USB cable pos 39601 039601 1,8 neg 39601 039601 Digital Hight Rod I/F cable pos 39601 039601 1,8 neg 39601 0
Max. Error: 1Passed / Failed: Passed
Tested in accordance with the method specified in OIML R76
Observer: CMTRemarks:
04-09-2013
Remarks
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 48 of 67 DANAK-1913746
TEST: Electrical Surge (OIML R76:2006 B.3.3)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 21Test item: Indicator Rel. H.[%]: 37Serial No.: E13613-0307 Time: 15:00Accuracy class: III Date:Max: 65000 [count] Bar.pres.: [hPa]Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Load cell applied
Input-voltage per VSI (∆umin): 0,83 [µV]
Note: See "Set(T)" page for the configuration of the test setup.
Load cell number and input impedance: 1 load cell(s) 350 [Ohm]Load: 40 [g]a) Power supply lines
Power supply lines: 3 positive and 3 negative surges
No disturbance Disturbance Polarity ResultIndication Applied on amplitude angle Indication Error Remarks
I1 I2 I2-I1[count] [kV] [°] [count] [count]
39680 L - N 0,5 0° 90° 180° 270° pos 39680 039680 L - N 0,5 0° 90° 180° 270° neg 39680 039680 L - PE 1 0° 90° 180° 270° pos 39680 039680 L - PE 1 0° 90° 180° 270° neg 39680 039680 N - PE 1 0° 90° 180° 270° pos 39680 039680 N - PE 1 0° 90° 180° 270° neg 39680 0
Max. Error: 0Passed / Failed: Passed
Tested in accordance with the method specified in OIML R76
Observer: MINRemarks:
b) I/O circuits and communication lines.
I/O signals, data and control lines: Only for industri and cables > 30m, test voltage 0.5 kV.
No disturbance Cable/Interface Test voltage Polarity ResultIndication Indication Error
I1 I2 I2-I1[count] [kV] [count] [count]
Connection to load cell pos not applicable< 3 m, shielded neg not applicable
RS-232C cable pos not applicable< 3 neg not applicable
USB cable pos not applicable1,8 neg not applicable
Digital Hight Rod I/F cable pos not applicable1,8 neg not applicable
Max. Error:Passed / Failed:
Tested in accordance with the method specified in OIML R76
Observer:Remarks:
18-11-2013
Remarks
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 49 of 67 DANAK-1913746
TEST: Electrostatic discharges (OIML R76:2006 B.3.4)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 21Test item: Indicator Rel. H.[%]: 37Serial No.: E13613-0307 Time: 14:00Accuracy class: III Date:Max: 65000 [count] Bar.pres.: [hPa]Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Load cell applied
Input-voltage per VSI (∆umin): 0,83 [µV]
Note: See "Set(T)" page for the configuration of the test setup.
Load cell number and input impedance: 1 load cell(s) 350 [Ohm]Load: 40 [g]
a) Direct application.
Contact discharges x Paint penetration
Air discharges x Polarity(*): pos x neg x(*) IEC 801-2 specifies that the test shall be conducted in the most sensitive polarity
No disturbance Disturbance ResultIndication Test Number of Repetition Indication Error Remarks
I1 Voltage discharges interval I2 I2-I1[count] [kV] ≥10 [s] [count] [count]39680 2 10 10 39680 039680 4 10 10 39680 039680 6 10 10 39680 039680 8 (air discharge) 10 10 39680 0
Max. Error: 0Passed / Failed: Passed
Tested in accordance with the method specified in OIML R76
Observer: MINRemarks:
b) Indirect application (contact discharges only)
Polarity(*): pos x neg x(*) IEC 801-2 specifies that the test shall be conducted in the most sensitive polarity
Horizontal coupling planeNo disturbance Disturbance Result
Indication Test Number of Repetition Indication Error RemarksI1 Voltage discharges interval I2 I2-I1
[count] [kV] ≥10 [s] [count] [count]39680 2 10 10 39680 039680 4 10 10 39680 039680 6 10 10 39680 0
Vertical coupling planeNo disturbance Disturbance Result
Indication Test Number of Repetition Indication Error RemarksI1 Voltage discharges interval I2 I2-I1
[count] [kV] ≥10 [s] [count] [count]39680 2 10 10 39680 039680 4 10 10 39680 039680 6 10 10 39680 0
Max. Error: 0Passed / Failed: Passed
Tested in accordance with the method specified in OIML R76
Observer: MINRemarks:
18-11-2013
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 50 of 67 DANAK-1913746
TEST: Radiated electromagnetic fields (OIML R76:2006 B.3.5)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 22Test item: Indicator Rel. H.[%]: 54Serial No.: E13613-0307 Time: 11:05Accuracy class: III Date:Max: 65000 [count] Bar.pres.: [hPa]Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Load cell applied
Input-voltage per VSI (∆umin): 0,83 [µV]
Note: See "Set(T)" page for the configuration of the test setup.
