today’s outline - january 28, 2020phys.iit.edu/~segre/phys570/20s/lecture_05.pdf · today’s...

178
Today’s outline - January 28, 2020 C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Upload: others

Post on 19-Aug-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 2: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 3: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 4: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 5: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 6: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs

• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 7: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 8: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors

• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 9: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters

• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 10: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors

• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 11: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 12: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 13: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Today’s outline - January 28, 2020

• Undulator harmonics

• Undulator coherence

• Emittance

• Time structure

• APS-U, ERLs and FELs• Detectors

• Gas detectors• Scintillation counters• Solid state detectors• Area detectors

Reading Assignment: Chapter 3.1–3.3

Homework Assignment #01:Chapter 2: 2,3,5,6,8due Thursday, January 30, 2020

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 1 / 33

Page 14: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1 and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 15: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1 and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 16: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1 and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 17: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1 and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 18: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1 and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 19: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1 and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 20: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1

and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 21: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1

and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)

Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 22: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1 and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)

Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 23: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator review

For an undulator of period λu we have derived the following undulatorparameters and their relationships:

the K parameter, a dimension-less quantity which represents the“strength” of the undulator

the electron path length throughthe undulator, Sλu

the fundamental wavelength of theundulator, λ1 and the origin of theodd and even harmonics

K =e

2πmcλuB0

= 0.934λu[cm]B0[T]

Sλu ≈ λu(

1 +1

4

K 2

γ2

)

λ1 ≈λu2γ2

(1 +

K 2

2

)Now let us look at the coherence of the undulator radiation

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 2 / 33

Page 24: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Diffraction grating

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

A diffraction grating consists of N coherent sources whose emission isdetected at a single point.

δL

δL

The radiation from each slit has totravel a slightly different distance toget to the detector. For consecu-tive slits this path length difference,Lm+1 − Lm = δL, gives rise to aphase shift, 2πε = 2πδL/λ. So at thedetector, we have a sum of waves:

N−1∑m=0

e i(~k·~r+2πmε) = e i

~k·~rN−1∑m=0

e i2πmε

the phase shift from each undulatorpole depends on the wavelength λu

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 3 / 33

Page 25: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Diffraction grating

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.A diffraction grating consists of N coherent sources whose emission isdetected at a single point.

δL

δL

The radiation from each slit has totravel a slightly different distance toget to the detector. For consecu-tive slits this path length difference,Lm+1 − Lm = δL, gives rise to aphase shift, 2πε = 2πδL/λ. So at thedetector, we have a sum of waves:

N−1∑m=0

e i(~k·~r+2πmε) = e i

~k·~rN−1∑m=0

e i2πmε

the phase shift from each undulatorpole depends on the wavelength λu

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 3 / 33

Page 26: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Diffraction grating

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.A diffraction grating consists of N coherent sources whose emission isdetected at a single point.

δL

δL

The radiation from each slit has totravel a slightly different distance toget to the detector.

For consecu-tive slits this path length difference,Lm+1 − Lm = δL, gives rise to aphase shift, 2πε = 2πδL/λ. So at thedetector, we have a sum of waves:

N−1∑m=0

e i(~k·~r+2πmε) = e i

~k·~rN−1∑m=0

e i2πmε

the phase shift from each undulatorpole depends on the wavelength λu

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 3 / 33

Page 27: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Diffraction grating

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.A diffraction grating consists of N coherent sources whose emission isdetected at a single point.

δL

δL

The radiation from each slit has totravel a slightly different distance toget to the detector. For consecu-tive slits this path length difference,Lm+1 − Lm = δL,

gives rise to aphase shift, 2πε = 2πδL/λ. So at thedetector, we have a sum of waves:

N−1∑m=0

e i(~k·~r+2πmε) = e i

~k·~rN−1∑m=0

e i2πmε

the phase shift from each undulatorpole depends on the wavelength λu

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 3 / 33

Page 28: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Diffraction grating

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.A diffraction grating consists of N coherent sources whose emission isdetected at a single point.

δL

δL

The radiation from each slit has totravel a slightly different distance toget to the detector. For consecu-tive slits this path length difference,Lm+1 − Lm = δL, gives rise to aphase shift, 2πε = 2πδL/λ.

So at thedetector, we have a sum of waves:

N−1∑m=0

e i(~k·~r+2πmε) = e i

~k·~rN−1∑m=0

e i2πmε

the phase shift from each undulatorpole depends on the wavelength λu

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 3 / 33

Page 29: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Diffraction grating

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.A diffraction grating consists of N coherent sources whose emission isdetected at a single point.

δL

δL

The radiation from each slit has totravel a slightly different distance toget to the detector. For consecu-tive slits this path length difference,Lm+1 − Lm = δL, gives rise to aphase shift, 2πε = 2πδL/λ. So at thedetector, we have a sum of waves:

N−1∑m=0

e i(~k·~r+2πmε)

= e i~k·~r

N−1∑m=0

e i2πmε

the phase shift from each undulatorpole depends on the wavelength λu

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 3 / 33

Page 30: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Diffraction grating

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.A diffraction grating consists of N coherent sources whose emission isdetected at a single point.

δL

δL

The radiation from each slit has totravel a slightly different distance toget to the detector. For consecu-tive slits this path length difference,Lm+1 − Lm = δL, gives rise to aphase shift, 2πε = 2πδL/λ. So at thedetector, we have a sum of waves:

N−1∑m=0

e i(~k·~r+2πmε) = e i

~k·~rN−1∑m=0

e i2πmε

the phase shift from each undulatorpole depends on the wavelength λu

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 3 / 33

Page 31: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Diffraction grating

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.A diffraction grating consists of N coherent sources whose emission isdetected at a single point.

δL

δL

The radiation from each slit has totravel a slightly different distance toget to the detector. For consecu-tive slits this path length difference,Lm+1 − Lm = δL, gives rise to aphase shift, 2πε = 2πδL/λ. So at thedetector, we have a sum of waves:

N−1∑m=0

e i(~k·~r+2πmε) = e i

~k·~rN−1∑m=0

e i2πmε

the phase shift from each undulatorpole depends on the wavelength λu

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 3 / 33

Page 32: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1 = 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 33: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km

= 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1 = 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 34: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1 = 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 35: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1 = 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 36: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1

and

SN = 1 + kSN−1 = 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 37: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1

= 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 38: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1 = 1 + k(SN − kN−1)

= 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 39: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1 = 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 40: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1 = 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 41: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1 = 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN

−→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 42: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Geometric series

The sum is simply a geometric series, SN with k = e i2πε

SN =N−1∑m=0

km = 1 + k + k2 + · · ·+ kN−2 + kN−1

We can develop a recursion relation by writing the expression for SN−1

SN−1 =N−2∑m=0

km = 1 + k + k2 + · · ·+ kN−2

so we can write that SN−1 = SN − kN−1 and

SN = 1 + kSN−1 = 1 + k(SN − kN−1) = 1 + kSN − kN

Solving for SN , we have

SN − kSN = 1− kN −→ SN =1− kN

1− k

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 4 / 33

Page 43: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN =1− e i2πNε

1− e i2πε=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2 =

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 44: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN

