Maintenance Manual - Revision 1.0 - English version BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Upload: others

Post on 09-Jan-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Maintenance Manual - Revision 1.0 - English version

BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Page 2: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

602922

This document has been prepared with the utmost care. However, AFI Centrifuge refuses to accept any responsibility in the event of errors or omissions. The same applies to any damage arising from the use of information contained in this manual. AFI Centrifuge 3, Rue Nicolas Copernic ZA Nord Bazouges 53200 CHATEAU-GONTIER FRANCE +33 (0) 2 43 06 66 76

Page 3: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 2/ 54

You have just purchased one of the finest Laboratory Centrifuges available. With proper care, maintenance and laboratory procedure, this centrifuge will provide you years of productive service. Please read this manual carefully to familiarize you with proper installation, and operation of the centrifuge. Other information are available on our

website, www.aficentrifuge.com .

Explanation of Symbols

Safety alert symbol indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury

CAUTION used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.

Safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.

Pinching: This warning symbol indicates the presence of a risk of pinching when handling the lid.

Note: Used for important information

Potential electrical hazard, only qualified person to access

Biological hazard,

Protective earth for user safety

Lead Free

CAUTION

CAUTION

WARNING

Page 4: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 3/ 54

Updates

Date Revision Modification Paragraph Author

10/03/2020 1.0 Creation All F. Boucard

Page 5: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 4/ 54

Table of contents

1 DESCRIPTION OF THE CENTRIFUGE .......................................................................................................... 3

1.1 TECHNICAL SPECIFICATIONS: ................................................................................................................................... 3 1.2 EXTERNAL PARTS DESCRIPTION: ............................................................................................................................... 5 1.3 EXPLODED VIEWS ................................................................................................................................................. 6 1.4 ACCESSORIES ....................................................................................................................................................... 7 1.5 INTERNAL PARTS DESCRIPTION ................................................................................................................................ 8 1.6 SERVICE TOOLS .................................................................................................................................................. 12 1.7 WIRING DIAGRAM .............................................................................................................................................. 13 1.8 REFRIGERATION CIRCUIT ........................................................................................................................................ 2 1.9 ELECTRONIC CONTROL CENTER / BLOCK ................................................................................................................. 20 1.10 TRANSFORMER .................................................................................................................................................. 21 1.11 LOCKING / UNLOCKING SYSTEM ............................................................................................................................. 22 1.12 MOTOR ........................................................................................................................................................... 23 1.13 SPEED SENSOR ................................................................................................................................................... 23 1.14 TACHOMETER BOARD .......................................................................................................................................... 23 1.15 TEMPERATURE PROBE ......................................................................................................................................... 23 1.16 IMBALANCE SENSOR............................................................................................................................................ 23 1.17 GAS SPRING ...................................................................................................................................................... 23 1.18 VIBRATIONS ABSORPTION / SUSPENSION ................................................................................................................. 24 1.19 REFRIGERATION SYSTEM: ..................................................................................................................................... 24

2 PROPER INSTALLATION ...........................................................................................................................25

2.1 CENTRIFUGE LOCATION ....................................................................................................................................... 25 2.2 LEVELING .......................................................................................................................................................... 25 2.3 ELECTRICAL INSTALLATION ................................................................................................................................... 25

3 TROUBLESHOOTING................................................................................................................................26

3.1 TROUBLESHOOTING ............................................................................................................................................ 26 3.2 ERROR MESSAGES .............................................................................................................................................. 26

4 CLEANING AND LUBRICATION .................................................................................................................29

4.1 CLEANING ......................................................................................................................................................... 29 4.2 LUBRICATION .................................................................................................................................................... 29

5 SETTINGS ................................................................................................................................................30

5.1 ACCESS TO THE SERVICE MENU ............................................................................................................................. 30 5.2 SETTING THE TEMPERATURE PROBE ........................................................................................................................ 32 5.3 UNBALANCING SENSITIVITY DETECTION .................................................................................................................. 32 5.4 SETTING THE LOCK .............................................................................................................................................. 34 5.5 AUTO-DIAGNOSTIC ............................................................................................................................................ 36 5.6 INFORMATION UPDATE ........................................................................................................................................ 37 5.7 COUNTERS MODIFICATION ................................................................................................................................... 37 5.8 OVERTEMPERATURE ALARM MENU ........................................................................................................................ 38 5.9 FIRMWARE UPDATE ............................................................................................................................................ 38 5.10 FACTORY RESET ................................................................................................................................................. 39 5.11 CLEARING COUNTERS .......................................................................................................................................... 40

6 STANDARD EXCHANGE OF COMPONENTS..............................................................................................41

6.1 SAFETY RULES ................................................................................................................................................... 41 6.2 SPARE PARTS & TOOLS ........................................................................................................................................ 41 6.3 SETTINGS FORM ................................................................................................................................................. 41

7 FUNCTIONAL CHECKS ..............................................................................................................................44

7.1 MECHANICAL SAFETY .......................................................................................................................................... 44 7.2 ELECTRICAL INSTALLATION – PROTECTIVE EARTH ...................................................................................................... 44

Page 6: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 5/ 54

7.3 SPEED CHECK .................................................................................................................................................... 45 7.4 CHECKING THE TIMER ......................................................................................................................................... 45 7.5 CHECKING THE TEMPERATURE .............................................................................................................................. 45

8 ANNEXES ................................................................................................................................................46

8.1 RETURN AUTHORIZATION ..................................................................................................................................... 46

Page 7: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Remarks

KATRINA must be connected to an electrical installation that conforms to all applicable safety standards, national or international prescriptions. In order to ensure user protection against electric shock, KATRINA protective earth conductor must be properly connected to the building earth. Do not operate a centrifuge until all necessary safety measures have been taken. KATRINA centrifuge is designed in such a way that its use does not expose the operator and his environment to any danger. This protection can only be guaranteed if the centrifuge is used in accordance with all the instructions in this manual. The rotors and buckets must be removed from use in the event of a trace of mechanical defect or corrosion. These elements have a lifetime, engraved on their visible part: To maintain good safety conditions, it is imperative to replace them when the recommended duration is reached. KATRINA can only be serviced by certified personal, informed and trained by AFI Centrifuge.

