CG2010 CytogardCytotoxic Drug Safety Cabinets
www.thermoline.com.au
Cytotoxic drug safety cabinets are
defined in Australian Standard AS2252.5
as the primary barrier against exposure to
aerosolsthatareproducedinthepreparation,
manipulation and dispensing of cytotoxic
drugs. Many of these drugs are known to
be mutagens and are suspected of being
carcinogens and teratogens. Recent work
involving prions has proven that exposure
may result in effects that are insidious and
maynotmanifestthemselvesforsomeyears.
The requirements for protection involve the
following:
› Protection of cabinet users and other
staff from exposure to aerosols or
vapourswhichmay be generated in the
preparation,manipulation,anddispensing
ofcytotoxicdrugs;
› Protectionofdrugproducts,so that they
maybepreparedinanenvironmentwhich
is essentially free from particulate and
biologicalcontamination;
› Protectionofcabinetmaintenance
personnelfromtheresidueofdrug
particleswhichcancontaminatefilters,
fansandothermechanicalcomponents.
Applications
CG2010 CytogardTM drug safety
cabinets are designed and manufactured in
Australiainthreenominalwidths –a900mm,
1200mm and 1800mm – and fully comply
with all requirements of Australian Standard
AS2252.5.
Cytotoxic cabinets are necessary for
operator and product protection, in addition
tothesafeguardsprovidedbyaneffectiveair
barrierandClyde-ApacMicrosealTMHEPAfilter
technology,CG2010 series is also equipped
withacarbonfiltertoremoveharmfulvapours
thatmaybereleasedduringthecompounding
process. CG2010 series are essential for
the protection of personnel, product and the
environment.
Description
AESEnvironmentalmaintainsanISO
9001:2015qualitymanagementsystemto
ensureprocessandproductconformance.
www.thermoline.com.au
CG2010 CytogardTM drug safety
cabinets are designed and manufactured to
comply with all requirements of Australian
StandardAS2252.5.
Each cabinet is factory-certified by
a NATA Accredited laboratory to meet the
specified performance requirements. These
cabinetsmayalsobeusedwherethehandling
of other drugs and chemical requires both
containmentandasepticmanipulation.
Cytotoxic safety cabinets are part-
recirculating laminar air flow enclosureswith
highefficiencyparticulateair(HEPA)filtration
of exhaust air andanair barrier at thework
opening.
HEPA-filtered vertical laminar airflow
whichisrecirculatedintheworkzonecreates
an ultra-clean work environment for product
protection.Anairbarrierbetweentheoperator
and the work zone is maintained by a flow
of roomair intoa fullwidthgrille in thework
opening.
The barrier air mixes with the
recirculated laminar flow air in a sump
underneaththeworksurfaceandisexhausted
from the cabinet via a HEPA filter which is
locateddirectlyundertheworktray.
Australian Standards
All positive pressure zones and filter
seals are surrounded by negative pressure
zones,soastocontainpotentiallyhazardous
aerosols.Cabinetsareavailablewiththework
zone width of 90cm, 120cm or 180cm and
arefreestandingunitsthatincorporateafloor
stand.
Standard cabinets have exhaust
dischargeontheright-handsidewithoptional
left-hand side or top exhaust available. Top
exhaust is typically specified where cabinet
exhaust air is to be entrained into the room
exhaustinaccordancewithAS2252.5.
These cabinets provide advanced
systemmonitoring technologyandanumber
ofuniquedesignfeaturesintendedtoenhance
safetyandeaseofuse.
www.thermoline.com.au
Construction
CabinetConstructed in electro-galvanised steel
withjointsweldedusingagasshieldedarcprocess.
Thismethodproducesarobust,leakfreehousing
that is able to withstand the rigours of transport
and handling. The exterior is finished in a high-
qualitypowdercoatwhichhasbeendevelopedfor
laboratoryequipment.
Work ZoneConstructed in grade 304 stainless steel
with2Bfinish.Cornersareradiusedandcrevice-
free for ease of cleaning and all surfaces are
carefully dressed to remove sharp edges. The
removableworktrayisdesignedtoallowcleaning
access to its undersidewithout removing it from
thecabinet.
A pneumatic mechanism opens and
closes the viewing window without the need for
externalfastenersorcatches.Thewindowisself-
supporting in the fully open position to facilitate
cleaningandaccessforlargeitems.Openingthe
window with the cabinet running automatically
engagesaboostmodeforenhancedcontainment
byactivatingamaximumexhaustairflowandthe
alarmsystems.
Hepa FiltersClydeApacMicrosealTMHEPAfilters,are
manufactured to meet requirements of AS4260.
Eachfilterisindividuallycertifiedtobeleakfreein
accordancewithAS1807.6.
Amanufacturerendorsed test label fittedwithan
extractofthetestreportisaffixedtoeachfilter.
Aprefilterextendsthelifeof theexhaust
HEPA filter and protects it from mechanical
damageduringcleaningoftheworkzone.
FansSeparatedirectdrivefansareprovidedfor
theexhaustand laminar flowHEPA filters.Fans
arefittedwithspeedcontrollerstoenableairflows
tobemaintainedthroughfilterlife.
Thefancontrolcircuitsareinterlockedso
thatthelaminarflowsystemwillnotoperateuntil
theexhaustsystemhasachievedacontainment
condition.
Audible and visible alarms with
rechargeablebatteryback-upsignifyanyreduction
inbarriercontainmentorlaminarairflow.
ElectricalCabinets operate on single-phase 240V,
50Hzpowerviaa10Aoutlet.Theelectricalsystem
complieswithAustralianStandardAS3100.
