ENInstallation and Operating Instructions

9000-608-67/30

*9000-608-67/30*

CR 7 VET

130905

EN

32013/09/06 9000-608-67/30

Content

Important information 1. General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

1.1 Noteonconformity.. . . . . . . . . . . . . . . . . 41.2 GeneralNotes. . . . . . . . . . . . . . . . . . . . . . 41.3 Disposalofappliance.. . . . . . . . . . . . . . . 51.4 CorrectUsage. . . . . . . . . . . . . . . . . . . . . . 51.5 IncorrectUsage.. . . . . . . . . . . . . . . . . . . . 51.6 UsingPeripheralDevices.. . . . . . . . . . . . 51.7 Usingimageplates.. . . . . . . . . . . . . . . . . 51.8 Useofsoftware. . . . . . . . . . . . . . . . . . . . 5

2. Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.1 Generalnotesonsafety.. . . . . . . . . . . . . 62.2 Electricalsafetyinstructions.. . . . . . . . . 6

3. Warnings and symbols . . . . . . . . . . . . . . . . 73.1 InstallationandoperatingInstructions..73.2 Adhesivelabel. . . . . . . . . . . . . . . . . . . . . . 73.3 Modelidentificationplate.. . . . . . . . . . . . 7

4. Delivery contents . . . . . . . . . . . . . . . . . . . . . . 84.1 Specialaccessories. . . . . . . . . . . . . . . . . 84.2 Disposablematerials. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 9

5. System requirements . . . . . . . . . . . . . . . . . 105.1 Monitor.. . . . . . . . . . . . . . . . . . . . . . . . . . . 105.2 PCsystem.. . . . . . . . . . . . . . . . . . . . . . . . 10

6. Technical data . . . . . . . . . . . . . . . . . . . . . . . . 116.1 CR7VET.. . . . . . . . . . . . . . . . . . . . . . . . . 116.2 Imageplate.. . . . . . . . . . . . . . . . . . . . . . . 11

7. Functional layout . . . . . . . . . . . . . . . . . . . . . 127.1 Functionofindividualcomponents.. . 13

8. Functional description . . . . . . . . . . . . . . . . 14

Mounting 9. Transport and packaging. . . . . . . . . . . . . . 15 10. Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

10.1 Roomforset-up.. . . . . . . . . . . . . . . . . . 1510.2 Set-upalternatives.. . . . . . . . . . . . . . . . 16

11. Electrical connection . . . . . . . . . . . . . . . . . 16 12. Start-up and first use . . . . . . . . . . . . . . . . . 17

12.1 CR7VET.. . . . . . . . . . . . . . . . . . . . . . . . 1712.2 X-raydevicesettings.. . . . . . . . . . . . . . 1812.3 Commissioningtests.. . . . . . . . . . . . . . 18

Use 13. Image plates . . . . . . . . . . . . . . . . . . . . . . . . . 19

13.1 General.. . . . . . . . . . . . . . . . . . . . . . . . . . 1913.2 Workingwithimageplates.. . . . . . . . 2013.3 Storingimageplates. . . . . . . . . . . . . . 20

14. Light protection covers . . . . . . . . . . . . . . . 21 15. Protective cover . . . . . . . . . . . . . . . . . . . . . . 21 16. X-ray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

16.1 PreparationforX-raying.. . . . . . . . . . . 2216.2 X-rayexposure.. . . . . . . . . . . . . . . . . . . 2316.3 Preparationforreadingoutthe

imagedata.. . . . . . . . . . . . . . . . . . . . . . . 2416.4 Readingouttheimagedata. . . . . . . . 2516.5 Erasingtheimagedata.. . . . . . . . . . . . 28

17. Disinfection and Cleaning . . . . . . . . . . . . 2917.1.Lightprotectioncovers.. . . . . . . . . . . . 2917.2 Imageplates.. . . . . . . . . . . . . . . . . . . . . 2917.3 Protectivecover.. . . . . . . . . . . . . . . . . . 2917.4 CR7VET.. . . . . . . . . . . . . . . . . . . . . . . . 30

18. Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . 31

Troubleshooting 19. Tips for Operators and Technicians . . . 32

19.1 CR7VETGeneralInformation.. . . . . 3219.2 ErrorMessagesontheDisplay.. . . . . 36

AppendixImage process times . . . . . . . . . . . . . . . . . . . . . 37File size (uncompressed). . . . . . . . . . . . . . . . . . 37

EN

4 9000-608-67/30 2013/09/06

Important information

1. General 1.1 Note on conformityTheproductwassubjecttoaconformityac-ceptanceprocessundertheEuropeanUnionguidelinescoveringthesetypesofdeviceandconformswiththeessentialrequirementsoftheseregulations.

ComplieswithFDAperformancestandardsforlaserproductsexceptfordeviationspursuanttoLaserNoticeNo.50,date(June,2007).

ETL approvalThisClassAdigitalapparatuscomplieswithCanadianICES-003.ThisequipmenthasbeentestedandfoundtocomplywiththelimitsforaclassAdigitaldevice,pursuanttoPart15oftheFCCrules.Theselimitsaredesignedtoprovidereasonableprotectionagainstharmfulinterferencewhentheequipmentisoperatedinacommercialenvironment.Thisequipmentgenerates,uses,andcanradiateradiofrequenceenergyand,ifnotinstalledandusedinaccordancewiththeinstructionmanual,maycauseharmfullinterfer-encetoradiocommunications.Operationoftheseequipmentinaresidentialareaislikelytocauseharmfulinterferenceinwhichcasetheuserwillberequiredtocorrecttheinterferenceathisownexpense.

1.2 General Notes•TheInstallationandOperatingInstructions

constitutepartoftheappliance.Theymustbemadeavailabletotheoperator.CorrectobservanceoftheInstallationandOperatingInstructionsisabasicrequirementforusingtheapplianceproperlyandsafely,andnewpersonnelmustbeinstructedaccordingly.TheseInstallationandOperatingInstructionsmustbehandedovertoanysubsequentowneroroperatorofthisappliance.

•Thesafetyoftheoperatorandtrouble-freeoperationoftheappliancecanonlybeensuredwhereoriginalengineeringmanufac-turedpartsareused.Furthermore,onlytheaccessoriesandconsumablesdescribedintheinstallationandoperatinginstructionsorapprovedbyDÜRRMEDICALforusewiththisappliancemaybeused.Ifotheraccessoriesorconsumablesareused,DÜRRMEDICALcanassumenowarrantyforthesafeopera-tionandreliablefunction.Noliabilityonthepartofthemanufacturewillbeacceptedinthecasethatdamagearisesthroughtheuseofnon-approvedaccessories.

•DÜRRMEDICALisresponsiblefortheap-pliancesfromthepointofviewofsafety,reliabilityandfunctiononlywheninstallation,newsettings,modifications,expansionsandrepairsarecarriedoutbyDÜRRMEDICALorbyaservicecentreauthorisedbyDÜRRMEDICAL,andwhentheapplianceisusedandoperatedinaccordancewiththeinstalla-tionandoperatinginstructions.

