chemical, proteomic and mechanical characterization of ... · sialolitos para explicar a...
TRANSCRIPT
Chemical, Proteomic and Mechanical Characterization of
Salivary Calculi
Ana Patrícia Baptista Rodrigues
Dissertação para obtenção do Grau de Mestre em
Engenharia de Materiais
Júri
Presidente: Profª. Drª. Fernanda Maria Ramos da Cruz Margarido
Orientador: Profª. Drª. Patrícia Maria Cristovam Cipriano Almeida de Carvalho
Co-orientador: Dr. António Pedro Alves de Matos
Vogais: Prof. Dr. Arlindo Pereira de Almeida
Outubro de 2010
ii
Acknowledgements
It would not be have been possible to produce this dissertation without the help of several
people who I would like to thank.
I would like to give my thanks to my supervisor Profª. Drª. Patrícia Carvalho for all her support,
endless patience, dedication and availability. Also, to my co-supervisor Prof. Dr. Antonio Alves
de Matos, for all of his knowledge of biology, for his patience and the time spent clarifying my
doubts.
I would also like to thank my colleague Lúcia Santos for the help in all the work undertaken, her
ability and availability to discuss ideas, dedication and all the support she gave me.
I must thank Daniela Nunes and Bruno Nunes for their availability and help in conducting the
tests requiring the use of an Atomic Force Microscope.
For their help in samples preparation for use with the Transmission Electron Microscope, I
would like to thank Drª. Cristina Lacerda of the Anatomic Pathology Department, Curry Cabral
Hospital.
I would also like to thank Richard Butcher for his clarification of certain points of the English
language and for all his support.
Lastly I would like to thank my parents for all the support they have given me throughout my
academic life.
iii
Agradecimentos
Não seria possível a realização desta dissertação sem a ajuda de várias pessoas a que
gostaria de agradecer.
Gostaria de agradecer à minha orientadora Profª. Drª. Patrícia Carvalho por todo o apoio,
dedicação e disponibilidade apresentada diariamente. Ao meu co-orientador Prof. Dr. António
Alves de Matos, por todos os seus ensinamentos na área de biologia, pela sua disponibilidade
e paciência para esclarecimento de dúvidas.
Gostaria de agradecer à minha colega Lúcia Santos pela ajuda em todo o trabalho realizado,
pela sua disponibilidade, discussão de ideias, dedicação e todo o apoio que me deu.
Não podia deixar de agradecer ao Bruno Nunes e Daniela Nunes pela disponibilidade e ajuda
na realização dos ensaios no Microscópio de Força Atómica.
Pela ajuda na preparação de amostras para o Microscópio Electrónico de Transmissão gostaria
de agradecer à Drª. Cristina Lacerda do Departamento de Anatomia Patológica do Hospital
Curry Cabral.
Queria igualmente agradecer ao Richard Butcher pelos seus esclarecimentos de dúvidas sobre
a escrita inglesa e por todo o apoio.
Por último gostaria de agradecer aos meus pais por todo o apoio que me deram ao longo da
minha vida académica.
iv
Abstract
The main goal of this work is the chemical, proteomic and mechanical characterization of
sialoliths in order to explain the existence of denatured collagen, to justify the presence of
sulfur, establish a protocol for the measurement of the mechanical properties of a sialolith and
to provide new information on the calculi formation mechanism.
This study used Scanning Electron Microscopy together with X-ray maps to investigate the
morphologic and chemical aspects of a sialolith. The results showed that the sialolith was
composed of both organic matter and inorganic minerals, organized in layers around a highly
mineralized core. Transmission Electron Microscopy together with electron diffraction was also
used to screen the presence of microorganisms and to investigate crystallographic aspects of
the mineral components of the sialolith. Bacteria and globules with nanometric hydroxyapatite
crystals were found. Atomic Force Microscopy was used to probe the mechanical behaviour and
adhesivity of the sialolith components. The Electrophoresis technique was used to detect and
identify the proteins existing in sialoliths. Several proteins from leukocytes and epithelial cells,
including proteolytic enzymes, were detected. From the results obtained it has been possible to
provide new information on the role of the globular mechanism in sialolith formation.
This work contributes to a better understanding of the sialolith structure and mechanical
properties, important for the development of new techniques for the treatment of sialolithiasis.
Keywords:
Sialolith, Sialolithiasis, Scanning Electron Microscopy, Transmission Electron Microscopy,
Atomic Force Microscopy, Electrophoretic analysis.
v
Resumo
O objectivo principal deste trabalho é a caracterização química, proteómica e mecânica de
sialolitos para explicar a existência de colagénio desnaturado, para justificar a presença de
enxofre, estabelecer um protocolo para a medição das propriedades mecânicas de um sialolito
e fornecer novas informações sobre o mecanismo de formação dos sialolitos.
Para este estudo foi utilizada a Microscopia Electrónica de Varrimento em conjunto com os
mapas de raios-X, a fim de investigar os aspectos morfológicos e químicos de um sialolito. Os
resultados mostraram que o sialolito era composto por matéria orgânica e mineral, organizado
em camadas em torno do núcleo altamente mineralizado. A Microscopia Electrónica de
Transmissão em conjunto com a difracção de electrões foi usada para observar a presença de
microorganismos e investigar aspectos cristalográficos dos componentes minerais existentes
no sialolito. Bactérias e glóbulos foram encontrados contendo cristais de hidroxiapatite
nanométricos. Usando a Microscopia de Força Atómica foi possível estudar o comportamento
mecânico e adesividade dos componentes num sialolito. A técnica de Electroforese foi
utilizada para detectar e identificar as proteínas existentes no sialolito. Foram detectadas
várias proteínas de leucócitos e de células epiteliais, incluindo enzimas proteolíticas. Com os
resultados obtidos, foi possível fornecer novas informações sobre o mecanismo globular na
formação dos sialolitos.
Este trabalho contribui para uma melhor compreensão da estrutura dos sialolitos e das suas
propriedades mecânicas, importantes para o desenvolvimento de novas técnicas para o
tratamento de sialolitíase.
Palavras-chave:
Sialolito, Sialolitíase, Microscopia Electrónica de Varrimento, Microscopia Electrónica de
Transmissão, Microscopia de Força Atómica, Análise Electroforética.
vi
Contents
Acknowledgements ............................................................................... ii
Agradecimentos .................................................................................... iii
Abstract ................................................................................................. iv
Resumo ................................................................................................... v
List of Figures ..................................................................................... viii
List of tables ........................................................................................... x
List of Abbreviations ............................................................................ xi
List of Symbols .................................................................................... xii
1. Introduction ........................................................................................ 1
2. Materials and Methods ...................................................................... 6
2.1 Scanning Electron Microscopy ....................................................................... 6
2.1.1 Beam-specimen interactions - signal types ............................................ 7
2.1.2 Energy Dispersive Spectroscopy and X-ray mapping ........................... 8
2.1.3 Sample preparation .................................................................................. 9
2.2 Transmission Electron Microscopy ................................................................ 6
2.2.1 Imaging techniques ................................................................................ 10
2.2.2 Diffraction techniques ............................................................................. 11
2.2.3 Sample preparation (Ultramicrotomy) ................................................... 12
2.3 Atomic Force Microscopy .............................................................................. 15
2.3.1 Young’s Modulus and Adhesion Force ................................................. 16
2.4 Proteomic analysis ........................................................................................ 19
2.4.1 Electrophoresis ....................................................................................... 19
2.4.2 Sample Preparation ................................................................................ 20
2.4.3 Electrophoretic analysis ......................................................................... 20
2.4.4 Processing of the polypeptide bands for sequencing .......................... 20
vii
3. Results and Discussion .................................................................. 22
3.2 Scanning Electron Microscopy ..................................................................... 22
3.2 Transmission Electron Microscopy .............................................................. 29
3.3 Atomic Force Microscopy .............................................................................. 35
3.4 Proteomic analysis ........................................................................................ 40
4. Globular Mechanism ........................................................................ 42
5. Conclusion ....................................................................................... 43
6. Future work ...................................................................................... 44
7. Bibliography ..................................................................................... 45
viii
List of Figures
Figure 1 – Salivary glands [1]. ................................................................................................... 1
Figure 2 – Schematic drawing representing the globular and sedimentary cycles in the sialoliths
formation mechanism [2]. .................................................................................................. 4
Figure 3 – Schematic representation of the electron column of SEM [18]. .................................. 7
Figure 4 – Embedding procedure yielding a mounted sample, which was subsequently polished
and used for an SEM observation. The non-embedded part subjected to the
Electrophoresis technique [2]. ............................................................................................ 9
Figure 5 – Imaging techniques: Bright Field Imaging and Off-axis Dark Field [23]. ................... 11
Figure 6 – Comparison between imaging mode and diffraction mode [23]. ............................... 12
Figure 7 – Ultramicrotomy using a diamond knife [24]. ............................................................. 13
Figure 8 – Schematic of an atomic force microscope [27]. ....................................................... 15
Figure 9 - Schematic position sensitive detector current signal (IPSD) vs. piezo position (Zp) curve
including approaching and retracting parts. Three types of hysteresis can occur: In the zero
force line (A), in the contact part (B) and adhesion (C) [26]. ............................................. 17
Figure 10 - Polished cross-section of Sialolith 1a (BSE mode). ................................................ 22
Figure 11 – Detail A of Sialolith 1a polished cross-section, from where X- ray maps shown in
Figure 12 were obtained (BSE mode). ............................................................................. 23
Figure 12 – (a), (b) and (c) are the Ca, P and S X-ray maps respectively of detail A of the
Sialolith 1a polished cross-section (BSE mode). .............................................................. 24
Figure 13 – Typical EDS spectra obtained at (a) bright and (b) dark regions [2]. ...................... 24
Figure 14 – Detail B of the Sialolith 1a polished cross-section display the highly mineralized
core. The arrows indicate the teardrop globules. .............................................................. 25
Figure 15 – Detail B of the Sialolith 1a polished cross-section. Parts (a) and (b) are convolutions
of mineralized and weakly mineralized globule layers. ..................................................... 25
Figure 16 – Detail C of the Sialolith 1a polished cross-section displays the core surrounded by
an incomplete layer of organic matter with spherical globules (indicated by the arrow) (BSE
mode).............................................................................................................................. 26
Figure 17 – Detail D of the Sialolith 1a polished cross-section shows (indicated by the arrow)
the globules surrounded by crescents of smaller teardrop-shaped structures (BSE mode).
