shell corrosion mapping tank no: t1234 report no: xxxx · (iso document number 652 current issue),...
TRANSCRIPT
SHELL CORROSION MAPPING TANK No: T1234
Report No: XXXX
Test Date: May 2009
Client: Client
Client order No.: Order No
PAL contract ref.: D0xxxx
Asset Location: Client site
Test Engineer: A. Engineer
UT Level II (PCN: 301xxx)
Reviewed by: B. Engineer
UT Level II (PCN: 301xxx)
Approved by: P T Cole
MHG Level III
Report No. FTxxxx Tank T1234
CONTENTS
1. INTRODUCTION AND BACKGROUND 2. TEST PROCEDURES 3. LIMITATIONS 4. RESULTS 5. SUMMARY TABLE OF RESULTS 6. THICKNESS DISTRIBUTIONS 7. FIGURES 8. TANK and TEST DATA
Report No. FTxxxx Tank T1234
1. INTRODUCTION AND BACKGROUND
This report describes work carried out on site and concerns the measurement of wall thickness using the Physical Acoustics LSI automated scanner. LSI is an automated ultrasonic mapping system with a 500mm wide bridge that is operated via an umbilical 20-50 metres long. LSI can climb tank walls by virtue of its magnetic tractors. The measurement result is shell thickness as a function of position for the areas scanned, and may be provided in pure numerical form (e.g. excel compatible files), as colour coded maps, or as a composite image comprising many scans. Unique benefits of the LSI system are its high speed, and its ability to obtain measurement data at critical points such as the weld heat affected zone, which is often subject to selective corrosion, and to obtain data on poor surfaces. The purpose of the measurement is to establish tank shell condition in order to help make decisions on tank maintenance timing and priority. 2. TEST PROCEDURES
All testing is carried out to standard Physical Acoustics Limited standard procedures, (ISO document number 652 current issue), the field worksheets (Doc.654) are kept as a quality control record together with the digital test data, plus an overall quality plan (Doc.653) for the individual tank, according to PAL ISO 9001:2000 procedures. Engineers are trained specifically in these procedures, as well as in operation of the LSI system. Data and measurement results for the tank are detailed in the results, figures and appendix. Typical tank shell corrosion mapping practice for smaller tanks is to carry out a scan at each of the four compass points, each scan being 250-600 mm wide and nearly the full height of the tank (subject to access). Larger tanks typically have an extra four scans in between those detailed above. Measurement position resolution used is typically 2-10 mm on the scan axis, and 5-10 mm on the index (forward) axis, though the system will work at 0.2-600 mm. Thickness measurement resolution is ~0.1mm. The exact specification for the measurements reported here was determined by the client based upon requirement and cost; the details are given in the test data section. 3. LIMITATIONS
The method is ultrasonic, and as such is limited by the probe used to a minimum thickness, typically 2-5mm depending upon probe, surface finish, and thickness range being measured. Depending upon surface, some measurement points may be missed. Measurement is not possible where the surface cannot be “wet” by the water jet coupling medium, for example if oil or other waxy deposits are present on the surface. Whilst average paint thickness may be used to correct the overall measurement data, on a point to point basis it is not practical to correct for the coating thickness. Plate lamination prevents thickness measurement at that point.
Report No. FTxxxx Tank T1234
4. RESULTS
4.1 Details of the measurements are given in the test data section. A summary of the results for each strake for each scan is given in the summary table. In the figures section the collection of all four scans for each strake are shown with the colour coded scale for thickness. Where full numerical results are provided in excel these are supplied separately on CD. Distribution analysis and composite images are extra cost analysis and reporting options.
4.2 General Measurement Observations: None
4.3 Erroneous data identified in the scans: On weld caps, loss of signal due to uneven surface. This is displayed in red and black dots. Black points can also be the result of bad paint (no signal transfer). All red points and lines are due to erroneous data. Poor water (couplant) quality resulted in erroneous results in other areas all well. These are shown in white.
4.4 Scan details:
Four (4) vertical scans were completed on Plate 1, 4, 6 and 9. Four (4) horizontal scans were completed at course 4 on Plate 1, 4, 6 and 9. One (1) horizontal scan was completed at course 5 on Plate 10 above ladder. These scans continued to the highest possible (accessible) level.
Scan on Plate 1 : Scan start : 2400m from right weld.
Scan on Plate 4 : Scan start : 8100m from right weld.
Scan on Plate 6 : Scan start : 8400m from right weld.
Scan on Plate 9 : Scan start : 8700m from right weld.
Report No. FTxxxx Tank T1234
5. SUMMARY TABLE OF RESULTS
Scan at
P1
P4
P6
P10
Course
Nominal 6.40 6.40 6.40 6.40
Average n/a n/a n/a 7.05
Minimum n/a n/a n/a 5.85 (0.22%)
6
Nominal 6.40 6.40 6.40 6.40
Average n/a n/a n/a 5.13
Minimum n/a n/a n/a 3.73 (1.37%)
5
*Course 5 & 6 scan at plate 10 above the stairway.
