Simulation of a PET-like detector for the ballistic control of a Pencil Beam Scanning (PBS) proton beam
Arthur BONGRAND&
Nicolas FISICHELLA
11/05/2017
OpenGate Meeting
2
1. Context
2. Geometry
3. Sources
4. Challenges
Arthur BONGRAND
Hadrontherapy requires a very precise positioning of Bragg Peak (incident beam range)
Need to monitor dose in real time and how it is deposited in tissues
AdvantagesMore precise ballistic + RBE HADRONS (12C) > PHOTONS
β+ isotopes
Prompt γ (up to ~ 10MeV)
11/05/2017 3
Context
Arthur BONGRAND
Target Produced β+ isotopes T1/2
C 𝟏𝟎𝑪, 11𝐶,𝟖𝑩
19.29 s, 20.33 min, 770 ms
N 13𝑁 9.96 min
O 𝟏𝟒𝑶, 15𝑂 70.61 s, 122.24 s
(Os…) P , Ca 30𝑃, 38𝐾 2.50 min, 7.64 min
11/05/2017
4
K. Parodi et al., IEEE MIC CR, 2002Carbon Beam 212 MeV – Proton Beam 110 MeV
• Produced β+ isotopes
Gold objective
Online activity profile along Beam direction
• Origin of β + formation-> Inelastic collisions: Proton Beam: target fragment only
• β + activity and dose deposition correlationProton Beam: no "trivial" correlation
Context
Arthur BONGRAND
𝜷+ + 𝒆− −> 𝟐 γ𝟓𝟏𝟏 𝒌𝒆𝑽
11/05/2017 5
Context
6
1. Context
2. Geometry
3. Sources
4. Challenges
11/05/2017 Arthur BONGRAND 7
Final aim :Simulation of complete
treatments
240x LYSO + PMT R1450 Hamamatsu
Geometry
Nicolas Fisichella
Arthur Bongrand
LAPD Large Acceptance Pixelized
Detector
11/05/2017 Arthur BONGRAND 8
Trapezium
Z shift
Geometry
11/05/2017 Arthur BONGRAND 9
Problem when using repeater function->Doesn’t work for trapezium + Z shift
Not convenient solution -> Manual positioning of each trapezium
Geometry
10
1. Context
2. Geometry
3. Sources
4. Challenges
11/05/2017 Arthur BONGRAND 11
Sources
PBS
2 sources
LYSO
Mono energy
Proton Beam
A = 700 Bq / Crystal
Image de la ligne de
recherche
MEDICYC® 65MeVResearch Beam
Proteus®ONE 230 MeV
12
1. Context
2. Geometry
3. Sources
4. Challenges
11/05/2017 Arthur BONGRAND 13
Challenges
LYSO
#1 Analyze each event according to the global time
Beam data
Pulse Pile-up
Saturated Interval
11/05/2017 Arthur BONGRAND 14
Proteus®ONE 230 MeV
Challenges
Blind Golfer PBS
Each layer is processed in several "passes“ : (3)*
Pass 1: Most of the dose is delivered in the voxels
Pass 2 and pass 3 : complete the dose
Note: pass 3 = minor correction
*Padro J. et al., « Heuristic optimization of the scanning path of particle therapy beams », Med. Phys. 36
2043–51 (2009)
Pass 3
Pass 2
Pass 1
Dose/bin- +
11/05/2017 Arthur BONGRAND 15
Proteus®ONE 230 MeV
Challenges
Pass 3
Pass 2
Pass 1
Dose/bin- +
Blind Golfer PBS
In each slice, the succession of the delivered spots is done
according to time with a
complex spatial distribution
Input : “Log file“
For each spot:
X, Y position at isocenter energy (=range)
weight (monitor units, total charge, number of protons)
start time end time
11/05/2017 Arthur BONGRAND 16
Proteus®ONE 230 MeV
Challenges
MEDICYC® 65MeVResearch Beam
Arthur BONGRAND11/05/2017 17
Simulation of multi-systems (hybrid) (β+, prompt γ…)
Comparison Beam data / Gate -> Publications
Arthur BONGRAND11/05/2017 18
Arthur BONGRAND11/05/2017 20
Clinic Setup
- LAPPD detection system Patent in progress
LAPPD (MCP) Christophe INSA
11/05/2017 Arthur BONGRAND 21
Arthur BONGRAND
Vers des systèmes multimodales (hybrid) (β+,γ) ?
11/05/2017 22
C Lang et al 2014 JINST 9 P01008
11/05/2017 Arthur BONGRAND 23
Arthur BONGRAND
5 systèmes TEP
Ère des essais (en condition) cliniques
Modulation temporelle du faisceau
Vers des systèmes multimodales (hybrid) (β+,γ) ?
Construction de systèmes d’acquisitions “temps réel”
Controle temps réel du plan de traitement ?