•TheseInstallationandOperatingInstructionsaccordwiththefeaturesoftheapplianceandthelevelofengineeringatthetimeoffirstintroductionofthemodel.Allcircuits,processes,names,softwareandappliancesquotedareprotectedunderindustrialpropertyrights.

•ThistranslationoftheInstallationandOperat-ingInstructionshasbeencarriedoutingoodfaith.Liabilityforincorrecttranslationwillnotbeaccepted.TheaccompanyingGermanversionoftheseInstallationandOperatingInstructionsaretobeusedasreference;ifyouhaveanydoubtastothecorrectinterpretationoftheinstructions,pleaseconsultyourdealer.

•ReprintoftheInstallationandOperatingIn-structions,orpartsthereof,isonlypermittedwiththewrittenapprovalofDÜRRMEDICAL.

•Keeporiginalpackingforpossiblereturnoftheappliancetothesupplier.Onlytheoriginalpackingensuresoptimumprotectionfortheapplianceduringtransport.Iftheappliancehastobereturnedtothemanufacturersduringthewarrantyperiod,DÜRRMEDICALwillacceptnoliabilityfordamageoccurringduringtransportresultingfrompoorpackaging!Ensurethatthepackagingiskeptoutofthereachofchildren.

EN

52013/09/06 9000-608-67/30

1.3 Disposal of applianceDevice•TheEUguideline2002/96/EC-WEEE

(WasteElectricandElectronicEquipment)of27January2003anditscurrentimplementa-tionintonationallawdeterminesthatproductsaresubjecttotheabove-mentionedguidelineandmustbedisposedofespeciallywithintheEuropeanEconomicArea.Questionsontheproperdisposaloftheproductcanbedi-rectedtoDÜRRMEDICALorthedentaltrade.

image plateTheimageplatecontainsbariumcompounds.

•Disposeoftheimageplateproperlyinaccord-ancewiththelocallyapplicableregulations.

•InEurope,disposeoftheimageplateinac-cordancewithwastecode090199"wastesnototherwisespecified".Disposalasdomes-ticwasteisnotpossible.

1.4 Correct Usage•Theapplianceisintendedexclusivelyforvisual

scanningandprocessingofimagedataonanimageplateinveterinarymedicine.

•CorrectUsageoftheappliancealsoinvolvesobservingtheInstallationandOperatingInstructionsandadheringtothoseinstructionsconcerningset-up,operatingandmainte-nance.

• EveryusagewhichisnotinaccordancewiththeCorrectUsageabovewillleadtotheguaranteesandwarrantiesofDÜRRMEDICALbecomingnullandvoid.

•Theoperatorisliableforanydamagearisingthroughincorrectusage.

•Correctusageincludescompliancewithalllegalprovisionsapplicableattheplaceofusewithregardworkandradiationprotection.

1.5 Incorrect UsageTheapplianceisnotintendedforvisualscan-ningofimageplatesinhumanmedicine.Anyuseoftheapplianceaboveandbeyondthatlaiddownheresconsideredtobeimproperuseoftheequipment.Anyresultingdamagethuscausedisthesoleliabilityoftheuser.Themanufacturercannotbeheldresponsiblefordamagecausedbyimproperuse.

This appliance must not be used in operating theatres or similar types of room, where there is a risk of inflam-mable materials catching fire.

1.6 Using Peripheral Devices•Onlyperipheraldevices(computer,monitor,

printer)maybeconnectedtothisappliancewhichconformtostandardIEC60950-1(EN60950/1).

•Appliancesmayonlybeconnectedtogetherorconnectedtoothersystempartswherethesafetyofpatients,operatorsandalsotheenvi-ronmentisensured.Wheretheabsolutesafetyforoperator,patientorenvironmentisnotapparentthentheoperatorisexpectedtocallonexpertadvice,e.g.fromthemanufactureroranexpert,thatsuchconnectionsaresafebeforeproceeding.

1.7 Using image platesThescannermayonlybeoperatedincom-binationwithcassettesfromDürrNDTorDÜRRMEDICAL.Seealsothesectiononimageplates.

1.8 Use of softwareOnlysoftwarespecificallyapprovedbyDÜRRMEDICAL(e.g.Vet-ExamPlus,Vet-ExamIntra)maybeused.UseofanyothersoftwaremustbeexpresslyapprovedbyDÜRRMEDICAL.

Thefollowingoperatingsystemsaresup-ported-WindowsXPProfessional32-bitfromServicePack3,WindowsVista32-bitHomePremiumorhigherandWindows732-bit/64-bit.

EN

6 9000-608-67/30 2013/09/06

2. Safety 2.1 General notes on safetyTheappliancehasbeendevelopedandde-signedinsuchawaythatdangersareeffectivelyruledoutifoperatedinaccordancewiththein-tendeduse.However,wefeelitisimportanttodescribethefollowingsafetymeasuresinordertoremoveanylikelihoodofdanger.•Whenoperatingthisappliance,alllocaland

nationalrulesandregulationsmustbeob-served!Thisappliancemustnotbeconvertedoralteredinanyway.DÜRRMEDICALcanas-sumenowarrantyorliabilityformodifiedoradaptedappliances.Intheinterestsofsafeusageoftheappliance,bothoperatorandownerareresponsibleforseeingthatallrel-evantregulationsandprovisionsareobserved.

•OpeninganappliancewhichisstillconnectedtothemainspowersupplyisdangerousandcouldleadtolaserradiationClass3B.

• Installationmustbecarriedoutbyatechnicalexpert.

•Theoperatormustcarefullychecktheappli-anceforsafetyoffunctionandproperworkingconditionbeforeeveryuse.

•Theoperatormustbetrainedinthecorrectoperationoftheappliance.

2.2 Electrical safety instructions•Theappliancemayonlybeconnectedtoa

correctlyinstalledelectricalsocket.

•Beforeconnectingtotheelectricitysupply,theappliancemustbeinspectedandcheckedtoensurethatthesupplyvoltageandthesupplyfrequencycorrespondtothatofthelocalelectricalsupply.

•Beforeinitialuseandstart-up,theapplianceandallsupplylinesmustbecheckedforanysignsofdamage.Damagedsupplylinesandconnectionsmustbereplacedimmediately.

•Nevertouchpersonsandopenconnectorsoftheappliancesimultaneously.

•Whenusingtheappliance,observealltherelevantelectricalsafetyprocedures.

Beforeopeningthedevice,un-plugatthemainsandremoveallpowermainsplug!

EN

72013/09/06 9000-608-67/30

3. Warnings and symbols 3.1 Installation and

operating InstructionsIntheoperatinginstructions,thefollowingwarn-ingsandsymbolshavebeenused:

Information and/or mandatory regu-lations or prohibitions for the preven-tion of personal injury or substantial property damage

Specialinstructionsconcerningeco-nomicusageoftheapplianceandothernotes.

Switchofftheappliance(i.e.unplugfrommains).

ON/OFFkey

OperationalLED,greenStandby

OperationalLED,blueDatatransmission

OperationalLED,yellowCleaningmode

Onlyuseonce(disposable)

Wearprotectivegloves

3.2 Adhesive label

WarningLaser

Warningdangerousvoltage.

