AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than simply collect data. It should provide a means for remotely checking the system and each individual analyzer. It should enable analyzers and calibration system to be configured remotely. It should reliably log data directly from the instruments in digital form. It should automatically warn the user of measurement exceedances or events such as calibration failures. It should be easy to use.

Upload: shea-luttman

Post on 15-Jan-2016

217 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

1

AAQMS – DAS Requirements

A Data Acquisition System utilized in air quality and/or emissions monitoring should do far

more than simply collect data. It should provide a means for remotely checking the system and each individual analyzer.

It should enable analyzers and calibration system to be configured remotely.

It should reliably log data directly from the instruments in digital form.

It should automatically warn the user of measurement exceedances or events such as calibration failures.

It should be easy to use.

Page 2: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

2

Teledyne – Win AQMS DAS WinAQMS is a Windows® based system which provides full control over the entire

system (remote and onsite) enabling automatic calibrations to be performed and system errors to be monitored.

WinAQMS interfaces to analyzers via a RS232 multi-drop serial link, through a USB link or via TCP/IP network link. This allows it to collect data directly from the instruments in digital format thus eliminating digital to analog and analog to digital conversion errors. The bi-directional interface also allows a remote central computer, running Teledyne WinCollect software, to access full control and diagnostics of the analyzers used and data to be collected remotely.

The Data Acquisition system has sufficient data storage to hold well over thirty years

worth of historical data.

Page 3: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

3

Main Features : Win AQMS (DAS) Supports remote communication from anywhere in the world enabling full control over the pollution monitoring

system and direct interfacing with supported analyzers

Full control over calibration cycle periods

Remote communication through dial-up, Ethernet, satellite telephone networks, 3G network as well as short haul modems

Solid state hard disk drive eliminates risk of mechanical failure due to shock, high temperatures, vibration and harsh environments.

Data retrieval via USB Memory Stick or Mini SD card in the event communication networks are unavailable

"Industry Standard" RS232 and USB Communication enables digital communication with a wide range of supported monitoring equipment

Captures minimum, maximum, average values and standard deviations

Creates calculated channels that allows the user to apply calculations to data (logged automatically)

Fifth Generation Data Acquisition & Control System based on 30 years of field experience

Data storage space for over thirty years of 5 minute historical data

Lightning & surge protection facilities

Windows XP PRO Compatible

Page 4: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

4

Specifications : Win AQMS (DAS) CPU: ........ 1.6 Ghz+

RAM: 1 Gbytes

Hard Drive Config.: Solid State 64 GbytesDisplay Driver: Supporting up to 1929 x 1200 (allows dual screen)

Operating System: Windows XP Data Storage Format: Data Stored in Ascii format

Data Averaging Periods: Up to 4 independent data sets Selectable between 1 minute and 7 days

Calibration control: Automatic Span and Zero control. Calibration data is stored separately

Communication ports: 1 RS232 /RS485 serial ports standard, 6 x USB2.0 Ports

Networking :1000 Base T Ethernet 802.11g WLAN

Power: 12 volts < 7 Watt power supply.

Size: 10.1 x 11.5 x 2.7 cm  370 grams

Page 5: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

5

AAQMS with DAS

Page 6: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

6

AAQMS – Sample Conditioning System

Page 7: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

7

AAQMS Analyzer Panel

Page 8: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

8

Mobile AAQMS Station

Page 9: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

9

AAQMS Mobile Shelter

Page 10: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

10

AAQMS Cabinet – Rear & Front Side

Page 11: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

11

AQMS Calibration Procedure

Page 12: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

12

AQMS Calibration Procedure

Page 13: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

13

AQMS Calibration Procedure

The following three-step procedure assumes that the instrument does not have any of the available zero/span (Z/S) or IZS valve options installed. Chapter 7 contains instructions for calibrating instruments with valve options. Chapter 8 contains directions for performing for EPA protocol calibrations.

The initial calibration should be carried out with the analyzer’s reporting range for SINGLE range mode with a range span of 500 PPB (factory default settings for most units). This will enable you to compare your results to the factory calibration.