=1− e i2πNε

1− e i2πε=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2 =

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 45: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN =1− e i2πNε

1− e i2πε

=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2 =

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 46: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN =1− e i2πNε

1− e i2πε=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2 =

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 47: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN =1− e i2πNε

1− e i2πε=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2 =

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 48: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN =1− e i2πNε

1− e i2πε=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2 =

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 49: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN =1− e i2πNε

1− e i2πε=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2 =

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 50: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN =1− e i2πNε

1− e i2πε=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2

=

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 51: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN =1− e i2πNε

1− e i2πε=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2 =

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 52: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Intensity from a diffraction grating

Restoring the expression for k = e i2πε, we have:

N−1∑m=0

e i2πmε = SN =1− e i2πNε

1− e i2πε=

(e−iπNε − e iπNε

e−iπε − e iπε

)e iπNε

e iπε

SN =

(sin (πNε)

sin (πε)

)e iπ(N−1)ε

Therefore, for the diffraction grating we can calculate the intensity at thedetector as

I =

∣∣∣∣∣e i~k·~rN−1∑m=0

e i2πmε

∣∣∣∣∣2

=∣∣∣e i~k·~rSN

∣∣∣2 =

∣∣∣∣e i~k·~r sin (πNε)

sin (πε)e iπ(N−1)ε

∣∣∣∣2

I =sin2 (πNε)

sin2 (πε)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 5 / 33

Page 53: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=0

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

units)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 54: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=0

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 55: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=5o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 56: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=10o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 57: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=15o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 58: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=20o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 59: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=25o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 60: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=30o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 61: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=35o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 62: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=40o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 63: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=45o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 64: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Beam coherence

An N period undulator is basically like a diffraction grating, only in thetime domain rather than the space domain.

2πε=45o

0 10 20 30 40

2πε [degrees]

Inte

nsity (

arb

un

its)

With the height and width of the peak dependent on the number of poles.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 6 / 33

Page 65: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator monochromaticity

2πε

Inte

nsity (

arb

un

its)

32 poles

16 poles

96 poles

The more poles in the undulator,the more monochromatic the beamsince a slight change in ε = δL/λimplies a slightly different wave-length λ

Another way to look at this is thatthe longer the undulator, the longerthe pulse train in time and the nar-rower the frequency distribution inits Fourier Transform

The APS has a 72 pole undulatorof 3.3 cm period

Higher order harmonics have nar-rower energy bandwidth but lowerpeak intensity

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 7 / 33

Page 66: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator monochromaticity

2πε

Inte

nsity (

arb

un

its)

32 poles

16 poles

96 poles

The more poles in the undulator,the more monochromatic the beamsince a slight change in ε = δL/λimplies a slightly different wave-length λ

Another way to look at this is thatthe longer the undulator, the longerthe pulse train in time and the nar-rower the frequency distribution inits Fourier Transform

The APS has a 72 pole undulatorof 3.3 cm period

Higher order harmonics have nar-rower energy bandwidth but lowerpeak intensity

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 7 / 33

Page 67: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator monochromaticity

2πε

Inte

nsity (

arb

un

its)

32 poles

16 poles

96 poles

The more poles in the undulator,the more monochromatic the beamsince a slight change in ε = δL/λimplies a slightly different wave-length λ

Another way to look at this is thatthe longer the undulator, the longerthe pulse train in time and the nar-rower the frequency distribution inits Fourier Transform

The APS has a 72 pole undulatorof 3.3 cm period

Higher order harmonics have nar-rower energy bandwidth but lowerpeak intensity

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 7 / 33

Page 68: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Undulator monochromaticity

2πε

Inte

nsity (

arb

un

its)

3rd

harmonic

5th

harmonic

1st harmonic

The more poles in the undulator,the more monochromatic the beamsince a slight change in ε = δL/λimplies a slightly different wave-length λ

Another way to look at this is thatthe longer the undulator, the longerthe pulse train in time and the nar-rower the frequency distribution inits Fourier Transform

The APS has a 72 pole undulatorof 3.3 cm period

Higher order harmonics have nar-rower energy bandwidth but lowerpeak intensity

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 7 / 33

Page 69: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Synchrotron time structure

There are two importanttime scales for a storagering such as the APS: pulselength and interpulse spacing

The APS pulse length in 24-bunch mode is 90 ps whilethe pulses come every 154 ns

Other modes include single-bunch mode for timing ex-periments and 324-bunchmode (inter pulse timing of11.7 ns) for a more constantx-ray flux

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 8 / 33

Page 70: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Synchrotron time structure

There are two importanttime scales for a storagering such as the APS: pulselength and interpulse spacing

The APS pulse length in 24-bunch mode is 90 ps whilethe pulses come every 154 ns

Other modes include single-bunch mode for timing ex-periments and 324-bunchmode (inter pulse timing of11.7 ns) for a more constantx-ray flux

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 8 / 33

Page 71: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Synchrotron time structure

There are two importanttime scales for a storagering such as the APS: pulselength and interpulse spacing

The APS pulse length in 24-bunch mode is 90 ps whilethe pulses come every 154 ns

Other modes include single-bunch mode for timing ex-periments and 324-bunchmode (inter pulse timing of11.7 ns) for a more constantx-ray flux

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 8 / 33

Page 72: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Emittance

Is there a limit to the brilliance of an undulator source at a synchrotron?

the brilliance is inversely proportional to the square of the product of thelinear source size and the angular divergence

brilliance =flux [photons/s]

divergence[mrad2

]· source size [mm2] [0.1% bandwidth]

the product of the source size and divergence iscalled the emittance, ε and the brilliance is thuslimited by the product of the emittance of theradiation in the horizontal and vertical directionsεxεy

this emittance cannot be changed but it can berotated or deformed by magnetic fields as theelectron beam travels around the storage ringas long as the area is kept constant

y

y

σ

σ

y

y

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 9 / 33

Page 73: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Emittance

Is there a limit to the brilliance of an undulator source at a synchrotron?

the brilliance is inversely proportional to the square of the product of thelinear source size and the angular divergence

brilliance =flux [photons/s]

divergence[mrad2

]· source size [mm2] [0.1% bandwidth]

the product of the source size and divergence iscalled the emittance, ε and the brilliance is thuslimited by the product of the emittance of theradiation in the horizontal and vertical directionsεxεy

this emittance cannot be changed but it can berotated or deformed by magnetic fields as theelectron beam travels around the storage ringas long as the area is kept constant

y

y

σ

σ

y

y

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 9 / 33

Page 74: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Emittance

Is there a limit to the brilliance of an undulator source at a synchrotron?