Warranty

AFI Centrifuge warrants this device does not have any assembly or material defects when it leaves the factory. In the event of a defect appearing during proper use, AFI Centrifuge undertakes to repair the device for a period of two years from the date of delivery, for malfunctions concerning the following parts:

- Motor - Electronic control center - Tachometer board. - Mains filter board - Interconnection board - Lid lock motor supply board

All parts claimed defective shall be returned, properly identified to the company at its factory, charges prepaid. Factory installed equipment or accessories are warranted only to the extent guaranteed by the original manufacturer, and this warranty shall not apply to any portion of the equipment modified by the user. Claims under this warranty should be directed to AFI Centrifuge. Setting forth in detail the nature of the defect, the date of the initial installation and the serial and model number of the equipment.

This warranty shall not apply to any AFI Centrifuge product or part thereof which has been subject to misuse, abuse, accident, shipping damage, improper installation or service, or damage by fire, or flood. If the serial number of this product is altered, removed or defaced as to be illegible, the Warranty shall be null and void in its entirety.

The warranty is for the sole benefit of the original purchaser and is not assignable or transferable. Prior to returning any item, for any reason, contact AFI Centrifuge for a Return Authorization Number. This number must accompany all returns. Any product shipped to AFI Centrifuge without this number will be returned refused shipment or collect freight. Please carefully read user manual instruction before using the centrifuge.

Page 8: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 2/ 54

Forbbiden use

KATRINA is not designed for the following uses and environments: The following are considered prohibited: • Flammable materials, • Explosive substances, • Materials that can react causing a danger, • Toxic materials • Radioactive materials • Contaminated material not contained in an aerosol-tight container. The use of this type of container is understood to be in a biosecurity system in accordance with national and international indications, they must not constitute the only protection of workers and the environment during the handling of pathogenic microorganisms. (IEC / EN61010-2-020) See for example the "Laboratory biosafety manual" guide by the World Health Organization. • Rotors and rotating accessories having exceeded their maximum service life. • The rotors and rotating accessories not being maintained in accordance with the instructions in this manual, or showing signs of wear and / or corrosion. • Rotors and rotating accessories not validated by AFI Centrifuge • Explosive environments • Radioactive environments

Decontamination

Depending on its use, and for example if potentially pathogenic substances have been spilled inside the bowl, KATRINA can expose service personal to a biological danger. In this case, or in case of any doubt, cleaning and decontamination are necessary as a preventative measure before the centrifuge, rotors and other accessories are maintained, repaired or transferred. The manufacturers can provide a form allowing the users to document the realization of this treatment. (IEC / EN 61010-2-020) Before any intervention, in order to guarantee the operator's safety, please refer to the laboratory decontamination protocol, and carefully apply the decontamination prescriptions. During decontamination operations, the required personal protective equipment must be worn. Radioactive materials are prohibited and cannot be centrifuged by KATRINA.

