Introduction Name: Reusable SpO2 Sensor with Disposable Wrap Model Number: 15040059 Compatibility: CMI PC-66 series ONLY Components: It consists of 2 parts. 1. Reusable rubber sensor 2. Single user disposable velcro wrap Recommended Patients: Pediatric (weight 33 lbs to 88 lbs age > 3 years), Infant (weight 6.6 lbs to 33 lbs age from 6 months to 3 years). Life Expectancy: 3 months for the reusable part. Attentions The operation of this sensor can ONLY be performed by the trained personnel. Different measuring position should be chosen for different patient type, forefinger (circumference > 1.4 inches) is preferred for pediatric and big toe (circumference < 2.3 inches) is preferred for infant. When applying this sensor on infant, the self-adhesive tape is recommended to be used for wrapping on the sole to fix the sensor cable. Make sure the light emitting end aims at the bottom of the nail, and wrapping is done with proper tightness. Make sure the measured finger or toe nail has no cosmetic painting (e.g. nail polisher), otherwise the measurement might be inaccurate or not produce readings. High emission of surrounding light sources, such as fluorescent light, ruby lamp, infrared heating lamp, and direct sunlight, can affect the measuring results. Excessive patient movement and the interference from an electro-surgical unit can also affect the measuring accuracy. Specifications SpO2 measuring range: 35%~100% SpO2 measuring accuracy: Arms value (defined in ISO 9919 / ISO 80601-2-61) is not greater than 3% in the range of 70%~100%. Pulse Rate measuring range: 30bpm~250bpm Pulse Rate measuring accuracy: ±2bpm or ±2%, whichever is greater. Wavelength: Red light: 663nm, Infrared light: 890nm Instructions It is intended to be used with a compatible pulse oximeter or capnography with oximetry function for measuring the functional oxygen saturation (SpO2) and pulse rate. This sensor should be used together with the compatible oximetry device, otherwise the sensor may not work or the measurement will be inaccurate. Although the biocompatibility evaluation has been performed on this sensor, some exceptional allergic patients may still have anaphylaxis. Do Not apply this sensor to those who has anaphylaxis. Change measuring site every 2 or 3 hours. When the ambient temperature is over 95 F, change the measuring site every 2 hours. When the ambient temperature is over 98.6 F, STOP using this sensor on the patient, since long term use of measurement can cause serious scalding or burn injury. The measuring site must be examined more carefully for special patients. Do not place the SpO2 sensor on the site with edema or fragile tissue is present. Misapplication of the sensor with excessive pressure for prolonged periods can induce pressure injury. Check the integrity of the sensor before use, discard and replace the sensor if any part of it is damaged. Note: This sensor is compatible only with oximeter and capnography sold by Creative Medical International or the CMI Store. Other information about this sensor, please refer to the user manual of its compatible device. Creative Medical International Inc. Address: 5975 Shiloh Road, Suite 114, Alpharetta, GA 30005, USA Toll-free: +1-888-985-1125 E-mail: info@cmihealth.com Website: www.cmihealth.com Self-adhesive tape Disposable velcro wrap latex free last updated: Feb 2016 Reusable SpO 2 Sensor with Disposable Wrap Directions Use M/N: Model Number Do not litter Refer to the instructions Warranty CMI offers 3 months warranty against manufacturing defects for the reusable part of this product, in its undamaged condition. Self-adhesive tape For infant For pediatric Warnings Light receiving end Light path Light emitting end facing the cuticle Reusable sensor

Upload: others

Post on 02-Jan-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Reusable SpO2 Sensor with Disposable Wrap · 2019-07-10 · functional oxygen saturation (SpO ) and pulse rate. This sensorshould be used together with the compatible oximetry device,

Introduction Name: Reusable SpO2 Sensor with Disposable WrapModel Number: 15040059 Compatibility: CMI PC-66 series ONLY Components: It consists of 2 parts.

1. Reusable rubber sensor2. Single user disposable velcro wrap

Recommended Patients: Pediatric (weight 33 lbs to 88 lbs age > 3 years), Infant (weight 6.6 lbs to 33 lbs

age from 6 months to 3 years). Life Expectancy: 3 months for the reusable part.

Attentions The operation of this sensor can ONLY be performed by

the trained personnel. Different measuring positionshould be chosen for different patient type, forefinger(circumference > 1.4 inches) is preferred for pediatricand big toe (circumference < 2.3 inches) is preferred forinfant.

When applying this sensor on infant, the self-adhesivetape is recommended to be used for wrapping on the soleto fix the sensor cable.

Make sure the light emitting end aims at the bottom ofthe nail, and wrapping is done with proper tightness.Make sure the measured finger or toe nail has nocosmetic painting (e.g. nail polisher), otherwise themeasurement might be inaccurate or not producereadings.

High emission of surrounding light sources, such asfluorescent light, ruby lamp, infrared heating lamp, anddirect sunlight, can affect the measuring results.

Excessive patient movement and the interference froman electro-surgical unit can also affect the measuringaccuracy.

Specifications SpO2 measuring range: 35%~100% SpO2 measuring accuracy: Arms value (defined in ISO 9919 / ISO 80601-2-61) is not greater than 3% in the range of 70%~100%. Pulse Rate measuring range: 30bpm~250bpm Pulse Rate measuring accuracy: ±2bpm or ±2%,

whichever is greater. Wavelength: Red light: 663nm, Infrared light: 890nm

Instructions It is intended to be used with a compatible pulse oximeter or capnography with oximetry function for measuring the functional oxygen saturation (SpO2) and pulse rate.

This sensor should be used together with the compatibleoximetry device, otherwise the sensor may not work orthe measurement will be inaccurate.

Although the biocompatibility evaluation has beenperformed on this sensor, some exceptional allergicpatients may still have anaphylaxis. Do Not apply thissensor to those who has anaphylaxis.

Change measuring site every 2 or 3 hours. When theambient temperature is over 95 F, change the measuringsite every 2 hours. When the ambient temperature is over98.6 F, STOP using this sensor on the patient, sincelong term use of measurement can cause seriousscalding or burn injury.

The measuring site must be examined more carefully forspecial patients. Do not place the SpO2 sensor on the sitewith edema or fragile tissue is present.

Misapplication of the sensor with excessive pressure forprolonged periods can induce pressure injury.

Check the integrity of the sensor before use, discard andreplace the sensor if any part of it is damaged.

Note: This sensor is compatible only with oximeter and capnography sold by Creative Medical International or the CMI Store. Other information about this s ensor, please refer to the user manual of its compatible device.

Creative Medical International Inc. Address: 5975 Shiloh Road, Suite 114, Alpharetta, GA 30005, USA

Toll-free: +1-888-985-1125 E-mail: [email protected] Website: www.cmihealth.com

Self-adhesive tape

Disposable velcro wrap

latex freelast updated: Feb 2016

Reusable SpO2 Sensor with Disposable Wrap Directions for Use

M/N: Model Number Do not litter Refer to the instructions

WarrantyCMI offers 3 months warranty against manufacturing defects for the reusable part of this product, in its undamaged condition.

Self-adhesive tape

For infant For pediatric

Warnings

Light receiving end

Light path

Light emitting end facing the cuticle

Reusable sensor