user manual
Contents
Chapter 1 1 Intended Use, Contraindications and General Precautions
Chapter 2 3 Description of the Inogen One® G2 Oxygen Concentrator 3 ImportantPartsoftheInogenOne®G2OxygenConcentrator 4 UserControls 5 UserInterfaces 5 Input/OutputConnections 6 PowerSupplyOptions 9 InogenOne®G2Accessories
Chapter 3 11 Operating Instructions 11 GeneralInstructions 15 AdditionalOperatingInstructions 18 BatteryOperatingInstructions 19 BatteryCareandMaintenanceChapter 4 21 Inogen One® G2 Oxygen Concentrator Audible and Visible Signals (including Alarms)
Chapter 5 29 Troubleshooting Chapter 6 31 Cleaning, Care and Maintenance 31 CannulaReplacement 31 CaseCleaning 32 FilterCleaningandReplacement 34 OtherServiceandMaintenance 34 Disposalofequipmentandaccessories
Chapter 7 35 Symbols Used on Concentrator and Accessories
Chapter 8 37 Inogen One® G2 System Specifications
English
1 Intended Use, Contraindications and General Precautions
IntendedUse,ContraindicationsandGeneralPrecautionsChapter1 1
Intended Use
TheInogenOne®G2OxygenConcentratorisusedonaprescriptivebasisbypatientsrequiringsupplementaloxygen.Itsuppliesahighconcentrationofoxygenandisusedwithanasalcannulatochanneloxygenfromtheconcentratortothepatient.TheInogenOne®G2maybeusedinhome,institution,vehicle,onanairplaneandvariousmobileenvironments.
CAUTION USAFederallawrestrictsthisdevicetosalebyorontheorderofaphysician.Mayalsobeapplicableinothercountries.
CAUTION Useofacannulaotherthanahighflowcannula(e.g.Salter1600Q)mayrestrictoxygendeliveryand/orattachmenttonozzlefittings.
wArNING Availabilityofanalternatesourceofoxygenisrecommendedincaseofpoweroutageormechanicalfailure.Consultyourequipmentproviderfortypeofback-upsystemrecommended.
CAUTION Itistheresponsibilityofthepatienttomakeback-uparrangementsforalternativeoxygensupplywhentraveling;Inogenassumesnoliabilityforpersonschoosingnottoadheretomanufacturerrecommendations.
! C
English
2
Contraindications
wArNING ThisdeviceisNOTINTENDEDtobelifesustainingorlifesupporting. CAUTION Undercertaincircumstances,theuseofnon-prescribedoxygen
therapycanbehazardous.Thisdeviceshouldbeusedonlywhenprescribedbyaphysician.
CAUTION Additionalmonitoringorattentionmayberequiredforpatientsusingthisdevicewhoareunabletohearorseealarmsorcommunicatediscomfort.Ifthepatientshowsanysignsofdiscomfort,aphysicianshouldbeconsultedimmediately.
CAUTION TheInogenOne®G2isnotdesignedorspecifiedtobeusedinconjunctionwithahumidifierornebulizer.Useofthisdevicewithahumidifierornebulizermayimpairperformanceand/ordamagetheequipment.
! C
! C
! C
General Precautions
wArNING Thedeviceproducesenrichedoxygengaswhichacceleratescombustion.DoNOTALLOWSMOKINGOROPENFLAMESwithin10feetofthisdevicewhileinuse.
wArNING DonotsubmersetheInogenOne®G2oranyoftheaccessoriesinliquid.Donotexposetowaterorprecipitation.Donotoperateinexposedrain.Thiscouldleadtoelectricalshockand/ordamage.
CAUTION Donotuseoil,grease,orpetroleum-basedproductsonornearthe
InogenOne®G2.
CAUTION NeverleavetheInogenOne®G2inanenvironmentwhichcanreachhightemperatures,suchasanunoccupiedcarinhightemperatureenvironments.Thiscoulddamagethedevice.
Important Parts of the Inogen One® G2 Oxygen Concentrator
Description of the Inogen One® G2 Oxygen Concentrator
DescriptionoftheInogenOne®G2OxygenConcentratorChapter2 3
2
On/Off Button
Alert/Alarm Light
Flow ControlDisplay
DisplayBacklight Button
Breath Detection Light
Audible Alarm Button
English
4
User Controls
ON / OFF ButtonPressoncetoturn“ON”;Pressandholdforonesecondtoturn“OFF”.
Audible Alarm Button PressingthisbuttonwilltoggletheInogenOne®G2’sbreathdetectionalertonandoff:
1. DefaultMode.WhentheInogenOne®G2ispoweredup,thebreathdetectionalertisdisabled.Thedisplay’smodeindicationareaisemptyinDefaultMode.
2. BreathDetectionAlertMode.TheInogenOne®G2willalertwithaudibleandvisualsignalsfor“nobreathdetected”whenthismodeisenabledandnobreathhasbeendetectedfor60seconds.Thedisplay’smodeindicationareawillshowabelliconwhenthealertisenabled.
Flow Setting Control Buttons Usethe–or+flowsettingcontrolbuttonstoselectthesettingasshownonthedisplay.Therearefivesettings,from1to5.
Display Backlight ButtonPushon;automaticallyturnsoffafter10seconds.
DescriptionoftheInogenOne®G2OxygenConcentratorChapter2 5
User Interfaces
DisplayThisscreendisplaysinformationregardingflowsetting,powerstatus,batterylifeanderrors.IfyouwouldliketochangethelanguageontheInogenLCDScreencontactInogenClientServices.
Indicator LightsAredlightindicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alarm).Aflashinglightishigherprioritythannon-flashing.Audible SignalsAnaudiblesignal(beep)indicateseitherachangeinoperatingstatusoraconditionthatmayneedresponse(alarm).Morefrequentbeepsindicatehigherpriorityconditions.
Input / Output Connections
Particle FilterThefiltermustbeinplaceattheintakeendoftheconcentratorduringoperationtokeepinputairclean.
Cannula Nozzle FittingThenasalcannulaconnectstothisnozzleforInogenOne®G2outputofoxygenatedair.