Page 9: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 3/ 54

1 Description of the Centrifuge

1.1 Technical specifications:

Page 10: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 4/ 54

Page 11: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 5/ 54

1.2 External parts description:

view N° Designation Service kit Cat. N°

Fig. 1 ¾ front view

1 Lid

2 Viewport

3 Control panel

4 Electronic ogive 71124005

5 Adjustable feet (x4)

Fig. 2 ¾ rear view

1 Main switch / circuit braker

2 Condensor grid

3 Hole for Emergency opening

tool

Fig. 3 Control panel

1 Membrane

Top ogive & membrane

2 Display screen / touch pad

3 Start / Stop button

4 Open button

1

2

3

4

5

1 2

3

1 2

3

4

Page 12: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 6/ 54

1.3 Exploded views

View N° Designation Service kit Cat. N°

3/4 front view

3/4 rear view

1 Front panel

2 Right side lateral panel

3 Ogive 230V / 50 hz 71124005

4 Left side lateral panel

5 Lid

6 Main switch / circuit breaker

71154012

Front view Front view

7 Feet (x4)

Bowl

8 Bowl seal 71124001

9 Motor seal 71124002

1

2

4

5

6

3

9

8

7

Page 13: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 7/ 54

1.4 Accessories

Page 14: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 8/ 54

For adapters range, please see commercial documentation in annex

1.5 Internal parts description

view N° Designation Service kit Cat. N°

Inner Lid

1 Latch for left side

2 Latch for right side

3 Viewport

4 Temperature sensor 71124007

5 Plastic inner lid

Right side

1 Electronic control center 71124005

2 Braking resistor

3 Filter board 71124009

4 Solenoid valve coil

5 Motor contactor 71124011

6 “Tachometer” board 71124008

1

2

3

4

5

2

3 4 6

4

5

1

Page 15: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 9/ 54

Right side – low part details

Motor and low suspension

1 Motor 71124004

2 Low suspension foam

Shock absorbers

3 Shock absorbers 71124003

High suspension

4 High suspension foam

Front side

1 Lid lock motor

2 Lid lock system 71124006

3

Compressor

5

6

3 4

1

2

3

4

1 2

3

Page 16: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 10/ 54

Left side

1 Acces to condenser, behind grid

Left side

1 Drier filter

2 Liquid receiver tank

3 Condenser

Lid locking system

Overview

Lid locking system details

1 Open position sensor

2 Motoreducer

1

3

1

3

2

3

3

3

1

3 2

3

Page 17: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 11/ 54

1 Locked position sensor 1

2 Latch presence sensor

3 Lid approach sensor

1 Locking motor supply board 71124010

1 Gas spring 71124000

1

3

2

3

3

3

1

3

1

3

Page 18: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 12/ 54

Bottom view

Speed sensor 71124013

1.6 Service tools Here are the required tools for KATRINA service opertions :

Tool Picture Tool Picture

3mm 4mm and 6mm hexagonal wrench set with spherical head

Condenser bruch

Insulated PZ2 screw dirver

Condenser fins rectifier

Circlip plier

Flat wrench

Insulated flat screw driver

Hammer

Putty gun

Needle tool

Felt pen

Page 19: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 13/ 54

Slotted flat head screwdriver

Ratchet

AngledL-shaped wrench

Articulated adapter for ratchet

Bent nose plier

Cutting pliers

antistatic brush or Air duster spray

Manual jack and chock

Measuring devices

Optical tachometer (*) Min reading <= 200 rpm

Max reading >= 5000 rpm Precision : 0.05% + 1 digit

K thermocouple thermometer Min reading : 0°C Max reading : 50°C Required precision : Classe 1

Chronometre (*) Min reading : 1 min Max reading : 1 heure Required precision : 1 second

Multimètre (*)

Specific gear

Programming cable

Mini USB cable

Delivered with KATRINA

6mm hexagonal wrench

8mm hexagonal wrench

(*)For accurate measurement, please use only properly calibrated measuring tools.

1.7 Wiring diagram See next page

Page 20: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml
Page 21: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 2/ 54

1.8 Refrigeration circuit

Page 22: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 20/ 54

1.9 Electronic Control Center / Block The electrical block, or Control Center (sometimes named “Ogive”), includes a touch screen, a Start/Stop button, a lid opening button, as well as the microcontroller electronic board, and the electrical power supply electronic board. There is no provision for changing the elements composing the Control Center. It is a complete separate part. In case of defective element, a standard exchange of the complete Control Center must be considered, which facilitates the restoration of the centrifuge.

Touch Screen

This display is based on a TFT type display its size is 5.7”. It offers 640x480 pixels, and 16 Million colors. It is also equipped with LED backlighting for bright background with low power consumption. Just in front of the display, a resistive touch pad sense for user finger press. It provides access to many features when using the interface. The display and display card unit are located in the upper part of the Control Center.

Start/Stop Button

The control button placed under the screen allows for controlling the startup and interruption of a centrifugation cycle.

Lid Opening Button

The lid opening is controlled by the push button on the right side of the Control Center.

Electronic microcontroller Board

The electronic microcontroller board is located under the display card. The data for controlling the various parts are stored in this board, in the form of software, also marked “Microcontroller fw” The connection for the update of this software is below the Control Center, via the connection kit part no. 71199004. The procedure is described in the “Software programming annex” annex. This board communicates with: • The tachometer board • Llid position sensors for detecting the lid position, • Touchpad display module • The programming connector • The temperature probe • The Start / Stop control button on the membrane • The lid opening button on the right side of the Control Center.

Electronic Power Board

This board is directly connected to the mains and provides supply for: • The microcontroller board • The display card • The centrifuge motor • The refrigeration group • The lock motor It also transmits the power accumulated during rotor braking to the load shedding resistance.

Lid locking motor supply board

This board supplies 24Vdc to the lid lock moto reducer when opening or closing is required. It receives 230V at mains input, this voltage is provided by the Electronic Control Center only when required.

Page 23: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 21/ 54

The board and especially the molded transformer are protected by a 2ATT 250V 5x20 fuse cartridge.

1.10 Transformer A toroid transformer is connected to the power board and provides 3 different voltages identified by the following colors:

Tension 26V 16V 10V 18.5 V

Colour Green Yellow Red Black

Page 24: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 22/ 54

1.11 Locking / unlocking system

Description :

1) Lid latches. The left one is equipped with 2 magnets, the right one is not.

2) Latch approach sensor (Magnetic switch). 3) Closed position sensor (Magnetic switch). It detects the

latch postion. 4) Crankshaft: This is the moving part driven by the motor

reducer. 5) Locked position sensor (microswitch contact). It detects

the deadbolt when the device is locked. 6) Motor reducerHook. They are part of the "welded shaft"

& deadbolt. This welded shaft is moved through the crankshaft, which moves in the oblong guide zone of the deadbolt.

7) Open position sensor. (microswitch contact). It detects when the device is unlocked.

8) Alignment marks of deadbolt and crankshaft. When aligned, the device is in locked position.

A safety system prevents the motoreducer from being energized if a speed signal is detected. This prevents the lid from

being open is rotor is rotating.

1

3

8

6

4

5

7

2

9

Page 25: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 23/ 54

1.12 Motor Three phase induction motor, variable frequency U/F MLI driver. An integrated thermoswitch automatically stops the rotation cycle when windings temperature reaches 130°C, in this case “ERROR 04” is displayed.

1.13 Speed sensor KATRINA Speed sensor is a Hall Effect magnetic sensor, located under the motor. Six (6) magnets with alternating poles are arranged around a rotating disc. The sensor measures the frequency of passage of the magnets, and microcontroller determines the speed.

