tek1psqt manual

19
PURE STEAM Pure Steam Quality Test Kit TEK1pSQT Operator’s Manual

Upload: kp

Post on 15-Nov-2015

212 views

Category:

Documents


0 download

DESCRIPTION

MANUAL

TRANSCRIPT

  • PURE STEAM

    Pure Steam Quality Test Kit

    TEK1pSQT

    Operators Manual

  • Table of Contents

    INTRODUCTION ........................................................................................................................ 4

    ADDITIONAL EQUIPMENT REQUIREMENTS ..................................................................... 5

    STEAM TEST POINTS ............................................................................................................... 6

    ASSEMBLING THE STEAM TEST KIT ................................................................................... 7

    NON-CONDENSABLE GAS KIT ........................................................................................... 7

    SUPER-HEAT TEST KIT ........................................................................................................ 7

    DRYNESS TEST KIT .............................................................................................................. 8

    NON CONDENSABLE GAS TEST ............................................................................................ 9

    S E T T I N G U P .................................................................................................................... 9

    P E R F O R M I N G T H E T E S T. .10

    A C C E P T A N C E C R I T E R I A .................................................................................. .11

    THE SUPERHEAT TEST .......................................................................................................... .12

    S E T T I N G U P .................................................................................................................. .12

    P E R F O R M I N G T H E T E S T ...................................................................................... .13

    A C C E P T A N C E C R I T E R I A .................................................................................. .13

    DRYNESS TEST........................................................................................................................ .14

    S E T T I N G U P ................................................................................................................... .14

    P E R F O R M I N G T H E T E S T ...................................................................................... .15

    A C C E P T A N C E C R I T E R I A .................................................................................. .16

    REPORT FORMAT.............17

    HEALTH AND SAFETY ........................................................................................................... .18

    CARE AND MAINTENANCE .................................................................................................. .19

  • File (Manual, SOPs, Calibration certificate,

    IQ/OQ)

    Carrying case

    Non condensable gas

    unit

    Laboratory balance

    Steam sampling (service) pipe

    Gloves

    Temperature meter & thermocouple

    Storage boxes Pitot tubes (0.4mm,

    0.6mm, 0.8mm)

    Sanitary diaphragmatic valve

    Gaskets (spare)

    TC clamp

    Expansion tube

    1 liter vacuum flask

    (with stand, rubber

    bung assembly and

    SS inlet tubes)

    SS hose pipe

    with TC end

    fitting

    Rubber tube

    Water pipe

  • 4

    INTRODUCTION

    Thank you for choosing the TEK1 Pure Steam Quality Testing Kit (TEK1

    pSQT).This kit is developed based on extensive practical experience and

    has been designed to provide reliable and consistent results that conform to

    the requirements of EN285/HTM2010/ISO11134.

    The Non-condensable gas test method exceeds the HTM 2010 standard

    equipment and also less dependent on operators skill. User will find the

    test equipment provides very consistent results.

    The Dryness test is similar to that described in EN285/HTM 2010, but it is

    modified to make the measurements easier and it is ready to test kit.

    The Superheat test equipment is identical to that in EN285/HTM2010.

    Warning

    Before conducting any tests the contents of this manual must be studied and

    any associated hazards must be considered. It is stressed that the test

    methods prescribed in this manual is defined in EN285 and HTM 2010 not

    by author. EN285 and HTM 2010 are the primary reference document for

    this manual. Sterilizer plant rooms are potentially hazardous areas and the

    test involve live steam. It is assumed by the author of this manual that staff

    performs the tests must be trained professional and can understand the

    potential of danger.

  • 5

    Warranty

    The TEK1pSQT is designed and manufactured to the highest standards using

    top quality material. A one year warranty * is provided.

    * Warranty is applicable for part only. Warranty will not be applicable if misuse or its

    use outside of the operating parameters defined within the manual.

    ADDITIONAL EQUIPMENT REQUIREMENTS

    To enable you to perform steam quality tests, additional equipments requirements are:

    Running cold water supply or bucket type water storage container may also be

    used.

    Extension pipe for water supply

  • 6

    STEAM TEST POINTS

    In order to test the steam quality, specific test points as per the guide line on the steam

    line are required.

    Fig. (i) Illustrate the location of the three test points on the steam supply pipe and they

    are fitted between the steam distribution system and sterilizer. It is expected that the

    pressure at this point would be 2-5bar.

    Fig: (i)

    Note: (i) It is important that the pitot tube entry point is level and parallel with the

    steam pipe as any deviation towards the edge of the pipe can influence the results

    detrimentally.

    (ii) All the entry points are 1/2" BSP female sockets with full size holes drilled

    through the pipes.

