analytical performance evaluation of a dedicated
TRANSCRIPT
Analytical performance evaluation of a
dedicated calibrator and controls for
emicizumab▼quantification
Narayanan Ramamurthy,1 Cheryl Kucharski,1 Mags McInerney,1
David Chen,2 Michael Morris1
1r2 Diagnostics, South Bend, IN, USA; 2Genentech Inc., South San Francisco, CA, USA
▼ This medicinal product is subject to additional monitoring. This will allow quick identification of new safety information. Healthcare professionals are asked to report any suspected adverse reactions.
These should be reported to the Regulatory authorities in your country according to your national requirements.
Disclosures for Narayanan Ramamurthy
2
Disclosures
Employment r2 Diagnostics
Emicizumab
– Bispecific antibody that mimics the
function of missing activated FVIII,
restoring haemostasis in PwHA1
– Much higher apparent FVIII activity than
expected with one-stage FVIII assays
MOSA for emi
– Calibration curve:
• Emi calibrator, FVIII DP, aPTT reagent
and CaCl2
– 1:8 predilution on analyser prior to assay
3
Background
Emicizumab
FIXaFX
aPTT, activated partial thromboplastin time; DP, depleted plasma; emi, emicizumab; F, factor; MOSA, modified one-stage assay;
PwHA, persons with haemophilia A1. Kitazawa T, et al. Thromb Haemost 2017;117:1348–1357.
Optimised on BCS XP (Siemens) using the
FVIII one-stage assay method
– Samples pre-diluted 1:8 on analyser
– Emi calibrator instead of standard
human plasma
– Emi controls instead of FVIII normal and
abnormal plasma controls
Reagents:
– Emi calibrator: 100 µg/mL (r2 Diagnostics)
– Emi controls: L1=25 µg/mL; L2=75 µg/mL
(r2 Diagnostics)
– Actin FSL, CaCl2 and FVIII deficient plasma
(Siemens)
Calibration range:
– 10–100 µg/mL emi
1.60
1.65
1.70
1.75
1.80
1.85
1.90
0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20
Lo
g c
lot
tim
e (
se
c)
Log [emi] (µg/mL)
Emicizumab MOSA calibration curve
4
MOSA for emicizumab
Performed per CLSI EP17-A2 guidelines:
– Samples: FVIII DP with and without spiked emi
– Reagents: 2 lots of all assay reagents
LoB:
– 5 blank samples x 4 runs/day x 3 days
LoD:
– 7 low level emi samples (2.5–17.5 µg/mL) x 4
runs/day x 3 days
LoQ:
– 7 low level emi samples x 4 runs/day x 3 days
5
Analytical sensitivity – LoB/ LoD and LoQ
Parameter Result (µg/mL)
LoB 1.8
LoD 3.2
LoQ 10.0
LoB, limit of blank; LoD, limit of detection; LoQ, limit of quantitation
CLSI EP17-A2, Clinical and Laboratory Standards Institute document EP17-A2—Evaluation of Detection
Capability for Clinical Laboratory Measurement Procedures; Approved Guideline—Second Edition.
Performed per CLSI EP06-A guidelines:
– Samples: FVIII DP spiked with emi
– Reagents: 3 lots of all assay reagents
Method:
– Sample series (0–250 µg/mL emi) tested
– 4 runs/ sample
– Samples >100 µg/mL run with reflex testing
Results
– Lot 1 slope=0.89, R2=0.99
– Lot 2 slope=0.95, R2=1.00
– Lot 3 slope=0.93, R2=0.99
6
CLSI EP06-A, Clinical and Laboratory Standards Institute document EP06-A—Evaluation
of the Linearity of Quantitative Measurement Procedures: A Statistical Approach
Assay linearity
Performed per EP05-A3 guidelines:
– Samples:
• FVIII DP spiked with emi
(10–200 µg/mL)
• Calibrator and controls (3 lots each)
– Analysis:
• Nested ANOVA
Repeatability:
– 20 days x 2 runs/day x 4 replicates/run
– 3 lots of reagents tested (1 lab)
Results:
– Excellent precision seen across
measurement range
– Total %CV values 6.3–10.2%
7ANOVA, analysis of variance; CV coefficient variation
CLSI EP05-A3, Clinical and Laboratory Standards Institute document EP05-A3—Evaluation of
Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition
Precision – repeatability within one site
Within
run
(%CV)
Between
run
(%CV)
Between
day
(%CV)
Between
calibration
(%CV)
Between
lot (%CV)
Total
(%CV)
Spiked
samples3.4–4.0 2.0–7.0 3.0–5.2 0.0–4.0 0.0–4.9 6.3–10.2
Calibrator 3.6 1.9 3.4 4.1 0.0 6.7
Control
L13.3 3.4 2.4 1.2 2.4 6.0
Control
L23.1 3.3 1.6 4.9 0.0 6.9
8
Precision – reproducibility across three sites
Within
run
(%CV)
Between
run
(%CV)
Between
day
(%CV)
Between
lab
(%CV)
Total
(%CV)
Spiked
samples5.1–9.0 0.8–1.4 0.9–2.4 0.9–3.6 5.5–10.2
Calibrator 5.2–6.7 0.8–2.5 0.6–2.8 4.6–6.1 7.5–9.9
Control
L16.1–6.5 0.9–1.1 1.4–1.8 0.9–1.5 6.5–6.9
Control
L24.8–5.1 0.5–1.4 1.0–2.3 0.9–2.1 5.3–6.1
Performed per EP05-A3 guidelines:
– Samples:
• FVIII DP spiked with emi
(10–200 µg/mL)
• Calibrator and controls (3 lots each)
– Analysis:
• Nested ANOVA
Reproducibility:
– 5 days x 2 runs/day x 3
replicates/run
– 1 reagent lot tested across 3 labs