DC Power InConnectionforexternalpowerfromtheUniversalPowerSupply.
English
6
Power Supply Options
Single and Double rechargeable Lithium Ion BatteriesThebatterywillpowertheInogenOne®G2withoutconnectiontoanexternalpowersource.Whenfullycharged,asinglebatterywillprovide2to4hoursofoperation;adoublebatterywillprovide4to8hoursofoperation.ThebatteryrechargeswhenproperlyinstalledintheInogenOne®G2andtheconcentratorisconnectedtoACorDCpower.Rechargingtimeisupto4hoursforasinglebatteryandupto8hoursforadoublebattery.Seeinformationinthe“BatteryCareandMaintenance”section.
UNIVErSAL POwEr SUPPLyOverviewTheInogenUniversalPowerSupply(BA-107/207)isusedtopowertheInogenOne®G2concentratorfromanACpowersourceoraDCpowersource.Thisallowstheusertopowertheconcentratorinthehome,inavehicleorotherplaceswhereACorDCpowerisavailable.
DescriptionTheInogenOne®G2UniversalPowerSupplyisspecificallydesignedforusewiththeInogenOne®G2OxygenConcentrator(IO-200).TheUniversalPowerSupplyprovidestheprecisecurrentandvoltagerequiredtosafelypowertheInogenOne®G2andisdesignedtooperatefromspecifiedACandDCpowersources.WhenusedwithACpowersources,thepowersupplyautomaticallyadaptstoinputvoltagesfrom100Vto240V(50-60HZ)permittingusewithmostpowersourcesthroughouttheworld.
TheUniversalPowerSupplywillchargetheInogenOne®G2BatterywhenusedwithACinputpowerorwhenusedwithaDCpowersupply,suchastheonefoundinyourcar.Duetoaircraftpowerlimitations,theUniversalPowerSupplycannotbeusedtochargetheInogenOne®G2Batterywhenusedonanaircraft.TheUniversalPowerSupplyisusedwiththefollowingcomponents:
•PowerSupplywithattachedpoweroutputcabletoconnecttotheInogenOne®G2•AutomobileDCcigarettelighterPowerInputCable•ACPowerInputCable
7DescriptionoftheInogenOne®G2OxygenConcentratorChapter2
Power Output Plug
Universal Power Supply
LED Indicator Light
AC Input
DC cigarette lighter Power Plug for use in Automobile/RV/Boat/Aircraft
DC Input
AC Power Plug*
wArNING DonotwrapcordsaroundPowerSupplyforstorage.Doingsomayleadtodamagedcordsandafailuretoprovidepowertotheconcentrator.
Model#BA-107
! C
*Actualproductappearancemayvary.
English
8
Power Output Plug
Universal Power Supply
LED Indicator Light
AC Input
DC cigarette lighter Power Plug for use in Automobile/RV/Boat/Aircraft
DC Input
AC Power Plug*
Model#BA-207
wArNING DonotwrapcordsaroundPowerSupplyforstorage.Doingsomayleadtodamagedcordsandafailuretoprovidepowertotheconcentrator.! C
*Actualproductappearancemayvary.
Inogen One® G2 Accessories
Nasal CannulaAnasalcannulamustbeusedwiththeInogenOne®G2toprovideoxygenfromtheconcentrator.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.
NOTE Increasingthecannulalengthmayreducetheperceivednoiseduringoxygenbolusdelivery.
Carry BagTheCarryBagprovidesaprotectivecoverwithahandleandshoulderstraptoenableyoutocarrytheInogenOne®G2.TheInogenOne®G2canbeoperatedusingbatterypowerduringtransportwiththeCarryBag.
CartTheCarthaswheelsandatelescopinghandletoprovideeasytransportoftheInogenOne®G2.TheInogenOne®G2canbeoperatedusingbatterypowerduringtransport.PlacetheCarryBagovertheCart.MakesuretheCarthandleisinsertedbetweentheelasticstrapinthebackofthecarrybagandthefrontofthecarrybag.
DescriptionoftheInogenOne®G2OxygenConcentratorChapter2 9
English
Inogen One® G2 Optional Accessory
External Battery Charger1. PlugtheExternalBatteryChargerACpowersupply
cordintoanelectricaloutlet.2. PlugtheExternalBatteryChargerACpowersupplyinto
thebatterycharger.3. SlideyourchargerontotheInogenOneG2Batteryby
clickingandlockingintothecharger.4. Whenthebatteryisinthecorrectposition,asolidredlight
willindicatethatthebatteryischarging.5. Whenthegreenlightilluminates,thebatteryis
fullycharged.
CAUTION AvoidtouchingtherecessedelectricalcontactsoftheExternalBatteryCharger;damagetocontactsmayaffectchargeroperation.
NOTE Thesecontactsarenotpoweredunlessabatteryisinplaceandcharging.
NOTE TocompletelyremovepowerfromtheExternalBatteryCharger,removetheplug.
Inogen One G2 BackpackAlternative/optionalwayofcarryingyourInogenOneG2,handsfree,morecomfort,outofyourwaywithextrapocketsforadditionalaccessories.
ToorderpleasecallInogenClientServices.
10
Operating Instructions
OperatingInstructionsChapter3 11
3
! C
Intake
Exhaust
General Instructions
1. Place the Inogen One® G2 in a well ventilated location.Airintakeandexhaustmusthaveclearaccess.LocatetheInogenOne®G2insuchawaythatanyauditoryalarmsmaybeheard.
wArNING AvoiduseoftheInogenOne®G2inpresenceofpollutants,smokeorfumes.DonotusetheInogenOne®G2inpresenceofflammableanesthetics,cleaningagentsorotherchemicalvapors.
CAUTION Donotobstructairintakeorexhaustwhenoperatingthedevice.
Blockageofaircirculationorproximitytoaheatsourcemayleadtointernalheatbuildupandshutdownordamagetotheconcentrator.
2. Ensure the particle filter is in place.
CAUTION DonotoperatetheInogenOne®G2withouttheparticlefilterinplace.Particlesdrawnintothesystemmaydamagetheequipment.