....................................................................................................................................... 26
ix
Figure 18 – Detail E of the Sialolith 1a polished cross-section. (a) Laminar layers (b) layers of
teardrop-shaped (indicated by the uppermost arrow) and spherical globules (indicated by
the lowermost arrow) (BSE mode). .................................................................................. 27
Figure 19 – Decalcified sample. (a) Organic matter without HA crystals, the arrows indicate the
spaces in organic matter where HA crystals were present before decalcification (b)
globules (Y) with a few HA crystals inside, the arrows indicate HA crystals. ..................... 29
Figure 20– Calcified sample. (a) Globules (Y) surrounded by HA crystals in organic matter and
(b) HA crystals inside a globule, the arrows indicate the HA crystals. ............................... 30
Figure 21 – Calcified sample. (a) Dark field and (b) Bright field images of HA crystals obtained in
the same region of the sample. ........................................................................................ 31
Figure 22 – The ring pattern diffraction of HA crystals obtained from a region similar to the one
presented in Figure 21. .................................................................................................... 31
Figure 23 – Calcified sample with bacteria. (a) and (b) bacteria (B). The arrow point to the
occurrence of cell division of bacteria. ............................................................................. 32
Figure 24 – Calcified sample. (a) Dark field and (b) Bright field images of HA crystals inside and
surrounding the bacteria. ................................................................................................. 33
Figure 25 – Region A of detail A of the Sialolith 1a polished cross-section obtained in BSE
mode by SEM. Region A is located in a highly mineralized area....................................... 35
Figure 26 – Region B of detail A of the Sialolith 1a polished cross-section obtained in BSE
mode by SEM. Region B is located in a weakly mineralized area. .................................... 36
Figure 27 – Topographic images of regions A (scan size: 13 μm x 13 μm) and B (scan size: 40
μm x 40 μm) from the polished cross-section. (x) and (y) are points selected in order to
obtain force-distance curves of each region. .................................................................... 36
Figure 28 – Example of the force-distance curve obtained at point x (Figure 21). Force-distance
curve describing a single approach-retract cycle of the AFM tip: (A) The AFM tip is
approaching the sample surface and the long range interactions are too small to give
measurable deflections; (B) the initial contact between the tip and the surface is mediated
by attractive Van der Waals forces (contact) that lead to an attraction of the tip toward the
surface; (C) the tip applies a constant and default force upon the surface that leads to
sample indentation, cantilever deflection and increase of repulsive forces; (D)
subsequently, the tip tries to retract and to break loose from the surface; (E) various
adhesive forces between the sample and the AFM tip, however, hamper tip retraction.
These adhesive forces can be taken directly from the force-distance curve; (F) the tip
looses contact with the surface upon overcoming of the adhesive forces [2]..................... 37
Figure 29 - Young’s (elastic) moduli of different materials. The diagram shows a spectrum from
very hard to very soft: steel > bone > collagen > protein crystals > gelatin, rubber > cells
[34]. ................................................................................................................................. 38
x
List of tables
Table 1 – Mean values (30 trials) of (N/m) and (N), and their standard deviations for the
bright and dark regions. ............................................................................................................ 37
Table 2 – Young’s modulus (MPa) for the bright and dark regions................................................ 38
Table 3 – Proteins with high molecular weight except Humam Cathepsin G (low molecular
weitht) and their accession number. ........................................................................................ 40
Table 4 – Proteins with high molecular weight and their accession number................................. 40
xi
List of Abbreviations
AFM Atomic Force Microscopy
BSE Backscattered Electrons
CPLX Calcium-Acidic Phospholipid-Phosphate Complexes
CRT Cathode Ray Tube
CT Computed Tomography
DEFA-4 Human Alpha-Defensin-4
DTT Dithiothreitol
EDS Energy Dispersive X-ray Spectroscopy
EDTA Ethylenediaminetetraacetic Acid
EISWL Endoscopic Intracorporeal Shock-Wave Lithotripsy
ESWL Extracorporeal Shock-Wave Lithotripsy
FS Force Spectroscopy
GAGs Glycosaminoglycans
HA Hydroxyapatite
HLE Human Leukocyte Elastase
HNP-4 Human Neutrophil Peptided-4
LASER or laser Light Amplification by Stimulated Emission of Radiation
LDS Lithium Dodecyl Sulphate
MALDI Matrix-Assisted Laser Desorption/Ionization
MOPS [3-(N-Morpholino) Propane Sulfonic acid]
MS Mass Spectrometry
NCBI National Center for Biotechnology Information
ND-4 Neutrophil Defensin 4
NuPAGE Novex Bis-Tris [Bis (2-hydroxyethyl) imino-tris (hydroxymethyl) methane-HCL]
SDS Sodium Dodecylsulfate
SDS-PAGE Sodium Dodecyl Sulfate Polyacrylamide Gel Electrophoresis
SE Secondary Electrons
SEM Scanning Electron Microscopy
TEM Transmission Electron Microscopy
TOF Time-Of-Flight mass spectrometer
xii
List of Symbols
contact radius (m)
interplanar spacing (m)
tip-sample distance (m)
, Young’s modulus of tip and sample material (Pa)
reduced Young’s modulus (Pa)
adhesion force (N)
spring constant cantilever (N/m)
effective spring constant (N/m)
sample Stiffness (N/m)
camera Constant
ring radius (m)
potential energy between tip and sample (J)
Hooke’s elastic potential of the cantilever (J)
tip-sample interaction potential (J)
Hooke’s elastic potential of the sample (J)
deflection of the cantilever at its end (m)
height position of the piezoelectric translator (m)
Greek letters
Indentation (m)
Wavelength (m)
Poisson’s ratio of tip and sample material (Pa)
xiii
“Godi dei tuoi successi e anche dei tuoi progetti.
Mantieni interesse per la tua professione, per quanto umile:
essa costituisce un vero patrimonio nella mutevole fortuna del tempo.”
Desiderata
1
1. Introduction
Salivary glands are the glands that secrete saliva. The three major salivary glands in the mouth
are the parotid, the submandibular and the sublingual glands, and there are two of each, one on
either side of the mouth (Figure 1). There are also other smaller salivary glands within the
cheeks and tongue. The largest salivary glands are the parotids, located below and in front of
each ear. Saliva secreted by them is discharged into the mouth through openings in the cheeks
opposite the upper teeth. The submandibular glands, located inside the lower jaw, discharge
saliva upward through openings into the floor of the mouth. The sublingual glands, beneath the
tongue, also discharge saliva into the floor of the mouth [1].
Figure 1 – Salivary glands [1].
Sialadenitis is a salivary inflammation caused by the obstruction of a salivary gland or duct, of
which one of the major causes is sialolithiasis. Sialolithiasis is the formation or presence of a
salivary calculus or sialolith in the intra or extra-glandular duct system. This disease represents
30% of all salivary pathologies, is more frequent in adults than in children, and affects 1.2% of
the general population [2].
A small sialolith that still allows saliva to flow through the duct may not cause any symptoms.
However, when a sialolith reaches a size that obstructs the passage completely, secretion by
the gland is impeded as a result of an increase in pressure, and this causes pain and swelling
2
[2, 3]. Sialoliths may also cause stasis of saliva, facilitating bacterial ascent into the gland and
subsequent infection. Long-term obstruction in the absence of infection can lead to atrophy of
the gland with resultant lack of secretory function and ultimately fibrosis [2].
Sialolithiasis is a common condition and occurs mainly in the submandibular gland (80-90% of
all cases), and to a lesser degree in the parotid gland (5-20%). The sublingual gland and the
minor salivary glands are very rarely affected [4, 5]. Single salivary calculus occur in 70-80% of
cases. Multiple sialolithiasis is rare although a few reports of it exist in the literature. Calculi can
be found in both salivary glands and ducts. Parotid calculi are found in the helium or
parenchyma of the gland in 50% of cases, while submandibular calculi are found mainly in the
duct system. The calculi vary in size and shape. Sialoliths are typically round or oval-shaped
when formed in the gland, and elongated when formed in the ducts. Parotid calculi are usually
smaller and more numerous [3]. In total, the recurrence rate of sialolithiasis has been estimated
to be 8.9% [2].