the brilliance is inversely proportional to the square of the product of thelinear source size and the angular divergence

brilliance =flux [photons/s]

divergence[mrad2

]· source size [mm2] [0.1% bandwidth]

the product of the source size and divergence iscalled the emittance, ε

and the brilliance is thuslimited by the product of the emittance of theradiation in the horizontal and vertical directionsεxεy

this emittance cannot be changed but it can berotated or deformed by magnetic fields as theelectron beam travels around the storage ringas long as the area is kept constant

y

y

σ

σ

y

y

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 9 / 33

Page 75: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Emittance

Is there a limit to the brilliance of an undulator source at a synchrotron?

the brilliance is inversely proportional to the square of the product of thelinear source size and the angular divergence

brilliance =flux [photons/s]

divergence[mrad2

]· source size [mm2] [0.1% bandwidth]

the product of the source size and divergence iscalled the emittance, ε and the brilliance is thuslimited by the product of the emittance of theradiation in the horizontal and vertical directionsεxεy

this emittance cannot be changed but it can berotated or deformed by magnetic fields as theelectron beam travels around the storage ringas long as the area is kept constant

y

y

σ

σ

y

y

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 9 / 33

Page 76: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Emittance

Is there a limit to the brilliance of an undulator source at a synchrotron?

the brilliance is inversely proportional to the square of the product of thelinear source size and the angular divergence

brilliance =flux [photons/s]

divergence[mrad2

]· source size [mm2] [0.1% bandwidth]

the product of the source size and divergence iscalled the emittance, ε and the brilliance is thuslimited by the product of the emittance of theradiation in the horizontal and vertical directionsεxεy

this emittance cannot be changed but it can berotated

or deformed by magnetic fields as theelectron beam travels around the storage ringas long as the area is kept constant

y

y

σ

σ

y

y

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 9 / 33

Page 77: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Emittance

Is there a limit to the brilliance of an undulator source at a synchrotron?

the brilliance is inversely proportional to the square of the product of thelinear source size and the angular divergence

brilliance =flux [photons/s]

divergence[mrad2

]· source size [mm2] [0.1% bandwidth]

the product of the source size and divergence iscalled the emittance, ε and the brilliance is thuslimited by the product of the emittance of theradiation in the horizontal and vertical directionsεxεy

this emittance cannot be changed but it can berotated or deformed by magnetic fields as theelectron beam travels around the storage ringas long as the area is kept constant

y

y

σ

σ

y

y

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 9 / 33

Page 78: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS emittance

For photon emission from a single electron in a 2m undulator at 1A

y

10µrad

-10µrad

10µm-10µm

y

σradiation = 9.1µm

σ′radiation = 7.7µrad

σγ=

√Lλ

4π= 1.3µm

σ′γ=

√λ

L= 7.1µrad

current APS electron beamparameters are

σy= 9.1µm

σ′y= 3.0µrad

must convolute to get pho-ton emission from entirebeam (in vertical direction)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 10 / 33

Page 79: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS emittance

For photon emission from a single electron in a 2m undulator at 1A

y

10µrad

-10µrad

10µm-10µm

y

σradiation = 9.1µm

σ′radiation = 7.7µrad

σγ=

√Lλ

4π= 1.3µm

σ′γ=

√λ

L= 7.1µrad

current APS electron beamparameters are

σy= 9.1µm

σ′y= 3.0µrad

must convolute to get pho-ton emission from entirebeam (in vertical direction)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 10 / 33

Page 80: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS emittance

For photon emission from a single electron in a 2m undulator at 1A

y

10µrad

-10µrad

10µm-10µm

y

σradiation = 9.1µm

σ′radiation = 7.7µrad

σγ=

√Lλ

4π= 1.3µm

σ′γ=

√λ

L= 7.1µrad

current APS electron beamparameters are

σy= 9.1µm

σ′y= 3.0µrad

must convolute to get pho-ton emission from entirebeam (in vertical direction)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 10 / 33

Page 81: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS emittance

For photon emission from a single electron in a 2m undulator at 1A

y

10µrad

-10µrad

10µm-10µm

y

σradiation = 9.1µm

σ′radiation = 7.7µrad

σγ=

√Lλ

4π= 1.3µm

σ′γ=

√λ

L= 7.1µrad

current APS electron beamparameters are

σy= 9.1µm

σ′y= 3.0µrad

must convolute to get pho-ton emission from entirebeam (in vertical direction)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 10 / 33

Page 82: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS emittance

For photon emission from a single electron in a 2m undulator at 1A

y

10µrad

-10µrad

10µm-10µm

y

σradiation = 9.1µm

σ′radiation = 7.7µrad

σγ=

√Lλ

4π= 1.3µm

σ′γ=

√λ

L= 7.1µrad

current APS electron beamparameters are

σy= 9.1µm

σ′y= 3.0µrad

must convolute to get pho-ton emission from entirebeam (in vertical direction)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 10 / 33

Page 83: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS emittance

For photon emission from a single electron in a 2m undulator at 1A

y

10µrad

-10µrad

10µm-10µm

y

σradiation = 9.1µm

σ′radiation = 7.7µrad

σγ=

√Lλ

4π= 1.3µm

σ′γ=

√λ

L= 7.1µrad

current APS electron beamparameters are

σy= 9.1µm

σ′y= 3.0µrad

must convolute to get pho-ton emission from entirebeam (in vertical direction)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 10 / 33

Page 84: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS emittance

For photon emission from a single electron in a 2m undulator at 1A

y

10µrad

-10µrad

10µm-10µm

y

σradiation = 9.1µm

σ′radiation = 7.7µrad

σγ=

√Lλ

4π= 1.3µm

σ′γ=

√λ

L= 7.1µrad

current APS electron beamparameters are

σy= 9.1µm

σ′y= 3.0µrad

must convolute to get pho-ton emission from entirebeam (in vertical direction)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 10 / 33

Page 85: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS emittance

For photon emission from a single electron in a 2m undulator at 1A

y

10µrad

-10µrad

10µm-10µm

y

σradiation = 9.1µm

σ′radiation = 7.7µrad

σγ=

√Lλ

4π= 1.3µm

σ′γ=

√λ

L= 7.1µrad

current APS electron beamparameters are

σy= 9.1µm

σ′y= 3.0µrad

must convolute to get pho-ton emission from entirebeam (in vertical direction)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 10 / 33

Page 86: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Evolution of APS parameters

Parameter 1995 2001 2005σx 334 µm 352 µm 280 µmσ′x 24 µrad 22 µrad 11.6 µrad

σy 89 µm 18.4 µm 9.1 µmσ′y 8.9 µrad 4.2 µrad 3.0 µrad

When first commissioned in 1995, the APS electron beam size anddivergence was relatively large, particularly in the horizontal, x direction