STEP ONE: Set/verify the analog output reporting range of the M100E:

1] SO2 CALIBRATION: 6400E (Refer manual for reference)

Page 14: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

14

AQMS Calibration Procedure

Page 15: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

15

AQMS Calibration Procedure

STEP TWO: Set the expected SO2 span gas concentration.

Page 16: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

16

AQMS Calibration Procedure

STEP THREE: Perform the zero/span calibration procedure:

Page 17: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

17

AQMS Calibration Procedure

Check the value of the SLOPE and OFFSET test functions (see Section 6.2.1) to verify that they are within the limits listed in Table 7-5)

Model 100E analyzer is now ready for operation.

Page 18: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

18

AQMS Calibration Procedure

The initial calibration should be carried out using the 500ppb range, a span gas with about 400 ppb NO and with the unit set for SINGLE range, which will enable you to compare your calibration results to the factory calibration as listed on the Final Test and Validation Data sheet. The following three-step procedure assumes that the instrument does not have any of the available zero/span (Z/S) valve options installed. Chapter 1 contains instructions for calibrating instruments with Z/S valve options, Chapter 1 for EPA protocol calibration.

STEP ONE: Set the analog output range of the M200E:

2] NOX CALIBRATION: 9110E (Refer manual for reference)

Page 19: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

19

AQMS Calibration Procedure

Page 20: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

20

AQMS Calibration Procedure

STEP TWO: Set the expected NO and NOX span gas concentration. If you supply NO gas to the analyzer, the values for expected NO and NO2 need to be identical.

Page 21: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

21

AQMS Calibration Procedure

Perform the zero/span calibration procedure:

Page 22: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

22

AQMS Calibration Procedure

Page 23: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

23

AQMS Calibration Procedure

The Model 200E analyzer is now ready for operation.

3] CO CALIBRATION: 7001E (Refer manual for reference) Set the expected CO span gas concentration

Page 24: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

24

AQMS Calibration Procedure Perform the Zero/Span Calibration Procedure

The Model GFC7001E Analyzer is now ready for operation.

Page 25: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

25

AQMS Calibration Procedure

4] O3 CALIBRATION: 7001E (Refer manual for reference)

The O3 span concentration value automatically to 400.0 PPB and it is recommended that an O3 calibration gas of that concentration be used for the initial calibration of the unit. To verify that the analyzer span setting is set for 400 PPB

Page 26: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

26

AQMS Calibration ProcedureTo perform an initial Calibration of the Model 400E photometric ozone

analyzer, press

Page 27: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

27

AQMS Calibration Procedure

Page 28: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

28

AQMS Calibration Procedure

The Model 400E ANALYZER IS NOW READY FOR OPERATION.

5] VOC CALIBRATION: 4060 (Refer manual for reference) The analysis has 2 modes: Sample Mode and Analysis Mode.

1) Sample Mode:

In this mode, the Analyzer configures the operational valve to back flush the column and charge the sample loop. The sample mode is programmed to continue for a 7 minute period by factory default; however, it is usually modified to match the needs of a particular process. It is possible to set the Sample Mode duration up to 30 minutes.

Page 29: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

29

AQMS Calibration Procedure

2) Analysis Mode

During this mode, the Analyzer configures the operational valve to feed the gas in the sample loop through the column to the detector. If configured to do so, the Analyzer may read the detector for a 5 second period programmed somewhere during this mode to obtain a baseline reading for further calculation. This mode is usually programmed to continue until the compound of interest is fully calculated from the column. It is possible to set it up to 10 minutes.

During the Analysis Mode, a ‘Peak Detect’ period is programmed at which time the Analyzer reads the Detector Output. The Analyzer stores the maximum reading during this time and uses the value in conjunction with the baseline value to calculate the concentration of the compound of interest. The calculation is performed at the end of the ‘Peak Detector’ period and the result is displayed at the end of Analysis Mode.

