Getting Started on the Rotor-Gene-6000

www.corbettresearch.com

Jennifer McMahon, PhD

corbett

LIFE SCIENCE

Page 2: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

2

www.corbettresearch.com

corbett

RESEARCH

the world’s only real-time rotary thermo-optical analyser

Page 3: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

4

www.corbettresearch.com

corbett

RESEARCH

Getting Started……

Click on advanced tab for step by step run set up

Choose run template depending on chemistry

Dual labeled probes = hydrolysis probes

SYBR = intercalating dyes

Click on Rotor-Gene icon

Page 4: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

5

www.corbettresearch.com

corbett

RESEARCH

Select the rotor you are using, put the rotor into the machine with the locking ring attached

The locking ring tick box must be ticked to proceed

Click “Next”

Page 5: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

6

www.corbettresearch.com

corbett

RESEARCH

Insert user name, any notes (primer conc, primer sequence mix used etc) and volume required (10-25uL)

This info is saved and can be printed in reports

Click “Next”

Page 6: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

7

www.corbettresearch.com

corbett

RESEARCH

click on “Edit Profile” to change run parameters

section to be altered is greyed out, click on “Hold Temperature” to change temperature, click on “Hold Time” to change time

Always read manufacturer’s instructions regarding Hot start Taq activation times

Page 7: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

8

www.corbettresearch.com

corbett

RESEARCH

Click on “Cycling” to adjust the cycling parameters

Default cycle number is 40

To change a step click on the temperature and adjust then click on the time and adjust

Page 8: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

9

www.corbettresearch.com

corbett

RESEARCH

Data is acquired each cycle (blue dots)

No blue dots = no data = very bad

Acquire at the end of the extension phase

Click on “Aquiring to Cycling A” to select colours

Move the colours you want from “Available Channels” to “Acquiring Channels” using arrows

Use Green channel for SYBR green

Page 9: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

10

www.corbettresearch.com

corbett

RESEARCH

With SYBR use “Melt” to look for primer-dimer or non-specific product

Melt should start from acquiring temperature

If you start the melt from annealing temp the melt may have a shoulder on it

Use default settings, make sure you are acquiring data (blue dots)

Page 10: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

11

www.corbettresearch.com

corbett

RESEARCH

Gain can be compared to the aperture on a camera and is used to fit data on raw scale

If your starting fluorescence signal is weak the gain should be high, if strong the gain should be low

Many people just accept default gain and go to “Next” on this screen, others optimise the gain for each run

Click on “Gain Optimisation” to begin

Page 11: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

12

www.corbettresearch.com

corbett

RESEARCH

Click on “Optimise Acquiring” and “Perform Calibration Before 1st Acquisition”

If you have different colours in different positions click “Edit”

The colour you want to use must be in the tube position for calibration to be successful

Don’t set gain on no template controls

Page 12: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

13

www.corbettresearch.com

corbett

RESEARCH

Click on “Start Run”

Save file name in directory of your choice

Default nomenclature has name of template file, date and version (1 for 1st run that day etc)

Page 13: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

14

www.corbettresearch.com

corbett

RESEARCH

To save as a template open template folder C:\Program Files\Rotor-Gene 6000 Software\Templates and save as template file (*.ret)

Run files are *.rex

Page 14: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

15

www.corbettresearch.com

corbett

RESEARCH

Choose to enter sample names now or click “Finish” and enter sample names later

Run will commence, while the experiment is running you will see:

Raw Channel - fluorescence collected cycle by cycle

Temperature - profile as temperature cycles

Profile Progress – tells you time remaining, shows you where you are. “Skip” allows you to skip to next step (e.g. from cycling to melt); “Add 5 cycles” lets you add cycles at the end

Page 15: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

16

www.corbettresearch.com

corbett

RESEARCH

Click on “Edit Samples” on RHS of screen

Can highlight blocks of cells for naming etc (like excel)

Samples with the same name will be treated as replicates regardless of position

Type can be standard, no template control, unknown (sample)

Adding sample names and types

Page 16: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

17

www.corbettresearch.com

corbett

RESEARCH

Drop down “Type” menu to view options

If you choose “Standard” you have to add a value for concentration

If you are running two standards define one set on page 1 then make a new page (page 2) to define the second set of standards

Groups can be used to define e.g HKG (housekeeper) and GOI (gene of interest

Page 17: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

18

www.corbettresearch.com

corbett

RESEARCH

Can choose colour from palette and can set colour gradient across range of samples/standards

Page 18: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

19

www.corbettresearch.com

corbett

RESEARCH

Post run analysis - basic

Raw data is displayed on screen during the run and at the end of the run.