By the end of the first decade of operation, the horizontal source sizedecreased by about 16% and its horizontal divergence by more than 50%

At the same time the vertical source size decreased by over 90% and thevertical divergence by nearly 67%

The next big upgrade (slated for 2022) will make the beam more square inspace and by choosing the undulator correctly, a higher performanceinsertion device.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 11 / 33

Page 87: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Evolution of APS parameters

Parameter 1995 2001 2005σx 334 µm 352 µm 280 µmσ′x 24 µrad 22 µrad 11.6 µrad

σy 89 µm 18.4 µm 9.1 µmσ′y 8.9 µrad 4.2 µrad 3.0 µrad

When first commissioned in 1995, the APS electron beam size anddivergence was relatively large, particularly in the horizontal, x direction

By the end of the first decade of operation, the horizontal source sizedecreased by about 16% and its horizontal divergence by more than 50%

At the same time the vertical source size decreased by over 90% and thevertical divergence by nearly 67%

The next big upgrade (slated for 2022) will make the beam more square inspace and by choosing the undulator correctly, a higher performanceinsertion device.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 11 / 33

Page 88: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Evolution of APS parameters

Parameter 1995 2001 2005σx 334 µm 352 µm 280 µmσ′x 24 µrad 22 µrad 11.6 µrad

σy 89 µm 18.4 µm 9.1 µmσ′y 8.9 µrad 4.2 µrad 3.0 µrad

When first commissioned in 1995, the APS electron beam size anddivergence was relatively large, particularly in the horizontal, x direction

By the end of the first decade of operation, the horizontal source sizedecreased by about 16% and its horizontal divergence by more than 50%

At the same time the vertical source size decreased by over 90% and thevertical divergence by nearly 67%

The next big upgrade (slated for 2022) will make the beam more square inspace and by choosing the undulator correctly, a higher performanceinsertion device.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 11 / 33

Page 89: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Evolution of APS parameters

Parameter 1995 2001 2005σx 334 µm 352 µm 280 µmσ′x 24 µrad 22 µrad 11.6 µrad

σy 89 µm 18.4 µm 9.1 µmσ′y 8.9 µrad 4.2 µrad 3.0 µrad

When first commissioned in 1995, the APS electron beam size anddivergence was relatively large, particularly in the horizontal, x direction

By the end of the first decade of operation, the horizontal source sizedecreased by about 16% and its horizontal divergence by more than 50%

At the same time the vertical source size decreased by over 90% and thevertical divergence by nearly 67%

The next big upgrade (slated for 2022) will make the beam more square inspace and by choosing the undulator correctly, a higher performanceinsertion device.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 11 / 33

Page 90: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Evolution of APS parameters

Parameter 1995 2001 2005σx 334 µm 352 µm 280 µmσ′x 24 µrad 22 µrad 11.6 µrad

σy 89 µm 18.4 µm 9.1 µmσ′y 8.9 µrad 4.2 µrad 3.0 µrad

When first commissioned in 1995, the APS electron beam size anddivergence was relatively large, particularly in the horizontal, x direction

By the end of the first decade of operation, the horizontal source sizedecreased by about 16% and its horizontal divergence by more than 50%

At the same time the vertical source size decreased by over 90% and thevertical divergence by nearly 67%

The next big upgrade (slated for 2022) will make the beam more square inspace and by choosing the undulator correctly, a higher performanceinsertion device.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 11 / 33

Page 91: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS upgrade

In 2022, the APS will shut down for a major rebuild with a totally newmagnetic lattice.

One of the biggest changes will be the beam (source)size and emittance

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

The beam will be nearly square and there will be much more coherencefrom the undulators

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 12 / 33

Page 92: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS upgrade

In 2022, the APS will shut down for a major rebuild with a totally newmagnetic lattice. One of the biggest changes will be the beam (source)size and emittance

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

The beam will be nearly square and there will be much more coherencefrom the undulators

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 12 / 33

Page 93: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS upgrade

In 2022, the APS will shut down for a major rebuild with a totally newmagnetic lattice. One of the biggest changes will be the beam (source)size and emittance

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

The beam will be nearly square and there will be much more coherencefrom the undulators

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 12 / 33

Page 94: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS upgrade

In 2022, the APS will shut down for a major rebuild with a totally newmagnetic lattice. One of the biggest changes will be the beam (source)size and emittance

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

The beam will be nearly square and there will be much more coherencefrom the undulators

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 12 / 33

Page 95: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APS upgrade

In 2022, the APS will shut down for a major rebuild with a totally newmagnetic lattice. One of the biggest changes will be the beam (source)size and emittance

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

Position (µm)

Positio

n (

µm

)

0

300

600

-300

-600

6003000-300-600

The beam will be nearly square and there will be much more coherencefrom the undulators

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 12 / 33

Page 96: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APSU undulator performance

The APS upgrade will bea diffraction-limited sourcewith a lower energy (6.0GeV) and doubled current(200 mA).

Since MRCAT’s science isprimarily flux driven, the goalwill be to replace the 2.4mundulator with one that out-performs the current 33mmperiod but with only modestincrease in power.

The APS is calling this a“4th” generation synchrotronsource

0 10 20 30 40Photon Energy (keV)

0

50

100

150

200

Pow

er

Density (

W/m

m2)

33 mm

31 mm

29 mm

27 mm

APS 33 mm

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 13 / 33

Page 97: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APSU undulator performance

The APS upgrade will bea diffraction-limited sourcewith a lower energy (6.0GeV) and doubled current(200 mA).

Since MRCAT’s science isprimarily flux driven, the goalwill be to replace the 2.4mundulator with one that out-performs the current 33mmperiod but with only modestincrease in power.

The APS is calling this a“4th” generation synchrotronsource

0 10 20 30 40Photon Energy (keV)

1e+13

1e+14

Ph

oto

n F

lux (

ph

/s/0

.1%

bw

)

APS 33 mm

31 mm

29 mm

27 mm

33 mm

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 13 / 33

Page 98: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

APSU undulator performance

The APS upgrade will bea diffraction-limited sourcewith a lower energy (6.0GeV) and doubled current(200 mA).

Since MRCAT’s science isprimarily flux driven, the goalwill be to replace the 2.4mundulator with one that out-performs the current 33mmperiod but with only modestincrease in power.

