x -ray c orrelation s pectroscopy instrument
DESCRIPTION
X -ray C orrelation S pectroscopy Instrument. Aymeric Robert – XCS Instrument Scientist LCLS Facilities Advisory Committee Meeting November 12, 2008. Team Leaders : Brian Stephenson Gerhard Gr übel Karl Ludwig Lead Engineer: Eric Bong. X-ray Photon Correlation Spectroscopy. - PowerPoint PPT PresentationTRANSCRIPT
Aymeric Robert
XCS Instrument [email protected]
11/12/20081
SLAC National Accelerator Laboratory
X-ray Correlation Spectroscopy
InstrumentAymeric Robert – XCS Instrument Scientist
LCLS Facilities Advisory Committee MeetingNovember 12, 2008
Team Leaders : Brian StephensonGerhard GrübelKarl Ludwig
Lead Engineer: Eric Bong
Aymeric Robert
XCS Instrument [email protected]
11/12/20082
SLAC National Accelerator Laboratory
X-ray Photon Correlation Spectroscopy
Speckles : scattering patterns produced by the coherent illumination of the sampleXPCS : observation of the time-fluctuation of speckle patternsCharacterization of the underlying dynamics of the system
Coherent X-rays
Coherent Scattering(Speckles)
XPCS is independent of the scattering geometry (SAXS, Diff., GI-SAXS, Refl.,…)XPCS probes dynamical phenomena
Aymeric Robert
XCS Instrument [email protected]
11/12/20083
SLAC National Accelerator Laboratory
Photon ShutterPrimary Slits
Diagnostics
Secondary SlitsDiagnostics
Split and Delay
Focusing Lenses
Attenuators
Diagnostics
Photon Shutter
Defining Slits
DiffractometerLarge AngleDetector Stage
MonochromatorTransport TunnelFEH
Hutch 4
Harm. Rej. Mirrors
Pulse Picker
Secondary SlitsDiagnostics
Goal Time-resolved observation of coherent diffraction patterns using LCLS coherent hard x-rays (6-25keV)
Tailor and characterize X-ray beam parametersSpatial profileIntensityRepetition RateSpectral Bandwidth
X-ray Photon Correlation Spectroscopy Requirements
Versatility of the instrumentation (SAXS, WAXS, GI)Coherence preservation of x-ray optics (to the extent possible)
Detector with small pixel size/Large sample-detector distance
Key Performance Parameters6-20 keV energy range
Using the fundamental and third harmonic0.1-0.01% energy resolution
Aymeric Robert
XCS Instrument [email protected]
11/12/20084
SLAC National Accelerator Laboratory
Photon ShutterPrimary Slits
Diagnostics
Secondary SlitsDiagnostics
Split and Delay
Focusing Lenses
Attenuators
Diagnostics
Photon Shutter
Defining Slits
DiffractometerLarge AngleDetector Stage
MonochromatorTransport TunnelFEH
Hutch 4
Harm. Rej. Mirrors
Pulse Picker
Secondary SlitsDiagnostics
Large Angle Detector Mover& Diffractometer
•Sample/detector distance 7-8m•Decoupled from diffractometer•SAXS, WAXS, GI• 2θ up to 55º, vertical tilt•Height adj.
Aymeric Robert
XCS Instrument [email protected]
11/12/20085
SLAC National Accelerator Laboratory
Previous FAC Reports
FAC Report Oct 2007The committee recommends that the XCS staff retain flexibility in their designs to facilitate change as opportunities and problems are discovered.