Load cell number and input impedance: 1 load cell(s) 350 [Ohm]Load: 20 [g]Size of test item: Height: 175 [mm] Width: 195 Depth: 45 [mm]Field strength: 10 [V/m]Antenna test distance: 2,5 [m]Rate of sweep: 0,005 [oct./s]Material load:Modulation: 80% AM, 1 kHz sine wave
Antenna: BilogNo disturbance Disturbance Result
Indication Frequency Polariza- Field lowered Antenna Indication Error RemarksI1 range tion to comply facing I2 I2-I1
[count] [Mhz] [V/m] EUT [count] [count]39529 Front 39536 739528 Right 39533 539528 Vertical Left 39534 639528 80-1000 Rear 39533 539526 Front 39533 739526 Right 39531 539527 Horizontal Left 39533 639528 Rear 39533 5
Antenna: HornNo disturbance Disturbance Result
Indication Frequency Polariza- Field lowered Antenna Indication Error RemarksI1 range tion to comply facing I2 I2-I1
[count] [Mhz] [V/m] EUT [count] [count]39523 Front 39524 139524 Right 39524 039524 Vertical Left 39524 039524 1000-2000 Rear 39525 139521 Front 39522 139522 Right 39522 039522 Horizontal Left 39522 039522 Rear 39523 1
Max. Error: 7Passed / Failed: Passed
Tested in accordance with the method specified in OIML R76
Observer: CMTRemarks:
04-09-2013
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 51 of 67 DANAK-1913746
TEST: Conducted electromagnetic fields (OIML R76:2006 B.3.6)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 21Test item: Indicator Rel. H.[%]: 37Serial No.: E13613-0307 Time: 13:00Accuracy class: III Date:Max: 65000 [count] Bar.pres.: [hPa]Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Load cell applied
Input-voltage per VSI (∆umin): 0,83 [µV]
Note: See "Set(T)" page for the configuration of the test setup.
Load cell number and input impedance: 1 load cell(s) 350 [Ohm]Load: 40 [g]Level: 10 [Vrms]Rate of sweep: 0,005 [oct./s]Modulation: 80% AM, 1 kHz sine wave
No disturbance Disturbance ResultIndication Cable / interface Frequency Level Indication Error Remarks
I1 range lowered to I2 I2-I1[count] [Mhz] [V] [count] [count]39681 Power supply cable 0.15 - 80 39672 -9
Connection to load cell 0.15 - 80 N.A., < 3 mRS-232C cable 0.15 - 80 N.A., < 3 mUSB cable 0.15 - 80 N.A., < 1.8 mDigital Hight Rod I/F cable 0.15 - 80 N.A., < 1.8 m
Max. Error: 9Passed / Failed: Passed
Tested in accordance with the method specified in OIML R76
Observer: MINRemarks:
18-11-2013
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 52 of 67 DANAK-1913746
TEST: Electrical transients from a road vehicle power supply(OIML R76:2006 B.3.7)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]:Test item: Indicator Rel. H.[%]:Serial No.: E13613-0307 Time:Accuracy class: III Date:Max: 65000 [count] Bar.pres.: [hPa]Verification scale interval, e: 10,83 [count]No. of verification scale intervals, n: 6000Resolution during test: 1 [count]Load cell applied
Input-voltage per VSI (∆umin): 0,83 [µV]
Note: See "Set(T)" page for the configuration of the test setup.