The APS is calling this a“4th” generation synchrotronsource

0 10 20 30 40Photon Energy (keV)

0

50

100

150

200

Pow

er

Density (

W/m

m2)

33 mm

31 mm

29 mm

27 mm

APS 33 mm

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 13 / 33

Page 99: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Energy recovery linacs

Undulators have limited peak brilliance

but the use of an energy recoverylinac can overcome this limitation and enhance peak brilliance by up tothree orders of magnitude

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 14 / 33

Page 100: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Energy recovery linacs

Undulators have limited peak brilliance but the use of an energy recoverylinac can overcome this limitation and enhance peak brilliance by up tothree orders of magnitude

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 14 / 33

Page 101: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Energy recovery linacs

Undulators have limited peak brilliance but the use of an energy recoverylinac can overcome this limitation and enhance peak brilliance by up tothree orders of magnitude

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 14 / 33

Page 102: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Free electron laser

Initial electron cloud, each electron emits co-herently but independently

Over course of 100 m, electric field of pho-tons, feeds back on the electron bunch

Microbunches form with period of FEL (andradiation in electron frame)

Each microbunch emits coherently withneighboring ones

Again, an alternative way to view this is that the pulse train from a 100mlong undulator is long enough in time to produce a monochromatic andcoherent frequency distribution when Fourier Transformed

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 15 / 33

Page 103: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Free electron laser

Initial electron cloud, each electron emits co-herently but independently

Over course of 100 m, electric field of pho-tons, feeds back on the electron bunch

Microbunches form with period of FEL (andradiation in electron frame)

Each microbunch emits coherently withneighboring ones

Again, an alternative way to view this is that the pulse train from a 100mlong undulator is long enough in time to produce a monochromatic andcoherent frequency distribution when Fourier Transformed

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 15 / 33

Page 104: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Free electron laser

Initial electron cloud, each electron emits co-herently but independently

Over course of 100 m, electric field of pho-tons, feeds back on the electron bunch

Microbunches form with period of FEL (andradiation in electron frame)

Each microbunch emits coherently withneighboring ones

Again, an alternative way to view this is that the pulse train from a 100mlong undulator is long enough in time to produce a monochromatic andcoherent frequency distribution when Fourier Transformed

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 15 / 33

Page 105: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Free electron laser

Initial electron cloud, each electron emits co-herently but independently

Over course of 100 m, electric field of pho-tons, feeds back on the electron bunch

Microbunches form with period of FEL (andradiation in electron frame)

Each microbunch emits coherently withneighboring ones

Again, an alternative way to view this is that the pulse train from a 100mlong undulator is long enough in time to produce a monochromatic andcoherent frequency distribution when Fourier Transformed

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 15 / 33

Page 106: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Free electron laser

Initial electron cloud, each electron emits co-herently but independently

Over course of 100 m, electric field of pho-tons, feeds back on the electron bunch

Microbunches form with period of FEL (andradiation in electron frame)

Each microbunch emits coherently withneighboring ones

Again, an alternative way to view this is that the pulse train from a 100mlong undulator is long enough in time to produce a monochromatic andcoherent frequency distribution when Fourier Transformed

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 15 / 33

Page 107: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Free electron laser

Initial electron cloud, each electron emits co-herently but independently

Over course of 100 m, electric field of pho-tons, feeds back on the electron bunch

Microbunches form with period of FEL (andradiation in electron frame)

Each microbunch emits coherently withneighboring ones

Again, an alternative way to view this is that the pulse train from a 100mlong undulator is long enough in time to produce a monochromatic andcoherent frequency distribution when Fourier Transformed

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 15 / 33

Page 108: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Free electron laser

Initial electron cloud, each electron emits co-herently but independently

Over course of 100 m, electric field of pho-tons, feeds back on the electron bunch

Microbunches form with period of FEL (andradiation in electron frame)

Each microbunch emits coherently withneighboring ones

Again, an alternative way to view this is that the pulse train from a 100mlong undulator is long enough in time to produce a monochromatic andcoherent frequency distribution when Fourier Transformed

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 15 / 33

Page 109: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

FEL emission

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 16 / 33

Page 110: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

FEL emission

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 17 / 33

Page 111: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

FEL emission

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 18 / 33

Page 112: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

FEL layout

An FEL has a single accelerator whose electron beam is shuntedsequentially through different undulators and end stations

The single pass of the electron beam permits a very low emittance to beachieved and thus higher coherence

The high brightness usually results in destruction of the sample during theillumination, thus the need for multiple samples and multiple shotexperiments

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 19 / 33

Page 113: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

FEL layout

An FEL has a single accelerator whose electron beam is shuntedsequentially through different undulators and end stations

The single pass of the electron beam permits a very low emittance to beachieved and thus higher coherence

The high brightness usually results in destruction of the sample during theillumination, thus the need for multiple samples and multiple shotexperiments

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 19 / 33

Page 114: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

FEL layout

An FEL has a single accelerator whose electron beam is shuntedsequentially through different undulators and end stations

The single pass of the electron beam permits a very low emittance to beachieved and thus higher coherence

The high brightness usually results in destruction of the sample during theillumination, thus the need for multiple samples and multiple shotexperiments

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 19 / 33

Page 115: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

FEL layout

An FEL has a single accelerator whose electron beam is shuntedsequentially through different undulators and end stations

The single pass of the electron beam permits a very low emittance to beachieved and thus higher coherence

The high brightness usually results in destruction of the sample during theillumination, thus the need for multiple samples and multiple shotexperiments

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 19 / 33

Page 116: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Compact sources

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 20 / 33

Page 117: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Lyncean CLS

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 21 / 33

Page 118: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 119: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 120: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation counters

Solid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 121: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 122: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 123: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 124: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 125: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 126: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 127: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 128: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 129: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 130: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 22 / 33

Page 131: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Gas Detectors

Gas detectors operate in several modes depending on the particle type, gascomposition and pressure and voltage applied

The most interesting arethe ionization, propor-tional, and Geiger-Muller

At a synchrotron, the par-ticle being detected ismost often a photon (γ)

The most useful regimeis the ionization regionwhere the output pulseis independent of the ap-plied voltage over a widerange

Applied Voltage

Puls

e S

ize

Recom

bin

ation

Ioniz

ation C

ham

ber

Pro

port

ional

Lim

ited P

roport

ional

Geig

er-

Muelle

r

Continuous D

ischarg

e

10-1

101

103

105

107

109

1011

1013

α

β

γ

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 23 / 33

Page 132: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Gas Detectors

Gas detectors operate in several modes depending on the particle type, gascomposition and pressure and voltage applied

The most interesting arethe ionization, propor-tional, and Geiger-Muller

At a synchrotron, the par-ticle being detected ismost often a photon (γ)

The most useful regimeis the ionization regionwhere the output pulseis independent of the ap-plied voltage over a widerange

Applied Voltage

Puls

e S

ize

Recom

bin

ation

Ioniz

ation C

ham

ber

Pro

port

ional

Lim

ited P

roport

ional

Geig

er-

Muelle

r

Continuous D

ischarg

e

10-1

101

103

105

107

109

1011

1013

α

β

γ

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 23 / 33

Page 133: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Gas Detectors

Gas detectors operate in several modes depending on the particle type, gascomposition and pressure and voltage applied

The most interesting arethe ionization, propor-tional, and Geiger-Muller

At a synchrotron, the par-ticle being detected ismost often a photon (γ)