Class3Blaser

WarningoflaserbeamsClosedappliance:Laserclass1Openappliance:Laserclass3B

3.3 Model identification plateThemodelidentificationplateislocatedontherearsideoftheCR7VET.Thefollowingterminologyand/orsymbolsareusedinconjunctionwiththisappliance,modelidentificationplateandpowerunit:

Dateofmanufacture

Manufacturer

ETL-CertificationCONFORMSTOULSTD61010-1CERTIFIEDTOCAN/CSASTDC22:2NO61010-1

Observenotesinaccompanyingdocu-mentation!

DisposeofcorrectlyaccordingtoEUdirective(2002/96/EG-WEE)

IP20 Typeofprotection

REF OrderNo./ModelNo.

SN SerialNo.

CE/labelingwithoutNotifiedBodyNumber

Notesonlasersource

Dangertocomponentsduetoelectrostaticdischarge(ESD)

EN

8 9000-608-67/30 2013/09/06

4. Delivery contents Thefollowingarticlesareincludedinthescopeofdelivery(possiblevariationsduetocountry-specificrequirementsand/orimportregulations):

CR 7 VET . . . . . . . . . . . . . . . . . . . . . . 2137-01- CR7VET

- Powerunit

- USBdatacable(3m)

- Datacable(RJ453m)

- Imaging-SoftwareVet-ExamIntra(DemoVersion)-Vet-ExamIntraX-raymodule-Vet-ExamIntravideomodule-Vet-ExamIntraDICOMPat-CD-Vet-ExamIntraEqualisationfilter

- Protectivecover

- IP-CleaningWipes(10pcs)

- Notesoninstallationandoperatinginstruc-tions

4.1 Special accessoriesThefollowingpartsarenotsuppliedwiththecontentsofdelivery.Pleaseorderasrequired.

Wallmountingbracket......... 2141-001-002DÜRRMEDICALImageplatesPLUS-Size0-2x3(22x35mm). . . . . . . 2134-040-502DÜRRMEDICALImageplatesPLUS-Size1-2x4(24x40mm). . . . . . . 2134-041-504DÜRRMEDICALImageplatesPLUS-Size2-3x4(31x41mm). . . . . . . 2134-042-502DÜRRMEDICALImageplatesPLUS-Size3-2.7x5.4mm. . . . . . . . . . 2134-043-501DÜRRMEDICALImageplatePLUS-Size4-5.7x7.6mm. . . . . . . . . . 2134-044-50VETDENTRASETSpecialimageplateforrabbits(2pcs). . . . . . . . . . . . . . . . . . . VHIP34SET112BiteprotectorsSize4(100pcs) . . . . . . . . . . . . . . . . . . 2130-074-03Datacable(USB5m). . . . . . . . 9000-119-027Datacable(RJ455m) . . . . . . . 9000-118-036

Furtherimageplateformatsondemand

Theusageofothercomponents,trans-formersandcablesasthosespecificallylistedunderaccessories,canleadtoeitherincreasedsignaloutputorreducetheresistancetointerferenceofthesystem.

EN

92013/09/06 9000-608-67/30

4.2 Disposable materialsLightprotectioncovers-Size0-2x3(100pieces) . . . . . . 2134-070-00Lightprotectioncovers-Size1-2x4(100pieces) . . . . . . 2134-071-00Lightprotectioncovers-Size2-3x4(300pieces) . . . . . . 2134-072-00Lightprotectioncovers-Size3-2.7x5.4(100pieces) . . . 2134-073-00Lightprotectioncovers-Size4-5.7x7.6(100pieces) . . . 2134-074-00IP-CleaningWipes(100pcs.) . . CCB351A0101

OthercleaninganddisinfectantscanbeobtainedfromthecompanyOrochemie(www.orochemie.de)underthefollowingnames:QuickdisinfectionB30DisinfectantwipesB60,

EN

10 9000-608-67/30 2013/09/06

5. System requirements TooperateCR7VETthecomponentslistedundersections5.1and5.2willberequired.ThearenotsuppliedwiththecontentsofdeliveryoftheCR7VET.

5.1 MonitorUsemonitorsfordigitalX-raywithhigherlightintensityandwidecontrastrange.Strongroomlighting,directsunlightandreflectionsallservetoreducethediagnosticcapabilitiesoftheX-rayexposures.InGermany,anacceptancetestinaccordancewithDINV6868-57canbecarriedout.Inothercountries,observetheapplicablelegalregulations!

5.2 PC systemRefertoleafletMinimumsystemrequirementsforcomputersystems,ordernumber9000-608-100/..orontheinternetunderwww.duerr-medical.de.

EN

112013/09/06 9000-608-67/30

6. Technical data 6.1 CR 7 VETElectrical Details power unitVoltage(V) . . . . . . . . . . . . . . . . . . . . 100-240Maximumfluctuationinratedcurrent(%). . . . . . . . . +/-10Frequency(Hz). . . . . . . . . . . . . . . . . . . 50-60Current(A) . . . . . . . . . . . . . . . . . . . . . max.0,8Currentconsumption(W) . . . . . . . . . . . . . .<40Typeofprotection . . . . . . . . . . . . . . . . . . IP20Protectionclass. . . . . . . . . . . . . . . . . . . . . . . IIOvervoltagecategory. . . . . . . . . . . . . . . . . . . II

Electrical Details CR 7 VETVoltage(V,DC) . . . . . . . . . . . . . . . . . . . . . . .24Maximumfluctuationinratedcurrent(%). . . . . . . . . +/-10Current(A) . . . . . . . . . . . . . . . . . . . . max.1,25Currentconsumption(W) . . . . . . . . . . . . . .<30

Duty cycleS2AccordingtoVDE0530-1(min) . . . . . . . .25S6AccordingtoVDE0530-1(%) . . . . . . . . .25

DimensionsW=226mm,H=234mm,D=243mm

Weight CR7VET(kg). . . . . . . . . . . . . . . . . approx.6,5

Sound level Stand-by(db(A)). . . . . . . . . . . . . . . approx.37Duringscanning(db(A)) . . . . . . . . . approx.55

General valuesExternal:laserclass . . . . . . . . . 1(EN60825-1)

Intern:laserclass . . . . . . . . . . . .3B(EN60825-1)

Wavelengthλ635nm10mWEN60825-1:2006-04+A1:2002-07+A2:2001-03

Heatradiation(W). . . . . . . . . . . . . . . . . . . .<40

Pixel size canbeselectedfromrange(µm). . . .12.5to50

Resolution (dependingonimageplate)linepairs/mm(Lp/mm). . . . . .max.ca.40

Conformity Acceptance Procedure

Lowvoltage . . . . . . . . . . . . . . . . . . . 06/95/EG

EMCConductedHF-disturbanceV1accordingtoIEC61000-4-6(Veff). . . . . . . . . . .3RadiatedHF-disturbanceV1accordingtoIEC61000-4-3(Veff). . . . . . . . . .4

Ambient temperature during operation

Temperature(°C). . . . . . . . . . . . . . . +10to+35Relativehumidity(%) . . . . . . . . . . . . . . .20-80Airpressure(hPA). . . . . . . . . . . . . . .750-1060Heightabovesealevel(m). . . . . . . . . . . <2000

