1
Sputtering System- Four targets or higher.
- Deposition rate up to: 0.25 Ao/s.
- Minimum four magnetrons.
- Integrated E-beam deposition with sweep pattern and process control.
- Integrated in situ Ellipsometry.
- Gas Control for Reactive Sputtering Upgrade
- Vacuum level up to: 9 x 10-8 Torr.
- Deposition software with unified machine and deposition control built
on .NET framework that communicates with an industrial PLC.
- Recipe controlled automated crucible indexer with torque sensing that
prevents jamming of hearth and provides precise pocket tracking at all
times.
- Fully integrated and unified operation of reactive gas control within
deposition software allows for the addition of reactive sputtering recipe
steps to a deposition
- Glove box ready attachment.
- Aluminum high vacuum box chamber
- Minimum five years warranty.
- Minimum five days onsite training.
- Power supply requirements: 220 V / 50 Hz
- Priority will be given to Jordanian providers have exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
2
FTIR system- Spectral range: (50 – 50000) cm-1
- Spectral resolution: Better than 0.06 cm-1
- Wavenumber accuracy: Better than 0.005 cm-1 at 2,000 cm-1
- Photometric accuracy: Better than 0.06% T
- Scan speed: 12 velocities standard, 1.6 – 160 kHz.
- Step Scan/ Slow Scan function: Continuously variable scanner velocity down to
10 Hz.
- Automatic shutters to seal optics while changing samples.
- All vacuum pumps to be included.
- Beam splitter with wide range beam splitter for FIR and MIR, wide range CaF2
for UV/VIS/NIR, KBr beam splitter for MIR, and four positions automatic beam
splitter exchanger unit for computer controlled exchange under vacuum
condition.
- Sources: Internal Mid and Far IR ceramic source, Internal near IR/Visible
tungsten source, and External Deuterium-UV lamp with air cooling.
- Detectors: Five detectors mounted in the same time including high sensitivity
DLaTGS detector for MIR range, wide range DLATGS detector for MIR and
FIR ranges, room temperature InGaAs detector for NIR range, Si diode detector
for NIR/Vis range, GaP diode detector for UV range, and an additional detector:
MCT detector, medium band Liquid N2 cooled with spectral range: 12,000 – 600
cm-1
- Cryostat: Variable temperature cell for liquids and solids including adaption for
vacuum spectrometer and temperature controller.
- Computer interface: standard Ethernet connection, TCP/IP protocol.
- Integrated advanced FTIR Microscope with:
High performance infrared microscope with motorized mapping stage for
mapping and integrated high resolution color LCD.
Motorized z-drive and autofocus.
Motorized stage provides an adjustment accuracy of ± 0, 1 μm.
Repeatability better than ± 1 μm.
Equipped with trinocular and video camera for visual observation of the
samples
The measurement area is visually selected by adjustment of apertures.
The microscope is optimized for analysis of samples with areas up to 170
μm.
The minimal spot size is determined by the diffraction limit of the analyzing
radiation and the numerical aperture of the objective used.
Spatial resolution : 8 μm
Objectives: 15x objective with 24 mm working distance
20x ATR-objective (NA=0.6) - Operates as a high quality reflecting
objective as well as an ATR objective (Ge crystal , tip size 100 μm )
Upgradable to integrated focal plane array detector
- Power supply requirements: 220 V / 50 Hz
- Priority will be given to Jordanian providers have exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
3
Ellipsometry Device- Reflection and Transmission Ellipsometry
- Spectral range: (240 – 1700) nm, upgradable to 3400 nm.
- Automated angle of incident between 15o – 90o, with fine resolution up to
0.0025o.
- Double chamber monochromator for stray light rejection.
- Fully automated Auto Retarder rotatable and tilt able wave plate
compensator.
- Ability to have variable angle spectroscopic Ellipsometry and
transmission and reflection intensity measurements.
- Theta-2Theta goniometer configuration
- Reflection intensity) measurements.
- Data acquisition rate: 0.1 – 3 seconds/wavelength.
- Ability to define number of wavelengths required for the measurement.
- Ability to measure depolarization, anisotropy, and Mueller-Matrix data.
- Operator computer with required software package.
- Vertical sample stage with vacuum mount
- Liquid prism cell.