The most useful regimeis the ionization regionwhere the output pulseis independent of the ap-plied voltage over a widerange

Applied Voltage

Puls

e S

ize

Recom

bin

ation

Ioniz

ation C

ham

ber

Pro

port

ional

Lim

ited P

roport

ional

Geig

er-

Muelle

r

Continuous D

ischarg

e

10-1

101

103

105

107

109

1011

1013

γ

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 23 / 33

Page 134: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Gas Detectors

Gas detectors operate in several modes depending on the particle type, gascomposition and pressure and voltage applied

The most interesting arethe ionization, propor-tional, and Geiger-Muller

At a synchrotron, the par-ticle being detected ismost often a photon (γ)

The most useful regimeis the ionization regionwhere the output pulseis independent of the ap-plied voltage over a widerange

Applied Voltage

Puls

e S

ize

Recom

bin

ation

Ioniz

ation C

ham

ber

Pro

port

ional

Lim

ited P

roport

ional

Geig

er-

Muelle

r

Continuous D

ischarg

e

10-1

101

103

105

107

109

1011

1013

γ

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 23 / 33

Page 135: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

window

x-rays

current collector A

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/I0 = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-ion pair(depending on the gas) makesthis useful for quantitativemeasurements.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 24 / 33

Page 136: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

window

x-rays

current collector A

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/I0 = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-ion pair(depending on the gas) makesthis useful for quantitativemeasurements.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 24 / 33

Page 137: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

window

x-rays

current collector A

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/I0 = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-ion pair(depending on the gas) makesthis useful for quantitativemeasurements.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 24 / 33

Page 138: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

window

x-rays

current collector A

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/I0 = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-ion pair(depending on the gas) makesthis useful for quantitativemeasurements.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 24 / 33

Page 139: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

window

x-rays

current collector A

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/I0 = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-ion pair(depending on the gas) makesthis useful for quantitativemeasurements.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 24 / 33

Page 140: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

window

x-rays

current collector A

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/I0 = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-ion pair(depending on the gas) makesthis useful for quantitativemeasurements.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 24 / 33

Page 141: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

window

x-rays

current collector A

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/I0 = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-ion pair(depending on the gas) makesthis useful for quantitativemeasurements.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 24 / 33

Page 142: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Getting a reading

The ionization chamber puts out a current in the nA range, this needs tobe converted into a useful measurement

Ionization

Chamber

Voltage to

Frequency

Transconductance

Amplifier

[nA] [kHz]Scaler

[V]

the tiny current is fed into an sensitive amplifier with gains of up to 1010

which outputs a voltage signal of 1-10 V that tracks the input with anadjustable time constant

the voltage is converted to a digital signal by a Voltage to Frequencyconverter which outputs 100 kHz (or more) pulse frequency per Volt ofinput

the digital pulse train is counted by a scaler for a user-definable length oftime

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 25 / 33

Page 143: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Getting a reading

The ionization chamber puts out a current in the nA range, this needs tobe converted into a useful measurement

Ionization

Chamber

[V] Voltage to

Frequency

Transconductance

Amplifier

[nA] [kHz]Scaler

the tiny current is fed into an sensitive amplifier with gains of up to 1010

which outputs a voltage signal of 1-10 V that tracks the input with anadjustable time constant

the voltage is converted to a digital signal by a Voltage to Frequencyconverter which outputs 100 kHz (or more) pulse frequency per Volt ofinput

the digital pulse train is counted by a scaler for a user-definable length oftime

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 25 / 33

Page 144: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Getting a reading

The ionization chamber puts out a current in the nA range, this needs tobe converted into a useful measurement

Ionization

Chamber

[V] Voltage to

Frequency

Transconductance

Amplifier

[nA] [kHz]Scaler

the tiny current is fed into an sensitive amplifier with gains of up to 1010

which outputs a voltage signal of 1-10 V that tracks the input with anadjustable time constant

the voltage is converted to a digital signal by a Voltage to Frequencyconverter which outputs 100 kHz (or more) pulse frequency per Volt ofinput

the digital pulse train is counted by a scaler for a user-definable length oftime

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 25 / 33

Page 145: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Getting a reading

The ionization chamber puts out a current in the nA range, this needs tobe converted into a useful measurement

Ionization

Chamber

[V] Voltage to

Frequency

Transconductance

Amplifier

[nA] [kHz]Scaler

the tiny current is fed into an sensitive amplifier with gains of up to 1010

which outputs a voltage signal of 1-10 V that tracks the input with anadjustable time constant

the voltage is converted to a digital signal by a Voltage to Frequencyconverter which outputs 100 kHz (or more) pulse frequency per Volt ofinput

the digital pulse train is counted by a scaler for a user-definable length oftime

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 25 / 33

Page 146: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Scintillation detector

Useful for photon counting experiments with rates less than 104/s

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.• Light strikes a thin photocathode which emits electrons into the

vacuum portion of a photomultiplier tube.• Photoelectrons are accelerated in steps, striking dynodes and

becoming amplified.• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 26 / 33

Page 147: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Scintillation detector

Useful for photon counting experiments with rates less than 104/s

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.• Photoelectrons are accelerated in steps, striking dynodes and

becoming amplified.• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 26 / 33

Page 148: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Scintillation detector

Useful for photon counting experiments with rates less than 104/s

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.• Light strikes a thin photocathode which emits electrons into the

vacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 26 / 33

Page 149: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Scintillation detector

Useful for photon counting experiments with rates less than 104/s

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.• Light strikes a thin photocathode which emits electrons into the

vacuum portion of a photomultiplier tube.• Photoelectrons are accelerated in steps, striking dynodes and

becoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 26 / 33

Page 150: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Scintillation detector

Useful for photon counting experiments with rates less than 104/s

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.• Light strikes a thin photocathode which emits electrons into the

vacuum portion of a photomultiplier tube.• Photoelectrons are accelerated in steps, striking dynodes and

becoming amplified.• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 26 / 33

Page 151: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Counting a scintillator pulse

∆E

E

the scintillator and photomultiplier put out avery fast negative-going tail pulse which is por-portional to the energy of the photon

a fast shaping and inverting amplifier is used toconvert this fast tail pulse to a more completepositive voltage pulse which can be processed

an energy discriminator with a threshold, E , andwindow ∆E is used to detect if the voltage pulsehas the desired height (which corresponds to thex-ray energy to be detected)

if the voltage pulse falls within the discriminatorwindow, a short digital pulse is output from thediscriminator and into a scaler for counting

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 27 / 33

Page 152: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Counting a scintillator pulse

∆E

E

the scintillator and photomultiplier put out avery fast negative-going tail pulse which is por-portional to the energy of the photon

a fast shaping and inverting amplifier is used toconvert this fast tail pulse to a more completepositive voltage pulse which can be processed

an energy discriminator with a threshold, E , andwindow ∆E is used to detect if the voltage pulsehas the desired height (which corresponds to thex-ray energy to be detected)

if the voltage pulse falls within the discriminatorwindow, a short digital pulse is output from thediscriminator and into a scaler for counting