1.14 Tachometer board Speed sensor signal is filtered, shaped, and sent back to the Electronic Control Center, as well as the “rotating” signal. This “rotating” signal is activated when a speed signal is detected and used for the lid safety feature as described in §1.11 1) Connection to the tachometer sensor

2) Connection to the Control Center

1.15 Temperature probe Located on the lid internal panel, making it closed to the buckets once the lid is closed. Measurements are only taken during rotation. Sensor is an encapsulated PT500 class A element. During a centrifugation cycle, KATRINA displays the sample temperature, computed from the air temperature measurement, and assuming the samples enter the centrifuge at the program target temperature. Display accuracy is warranted only if samples enter the centrifuge at the target temperature. KATRINA is designed to maintain the desired temperature during a rotation cycle, this also can be achieved only if samples enters the centrifuge at the correct temperature.

1.16 Imbalance sensor In the case of an incorrect balancing of the buckets, oscillations and vibrations can reach an unacceptable level. Vibrations are permanently measured by an electronic accelerometer located inside the Electronic control center. If the measured level becomes higher than the acceptable threshold, centrifugation cycle is automatically interrupted and ERROR n°2 is displayed. The default detection threshold is: 50 gr This threshold is adjustable according to the media on which the centrifuge rests, in order for the imbalance sensor to always be triggered with a minimum imbalance of 50 gr see Section 3.3 for the procedure to adjust the imbalance detection threshold.

1.17 Gas spring A double gas cyclinder / spring system allows the opening of the lid when it is released by the lock. The spring allows a better relaxation at the opening start. The cylinder should open the lid without causing rebound at the end of travel.

1 2

3

Page 26: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 24/ 54

1.18 Vibrations absorption / suspension KATRINA is designed with an optimized suspension system allowing vibrations and parasitic movement’s absorption in normal operational conditions, provided the rotor load is well balanced around the rotation axis. This suspension is made of 5 shock absorbers at the base of the motor ballast and 2 foam disks located at the top and bottom of the motor. A good balancing of the accessory is a key point to achieve satisfying separation quality.

1.19 Refrigeration System: KATRINA refrigeration system allows to maintain the samples temperature at their initial temperature while rotating. Temperature target can be set between -10°C and 40°C. Maximum performance is achieved for a 20 +/-2°C ambient temperature:

4°C can be maintained +/-2°C at 4300 rpm with windshield lid,

4°C can be maintained +/-2°c at 3800 rpm without windshield lid. These results can be achieved only if samples enter the centrifuge at the target temperature. The centrifuge air inlet / outlet must be free of any obstacle.

Page 27: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 25/ 54

2 Proper installation

2.1 Centrifuge location

In order to install KATRINA, it is necessary to have a

free space of 1406 x 1560 mm.

The soil at this location must be able to withstand

twice the weight of the machine in use (about 350kg),

i.e. 700 kg,

and allow a vibration-less operation.These vibrations

could degrade the results of the centrifugation.

A free space of 30 cm all around the centrifuge must

absolutelybe provided, according to standard IEC

61010-2-020:

No potentially dangerous material, no person, nor any

object must be inside the free space during

centrifugation.

Leave a minimum space of 60 cm between 2 adjacent centrifuges. This space guarantees the required clearance area for

each machine and sufficient ventilation between them.

The reserved space must not be exposed to direct sunlight, nor placed near a heat source. The room will be well

ventilated to avoid overheating and electrical overconsumption.

2.2 Leveling Before being used, KATRINA must be properly leveled using its 4 adjustable feet. The centrifuge accessory must rotate in a perfectly horizontal plan, please see the leveling procedure available in the annex. A bad leveling can lead to:

drastically shorten lifetime of the motor and suspension system

poor centrifugation quality in reason of a bad behavior of the suspension system

2.3 Electrical installation

The centrifuge mains supply must be 230V / 50 Hz, single phase, in accordance with the manufacturer's plate on the

back of the device.

- 17Arms can be drawn continuously during the centrifuges operation

- 21Arms can be drawn for less than 1mn during centrifuges acceleration under heaviest load and with maximal

acceleration rate.

KATRINA must be directly connected to the distribution board and protected by its own protection device.

This device must be identified as the centrifuges' disconnector device and easily accessible in the immediate vicinity of

the centrifuge.

The protective earth conductor must imperatively be connected to the installation earth.

The centrifuge's power supply must be protected by a safety device ensuring the automatic cut-off in the event of an

insulation fault, such as a correctly sized differential circuit breaker.

The electrical installation must comply with all regulations in force at the place of installation.

1560

1406

Zones

d'accès

Page 28: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 26/ 54

3 Troubleshooting

3.1 Troubleshooting

Problem Cause Solution

No display: the screen remains black.

No voltage Operate the switch under the Control Center.

Connect the mains power cable.

Check the electrical supply / protection of the laboratory and restore it.

Centrifuge display remains frozen at turn On, AFI logo is displayed but logging menu is never displayed

SUB D9 cap improperly connected to the programming slot at the bottom of the Electronic Control Center

Turn Centrifuge OFF. Remove the translucent panel at the bottom of the Control Center in order to access and check the SUB D9 cap connection. Once properly connected, put the protective panel back before turning centrifuge ON.

The lid does not close: the mechanism does not start.

The lid has been opened with the manual opening key, Or centrifuge supply was turned off before the completion of locking / unlocking The lid motor supply fuse has blown

Press the open button, the locking motor starts in order to reset the system mechanical position. Once done, the lid can normally be closed. Turn Centrifuge OFF Check for any short cut between the board and the motor, check there is no mechanical blocking point preventing the locking or unlocking movement to complete. Once this points checked and eventually fixed, replace the fuse. (2A TT, 250V, 5x20mm)

Displayed speed value is decreasing, display is blinking.

After a power outage while centrifuging, the cycle is interrupted. The rotor slows down in free wheel.

Wait for the complete stop of the rotor and cycle the power

3.2 Error Messages

Error / Message No.

Cause Solution

01 LOCK FAIL The lid is open at startup Press Start / Stop: The message disappears. Close the lid. Wait for the complete closure of the lid before starting a cycle.