Analysis is needed to correct for background etc

Click on “Analysis” on the main toolbar at the top

Click on “Quantitation” tab

Click on channel so that it gets greyed out

Click on “Show”

Page 19: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

20

www.corbettresearch.com

corbett

RESEARCH

The software looks at the standards and draws a threshold, calculates the standard curve and the replicates

Page 20: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

21

www.corbettresearch.com

corbett

RESEARCH

“Dynamic Tube” normalisation should be on to correct for background

“Slope Correct” should be on for dual labeled probes

“Ignore first” can be useful if the first few cycles are different (usually higher) than the rest of the run

Page 21: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

22

www.corbettresearch.com

corbett

RESEARCH

This option is ticked by default and is used to determine the average background of each individual sample just before amplification commences. Standard Normalization simply takes the first five cycles and uses this as an indicator for the 'background' level of each sample. All data points for the sample are then divided by this value to normalize the data. This process is then repeated for all samples. This can be inaccurate as for some samples the background level over the first five cycles may not be indicative of the background level just prior to amplification. Dynamic Tube Normalization uses the second derivative of each sample trace to determine a starting point for each sample. The background level is then averaged from cycle 1 up to this starting cycle number for each sample. This method gives the most precise quantitation results. Alternatively with some data sets it may be necessary to disable the dynamic tube normalization. If this is the case the average background for each of the samples is only calculated over the first 5 cycles. If the background is not constant over the cycles before amplification it will result in less precise data.

Dynamic Tube Normalisation

The background fluorescence (Fl) of a sample should ideally remain constant before amplification. However, sometimes the Fl-level can show an increase or decrease due to the effect of the chemistry being run and produce a skewed noise level. The Noise Slope Correction option uses a line-of-best-fit to determine the noise level instead of an average, and normalizes to that instead. Turning on this option can tighten replicates if your sample baselines are noticeably sloped. This function improves the data when raw data backgrounds are seen to slope upward or downward before the amplification Takeoff point (CT). It is very helpful for runs when for example the FAM background is seen to creep upwards due to gradual probe autohydrolysis.

Slope correct

The first couple of cycles in a quantitation run are not usually representative of the rest of the run. For this reason, you may get better results if you select to ignore the first few cycles. If the first cycles look similar to cycles after them, you will gain better results by disabling this function, as the normalization algorithm will have more data to work with. You can ignore up to ten cycles.

Ignore first

Page 22: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

23

www.corbettresearch.com

corbett

RESEARCHIf there are no standards in the run you have to set the threshold manually.

Click on “Threshold” button

“Click to select the threshold height” comes up, move the line up or down to select the threshold

Do this in log scale

Linear scale view, click on “log scale” button to get log scale

Log scale view, click on “linear scale” button to get linear scale

Page 23: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

24

www.corbettresearch.com

corbett

RESEARCH

Standards are plotted in blue, samples as red

Software calculates relationship value (R2) – ideally 0.99

Software calculates the efficiency out of 1

1 means 100% efficient or doubling every cycle

Page 24: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

25

www.corbettresearch.com

corbett

RESEARCH

Software calculates “Rep. Ct” to get replicate Ct from triplicates (geometric mean)Software calculates “Rep. Ct Stc” to get standard deviation for triplicatesSoftware calculates “Rep. Calc Conc.” to calculate concentration from a standard

curveColumns can be switched off by right clicking on grey bar and changing optionsTable can be exported to MS Excel by right clicking on table text and choosing

export to excel

Page 25: Www.corbettresearch.com corbett RESEARCH 1 Getting Started on the Rotor-Gene-6000  Jennifer McMahon, PhD corbett LIFE SCIENCE

26

www.corbettresearch.com

corbett

RESEARCH

Raw melt data can be viewed at end of run. Fluorescence decreases as temperature increases

The software plots a mathematical derivative to turn the raw data into a peak

Click on “Analysis” then highlight “Melt” then click “Show”