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 27 / 33

Page 153: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Counting a scintillator pulse

∆E

E

the scintillator and photomultiplier put out avery fast negative-going tail pulse which is por-portional to the energy of the photon

a fast shaping and inverting amplifier is used toconvert this fast tail pulse to a more completepositive voltage pulse which can be processed

an energy discriminator with a threshold, E , andwindow ∆E is used to detect if the voltage pulsehas the desired height (which corresponds to thex-ray energy to be detected)

if the voltage pulse falls within the discriminatorwindow, a short digital pulse is output from thediscriminator and into a scaler for counting

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 27 / 33

Page 154: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Counting a scintillator pulse

∆E

E

the scintillator and photomultiplier put out avery fast negative-going tail pulse which is por-portional to the energy of the photon

a fast shaping and inverting amplifier is used toconvert this fast tail pulse to a more completepositive voltage pulse which can be processed

an energy discriminator with a threshold, E , andwindow ∆E is used to detect if the voltage pulsehas the desired height (which corresponds to thex-ray energy to be detected)

if the voltage pulse falls within the discriminatorwindow, a short digital pulse is output from thediscriminator and into a scaler for counting

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 27 / 33

Page 155: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Solid state detectors

The energy resolution of a scintillator is very poor and it is often necessaryto be able to distinguish the energy of specific x-rays. The semiconductordetector is ideal for this purpose.

a semiconductor such as Si or Ge has a com-pletely filled valence band and an empty con-duction band separated by an energy gap ∆Eof ∼ 1.2 eV for Si and ∼ 0.7 eV for Ge

when a photon is absorbed by the semicon-ductor, it promotes electrons from the va-lence to the conduction band creating anelectron-hole pair

Valence Band

Conduction Band

∆E

because of the small energy required to produce an electron-hole pair, onex-ray photon will create many and its energy can be detected with veryhigh resolution

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 28 / 33

Page 156: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Solid state detectors

The energy resolution of a scintillator is very poor and it is often necessaryto be able to distinguish the energy of specific x-rays. The semiconductordetector is ideal for this purpose.

a semiconductor such as Si or Ge has a com-pletely filled valence band and an empty con-duction band separated by an energy gap ∆Eof ∼ 1.2 eV for Si and ∼ 0.7 eV for Ge

when a photon is absorbed by the semicon-ductor, it promotes electrons from the va-lence to the conduction band creating anelectron-hole pair

Valence Band

Conduction Band

∆E

because of the small energy required to produce an electron-hole pair, onex-ray photon will create many and its energy can be detected with veryhigh resolution

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 28 / 33

Page 157: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Solid state detectors

The energy resolution of a scintillator is very poor and it is often necessaryto be able to distinguish the energy of specific x-rays. The semiconductordetector is ideal for this purpose.

a semiconductor such as Si or Ge has a com-pletely filled valence band and an empty con-duction band separated by an energy gap ∆Eof ∼ 1.2 eV for Si and ∼ 0.7 eV for Ge

when a photon is absorbed by the semicon-ductor, it promotes electrons from the va-lence to the conduction band creating anelectron-hole pair

Valence Band

Conduction Band

∆E

because of the small energy required to produce an electron-hole pair, onex-ray photon will create many and its energy can be detected with veryhigh resolution

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 28 / 33

Page 158: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Solid state detectors

The energy resolution of a scintillator is very poor and it is often necessaryto be able to distinguish the energy of specific x-rays. The semiconductordetector is ideal for this purpose.

a semiconductor such as Si or Ge has a com-pletely filled valence band and an empty con-duction band separated by an energy gap ∆Eof ∼ 1.2 eV for Si and ∼ 0.7 eV for Ge

when a photon is absorbed by the semicon-ductor, it promotes electrons from the va-lence to the conduction band creating anelectron-hole pair

Valence Band

Conduction Band

∆E

because of the small energy required to produce an electron-hole pair, onex-ray photon will create many and its energy can be detected with veryhigh resolution

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 28 / 33

Page 159: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Semiconductor junctions

Producing the electron-hole pairs is not sufficient. It is necessary toextract the charge from the device to make a measurement

start with two pieces of semiconductor,one n-type and the other p-type

if these two materials are brought intocontact, a natural depletion region isformed where there is an electric field ~E

p n

depletion region

+

this region is called an intrinsic region and is the only place where anabsorbed photon can create electron-hole pairs and have them be swept tothe p and n sides, respectively

by applying a reverse bias voltage, it is possible to extend the depletedregion, make the effective volume of the detector larger and increase theelectric field to get faster charge collection times

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 29 / 33

Page 160: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Semiconductor junctions

Producing the electron-hole pairs is not sufficient. It is necessary toextract the charge from the device to make a measurement

start with two pieces of semiconductor,one n-type and the other p-type

if these two materials are brought intocontact, a natural depletion region isformed where there is an electric field ~E

p n

depletion region

+

this region is called an intrinsic region and is the only place where anabsorbed photon can create electron-hole pairs and have them be swept tothe p and n sides, respectively

by applying a reverse bias voltage, it is possible to extend the depletedregion, make the effective volume of the detector larger and increase theelectric field to get faster charge collection times

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 29 / 33

Page 161: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Semiconductor junctions

Producing the electron-hole pairs is not sufficient. It is necessary toextract the charge from the device to make a measurement

start with two pieces of semiconductor,one n-type and the other p-type

if these two materials are brought intocontact, a natural depletion region isformed where there is an electric field ~E

p n

depletion region

+

ε

this region is called an intrinsic region and is the only place where anabsorbed photon can create electron-hole pairs and have them be swept tothe p and n sides, respectively

by applying a reverse bias voltage, it is possible to extend the depletedregion, make the effective volume of the detector larger and increase theelectric field to get faster charge collection times

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 29 / 33

Page 162: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Semiconductor junctions

Producing the electron-hole pairs is not sufficient. It is necessary toextract the charge from the device to make a measurement

start with two pieces of semiconductor,one n-type and the other p-type

if these two materials are brought intocontact, a natural depletion region isformed where there is an electric field ~E

p n

depletion region

+

ε

this region is called an intrinsic region and is the only place where anabsorbed photon can create electron-hole pairs and have them be swept tothe p and n sides, respectively

by applying a reverse bias voltage, it is possible to extend the depletedregion, make the effective volume of the detector larger and increase theelectric field to get faster charge collection times

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 29 / 33

Page 163: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Semiconductor junctions