Load cell number and input impedance: 1 load cell(s) 350 [Ohm]Load: 40 [g]
a) Conduction along supply lines of external 12 V and 24 V batteries (ISO 7637-2:2004)
12 V battery voltageNo disturbance Disturbance Result
Indication Cable / interface Test Conducted Indication Error RemarksI1 pulse voltage I2 I2-I1
[count] [V] [count] [count]2a +502b +103a -1503b +1004 -75 +87
24 V battery voltageNo disturbance Disturbance Result
Indication Cable / interface Test Conducted Indication Error RemarksI1 pulse voltage I2 I2-I1
[count] [V] [count] [count]2a +502b +203a -2003b +2004 -165 +173
Max. Error:Passed / Failed:
Tested in accordance with the method specified in OIML R76
Observer:Remarks:
Power supply cable
Power supply cable
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 53 of 67 DANAK-1913746
b) Capacitive and inductive coupling via lines other than supply lines (ISO 7637-3:2007)
12 V battery voltageNo disturbance Disturbance Result
Indication Test Conducted Indication Error RemarksI1 pulse voltage I2 I2-I1
[count] [V] [count] [count]a -60b +40a -60b +40a -60b +40a -60b +40a -60b +40
24 V battery voltageNo disturbance Disturbance Result
Indication Test Conducted Indication Error RemarksI1 pulse voltage I2 I2-I1
[count] [V] [count] [count]a -80b +80a -80b +80a -80b +80a -80b +80a -80b +80
Max. Error:Passed / Failed:
Tested in accordance with the method specified in OIML R76
Observer:Remarks:
Conditional Passed:
RS-232C cable
USB cable
RS-232C cable
USB cable
Digital Hight Rod I/F cable
Ethernet cable
Kind or type of other lines
Digital Hight Rod I/F cable
Ethernet cable
Kind or type of other lines(no power supply lines)
(no power supply lines)
Connection to load cell
Connection to load cell
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 54 of 67 DANAK-1913746
Photographs from the electrical disturbance tests.
Radiated electromagnetic susceptibility Radiated electromagnetic susceptibility
Bursts Voltage dips & interruptions
Conducted electromagnetic susceptibility
Remarks:
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 55 of 67 DANAK-1913746
Calibration of the electromagnetic field.
Before the test item is placed in the anechoic chamber, an isotropic E-field sensor is placed on the spot where the test item is placed later on. An automatic measurement of the generator output, necessary to establish the specified field strengthreading on the E-field sensor is performed. The data of the corresponding frequency and generator output is stored for use during the exposure. These data are printed out, saved and kept.
The frequency range is scanned at a maximum rate of 0.005 oct./sec. (approx. 1 octave/2 minutes) while the field strengthis maintained using the generator settings achieved during the previous measurement with the E-field sensor installedinstead of the test object.
Remarks:
List of instrumentation.
No. Description Manufacturer Type No.29342 ZPD 29646 8308-200-N 29691 83620A 29694 AN 8200 F 29744 BN 52 76 76 29751 EFT 50029898 M329844 SAMPLER_VER_229846 202429856 34401A29859 6842A29914 VCS 50029984 AS0822-2002998549000 NRVS49001 NRV-Z5149009 RG 214/U49173 202449518 SHIELD-LINE CDN 49521 SHIELD-LINE CDN 49528 8471D49562 NSG43849617 CBA1G-500
REFLECTOMETER COUPLER, 600-4200
RF DIRECTIONAL COUPLER, 26-1000 Electrical fast transient (burst) generator3-LINE CDN NETWORK, IEC 61000-4-6 -40 dBc VOLTAGE SAMPLER, DC-100 MHz RF generator, 9 kHz-2.4 GHz
HEWLETT-PACKARD LOGIMETRICS 1-12 GHz. HORN ANTENNA.