3. Install the battery. InserttheInogenOne®G2batterybyslidingbatteryintoplaceandlockingatbottomofconcentrator.
English
12
CAUTION TheInogenOne®G2batteryactsasasecondarypowersupplyintheeventofaplannedorunexpectedlossoftheACorDCexternalpowersupply.WhenoperatingtheInogenOne®G2fromanACorDCexternalpowersupply,aproperlyinsertedInogenOne®G2batteryshouldbemaintainedintheunit.Thisprocedurewillensureuninterruptedoperationandwilloperateallalarmsandalertsintheeventofalossoftheexternalpowersupply.
NOTE TheInogenOne®G2BatteryrequiresaninitialfullchargeofuninterruptedchargingfromanemptystateusingtheInogenOne®G2ontheACpowersource.DonotruntheInogenOne®G2onbatterypoweruntilthisinitialcharginghasbeencompleted.Oncetheinitialchargeiscompleted,thebatterymaybeusedwithanyamountofchargelevel.
4. Connect the Universal Power Supply. ConnecttheACPowerplugtotheUniversalPowerSupply.ConnecttheACPowerPlugtothepowersourceandconnectthePowerOutputPlugtotheInogenOne®G2.ThegreenLEDontheUniversalPowerSupplywillbeilluminatedandabeepwillsoundfromtheconcentrator.
CAUTION EnsuretheUniversalPowerSupplyisinawellventilatedlocationasitrelies onaircirculationforheatdissipation.TheUniversalPowerSupplymaybecome hotduringoperation.MakesuretheUniversalPowerSupplycoolsdownbefore handling.
CAUTION TheUniversalPowerSupplyisnotwaterresistant.
CAUTION DonotdisassembletheUniversalPowerSupply.Thismayleadtocomponentfailureand/orsafetyrisk.
CAUTION DonotplaceanythingintheUniversalPowerSupplyportotherthanthesuppliedwallcord.AvoidtheuseofelectricalextensioncordswiththeInogenOne®G2.Ifanextensioncordmustbeused,useanextensioncordthathasanUnderwritersLaboratory(UL)Markandaminimumwirethicknessof18gauge.Donotconnectanyotherdevicestothesameextensioncord.
OperatingInstructionsChapter3 13
NOTE Undercertainconditions(seeTechnicalSpecifications)theUniversalPowerSupplymayshutdown.ThegreenLEDwillblinkorwillnolongerbeilluminated.Ifthisoccurs,disconnectthePowerSupplyforatleast10secondsandreconnect.
NOTE WhentheUniversalPowerSupplyisdisconnectedfromtheACoutlet,alsodisconnectitfromtheconcentratortoavoidunnecessarybatterydischarge.
5. Connect the nasal cannula tubing to the nozzle fitting.NozzlefittingislocatednexttothehandleoftheInogenOne®G2.Useofasinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.Additionaltitrationmaybeneededtoensureproperoxygendeliverywhenusingaparticularcannula.
CAUTION Toensureoxygenflow,ensurethatthenasalcannulaisproperlyconnectedtothenozzlefittingandthatthetubingisnotkinkedorpinchedinanyway.
CAUTION Replacethenasalcannulaonaregularbasis.Checkwithyourequipmentproviderorphysiciantodeterminehowoftenthecannulashouldbereplaced.
6. Turn on your Inogen One® G2 by pressing the ON/OFF Button. AsingleshortbeepwillsoundaftertheInogenlogoisdisplayed.“PleaseWait”willappearwhiletheconcentratorstartsup.Thedisplaywillindicatetheselectedflowsettingandpowercondition.Followingabriefstart-upsequence,awarmupperiodupto2minuteswillinitiate.Duringthistimeperiodtheoxygenconcentrationisbuildingtobutmaynothavereachedspecification.AdditionalwarmuptimemaybeneededifyourInogenOne®G2hasbeenstoredinextremelycoldtemperatures.
English
! C
14
7. Set the Inogen One® G2 Concentrator to the flow rate prescribed by your physician orclinician.Usethe+or–settingbuttonstoadjusttheInogenOne®G2tothedesiredsetting.Thecurrentsettingcanbeviewedonthedisplay.
8. Position the nasal cannula on your face and breathe through your nose. TheInogenOne®G2willsensetheonsetofinhalationanddeliveraburstofoxygenataprecisetimewhenyouinhale.TheInogenOne®G2willsenseeachbreathandcontinuetodeliveroxygeninthismanner.Asyourbreathingratechanges,theInogenOne®G2willsensethesechangesanddeliveroxygenonlyasyouneedit.Attimes,ifyouinhaleveryquicklybetweenbreathstheInogenOne®G2mayignoreoneofthebreaths,givingtheappearanceofamissedbreath.ThismaybenormalastheInogenOne®G2sensesandmonitorsthechangesinyourbreathingpattern.TheInogenOne®G2willnormallysensethenextbreathanddeliveroxygenaccordingly.
Agreenlightwillflasheachtimeabreathisdetected.Makecertainthatthenasalcannulaisproperlyalignedonyourfaceandyouarebreathingthroughyournose.
wArNING Ifyoubegintofeelillorareexperiencingdiscomfortwhileusingthisdevice,consultyourphysicianimmediately.
CAUTION TheInogenOne®G2isdesignedtoprovideaflowofhighpurityoxygen.
Anadvisoryalarm,“OxygenLow”,willinformyouifoxygenconcentrationdrops.Ifalarmpersists,contactyourequipmentprovider.
GeneralToremovepower,unplugtheinputcordfromitssource(i.e.,ACwalloutlet,DCcarcigarettelighteradapter).
CAUTION EnsuretheUniversalPowerSupplyispoweredfromonlyonepowersource(ACorDC)atanygiventime.TheUniversalPowerSupplymaynotoperateproperlyifsimultaneouslypoweredusingbothACandDCsources.
OperatingInstructionsChapter3 15
Additional Operating InstructionsFor Use In Home with AC PowerForoperationofthePowerSupplyusinganACpowersource,followtheseinstructions:
1. ConnecttheACInputtothepowersupply.2. ConnecttheACPowerPlugtothepowersourceandconnectthePowerOutput
PlugtotheInogenOne®G2.ThegreenLEDwillbeilluminated,indicatingtheUniversalPowerSupplyhasinputpower.