Individual differences in the structure and mineralization of sialoliths exist [2, 3]. In general,
sialoliths are composed of an organic matter and inorganic minerals distributed by a central
core and a laminar peripheral structure [3]. The nucleus of the sialolith is mostly inorganic,
predominantly made up of calcium phosphate and calcium carbonate in the form of
hydroxyapatite [Ca10(PO4)6(OH)2] although other crystallographic forms of calcium phosphate
may be present [6]. Around this amorphous nucleus laminar layers of organic and inorganic
substances (whose content varies within a single sialolith) accumulate. The organic material is
predominantly concentrated in the outer shell of a sialolith and its components are mostly
glycoproteins, mucopolysaccharides, lipids and cell detritus [2, 6]. In this region a large zone
(50-210 μm) of connective tissue and mataplastic squamous epithelium has been detected,
along with the enzymatic activity of acid phosphatase, lactate dehydrogenase, succinate
dehydrogenase and maleate dehydrogenase. Uric acid calculi may form in patients with gout.
Human submandibular-gland sialoliths contain lipids whose composition resembles that of
plasma membranes, suggesting that these lipids may originate from cell debris. Associated with
the mineral phase of submandibular sialoliths are specific lipid components, calcium-acidic
phospholipid-phosphate complexes (CPLX). These complexes induce hydroxyapatite deposition
[7]. In a recent investigation, the organic matter in the globular structures and organic strata was
found to have a high concentration of denatured collagen, probably originating from fibroblast
cell invasion and connective tissue formation in the sialolith precursor structures [2].
Salivary calculi are predominantly composed of elements including Ca, P, O, C and N, and
small amounts of Mg, Na, Cl, Si, Fe, K [2, 3]. The presence of Mg in the low concentration
facilitates precipitation of apatite crystals, which may explain the very high mineral content of
some calculi. Although a higher content of Mg inhibits crystal formation [8]. The presence of a
high sulfur content in the organic regions has been reported [2, 9] but remains unexplained.
Sulfur is an important constituent of Glycosaminoglycans (GAGs), which are present in saliva
and may be relevant in the calcification process [9].
3
The diagnosis of sialolithiasis is based on the medical history and a physical examination of the
patient. Recurrent pain and swelling of the gland, especially during meals, are common
symptoms of sialolithiasis. Imaging techniques are of fundamental importance for a positive
diagnosis: occlusal radiographs reveal radiopaque calculi, while radiolucent calculi are detected
through sialography. Ultrasound and Computed Tomography (CT) scans are other techniques
used in the detection of salivary stones [2].
Stones of the submandibular gland have a greater inorganic percentage (about 82%) than those
of the parotid gland (about 50%) and the larger the stone, the smaller the organic fraction. Many
parameters are important for the choice of therapy: the history of complications and complaints,
the position, size, and number of stones, their echogenity and adherence to the duct, and the
diameter of the duct between the stone and the papilla [2, 9]. Sialoliths can be successfully
treated by radiologically, fluoroscopically or sialendoscopically-based methods in approximately
80% of cases. Endoscopy techniques reveal a success rate of 86% in parotid sialolithotomy and
89% in submandibular sialolithotomy [2, 10]. Other non-invasive therapeutic techniques
available to treat sialolithiasis are Extracorporeal Shock-Wave Lithotripsy (ESWL) (which is
successful in up to 50% of cases) and Endoscopic Intracorporeal Shock-Wave Lithotripsy
(EISWL) (30-45%). These procedures use shock-waves to shatter salivary calculi, enabling the
fragments to pass out the duct [11] and the therapy was adopted from the lithotripsy of renal
and bladder calculi [12]. However, this method has success rates for treatment of sialolithiasis
typically lower than those achieved in the treatment of renal calculi which are essentially (95%)
calcium stones [13]. The reasons underlying this fact are still unexplained and require further
investigation. Preliminary work carried out suggests that this behaviour is related to the
ultrastructural features of salivary calculi, particularly with the presence of collagen [2, 9].
Transoral Duct Slitting is another important method used on extraparenchymal submandibular
stones, which has a success rate of 90%. Operative duct procedures and the combined
Endoscopic-Transcutaneous approach complete the spectrum of treatment possibilities for the
Parotid gland. Sialendoscopy plays a central role in the treatment of obstructive salivary gland
diseases, but maximum success can only be attained through a combination of all of these new,
minimally invasive, techniques. Through adopting a minimally invasive and multimodal policy, a
significant number (74%–100%, technique dependent) of salivary calculi can be safely and
successfully retrieved while leaving an intact and functional salivary gland system. Only 2% to
5% of patients require gland excision [2, 11, 14].
Many theories have been put forward to explain the etiology and pathogenesis of sialolithiasis.
The exact cause of the formation of a calculus is still unknown but a possible mechanism for
sialolith formation has been proposed, based on alternating sedimentary and globular cycles
(Figure 2). The globular cycle involves connective tissue invasion followed by the calcification of
the organic structures. A globular/sedimentary cycles mechanism for sialoliths formation was
proposed by Olga Santo [2] comprehending the following stages:
(1) Initial foreign body or luminal organic nidus formation.
4
(2) Fretting of foreign body/nidus damages duct/gland epithelial cells.
(3) Damage of duct/gland walls exposing the foreign body/nidus to connective tissue. Bacterial
infection may develop in the damaged tissue causing an inflammation.
(4, 5) Foreign body/nidus invades the connective tissue, followed by scar tissue. This tissue
contains collagen and an extracellular matrix. The collagen may be synthesized by fibroblast
cells, which are a constituent of connective tissue.
(6) Separation of the foreign body/nidus which interrupts vascularization of the living connective
tissue.
(7) Death of cells and living tissue (necrosis).
(8) Calcification of organic matter.
The structures resulting from stage (2) to (8) are globular. Alternatively, sialolith growth can
develop through the sedimentary precipitation of highly mineralized strata intercalated with thin
organic layers through a sedimentary cycle (9). The foreign body/nidus formation stage may be
followed directly by the sedimentary cycle. The globular cycle may occur subsequent to the
sedimentary cycle, or not at all. It may also occur directly after the foreign body/nidus formation
stage if the sedimentary cycle (9) does not take place [2, 9].
Figure 2 – Schematic drawing representing the globular and sedimentary cycles in the sialoliths formation
mechanism [2].
The objective of the present work is the chemical, proteomic and mechanical characterization of
two sialoliths using Scanning Electron Microscopy (SEM) coupled with X-ray mapping,
Transmission Electron Microscopy (TEM) combined with electron diffraction, Atomic Force
Microscopy (AFM) and Electrophoresis. The work aimed to explain the existence of denatured
5
collagen, justify the presence of sulfur, establish a protocol for the measurement of the
mechanical properties of a sialolith, show and provide new information on the formation
mechanism.
6
2. Materials and Methods
The sialoliths were extracted in the Maxillofacial Surgery of S. José Hospital, Lisbon and kept
dry in medical containers. Two sialoliths were used for this study: Sialolith 1 was divided into
two parts; Sialolith 1a was a polished cross-section examined with SEM and AFM and the other
part of the sialolith, Sialolith 1b, was prepared for TEM. Sialolith 2 was powered for the
electrophoresis procedure.
2.1 Scanning Electron Microscopy
The scanning electron microscope is capable of producing images with a resolution of 1 nm of a
sample surface, by scanning it with a high-energy beam of electrons. The two major
components of an SEM are the electron column (Figure 3) and the control console. The electron
column consists of an electron gun, a lens system, a scan deflection system, a specimen stage
and a secondary electron detector. The electron gun produces a beam of electrons (which can
be thermionically emitted from a tungsten or lanthanum hexaboride cathode, or alternatively
generated via field emission) which are accelerated towards an anode. The lens system
(condenser lenses and objective lens), composed of electromagnetic lenses, is involved in
producing a small, focused spot of electrons that are then rastered over a specimen’s surface
by means of a scan deflection system (scanning coils), which deflects it horizontally and
vertically so that it scans in a raster fashion, in a single rectangular area, over the sample
surface. A specimen stage is needed so that the specimen may be inserted and suitably
situated relative to the beam. A secondary electron detector is used to collect the electrons and
to generate a signal that is processed by electronics and ultimately displayed on viewing and
recording monitors. The base of the column is usually taken up with vacuum pumps that
produce a vacuum to remove air molecules that might impede the passage of the high-energy
electrons down the column. The control console consists of a Cathode Ray Tube (CRT) viewing
screen or a computer that digitally displays the detectors signal, knobs and a computer
keyboard that control the electron beam [2, 15-17].
7
Figure 3 – Schematic representation of the electron column of SEM [18].
2.1.1 Beam-specimen interactions - signal types
When the electrons, accelerated in the SEM to 15 to 25 KV, strike the specimen, they give rise
to a number of signs. Depending on the speed of the electrons, as well as the density of the
specimen, the beam may penetrate to a variable depth in the specimen. A longitudinal section
across the point of entry of the electron beam will reveal a teardrop-shaped area. When the
primary electron beam is focused on the material, the electrons lose energy from repeated
scattering and absorption within a teardrop-shaped volume of the specimen, known as the
interaction volume, which extends from less than 100 nm to around 5 μm into the specimen.
The spatial resolution is primarily defined by the size of this interaction volume, which depends
on the beam accelerating voltage, the average atomic number of the specimen and the
specimen’s density [2, 17].
