sudden jumps in the absolute energy scale and their origin

40
K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale EPIC Calibration Meeting, Mallorca Sudden jumps in the Sudden jumps in the absolute energy scale absolute energy scale and their origin and their origin

Upload: august-hendrix

Post on 04-Jan-2016

16 views

Category:

Documents


0 download

DESCRIPTION

Sudden jumps in the absolute energy scale and their origin. FF Mn-K Pos Quadrant 0. ± 5 eV. Effect of the particle background. FF Mn-K Pos Quadrant 0. + 41.5 eV. ± 5 eV. 2005 Dec 20 – 2006 Jan 01. FF AL-K Pos Quadrant 0. ± 5 eV. + 10.5 eV. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Sudden jumps in the Sudden jumps in the

absolute energy scale and absolute energy scale and

their origintheir origin

Page 2: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

FF Mn-K Pos Quadrant 0

± 5 eV

Page 3: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Effect of the particle Effect of the particle

backgroundbackground

Page 4: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

FF Mn-K Pos Quadrant 0

± 5 eV

2005 Dec 20 – 2006 Jan 01

+ 41.5 eV

Page 5: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

FF AL-K Pos Quadrant 0

± 5 eV

2005 Dec 20 – 2006 Jan 01

+ 10.5 eV

Page 6: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Al-K: 1.486 keV, +10.5 eV +0.707%

Mn-K: 5.894 keV, +41.5 eV +0.704% Change in the gain,

not in the offset

Page 7: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Energy resolution somewhat improved during high background

reduced charge loss noise reduced charge loss

increased trap saturation

Energy shift occurs also for first singles

precursors ‘invisible’

precursors below lower threshold or in preceding frame(s)

low background

high background

Page 8: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

rev 711, high background

Al-K

Mn-K

Al-K

Mn-K

The CTI during low and high background

rev 709, low background

decreased CTI during high

background (Mn-K)

Page 9: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

low

back

gro

un

d

Page 10: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

hig

h

back

gro

un

d

Page 11: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

During episodes of high particle background the CTI is reduced.

This is most likely caused by saturation of traps due to precursors.

The charge of the precursors is below the 20 adu threshold and thus not available.

The usual CTI correction cannot be used.

But there may be a correlation with the background flux which can be used for the

correction..

Page 12: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Effect of the quadrant box Effect of the quadrant box

temperaturetemperature

Page 13: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Gain/temperature correlation: earlier investigations

Page 14: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα & Al-K position and quadrant box temperature

Page 15: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Quadrant box temperatures

(yellow circles: no FF mode)

Page 16: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα position vs. temperature: CCD 1 & 2

Page 17: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα position vs. temperature: CCD 3 & 4

Page 18: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα position vs. temperature: CCD 5 & 6

Page 19: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα position vs. temperature: CCD 7 & 8

Page 20: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα position vs. temperature: CCD 9 & 10

Page 21: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα position vs. temperature: CCD 11 & 12

Page 22: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα position vs. Al-K position

Page 23: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Deviations of the Al-K and Mn-Kα positions

Page 24: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα position vs. Al-K position

Page 25: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα positions (after gain and CTI correction)

Page 26: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mn-Kα positions after temperature (and dCTI/dt) correction

Page 27: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

“CCD offsets”

Page 28: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Doy 2000

Quadrant box temperatures

Page 29: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Page 30: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Page 31: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Standard correction

Page 32: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

with 1 adu / 2000 d drop

Page 33: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

with 1 adu / 2000 d drop

Page 34: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Mean quadrant box temperature [C]

(F1576..F1876)

Mn

-Kα p

osi

tion

[a

du

]

(in q

uadra

nt

0, aft

er

1 a

du /

20

00

d d

rop)

slope: +

0.43 adu / C

Page 35: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

standard correction

Page 36: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

after temperature.. correction

Page 37: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

standard correction

Page 38: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

after temperature.. correction

Page 39: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca

Multiply all energies with a factor F:

F = ( 1 - A(t) - B(Tq) ) C

A = 4.24 * 10-7 (t-t0)/[d]

B = 3.65 * 10-4 Tq/[oC]

C = 1.0057

t0 : 2000-Jan-01

Tq: mean of T(F1576)..T(F1876)

Page 40: Sudden jumps in the absolute energy scale and their origin

K. Dennerl, 2006 October 26EPIC pn: Jumps in the energy scaleEPIC Calibration Meeting, Mallorca