Traveling with your Inogen One® G2 System
TheInogenOne®G2Systemmakestravelbyplane,boat,carortrainmoreconvenientforoxygenusersthaneverbefore.Nowyougetthesamequalityperformanceandconveniencewhileonthegothatyou’reusedtoreceivingfromyourInogenOne®G2athome.HerearesomeusefulandimportantinstructionsformaximizingperformanceandconveniencewhenusingyourInogenOne®G2whiletraveling.
Youshouldbeginplanningforatripwithachecklistofitemstoremember.Thischecklistshouldinclude:
✓ UniversalPowerSupply
✓ ExtraBattery(s)ifrequired
✓ Importantphonenumbers,suchasthoseofyourphysicianandhomehealthcare provider,orprovidersintheareayouwillbetraveling
✓ Planforbackupoxygenintheeventofaprolongedpoweroutageormechanicalfailure.
For Use In Automobile/rV/BoatForoperationoftheUniversalPowerSupplyusingaDCpowersource,followtheseinstructions:
1. ConnecttheDCInputtothepowersupply.2. ConnecttheDCPowerPlug(CigaretteLighterAdapter)to
thepowersourceandconnectthePowerOutputPlugtotheInogenOne®G2.ThegreenLEDwillbeilluminated,indicatingthepowersupplyhasinputpower.
English
3.Theplugshouldinsertintothesocketwithoutexcessiveforceandsecurelyremaininplace.TheCigaretteLighterAdaptercomeswithaslideswitchpre-adjustedtothenarrowposition.Thiswillfitmostautomobilecigarettelightersockets.IftheCigaretteLighterAdapterislooseinthesocket,adjusttheswitchtothe"wide"position.
WARNING EnsurethattheautomobilepowersocketisadequatelyfusedfortheInogenOne®G2powerrequirements(minimum15Amp).Ifthepowersocketcannotsupporta15Ampload,thefusemaybloworthesocketmaybedamaged.
WARNING ThetipoftheCigaretteAdapterPlugbecomesHOTwheninuse.Donottouchthetipimmediatelyafterremovalfromanautocigarettelightersocket.
CAUTION Ensuretheautomobilepowersocketiscleanofcigaretteashandtheadapterplugfitsproperly,otherwiseoverheatingmayoccur.
CAUTION DonotusetheUniversalPowerSupplywithacigaretteplugsplitterorwithanextensioncable.ThismaycauseoverheatingoftheDCpowerinputcable.
CAUTION DonotjumpstarttheautomobilewiththeUniversalPowerSupplyconnected.Thismayleadtovoltagespikeswhichcouldshutdownand/ordamagetheUniversalPowerSupply.
CAUTION WhenpoweringtheInogenOne®G2inanautomobileensuretheengineisrunning.Operatingthedevicewithouttheenginerunningmaydrainthecarbattery.
Traveling By AirTheFAAallowstheInogenOne®G2onboardallU.S.aircraft,hereareafewpointstomakeairtraveleasy.
Planning your FlightWhenflyingwiththeInogenOne®G2,youmustinformtheairlinethatyouwillbeusingyourInogenOne®G2onboardtheaircraft.Youmustalsokeepasignedstatementfromyourphysicianwithyouthatincludes:
•Yourabilitytosee/hearalarmsandappropriatelyrespond.•Whenoxygenuseisnecessary(alloraportionofthetrip).•Maximumflowratecorrespondingtothepressureinthecabinundernormaloperatingconditions.16
! C
! C
•Anewstatementwillnotbenecessaryeachtimeyoufly,butthestatementshouldbeavailableduringeveryflight.
•Someairlinesmayequiptheiraircraftwithonboardelectricalpower.YoumayhaveanopportunitytorequestaseatwithapowerportwhichcanbeusedtopoweryourInogenOne®G2.However,availabilityvariesbyairline,typeofaircraftandclassofservice.Youshouldcheckwithyourairlinesforavailabilityandalwaysplanonhavingsufficientbatterypowerforthedurationofyourflight,plusaconservativeestimateofunanticipateddelays.
•Youruniversalpowersupplyisequippedwithacommonlyusedcigarettelighteradapter.However,aircraftsusedifferentpowerportconfigurationsanditisdifficulttodeterminewhichtypeofpowerportcompatibilityyouraircraftmaysupply;it’sagoodideatopurchaseanadaptersuchasMagellan’sEmPowerSocketConnectorItem#EA270.ContactMagellan’sat800-962-4943orvisitwww.magellans.comforastorelocationnearyouorpurchaseop-tionsonline.
Before your FlightHerearesomethingstokeepinmindthedayyourflightdeparts:
•EnsureyourInogenOne®G2isclean,ingoodconditionandfreefromdamageorothersignsofexcessivewearorabuse.
•BringenoughchargedbatterieswithyoutopoweryourInogenOne®G2forthedurationoftheflightplusaconservativeestimateofunanticipateddelays.
•Regional/Commuterairlinesdonotofferonboardelectricalpower.Ifyourtravelplanscallforflightsonregionalairlines,youwillneedsufficientbatterypowerforthedurationoftheflight,plusaconservativeestimateofunanticipateddelays.
CAUTION Airlinesmaynotbeequippedtoprovidebackupoxygen.
•Arriveattheairportearly.AirportsecurityscreeningpersonnelmayrequireextratimetoinspectyourInogenOne®G2.
•Whilewaitingtoboardyourflight,youmaybeabletoconservebatterypowerbyusingtheUniversalPowerSupplytopoweryourInogenOne®G2fromanelectricaloutletintheairportterminalifavailable.
•YoushouldinformtheairlineyouwillbeusingyourInogenOne®G2.Haveyourphysicianletterwithyouandreadyforinspectionifrequested.
OperatingInstructionsChapter3 17
English
During your Flight1. Ifusingairlinepowerport,removetheBatteryfromtheInogenOne®G2Oxygen
Concentrator.Duetoaircraftpowerlimitations,theUniversalPowerSupplycannotbeusedtochargetheInogenOne®G2Batterywhenonboardaircraft.