Ambient conditions during storage and transport

Temperature(°C) . . . . . . . . . . . . . . -20to+60Relativehumidity(%) . . . . . . . . . . . . . . .10-95Airpressure(hPA). . . . . . . . . . . . . . .750-1060Heightabovesealevel(m). . . . . . . . . . <16000

6.2 Image plateAmbient conditions during storage and transport

Temperature(°C). . . . . . . . . . . . . . . . . . . . .<33Relativehumidity(%) . . . . . . . . . . . . . . . . . .80

Ambient temperature during operation

Temperature(°C). . . . . . . . . . . . . . . . . . 18-45Relativehumidity(%) . . . . . . . . . . . . . . . . . .80

Dimensions of intraoral image platesSize0(mm). . . . . . . . . . . . . . . . . . . . . 22x35Size1(mm). . . . . . . . . . . . . . . . . . . . . 24x40Size2(mm). . . . . . . . . . . . . . . . . . . . . 31x41Size3(mm). . . . . . . . . . . . . . . . . . . . . 27x54Size4(mm). . . . . . . . . . . . . . . . . . . . . 57x76

EN

12 9000-608-67/30 2013/09/06

7. Functional layout

3

65

4

11

7

10

8

9

12

1 2

EN

132013/09/06 9000-608-67/30

7.1 Function of individual components

1. Image plateTheimageplateservesastheexposuremediumforX-raysandstorestheimagedata.

2 Light protection covers•Protectstheimageplatefromcontamination

anddirt.

•Preventimagedataontheimageplatefrombeingerased.

3. Operational LEDsGeneralinformationconcerningtheoperatingstateoftheapplianceisdisplayed.

LEDgreen:readyforoperationisdisplayed

LEDblue:statusofdatacommunicationisdisplayed

LEDyellow:statusofthecleaningoperationisdisplayed

4. Cleaning key

Pressingthekeywillsettheunitintotheclean-ingpositionorreturntheunitfromthecleaningpositionbackintopositionreadyforoperation.

5. On / Off keySwitchon:presstheOn/Offkey.

Switchoff:presstheOn/Offkeyfor2seconds.Theapplianceshutsdown,thegreenandblueLEDsblinkalternately.

Completeshutdown:presstheOn/OffkeycontinuouslyuntilalltheLEDsgoout.

6. Status LEDsDisplaysinformationonstatusofimageplatetransport.

LEDred:Fault

LEDyellow:Imageplatetransportoccupied

LEDgreen:Newimageplatescanbefedin

7. Cover

Coversthemovingpartsofthefilmfeedsection.Thiscanberemovedforcleaninganddisinfect-ing.

8. Display

Thisdisplaymodulepresents,dependingonthesoftwareused,variousinformationontheap-pliancestatus,e.g.thename,roomusedorthetoothnumberwhenusingVet-ExamIntra.

Inthecasethatanerrormessageisdis-played,seesectiononTipsforOpera-torsandServiceTechnician.

9. Cover lock

Unlocksthecoverforopening.

10. Film collection trayAfterscanninganderasuretheimageplatesfallintothedetachablefilmcollectiontray.

11. Interior image transportInsidetheunittheimageplatesaretransportedthroughthefilmscanningandfilmerasureunits.

12. Feed slotThefilmfeedslotiswheretheimageplates,insidetheiropenedlightprotectioncovers,areplacedsothatthefilmcatchtakesholdandthefilmisfedtotheimageplatetransportsystem.

EN

14 9000-608-67/30 2013/09/06

8. Functional description CR7VETtransformstheintraoralimagedatastoredontheimageplateintoadigitalX-rayim-age.ThisimagecanthenbeviewedandeditedusingtheconnectedPCsystem.

X-raysTheimageplate(1)ispacked,disinfectedandexposedinsidethelightprotectioncover(2).Afterexposurethelightprotectioncoverisagaindisinfected.

Scanning the image dataPressingtheOn/Offkey(5)willswitchontheCR7VET.Thisactivatesthescanningmodeusingx-raysoftware,e.g.Vet-ExamIntra.OncethegreenstatusLED(6)islit,theimageplateinitsprepared(opened)lightprotec-tioncovercanbeplacedcentrallyonthefilmfeedtray(12).

Theimageplateisnowpushedcarefullyoutofitssecuredandopenedlightprotectioncoverintothefilmtransport(11)usingthefingers.Thelightprotectioncoverstaysinpositionatthefilmfeedslot.Assoonasthelightprotectioncoverisreleased,itcanberemoved.

Oncetakenupbythefilmtransport(11)theim-ageplateisfedtothescannerunit.Heretheim-agedataontheimageplateisscannedusingalaserandtheinformationtransmittedtothePC.Awindowonthemonitorshowstheprogressofthisprocedure.

Afterscanningtheimageplateproceedstotheerasureunit,whereanyremainingimagedataiserased.Afterwardstheimageplatefallsintothecollectiontray(10).

Further X-ray imagesAfterremovingtheimageplatesfromthefilmcollectiontrayitisrecommendedthattheyberepackedintoalightprotectioncover.

Ifusedcarefullyitispossibletoreusetheimageplatesseveralhundredtimes.Beforereusetheimageplatesshouldbecheckedforsignsofdirtordamage.Damagedimageplatesmustnotbereused.

Imageplate

Parabolicmirror

Photomultiplier

Lasersource

Reflectionlayer

Pentaprism

Rotatingdisc

EN

152013/09/06 9000-608-67/30

Mounting

Only fully-qualified personnel may set-up, install or operate this appli-ance.

9. Transport and packaging•Checkthatthedeliverycontentsarecomplete

andcheckforanysignsoftransportdamage.

•Completelyemptythepackaging.

•CR7VETDonotexposetheVistaScanMinitoanystrongshockswhichcoulddamagetheappliance.

Damage through shock. During operation and the switching off process (blue and green op-erational LEDs blinking alternately) ensure that the appliance is not moved in any way whatsoever, as the moving parts must be allowed to stop moving completely.

10. Set-up 10.1 Room for set-up• InstalltheCR7VETonlyinaclosed,dryand

wellventilatedroom.

•Theroomtemperatureshouldnotfallbelow+10°Corexceed+35°C.

Danger of short circuit due to build up of condensation. The system can only be put into operation once the appliance has warmed up to room temperature and it is dry.

• Installationinapurpose-builtroom,e.g.inaboilerroomorinawetroomisnotpermitted.

•DonotexposetheCR7VETtodirectsunlight.Theambientlightmustbereducedifnecessary.Themaximumlightintensityis1000Lux.

•Theremustbenosourceofinterferenceintheroomforset-up(e.g.strongmagneticfield).Theseinterferencefieldsmaycausemalfunc-tionsintheCR7VET.

EN

16 9000-608-67/30 2013/09/06

1

10.2 Set-up alternativesTable set-up

CR7VETcanbesetupasatabletopappli-ance.

Inordertoavoidproblemswhenscan-ningtheimagedata,theCR7VETshouldbesetupsothatwillnotbesubjecttoknocks,vibrationorsimilar.

Wall mounting

TheCR7VETcanbemountedontoanywallusingthespecialwallmountingbracket(ordernumber2141-001-00).