Alowvoltagetouchcontrolpanelislocated
on the frontof thecabinet.TheOptima2000TM
microprocessor-controlled control and diagnostic
systemprovidescontinuousmonitoringofcritical
cabinet functionswith a digital display indicating
thenatureofanymalfunction.
www.thermoline.com.au
Features
Standard
Features
› Optima2000TMprorammablecontrolanddiagnosticsystemwithdigitaldisplay
› Lowvoltagetouchcontrols
› Alarmsandboostmodeautomaticallyengagedwhenviewingwindowisopen
› Boostmodeselectableatcontrolpanel
› Selectablepost-useover-runtimer
› Hourmetertorecordoperatingtime
› Provisionforinterfacewithbuildingenergymanagementsystems
› Pneumaticallyassistedviewingwindow
› Magnahelicgaugetomonitorfansystems
› Fully-sealedworkopeningcoverfortestingprocedures
› Comprehensiveoperationandmaintenancemanual
Options
› Activatedcharcoalexhaustfilter
› Workareapoweroutlet
› Ultra-violetgermicidallamp
› Gastap(solenoid-interlocked)
› Servicetaps(vac,air,CO2,etc.)
› Hangingrailinworkarea
Exhaustfan
LaminarflowHEPAfilter
Exhaustfanplenum
Prefilter
ExhaustHEPAfilter
Kneespace
Airbarrier
Positivepressureplenum
Negativepressureplenum
Laminarflowfan
www.thermoline.com.au
Personnel Protection
Cytotoxic drugs and prions are
hazardoustohumanhealth. It isnecessary
toprotectbothusersandservicepersonnel.
Forthoseapplicationswherepersonneland
environmentalprotectionisrequired,Clyde-
Apac Class I or Class II biological safety
cabinets, or cytotoxic drug safety cabinets
(asapplicable)shouldbeconsidered.
ModelOverall
Dimensions (mm)Work Zone
Dimensions (mm)Weight
(kg)
W D H W D H
CGA90SIDEEXHAUST 1135 770 2310 880 560 610 326
CGA90TOPEXHAUST 1035 770 2410 880 560 610 326
CGA120SIDEEXHAUST
1440 770 2310 1180 560 610 372
CGA120TOPEXHAUST 1340 770 2410 1180 560 610 372
CGA180SIDEEXHAUST
2050 770 2310 1790 560 610 487
CGA180TOPEXHAUST 1950 770 2410 1790 560 610 487
ModelRHS
exhaustLHS
exhaustTop
exhaust
CGA90 2030021 2030022 2030023
CGA120 2030201 2030202 2030203
CGA180 2031201 2031202 2031203
WARNING
www.thermoline.com.au
Operation
ControlPanel
1. Fan/post-useover-runswitch2. Poweroutletswitch3. Gasresetswitch*4. Fluorescentlightswitch5. UVlampswitch*6. Boostmodeswitch7. Displaypanel8. Mainspowerindicator9. Alarmindicator
*optionalfunction
1 2 3 8 9
4 5 6 7
High-efficiency filters and fans
deliver quiet operation and safety.
Negative pressure zones surround all
positive pressure areas, eliminating the
possibility of contaminated air bypassing
thefilterorescapingfromthecabinet.
Inoperation,verticallaminarairflow
gently passes from the Laminar HEPA
filter to the sumpHEPA filter to createa
biologicallycleanworkarea.
In Cytotoxic cabinets:
An air barrier across the work
access opening, into the sump, reduces
potential risks topersonnel fromairborne
contaminantsintheworkzone.
In Cytotoxic models, the airflows
combine in the sump area beneath the
work floor and pass through an extra
HEPAfilterbeforerecirculationviaareturn
airplenum,tothetophousing.
Separate fan/filter arrangements
allowindependentadjustmenttomaintain
aneffectiveairbarrier.
A microprocessor is used to
control the speed of the blower motors.
Thismicroprocessor also allows fingertip
controloffunctionsandstatusincluding:
› Cabinetperformanceandstatusclearly
displayedinplainEnglish.
› Boostmode.
› Built-instopwatch.
www.thermoline.com.au
© 2019 AES Environmental | Bulletin No. 14N.11.17 AES114 REV3In keeping with our policy of continuing product improvement, we reserve the right to alter specifications without notice.
THE COMPANY AESEnvironmentalisanAustralianownedmanufacturingbusinessproducingproductsunderClyde-Apac,EmailAirHandlingandVokesbrandnamesforindustriesthatareasvariedasindustrialplants,commercialbuildings,powerstations,foodprocessing,healthcare,scienceandelectronics.AESEnvironmentalconsiderstheAustralianStandardsasacorecomponentofitsproductmixandhasdevelopedanexportmarketin25countries,promotingAustralianStandards,engineeringandmanufacturingsolutions.AESEnvironmental,atrustedmanufacturercapableofdeliveringreliableproductsolutionstohighly-criticalapplications,wherethecontrolofhazardousairbornecontaminationisoftencriticaltoprocessandpersonnel.
› HWSSeriesTMhorizontallaminarflowcabinets.
› VWSSeriesTMverticallaminarflowcabinets.
› BSC2000TMClassIbiologicalsafetycabinets.
› BH2000TMClassIIbiologicalsafetycabinets.
› PCRlaminarflowcabinets.
› Recirculatingfumecabinets.
› TFPTMSeriesHEPAfiltercleanroommodules.
› ExhaustCaptureHoodforCytotoxicSuite.
› Passthroughhatches.
Other Products On-Site Testing
Cytotoxicdrugsafetycabinetsare
factorytestedandcertifiedbyaNATA-
Accreditedlaboratory.Additionaltestingand
certificationisrecommendedasfollows:
› Onsitepriortouse
› Aftermaintenance
› Afterfilterreplacement
› Afterre-location
› Atleastannually
› Inspecialcircumstances,e.g.iffaulty
operationissuspected.
www.thermoline.com.au