The beam electrons that interact with the atoms of the specimen, resulting in the production of
low energy, are called Secondary Electrons (SE). SE have energy ranges from 0 to 50 eV and
are the electrons most commonly used to generate the three-dimensional image. Generally the
SE are produced as a result of interactions between energetic primary electrons and weakly
bound conduction electrons in the sample. An important characteristic of SE is their shallow
sampling depth, a direct consequence of the low kinetic energy with which they are formed, and
that justifies the high spatial resolution attainable with this signal type typically within 1-2 nm [2,
17, 19]. After the beam electrons strike the specimen, low energy secondary electrons leave the
8
specimen from many different angles. Because they are weakly negative, the SE will be
attracted to any positive source. The secondary electron detector uses this phenomenon to
gather electrons. The brightness of the SE signal depends on the number of electrons reaching
the detector. If the beam enters the sample perpendicularly to the surface, then the activated
region is uniform about the axis of the beam and a certain number of SE “escape” from within
the sample. As the angle of incidence increases, the “escape” distance of one side of the beam
will decrease, and more secondary electrons will be emitted. Thus steep surfaces and edges
tend to be brighter than flat surfaces, and this result in topographical images with well-defined,
three-dimensional appearance [2, 17].
Backscattered Electrons (BSE) are high-energy electrons and can escape from a much larger
volume than SE. This type of scattering occurs when a beam electron collides with, or passes
close to, the nucleus of an atom of the specimen. The intensities of the different types of signal
depend on the average atomic number of the specimen, but are less dependent on topography.
BSE images have lower spatial resolution than SE images due to the larger volume of their
origin, but can show contrast between areas with different chemical compositions, since the
signal intensities increase with the average atomic number in the sample [2, 17].
2.1.2 Energy Dispersive Spectroscopy and X-ray mapping
Energy Dispersive X-ray Spectroscopy (EDS) is a standard technique for element identification
in material analysis. EDS systems are mounted on SEM and use the primary beam of the
microscope to generate characteristic X-rays. The composition of the sample is found by
analyzing the energy of the characteristic X-rays. The spatial resolution of EDS depends on the
sample material and the energy of the primary beam of the SEM [19]. When the sample is
bombarded by the SEM's electron beam, electrons are ejected from the atoms comprising the
sample's surface. The resulting electron vacancies are filled by electrons from a higher state,
and an X-ray is emitted to balance the energy difference between the two electrons' states. The
X-ray energy is characteristic of the element from which it was emitted. The EDS X-ray detector
measures the relative abundance of emitted X-rays in relation to their energy. The detector is
typically a lithium-drifted silicon, solid-state device [19, 20].
A qualitative analysis is possible: the sample X-ray energy values from the EDS spectrum are
compared with known characteristic X-ray energy values to determine the presence of an
element in the sample. Elements with atomic numbers ranging from that of Beryllium to Uranium
can be detected.
It is possible to obtain an elemental mapping (X-ray mapping): characteristic X-ray intensity is
measured relative to lateral position on the sample. Variations in X-ray intensity at any
characteristic energy value indicate the relative concentration for the applicable element across
the surface. One or more maps are recorded simultaneously using image brightness intensity
9
as a function of the local relative concentration of the element(s) present. About 1 μm lateral
resolution is possible [20].
In this present work, Scanning Electron Microscopy observations in BSE modes, as well as X-
ray mapping were carried out with a Hitachi S-2400 microscope operated at 25 kV and
equipped with a standard Rontec EDS detector. The SEM work consisted of an observation of
the polished sialolith cross-section to obtain structural and chemical information.
2.1.3 Sample preparation
The specimen (Sialolith 1) was embedded in an epoxy resin by cold mounting. After a
polymerization period of 24 hours the non-embedded part of the sample (Sialolith 1b) was cut
and kept for subsequent TEM observation (Figure 4). Special care was put into this operation to
avoid contamination of the sample. The resin-embedded cross-section was grounded with SiC
paper and polished using alumina suspensions in water.
Figure 4 – Embedding procedure yielding a mounted sample, which was subsequently polished and used
for an SEM observation. The non-embedded part subjected to the TEM [2].
The polished sample (Sialolith 1a) was mounted in metallic supports and coated with carbon in
a Polaron E-5100 coating unit prior to observation to prevent charge accumulation.
10
2.2 Transmission Electron Microscopy
Transmission electron microscope is an analytical tool that allows detailed investigation of the
morphology, structure and local chemistry of metals, ceramics, polymers, biological material and
minerals. It also enables the investigation of crystal structures, crystallographic orientations
through electron diffraction, as well as second phase precipitates’ and contaminants’
distributions by EDS. Magnifications of up to 500,000x and detailed resolutions below 1 nm are
achieved routinely. Quantitative and qualitative elemental analysis can be provided from
features smaller than 10 nm [16].
The instrument includes an electron gun which emits electrons into the vacuum and accelerates
them between the cathode and anode. The most common types of TEM have thermionic guns
capable of accelerating the electrons through a selected potential difference in the range 60 to
120 kV. Schottky-emission and field emission guns are newer alternatives for which the energy
spread is less and the gun brightness higher. The condenser lens system of the microscope
controls the specimen illumination, which ranges from uniform illumination of a large area at low
magnification, through a stronger focusing for high magnification. TEM specimen stage designs
include airlocks to allow for the insertion of the specimen holder into the vacuum with a minimal
increase in pressure in other areas of the microscope. The specimen holders are adapted to
hold a standard size of grid upon which the sample is placed. The standard TEM grid size is a
3.05 mm diameter ring, with a thickness and mesh size ranging from a few to 100 μm. The
sample is placed onto the inner meshed area which has a diameter of approximately 2.5 mm.
The grid is typically made of copper, except when analytical methods are employed and in this
case it can be made of molybdenum, gold or platinum and is placed into the sample holder.
Observation of the images on a fluorescent screen, and the recording of image on photographic
emulsions are currently replaced by techniques that allow digital, parallel, and quantitative
recording of the image intensity [21, 22].
2.2.1 Imaging techniques
Bright Field imaging is used for the examination of most ultrastructures. In order to examine
samples in Bright Field, the objective aperture must be inserted (Figure 5). The objective
aperture is a metal plate with holes of various sizes drilled into it. The aperture is inserted into
the back focal plane, the same plane at which the diffraction pattern is formed. The back focal
plane is located just below the sample and the objective lens. When the aperture is inserted, it
only allows the electrons in the transmitted beam to pass and contribute to the resulting Bright
Field image. When an electron beam strikes a sample, some of the electrons pass directly
through, while others may undergo slight inelastic scattering from the transmitted beam. The
11
differences in scattering create the contrast in an image. By inserting an aperture in the back
focal plane, an image can be produced with these transmitted electrons [16].
Dark Field images are produced when the objective aperture is positioned off-axis from the
transmitted beam in order to allow only a diffracted beam to pass (Figure 5). In order to
minimize the effects of lens aberrations, the diffracted beam is deflected along the optical axis.
Dark Field images are particularly useful in examining ultrastructural detail in single crystalline
phases [16].
Figure 5 – Imaging techniques: Bright Field Imaging and Off-axis Dark Field [23].
2.2.2 Diffraction techniques
The objective lens simultaneously generates the diffraction pattern and the first intermediate
image. The ray paths are identical until the intermediate lens, where the field strengths are
changed, depending on the desired operation mode (Figure 6). The field strengths can be
changed by setting the focal lengths (the distance from the lens to the ray crossover). Higher
field strength (shorter focal length) is used for imaging, whereas weaker field strength (longer
focal length) is used for diffraction [16].
12
Figure 6 – Comparison between imaging mode and diffraction mode [23].
Electron diffraction was used to produce ring patterns of crystalline material in sialolith. This
type of pattern is created when electron diffraction occurs simultaneously from many grains with
different orientations relative to the incident electron beam. Ring patterns can be used to identify
unknown phases or characterize the crystallography of a material. The radii and spacing of the
rings are governed by [16]:
(1)
Where is the interplanar spacing, is the ring radius, and is known as the camera constant.
2.2.3 Sample preparation (Ultramicrotomy)
It is not straight-forward to make a specimen thin enough for TEM (from a few tenths of a
nanometre, to a micron in thickness). The task is made harder still by the need to avoid
changes in the specimen due to the preparation technique, and to obtain a representative
region. The sample must also be strong enough to handle, and last at least long enough to be
13
examined in the microscope [22]. Specimen preparation techniques can be divided into two
basic approaches. First: the removal of unwanted material, by either chemical or mechanical
means, until only a very thin specimen is left behind. Second: cutting the sample is either cut
with a knife or cleaving along crystallographic planes so that a very thin specimen, or region of a
specimen, is produced [22]. In this present work the second was used: the material was cut with
a knife using an ultramicrotome. An ultramicrotome is a slicing instrument developed from the
larger-scale devices used for cutting tissue sections for biological materials. In an
ultramicrotome a firmly mounted or embedded specimen with an area less than 1 mm x 1mm is
moved past a fixed knife made of diamond (Figure 7). The resulting slices are collected in a
liquid-filled trough and are mounted on grids before being inserted into the microscope [22]. In
this present work, specimens needed to be embedded in epoxy resin before sectioning, in order
to provide support during cutting.
Figure 7 – Ultramicrotomy using a diamond knife [24].