Installationofthewallmountingbracketcanbefoundintheinstructionleafletordernumber9000-618-162/01

11. Electrical connection Beforeinitialstart-upcomparetheoperat-ingvoltageof24VDCandthesupplyvoltage100V-240Vareconsistentwiththeinformationonthemodelidentificationplate.Theconnectionportsareintherecessesontherearsideoftheappliance.•Whereinconsistentremoveanyincorrect

(country-specific)adapter/plug.•Connectappropriatecountry-specificadapter/

plug.• RemovetherearcoveroftheCR7VET.• Insertconnectorofthepowerunittothecon-

nectionportsoftheCR7VET.•Placethepowercableintothepoweroutlet

socket.•Replacethecover.

Whenoperatingtheappliance,therearcovermustbeinplace.

EN

172013/09/06 9000-608-67/30

12. Start-up and first use12.1 CR 7 VET

InstallationandconfigurationmayonlybecarriedoutbyaDÜRRDENTALtrainedandcertifiedpersonoroneofourownservicetechnicians.

TheCR7VETisequippedwithfourconnectionportswhicharelocatedunderthecoverontherearsideoftheunit:

14. 24 V - ConnectionThe24V-connectionprovidesthepowertooperatetheCR7VET.

15. Reset buttonPressingandholdingtheResetbuttonfor15-20secondswhileswitchingonreturnsallnet-worksettingsoftheunittotheoriginalfactorysettings.ResetisconfirmedandtheCR7VETPlusdisplayshows"FactoryReset".

16. AUX-connectionTheAUX-connectionisusedfordiagnosispurposesbyourServiceTechnician.TheAUX-connectioncanbeusedtoconnectauxiliarydevicesfortestingandfortrouble-shooting.

OnlyauxiliarydevicesapprovedbyDÜRRMEDICALmaybeused.

17. Network connection status LEDsThestatusLEDsfornetworkingshowwhetherconnectiontoanetworkhasbeenestablishedandwhetherdatacanbetransmitted.

18. Network connectionCR7VETcanbeoperatedinanetwork.

19. USB-connectionCR7VETcanbeoperatedthroughanUSBport.

WhereCR7VETisconnectedtobothUSBand anetwork,thenetworktakespriority.

14

16

18

19

17

2

3

15

EN

18 9000-608-67/30 2013/09/06

12.2 X-ray device settingsCheckthefollowingappliance-specificstandardvaluesandadaptasrequired.

Reference conditions:

•Housecat(approx.6kg)tomediumsizeddog(approx.20kg)

Theexposuretimesindicatedinthetablefortubelength20cmweredeterminedwithadentalx-rayunitwithDCemitter(focalpoint0.7mm;tubelength20cm).Theexposuretimesfortubelength30cmwerecalculatedfromtheexposuretimesfortubelength20cm.

Parameter DC-emitter, 7 mA, tube length 20 cm

DC-emitter, 7 mA, tube length 30 cm

recommended exposure time (s)

recommended exposure time (s)

Upper jaw 60kV 70kV 60kV 70kV

Incisor 0,1 0.08 0.2 0.16

Premolar 0,125 0,1 0.25 0.2

Molar 0.16 0,125 0.32 0.25

Lower jaw bone 60kV 70kV 60kV 70kV

Incisor 0,1 0.08 0.2 0.16

Premolar 0,125 0,1 0.25 0.2

Molar 0,125 0,1 0.25 0.2

Whereitispossibletosetthex-raytubeto60kV,thissettingshouldbeselected.TheusualexposurevaluesforF-Film(e.g.KodakInsight)canbeused.

12.3 Commissioning testsTherequiredtesting(e.g.commissioningtests)mustbecarriedoutaccordingtolocalrulesandregula-tions.•Obtaininformationastowhichtestsarerequired.

•Carryouttestingaccordingtolocalrulesandregulations.

Electrical safety check

•Carryoutanelectricalsafetycheckaccordingtonationalregulations(e.g.housingleakagecurrent).

•Documenttheresults.

EN

192013/09/06 9000-608-67/30

Use

13. Image plates

Image plates are toxic: Image plates are therefore to always be used in a light protection cover. If the image plates or parts thereof are swal-lowed, the image plate must be removed immediately. If a light protection cover is dam-aged when biting on it, then the mouth must be rinsed out thoroughly with copious amounts of water.

13.1 GeneralTheCR7VETmayonlybeoperatedincombi-nationwithimageplatesPLUSfromDürrNDTorDÜRRMEDICAL.UseofanyotherimageplatesmustbeexpresslyapprovedbyDÜRRMEDICAL.Exposureoftheimageplatemustalwaysbecarriedoutontheactiveside:

activeside:•theactivesideislightblue.Thissideisalso

markedwitha positioningguide.

inactiveside:•theinactivesideisblack.Thisisprintedwith

thesize,manufacturer’sinformationandtheword"Back".Furthermore,theinactivesideissurroundedbyasilveredgewhichmakespositioningtogetherwiththelightprotectioncovereasier.Inonecornerthereisasilverdot;thisreplacestheslightindentationfoundonconventionalX-rayfilm.Thisdotshouldalwaysbesetinthedirectionoftheocclusalplane.ThisallowseasierorientationoftheX-rayexposure.

Inactiveside:

Activeside:

4

EN

20 9000-608-67/30 2013/09/06

13.2 Working with image plates Imageplatesmustbevisiblycheckedforsignsofdamage.Damagedimageplates(e.g.atornouterlayer,orvisiblescratches,whichwillaffectthediagnosticassessment)mustbereplaced.

• imageplatesareflexiblelikeX-rayfilm.

•donotbendorsubjecttopressure.

•donotscratchorgetdirty.

•donotputinanautoclave

13.3 Storing image plates •storeimageplatesinadarkplace.

•Donotstoreinhotormoistconditions.

•storeintraoralimageplatesinalightprotec-tioncoveroftheappropriatesize.

• ifanimageplateisstoredforlongerthan24hours,thenitshouldbeerasedagainbeforethenextX-rayexposure.Seesection16.5

•protectfromdirectsunlightandultravioletbeams.

• ifstoredintheX-rayroomitselfthentheim-ageplatesmustbeprotectedfrompre-expo-surecausedbyX-rayradiation.

5

6

EN

212013/09/06 9000-608-67/30

7

14. Light protection covers lightprotectioncover• lightprotectioncoversprotectthepatientfrom

infectionbycontaminatedimageplates.

•protecttheimageplatefromcontaminationanddirt.

•protecttheimageplatesfromstraylight.

• isavailableindifferentsizesjustlikeimageplates.Choosethemostappropriatesize.

The light protection cover is a disposable article and must not be reused.

Forinformationonusinglightprotectioncoversseetheappropriateinstructionsforuse,ordernumber9000-608-29/...

15. Protective coverTheprotectivecoverisdesignedtoprotecttheCR7VETfromdustanddirtwhenstandingforlongperiods.

•Pulltheprotectivecover,withtheyellowmark-ingtowardsthefront,forwardsovertheunituntilitiscompletelycovered.

•Whennotinusestoretheprotectivecoverinacleandryplace.