    Socket for pitot tube for Dryness and

    Superheat test

  • 7

    ASSEMBLING THE STEAM TEST KIT

    NON CONDENSABLE GAS KIT

    Non condensable gas detection unit

    Hose pipe with TC end fitting

    Cold water supply (user to provide)

    Radiant Flow Valve

    Cooling Water Inlet pipe

    Cooling Water Outlet pipe

    Steam inlet

    Drain Valve Fig: (ii)

    SUPER-HEAT TEST KIT

    Superheat expansion tube

    Pitot Tubes(0.8mm,0.6mm & 0.4mm)

    Temperature probe entry gland

    Dual temperature meter

    Sampling (service) pipe

    Fig: (iii)

  • 8

    DRYNESS TEST KIT

    Pitot Tube (0.8mm, 0.6mm & 0.4mm)

    Rubber tubing

    Rubber bung and SS tube assembly

    1 Liter vacuum flask

    Weighing balance

    Dual temperature meter

    Sampling (service) Pipe

    Fig: (iv)

  • 9

    NON CONDENSABLE GAS TEST

    SETTING UP

    Assemble equipment together as illustrated in Fig. (v) place on table top. Check that the

    steam valve on the test equipment is fully closed and that the cooling water valve is fully

    open.

    Fig: (v) Fig:(vi)

    Connect the cooling water supply pipe to the running water source at one end and the

    cooling water inlet on the condensing unit at the other (Fig. v).Connect another pipe from

    the cooling water outlet to drain or a bucket. Please note that the water from the water

    outlet may be very hot and care should be taken to ensure that it can run to drain without

    blockage and spillage.

    Isolate the steam supply and after checking that no residual pressure is present, connect

    the hose pipe for steam supply tubing to the non condensable gas take off point steam

    control valve. Take care, as the steam pipe may be hot and residual steam may be

    present. Connect the other end of the hose tube to the steam connection of condensing

    unit and turn the steam supply back on. Carefully open the Steam control valve a small

    amount. While it is insulated, this is to protect the user against casual contact. Even a

    short contact time will result in a burn, unless the necessary precautions have been taken.

    Needle valve

    Drain

    Water Inlet

    Water Outlet

  • 10

    Do not block the flow of cooling water .The cooling water outlet must be kept clear at all

    times.

    PERFORMING THE TEST

    Turn on the water supply to supply the cold water to the cold water inlet. After ensuring

    that the cooling water is leaving the outlet pipe, the main steam valve may be slowly

    opened. Live steam and boiling water could discharge from the condense collecting

    chamber if insufficient cooling water is available may present a hazard. In any event,

    precautions should be taken. Do not look into collecting chamber and wear eye

    protection.

    Fill the condense collection chamber with water either by filling with water from an

    external source, or allowing condense to accumulate.

    Water level should be in Zero position

    Ensure that the sterilizer chamber is empty except the normal furniture etc. select a

    Porous load / equipment cycle and start a run.

    When the steam supply to chamber first opens, ensure the measuring cylinder is empty,

    by emptying it.

    Any non- condensable gases present in the steam being sampled will rise to the top of the

    Collection tube when closed the valve.

    When at least 100 ml of condense has been collected in the measuring cylinder note the

    volume of gas collected in the non condensable gas collection tube (V1) and the volume

    of water collected in the measuring cylinder(V 2).

    Calculate the fraction of non- condensable gases as a percentage as follows:

    Percentage of non condensable gas = 100 X (V1/V2).

  • 11

    ACCEPTANCE CRITERIA

    The test should be considered satisfactory if the fraction of non condensable gases does

    not exceed 3.5%.

    The test should be done three times in total to check consistency. If the results of the

    tests differ significantly, then the cause should be investigated before proceeding further.

    NOTE: Calculations are included on an enclosed pen drive.

  • 12

    THE SUPERHEAT TEST

    SETTING UP

    Fig:(vii)

    Isolate the steam supply and after checking that no residual pressure is present, insert a

    temperature sensor entry gland into the fitting on the steam pipe. The steam pipe may

    be hot and precautions should be taken against both burning and the presence of

    residual steam. The temperature probe should be at the geometric centre of the steam

    pipe. Insert pitot tube into the steam supply. Turn the steam supply back on, taking the

    necessary precautions against burning/scalding from the steam that will issue from the

    pitot tube.

    Insert temperature sensor through gland in expansion tube and position it so the sensing

    point of the thermocouple element is in the geometric centre of the expansion tube.

    Push the expansion tube onto the pitot tube taking the necessary precautions. This will

    require the tester to be in close proximity to the steam issuing from the pitot tube and

    extreme care is required to avoid scalding and/ or burning. Gloves, overalls and eye

    protections must be worn.

  • 13

    PERFORMING THE TEST

    Select and start a porous load / equipment cycle.

    From the measured temperature, note the temperature in the steam service pipe (for use

    in the dryness test) and in the expansion tube (T e) when the steam supply to the chamber

    first opens.