2. ConnecttheDCPowerPlugforavailableairlinepower.Checkwithairlinepersonneltoensurecompatibility.
•Duringtaxi,takeoffandlanding,stowyourInogenOne®G2undertheseatinfrontofyou.YourInogenOne®G2willfituprightundermostairlineseats.However,ifitdoesn’tfityoumayturnitonitsside.
•ItisnotnecessarytoturnoffyourInogenOne®G2duringtaxi,takeoffandlandingifyourphysician’swrittenstatementrequiresyoureceiveoxygenduringtheseperiods.
CAUTION Achangeinaltitude(forexample,fromsealeveltomountains)mayaffecttotaloxygenavailabletothepatient.TheInogenOne®G2hasbeenverifiedtoprovideoxygentospecificationupto10,000ft(3048m).Consultyourphysicianbeforetravelingtohigherorloweraltitudestodetermineifyourflowsettingsshouldbechanged.
After your Flight•Remembertorechargeadditionalbatteriesyoumayhaveusedpriortoyournextflight.
Traveling By Bus, Train or BoatContactyourcarriertofindoutaboutpowerportavailability.
Battery Operating InstructionsEnsurethatthebatteryisinplaceandcharged.DisconnecttheInogenOne®G2fromitspowersource.WhiletheInogenOne®G2isoperatingonbatterypower,thebatterywilldischarge.Thedisplaywillindicatetheestimatedremainingpercentage(%)orminutesofuse.
Whentheconcentratordetectsthatthebatterylifeisbelow10%remaining,alowpriorityalertwillsound.Whenthebatteryisempty,thealertwillchangetoahighpriority.
Whenbatterylifeislow,dooneofthefollowing:•PlugtheInogenOne®G2intoanACorDCpowersourceusingtheUniversalPowerSupply.•ReplacethebatterywithachargedbatteryafterturningofftheInogenOne®G2(bypressingtheON/OFFbutton).•Ifthebatteryisdrained,chargethebatteryorremoveitfromtheconcentrator.18
OperatingInstructionsChapter3 19
IftheInogenOne®G2isbeingpoweredbytheUniversalPowerSupply,batterieswillchargeduringoperation.LeavingyourInogenOne®G2pluggedinpastthefullchargetimewillnotharmtheconcentratororthebattery.
Normal Battery ChargingToensurethatyourbatteryisproperlycharging,inspectthatthecorrectACandDCpoweroutputplugadapterisbeingusedandthattheadapterisproperlyinsertedintothepoweroutlet.Observethedisplayorlightsthatindicatechargingstatus.
NOTE Whenstartingtochargeafullydischargedbattery,thechargingprocessmay startandstopduringthefirstfewminutes.
Battery Care and MaintenanceYourInogenOne®G2LithiumIonBatteryrequiresspecialcaretoensureproperperformanceandlonglife.UseonlyInogenOne®G2BatterieswithyourInogenOne®G2Concentrator.
Keep DryAlwayskeepliquidsawayfrombatteries.Ifbatteriesbecomewet,discontinueuseimmediatelyanddisposeofbatteryproperly.
Effect of Temperature on Battery PerformanceTheInogenOne®G2singlebatterypowerstheInogenOne®G2Concentratorfrom2to4hoursundermostenvironmentalconditions.Toextendtherun-timeofyourbattery,avoidrunningintemperatureslessthan41˚F(5˚C)orhigherthan95˚F(35˚C)forextendedperiodsoftime.
Battery Time remaining ClockTheInogenOne®G2continuouslydisplaysbatterytimeremaining.Thisdisplayedtimeisonlyanestimateandtheactualtimeremainingmayvaryfromthisvalue.
Please Follow These Important Guidelines to Maximize Battery Performance and Life:•Storebatteryinacool,dryplace.Storewithachargeof40-50%.•Ifusingmultiplebatteries,makesurethateachbatteryislabeled(1,2,3orA,B,C,etc.)androtateonaregularbasis.Batteriesshouldnotbeleftdormantforextendedperiodsoftime.
English
Mode
PowerText
Display Icons
TheInogenOne®G2displayisdividedintothreeareas.Theupperleftcornerofthedisplayshowsthebreathdetectionalertstatus.Thelowerleftcornerindicatespowersourceandbatterychargelevel.Therightsideofthedisplaycontainstextinformation,suchasflowsetting,batterytimeremaininganderrornotifications.
Power Status Icons
Theseiconsareexamplesofthoseshowninthedisplay’spowerstatuswindowwhentheInogenOne®G2isoperatingonbatterypower.
Inogen One® G2 Oxygen Concentrator Audible and Visible Signals
AudibleandVisibleSignalsChapter4
! C
! C
! C
! C
Meaning
Batteryisempty.
Batteryhaslessthan10%chargeremaining.Thisiconflashes.
Batteryhasapproximately40%to50%chargeremaining.
Batteryisfull.
Icon
21
4English
Power Status Icons (continued)TheiconsbelowareexamplesofthoseshownwhentheInogenOne®G2isoperatingfromanexternalpowersupplyandchargingthebattery.Thelighteningboltindicatesthatanexternalpowersupplyisconnected.
! C
! C
! CMeaning
Batteryischargingwithchargelevelbetween60%and70%.
Thebatteryisfullychargedandischargingasnecessarytomaintainitscharge.
Batteryischargingwithchargelevellessthan10%.
TheInogenOne®G2isoperatingfromanexternalpowersourcewithnobatterypresent.
Icon
22
Mode IconsThesearetheiconsshowninthedisplay’smodewindow.
Display Text
NOTE Whentwoconditionsoccursimultaneously,theconditionwiththehigherprioritywillbedisplayed.
Meaning
Thebreathdetectionalarmhasbeenenabled.
Theconcentratorisinthestandardbreathdetectmode,andthebreathdetectionalertisdisabled.Thisisthedefaultcondition.
Icon
! C
[blank]
! C
AudibleandVisibleSignalsChapter4 23
Informational MessagesThefollowinginformationdisplaysarenotaccompaniedbyanyaudiblefeedbackoranyvisualchangeintheindicatorlights.
Condition/Action/Explanation
TheInogenlogoisdisplayedatstartup.