Page 30: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

30

AQMS Calibration Procedure

Page 31: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

31

AQMS Calibration Procedure

Page 32: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

32

AQMS Calibration Procedure

Page 33: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

33

SO2 Analyzer – Model 6400EMaintenance Schedule

Item Action Priority Frequency Cal Check

Particulate filter Change particle filter Mandatory Weekly or As Needed No

Flow controller Replace sintered filter & O-rings

Mandatory AnnuallyYes

Pump Rebuild-Kit Mandatory Annually Yes

Zero/span calibration Zero/span calibration As Needed Annually or as needed Yes

Sample Chamber windows and Filters

Clean As Needed As necessary Yes

TEST functions Review and evaluate Recommended Weekly No

Pneumatic sub-system Check for leaks in gas flow paths

Recommended Annually or after repairs involving pneumatics

Yes on leaks, else no

PMT Sensor Hardware Calibration

Low-level hardware calibration

Recommended On PMT or preamp changes

Yes

Page 34: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

34

NO x Analyzer - 9110 Maintenance Schedule

Page 35: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

35

CO Analyzer – GFC 7001E Maintenance Schedule

Page 36: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

36

O3 Ozone Analyzer – Model 400E Maintenance Schedule

Page 37: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

37

Zero Air Generator - Model 701Maintenance

Before any maintenance cycle power on/off, drain any water in bowl

Do not loosen any tubing connection while the 701 is running.

Before working on the 701 plumbing turn off the 701, wait for the pressure gauge to read zero. (This may require that you loosen the rear panel bulkhead union to allow the air to bleed away.)

Even though the front panel power switch is off, there is line voltage present at the power entry terminals, power switch terminals and on the control board. For added safety, remove the power cord from the rear panel receptacle.

Occasionally, depending upon the local conditions, check the inside of the 701 for excessive dirt or dust.

Particularly, check the cooling fan, cooling coil and compressor fan inlet.

Remove any dirt or dust with a vacuum cleaner. Do not use an air jet. This will only redistribute the dirt and will not remove it.

Page 38: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

38

VOC Analyzer – Model 4060 GC Routine Maintenance

The pressure gauges and flow rate should be checked weekly.

All support gases need to be replaced when the primary gauge reaches 100PSIG.

Span calibration required whenever Air/Fuel cylinder is replaced or flame is re-ignited.

Page 39: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

39

VOC Analyzer – Model 4060 GC Routine Maintenance

Item Action Frequency Remarks

Front Panel Pressure Gauge

Setting

Air Set pressure as recommended Daily/Weekly Ensure that press setting is same as

per previous calibration.

Fuel Set pressure as recommended Daily/Weekly Ensure that press setting is same as

per previous calibration.

Carrier Set pressure as recommended Daily/Weekly

Very critical. Do not disturb on front gauge. Ensure that Cylinder outlet pressure is at 80 to 90psig.

Support Gases Cylinder press. Fuel

Replace cylinder when primary pressure reaches 100psig.

As on required Whenever cylinder replaced, perform span calibration.

Page 40: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

40

VOC Analyzer – Model 4060 GC Routine Maintenance

Item Action Frequency Remarks

Front Panel Pressure Gauge

SettingAir Set pressure as

recommended Daily / Weekly Ensure that press setting is same as per previous calibration.

Calibration Gas Cylinders Span Calibration Weekly Whenever cylinder replaced,

perform calibration.

SampleBypass flow

Sample / Zero / Span

Set flow @1LPM as recommended Daily / Weekly Ensure that flow setting is same as

per previous calibration.

Temp. FID / GC/SampleCheck PV/SV values. All temp. should maintained as per SV.

Daily / Weekly Ensure that GC Temperature is as per PV value.

Page 41: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

41

Zero Air Generator - Model 701Maintenance Schedule

Page 42: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than

42

MFC – Model 700E Maintenance Schedule

Page 43: AAQMS – DAS Requirements AAQMS – DAS Requirements 1 A Data Acquisition System utilized in air quality and/or emissions monitoring should do far more than