The samples used for TEM observation were cut from Sialolith 1b. Two types of samples were
prepared, calcified and decalcified. To obtain these samples the following steps were
performed:
Part of the Sialolith 1b was fractured into small pieces.
For decalcification some fragments were immersed in EDTA at 25% in glutaraldehyde
0.1 M at 3%, pH 7.2 for approximately one month, changing the solution every two days
until the solution becomes translucent. After this procedure were decalcified and two
fixative procedures were necessary. The first fixation was carried out with
glutaraldehyde 3% in sodium cacodylate 0.1 M, pH 7.3. The second fixation was the
immersion in osmium tetroxide 1% in sodium cacodylate to cross-link and stabilize cell
and organelle membrane lipids.
14
For the calcified sample, only the fixation (mentioned above) procedure was performed.
Both calcified and decalcified fragments were subsequently dehydrated in ethanol,
followed by embedment in an epoxy resin by cold mounting for 24 hours.
The embedded samples were cut using an Ultramicrotome Reichert.
The samples were stained with uranyl acetate 2% in bidistilled H2O for 20 minutes and
lead citrate (Reynolds) for 4 minutes.
The samples were finally placed on a copper grid with an amorphous carbon
membrane.
The calcified and decalcified samples were both observed using a JEOL - JEM 100 SX TEM
operated at 100kV and by Hitachi H-8100 TEM was operated at 200 kV. Ring diffraction
patterns were obtained with the latter instrument.
15
2.3 Atomic Force Microscopy
The high-resolution images of the AFM are obtained by measuring and controlling the
interaction force between tip and sample. First, a tip at the end of a cantilever approaches the
sample, and then scans over the surface, driven by a piezoelectric actuator (Figure 8). While
scanning, the force between the tip and the sample is measured by monitoring the deflection of
the cantilever. The deflection is detected by a position-sensitive photodiode detector, onto which
the cantilever reflects a laser light. Finally, a topographic image of the sample is obtained by
plotting the defection of the cantilever in relation to its position on the sample [25, 26].
Figure 8 – Schematic of an atomic force microscope [27].
According to the different modes of interaction between tip and sample, the operation modes
could be classified into: non-contact mode, contact mode, and intermittent contact mode. In this
present work contact mode was used. In this mode, the tip remains in contact with the target
sample.
Force Spectroscopy (FS) is used to measure forces of interaction between the probe tip and the
surface sample as a function of their distance. The base of the cantilever is moved slowly in the
vertical-z direction towards the sample surface and retracted again after contact or indentation
(or vice versa). This technique provides information on local elasticity and adhesion force [25].
16
2.3.1 Young’s Modulus and Adhesion Force
From the contact lines of force-displacement curves it is possible to retrieve information on the
elastic–plastic behavior of materials.
In the contact part of force curves, both in the approach and in the retraction stages, the elastic
deformation of the sample can be related to its Young’s modulus. In order to relate the
measured quantities to the Young’s modulus, it is necessary to consider the deformation of the
sample . For elastic deformation it is useful to describe the system by means of a potential
energy [26]:
(2)
Where is the tip-sample interaction potential caused by surface forces, the energy due to
bending of the cantilever, the elastic deformation energy of the sample, is the so-called
sample stiffness (slope values of force-distance curve), the spring constant cantilever, the
deflection of the cantilever at its end and the indentation. In general, it can be written thus
[26]:
(3)
In contact , is the tip-sample distance, and if the system is in equilibrium, also
. Substituting, it obtains:
. (4)
This simple relation shows that the slope of the force-displacement curve is a measure of the
stiffness of the sample. If the sample is much stiffer than the cantilever, that is for then
, whereas when , i.e., when the sample is much more compliant than
the cantilever. This gives also a rule of thumb for the choice of the cantilever spring constant in
experiments dealing with the elastic properties of the sample: if the cantilever spring constant is
much lower than the sample spring constant, the force curve will probe primarily the stiffness of
the cantilever, and not that of the sample [26].
17
The stiffness of the sample is related to its Young’s modulus by:
with
. (5)
Here, and are the Poisson’s ratio and the Young’s moduli of tip and sample,
respectively, the reduced Young’s modulus, and is the tip-sample contact radius.
AFM force-distance curves have become an important method for studying the adhesion force.
Adhesion occurs when retracting the tip from the sample surface. The tip stays in contact with
the surface until the cantilever force overcomes the adhesive tip-sample interaction [26].
In the most general sense adhesion force is a combination of electrostatic force, van der Waals
force, meniscus or capillary force and forces of chemical bond or acid–base interactions. In
many of the AFM studies on adhesion force ( ), conditions were chosen such that the van der
Waals forces were expected to dominate. In this case should be determined by the
Hamaker constants of the AFM probe, the sample and the contact geometry. Quantitative
comparison of such experiments with theoretical predictions is hampered by several factors:
Surface roughness has a pronounced influence on adhesion force that is hard to quantify; the
precise contact geometry is often hard to determine and adsorption of contaminants on high
energy solid surfaces leads to chemical inhomogeneities of the surfaces [26]. The adhesion
force can be calculated from the force-distance curve and it is represented in Figure 9.
Figure 9 - Schematic position sensitive detector current signal ( ) vs. piezo position ( ) curve
including approaching and retracting parts. Three types of hysteresis can occur: In the zero force line (A),
in the contact part (B) and adhesion (C) [26].
18
Variations in the shape of force curves taken at different locations indicate variations in the local
nano-scale properties of the specimen surface. The shape of the curve is also affected by
contaminants and surface lubricants, as well as the water content of the surface layer of the
specimen when operating an AFM in air [28].
Topographic images of regions mapped by SEM and X-ray maps were obtained under contact
mode in specific areas of Sialolith 1a (polished cross-section). FS was used to obtain force-
distance curves at selected points of the sample, enabling the assessment of the variations of
the adhesion force and Young’s modulus. AFM/FS work was performed at room temperature
and humidity with a Veeco DI CP-II Atomic Force Microscope using commercially available
silicon cantilevers and coating with Pt/Ir (SCM-PIT, Veeco), with a nominal spring constant of
2.5 N/m and a nominal tip radius of 20 nm after sensitivity calibration.
19
2.4 Proteomic analysis
2.4.1 Electrophoresis
Electrophoresis is a simple, rapid, and sensitive analytical tool for separating proteins. An
electric current is passed through a medium containing the target mixture. Each different type of
molecule travels through the medium at a different rate, depending on its electrical charge
and/or size. Acrylamide gel is the most common media for the electrophoresis of proteins. In the
present work, a NuPAGE system was used: the NuPAGE Novex Midi Gel system, which is a
pre-cast polyacrylamide gel system. NuPAGE Novex Midi Gels are used with the NuPAGE Bis-
Tris SDS Buffer System to produce a discontinuous SDS-PAGE system operating at neutral pH.
The neutral pH 7.0 environment of electrophoresis results in the maximum stability of proteins
and gel matrix, providing better band resolution than other gel systems, including the traditional
SDS-PAGE Laemmli system. For many years the Laemmli system has been the standard
method used to perform SDS-PAGE Laemmli-type gels. These gels are useful for a broad
range of protein separations [29, 30].
The NuPAGE Novex Midi Gels consists of NuPAGE Novex Bis-Tris [Bis (2-hydroxyethyl) imino-
tris (hydroxymethyl) methane-HCL] Midi Gels (for separating small to mid-size molecular weight
proteins), NuPAGE LDS (Lithium Dodecyl Sulphate) Sample Buffer, NuPAGE Sample Reducing
Agent and NuPAGE MOPS [3-(N-morpholino) propane sulfonic acid] SDS (Sodium
Dodecylsulfate) Running Buffer for NuPAGE Noves Bis-Tris Midi Gels.
The NuPAGE Bis-Tris discontinuous buffer system involves three ions [30]:
Chloride is supplied by the gel buffer and serves as a leading ion due to its high affinity to the
anode compared to other anions in the system. The gel buffer ions are Bis-Tris and Cl- (pH 6.4).
MOPS serves as the trailing ion. The running buffer ions are Tris, MOPS and dodecylsulfate (pH
7.3-7.7).
Bis-Tris is the common ion present in the gel buffer and running buffer. The combination of a
lower pH gel buffer (pH 6.4) and running buffer (pH 7.3-7.7) results in a significantly lower
operating pH of 7 during electrophoresis.
The hydrophobic tail of the dodecylsulfate in SDS interacts strongly with polypeptide chains and
it is also a detergent that disrupts protein folding. The NuPAGE LDS Sample Buffer is used to
prepare samples for denaturing gel electrophoresis with the NuPAGE Novex Midi Gels, and it is
formulated to reliably provide complete reduction of the disulfides under mild heating conditions
and eliminate any protein cleavage during sample preparation. The NuPAGE Sample Reducing
Agent contains Dithiothreitol (DTT) and prepares samples for reducing gel electrophoresis [29,
30].
20
2.4.2 Sample Preparation
1. Part of Sialolith 2 was broken down mechanically. These samples were kept in EDTA at 25% in
glutaraldehyde 0.1 M at 3% at room temperature for about 6 months for removal of calcified
matrix, changing the solution every two days until the solution became translucent and gel-like.
Then they were kept at 4 oC.
2. Aliquots of 1 ml containing fragments of the sample were removed and centrifuged at 14000 Xg
for about 15 minutes.