Defaultdisplaywhenoperatingonbatterypower.“X”representstheselectedflowsetting(e.g.,Setting2).“HH:MM”representstheapproximatetimeremainingonthebatterycharge(e.g.,1:45).
Defaultdisplaywhenoperatingonanexternalpowersupplyandthebatteryischarging.“xx%”representsthepercentbatterycharge(e.g.,86%).
Defaultdisplaywhenthebatteryisnotchargingorwhenthetimeremainingisnotavailablefromthebattery.
Displaywhentheconcentratorispluggedinandbeingusedtochargeabattery(notbeingusedforoxygenproduction).Itisnormaltoseeafullychargedbatteryreadbetween95%and100%whenexternalpowerisremoved.Thisfeaturemaximizestheusefullifeofthebattery.
Message Display & Text
SettingXBatteryHH:MM
SettingXChargingxx(or)BatteryFull
SettingXBatteryxx%
Chargingxx%(0r) BatteryFull
NotificationsThefollowingnotificationmessagesareaccompaniedbyasingle, short beep.
Condition/Action/Explanation
On/Offbuttonhasbeenpressedfortwoseconds.Concentratorisperformingsystemshutdown.
AudibleAlarmhasbeenpressedforfiveseconds.
Message Display & Text
InogenOne® ShuttingDown
HH:MM Vx.x:SerialNumber
English
Low Priority AlertsThefollowinglowpriorityalertmessagesareaccompaniedbyadouble beepandasolid red light.
Condition/Action/Explanation
Externalpowersupplyistoolowtoruntheunit,whichisrunningonbatterypowerdespitebeingpluggedin.Checkexternalpowersupplyconnections.Ifconditionpersists,contactyourequipmentprovider.
Externalpowersupplyistoohigh.Checkexternalpowersupplyconnections.Ifconditionpersists,contactyourequipmentprovider.
Batterypowerislow,withlessthan6%remaining.Attachexternalpowersupplyorpowerdownandinsertafullychargedbattery.
Batteryerrorhasoccurred.Switchtoanewbatteryorremovebatteryandoperateconcentratorusingexternalpowersupply.Ifbatteryerrorrecurswithsamebattery,stopusingthebatteryandcontactyourequipmentprovider.
Concentratorisproducingoxygenataslightlylowlevel(<82%).Ifconditionpersists,contactyourequipmentprovider.
Message Display & Text
External PowerLow
External PowerHigh
BatteryLow AttachPlug
BatteryError SeeManual
OxygenLow SeeManual
24
Low Priority Alerts (continued)
Condition/Action/Explanation
Batteryhasexceededitschargingtemperatureandcharginghasstopped.Thebatterywillnotchargewhilethisalertispresentbutwillbegintochargewhenthebatterytemperaturereturnstothenormaloperatingrange.Ifbatterychargingisdesiredsooner,removethebatteryfromtheconcentratorandallowittocoolinanopenareaforapproximately10-15minutes.Then,re-insertthebatteryintotheInogenOne®G2.Iftheproblemstillpersists,contactyourequipmentprovider.
Concentratorisproducingoxygenbutcannotreportbatterystatus.Replacebattery.Ifconditionpersists,contactyourequipmentprovider.
Theconcentratorrequiresservicingattheearliestconvenience.Theconcentratorisoperatingtospecificationandmaycontinuetobeused.Contactyourequipmentprovidertoarrangeforservice.
Theconcentrator’soxygensensorhasmalfunctioned.Youmaycontinuetousetheconcentrator.Iftheconditionpersists,contactyourequipmentprovider.
Message Display & Text
RemoveBatterytoCool
CommError SeeManual
ServiceNeeded
02SensorFailSeeManual
AudibleandVisibleSignalsChapter4 25
English
Condition/Action/Explanation
Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Ifpossible,moveconcentratortoacoolerlocationorpowerunitwithanexternalpowersupplyandremovebattery.Ifconditionpersists,contactyourequipmentprovider.
Concentratorisnotdetectingabreath.•Checkthatcannulaisconnectedtoconcentrator,therearenokinksintubingandcannulaispositionedproperlyinyournose.
Concentratorisexperiencinganerrorbutiscapableofcontinuedoperation.
•Removeandre-insertbatteryand/or•Checkexternalpowersupplyconnections.
Ifconditionpersists,contactyourequipmentprovider.
Oxygenoutputconcentrationisbelow50%.Ifconditionpersists,switchtoyourbackupoxygensourceandcontactyourequipmentprovidertoarrangeforservice.
Message Display & Text
BatteryHOT Warning
NoBreathDetect CheckCannula
SystemError SeeManual
OxygenErrorServiceNeeded
Medium Priority AlertsThefollowingmediumpriorityalertmessagesareaccompaniedbyatriple beep,repeatedevery25seconds,andaflashing red light.
26
Condition/Action/Explanation
Concentratorhasinsufficientbatterypowertoproduceoxygen.Attachexternalpowersupplyorexchangebattery,thenrestartunitifnecessarybypressingOn/Offbutton.
Batteryhasexceededtemperaturelimitwhileconcentratorisrunningonbatterypower.Concentratorhasstoppedproducingoxygen.Ifpossible,moveconcentratortoacoolerlocation,thenturnpoweroffandbackon.Ensureairintakeandoutletventshaveclearaccessandparticlefilterisclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.
Concentratortemperatureistoohighandoxygenproductionisshuttingdown.Ensureairintakeandoutletventshaveclearaccessandparticlefilterisclean.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.
Message Display & Text
BatteryEmpty AttachPlug
BatteryHOT ShutDown
SystemHOT ShutDown
High Priority Alerts
CAUTION IfyouarenotneartheInogenOne®G2youmaynotbeabletohearorseethehighpriorityalerts.MakesuretheInogenOne®G2isinalocationwherethealertsandalarmswillberecognizediftheyoccur.
Thefollowinghighpriorityalertmessagesareaccompaniedbyafive beep pattern,repeatedevery10secondsandaflashing red light.