Results:
– Excellent precision seen across
measurement range
– Total %CV values 5.3–10.2%
CLSI EP05-A3, Clinical and Laboratory Standards Institute document EP05-A3—Evaluation of
Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition
Performed per EP07-A2 guidelines:
– Samples: FVIII DP spiked with emi
– Analysis: paired difference method
– Reagents: 3 lots of all assay reagents
Method:
– 10, 100 and 200 µg/mL emi spiked with 4
levels of each interferent + one 0
interferent level
– Each sample tested in quadruplicate
Results
– aPCC, standard rFVIII and porcine rFVIII
falsely elevate the emi measurement
– No interference from rFVIIa and
heparin (up to 1 U/mL)
9aPCC, activated prothrombin complex concentrate; rFVIIa, activated
recombinant factor VII; rFVIII, recombinant factor VIII
CLSI EP07-A2, Clinical and Laboratory Standards Institute document EP07-A2—
Interference Testing in Clinical Chemistry; Approved Guideline—Second Edition
Analytical specificity – exogenous interferences
Interferent Target range Effect
aPCC 0–6 U/mL Emi falsely elevated
rFVIIa 0–5 µg/mL No interference
Standard rFVIII 0–2 U/mL Emi falsely elevated
porcine rFVIII 0–2 U/mL Emi falsely elevated
Unfractionated
heparin0–10 U/mL No interference up to 1 U/mL
Maximum allowable difference for each combination ≤10%
Performed per EP07-A2 guidelines:
– Samples: FVIII DP spiked with emi
– Analysis: paired difference method
– Reagents: 3 lots of all assay reagents
Method:
– 10, 100 and 200 µg/mL emi spiked with 4
levels of each interferent + one 0
interferent level
– Each sample tested in quadruplicate
Results
– No interference from endogenous interferents
10CLSI EP07-A2, Clinical and Laboratory Standards Institute document EP07-A2—Interference Testing
in Clinical Chemistry; Approved Guideline—Second Edition
Analytical specificity – endogenous interferences
Interferent Target range Effect
Hemolysis
(hemoglobin)0–500 mg/dL No interference
Lipemia
(triglycerides)0–2000 mg/dL No interference
Icterus (bilirubin) 0–20 mg/dL No interference
Lupus anticoagulant
(high titer)
0–73.2 DSec StaClot
LANo interference
FVIII inhibitor
(high titer)0–108 BU/mL No interference
Maximum allowable difference for each combination ≤10%
Performed per EP09-A3 guidelines:
– Patient samples: PwHA on emi treatment
Method:
– MOSA tested at 3 sites on fresh and frozen
samples (ongoing)
– Samples tested on emi specific ELISA (used
in phase 3 studies of emi)
– 146 samples tested across all sites
Results
– Slope=0.97; R2=0.9
– Excellent agreement between ELISA and
MOSA across the range 0–150 µg/mL
11
CLSI EP09-A3, Clinical and Laboratory Standards Institute document CLSI EP09-A3—Measurement Procedure
Comparison and Bias Estimation Using Patient Samples; Approved Guideline—Third Edition
Method comparison – MOSA vs ELISA
y = x
y = 0.9727 + 1.9628
R² = 0.9007
0
20
40
60
80
100
120
140
160
0 20 40 60 80 100 120 140 160
MO
SA E
mic
izu
mab
Co
nce
ntr
atio
n (μ
g/m
L)
ELISA Emicizumab Concentration (μg/mL)
Linear Regression Plot: ELISA vs MOSA All sites
Stability:
– Sample stability: emi treated PwHA plasma
– Shelf-life stability: emi calibrator and controls
– Reconstituted stability: emi calibrator
and controls
Method:
– Samples tested at T0 and at various
timepoints at each storage condition
– Timepoint results compared to T0
– Allowable drift: ±2 SD
Results:
– Sample, calibrator and controls demonstrated
acceptable stability for all conditions tested
12h, hour; m, month; RT, room temperature; SD standard deviation; w, week
Stability
Stability
type
Stability
condition
Stability
duration
Sample
RT
-20°C
-80°C
4h
up to 3w
up to 4m
Shelf-life* 2–8°C 12m
Reconstituted
20–25°C capped
8–15°C uncapped
2–8°C capped
8h
8h
24h
*Testing ongoing
A MOSA was developed along with a dedicated calibrator and controls for the measurement of
emi in plasma samples
MOSA shows good sensitivity and linearity across a measurement range of 10–250 µg/mL emi
MOSA shows good repeatability and reproducibility across a measurement range of
10–200 µg/mL (%CV 5.3–10.2)
No measurable impact on emi recovery seen in the presence of endogenous interferences
rFVIIa and unfractionated heparin (up to 1U/mL) show no effect on emi recovery
Standard rFVIII, aPCC and porcine rFVIII at prophylactic levels result in falsely elevated emi
level
MOSA shows excellent agreement with an emi-specific ELISA assay
Sample stability, calibrator and control shelf-life and reconstituted stability are acceptable
13
Summary
Download this presentation: http://bit.ly/2FOk18H