11/05/2017 24
Arthur BONGRAND
TEP “hors salle”
TEP “en salle”
TEP “en ligne”
BASTEI (1997-2008)
Prérequis indispensables
Rejet du bruit de fondHaute sensibilité
Champ de vue
TOF
Synchronisationfaisceau / acquisition
des données Modulation
temporelle du faisceau
11/05/2017 25
Arthur BONGRAND
OpenPETDoPET-L INSIDE
TOF PETDPC
DPGA
Open Close
Scintillateur + PM
Sensibilité(efficacité x acceptance)
Algo
ResolutionSpatiale (mm)
TOF
Faisceau
Online27
11/05/2017
Scintilateur
Zr-doped GSO (GSOZ) 2.8 × 2.8 × 7.5 mm3
4-layergrille 16 × 16
PMT H8500 (Hamamatsu)
Close Open
11/05/2017 28Hideaki Tashima et al 2016 Phys. Med. Biol. 61 1795
Arthur BONGRAND
29
Structure temporelle du faisceau :3.3s beam on (in-spill)
1.8s beam off (inter-spill) Intensité 5.106 carbon par spill
Reconstruction : 3D LM MLEM après irradiation3D OSEM en cours de développement
Hideaki Tashima et al 2016 Phys. Med. Biol. 61 179511/05/2017 Arthur BONGRAND
OpenPET
Open Close
Scintillateur + PM GSOZ / PMTmulti-anodes
Sensibilité(efficacité x acceptance)
5.1% 7.3%
Algo 3D LM-MLEM
ResolutionSpatiale (mm)
2.6±0.2
2.1±0.1
TOF Non
Faisceau11C (340 MeV)
2.5Gy
Online Non30
11/05/2017 Arthur BONGRAND
11/05/2017 31G. Sportelli et al 2016 JINST 11 C02089
Aire de détection : 2 detecteurs opposés (15x15 cm2) placés à 30 cm l’un de l’autre.
9 blocs
Composition d’un bloc :23x23 LYSO pixel de 2x2x16 mm
PMT H8500 Hamamatsu
Arthur BONGRAND
11/05/2017 32
RANDO
G. Sportelli et al 2016 JINST 11 C02089Arthur BONGRAND
11/05/2017 33
Y
X
Z
X
Z
Y18-FDG
G. Sportelli et al 2016 JINST 11 C02089Arthur BONGRAND
Arthur BONGRAND
OpenPETDoPET-L
Open Close
Scintillateur + PM GSOZ / PMTmulti-anodes
LYSO / PMTmulti-anodes
Sensibilité(efficacité x acceptance)
5.1% 7.3%?
FOV10x16x16 cm3
Algo 3D LM-MLEM 3D MLEM
ResolutionSpatiale (mm)
2.6±0.2
2.1±0.1
?
TOF Non Non
Faisceau11C (340 MeV)
2.5Gy
P (98.3 MeV)
C (178-260 MeV)
2 Gy
Online Non Non34
11/05/2017
Aire de détection : 2 modules opposés (5.12 * 5.12 cm2)
placés à 50 cm l’un de l’autre
Composition d’un module :Grille de 16 × 16 pixels (3.2 × 3.2 x 20 mm3)
Scintillateur et PMs associés
Cristaux de Lutetium Fine Silicate (LFS)couplés un par un à 16 × 16 SiPM
(Hamamatsu)
11/05/2017 35M A Piliero et al 2016 Phys. Med. Biol. 61 N650
DOSE PROFILER (gamma prompt)Tracker + Calorimètre
PET System (β+ )
Arthur BONGRAND
Test d’irradiation :
Fantôme de PMMA (4.9 × 4.9 x 7 cm3)Placement parrallèle à l’axe du faisceau
Type d’irradiation :
Proton 68 et 72 MeV
CNAO synchrotron Structure temporelle du faisceau :
1s beam on (in-spill) 4s beam off (inter-spill)
Intensité (1 − 2). 109 protons par spill
11/05/2017 36M A Piliero et al 2016 Phys. Med. Biol. 61 N650
Arthur BONGRAND
Arthur BONGRAND
OpenPETDoPET-L INSIDE
Open Close
Scintillateur + PM GSOZ / PMTmulti-anodes
LYSO / PMTmulti-anodes
LFS* / SiPM*Lu Fine Silicate
Sensibilité(efficacité x acceptance)
5.1% 7.3%?
FOV10x16x16 cm3
?FOV
5.12x5.12x5.12 cm3
Algo 3D LM-MLEM 3D MLEM 3D MLEM
ResolutionSpatiale (mm)
2.6±0.2
2.1±0.1
? ?