AudibleandVisibleSignalsChapter4 27
English
Condition/Action/Explanation
Thismayresultfromtheconcentratorbeingstoredinacoldenvironment(below0ºC(32ºF)).Movetoawarmerenvironmenttoallowtheunittowarmupbeforestartingit.Ifconditionpersists,switchtoabackupsourceofoxygenandcontactyourequipmentprovider.
Concentratorhasstoppedproducingoxygenandisshuttingdown.Youshould:
1. Noteerrornumber2.Switchtobackupoxygensource3.Contactyourequipmentprovider
Message Display & Text
SystemCOLD ShutDown
Error(###) ServiceNeeded
High Priority Alerts (continued)
28
Troubleshooting
TroubleshootingChapter5
Useexternalpowersupplyorreplacebatterywithonethatisfullycharged
Checkpowersupplyconnectionandverifygreenlightissolid
Contactyourequipmentprovider
PressOn/Offbuttontopowerconcentrator
Checkcannulaanditsconnectiontoconcentratornozzle
Batteryisdischargedornobatteryispresent
Powersupplyisnotconnectedproperly
Malfunction
Concentratorisnotpoweredon
Cannulaisnotconnectedproperlyoriskinkedorobstructed
ConcentratordoesnotpoweronwhenOn/Offbuttonispressed
Nooxygen
recommended Solution
RefertoChapter4
Possible Cause
RefertoChapter4
Problem
Anyproblemaccompaniedbyinformationonconcentratordisplay,indicatorlightsand/oraudiblesignals
Solutionstosomepossibleissuesyoumayencounteraredescribedinthissection.
Inogen One® G2 Oxygen Concentrator
29
5English
Cannula replacement
Yournasalcannulashouldbereplacedonaregularbasis.Consultwithyourphysicianand/orequipmentproviderand/orcannulamanufacturer’sinstructionsforreplacementinformation.Asinglelumencannulaupto25feetinlengthisrecommendedtoensureproperbreathdetectionandoxygendelivery.
CAUTION Useofacannulaotherthanahighflowcannula(e.g.Salter1600Q)mayrestrictoxygendeliveryand/orattachmenttonozzlefittings.
Case CleaningYoumaycleantheoutsidecaseusingaclothdampenedwithamildliquiddetergent(suchasDawnTM)andwater.
wArNING DonotsubmersetheInogenOne®G2oritsaccessoriesinwaterorallowwatertoenterintothecase;thismayleadtoelectricalshockand/ordamage.
wArNING DonotusecleaningagentsotherthanthosespecifiedinthisUserManual.Donotusealcohol,isopropylalcohol,ethylenechlorideorpetroleumbasedcleanersonthecasesorontheparticlefilters.
Cleaning, Care and Maintenance
31Cleaning,CareandMaintenanceChapter6
6
! C
! C
English
Filter Cleaning and replacementTheparticlefiltermustbecleanedweeklytoensuretheeaseofairflow.Removefilterfromthefrontofthedevice.Cleantheparticlefilterwithamildliquiddetergent(suchasDawnTM)andwater;rinseinwateranddrybeforereuse.
NOTE Itmaybenecessarytocleantheparticlefiltermoreoftenindustyenvironments.
TopurchaseadditionalparticlefilterscontactyourequipmentproviderorInogen.
Output FilterTheoutputfilterisintendedtoprotecttheuserfrominhalationofsmallparticlesintheproductgasflow.TheInogenOne®G2includesanoutputfilterconvenientlylocatedbehindtheremovablecannulanozzlefitting.Inogenrequiresthatthisfilterbereplacedbetweenpatients.
TheoutputfiltermaybereplacedbytheequipmentproviderorbytheownerusingtheOutputFilterReplacementKit(RP-107).
DC Input Cable Fuse replacementTheCigaretteLighterDCPowerPlugcontainsafuse.IftheDCInputCableisbeingusedwithaknowngoodpowersourceandthepowersupplyisnotoperative(nooutputpowerandthegreenLEDisnotilluminated),thefusemayneedtobereplaced.
32
FuseCigarette Adapter Plug retainer Tip retainer
CAUTION Forcontinuedprotectionagainstriskoffire,onlyusefusespecified.
33Cleaning,CareandMaintenanceChapter6
Toreplacethefuse,followtheseinstructionsandrefertothephotographbelow.1. Removethetipbyunscrewingtheretainer.Useatoolifnecessary.2. Removetheretainer,tipandfuse.3. ThespringshouldremaininsidetheCigaretteLighterAdapterhousing.Ifthespring
isremoved,makesuretoreplacethespringfirstbeforeinsertingthereplacementfuse.
4. Installareplacementfuse,InogenRP#125(BUSSMDA-12)andreassemblethetip.Ensuretheretainerringisproperlyseatedandtightened.
English
34
Disposal of Equipment and Accessories
FollowyourlocalgoverningordinancesfordisposalandrecyclingoftheInogenOne®G2andaccessories.IfWEEEregulationsapply,donotdisposeofinunsortedmunicipalwaste.WithinEurope,contacttheEUAuthorizedRepresentativefordisposalinstructions.Thebatterycontainslithiumioncellsandshouldberecycled.Thebatterymustnotbeincinerated.
Maintenance Items List• InogenOne®G2singlebattery(model#BA-200)• InogenOne®G2doublebattery(model#BA-224)• Replacementintakeparticlefilters(model#RP-200)• OutputFilterReplacementKit(model#RP-107)
Other Service and Maintenance wArNING DonotdisassembletheInogenOne®G2oranyoftheaccessories
orattemptanymaintenanceotherthantasksdescribedinthetroubleshootingsection;disassemblycreatesahazardofelectricalshockandwillvoidyourwarranty.Donotremovetamperevidentlabel.Foreventsotherthanthosedescribedinthismanual,contactyourequipmentproviderforservicingbyauthorizedpersonnel.
CAUTION DonotuselubricantsontheInogenOne®G2oritsaccessories.
! C
Symbols Used On Concentrator and Accessories
Meaning
Awarningindicatesthatthepersonalsafetyofthepatientmaybeinvolved.Disregardingawarningcouldresultinsignificantinjury.
Acautionindicatesthataprecautionorserviceproceduremustbefollowed.Disregardingacautioncouldleadtoaminorinjuryordamagetoequipment.