FAC Report Jun 2008“ … the committee endorses the decisions that have been made to date.” “ The large offset monochromator also requires design attention and carries risk. This has been recognized by the project and appropriate efforts are being made to create a solid design”
Aymeric Robert
XCS Instrument [email protected]
11/12/20086
SLAC National Accelerator Laboratory
Photon ShutterPrimary Slits
Diagnostics
Secondary SlitsDiagnostics
Split and Delay
Focusing Lenses
Attenuators
Diagnostics
Photon Shutter
Defining Slits
DiffractometerLarge AngleDetector Stage
MonochromatorTransport TunnelFEH
Hutch 4
Harm. Rej. Mirrors
Pulse Picker
Secondary SlitsDiagnostics Monochromatization
scheme for XCS
Aymeric Robert
XCS Instrument [email protected]
11/12/20087
SLAC National Accelerator Laboratory
Photon ShutterPrimary Slits
Diagnostics
Secondary SlitsDiagnostics
Split and Delay
Focusing Lenses
Attenuators
Diagnostics
Photon Shutter
Defining Slits
DiffractometerLarge AngleDetector Stage
MonochromatorTransport TunnelFEH
Hutch 4
Harm. Rej. Mirrors
Pulse Picker
Secondary SlitsDiagnostics
Motivations•XCS Science (i.e Early Science)•XCS Science NEEDS a monochromator•Satisfaction of the XCS User Community
First Coherence WorkshopApplication of Coherent X-rays at the LCLS
10/17/2008
Aymeric Robert
XCS Instrument [email protected]
11/12/20088
SLAC National Accelerator Laboratory
Transport TunnelFEH
Hutch 4
Original Plan for XCS
X-rayCorrelationSpectroscopy
CoherentX-rayImaging
HighEnergyDensity
XCS
Near Experimental Hall
Far Experimental Hall
X-ray Transport Tunnel(200m long)
Source to Sample distance : ~ 420 m
Aymeric Robert
XCS Instrument [email protected]
11/12/20089
SLAC National Accelerator Laboratory
Transport TunnelFEH
Hutch 4
Original Plan for XCS Monochromator
Double crystal Large Offset Monochromator•Energy Range : 6-25keV•Offset : 600mm•Design similar to XPP•Location at the entrance of the XRT (200m)•Possibility of changing the crystal•Preserve “future” capability of multiplexing (i.e. thin crystals
Total Budgeted cost: 1.2M$
Delivery date : 2012Associated XRT cost : ~0.8M$
Aymeric Robert
XCS Instrument [email protected]
11/12/200810
SLAC National Accelerator Laboratory
Transport TunnelFEH
Hutch 4
New Plan for XCS Monochromator
Double crystal Large Offset Monochromator•Energy Range : 6-25keV•Offset : 600mm•Design similar to XPP•Location at the entrance of the XRT (200m)•Possibility of changing the crystal•Preserve “future” capability of multiplexing (i.e. thin crystals
Total Budgeted cost: 1.2M$Delivery date : 2012
Associated XRT cost : ~0.8M$
BUT
~35m
Aymeric Robert
XCS Instrument [email protected]
11/12/200811
SLAC National Accelerator Laboratory
Transport TunnelFEH
Hutch 4
Plan for XCS MonochromatorRelease LO-DCM distance from XCS
Advantages Disadvantages
Relax design parameters for stability
No change in delivery date (2011)
Identical performances More but moderate interferences with ME2 operation
Save the XRT transport cost (0.8M$)
Still not acceptable by XCS User Community
All XCS optical components located on the same concrete slab ( gain in stability)
How to satisfy the request for early science from the XCS User Community and provide
early science capabilities ?
Aymeric Robert
XCS Instrument [email protected]
11/12/200812
SLAC National Accelerator Laboratory
Transport TunnelFEH
Hutch 4
Plan for XCS MonochromatorS
• Use part of the saving from the XRT transport cost to purchase a commercially available High Resolution Double Crystal Monochromator ( pseudo channel cut in vertical scattering geometry )* with small “almost” fixed exit allowing the use of multiple standard reflections and covering easily the LCLS hard x-ray range 6-25keV•Ease polishing of each monochromator crystal • This can be purchased early (very early)
* For LCLS hard x-ray beam, there is no heat load issue. The major issue of that type of monochromator is known to be the cooling system which in our case is irrelevant
Aymeric Robert
XCS Instrument [email protected]
11/12/200813
SLAC National Accelerator Laboratory
Transport TunnelFEH
Hutch 4
Plan for XCS MonochromatorImpact on XCS instrument design :almost NONE
•Operation for some time with just the HR-DCM
•Operation of XCS in CXI line (LCLS main line)
•The table supports for optics/diagnostics are identical to XPP instrument and allow that operation with a 600mm offset•The diffractometer can be placed in either line (CXI/XCS)•The Large Angle detector mover can have its center of rotation on both the CXI and XCS line
Provide earlier science capability for the XCS Instrument
Which other benefits ?