3. The sediments were re-suspended in 50 l of supernatant and diluted 1:2 in NuPAGE sample
buffer (2X) with LDS and a reducing agent. Then they were denatured by heating at 70 °C for
10 minutes, and afterwards they were subjected to electrophoresis.
2.4.3 Electrophoretic analysis
1. The samples were analysed in NuPAGE Novex Bis-Tris gradient gels (4-12%) with NuPAGE
MOPS SDS running buffer, to which a reducing agent was added Gels with 20 wells were
loaded with samples of 25 μl.
2. The electrophoiresis was then processed in a XCell4 SureLock midi cell system at 200 volts for
55 minutes. A sample of a molecular marker (Novex Sharp protein standard) was run in the
same gel.
3. At the end of the procedure, the cassette containing the gel was dismantled and the gel placed
in a box of appropriate size, to be equilibrated in ultrapure water for 30 minutes with shaking.
The gel was then stained with Bio-Safe Coomassie Stain for 1 hour. The dye was replaced by
ultrapure water, which kept the gel preserved until further use.
2.4.4 Processing of the polypeptide bands for sequencing
1. The polypeptide bands visualized on the gel were excised in order to be sent for identification.
The excision was achieved by placing the gel on a transilluminator, the surface of which was
cleaned with 70 % ethanol. To avoid contamination it was necessary to use a disposable sterile
scalpel for each band. It was also necessary to work carefully with gloves, and to have hair
covered with a cap, in order to prevent contamination of skin and hair.
2. The excised protein bands were placed in sterile microtubes and were sent to the Mass
Spectrometry Laboratory, Analytical Services Unit, Instituto de Tecnologia Química e Biológica,
Universidade Nova de Lisboa, in order to be identified. There they were subjected to proteolytic
digestion with trypsin. The peptides resulting from this digestion were purified, concentrated and
21
eluted directly onto a MALDI plate. The mass spectra of the different peptides were obtained
using a MALDI-TOF/TOF MS instrument. MS and MS/MS spectra obtained were analyzed using
a combined of the Mascot search engine and database of NCBI.
22
3. Results and Discussion
3.2. Scanning Electron Microscopy
The use of SEM, in the BSE imaging mode, resulted in the formation of images of the polished
cross-section (Sialolith 1a). X-ray maps of the images show the elements present in the bright
and dark regions.
Figure 10 displays the polished cross-section (Sialolith 1a) with some highlighted details, which
are discussed in this section. The sialolith exibits one central core, which has an irregular shape
and is surrounded by layers.
Figure 10 - Polished cross-section of Sialolith 1a (BSE mode).
X-ray maps (Figure 12) were made of detail A (Figure 11). These prove that, in the images
produced in the BSE imaging mode, the bright regions are Ca and P-rich and that darker
regions contained S. Typical EDS spectra obtained by Olga Santo [2] from both dark and bright
regions are shown in Figure 13. The dark region spectrum contains peaks of Ca and P, an
23
intense S peak and minor Cl and K peaks. The bright region spectrum also has only intense Ca
and P peaks.
Figure 11 – Detail A of Sialolith 1a polished cross-section, from where X- ray maps shown in Figure 12
were obtained (BSE mode).
24
Figure 12 – (a), (b) and (c) are the Ca, P and S X-ray maps respectively of detail A of the Sialolith 1a
polished cross-section (BSE mode).
Figure 13 – Typical EDS spectra obtained at (a) bright and (b) dark regions [2].
25
Detail B shows part of the core (highly mineralized) with organic teardrop globules (organic
matter) and spherical globules (partially mineralized) (Figure 14). Figure 15 displays layers of
mineralized and weakly mineralized globule convolutions.
Figure 14 – Detail B of the Sialolith 1a polished cross-section display the highly mineralized core.
The arrows indicate the teardrop globules.
Figure 15 – Detail B of the Sialolith 1a polished cross-section. Parts (a) and (b) are convolutions
of mineralized and weakly mineralized globule layers.
26
Detail C displays the core surrounded by incomplete layers of organic matter with partially
spherical mineralized globules (Figure 16). Detail D shows partially mineralized globules
surrounded by crescents of smaller teardrop-shaped structures, which are found in the layers
around the core (Figure 17).
Figure 16 – Detail C of the Sialolith 1a polished cross-section displays the core surrounded by an
incomplete layer of organic matter with spherical globules (indicated by the arrow) (BSE mode).
Figure 17 – Detail D of the Sialolith 1a polished cross-section shows (indicated by the arrow) the globules
surrounded by crescents of smaller teardrop-shaped structures (BSE mode).
27
In detail E, close to the peripheral of the sialolith it is possible to observe laminar layers (Figure
18a) and incomplete layers of spherical globules (highly mineralized) intercalated with
incomplete layers of teardrop-shaped globules (weakly mineralized) (Figure 18b).
Figure 18 – Detail E of the Sialolith 1a polished cross-section. (a) Laminar layers (b) layers of teardrop-
shaped (indicated by the uppermost arrow) and spherical globules (indicated by the lowermost arrow)
(BSE mode).
28
From the results it can be inferred that the bright regions are highly mineralized (Ca and P-rich).
Darker regions, which are essentially composed of organic matter, contain S. In summary, the
salivary calculus exhibits a highly mineralized irregular core, in which organic and partially
organic globules are sparsely distributed, or exist in the form of layer convolutions. Around the
core, incomplete layers of organic matter in which partially mineralized spherical globules are
present. These layers are loosely connected to the subsequent layers. The subsequent layers
present either laminar or globular structures [2]. The laminar layers consist of fine mineralized
strata (1-10 μm) intercalated with fine organic strata (1-5 μm) [2, 9]. However, the sialolith also
contains layers of mineralized globules intercalated with layers of organic globules. These layer
types alternate thereafter in succession following a chronologic sequence. Sialolith 1a has a
core surrounded by a globular layer which is in turn surrounded by a laminar layer. Some
sialoliths previously characterized have a core surrounded by a laminar layer, which is
surrounded by an outer globular layer [2, 6, 9].
The teardrop globules do not always have the same morphology in these images, but the 3-D
globule morphologies are close to teardrop shapes, with a circular configuration in transverse
cross-sections and an elongated configuration in longitudinal cross-sections.
In the layers around the core, partially mineralized globules are surrounded by crescents of
smaller teardrop-shaped structures which may be result from a “squeezing process” induced by
pressure build-up [2].
29
3.2 Transmission Electron Microscopy
TEM was used to observe the components present in the sialolith (Sialolith 1b). Figure 19 was
obtained from the decalcified sample. Figure 19a shows only the organic matter present in the
sialolith and Figure 19b displays the presence of globules with a few crystals inside, which were
supposed to be hydroxyapatite (HA), inside that were not possible to remove with the
decalcification operation.
Figure 19 – Decalcified sample. (a) Organic matter without HA crystals, the arrows indicate the
spaces in organic matter where HA crystals were present before decalcification (b) globules (Y) with a few
HA crystals inside, the arrows indicate HA crystals.
30
In the calcified sample it is possible to observe organic globules surrounded by many HA
crystals (Figure 20a) and HA crystals inside the globules (Figure 20b).
Figure 20– Calcified sample. (a) Globules (Y) surrounded by HA crystals in organic matter and (b)
HA crystals inside a globule, the arrows indicate the HA crystals.
Dark field and Bright field imaging were used to roughly determine the HA crystals size, which
ranged from a few tens to hundreds of nanometres (Figure 21). The presence of HA was
confirmed by indexation of the ring diffraction pattern (Figure 22).
31
Figure 21 – Calcified sample. (a) Dark field and (b) Bright field images of HA crystals obtained in the same
region of the sample.
Figure 22 – The ring pattern diffraction of HA crystals obtained from a region similar to the one presented
in Figure 21.
32
Evidence for bacteria existence was found in the sialolith observed (Figure 23). These bacteria
have thick cell walls and rod-shapes which suggest that they are Gram-positive. Dark field and
Bright field images were also used to observe the HA crystals inside and surrounding the
bacteria (Figure 24).
Figure 23 – Calcified sample with bacteria. (a) and (b) bacteria (B). The arrow point to the occurrence of
cell division of bacteria.
33
Figure 24 – Calcified sample. (a) Dark field and (b) Bright field images of HA crystals inside and
surrounding the bacteria.
These bacteria have thick cell walls and are rod-shaped. The bacteria size is 1 μm
transversely and 1 – 1.5 μm longitudinally. It was possible to detect bacteria in the process of
division. Bacteria grow to a fixed size and then reproduce through binary fission, a form of
asexual reproduction [31]. Bacteria can grow and divide extremely rapidly [32]. The high sulfur
content in the organic matter (previously discussed in section 3.1), might be produced by
bacteria [33]. To the author best knowledge evidence for a possible S origin had not been
previously reported.
Organic matter has an irregular distribution and may be composed of denatured collagen [2].
HA crystals are present inside some of the organic globules structures. A hypothesis for this is
that the mineralization starts with the diffusion of the Ca and P into globule interior, followed by
34
the precipitation of mineralized flakes, which grow or agglomerate inside the globules and
transform during the process into HA [2]. In summary, the results obtained allow to conclude
that the main component of the mineral matter is hydroxyapatite and that bacteria may be an
important factor in sialolith formation.