SeeUserManualforInstructions
U.S.FederalRegulationRestrictsthisDevicetoSalebyOrderofPhysician.MayalsobeapplicableinotherCountries.
ACPower
DCPower
NoSmokingwhiledeviceisinuse.
NoOpenFlames(Concentrator);Donotincinerate(Battery)
Symbol
WARNING
CAUTION
ONLY
ONLY
ONLY
ONLY
ONLY
SymbolsUsedonConcentratorandAccessoriesChapter7 35
7English
Meaning
KeepDry
IndoororDryLocationUseOnly,DoNotGetWet
UseNoOilorGrease
DoNotDisassemble
DoNotDisposeofInUnsortedMunicipalWaste
TypeBFAppliedPart,NotIntendedforCardiacApplication
ClassIIDevice
ElectricalSafetyAgencyCertificationLogo
CompliesWithApplicableEUDirectivesIncludingMedicalDeviceDirective
Symbol
ONLY
User Interface Label
Meaning
ON/OFFButton
DisplayBacklightButton
IncreaseFlowSetting
DecreaseFlowSetting
AudibleAlarmButton
Symbol
36
ONLY
ONLY
ONLYONLY
0123
93-01213-00-00 rev1
0123
COLOR SUMMARY: 1. Black
Inogen One® G2 ConcentratorDimensions:Withsinglebattery:
L/W/H: 10.7in(27.2cm)/3.9in(9.9cm)/8.75in(22.2cm)L/W/H: 10.7in(27.2cm)/3.9in(9.9cm)/9.5in(24.2cm)
Weight: 7.25pounds(3.3kg)(includessinglebattery)Noise: Lessthan38dBA(aspackaged)onsetting2Warm-UpTime: 2minutesOxygenConcentration: 90%-3%/+6%atallsettingsFlowControlSettings: 5settings:1to5Power: UniversalPowerSupply:
ACInput: 100to240VAC50to60Hz
Auto-Sensing: 1.0A
DCInput: 13.5-15VDC,10ADCOutput: 19VDC,5.0Amax.
RechargeableBattery: Voltage: 12.0to16.8VDC
BatteryDuration: Upto4hourswithsinglebatteryUpto8hourswithdoublebattery
BatteryChargingTime: Upto4hoursforasinglebatteryUpto8hoursforadoublebattery
EnvironmentalRangesIntendedforUse:
Temperature:41to104˚F(4to40˚C)Humidity:0%to95%,non-condensingAltitude:0to10,000ft(0to3048meters)
EnvironmentalRangesIntendedforStorage:
Temperature:-4to140˚F(-20to60˚C)Humidity:0%to95%,non-condensingStoreinadryenvironment
Transportation: KeepDry,HandleWithCare
InogenOne®G2SystemSpecificationsChapter8
Inogen One® G2 System Specifications8
37
English
Inogen One® G2 Concentrator (continued)
ClassificationsModeofOperation: ContinuousDutyTypeofProtectionAgainstElectricalShock:
ClassII
DegreeofProtectionAgainstElectricalShock:
TypeBFNotintendedforcardiacapplication
DegreeofProtectionAgainstIngressofWater:
IP20
DegreeofSafetyforApplicationinPresenceofAnestheticGases:
Notsuitableforsuchapplication
38
ELECTrOMAGNETIC COMPATIBILITy
ThisCEMarkedequipmenthasbeentestedandfoundtocomplywiththeEMClimitsfortheMedicalDeviceDirective93/42/EEC[EN55011ClassBandEN60601-1-2].Theselimitsaredesignedtoprovidereasonableprotectionagainstharmfulinterferenceinatypicalmedicalinstallation.
Theequipmentgeneratesusesandcanradiateradiofrequencyenergyandifnotinstalledandusedinaccordancewiththeinstructions,maycauseharmfulinterferencetootherdevicesinthevicinity.
TestedbyIndependentLaboratory:
Safety:IEC60601-1CAN/CSAC22.2No.60601-1ElectromagneticCompatibility:IEC60601-1-2RTCADO160
InogenOne®G2SystemSpecificationsChapter8 39
However,thereisnoguaranteethatinterferencewillnotoccurinaparticularinstallation.Ifthisequipmentdoescauseharmfulinterferencewithotherdevices,whichcanbedeterminedbyturningtheequipmentoffandon,theuserisencouragedtotrytocorrecttheinterferencebyoneormoreofthefollowingmeasures:
1. Reorientorrelocatethereceivingdevice.2. Increasetheseparationbetweentheequipment.3. Connecttheequipmentintoanoutletonacircuitdifferentfromthatto
whichtheotherdevice(s)isconnected.4. ConsultInogenoryourequipmentproviderforassistance.
CAUTION PortableandmobileRFcommunicationsequipmentcanaffectmedicalelectricalequipment.
CAUTION Stackingorplacingequipmentadjacenttootherdevicesisnotrecommended,wheresuchconfigurationsarenecessary,allequipmentinquestionshouldbecarefullyobservedtoensurethatEMIdoesnotdegradeperformance.
CAUTION Useofaccessories,transducersandcablesotherthanthosespecified,withtheexceptionoftransducersandcablessoldbythemanufactureroftheequipmentasreplacementpartsforinternalcomponents,mayresultinincreasedemissionsanddecreasedimmunityoftheequipmentorsystem.
English
Guidance and Manufacturer’s Declaration – Electromagnetic EmissionsTheInogenOne®G2OxygenConcentratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.ThecustomerortheuseroftheInogenOne®G2OxygenConcentratorshouldassurethatitisusedinsuchanenvironment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RFemissionsCISPR11
Group1 TheInogenOne®G2OxygenConcentratorusesRFenergyonlyforitsinternalfunction.ThereforeitsRFemissionsareverylowandnotlikelytocauseanyinterferenceinnearbyequipment.
RFemissionsCISPR11
ClassB TheInogenOne®G2OxygenConcentratorissuitableforuseinallestablishmentsincludingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
HarmonicEmissionsIEC61000-3-2
ClassA
Voltagefluctuations/flickeremissionsIEC61000-3-3
Complies
40