Producing the electron-hole pairs is not sufficient. It is necessary toextract the charge from the device to make a measurement

start with two pieces of semiconductor,one n-type and the other p-type

if these two materials are brought intocontact, a natural depletion region isformed where there is an electric field ~E

p n

depletion region

+

ε

this region is called an intrinsic region and is the only place where anabsorbed photon can create electron-hole pairs and have them be swept tothe p and n sides, respectively

by applying a reverse bias voltage, it is possible to extend the depletedregion, make the effective volume of the detector larger and increase theelectric field to get faster charge collection times

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 29 / 33

Page 164: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Silicon Drift Detector

Same principle as intrinsic or p-i-n detector but much more compact andoperates at higher temperatures

Relatively low stopping power is a drawback

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 30 / 33

Page 165: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Detector operation

Li metal

P-doped+Vbias

Vout

A A A A1 2 3 4pile up

icr

ocr

output current is integrated into voltage pulses by a pre-amp, whenmaximum voltage is reached, output is optically reset

fast channel “sees” steps and starts integration; pulse heights are measuredwith slow channel by integration; and detected pulse pileups are rejected

electronics outputs input count rate (icr), output count rate (ocr), andareas of integrated pulses (An)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 31 / 33

Page 166: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Detector operation

Li metal

P-doped+Vbias

Vout

A A A A1 2 3 4pile up

icr

ocr

output current is integrated into voltage pulses by a pre-amp, whenmaximum voltage is reached, output is optically reset

fast channel “sees” steps and starts integration; pulse heights are measuredwith slow channel by integration; and detected pulse pileups are rejected

electronics outputs input count rate (icr), output count rate (ocr), andareas of integrated pulses (An)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 31 / 33

Page 167: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Detector operation

Li metal

P-doped+Vbias

Vout

A A A A1 2 3 4pile up

icr

ocr

output current is integrated into voltage pulses by a pre-amp, whenmaximum voltage is reached, output is optically reset

fast channel “sees” steps and starts integration;

pulse heights are measuredwith slow channel by integration; and detected pulse pileups are rejected

electronics outputs input count rate (icr), output count rate (ocr), andareas of integrated pulses (An)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 31 / 33

Page 168: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Detector operation

Li metal

P-doped+Vbias

Vout

A A A A1 2 3 4pile up

icr

ocr

output current is integrated into voltage pulses by a pre-amp, whenmaximum voltage is reached, output is optically reset

fast channel “sees” steps and starts integration; pulse heights are measuredwith slow channel by integration; and detected pulse pileups are rejected

electronics outputs input count rate (icr), output count rate (ocr), andareas of integrated pulses (An)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 31 / 33

Page 169: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Detector operation

Li metal

P-doped+Vbias

Vout

A A A A1 2 3 4pile up

icr

ocr

output current is integrated into voltage pulses by a pre-amp, whenmaximum voltage is reached, output is optically reset

fast channel “sees” steps and starts integration; pulse heights are measuredwith slow channel by integration; and detected pulse pileups are rejected

electronics outputs input count rate (icr), output count rate (ocr), andareas of integrated pulses (An)

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 31 / 33

Page 170: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Dead time correction

Input Count Rate (kcps)

Ou

tpu

t C

ou

nt

Ra

te (

kcp

s)

0 50 100 150 2000

50

100

150

200

actual ocr

ideal ocr

the output count rateis significantly lowerthan the input countrate and gets worsewith higher photonrate

if these curves areknown, a dead timecorrection can be ap-plied to correct for roll-off

if dead time is toolarge, correction willnot be accurate!

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 32 / 33

Page 171: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Dead time correction

Input Count Rate (kcps)

Ou

tpu

t C

ou

nt

Ra

te (

kcp

s)

0 50 100 150 2000

50

100

150

200

actual ocr

ideal ocr

the output count rateis significantly lowerthan the input countrate and gets worsewith higher photonrate

if these curves areknown, a dead timecorrection can be ap-plied to correct for roll-off

if dead time is toolarge, correction willnot be accurate!

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 32 / 33

Page 172: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Dead time correction

Input Count Rate (kcps)

Ou

tpu

t C

ou

nt

Ra

te (

kcp

s)

0 50 100 150 2000

50

100

150

200

actual ocr

ideal ocr

the output count rateis significantly lowerthan the input countrate and gets worsewith higher photonrate

if these curves areknown, a dead timecorrection can be ap-plied to correct for roll-off

if dead time is toolarge, correction willnot be accurate!

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 32 / 33

Page 173: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

Dead time correction

Input Count Rate (kcps)

Ou

tpu

t C

ou

nt

Ra

te (

kcp

s)

0 50 100 150 2000

50

100

150

200

actual ocr

ideal ocr

the output count rateis significantly lowerthan the input countrate and gets worsewith higher photonrate

if these curves areknown, a dead timecorrection can be ap-plied to correct for roll-off

if dead time is toolarge, correction willnot be accurate!

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 32 / 33

Page 174: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

SDD spectrum

0 5000 10000 15000

Energy (eV)

100

1000

10000

Counts

fluorescence spectrumof Cu foil in air using9200 eV x-rays

Compton peak is visi-ble just above the CuKα fluorescence line

Ar fluorescence near3000 eV

a small amount ofpulse pileup is visiblenear 16000 eV

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 33 / 33

Page 175: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

SDD spectrum

0 5000 10000 15000

Energy (eV)

100

1000

10000

Counts

fluorescence spectrumof Cu foil in air using9200 eV x-rays

Compton peak is visi-ble just above the CuKα fluorescence line

Ar fluorescence near3000 eV

a small amount ofpulse pileup is visiblenear 16000 eV

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 33 / 33

Page 176: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

SDD spectrum

0 5000 10000 15000

Energy (eV)

100

1000

10000

Counts

fluorescence spectrumof Cu foil in air using9200 eV x-rays

Compton peak is visi-ble just above the CuKα fluorescence line

Ar fluorescence near3000 eV

a small amount ofpulse pileup is visiblenear 16000 eV

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 33 / 33

Page 177: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

SDD spectrum

0 5000 10000 15000

Energy (eV)

100

1000

10000

Counts

fluorescence spectrumof Cu foil in air using9200 eV x-rays

Compton peak is visi-ble just above the CuKα fluorescence line

Ar fluorescence near3000 eV

a small amount ofpulse pileup is visiblenear 16000 eV

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 33 / 33

Page 178: Today’s outline - January 28, 2020phys.iit.edu/~segre/phys570/20S/lecture_05.pdf · Today’s outline - January 28, 2020 Undulator harmonics Undulator coherence Emittance Time structure

SDD spectrum

0 5000 10000 15000

Energy (eV)

100

1000

10000

Counts

fluorescence spectrumof Cu foil in air using9200 eV x-rays

Compton peak is visi-ble just above the CuKα fluorescence line

Ar fluorescence near3000 eV

a small amount ofpulse pileup is visiblenear 16000 eV

C. Segre (IIT) PHYS 570 - Spring 2020 January 28, 2020 33 / 33