TOF Non Non Oui (1.7 ns*)* en ligne direct
Faisceau11C (340 MeV)
2.5Gy
P (98.3 MeV)
C (178-260 MeV)
2 Gy
P (68 & 72 MeV)
2Gy
Online Non Non Non37
11/05/2017
Aire de détection2 modules opposés (6.6 * 6.6 cm2)
placés à 30 cm l’un de l’autre
Composition d’un module Scintillateur et PMs associés :
256 polished LYSO (3.8mm × 3.8mm × 22mm)2 × 2 DPC3200-22–44 (Philips)
11/05/2017 38P Cambraia Lopes et al 2016 Phys. Med. Biol. 61 6203
Digital Photo Counter (dSiPM)DPC3200-22–44
Sensibilité40% à 420 nm
Tunable dark count levels due to digital controlSpeed
Temps de montée 60 ps (intrinsèque)Robustness
Insensible aux champs magnétiques
Arthur BONGRAND
11/05/2017 39P Cambraia Lopes et al 2016 Phys. Med. Biol. 61 6203
Test d’irradiation :
Fantôme de PMMA (9 × 9 x 30 cm3)Fantôme de Polyethylene PE (9 × 9 x 30 cm3)
Type d’irradiation :
Proton 125.67 MeV
Hit synchrotron Structure temporelle du faisceau :
0.62s beam on (in-spill) 3.8s beam off (inter-spill)
Intensité 4.3.108 protons/s Focus faisceau 12.9mm
Arthur BONGRAND
11/05/2017 40P Cambraia Lopes et al 2016 Phys. Med. Biol. 61 6203
Arthur BONGRAND
Arthur BONGRAND
OpenPETDoPET-L INSIDE
TOF PETDPC
Open Close
Scintillateur + PM
GSOZ / PMTmulti-anodes
LYSO / PMTmulti-anodes
LFS* / SiPM*Lu Fine Silicate
LYSO / DPCDigital Photon Counter
Sensibilité(efficacité x acceptance)
5.1% 7.3%?
FOV10x16x16 cm3
?FOV
5.12x5.12x5.12 cm3
?FOV
6.6x6.6 cm2
Algo 3D LM-MLEM 3D MLEM 3D MLEM 3D MLEM
ResolutionSpatiale (mm)
2.6±0.2
2.1±0.1
? ?PE 1.4
PMMA 0.3
TOF Non Non Oui (1.7 ns*)* en ligne direct
Oui (382 ps)**FWHM
Faisceau11C (340 MeV)
2.5Gy
P (98.3 MeV)
C (178-260 MeV)
2Gy
P (68 & 72 MeV)
2GyP (125.67 MeV)
10Gy
Online Non Non Non Non41
11/05/2017
240x LYSO + PMT R1450 HamamatsuSurface couverte 175°
Electronique frontale Echantillonnage jusqu’à 5 GHz
ADC 33 MHz
DAQ µTCA [en cours]Taux de transfert jusqu’à 3 GB/s
Reconstruction
1) Début de traitementUtilisation des LORs pour un contrôle rapide
2) Fin de traitementImagerie (3D MLEM)
11/05/2017 42
Test faisceauProton 65 MeV
Thèse Arnaud ROZESArthur BONGRAND
Arthur BONGRAND
OpenPETDoPET-L INSIDE
TOF PETDPC
DPGA
Open Close
Scintillateur + PM
GSOZ / PMTmulti-anodes
LYSO / PMTmulti-anodes
LFS* / SiPM*Lu Fine Silicate
LYSO / DPCDigital Photon
Counter
LYSO / PMT
Sensibilité(efficacité x acceptance)
5.1% 7.3%?
FOV10x16x16 cm3
?FOV
5.12x5.12x5.12 cm3
?FOV
6.6x6.6 cm2
? Surface
couverte175°
Algo 3D LM-MLEM 3D MLEM 3D MLEM 3D MLEM LOR +
3D MLEM ?
ResolutionSpatiale (mm)
2.6±0.2
2.1±0.1
? ?PE 1.4
PMMA 0.3?
TOF Non Non Oui (1.7 ns*)* en ligne direct
Oui (382 ps)**FWHM
Non
Faisceau11C (340 MeV)
2.5Gy
P (98.3 MeV)
C (178-260 MeV)
2Gy
P (68 & 72 MeV)
2GyP (125.67 MeV)
10Gy
P (65 MeV)
2017
Online Non Non Non Non Oui (µTCA)4311/05/2017
Arthur BONGRAND
5 systèmes TEP
Ère des essais (en condition) cliniques
Modulation temporelle du faisceau
11/05/2017 44
Arthur BONGRAND
Vers des systèmes multimodales (hybrid) (β+,γ) ?
11/05/2017 45
C Lang et al 2014 JINST 9 P01008
Arthur BONGRAND
5 systèmes TEP
Ère des essais (en condition) cliniques
Modulation temporelle du faisceau
Vers des systèmes multimodales (hybrid) (β+,γ) ?
Construction de systèmes d’acquisitions “temps réel”
Controle temps réel du plan de traitement ?
11/05/2017 46
11/05/2017 Arthur BONGRAND 47
11/05/2017 Arthur BONGRAND 48
11/05/2017 Arthur BONGRAND 49
11/05/2017 Arthur BONGRAND 50
Cartographie 3D des émetteurs β+Noyaux β+
Profile d’émission des γ prompts γ prompts
Arthur BONGRAND
Système multimodaleContrôle temps réel du traitement
DPGADétecteur Pixellisé de Grande Acceptance
Caméra collimatée
11/05/2017 51