Danger of suffocationKeep the protective cover away from children.

8

EN

22 9000-608-67/30 2013/09/06

16. X-ray

Observethenotesonhygieneregula-tionsundersection17.CleaningandDisinfection.

16.1 Preparation for X-raying

ImageplatescanbeprematurelyexposedbynaturalX-rayradiationorbystrayX-rayradiation.

•Theimageplateshouldbeerasedbeforeus-ingitforthefirsttimeorafterstorageofmorethan24hours.Refertosection16.5Erasureofimagedata

•Slidetheimageplatecompletelyintothelightprotectioncover.Theblack(inactive)sideoftheimageplatemustbevisible.

•Peeloffthewhitestickystripandpressfirmlytoclosetheopeningofthelightprotectioncover.

The light protection cover is a disposable article and must not be reused.

•Disinfectthelightprotectioncoverusingadisinfectionwipe(e.g.OrochemieB60)beforepositioning.Alternatively,useadisinfectantspray(e.g.OrochemieB30)onasoft,fluff-freeclothinsteadofadisinfectantwipe.

9

10

11

EN

232013/09/06 9000-608-67/30

16.2 X-ray exposure

Dangerofinfection!Wearprotectivegloveswhenpositioningtheimageplate.

Theimageplatesaremarkedwithasymbol .ThesymbolisvisibleintheX-rayexposure,andmakesorientationduringdiagnosiseasier.TheimageplateshouldbepositionedsothatthesymbolwillnotaffectthediagnosticusefulnessoftheX-rayexposure.

•Placetheimageplateinsidethelightprotec-tioncoverasnormal.Thesilverpointontheinactivesidemustfacetheocclusalarea.TheactivesideoftheimageplatemustpointinthedirectionoftheX-raytube.

Damage to the image plate Only use handling systems which will not cause any damage to either the light protection cover or the im-age plates. Handling systems which have sharp edges should be avoided where possible.

•SetexposuretimeandsettingsontheX-rayappliance.

•TakingtheX-rayexposure.

The image data present on the image plate must be processed and scanned within 30 minutes of exposure.

12

EN

24 9000-608-67/30 2013/09/06

16.3 Preparation for reading out the image data

Dangerofinfection!Wearprotectivegloveswhenremovingimageplates.

•Removetheimageplateinthelightprotectioncover.

• Ifthelightprotectioncoverortheglovesareconsiderablydirty,e.g.fromblood,thenbeforedisinfectionthecovermustbegivenaquickdryclean,e.g.bywipingwithacleanpieceoflint.

•Cleananddisinfectthelightprotectioncoverandthegloveswithadisinfectantwipe(e.g.OrochemieB60).Alternatively,useadisinfectantspray(e.g.OrochemieB30)onasoft,lint-freecloth.

•Laytheimageplatedowninthelightprotec-tioncoverinthedisinfectantwipe.

Damage to the CR 7 VET Before handling the image plates, hands should be completely clean and free of any traces of powder from protective gloves. This powder can damage the optical scanning unit of CR 7 VET during the scanning procedure of the image plate.

13

14

EN

252013/09/06 9000-608-67/30

16.4 Reading out the image data

Contamination of the CR 7 VET The image plates must be clean and absolutely hygienic before scanning of the image data. Otherwise there could be a loss of image data or even to the transfer of bacteria.

•SwitchonPCandmonitor.

•SwitchontheCR7VETattheOn/Offswitch(5).

Afterswitchingontheapplianceneedstogothroughinitialisation.ThedisplayontheCR7VETwillshow"INITSYS-TEM".WhenreadythegreenoperatingLEDwillbecomeilluminated."READY"willalsobeshownonthedisplay.

•Startthesoftware.

Thesequenceoftheread-outpro-cedurecandiffer,dependingonthesoftwareused.Seealsothemanualofthesoftwareused.

•Selectanimalowner.•Selectmeasurefortheanimal.•ActivatetheCR7VETinthesoftware.

Confusing patient X-rays The software ensures that the x-rays are always allotted to patients one after the other in the order of acti-vated patients. In order to avoid confusing these X-rays from those of other patients only X-ray exposures for this (one) patient should be scanned at one time.

Operational LEDs

Readyforoperation

Notreadyforcommunication

Cleaningposition

Coverorfilmfeednotcorrectlymounted

Shutdownprocess

On OFF Flashing

Transport slot status LEDs

Problem

Imageplatecurrentlybeingproc-essed

Filmfeedslotready

Radiationdeflectorswitchesfromstandbymodetosetvalue

On Off Flashing

EN

26 9000-608-67/30 2013/09/06

OperationalLEDandstatusLEDarebothgreen:

WhentheCR7VETisstartedfromstand-bymodethegreenstatusLEDblinks.Additionally,aprogressbarwillappearonthedisplay.Assoonastheapplianceisreadyforuse,thestatusLEDwillshowgreen.

•Openthelightprotectioncoverstraight alongtheadhesiveedge(V-cut).

Loss of image data Image data on the image plate will be lost when exposed to light. Never handle exposed image plates with-out the light protection cover.

•Placethelightprotectioncoverwithitsimageplatecentrallyandparallel(openedsideofthelightprotectioncovershouldfacedown,inac-tivesideoftheimageplatetofaceoperator)ontothefeedinserttransport.Thefilmgripmechanismautomaticallymovesforwardsandtakesholdofthelightprotectioncoverwiththeimageplate.

•Slidetheimageplateoutofitslightprotectioncoverdownwardsintotheapplianceuntiltheimageplateisautomaticallytakenupbythetransport.Thelightprotectioncoverisheldbythegripmechanismandisnottakenupbytheappliance.

ScannerprogresscanbemonitoredonasmallinnerwindowonthePCscreen.

16

15

EN

272013/09/06 9000-608-67/30

StatusLEDisyellow:•Donotfeedanymoreimageplates.

•Oncethescanningprocesshasbeencompletedthegripmechanismautomati-callyreturnstostart.Removetheemptylightprotectioncover.StatusLEDchangesfromyellowtogreen.

•SavetheX-rayimageinVet-ExamIntra.

•EitherinsertanewimageplateforscanningorturnofftheapplianceattheOn/Offswitch.

WhentheOn/Offswitchispressedfor3secondsthentheapplianceshutsdown.Thiscantakeafewmoments(greenandblueLEDsblinkalternatelyandthedisplaywillshowaprogressbar).Theappliancewillshutdowncompletely.Donotmovetheapplianceduringthisphase.

WhentheCR7VETisswitchedoff,itcanbecoveredandprotectedusingtheprotectivecover.

17

EN

28 9000-608-67/30 2013/09/06

16.5 Erasing the image dataAfterscanning,theimageplateautomaticallyrunsthroughtheerasureunit.Imagedatastillheldontheimageplatewillbeerased.After-wardstheimageplatefallsintothecollectiontray.

Special mode "Erase"

Theimageplatemustbeerasedagainif:• theimageplatehasbeenstoredandnotused

forlongerthan24hours.

•TheimagedataontheimageplatewerenoterasedduetoafaultintheCR7VET.(Errormessagefromsoftware)

ThespecialmodeenablestheerasureunitinCR7VETtobeactivatedalone.Noimagedatawillbetransmittedtothesoftware.

