spectro q5800 - photos.labwrench.com
TRANSCRIPT
The Q5800 is a multi-purpose, portable measurement tool combining abnormal wear
metals analysis, particle counting, viscosity and IR spectroscopy into a compact,
field-based system. The battery-powered device is an ideal tool for maintainers of
mobile, high cost equipment who need a complete lubricant assessment for condition
monitoring and immediate results for time-based decisions. Its rugged design
accommodates a variety of environments requiring transport in the field.
Q5800 ComponentsThe Q5800 is comprised of the following four major modules and the associated
technology.
Features and Benefits•Solventfreewithminimalsamplewaste
•Testsalltypesoflubricatingfluidsand
asset compartments
•Accuratelydetectspotentialfailures
before they happen
•Easy-to-use,integratedsoftwarewith
touch screen
•Easytotransportwithacustombackpack
designed to move the device and all
accessories
•Ruggeddesignforfielduse
Critical asset owners in the military,
mining and marine industries need
immediate information about oil and
equipment condition. The portable
Q5800 eliminates the need to wait for
lab results, miss early signs of abnormal
wear and use messy hazmat test kits.
This integrated, mobile technology is the
ideal device for maintaining readiness
and lowering costs.
Spectro Q5800 EXPEDITIONARY FLUID ANALYSIS SYSTEM
The Q5800 brings mobile fluid sampling to the equipment operator in the field.
FLUID CHEMISTRY VISCOSITY PARTICLE COUNT ELEMENTAL ANALYSIS
FluidScan® Q1000 infrared spectrometer with fliptopcelldesign;testsforTAN/TBN,water content, soot, oxidation, mixed up fluids
SpectroViscQ3000Serieskinematicviscometer @40°C (cSt);solvent-free;low sample volume needed
Fitration Particle Quantifier(FPQ);solvent-free particle counting;handlesthe dirtiest and wettestsamples;filter patch for evaluation of debris
Wear metals analysis for abnormal wear conditions and contaminants using X-ray Fluorescence (XRF)technology
Spectro Q5800EXPEDITIONARY FLUID ANALYSIS SYSTEM
SPECTRO, INC. • One Executive Drive, Suite 101, Chelmsford, MA 01824-2563 • 978-431-1120 • www.spectroinc.com • [email protected] • An ISO 9001:2008 company
Spectro Q5800 Specifications
Copyright©2013SpectroInc.Allrightsreserved.Whileeveryeffortismadetoassuretheinformationinthisdocumentisaccurate,Spectrodoesnotacceptliabilityforanyerrorsormistakesthatmayarise.Specificationsaresubjecttochangewithoutnotice. Q5800_DSv1_2013-09-25
PRODUCT INFORMATION
Part Number Q5800EFA
Applications Mineral and synthetic lubricants including gear, engine, transmission, hydraulics, turbine as well as military, marine and mining applications.
Output Elemental:Silicon(Si);Aluminum(al);Chromium(Cr);Titanium(Ti);Iron(Fe);Nickel(Ni;Lead(Pb);Copper(Cu);Tin(Sn);Molybdenum(Mo);Silver(Si);Zinc(Zn);Vanadium(V)
Fluid Chemistry:TAN&TBN(mgKOH/g);Oxidation(abs/mm2);Nitration(abs/mm2);Sulfation(abs/mm2);Water(partspermillion);Glycol(%byweight);Soot(%byweight);Incorrectfluid(%byweight)AntioxidantDepletion(%remaining);AntiwearDepletion(%byweight)
Viscosity:@40°C(cSt)Particle Count: >micron(p/ml);soliddebris>4micronson filter
Methodology ModASTME2412(IR);ModASTMD7279(visc); ModISO21018-3(ParticleCount)
Accuracy <+15%ofmeasuredvalue,calibrated
Repeatability <+15%ofmeasuredvalue,typical
Test Temperature Control(Viscosity Measurement)
+ 0.1°C
Calibration Factory per module
Check Standards Provided
ELEMENTAL SYSTEM
Detector 10 mm2SDDDetector;peltiercooled
Resolution 145eVat10,000cps
Excitation Source X-raytubewithRhodiumtarget;maxvoltage40kV
Calibration Factory preset per application
OPERATIONAL SPECIFICATIONS
Sample Volume 10 ml
Sample Time Lessthan7minutes
Solvents/Reagents None
Ambient Operating Temperature
0°C to +40°C
Relative Humidity 10%to80%r.h.non-condensing
Ambient Altitude Upto5,000meters(16,404feet)
USER INTERFACE SPECIFICATIONS
Software/Operating System
Windows®CE(built-in)
Display Fixed angle color touchscreen display
Data Storage Internalflashmemory(SDCardExpansion)
Data Transfer Ethernet,USB
Security Password protected
Data Entry Touchscreendesktopsoftware
Communication WIFI(optional);Bluetooth(optional)
POWER REQUIREMENTS
Battery Power Source Exchangeablelithium-ionbatterypack
Charge Power AC110/240V,50/60Hz,10watts
Typical Runtime 4-6 hours
Recharge Time 2.5 hours
MECHANICAL SPECIFICATIONS
Dimensions 487.68mm(L)x386.08(W)x228.6(H);19.2”x15.2x9
Weight 16.5kg(36.4lbs)
COMPLIANCE
RoHS,CE,FCC,CSA,us.IEC61010-1:2001/EN61010-1:2001,CE,cTUV.IEC61010-1:2002
The easily transportable Q5800 is ideal for on-site analysis.