Scan at P1 P4 P6 P9 Course
Nominal 6.40 6.40 6.40 6.40
Average 5.12 4.90 4.76 4.96
Minimum 2.98 (0.53%) 2.79(0.36%) 2.95 (1.00%) 3.12 (1.67%)
4
Nominal 8.30 8.30 8.30 8.30
Average 7.02 7.27 6.97 6.76
Minimum 5.33 (0.22%) 5.45 (0.19%) 5.31 (0.91%) 5.25 (0.89%)
3
Nominal 10.60 10.60 10.60 10.60
Average 10.69 10.58 10.66 10.68
Minimum 9.57 (0.22%) 9.44 (1.34%) 9.32 (0.20%) 9.70 (1.57%)
2
Nominal 13.10 13.10 13.10 13.10
Average 13.73 13.67 13.54 13.62
Minimum 13.09 (1.96%) 12.91 (2.29%) 12.32 (0.20%) 12.80 (0.57%)
1
Corrosion allowance from general assembly drawing Drawing No is 0.8mm. Figures in red represent thicknesses below the corrosion allowance. The percentage of scan area at the minimum thickness is also shown.
Report No. FTxxxx Tank T1234
T7701 (P1)
(mm) (P4) (mm)
(P6) (mm)
(P9) (mm)
Course
Nominal 6.40 6.40 6.40 6.40
Average 4.58 4.73 4.67 4.72
Minimum 2.97 (1.00%) 3.30 (0.38%) 3.30 (0.20%) 2.94 (0.30%)
4
*Course 4 Horizontal Scans 5000mm at plate 1, 4, 6 and 9.
T7701 (P1) (mm)
(P4) (mm)
(P6) (mm)
(P10) (mm)
Course
Nominal 6.40 6.40 6.40 6.40
Average n/a n/a n/a 6.35
Minimum n/a n/a n/a 3.32 (0.48%)
5
*Course 5 Horizontal Scans 5000mm at plate10 above the stairway. Corrosion allowance from general assembly drawing Drawing No is 0.8mm. Figures in red represent thicknesses below the corrosion allowance. The percentage of scan area at the minimum thickness is also shown.
Report No. FTxxxx Tank T1234
6. Thickness Distributions
Plate 1, 4, 6 and 9 of course 1 scans.
Plate 1, 4, 6 and 9 of course 2 scans.
P1C1 P4C1
P6C1 P9C1
P2C1 P4C2
P6C2 P9C2
Report No. FTxxxx Tank T1234
Plate 1, 4, 6 and 9 of course 3 scans.
Plate 1, 4, 6 and 9 of course 4 scans.
P1C3 P4C3
P6C3 P9C3
P1C4 P4C4
P6C4 P9C4
Report No. FTxxxx Tank T1234
7. FIGURES
Collection of all four scans for each strake with the colour coded scale for thickness. All scans start from top to bottom.
Figure 1. Course 1 – All scans from left to right. Plate 1, 4, 6 and 9 respectively
Figure 2. Course 2 – All scans from left to right. Plate 1, 4, 6 and 9 respectively
Report No. FTxxxx Tank T1234
Figure 3. Course 3 – All scans from left to right. Plate 1, 4, 6 and 9 respectively
Figure 4. Course 4 – All scans from left to right. Plate 1, 4, 6 and 9 respectively
Report No. FTxxxx Tank T1234
Figure 5. Course 5 – Above ladder scan at plate 10.
Figure 6. Course 6 – Above ladder scan at plate 10.
Report No. FTxxxx Tank T1234
Figure 7. Course 4 – Horizontal scans 5000mm. Plate 1, 4, 6 and 9 respectively. All scans from right to left.
Plate 1, 4, 6 and 9 of course 4 horizontal scans.
P1C4H P4C4H
P6C4H P9C4H
Report No. FTxxxx Tank T1234
Figure 8. Course 5 – Horizontal scans 5000mm. Plate 10 above ladder. Scan from right to left. Plate 10 of course 5 horizontal scans
P10C5H
Report No. FTxxxx Tank T1234
8. TANK and TEST DATA
Tank No. : T1234 Diameter: : 29.02m
Contents: : Light Slop Height: : 13.67m
Date of test: : May 2009 Duration: : 1 day
Client: : Client
Client order No. : Order No
PAL contract ref: : D0xxxx
Asset location: : Client Site
Asset owner: : Client
Responsible test eng: : A.Engineer
Certification: : UT PCN Level II
Test system I.D.: : FSI-4
UT probe and SN: : Focussed Probe 5 MHz – 0111B1
Water path: : 1”
Index resolution: : 15 mm
Scan resolution: : 4 mm
Scan width: : 450 mm
Number of scans: : 4 locations – Plate 1, Plate 4, Plate 6 and Plate 9
Colour palette: : 16 colour
DAC details: : N/A
Gain: : 22 dB
MSPS: : 100
Filter: : 2.0 – 12.5 MHz
SET file name: : Course1.set, Course2.set, Course3.set, Course4.set
Course5.set and Course6.set
Calibration: : PAL Calibration Blocks 5mm and 10mm.