RMN Quantitative
Dr. Fabrice Moriaud, Product Manager 31ème Réunion d’Utilisateurs RMN, 28 novembre 2017, Lyon, France
Innovation with Integrity
Quantification by NMR
Quantitation by NMR has many faces and can answer many questions
• Absolute concentration (mM) • Potency / Purity (%)
– … in full automation (but fully editable) or manual – … with internal or external standards – … high throughput or single samples
– Bruker provides solutions for all of these
Innovation with Integrity
Popular qNMR Applications
qNMR method: Internal standard (IS) External standard (ES) with PULCON Units: wt/wt% - potency mM Error (accuracy): < 1 % > 2% (typically 5%) Workflow: IS + analyte in 1 tube 1) calibrate with ES 2) measure analyte Sample preparation
(1) (2) 10-20 minutes: because of
manual and accurate weighing of two compounds < 1 minute and can be fully automatised
In automation with PotencyMR : typically 2 %
Innovation with Integrity
Potency: Round Robin Test
• 15 / 25 participants measured true potency with +- 1% • Using a standard operating procedure (SOP) • Deviation > 2% are traced back to problems in the sample preparation
NMR spectroscopy in pharmaceutical analysis, Editors U. Holzgrabe, I. Wawer, B. Diehl, 2008
Innovation with Integrity
that uses External Standard Method
Concentration Determination
Innovation with Integrity
Rapid and Automated Workflow
1H Experiment (with or without solvent suppression) : automatic processing
Result is the concentration:
• Submission in IconNMR
• Results: pdf
The external standard methodology is based on PULCON (Wider G, Dreier L. J. Am. Chem. Soc. 2006; 128: 2571.)
Innovation with Integrity
Today´s Bruker Automated Solution
SampleJet: 5 x 96 samples
SamplePro Tube Liquid handler
Innovation with Integrity
High Throughput Workflow
• Medchem compounds in NMR solvents (deuterated or not)
• E.g. for FBS ligands in well plates reference & QC
Samples
• Any Bruker sample changer SampleJet for high throughput
• Optimized parametersets
• Customization easy
Automation
• Fully automated, on-the-fly, analysis:
• Concentration, purity, ID, water
Analysis
• All information at a glance
• Easy manual cross check and editing
• Spread sheets, pdfs…
Results
Innovation with Integrity
Potency Determination
that uses Internal Standard Method
Sample 20 mg 18 mg API, 2 mg of anything else , detectable or not by NMR Potency = 90 %
Innovation with Integrity
Popularity of Potency Determination
Gaining popularity and adoption:
• Introduction Pharmacopeia
• Used in industry (Pharma, food, chem.)
• Deployment to open access (ease of use)
Planta Med 2013; 79 - SL79 DOI: 10.1055/s-0033-1351904
Innovation with Integrity
1.0208 a.u.
4.6817 a.u.
NMR: a Primary Ratio Method
8.56 mg 6.22 mg
1
1
3
Analyte Internal Standard
MW: 168.19 g/mol MW: 356.41 g/mol
I can quantify one compound with respect to another compound without a response factor!
2
𝑃𝑃𝑎𝑎 = 𝑃𝑃𝐼𝐼𝐼𝐼 ×𝐴𝐴𝑎𝑎𝐴𝐴𝐼𝐼𝐼𝐼
×𝑁𝑁𝐼𝐼𝐼𝐼𝑁𝑁𝑎𝑎
×𝑊𝑊𝑊𝑊𝐼𝐼𝐼𝐼𝑊𝑊𝑊𝑊𝑎𝑎
×𝑀𝑀𝑊𝑊𝑎𝑎𝑀𝑀𝑊𝑊IS
Sample prep. Known compounds NMR From supplier
Innovation with Integrity
NMR `One-Stop Shop´
potency determination purity assessment relative response factor calculation residual solvent moisture analysis identity testing
‘Potency determination by qNMR has been shown to be a single point replacement for routine development testing which previously involved several experiments and techniques.’*
1 single experiment -> qNMR
Efficiency, Economy Selectivity
* Webster G.F. and Kumar S., Anal. Chem., 86, 11474 (2014)
Innovation with Integrity
Fully Automated Potency Determination
Acquisition + Analysis + Reporting
Innovation with Integrity
Setup of Potency Determination in Icon
Innovation with Integrity
Results: PDF Report
Robustness and reproducibility Internal standard peaks ID and integration Analyte ID, check and quantification Potency calculation
Consistency check is done as an indicative statement: potency determination is independent of the consistency check.
Innovation with Integrity
Results in Excel
Quality Duplicates, triplicates … Error analysis – Intra and Inter sample Easy review
What if there would be an error in weighing ? inter sample RSD:
Triplicate measurement:
What if there would be an error in analyte integral (overlap with impurities, satellites) ? intra and inter sample RSD:
Innovation with Integrity
Reproducibility at 2 Magnetic Fields Sulindac API with 1,3,5-trimethoxybenzene as IS
500MHz 300MHz
Innovation with Integrity
Example of Drug Product
The potency of the formulated ibuprofen is: Potency = 400 𝑚𝑚𝑚𝑚
533 𝑚𝑚𝑚𝑚 = 75.0%
PotencyMR (300MHz) fully automated
The quantification is robust because ibuprofen aromatic integrals are separated from all excipients signals “Separation without physical separation“
Correct result in full automation:
The tablet weighs 533 mg Sample is a clear solution in dDMSO after centrifugation
Innovation with Integrity
Manual Interaction at any Time
Expert Analysis Interactively define integrals Use predefined regions
– Create and copy methods – Run methods in automation
• Make “methods” for re-occurring tasks • Set molecular mass • Define integral regions
• Use these in automation • Auto-report with your definitions
Innovation with Integrity
PotencyMR Software Summary
Problems it solves: streamlines potency/purity determination, which can now be run in automation by both NMR experts and non-experts.
• Easy sample submission through IconNMR, including replicate samples,
acquired and analysed in automation
• Excel report generated automatically after measurement. Describes how potency is computed and gives intra and inter sample RSD
• Easy-to-use GUI enables user interaction at any point
• Automatic analysis of the 1H-spectrum: analyte quantification, consistency analysis, overlap between reference and analyte determination.
• Flexible functionality allows to run a potency determination method with pre-defined analyte's integral regions
Software is available with TopSpin acquisition license (3.5.7 and 4.0.0)
Innovation with Integrity
Tips: FILCOR in automation, DE, local baseline correction in TS
• Optimize FILCOR is very important to get good automated phase correction with apk0
• When using a He
cryoprobe, make DE as short as possible (without ring down signal). Try back prediction on analog FID
• Integral definition with
local baseline correction in TopSpin : screenshots ->
Mephaquin Drug tablet
Innovation with Integrity
Thank you! …and please talk to us for demos and discussions