    Calculate the superheat in o C from the following equation. Superheat = T e - T o

    Where T o is the boiling point of water at local atmospheric pressure.

    ACCEPTANCE CRITERIA

    The temperature is considered to be acceptable if it is less than 25 C above that of

    the local temperature of boiling water

  • 14

    DRYNESS TEST

    SETTING UP

    Assemble the apparatus as per Fig. (viii)

    Fig: (viii) Fig: (ix)

    Isolate the steam supply and after checking that no residual pressure is present, insert a

    temperature sensor entry gland into the fitting on the steam pipe. The steam pipe may be

    hot and precautions should be taken against both burning and the presence of residual

    steam. The temperature probe should be at the geometric centre of the steam pipe. Insert

    pitot tube into the steam supply. Turn the steam supply back on, taking the necessary

    precautions against burning/scalding from the steam that will issue from the pitot tube.

  • 15

    PERFORMING THE TEST

    Important

    Before performing the test ensure the pitot hole is not restricted as this will lead to false

    results. This is best done prior to fitment for safety reasons. A long heat up time for the

    flask reaching ~ 80 o C is not normal. It should take no more than 5 minutes.

    Weigh the whole assembly including pipe and clips and note the weigh (ME).

    Remove the stopper and tube assembly and pour 650 50 ml of cold water (below 27 o

    C) into the flask. Replace the stopper and tube assembly, weigh the flask and record the

    mass (MS).

    Support the flask close to the pitot tube taking care to avoid the issuing steam. Ensure

    that the rubber tube and flask are protected from excess heat and draughts. Do not

    connect it to the pitot tube yet.

    Introduce the second temperature sensor through the shorter of the two pipes into the

    water in the flask. Agitate the flask and note the temperature of the water in the flask

    (T1).

    Ensure the sterilizer is empty for the usual chamber furniture. Select and start a porous

    load / equipment cycle.

    When the steam supply valve to the chamber first opens, connect the rubber tube to the

    pitot tube. This will require the tester to be in close proximity to the steam issuing from

    the pitot tube and extreme care is required to avoid scalding and/ or burning. Gloves

    overalls and eye protection must be worn.

    Observe / record the steam temperature for the duration of the test and on completion of

    the test calculates the average temperature (T2).

    T3 = Temperature of dry saturated steam in C

  • 16

    Weigh the flask and stopper assembly including pipe and clips and note the mass (MF).

    Calculate the dryness valve by using the following formula.

    D = [(T1-T2) (CPW (MS-ME) +A] / [L (MF-MS)] - [(T3-T2) CPW] / [L]

    Where

    D = Dryness value

    L = Latent heat of the dry saturated steam at T3 temperature in kj/kg.

    ME = Mass of flask in Kg.

    MS = Mass of (flask + water charge) in Kg.

    MF = Mass fo (flask + water + condensate) in Kg.

    T1 = Initial Temperature of water in flask, in C

    T2 = Final temperature of water and condensate in the flask, in C

    T3 = Temperature of dry saturated steam in C

    CPW = Specific heat capacity of water (4.18 kj/kg. K)

    A = Effective heat capacity of apparatus (0.24 kj/K)

    Note. The formula has been modified for stainless steel flasks and dip tubes instead of

    glass, as detailed in HTM 2010.

    ACCEPTANCE CRITERIA

    The dryness value is not less than 0.90 (If metal loads are processed, the dryness value

    should not be below 0.95).

    Throughout the operating cycle, the temperature measured in the steam service pipe is

    within 3o C of that measured during the superheat test.

    NOTE: Calculations are included on an enclosed pen drive.

  • 17

    REPORT FORMAT

    Fig: (x)

  • 18

    HEALTH AND SAFETY

    Care should always be taken when working on or near steam pipes due to the very high

    temperatures involved. Thermal gloves, overalls that cover arms and eye protection must

    be used.

    While every effort has been made to provide protection from the extreme temperatures,

    care should be taken when testing with equipment.

    Do not plug or restrict the Cooling out flow as this may pressurize the heat exchanger,

    which it has not been designed for.

    Do not look into the condense collection cylinder while the steam valve is open, as hot

    condense and / or steam could be ejected.

    The insulation surrounding the steam pipe is made from glass fiber and can cause

    irritation to the skin in some instances. It is therefore recommended that gloves or a good

    quality barrier cream be applied as precaution before handling the material. The

    insulation is designed to prevent burns from transient contact with the pipe and is not

    designed to protect against prolonged contact.

  • 19

    CARE AND MAINTENANCE

    The Hose Pipe, will with bending, become work hardened, which will lead to difficult in

    returning it to the tight radius required for storage. In this Instance, it may be easily

    replaced.

    Cleaned and dried the Equipments after completion of the test.

    Electronic equipments should protect from water/steam.