To erase again:

•Choosethespecialmode"Erase"inthesoft-ware.

•Placetheopenlightprotectioncoverwithitsimageplateintheinsertslot.

•Slidetheimageplateoutofitslightprotectioncoverdownwardsintotheapplianceuntiltheimageplateisautomaticallytakenupbythetransport.

18

EN

292013/09/06 9000-608-67/30

17. Disinfection and Cleaning

Onlyusedisinfectionandcleaningagentsspecificallyapprovedormanu-facturedbyDÜRRMEDICAL.OthercleaningagentsanddisinfectantscoulddamagetheCR7VETanditsaccessories.

17.1. Light protection coversDisinfectthelightprotectioncoverusingadis-infectionwipe(e.g.OrochemieB60)beforeandafterpositioning.Alternatively,useadisinfectantspray(e.g.OrochemieB30)onasoft,fluff-freeclothinsteadofadisinfectantwipe.Observetheactiveworkingtimeofthesolutioncarefully.Thefluidsmustbecompletelydried.

17.2 Image plates

Damage to image plates Do not put image plates in the auto-clave! Disinfectant baths will destroy image plates!

Westronglyrecommendtheuseofspecialcleaningclothsforimageplates,ordernumberCCB351A0101

•Dirtoneithersideshouldbecleanedoffusingasoftnon-fluffycloth.Remainingdirtcanberemovedwithaspecialcleaningagentfromthecleaningsetandbyapplyingacircularwipingmotion.

•Beforere-usage,theimageplatemustbecompletelydry.

17.3 Protective coverTheprotectivecovercanbecleanedusingadampcloth,andthendisinfectedusingadisin-fectionwipe(e.g.OrochemieB60).

20

19

EN

30 9000-608-67/30 2013/09/06

17.4 CR 7 VET

Damage to the CR 7 VET Fluids must not be allowed to enter the inner parts of the appliance, therefore do not use spray disinfec-tion or cleaning solutions on the CR 7 VET!

Appliance surfaces

Wheretheoutersurfacesoftheapplianceshowsignsofcontaminationorvisibledirt,cleananddisinfecttheunit.•CleanthesurfacesoftheCR7VETusinga

dampcloth,andthenuseadisinfectionwipe(e.g.OrochemieB60).

•Anydirtpresentattheentryandexitareasmustbecleanedbeforescanning.

Input unitWherethefilmfeedslotshowssignsofcon-taminationorvisibledirt,cleananddisinfecttheunit.

•Presscleaningbutton.Thefilmgripmovestoacentralposition.TheyellowLEDislit.ThedisplayontheCR7VETPluswillshow"CLEANINGPOSITION".

•Pressthereleasebuttonelementandsimul-taneouslyliftthecoverupandawayfromtheunit.

•Liftthelockingplateupandoffasshown.

Damage to the CR 7 VET No parts of the CR 7 VET are to be disinfected with a thermal disinfector or autoclave. The plastic parts will be destroyed by the heat!

• Internalpartscanbecleanedusingadampcloth,andthendisinfectedusingadisinfectionwipe(e.g.OrochemieB60).

•Thecoverandthefilmcatchmechanismcanbedisinfectedusinganinstrumentdisinfectant(e.g.OrochemieA20).

•Replacefilmcatchmechanism.

•Replacecover.

•Pressbuttonforcleaning.Thefilmgripmovestoacentralposition.ThegreenLEDislit.

23

22

21

1.

2.

EN

312013/09/06 9000-608-67/30

18. MaintenanceOnlyspecialiststafforpersonneltrainedbyDÜRRMEDICALmaymaintaintheappliance.

Priortoworkingontheapplianceorincaseofdanger,disconnectitfromthemains(e.g.pulltheplug).

Therecommendedmaintenanceintervalsarebasedonusingtheapplianceforsome15intraoralim-agesperdayand220workingdaysperyear.

Maintenance interval

Maintenance activity

Annual •Visuallyinspectappliance.

•Checktheimageplatesforsignsofscratchesandchangeifnecessary.

Every3years •Changethewiperbladeset.

•Changetherollerfixtures.

•Changethedrivebelt.

EN

32 9000-608-67/30 2013/09/06

Troubleshooting

19. Tips for Operators and Technicians Repairs may only be carried out by qualified personnel or by our After-Sales Service.

Before starting repairs or opening the appliance unplug at the mains and remove all power, remove the mains plug from the plug socket.

19.1 CR 7 VET General Information

Problem Possible cause Solution1. CR 7 VET does not

switch on•Nosupplyvoltage •Checkthemainscableandall

connections•Checkthemainfuseboxto

building•Checkthepowerunit,whether

greenLEDlights

•On/Offswitchdefective • Informyourservicetechnician

2. CR 7 VET switches off after a short period

•Mainscableoroneoftheplugshasnotbeeninsertedcorrectly

•Checkthemainscableandallconnections

•Hardwaredefective • Informyourservicetechnician

•Mainssupplyvoltagetoolow •Checkthesupplyvoltage

3. Status or operating LED does not light up

•LEDdefective • Informyourservicetechnician

4. Loud noises dur-ing operation after switching on

• Ifthenoisespersistlongerthan30seconds,thentheradiationdeflectorisdefective

• Informyourservicetechnician

5. Image editing software does not recognize CR 7 VET

•CR7VETisnotswitchedon •SwitchonCR7VET

•ConnectingcablebetweenCR 7VETandPCisloose

•CheckUSBcable/networkcable

•PCdoesnotrecogniseconnec-tiontoCR7VET

•USB-drivernotinstalledorinac-tive

•CheckUSBcable/networkcable•ChecktheIP-addressesandthe

subnet-masks

•Hardwarefault • Informyourservicetechnician

EN

332013/09/06 9000-608-67/30

Problem Possible cause Solution6. Appliance does not

react• CR7VEThasnotyetcompleted

bootprocedure•Theappliancetakesapprox.

20-30secondstobootafterswitchingon.Repeattheprocedureagainafterthistime.

•CR7VETisblockedbythefire-wall.

•Enabletheportsfortheappli-anceinthefirewall.

7. X-ray exposure does not appear on the monitor after scanning

• Imageplatehasbeenputinlightprotectioncoverupsidedownandtheinactivesidescanned

•Rescantheimageplateasquicklyaspossible

• Imagedataontheimageplatehasbeenerased,e.g.byambientlight

• Imagedataontheimageplatemustbereadoutasquicklyaspossibleafterremoval

•Problemwiththehardware • Informyourservicetechnician

•X-raydeviceisfaulty • Informyourservicetechnician

•Noimageinformationontheim-ageplate

•Exposetheimageplate

8. X-ray too dark / light

•Wrongbrightness/contrastsettinginthesoftware

•SetthebrightnessoftheX-rayimagewiththesoftware

•Theexposedimageplatehasbeencontaminatedbyambientlight

•Placetheimageplateintothelightprotectioncoverandinsertatfilmfeedslot.