The locked position detector is faulty.

Check that the micro contact blade is present and not folded. Check that this micro contact is wired properly. Check the operation of this micro contact, and change it if necessary. Check the magnetic sensor is properly wired and operational

Page 29: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 27/ 54

02 IMBALANCE The rotor is non-symmetrically loaded.

Wait full stop and open the lid by pressing the button. Press Start/Stop: The message disappears. Balance the rotor loads Restart the spin cycle

The centrifuge is installed on a wrong type of support.

Place the centrifuge on a compliant media.

Id displayed despite a balancing error < 50 gr: the detection threshold is set incorrectly.

Carry out a calibration of the imbalance sensitivity threshold.

03 BOWL OVERTEMPERATURE

The temperature in the bowl has exceeded 43°C, the temperature of the room being too high.

Wait until the end of the automatic shutdown of the rotor. Activate the air conditioning system for an ambient temperature under the conditions described in the user’s manual.

The programmed temperature setpoint is inappropriate

Change speed / temperature pair of the program, which is not compatible with normal use.

The initial loading temperature is too high.

Place the samples before centrifugation in an environment between 4 and 37°C, for 1h.

The cooling unit no longer works.

Check the operation of the chiller with the menu “Auto-Diagnostic “ (Startup and performance)

The temperature probe is defective

In the probe calibration menu, the value displayed is different from the value measured by the probe, with a gap > 2°C.

The difference between the air temperature and the over temperature alarm set point is greater than the set value.

Check the value of the temperature alarm threshold. Change program settings: speed and/or temperature.

04 MOTOR OVERTEMPERATURE

The motor temperature is too high.

A stop in “freewheel” mode occurs. Wait 30 minutes (software reset) before opening the lid.

Activate the air conditioning system of the room to obtain a lower ambient temperature.

The motor is damaged Change the motor.

Motor temperature sensor connection malfunction

Change the motor.

05 ERROR MICROCONTACTOR ON LOCKING MECHANISM

One of the micro contacts / magnetic sensor is damaged or not working. There is a bad connection to the micro contacts.

Contact a certified technician.

06 MOTOR CONTACTOR INCONSISTENT STATE: NOT SEEN OPEN

The contactor does not work. The auxiliary contact is not working. The connection of the auxiliary micro contact is not good.

At centrifuge start, motor contactor is seen closed when it should be open. Contact a Certified technician.

07 MOTOR CONTACTOR INCONSISTENT STATE: NOT SEEN CLOSED

The motor contactor doesn’t work. The open position switch was released after the lid closed, The contactor coil is not energized. The coil of the micro contactor is not working properly.

At centrifuge start, motor contactor is seen open when it should be closed. Contact a certified technician.

08 RESERVED RESERVED RESERVED

09 NO SPEED SIGNAL AT START

The speed sensor is working improperly. The tachometer board is working improperly.

The speed cannot be read within the first 30s of the cycle. Contact a certified technician

Page 30: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 28/ 54

The connection to these parts is not good. At least one of the 3 motor power wires is not properly connected to the contactor. The power board is faulty, motor is not energized

10 SPEED SIGNAL LOST DURING ROTATION

The speed sensor is working improperly. The board with tachometer / accelerometer is working improperly. The connection to these parts is not good.

The speed signal was lost during more than 6s during rotation. Wait for centrifuge full stop, acknowledge the error, cycle power and try again. If Error 10 is still detected call a certified technician

11 RESERVED RESERVED RESERVED

12 OVERSPEED Major error, the centrifuge will stop.

The measured speed is higher than the maximum allowed speed allowed. Contact a Certified Technician.

13 Communication issue with accelerometer

The microcontroller board is working improperly.

Wait for full stop, acknowledge the error and cycle the power supply. Try again, if error 13 is still detected, call a certified technician

14 Communication issue with temperature sensor amplifier

The microcontroller board is working improperly.

Wait for full stop, acknowledge the error and cycle the power supply. Try again, if error 14 is still detected, call a certified technician

15 LID OPEN DURING ROTATION

Manual lid opening during centrifugation. Wrong calibration of locking mechanism.

The lid was manually opened or the hook was released during the centrifugation. Contact a Certified Technician.

NO DISPLAY No power. Verify the power cable connection. Check the power in the laboratory. Turn the switch on.

16 TEMPERATURE CONTROL ISSUE

Temperature probe out of service.

Verify that the temperature sensor is properly installed under the lid and connected Disconnect the probe from the micro controller board and measure the resistance. The value must be 500+ - 100 Ohm (PT500 sensor)

17 RESERVED RESERVED RESERVED

Page 31: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 29/ 54

4 Cleaning and lubrication

4.1 Cleaning After proper decontamination of the centrifuge, Open the lid, remove the buckets and place them on a stable support.

Buckets and liners

- For each bucket, clean the trunnions bearing area (rotation axis) with a dry cloth

- Clean the bottom of the bucket compartments

- Clean the bottom of the liner compartments and their structural grooves.

Bowl and bowl seal :

- Remove the rotor from the motor axis, using the provided wrench.

- Remove any liquid that could be present at the bottom of the centrifugation bowl, clean with water

and soap or an hydroalcoholic solution or a quaternary ammonium solution.

- NEVER USER ANY CHLORINATED PRODUCT (bleach, quaternary ammonium chloride…).

- Dry well after cleaning.

- Do not use any fluffy cloth.

- Reassemble the rotor on the motor shaft with the provided wrench.

4.2 Lubrication

Rotor

- Clean the 12 trunnions (bucket rotation axis)

- lay a thin layer of alimentary grease on the top of each trunnion

Buckets

- The trunnion bearing area does not require any lubrication, it is always lubricated by the trunnions

themselves.