35
3.3 Atomic Force Microscopy
This work aimed at establishing an experimental protocol for mechanical behaviour assessment
and some preliminary results are presented here. FS was used to obtain force-distance curves
at selected points in specific areas (detail A) of the Sialolith 1a polished cross-section (Figure
11). Through these force-distance curves it is possible to obtain the Young’s modulus and
adhesion force values.
The measurements were carried out in the bright and dark regions corresponding to A and B in
Figures 25 and 26, respectively. Figure 27 displays the topographic images of regions A and B.
Figure 28 shows a typical force-distance curve.
Figure 25 – Region A of detail A of the Sialolith 1a polished cross-section obtained in BSE mode by SEM.
Region A is located in a highly mineralized area.
36
Figure 26 – Region B of detail A of the Sialolith 1a polished cross-section obtained in BSE mode by SEM.
Region B is located in a weakly mineralized area.
Figure 27 – Topographic images of regions A (scan size: 13 μm x 13 μm) and B (scan size: 40 μm x 40
μm) from the polished cross-section. (x) and (y) are points selected in order to obtain force-distance curves
of each region.
37
Figure 28 – Example of the force-distance curve obtained at point x (Figure 21). Force-distance curve
describing a single approach-retract cycle of the AFM tip: (A) The AFM tip is approaching the sample
surface and the long range interactions are too small to give measurable deflections; (B) the initial contact
between the tip and the surface is mediated by attractive Van der Waals forces (contact) that lead to an
attraction of the tip toward the surface; (C) the tip applies a constant and default force upon the surface
that leads to sample indentation, cantilever deflection and increase of repulsive forces; (D) subsequently,
the tip tries to retract and to break loose from the surface; (E) various adhesive forces between the sample
and the AFM tip, however, hamper tip retraction. These adhesive forces can be taken directly from the
force-distance curve; (F) the tip looses contact with the surface upon overcoming of the adhesive forces [2]
is the slope of the force-distance curve. The adhesive force ( ) is the difference between
the snap-in point (the point of contact between the cantilever tip and specimen surface) and the
snap-out point (the point of separation or detachment between the cantilever tip and specimen
surface). Thirty force-distance curves were obtained for each region, (N/m) and mean
values were determined (Table 1).
Table 1 – Mean values of (N/m) and (N), and their standard deviations for the bright and dark
regions.
Bright region (region x)
Dark region (region y)
(N/m) -2.92 ± 7.00 10
-2 -2.90
± 1.32 10
-1
(N) 1.22 10-7
± 7.88 10-8
3.22 10-8
± 1.68 10-8
The mean values were used to calculate the Young’s modulus values presented in Table 2
using the equation (5) and by considering that the Young’s modulus of the silicon tip ( ) is 150
GPa, the nominal tip radius ( ) is 20 nm, and that the Poisson’s ratio of tip to sample (
38
respectively) was considered as 0.3 it was possible to calculate the Young’s modulus of each
region ( ).
Table 2 – Young’s modulus (MPa) for the bright and dark regions
Bright region (region x)
Dark region (region y)
Young’s modulus, (MPa)
66.4 66.0
The AFM results show that the average Young’s modulus of the sialoliths is close to 66 MPa.
Nevertheless, a higher number of curves would be required to differentiate the two regions. The
same applies to the adhesion force values, which exhibited a large dispersion. Indeed a higher
adhesion force would be expected for the dark regions due to higher content of adhesive
substances (may be the presence of collagen) [2]. However since the measurements have been
carried out first in the dark region, contamination of the tip with organic matter may have
influenced the adhesion results.
Figure 29 shows the Young’s moduli of different materials [34]. The sialolith Young’s modulus is
situated between those of gelatin and protein crystals (10 MPa and 120 MPa, respectively). This
means that the sialolith material is very soft and that the hydroxyapatite crystals are not having
a strong reinforcement effect.
Figure 29 - Young’s (elastic) moduli of different materials. The diagram shows a spectrum from very hard
to very soft: steel > bone > collagen > protein crystals > gelatin, rubber > cells [34].
As suggested by Olga Santo [2] some regions of the sialolith showed a fracture surface
characteristic of a ductile material containing hard inclusions. When a material contains
relatively hard inclusions which do not deform at the same rate as the matrix, voids are
nucleated to accommodate this incompatibility. The nucleation involves decohesion at the
inclusion-matrix interface and when a ductile fracture surface is examined, it shows dimples,
39
that correspond to the voids. Therefore the relatively high adhesion forces measured and the
ductile behaviour previously observed may justify the modest success rates obtained by shock-
wave lithotripsy with salivary calculi [8, 9, 35] when compared with renal calculi which contain
less organic material [36].
A protocol for assessment of the mechanical properties of sialolith components has been
established. The Young Modulus and adhesivity of the sialolith can be measured by FS yet care
must be taken so that mineralized regions are probed first in order to minimize the tip
contamination with organic matter.
40
3.4 Proteomic analysis
Several gels were made from the samples and then four bands were selected for analysis by
Mass Spectrometry. These four bands were visible after staining with Coomassie Blue (Bio-Safe
Coomassie Stain, Bio Rad). Two bands had molecular weights exceeding 260 kDa and the
other two bands had molecular weights between 20 and 30 kDa. The two bands weighing 260
kDa had been taken from one gel, whereas the other two bands had been taken from a second
gel. Table 3 and 4 shows the proteins obtained.
Table 3 – Proteins with high molecular weight except Humam Cathepsin G (low molecular weitht) and their
accession number.
Proteins Human Cathepsin G
Crystal Structure Of Human Alpha-
Defensin-4
Medullasin
Acession number
gi|3891975 gi|109156990 gi|233229
Table 4 – Proteins with high molecular weight and their accession number.
Proteins The Complex Of Human Leukocyte
Elastase
Keratin 1 Keratin 10
Acession number
gi|809343 gi|7331218 gi|186629
Cathepsin G and Human Leukocyte Elastase (HLE) (also called Medullasin or Neutrophil
elastase) are serine proteases. Serine protease is a protease (enzymes that cut peptide bonds
in proteins) in which one of the amino acids at the active site is serine. Neutrophilic
polymorphonuclear leukocytes contain specialized azurophil granules which contains the serine
proteases catheptin G and HLE [37, 38]. These proteolytic enzymes are secreted by neutrophils
during inflammation, they destroy bacteria and are capable of degrading a wide variety of
substrates including elastin, collagen and proteoglycans [37-39]. Due to the existence of these
collagen-degrading proteins in a sialolith it is possible that collagen is present in a sialolith in its
denatured form as suggested by the spectroscopy data [2].
Human Alpha-Defensin-4 (DEFA-4), also called as Neutrophil Defensin 4 (ND-4) or Human
Neutrophil Peptided-4 (HNP-4), is a human peptide that is encoded by the DEFA4 gene [40,
41]. HNP-4 belongs to the alpha-defensin family. Alpha-defensins are particularly abundant in
primary (azurophil) granules of neutrophils and its function is antimicrobial activity against
Gram-negative bacteria, and to a lesser extent, also against Gram-positive bacteria and fungi
[40].
41
Keratins are heteropolymeric structural proteins which form intermediate filaments. These
filaments, along with actin microfilaments and microtubules, compose the cytoskeleton of
epithelial cells. Keratin 1 is a member of the type II cytokeratins family. The type II cytokeratins
consist of basic or neutral proteins which are arranged in pairs of heterotypic keratin chains
coexpressed during the differentiation of simple and stratified epithelial tissues. Keratin 10 is a
member of the type I (acidic) cytokeratin family. Epithelial cells almost always coexpress pairs of
type I and type II keratins, and the pairs that are coexpressed are highly characteristic of a
given epithelial tissue [42, 43]. Keratin 1 and 10 were found in the sialolith, which means that
may be keratins are present in epithelial cells of the duct and they are incorporated in the
sialolith.
42
4. Globular Mechanism
The results obtained provide new information on the globular mechanism discussed by Olga
Santo [2]. Using Figure 2 (globular mechanism) it is possible to rewrite some of the stages.
In stage 3, damage occurs to the duct/gland walls, exposing the foreign body/nidus to
connective tissue. At this point it is possible to prove the presence of bacteria. Bacterial
infection develops in the damaged tissue causing an inflammation. As a result of inflammation,
neutrophils appear, destroying the bacteria.
In stages 4 and 5, a foreign body/nidus invades the connective tissue, followed by the scar
tissue. This tissue contains collagen and an extracellular matrix. The collagen may be
synthesized by fibroblast cells which are a constituent of the connective tissue. Collagen in
sialoliths has a denatured form for which cathepsin G and HLE are responsible.
Figure 2 – Schematic drawing representing the globular and sedimentary cycles in the sialoliths formation
mechanism [2].
43
5. Conclusion
The salivary calculi exhibit a mineralized core with organic and partially organic globules. The
core is surrounded by organic matter layers with partially mineralized globules. These layers are
loosely connected to the subsequent layers. These subsequent layers can be made of laminar
or globular structures. The laminar layers consist of fine mineralized strata intercalated with fine
organic strata, whereas the globular structure layers can be either organic or mineralized. The
inorganic matter consists essentially of hydroxyapatite. The organic matter which has an
irregular distribution may be contains proteins and denatured collagen.
The existence of bacteria in the sialolith has been inferred from the results. The high sulfur
content in the organic matter may have been produced by bacteria. The presence of bacteria
proves the existence of a bacterial infection causing inflammation. The existence of neutrophils,
which destroy bacteria and contain proteolytic enzymes that degrade collagen, was also proved.