PWR. ATT. 20 dB, 75W, DC-2GHz. 0.01 - 20 GHz. SYNTH. SWEEPER
6140A
THERMAL POWER SENSOR, DC-18 GHzCABLE#36, 3,25 M, 50 OHM, Coax cableRF GENERATOR, 9 kHz-2.4 GHz
AC source w. harmonic/flicker test optionELECTRONIC SURGE GENERATORRF POWER AMPLIFIER, 0.8-2.2 GHz, 200W
SPINNER EM TESTMEBDELTA EMC DEPT. Marconi
MARCONI
ROHDE & SCHWARZ BIRD ELECTRONIC
BILOG ANTENNA 26-2000 MHzSINGLE CHANNEL POWER METER
HIGH POWER RF AMPLIFIER, 80-1000
HEWLETT-PACKARDHewlett-PackardEM TESTMILMEGASCHAFFNER/CHASEROHDE & SCHWARZ ROHDE & SCHWARZ SUHNER
DIGITAL MULTIMETER W. HPIB
TESEQ
SHIELD-LINE CDN NETWORK, IEC 61000-4-SHIELD-LINE CDN NETWORK, IEC 61000-4-COAX RF DIODE DETECTOR, NEG.
SHAFFNERESD GENERATOR, AIR AND CONTACT
DELTA EMC DEPT. DELTA EMC DEPT. AGILENT
Issued by DELTA EMC(T) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 56 of 67 DANAK-1913746
TEST: Damp heat, steady state (OIML R76: B.2.2)WELMEC 2.1 Begin Max
Application / task: T205886 Room temp.[°C]: 20
Type designation: MV2 Temp.[°C]: 20
Test item: Indicator Rel. H.[%]: 50Serial No.: E13613-0307 Time: 14:30Accuracy class: III Date:Max: 65000 [count] Ref. Test before humidity:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count] Simulator applied: HBM, K3608, S/N: 90220004 Placed outside the climatic chamber.Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Out of working range
1 load cell(s) 350 [Ohm]
Length of cable outside the climatic chamber: 2.5 [m] 4-wires
E = I + ½e - L - L Ec = E - E0 pi = 0,5
MPE P* pi /
↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ F% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,000000 0 0 0 0 2,7 P
10 0,100000 6500 6499 -1 -1 5,4 P20 0,200000 13000 12999 -1 -1 5,4 P30 0,300000 19500 19498 -2 -2 5,4 P40 0,400000 26000 25997 -3 -3 8,1 P50 0,500000 32500 32496 -4 -4 8,1 P60 0,600000 39000 38995 -5 -5 8,1 P70 0,700000 45500 45495 -5 -5 8,1 P80 0,800000 52000 51994 -6 -6 8,1 P90 0,900000 58500 58493 -7 -7 8,1 P100 1,000000 65000 64993 -7 -7 8,1 P
Passed / Failed: PASSED
Observer:Remarks:
06-09-201
LCorrected errorError
1
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Jens Hovgård
I LIndication Add. Load
E Ec
(Initial Intrinsic values)
Load
-10-8-6-4-202468
10
0 10000 20000 30000 40000 50000 60000
Err
or [c
ount
]
Load [count]
Error Curve
Up loading Down loading Mpe
Issued by DELTA H(1) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 57 of 67 DANAK-1913746
TEST: Damp heat, steady state (OIML R76: B.2.2)WELMEC 2.1 Begin Max
Application / task: T205886 Room temp.[°C]: 20
Type designation: MV2 Temp.[°C]: 40
Test item: Indicator Rel. H.[%]: 45Serial No.: E13613-0307 Time: 08:40Accuracy class: III Date:Max: 65000 [count] Test during humidity:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004 Placed outside the climatic chamber.Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Out of working range
1 load cell(s) 350 [Ohm]
Length of cable outside the climatic chamber: 2.5 [m] 4-wires
E = I + ½e - L - L Ec = E - E0 pi = 0,5
MPE P* pi /
↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ F% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,000000 0 -1 -1 -1 -1 0 0 2,7 P