Driving shaft:

- Remove the rotor from the motor axis, using the provided wrench.

- Clean any durt on the shaft and under the rotor.

- lay a thin layer of alimentary grease on the motor shaft

- Reassemble the rotor on the motor shaft with the provided wrench.

Page 32: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 30/ 54

5 Settings

Depending on its use, a centrifuge can be contaminated, Before any service operation, please refer to § DECONTAMINATION at the beginning of this manual.

5.1 Access to the Service Menu

Log in as “Maintenance”. Initial password is “0000” and shall be changed once centrifuge installed

Access the “Settings” menu using the following keys:

This menu is reserved for authorized technicians.

1st page of settings : Next page

Setting the temperature probe

Setting the imbalance sensitivity

Setting the time delay of the lock

Auto-Diagnostic

Reset to factory settings

Page 33: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 31/ 54

Clear the counters

2nd settings page Previous page

Changing the internal numbers

Changing the counters

Over Temperature Alarm

Acceleration profiles edition

Deceleration profiles edition

Page 34: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 32/ 54

5.2 Setting the temperature probe Access the temperature probe calibration

menu.

The upper value “Probe Temp.” corresponds to the physical value measured by the temperature sensor, located in the lid. The lower value “Displayed Temp.” is the value displayed on the screen. This is the temperature value of the samples in the buckets, during a centrifugation cycle. A correction of + /- 2°C is possible in the case of deviation with the temperature of the sample in the buckets.

Press the “-” or “+” keys to adjust the value.

Confirm by pressing ok.

5.3 Unbalancing Sensitivity Detection The centrifuge has a reaction to the load balancing faults which may be different depending on the media on which the centrifuge is placed. A calibration of the imbalance sensitivity is therefore necessary during installation. Required equipment: Use the reference balancing kit 71122002.

Access the unbalance sensitivity calibration menu.

Equip a rotor swing-out and its 4 buckets in the centrifuge. Remove the inserts of the buckets.

Low threshold

1. Place the round 50 gram weight in one of the buckets and close the lid,

2. Press the button “50gr” 3. The rotor accelerates until it stabilizes. 4. The centrifuge shakes, then the rotor

slows down. 5. The low threshold is then determined. It

appears under the button “50 gr”. (e.g. 480)

6. Open the lid and remove the 50 gram weight.

High threshold

7. Place the 100 gram weight in the same bucket.

8. Proceed in the same way as for the 100 gram weight.

9. The low threshold is then determined. (e.g. “520”)

Page 35: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 33/ 54

10. Open the lid and remove the 25 gram weight.

Average

11. Press on the Next button to average the setting. (e.g. [480+520]/2 = 500 ) The average value is displayed. By default, the factory setting value is 500.

Check The following steps are compulsory. High threshold:

Do not adjust this threshold with a weight greater than 100 gr. This would cause a breach of warranty. The appliance has been dimensioned for a maximum threshold of 100gr.

12. Place the 100 gr weights in a bucket. 13. Start the cycle. 2000 rpm / 1 minute / 9

Acceleration / Braking 9 14. The centrifuge must stop during the

acceleration, and display the message ERROR1: The 100 gr imbalance is not tolerated. Repeat the High threshold check 3 times: ERROR1 must appear each time.

Low threshold:

15. Leave only a weight of 50 gr in a bucket. 16. Start the cycle. 2000 rpm / 1 minute / 9

Acceleration / Braking 9 17. The centrifuge should achieve the

desired speed and then slow down and stop without the following error message appearing: The 50 gr imbalance must be tolerated. Repeat the Low threshold check 3 times: Each cycle should take place normally.

Note : This setting value is viewable in HOME -> INFORMATION menu

Page 36: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 34/ 54

5.4 Setting the lock Lid lock is an electromechanical system including moving parts. Keep your fingers / hand away from moving parts when operating it.

Locking steps : :

The lid is lowered manually. The left strike of the cover (1) is detected by the approach sensor (2). The motor reducer (6) starts. The open position sensor (8) changes to state 0.

The hooks (7) drive the 2 latches down. When the left strike is detected, the closed position sensor of the strike (3) goes to state 1.

The locked position sensor of the lock (5) goes to state 1. The motor reducer stops: The markings (9) of the crankshaft and the deadbolt are aligned: the lock is in the locked position. The lock is in closed position when the marks on the deadbolt and the crankshaft are aligned.

Unlocking steps : The steps are in the reverse order of the locking operation. The motor reducer rotates in the opposite direction.

6

2

1

8

7

3

7

9

5

Page 37: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 35/ 54

Setting the Time Delay of the lock

Access the unbalance sensitivity calibration menu.

Locking timer: The motor reducer timer when closing. Adjust with the “-” and “+” buttons up to the position described above, in Section 3.4

Unlocking timer: Motor reducer time delay when closing. Adjust with the “-” and “+” buttons until the bolt is completely clear at the opening, in order to let the lid strike plate escape freely.

Note : These setting values are viewable in HOME -> INFORMATION menu.

Page 38: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 36/ 54

5.5 Auto-Diagnostic

Access auto-diagnostic menu.

Action buttons :

Contactor

Compressor

Lid mot.

Ogive bottom LED indicator Indicators :

Speed : xx RPM

Temperature : yy °C

Contactor State

Lid locked sw

Lid unlocked sw

Lid Approach

Open button

Start button

Motor Temp

Functions Use Action buttons :

"Contactor"

Close the lid. Access the Auto-Diagnostic menu Press “Contactor”: the button turns green and the motor contactor is energized. The return of information “contactor state” turns green meaning the contactor and its auxiliary contact work properly.

"Compressor" Manual start of the compressor and the chiller fan Performance: By closing the lid and activating the compressor (pressing on compressor), after 5 minutes, ice forms on the sides of the bowl. Coolant is always present in the circuit.

