selecting and managing contract sterilizers and testing laboratories pacific biolabs inc. (510)...
Post on 19-Dec-2015
215 views
TRANSCRIPT
SELECTING AND MANAGING CONTRACT STERILIZERS AND
TESTING LABORATORIES
Pacific BioLabs Inc.(510) 964-9000
2
CONTRACT STERILIZERS
Written agreement between the contract sterilizer and the device manufacturer is required by 21 CFR- 801-150(e), Medical Devices, processing, labeling or repacking
3
WRITTEN AGREEMENT MUST INCLUDE:
Instructions for maintaining proper records
Acknowledgement that the device is non-sterile and awaiting processing
Detailed specifications for the sterilization process
Division of responsibilities
4
DIVISION OF RESPONSIBILITY
The device manufacturer is ultimately responsible for the sterility of its products
The contract sterilizer is responsible for a quality control system for steps under its control
Therefore, ……..
5
RESPONSIBILITIES OF MANUFACTURER
Ensure devices are sterile by confirming quality control system of contractor
Validation of the process
Ensure process controls are appropriate
Maintain records of specification & evidence they were followed
Product release
6
RESPONSIBILITIES OFCONTRACT STERILIZER
All requirements of the quality system regulation that apply to its operations Equipment installation, maintenance, and
equipment calibration Installation and operation qualifications Building adequacy Personnel training Environmental controls
7
RESPONSIBILITIES OFCONTRACT STERILIZER (cont.)
Quality system requirements: Process controls Documentation and record keeping Acceptance of manufacturing materials Procedures for handling process deviations
Any other requirement of the written agreement
8
COMMON PROBLEMS FOUND BY FDA
Manufacturer does not assume responsibility for sterilization performed by the contract sterilizer
Communication between the two parties of changes
Product is released following sterilization without approval of device manufacturer
9
TESTING LABORATORY
Manufacturing Facility
Contract Laboratory
Contract Sterilizer
10
LAB ATTRIBUTES
Quality System SOP Calibrated equipment Training
FDA Track Record
Experienced Personnel
Industry Reputation
Committed Client Services
11
STERILIZATION QUALITY TESTS
Bioburden
Bioburden Recovery Validation
Sterility
Bacteriostasis and Fungistasis
Biological Indicators
Pyrogen or Bacterial Endotoxin
Inhibition and Enhancement
12
STERILIZATION QUALITY TESTS (cont.)
Package Integrity
Chemical Residuals
Environmental Monitoring
Microbial Identification
13
BIOBURDEN
Webster’s bio (Greek) – life, living burden – that which is carried, load
Medical Device Population (number and types) of viable
microorganisms on a product and/or a package
14
BIOBURDEN TEST
Three steps Removal of microbes from a product Culturing on a medium appropriate for growth Enumeration
Bioburden Recovery Validation Demonstrates the adequacy of the steps above
15
BIOBURDEN METHODS
Pour plating
Spread plating
Membrane filtration
Automated microbiology systems
16
STERILITY TESTS
Product Sterility Membrane filtration Direct transfer
Biological Indicators (BI) Sterility Testing the BI by placing it in microbiological
growth medium
17
STERILITY TEST VALIDATION
Bacteriostasis and Fungistasis (B & F) Test using known organisms to demonstrate
whether the combination of the product and sterilization method do or do not inhibit microbial growth
18
STERILITY TESTUSP vs. AAMI/ISO
USP Two microbiological media
– SCDM and FTM B&F (6 organisms)
AAMI/ISO One microbiological medium
– SCDM B&F (3 organisms)
19
BIOLOGICAL INDICATOR TESTS
Three quality attributes may be measured: Species identification Population Resistance
– D-Value– Z-Value
20
BACTERIAL ENDOTOXIN TEST
Endotoxin = Pyrogen
Bacterial Endotoxin Test = LAL Test Gel clot method Photometric techniques
– Photochromogenic– Turbidimetric
21
INHIBITION AND ENHANCEMENT TEST
Validation of the LAL test
3 lots of product must be tested
22
DEVICE PACKAGING21 CFR 820.130
“Each manufacturer shall ensure that device packaging and shipping containers are designed and constructed to protect the device from alteration or damage during the customary conditions of processing, handling, and distribution”
23
DEVICE PACKAGING21 CFR 820.130
Package integrity is required to maintain the microbial barrier and sterility of the product
Expiration dating Required for IVD’s and certain other devices
that may degrade over time If label includes expiration, you must have data
to back it up
24
PACKAGE INTEGRITY TESTS
Dye Penetration
Microbial Challenge
Burst
Creep
Seal peal
25
CHEMICAL RESIDUALS(EO ONLY)
EO (ethylene oxide)
ECH (ethylene chlorohydrin)
EG (ethylene glycol) No longer required
26
ENVIRONMENTAL MONITORING
Conducted in-house
Laboratory field technicians conduct sampling at manufacturing facility
Samples collected by sponsor’s personnel and incubated/counted at testing laboratory
27
ENVIRONMENTAL MONITORING TESTS
Non-viable particles
Viable particles Bacteria
– Soybean Casein Digest Medium (SCDM)– 30 – 35 °C
Yeast and Mold– Sbouraud Dextrose Agar (SDA)– 20 – 25 °C
Growth promotion required
28
MICROBIAL IDENTIFICATION
Morphological characterization Colony description, Gram stain
Identification to genus and species level Phenotypic Genotypic
29
MICROBIAL IDENTIFICATION
Biochemical Tests Vitek
Fatty Acid Analysis MIDI
Carbon Utilization Biolog
Genetic Based Technology (preferred) RiboPrinter
30
MICROBIAL WATER TESTS
Total Heterotrophic Counts
Yeast and Mold Counts
Coliform Determinations
Thermophiles Counts
Anaerobes counts
Pseudomonas screening
31
WATER TESTS
Choice of Media Plate Count Agar Soybean Casein Digest Agar R2A Tryptone Glucose Extract
Choice of Incubation Temperature30-35°C, 20-25°C
Choice of Incubation Times2, 3, 5 or 7 days
THANK YOU
Q & A