10 0,100000 6500 6496 6496 -4 -4 -3 -3 5,4 P20 0,200000 13000 12995 12995 -5 -5 -4 -4 5,4 P30 0,300000 19500 19494 19494 -6 -6 -5 -5 5,4 P40 0,400000 26000 25993 25993 -7 -7 -6 -6 8,1 P50 0,500000 32500 32499 32499 -1 -1 0 0 8,1 P60 0,600000 39000 38991 38991 -9 -9 -8 -8 8,1 P70 0,700000 45500 45491 45491 -9 -9 -8 -8 8,1 P80 0,800000 52000 51991 51991 -9 -9 -8 -8 8,1 P90 0,900000 58500 58491 58491 -9 -9 -8 -8 8,1 P100 1,000000 65000 64991 64991 -9 -9 -8 -8 8,1 P
Passed / Failed: PASSED
Observer:Remarks:
(Initial Intrinsic values)
Jens Hovgård
2 days
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
09-09-201
ErrorE
Add. Load Corrected errorLoadL I L
IndicationEc
-10-8-6-4-202468
10
0 10000 20000 30000 40000 50000 60000
Err
or [c
ount
]
Load [count]
Error Curve
Up loading Down loading Mpe
Issued by DELTA H(2) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 58 of 67 DANAK-1913746
TEST: Damp heat, steady state (OIML R76: B.2.2)WELMEC 2.1 Begin Max
Application / task: T205886 Room temp.[°C]: 20Type designation: MV2 Temp.[°C]: 20Test item: Indicator Rel. H.[%]: 50Serial No.: E13613-0307 Time: 15:00Accuracy class: III Date:Max: 65000 [count] Test after humidity:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004 Placed outside the climatic chamber.Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Out of working range
1 load cell(s) 350 [Ohm]
Length of cable outside the climatic chamber: 2.5 [m] 4-wires
E = I + ½e - L - L Ec = E - E0 pi = 0,5
MPE P* pi /
↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ F% mV/V [count] [count] [count] [count] [count] [count] [count] [count] [count] [count]0 0,00000 0 0 0 0 0 0 0 2,7 P10 0,10000 6500 6499 6499 -1 -1 -1 -1 5,4 P20 0,20000 13000 12999 12999 -1 -1 -1 -1 5,4 P30 0,30000 19500 19498 19498 -2 -2 -2 -2 5,4 P40 0,40000 26000 25998 25998 -2 -2 -2 -2 8,1 P50 0,50000 32500 32497 32497 -3 -3 -3 -3 8,1 P60 0,60000 39000 38996 38996 -4 -4 -4 -4 8,1 P70 0,70000 45500 45496 45496 -4 -4 -4 -4 8,1 P80 0,80000 52000 51995 51995 -5 -5 -5 -5 8,1 P90 0,90000 58500 58495 58495 -5 -5 -5 -5 8,1 P
100 1,00000 65000 64994 -6 -6 8,1 P
Passed / Failed: PASSEDObserver:Remarks:
(Initial Intrinsic values)
Jens Hovgård
3
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
09-09-201
ErrorE
Add. Load Corrected errorLoadL I L
IndicationEc
-10-8-6-4-202468
10
0 10000 20000 30000 40000 50000 60000
Err
or [
coun
t]
Load [count]
Error Curve
Up loading Down loading Mpe
Issued by DELTA H(3) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 59 of 67 DANAK-1913746
TEST: Span Stability (OIML R76: B4)WELMEC 2.1
Application / task: T205886 Begin EndType designation: Temp.[°C]: 21,0Test item: Indicator Rel. H.[%]: 51Serial No.: E13613-0307 Time: 15:00Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]: 1011Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
Span measurement No: 1
Conditions of the measurement: Reference condition
E = I + ½e - ∆L - L Ec = E - E0 pi = 1MPE P
/L0 L I0 IL E0 E F
[count] [count] [count] [count] [count] [count] [count] [count] [count]0 0 0
65000 65000 0 16,2 P0 0 0
65000 65000 0 16,2 P0 0 0
65000 65000 0 16,2 P0 0 0
65000 65000 0 16,2 P0 0 0
65000 65000 0 16,2 P
Result: Average error: 0 [count]Ecmax - Ecmin: 0 [count]
0.1e = 1,1 [count]Note: One loading is sufficient for the remaining span measurements.