Aymeric Robert
XCS Instrument [email protected]
11/12/200814
SLAC National Accelerator Laboratory
Monochromator for LCLS Instruments : Phase ICX
ICX
IXC
SXC
SXP
PXP
P
X
Provide the High Resolution Double Crystal Monochromator to the XPP Instrument
•Provide the first early LCLS monochromatic beam
•Test the monochromator
•Provide temporarily monochromatic beam to XPP
•Gives XPP more time to get their own monochromator
•Allow EARLY Coherent Scattering experiments in the XPP experimental hutch with control of the longitudinal coherence length (CXI/XCS)
Aymeric Robert
XCS Instrument [email protected]
11/12/200815
SLAC National Accelerator Laboratory
Monochromator for LCLS Instruments : Phase IICX
ICX
IXC
SXC
S Provide the High Resolution DCM to hard X-
ray Coherence Instruments XCS or CXI
•Provide the first early LCLS monochromatic beam in FEH
•Test the monochromator close to FEH location (far from source and close to definite monochromator location)
•Provide temporarily monochromatic beam to CXI
•Allows early commissioning of XCS Instrument
•Evaluation of the impact of very hard coherent x-rays on XCS Science Program
•Allow EARLY Static and Dynamic Coherent Scattering Science in the CXI or XCS experimental hutches with control of the longitudinal coherence length
Aymeric Robert
XCS Instrument [email protected]
11/12/200816
SLAC National Accelerator Laboratory
Monochromator for LCLS Instruments : Phase IIICX
ICX
IXC
SXC
S Return High Resolution DCM to XCS
•Use of HR-DCM as Post Monochromator for experiments requiring very large longitudinal coherence lengths
•Allow to relax the parameters of the LO-DCM to use a single type of crystal (Si (220) for example) as post monochromatization can be achieved with the HR-DCM
•The compensation in terms of lower ΔE/E as compared to Si (111) could be obtained with proper focusing optics (i.e CRL’s)
•If multiplexing : a single and identical crystal in CXI beam (i.e easy to correct if necessary)
Aymeric Robert
XCS Instrument [email protected]
11/12/200817
SLAC National Accelerator Laboratory
Monochromator for LCLS Instruments
XCS Team Leader Meeting Oct 19, 2008
Move the large-offset and split/delay from about 200m upstream of experiment to about 40m upstream of experiment• We therefore endorse moving the large-offset monochromator to the new location.
Adding post-monochromator to scope
• Such a double-crystal monochromator is commercially available, and could be procured immediately for initial use in providing monochromatic beam to all of the early hard x-ray experiments. This will be extremely important for the scientific program. In particular, this will allow the early commissioning of the XCS experiments with the required monochromatic beam prior to the availability of the large-offset monochromator. All of the other XCS beamline components are compatible with operation in the in-line configuration without the large-offset monochromator.• We therefore endorse the concept of adding to the scope a double-crystal, small-vertical-offset monochromator with either multiple or easily changeable crystals.
The CXI team was involved in the discussion and supports the proposed plan
Aymeric Robert
XCS Instrument [email protected]
11/12/200818
SLAC National Accelerator Laboratory
Summary The cost of the post monochromator (2) should be significantly less than the cost savings associated with moving the location of the large offset monochromator downstream (1). Thus the net result of these changes will be to improve performance, improve capability, and making monochromatic x-ray beam available earlier to all experiments, while decreasing cost.*
* XCS Team Leader Meeting Summary Oct 2008
•Early Science for XCS, CXI, XPP•No increase of cost ( i.e even slight cost saving)•XCS instrument commissioned early (with HR-DCM)•Satisfies the expectation of the XCS User Community•Serve the Science Mission of the 3 LCLS hard x-ray instruments
Aymeric Robert
XCS Instrument [email protected]
11/12/200819
SLAC National Accelerator Laboratory
Additional Slides
XCS Science
Aymeric Robert
XCS Instrument [email protected]
11/12/200820
SLAC National Accelerator Laboratory
Sequential XCS
Time-average Brilliance
10 ms < C < hrs
Large Q’s accessible
tt
t
tt
t
12
34
43
21
Intensity autocorrelation function
Aymeric Robert
XCS Instrument [email protected]
11/12/200821
SLAC National Accelerator Laboratory
Ultrafast XCS : Split & Delay
Peak Brilliance & Pulse Duration
pulse duration < C< several ns
Large Q’s accessible
Detector
Aymeric Robert
XCS Instrument [email protected]
11/12/200822
SLAC National Accelerator Laboratory
XCS Science @ LCLS
LCLS ParametersFull Transverse Coherence
High Peak BrillianceShort pulse duration 230fs
Ultra Fast Mode
Dedicated 2D-Detector
8 and 24 keV
High Time–average BrillianceRep. Rate 120 Hz
Sequential Mode
Aymeric Robert
XCS Instrument [email protected]
11/12/200823
SLAC National Accelerator Laboratory
Crystal /Scattering
Angles (2θ)Coherence
Length ξL [μm]
1.5 Å 0.5 Å 1.5 Å 0.5 Å
Si(111) 1.4x10-4 27.6 9.1 1 0.3
Si(220) 6.1x10-5 45.8 14.9 2.45 0.8
C*(111) 5.9x10-5 42.5 13.9 2.54 0.9
C*(220) 2.3x10-5 -- 22.8 6.5 2.2
Multi-crystal XCS Monochromator