Lid mot. The locking system motor is energized. The mechanism runs without stopping. After this test

LEDs The leds under the control unit flash when pressed. (for control unit equipped with leds).

Indicators :

"Speed" Rotate the rotor by hand (no matter what direction): The rotation value is displayed in RPM.

"Lid Locked sw"

With lid open before entering the diagnostic menu. Lower the lid and with the manual tool rotate the locking system till the locked position: the button changes state (red-> tick green) meaning both the “locked” micro switch and magnetic detector work fine

"Lid unlocked sw"

Operate the lock with the manual opening key: ¼ turn clockwise, then ¼ turn anti-clockwise”. When the “unlock” micro contact engages, change of state.

"Lid approach" Passes in green checkmark when the lid is lowered (magnet detection).

"Open button" Press the button to open lid: becomes green

"Start button" Press Start/Stop: Change of state.

"Motor Temp." Disconnect the motor over temperature sensor, (intermediate connector in front of the motor contactor) Check ths state change

Exit with the cross.

Page 39: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 37/ 54

5.6 Information update Caution: This information concerns the serial numbers of the electronic boards fitted on the centrifuge. In case of changing an electronic board, these values should be changed:

Access the menu using the following key:

SERIAL NUMBER: Unique serial number of the centrifuge SOFTWARE: Information in area B and E aure updated automatically and related to the software configuration of KATRINA

Centrifuge fw: Microcontroller software version

Display fw: Screen software version HARDWARE: In case of part replacement, information of area D have to be updated:

Electronic Control Center serial n° (Ogive)

Tachometer board serial n°

Motor serial number

Exit

5.7 Counters modification The values are used to determine the age of the device and the rotating accessories. Changing values should ONLY be done in the case of an exchange of the control center. The values to be set must be identical to those recorded in the faulty control center. It is therefore necessary to record the values before removing the faulty control center. If the values are lost monitoring the life of the rotor will not be correct. A rotor may therefore be used beyond their maximum lifespan. This causes a DANGEROUS situation for users and machine. In case of inability to access data from the defective control center these values can be retrieved by the factory.

Enter the counters modification menu

Press the item to change. The input keyboard then appears. Fill in the value recorded from the old control center. Repeat this for all counts.

Exit

Page 40: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 38/ 54

5.8 Overtemperature alarm menu This alarm can protect the samples subjected to excessive heating of the air with respect to the set temperature. In case of triggering of the alarm, the cycle is aborted. ERROR 3 message is displayed and buzzer starts.

Entering the overtemperature alarm menu

Adjust the desired value. The default is + 10° C. The value is adjustable between + 5° C to + 20° C. Example: for a 10 ° C setpoint when the over temperature threshold is set at 5 ° C, the alarm will sound if the air temperature is> 15 ° C

Exit

Note: If the sample temperature exceeds 43 ° C, but also if the temperature of the air inside the bowl becomes higher than 50 ° C, then the cycle is aborted with the message ERROR 3.

5.9 Firmware update KATRINA centrifuge is controlled by 2 firmwares: • A firmware application called “Display firmware”, loaded into the screen board. It manages a HMI (Human Machine Interface), and visual screens. • A firmware application called “Microcontroller firmware”, loaded into the microcontroller board. It defines the behavior, and the interactions between the different parts of the centrifuge. It is necessary to use a specific programming cable, supplied with reference 71199004 “Software Programming Kit”. See detailed update process in annex.

Page 41: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 39/ 54

5.10 Factory Reset After loading a new version of centrifuge firmware, it may be necessary to perform a factory reset, which returns the device to factory configuration. This deletes the following data: - Lock setting (a default is recorded) - The imbalance calibration - User programs - The error counters (reset to 0) - The serial numbers of the unit and electronic cards

Access the factory reset menu

"Warning: Return to factory settings All settings of the user, and programs will be lost. Continue?" OK to continue

"Do not power off unit during reset." Wait until the following message.

Turn off & turn on the unit.

Page 42: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 40/ 54

5.11 Clearing counters The counters are used to determine the age of the device, and the rotating accessories. In case of resetting these counters, monitoring the life of the rotor will not be correct. A rotor may therefore be used beyond their maximum lifespan. This causes DANGER for users and machine. This menu is primarily used in the factory.

Access the factory reset menu

"Warning: This will erase all internal counters, including centrifugation cycles counters, and error counters. This should be achieved in specific cases. Keep a record of the current settings before proceeding. Continue?" OK to continue

Page 43: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 41/ 54

6 Standard Exchange of Components

6.1 Safety Rules

Before any repair intervention, observe the safety rules by turning the centrifuge off (rear circuit breaker). In the event of non-compliance with these instructions, there is a risk of electrical shock.

Depending on its use, a centrifuge can be contaminated. Before any service operation, please refer to § DECONTAMINATION at the beginning of this manual.

6.2 Spare parts & tools

Technote References Name

soon 71124000 Gas spring kit (set of 2)

soon 71124001 Bowl seal kit

soon 71124002 Motor seal kit

soon 71124003 Shock absorbers kit (set of 5)

soon 71124004 Motor Kit

soon 71124005 Electronic Control Center kit

soon 71124006 Lid lock système kit

soon 71124007 Temperature sensor kit

soon 71124008 Tachometer board kit

soon 71124009 Mains filter board kit

soon 71124010 Locking motor supply board kit

soon 71124011 230V 50Hz Motor contactor kit

soon 71124012 25A circuit breaker kit

soon 71124013 Speed sensor kit

Every service kit is delivered with an associated « technote ». This document describes every step of the replacement process as well as the required tools. “Tutorials” video will shortly be available for certified service people.