Keratins were also found in the sialolith, which probably originated from the epithelial cells of the
duct. The results obtain reinforce the validity of the globular mechanism for the formation of
sialoliths previously proposed.
In relation to the mechanical properties assessment, an experimental protocol has been
established. No significant differences in Young’s modulus were detected between the organic
and calcified regions, but the sialolith was found to present a relatively high adhesion force and
a ductile behaviour.
Fundamental knowledge on sialolith formation and experimental sialolith characterization are
essential both for increasing the probability of a successful treatment of patients and to prevent
further future re-occurrences of sialolithiasis. This research has resulted in important information
on salivary calculi, yet further studies will be required in order to develop a deeper knowledge.
44
6. Future work
Existing studies on sialoliths leave much to be known. More proteomic analysis and mechanical
tests on sialoliths are necessary in order to uncover more information on proteins and to
compare mechanical properties which will allow a better understanding of the sialolith formation
mechanism and refinement of the existing proposed models. Our results demonstrate that a
combined Biology and Materials Science interdisciplinary approach is effective for a deeper
understanding of sialolith pathogenesis and etiology.
45
7. Bibliography
[1] Dorland, W. (2007). Dorland’s Medical Dictionary for Health Consumers, Saunders (an
imprint of Elsevier).
[2] Santo, O. L. (2008). Structural and chemical diversity of sialoliths, Master thesis, Instituto
Superior Técnico (IST), Departament of Materials Engineering.
[3] Kasaboğlu, O., Er N., Tümer, C., Akkocaoğlu, M. (2004). Micromorphology of Sialoliths in
Submandibular Salivary Gland: A Scanning Electron Microscope and X-ray Diffraction Analysis.
Journal of Oral Maxillofacial Surgery, 62: 1253-1258.
[4] Hong, K. W., Yang, Y. S. (2003). Sialolithiasis in the sublingual gland. Journal of
Laryngology & Otology, 117: 905-907.
[5] Teymoortash, A., Wollstein, A. C., Lippert, B. M., Peldszus, R., Werner, J. A. (2002).
Bacteria and Pathogenesis of Human Salivary Calculus. Acta Otolaryngol, 122: 210-214.
[6] Teymoortash, A., Buck, P., Jepsen, H., Werner, J. A. (2003). Sialolith crystals localized
intraglandularly and in the Wharton's duct of the human submandibular gland: an X-ray
diffraction analysis. Archives of Oral Biology, 48: 233-236.
[7] Boskey, A. L., Burstein, L. S., Mandel, D. (1983). Phospholipids associated with human
parotid gland sialoliths [short communication]. Archives of Oral Biology. Vol. 28, No. 7, pp. 655-
657.
[8] Ekberg, O., Isacsson, G. (1981). Chemical analysis of the inorganic component of human
salivary duct calculi. Archives of Oral Biology. Vol. 26, pp. 951-953.
[9] Alves de Matos, A. P., Carvalho, P. A., Almeida, A., Duarte, L., Vilar, R., Leitão, J. (2007).
On the structural diversity of sialoliths. Microscopy and Microanalysis, 13 (5): 390-396.
[10] Urban, W., Geisthoff, M. D. (2007). Tecniques for multimodal salivary gland stone therapy.
Operative techniques in Otolaryngology, 18: 332-340.
[11] Capaccio, P., Torretta, S., Pignataro, L. (2009, December). The role of adenectomy for
salivary gland obstructions in the area of sialendoscopy and lithotripsy. Otolaryngologic Clinics
of North America, 42(6): 1161-71.
[12] Nakagawa, Y., Ahmed, M., Hall, S. L., Deganello, S., Coe. F. L. (1987, June). Isolation from
human calcium oxalate stones of nephrocalcin, a glycoprotein inhibitor of calcium oxalate
crystal growth. Journal of Clinical Investigation. Vol. 79, 1782-1787.
[13] Escudier, M. P., Brown, J. E., Drage, N. A., McGurk, M. (2003). Extracorporeal shockwave
lithotripsy in the management of salivary calculi. British Journal of Surgery. 90: 482-485.
46
[14] Koch, M., Zenk, J., Iro, H. (2009, December). Algorithms for treatment of salivary gland
obstructions. Otolaryngologic Clinics of North America, 42(6): 1173-92.
[15] Goldstein, J., Newbury, D. E., Joy, D. C., Echlin, P., Lyman, C. E., Lifshin, E., et. al. (2003,
January). Scanning electron microscopy and x-ray microanalysis, Vol. 1 (3rd
ed.) Germany:
Springer.
[16] Williams, D. B., Carter, C. B. (1996). Transmission Electron Microscopy: A Textbook for
Materials Science, Vol.4 (1st Ed.) New York: Plenum Press.
[17] Bozzola, J. J., Russell, L. (1998). Electron microscopy: principles and techniques for
biologists (2nd
Ed.). London: Jones & Bartlett Learning.
[18] Central Microscopy. http://www.uiowa.edu/~cemrf/methodology/sem/index.htm, (last
accessed October 2010).
[19] Holleritha, C., Wernicke, D., Bühler, M., Feilitzsch,F., Huber, M., Hӧhne, J. et. al. (2004).
Energy dispersive X-ray spectroscopy with microcalorimeters. Nuclear Instruments and
Methods in Physics Research A 520: 606-609.
[20] Material Evaluation and Engineering, Inc. http://mee-inc.com/eds.html, (last accessed
October 2010).
[21] Reimer, L., Kohl, H. (2008). Transmission electron microscopy: physics of image formation
(5 Ed.) Germany: Springer.
[22] Goodhew, P. J., Humphreys, F. J., Beanland, R. (2001). Electron microscopy and analysis
(3rd
Ed.). New York: Taylor & Francis Group.
[23] Transmission Electron Microscope.
http://www.mete.metu.edu.tr/pages/tem/TEMtext/TEMtext.html, (last accessed October 2010).
[24] Washington University in St. Louis. http://presolar.wustl.edu/work/instrumentation.html, (last
accessed October 2010).
[25] Braga, P., Ricci, D. (2004). Atomic force microscopy: biomedical methods and applications.
Totowa, New Jersey: Humana Press Inc.
[26] Butt, H., Cappella, B., Kappl, M. (2005, August). Force measurements with the atomic force
microscope: Technique, interpretation and applications. Surface Science Reports, 59:1–152.
[27] Michigan Tech. http://nano.mtu.edu/afm.htm, (last accessed October 2010).
[28] Fawzy, A., Farghaly, A. M. (2009). Probing nano-scale adhesion force between AFM and
acid demineralized intertubular dentin: Moist versus dry dentin. Journal of Dentistry, 37: 963-
969.
[29] Invitrogen. www.invitrogen.com, (laste accessed August 2010).
[30] Manual of NuPAGE Novex Midi Gel System (2005). Invitrogen Corporation.
47
[31] Koch, A. (2002). Control of the bacterial cell cycle by cytoplasmic growth. Critical Reviews
in Microbiology, 28 (1): 61–77.
[32] Eagon, R. G. (April 1962). Pseudomonas natriegens, a marine bacterium with a generation
time of less than 10 minutes. Journal of Bacteriology, 83 (4): 736–7.
[33] Washio, J., Sato, T., Koseki, T., Takahashi, N. (2005). Hydrogen sulfide-producing bacteria
in tongue biofilm and their relationship with oral malodour. Journal of Medical Microbiology, 54:
889-895.
[34] Alonso, J. L., Goldmann, W. H. (2002). Feeling the forces: atomic dorce microscopy in cell
biology. Life Sciences, 72: 2553-2560.
[35] Eggers, G., Chilla, R. (2005). Ultrasound guided lithotripsy of salivary calculi using an
electromagnetic lithotripter. Journal of Oral Maxillofacial Surgery, 34: 890-894.
[36] Kaneko, K., Yamanobe, T., Nakagomi, K., Mawatari, K., Onoda, M., Fujimori, S. (2003).
Proteomic in renal research. Analytical Biochemistry, 324: 191-196.
[37] Sinha, S., Watorek, W., Karr, S., Giles, J., Bode, W., Travis, J. (1987). Primary structure of
human neutrophils elastase, Proceedings of the National Academy of Science. Vol. 84, pp.
2228-2232.
[38] UniProt, http://www.uniprot.org/uniprot/P08246, (last accessed September 2010).
[39] Bode, W., Wei, A., Huber, R., Meyer, E., Travis, J., Neumann, S. (1986). X-ray crystal
structure of the complex of human leukocyte elastase (PMN elastase) and the third domain of
the turkey ovomucoid inhibitor. The EMBO Journal. Vol. 5, no.10, pp. 2453-2458.
[40] UniProt, http://www.uniprot.org/uniprot/P12838, (last accessed September 2010).
[41] Wilde, C. G., Griffith, J., Marra, M., Snable, J., Scott, R. (1989). Purification and
Characterization of Human Neutrophil Peptide 4, a Novel Member of the Defensin Family.
Journal of Biological Chemistry. Vol. 264, No. 19, pp. 11200-11203.
[42] UniProt, http://www.ncbi.nlm.nih.gov/omim/148080, (last accessed September 2010).
[43] Kierszenbaum, A., L. (2004). Histologia e Biologia Celular: Uma Introdução à Patalogia. Rio
de Janeiro: Elsevier.