•X-raydosetoolow •ChecktheX-rayparameters

9 .Image presented as mirror image

• Imageplateincorrectlyplacedinlightprotectioncover

•Placeimageplatecorrectly.Theblackside(printed)oftheimageplatemustbevisiblethroughthetransparentsideofthelightprotectioncover.

• Imageplatewasplacedinthelightprotectioncoverorinmouthofpatientupsidedown

•Placeandaligntheimageplatecorrectly

10. Ghost images and shadows on the X-ray exposure

• Imagedataontheimageplatehasnotbeensufficientlyerased

•Checktheerasuresettings• Informyourservicetechnician•Replacetheerasureunit

11 . Image is very faint on the scanning window software cannot piece the data together to form a complete picture

•TheX-raydosisontheimageplatewasinsufficient

• Informyourservicetechnician•Replacetheerasureunit

•Theamplificationissettoolow(HVvalue)

• Increaseamplification(HVvalue)

•Theselectedscanmodeisnotsuitable

•Selectsuitablescanmode

•Thesettingforthethresholdvalueistoohigh

•Reducesetting

EN

34 9000-608-67/30 2013/09/06

Problem Possible cause Solution12. CR 7 VET does

not appear in the selection list in CRScanConfig

•Applianceisconnectedbehindanetworkrouter.

•ConfiguretheIPaddressattheCR7VETNetwithoutaninter-mediaterooter.

•Connecttherouteragain.•ManuallyentertheIPaddressin

CRScanConfigandregistertheappliance.

13. CR 7 VET does not appear in the selection list in CRScanConfig, connection is not possible however

•ThesubnetmasksofthePCandCR 7VETdonotmatch

•Checksubnet-masksandalterifnecessary

14. When starting the X-ray module in the software, a fault message appears

•CR7isconflictingwithanotherdevice

• In"CRScanConfig",settheCR7ortheliketo"DemowithoutScanner".

15. X-ray exposure has stripes on image

• Imageplatehasbeenexposed •Onlyusetheimageplatewithin24hoursofdataerasure.

•Partoftheimageplatewasex-posedtolightduringhandling

•Donotexposeusedimageplatestolight.Exposureandthescanningoftheimagedatashouldbecarriedoutwithinhalfanhour.

• Imageplatewasdirtyorscratched

•Cleantheimageplatebywipingcarefullyusingasoftclothandspecialcleaningagenttakenfromthecleaningset.Allimageplateswithscratchesmustbereplacedimmediately

16. Warning message: "Too much ambient light"

•CR7VETexposedtotoomuchlight

•Darkentheroom•TurntheCR7VETsothatthe

lightdoesnotfalldirectlyontotheinputunit

17. Pale stripes in scanning window

•Toomuchstraylightentereddur-ingscanning

•Darkentheroom•TurntheCR7VETsothatthe

lightdoesnotfalldirectlyontotheinputunit

EN

352013/09/06 9000-608-67/30

Problem Possible cause Solution18. Blue LED blinks •NoconnectionbetweenCR7VET

andPC•ActivatetheCR7VETviathe

software•USB-drivernotinstalledorinac-

tive•Checkdatacable

•CR7VETdatamemoryisfull •Usesoftwaretosetupstateofreadiness.TheimagedatawillautomaticallybetransmittedfromCR7VETtothesoftware

19. Error message "E2490"

•AconnectionwasbrokenoffwhilethesoftwarewasaddressingtheCR7VET.

•Reconnecttheunitsandrepeattheoperation.

EN

36 9000-608-67/30 2013/09/06

19.2 Error Messages on the Display

Problem Possible cause Solution1. Error code -1010

on display.•Appliancetemperaturetoohigh •Allowtheappliancetocool

• Informyourservicetechnician

2. Error code -1022 on display.

•Assemblyhasnotbeeninitialised •Probleminsoftware,updatesoftwareifnecessary

• Informyourservicetechnician

3. Error code -1100 on display.

•Scanprocesshasexceededallot-tedtime

• Informyourservicetechnician•Checkthebeltdrive,blockage

duringscanning(imageplatestuckinunit)

4. Error code -1104 on display.

•Problemwitherasureunit • Informyourservicetechnician•Replacetheerasureunit

5. Error code -1116 on display.

•Transportdriveblocked •Clearblockage• Informyourservicetechnician

6. Error code -1117 on display.

•Positionnotcorrectoninsertion • Informyourservicetechnician•Checktransport(smoothrunning,

drivebelt)

7. Error code -1118 on display.

•Housingopen •Closehousing•Pressbuttonforcleaning.

8. Error code -1121 on display.

•Transportguideslidenotinplace •Putslideinposition

9. Error code -1153 on display.

•Probleminsub-assembly •Switchtheapplianceoffandonagain

10. Error code -1154 on display.

•Problemwithinternaldatacommunication

•Switchtheapplianceoffandonagain

11. Error code -1160 on display.

•FinalRPMforradiationdeflectornotachieved

• Informyourservicetechnician•Replacetheradiationdeflector

sub-assemblyifproblemarisesfrequently

12. Error code -1170 on display.

•SOL-SensorTimeout•Faultatlaser,SOL-sensororthe

deflectorassembly

• Informservicetechnician.•Firmwareupdate

13. Error code -1171 on display.

•Faultatlaser •Sendapplianceforrepair.

14. Error code -1172 on display.

•DeflectorassemblyortheSOL-sensordefect

•Switchtheapplianceoffandon.• Informservicetechnician.•Replacethedeflectorassembly

EN

372013/09/06 9000-608-67/30

Appendix

Image process timesScanningtimeistimetakenforcompletescanningofimagedataanddependsonimageplateformatandpixelsize.ThetimetoimagewilldependlargelyonthePCsystememployedandtheworkload.Timesstatedareapproximate.

Theor. resolution (LP/mm) 40 25 20 10

Pixel size (μm) 12.5 20 25 50

IntraSize0(2x3) 26s 16s 13s 6s

IntraSize1(2x4) 32s 20s 16s 8s

IntraSize2(3x4) 32s 20s 16s 8s

IntraSize3(2.7x5.4) 40s 25s 20s 10s

IntraSize4(5.7x7.6) 53s 33s 27s 14s

File size (uncompressed)Theactualfilesizewilldependontheimageplateformatandthepixelsize.Filesizesstatedareap-proximateandhavebeenroundedupwards.Suitablecompressionmethodscanconsiderablyreducethefilesizewithoutlossofdata.

Theor. resolution (LP/mm) 40 25 20 10

Pixel size (μm) 12.5 20 25 50

IntraSize0(2x3) 9.86MB 3.85MB 2.46MB 0.62MB

IntraSize1(2x4) 12,29MB 4.80MB 3.07MB 0.77MB

IntraSize2(3x4) 16.27MB 6.36MB 4.07MB 1.02MB

IntraSize3(2.7x5.4) 19.01MB 7.43MB 4.75MB 1.19MB

IntraSize4(5.7x7.6) 55.45MB 21.66MB 13.86MB 3.47MB

DÜRR NDT GmbH & Co. KGHöpfigheimer Strasse 22 74321 Bietigheim-BissingenGermanyTel: +49 7142 99381-0Fax: +49 7142 99381-299Division DÜRR MEDICALwww.duerr-medical.de [email protected]