- One year warranty.
- Power supply requirements: 220 V / 50 Hz
- Priority will be given to Jordanian providers have exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
4
3D Optical Profiler- Vertical scanning (VSI) and phase shifting interferometry (PSI)
in the same instrument
- 100 mm Motorized XY stage, and 100 mm motorized Z axis
range.
- integrated closed loop z-measurement flexure mechanism
- Piezo Z-scan range up to 500 microns in the VSI mode.
- Sub-nanometer vertical repeatability with 0.1 nm in PSI mode
- integrated tip-tilt stage
- Active vibration isolation system
- High Resolution Camera with minimum 5 Megapixels.
- Manual 4-place turret.
- 10 x – 100 x focal lenses
- Sample reflectance range 0.5 – 100 %.
- Software controlled process with required PC.
- Ability to analyze different surfaces in different settings.
- One year minimum warranty.
- Power supply requirements: 220 V / 50 Hz
- Priority will be given to Jordanian providers have exclusive
agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
5
Drop shape analyzer- USB 3.0 camera with max. speed not lower than 1900 fps and resolution
of the camera 1200 x 800 pix for up to 200 fps.
- Optic with 6.5 fold zoom and field of view not lesser than: 3.2 x 3.2 mm
(max zoom)
- X, Y and Z axes software controlled.
- Standard sample stage of 105 x 105 cm with max sample weight not lesser
than 10 kg
- Monochromatic LED illumination with 470 nm and field of light not lesser
than 42 mm in diameter
- Single direct dosing system with Software controlled dosing and drop
deposition
- Software controlled Pressure dosing system for simultaneous dosing and
evaluation of two droplets by using liquid needle
- Maximum sample space: 320 x 275 mm (W x H)).
- Ports for two temperature sensors.
- Keyboard for instrument control.
- Fully automated process with software and required PC, and Permanent
video recording.
- Minimum 4 years of warranty.
- Interface PC and onsite training to be included in the offer.
- Power supply requirements: 220 V / 50 Hz
- Priority will be given to Jordanian providers have exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
6
Wire Bonder- Touch screen operation: minimum 7”.
- Wire capacity: 0.0007" to 0.003" diameter
- Load/store wire bonding profiles with disk on
key backup.
- Bonding profiles internal library.
- Ability to wire Gold, Aluminum, and Copper,
more options are preferred.
- Z axis DC servo motion with closed loop control
- Phase Locked Loop (PLL) ultrasonic generator
and high-Q transducer
-RoHS compliance
- Bonding Types: Wedge, Tab Stitch and Ribbon.
- Advanced Wedge Automatic Wire re-Feed
- Integrated Microscope.
- Spot light targeting.
- Deep access 90 degree Feed Clamp for wires
1mil to 3mil
- One year warranty.
- Power supply requirements: 220 V / 50 Hz
- Priority will be given to Jordanian providers
have exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
7
Parameter Analyzer
- Customizable and fully integrated parameter analyzer.
- I-V source measurement unit with +/- 210 V/100 mA modules.
- Two and four wire connections.
- C-V multi frequency Capacitance unit with Ac impedance
measurements (C-V, C-f, C-t), and 1kHz – 10 MHz frequency
range.
- I-V/C-V switch modules to easily switch between tests without
changing any physical setup.
- GUI based software with data analysis.
- 15.6 inch minimum display screen (Touch screen is preferred)
- Built in ground unit.
- Standard data communication and connections ports included.
- Power supply requirements: 220 V / 50 Hz
- Five years minimum warranty.
- Priority will be given to Jordanian providers have exclusive
agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
8
Solar Simulator specs
- LED based solar simulator to cover the
range between 400 – 1100 nm.
- System should be able to tune the output
solar radiant in different wavelength bands
in the specified range.
- System output power can be tuned
between 0.1 – 1 Sun.
- Minimum 1 inch solar illuminated area.
- System should include a reference cell for
calibration purposes.
- USB controlled LED.
- Power supply requirements: 220 V / 50 Hz
- One year warranty
- Priority will be given to Jordanian
providers have exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
9
High output Power Solar Simulator
- High efficiency illumination up to 12 suns.
- Standard A spectral match.
- Class A or B uniformity.