Passed / Failed: Passed
Observer:
Remarks:
MV2
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
ErrorEI ∆L
Indication(Calibrated values)
19-08-2013
Corrected error
jbn
4
Add. LoadLoadEc
0
0
1
5
[count]
0
0
0
2
3
Issued by DELTA SP(1) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 60 of 67 DANAK-1913746
TEST: Span Stability (OIML R76: B4)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 20,7Test item: Indicator Rel. H.[%]: 52Serial No.: E13613-0307 Time: 15:50Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
Span measurement No: 2
Conditions of the measurement: After temperature test
E = I + ½e - ∆L - L Ec = E - E0 pi = 1MPE P
/L0 L I0 IL E0 E F
[count] [count] [count] [count] [count] [count] [count] [count] [count]0 0 0
65000 65000 0 16,2 P
Result: Average error: 0 [count]
Note: One loading is sufficient for the remaining span measurements.
Passed / Failed: Passed
Observer:
Remarks:
EcE
Jens Hovgård
I ∆L
5
[count]
01
4
2
(Calibrated values)Error Corrected error
3
22-08-2013
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Indication Add. LoadLoad
Issued by DELTA SP(2) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 61 of 67 DANAK-1913746
TEST: Span Stability (OIML R76: B4)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 19,9Test item: Indicator Rel. H.[%]: 50Serial No.: E13613-0307 Time: 11:00Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
Span measurement No: 3
Conditions of the measurement: No change of condition
E = I + ½e - ∆L - L Ec = E - E0 pi = 1MPE P
/L0 L I0 IL E0 E F
[count] [count] [count] [count] [count] [count] [count] [count] [count]0 0 0
65000 65001 1 16,2 P
Result: Average error: 1 [count]
Note: One loading is sufficient for the remaining span measurements.
Passed / Failed: Passed
Observer:
Remarks:
27-08-2013
Corrected errorEc
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
ErrorEI ∆L
Indication Add. LoadLoad(Calibrated values)
Jens Hovgård
5
[count]
1
2
3
1
4
Issued by DELTA SP(3) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 62 of 67 DANAK-1913746
TEST: Span Stability (OIML R76: B4)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 20,4Test item: Indicator Rel. H.[%]: 50Serial No.: E13613-0307 Time: 14:02Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]: 1013Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
Span measurement No: 4
Conditions of the measurement: After power interruption
E = I + ½e - ∆L - L Ec = E - E0 pi = 1MPE P
/L0 L I0 IL E0 E F
[count] [count] [count] [count] [count] [count] [count] [count] [count]0 0 0
65000 65002 2 16,2 P
Result: Average error: 2 [count]
Note: One loading is sufficient for the remaining span measurements.
Passed / Failed: Passed
Observer:
Remarks:
5
∆L EcIndication Add. Load
[count]
2
Jens Hovgård
02-09-2013
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
1
2
Corrected error
3
ErrorE
Load(Calibrated values)
4
I
Issued by DELTA SP(4) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 63 of 67 DANAK-1913746
TEST: Span Stability (OIML R76: B4)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 20,0Test item: Indicator Rel. H.[%]: 50Serial No.: E13613-0307 Time: 14:40Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]: 1013Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
Span measurement No: 5
Conditions of the measurement: No change of condition
E = I + ½e - ∆L - L Ec = E - E0 pi = 1MPE P
/L0 L I0 IL E0 E F
[count] [count] [count] [count] [count] [count] [count] [count] [count]0 0 0
65000 64995 -5 16,2 P
Result: Average error: -5 [count]
Note: One loading is sufficient for the remaining span measurements.
Passed / Failed: Passed
Observer:
Remarks:
06-09-2013
Corrected errorEc
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
ErrorEI ∆L
Indication Add. LoadLoad(Calibrated values)
Jens Hovgård
5
[count]
-5
2
3
1
4
Issued by DELTA SP(5) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 64 of 67 DANAK-1913746
TEST: Span Stability (OIML R76: B4)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 20,0Test item: Indicator Rel. H.[%]: 50Serial No.: E13613-0307 Time: 15:05Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]: 1006Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Out of working range
4 load cell(s) 87,5 [Ohm]
Span measurement No: 6
Conditions of the measurement: After humidity test
E = I + ½e - ∆L - L Ec = E - E0 pi = 1MPE P
/L0 L I0 IL E0 E F
[count] [count] [count] [count] [count] [count] [count] [count] [count]0 0 0
65000 64994 -6 16,2 P
Result: Average error: -6 [count]
Note: One loading is sufficient for the remaining span measurements.