6.3 Settings form In case of electronic control center replacement, If the state of the control center allows to get them, record the values in the tables below.

Programs :

Name Speed Duration Temperature Acceleration Braking

1

2

3

4

5

6

7

8

9

10

Page 44: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 42/ 54

Parameters :

Access the following menus

Accessing menus

Access the parameters

Function Description Value

Information menu :

SERIAL NUMBER (Also on the machine label)

CONTROL CENTER ID (Also on label attached to control center)

ACCELEROMETER ID (Visible on tach board)

MOTOR ID (Visible also on tab.)

User Preferences menu:

Mode RPM/RCF

Frequency of grease

CYCLES COUNT

SWING OUT CYCLES COUNT ≤ 4000 RPM

SWING OUT CYCLES > 4000 RPM

RUNNING HOURS

COLD MODULE RUNNING HOURS

COLD MODULE STARTS

LID OPENING

Unit of temperature °C

°F

Timing mode

ON

OFF

Error Counter 01 : / 02: / 03: / 04: / 05: / 06: / 07: / 08: / 09: / 10 : 11 : / 12: / 13: / 14: / 15: / 16: / 17: / 18: / 19: / 20 :

Postcool ON

OFF

Pre-cooling Mode AIR

ROTOR

Enter service

LOCKING TEMPO

1

2

1 3

3

2

Page 45: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 43/ 54

UNLOCKING TEMPO

IMBALANCE SETTING

Page 46: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 44/ 54

7 Functional checks

7.1 Mechanical Safety

Hinge: (lid) Check the tightness of the screws, and the integrity of the hinge by moving side-to-side. There should be no movement other than that of the hinge rotation.

Control Center: The Control Center must be secured by a nut located near the switch and fixed to the body of the

centrifuge. The lower protection (translucent plastic panel) must be screwed in place under the SUBD9 programing port.

Accessories: Check that the maximum cycle number marked on the free rotor is far superior to the value

recorded on the counter “SWING OUT CYCLES COUNT”

Conduct a visual inspection:

1) The surface of the rotors should be:

Smooth

Regular

Circular in shape on the periphery of the trunions

Check the absence of:

Trace of impact

Corrosion

Cracking

Wear

For correct surface condition of the trunions

7.2 Electrical installation – Protective earth In order to protect user against electric shock, KATRINA protective earth conductor must always be properly connected to the installation. This must be checked at least once a year, using a certified continuity tester.

3

Page 47: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 45/ 54

7.3 Speed Check Required equipment: Calibrated tachometer and reflective adhesive optical tachometer. Place a reflective strip on the central part of the rotor. Remove the central porthole located in the middle of the lid. Close the lid. Start a cycle at the desired speed When the target speed is reached, place the speedometer on the central porthole. The speed displayed by the measurement device must be equal to that displayed on the screen, within a tolerance of ± 10 rpm If this is not the case, check that the reflective tape is well placed in the central part of the rotor, and firmly hold the optical tachometer, without movement during the measurement.

7.4 Checking the Timer Required equipment: Calibrated stopwatch. Define a centrifugation cycle of the desired duration (e.g.: 10 min, 3000 rpm, acceleration 9, braking 9). Start the cycle by pressing start. To start the centrifuge stopwatch, operate the stopwatch. Stop the stopwatch at the end of the programmed time. Check that the centrifuge starts to brake well.

7.5 Checking the Temperature This test must be done starting from ambient temperature for both centrifuge and accessory. If you are using a windshielded accessory, windshield lid shall not be used. - A reference thermometer is required. - If you are using RHD6 accessory, Tubes with water already cooled down to 4°C are required - If you are using RX6-1000 accessory, water already cooled to 4°C will be poured directly in the buckets - A test program (or routine) must be created with the following parameters

o Speed: 2700RPM for RX6-1000, 2300RPM with RHD6

o Temperature 4°

o Duration 15mn

1. Insert your accessory, close the centrifuge’s lid, and start the precool cycle. 2. Wait for the cycle’s end and full stop 3. Open the lid and, without waiting, insert the tubes already preconditioned to 4°C, close the lid, pour 200ml

of water in each buckets if you are using RX6-1000 accessory. 4. Start the previously defined test cycle 5. Wait for the end of the cycle and centrifuge’s full stop. 6. Open the lid and check the tubes water temperature, it must be 4°C +/-2°C

Page 48: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 46/ 54

8 Annexes

Return authorization Return authorization N°

A return number will be provided after contacting us. In order to process your file, we ask you to return this

document duly completed by you. This document must be affixed outside the package. Within 72 hours upon

receipt of material, we will send you an acknowledgment of receipt.After diagnosis by our technicians, we will

send you a quote.

Your contact information

Company / Laboratory:

Interloculator:

Website address

Postal code Town

Country

Phone

Email Address

Billing address(if different from

the site address)

Postal code Town

Return address

Society

Address :

Postal code Town

Description of your material

Designation Catalog Ref.

Serial No

List of accessories attached to

your equipment

Reason for return

Decontamination

Agent used Duration of application

Operator

Remarks

1. It is your responsibility to properly protect the material and use suitable packaging. Any equipment received

damaged or damaged as a result of improper packaging will not be warranted. Taking under guarantee of the

equipment will be validated by our technical service only after inspection of the device. The guarantee will

not be applicable in case of mishandling of the user or deteriorated items due to misuse (for more details, see

our general conditions of sale).

Page 49: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 47/ 54

Page 50: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 48/ 54

Page 51: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 49/ 54

Page 52: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 50/ 54

Page 53: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 51/ 54

Page 54: BLOOD BANK CENTRIFUGE 6x1000ml 4x2000ml

Original Instructions - Rev.1.0 Service Manual– KATRINA centrifuge 52/ 54