- DC Xe arc lamp, up to 3000 Watt.
- Wide range of working distances.
- Digital shutter timer.
- HEPA filtered cooling
- Tunable output with Neutral Density filters.
- Certified by ASTM, IEC or JIS
- Should allow for variable range of working
distances.
- Power supply requirements: 220 V / 50 Hz
- Minimum one-year warranty
- Priority will be given to Jordanian providers
have exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
10
Nanomechanical Test Instrument- Advanced control module with performance leading force and
displacement feedback control algorithms with a 78 kHz feedback loop
rate and user definable data acquisition rates up to 30 kHz for superior
control over all Nanomechanical testing techniques
- Dual head testing capability: range from ≤30 nN to 10 N for true
Nano–micro scale connectivity
- Active vibration dampening.
- Top-down, high resolution color optics for viewing and selection of
testing sites
- 500 nm resolution staging for precise sample positioning.
- Transducer load specs: Resolution < 1 nN, Noise floor < 30 nN, image
contact resolution < 70 nN.
- Displacement specs: Resolution < 0.02 nm, Noise floor < 0.2 nm, Drift
< 0.05 nm/sec.
- XY stage specs: Travel (250 x 150) mm, and Encoder resolution up to
500 nm.
- Ability to perform the following tests: Quasistatic Nanoindentation,
Scratch testing, Scanning Wear, Modulus mapping, Nano DMA, Nano
ECR, Electrochemical cell, Tribol Image,
- Power supply requirements: 220 V / 50 Hz
- Five years minimum warranty.
Priority will be given to Jordanian providers have exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
11
Langmuir and Langmuir-Blodgett Deposition- Surface area up to 273 cm2 (Larger areas are accepted)
- Maximum compression ratio 10.8 (Higher options are accepted)
- Barrier speed: 0.1- 270 mm/min
- Balance measuring range: 0 – 300 mN/m
- Dipping speeds: 0.1 – 108 mm/min.
- Compression/relaxation isotherms: measuring surface pressure as a function of
mean molecular area, remaining area, time or any other measured parameter.
- Analysis of monolayer kinetics
- Analysis of monolayer penetration, solubility and binding of biomolecules
- Isochores and Isobars: increase or decrease of surface pressure/ temperature,
surface pressure/time, or surface pressure/any desired measurable parameter can
be plotted.
- Dilational rheology: oscillating barriers for monitoring viscoelastic properties at
desired surface pressure.
- Dipping: both Langmuir-Blodgett and Langmuir-Schaefer modes allow the
control and monitoring of surface pressure, dipping speed, stroke length,
deposition profiles and transfer ratio.
- Power supply requirements: 220 V / 50 Hz
- Five years minimum warranty.
- Priority will be given to Jordanian providers have exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
12
Laser lithography system- Substrates up to 200 x 200 mm²
- Structures down to 0.3 µm
- Address grid down to 7 nm
- Multiple write modes
- Vector and Raster exposure mode
- Multiple data input formats
- Back side alignment
- Climate chamber
- Customer specific laser
- Optical and air-gauge auto focus
- Scripting capability
- 3D exposure mode
- Camera system for alignment
- 5 years minimum warranty.
- Power supply requirements: 220 V / 50 Hz
- Priority will be given to Jordanian providers have
exclusive agents
Notes:-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Tender No:
Item No:
Our Instrument specification comply Non comply
13
Glove Box- A glove box with 3 gloves access.
- Dimensions L/H/P: 1500x900x725mm
- Stainless Steel frame on casters and lock out cylinders
- 1 manometer -125/+125mm CE + protection filter
- BUTYL glove size 7.5 or 8.5
- Electrical feedthrough 220V
- 2 blind feedthroughs ISO KF40
- One gas input (needle valve)
- One gas output (ball valve)
- Three stainless steel shelves
- Sapphire coated panel
- UV protection for front panel and lighting
- Vacuum chamber with automatic cycles
- HEPA (THE) 5M3/H filter
- O2 Analyzer IN % (0-25)
- Humidification sensor kit.
- Automatic regulation on humidity.
- Pressure regulation
- In deck (built in) spin coater.
- Power supply requirements: 220 V / 50 Hz
- Priority will be given to Jordanian providers have
exclusive agents