Passed / Failed: Passed
Observer:
Remarks:
EcE
Jens Hovgård
I ∆L
5
[count]
-61
4
2
(Calibrated values)Error Corrected error
3
09-09-2013
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Indication Add. LoadLoad
Issued by DELTA SP(6) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 65 of 67 DANAK-1913746
TEST: Span Stability (OIML R76: B4)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 19,8Test item: Indicator Rel. H.[%]: 50Serial No.: E13613-0307 Time: 08:50Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Out of working range
4 load cell(s) 87,5 [Ohm]
Span measurement No: 7
Conditions of the measurement: After power interruption
E = I + ½e - ∆L - L Ec = E - E0 pi = 1MPE P
/L0 L I0 IL E0 E F
[count] [count] [count] [count] [count] [count] [count] [count] [count]0 0 0
65000 65000 0 16,2 P
Result: Average error: 0 [count]
Note: One loading is sufficient for the remaining span measurements.
Passed / Failed: Passed
Observer:
Remarks:
10-09-2013
Corrected errorEc
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
ErrorEI ∆L
Indication Add. LoadLoad(Calibrated values)
Jens Hovgård
5
[count]
0
2
3
1
4
Issued by DELTA SP(7) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 66 of 67 DANAK-1913746
TEST: Span Stability (OIML R76: B4)WELMEC 2.1
Application / task: T205886 Begin EndType designation: MV2 Temp.[°C]: 19,7Test item: Indicator Rel. H.[%]: 48Serial No.: E13613-0307 Time: 08:15Accuracy class: III Date:Max: 65000 [count] Bar.pres.[hPa]:Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
Not in operation
4 load cell(s) 87,5 [Ohm]
Span measurement No: 8
Conditions of the measurement: No change of condition
E = I + ½e - ∆L - L Ec = E - E0 pi = 1MPE P
/L0 L I0 IL E0 E F
[count] [count] [count] [count] [count] [count] [count] [count] [count]0 0 0
65000 65000 0 16,2 P
Result: Average error: 0 [count]
Note: One loading is sufficient for the remaining span measurements.
Passed / Failed: Passed
Observer:
Remarks:
EcE
Jens Hovgård
I ∆L
5
[count]
01
4
2
(Calibrated values)Error Corrected error
3
11-09-2013
Aut. Zero-setting and zero-tracking device is:
Load cell number and input impedance:
Indication Add. LoadLoad
Issued by DELTA SP(8) DTPR76-v5#3d.xlsm
Annex for report R76 - Report page 67 of 67 DANAK-1913746
TEST: Span Stability (OIML R76: B4 & WELMEC 2.1)Summary
Application / task: T205886
Type designation:
Test item: IndicatorSerial No.: E13613-0307Accuracy class: IIIMax: 65000 [count]Verification scale interval, e: 10,83 [count] 1 interval selectedNo. of verification scale intervals, n: 6000Resolution during test: 1 [count]Simulator applied: HBM, K3608, S/N: 90220004Simulator range: 1 [mV/V], 0 [%] to 100 [%]
4 load cell(s) corresponding to 87,5 [Ohm]
Result of span stability testpi = 1
Error Variation mpe[count] [count] [count]
0 6 16,20 6 16,2
1 7 16,2
2 8 16,2-5 1 16,2
-6 0 16,20 6 16,20 6 16,2
8 All:
pi = 1
Maximum permissable difference, mpd: 8,1 [count]
Passed / Failed: Passed
Remarks:
MV2
PNo change of conditionP
After power interruption11-09-2013
P
Maximum - minimum =8
Passed /
P
P
Failed
P
P
P
P
Load cell number and input impedance:
12
3
Measurementno.
Condition
No change of condition
Reference condition
Date of test
7After humidity test
No change of condition
09-09-201310-09-2013
45
6
After temperature test
After power interruption
22-08-2013
27-08-2013
02-09-201306-09-2013
19-08-2013
0
1
2
3
4
5
6
7
8
9
1 2 3 4 5 6 7 8
Err
or
vari
atio
n [
cou
nt]
Measurements in chronological order
Span Stability
Issued by DELTA SP(res) DTPR76-v5#3d.xlsm