Page 1: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

The new inflatable external lower leg compression improves stroke volume and

peripheral perfusion during central hypovolemia in healthy volunteers

M. Helmi, A. Lima, D. Gommers, J. Bakker, J. van Bommel.

Page 2: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Shock

Page 3: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

HypotensionTissue hypoperfusionEnd-organ damage

Shock

Page 4: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

To reverse hemodynamic alterations

Fluid overloading

Fluid resuscitation

Page 5: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

External leg compression

Reverse hypotensionNo risk of fluid overloading

Use in intensive care ??

Page 6: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers
Page 7: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Hemodynamic effectHemodynamic effectExternal leg compression

Venous

Emptying

Venous return Venous pressure

Soft tissue blood flow Tissue oxygenation

Arterial

Arteriovenous gradient

Stroke volume Cardiac output

Mean arterial pressure Heart rate

Page 8: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

What is the potential effect of ELC around the legs during mild hypovolemia on systemic hemodynamic and peripheral perfusion parameters?

Question

?? ?

Page 9: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Prevent central hypovolemia when applied before HUT

The application of a new inflatable ELC around the upper and lower legs is able to:

Hypothesis

Restore central hypovolemia when applied after HUT.

!!

Page 10: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

• 20 healthy volunteers:– Free of cardiovascular or neurological disease– Not taking any medication

Subjects

Page 11: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

External leg compression

• Around upper and lower legs

• Inflation pressure can be set at levels of 0, 20, 40 and a maximum of 50 cmH2O

Page 12: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

The head up tilt

Page 13: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

One maneuver

Page 14: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Measurements• Global hemodynamic with Finometer®

• Perfusion index (PI) with Masimo®

• Tissue oxygen saturation (StO2) with InSpectra®

Page 15: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Each intervention was done in three minutes

To find the ideal ELC compression pressure (the inflation pressure related to the best SV and HR values).

SUP

Pre-Study

HUT HUT50 HUT40 HUT20

Page 16: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Supine HUT0 HUT50 HUT40 HUT20

• SV, mL 86±16 62±14 70±19 67±19 65±18• HR, bpm 71±13 85±12 77±10 79±9 82±11

Results pre-study

Page 17: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Each intervention was done in three minutes

Study protocol

Page 18: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Control model

SUP

HUT SUP’

Page 19: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Prevention model

SUP50 HUT50 SUP50’

Prevent hemodynamic alteration.-Mechanical ventilation-Hemodialysis

Page 20: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Restore model

SUP HUT50 HUT’

Restore hemodynamic alteration.(Circulatory shock)

HUT

Page 21: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

*: P < .05 against previous time point.

Global HemodynamicsControl model Prevention model

*

*-27%

**-17%

*

*+19%

* *+9%

Page 22: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

*: P < .05 against previous time point.

*

**

Restore model

+24%

-9%

Page 23: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

*: P < .05 against previous time point.

Control model Prevention model

*

*

*

*

*

**

*

Peripheral perfusion

-79% -65%

-9% -4%

Page 24: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

*: P < .05 against previous time point.

*

**

Restore model

+117%*

+30%

Page 25: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

It may be used to prevent or to restore hemodynamic alterations in critically ill patients.

ELC was shown to be able to prevent and restore hemodynamic alterations during HUT.

ELC produced an auto transfusion effect in normovolemic subjects in a 60o HUT at 50 cm H2O compression pressure.

Conclusion

Page 26: The new inflatable external lower leg compression improves stroke volume and peripheral perfusion during central hypovolemia in healthy volunteers

Thank you!!!