    TABLE-TOP CENTRIFUGE

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500N

    MANUAL

    V i s i o n S c i e n t i f i c C o . , L t d .W W W . V I S I O N B I O N E X . C O M

    Tel. 82-32-672-3091 FAX:82-32-672-3098

    Please read this manual before using.

  • 8/10/2019 PRP machine manual.pdf

    2/14

    - 2 -

    Prohibit Caution Indication

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NPrecautionPrecautionPrecautionPrecaution

    30cm from wall

  • 8/10/2019 PRP machine manual.pdf

    3/14

    - 3 -

    Caution in installationCaution in installationCaution in installationCaution in installation

    1. Do not install in high temperature & humidity. Install on well ventilated place.

    2. Install apart from wall by 30Cm. Please set on place which is plain & no vibration.

    3. Do not install on flammable or explosive place.

    4. Please clarify proper electronic source before using. (Voltage, Power)

    5. Please use in place where the voltage variable rate is within 10 %.

    6. Please use the ground outlet.

    Caution in operationCaution in operationCaution in operationCaution in operation

    1. Please keep the circumstantial conditions as temperature from 2 to 40 , and

    humidity lower than 80%. Unit may not operate correctly.

    2. Do not put the plug with wet to prevent the electric shock.

    3. Please keep the circumstantial temperature not to changed rapidly.

    Humidity may caused on unit inside. And it can cause trouble on unit.

    (If dew is foamed on chamber, clean with soft dry fabric.)

    4. Please put the contents into ROTOR symmetrically with same weight of contents.

    If not, unit is operated in UNBALANCE and causes the trouble.

    5. If you use under room temperature, please operate the refrigerated before 30 minutes

    from input contents.

    If you put the contents into CHAMBER before getting the aimed temperature, contents

    can be degenerated.

    6. Once refrigerator is operated, do not touch the chamber in wet. Hand can be frozen or

    stuck on chamber by cold.

    7. Door is closed during rotation. Do not open by forced for safety.

    After end of operation, DOOR can be opened but chamber is still rotating for a while.

    Therefore, please put out the contents after rotor is fully stopped.

    8. When move the unit or do not use for a long time, please separate the rotor from unit.

    If do not used for a long time, rotor may cannot be separated form shaft.

    Additional cautions.Additional cautions.Additional cautions.Additional cautions.1. Use mild detergent for cleaning.

    2. Please use proper tubes.

    - Plastic tubes of Nalgene, Falcon, Corning, Nunc, Sastedit.

    - It it is possible, don't use glass tubes. In case of using glass tubes, take Corex

    tubes or Firex tubes and don't set speed higher than 2,500RPM.

    3. In case of smoke, strange smell or Error occurs, cut off main power switch and inform

    to contact our service center.

    4. Do not alternate or repair by an unauthorized person.

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NCAUTIONCAUTIONCAUTIONCAUTION

  • 8/10/2019 PRP machine manual.pdf

    4/14

    CONTENTS

    - 4 -

    PrecautionPrecautionPrecautionPrecaution ................................................. 2

    CautionCautionCautionCaution .................................................... 3

    CompositionCompositionCompositionComposition ................................................ 5

    - Contents ................................................. 5

    - Installation ............................................. 5

    - Composition of each part ................................. 6

    FeatureFeatureFeatureFeature .................................................... 7

    - Meaning of word .......................................... 7

    - Feature ................................................. 7

    - Specification ........................................... 8

    - Type of rotor ........................................... 8

    OperationOperationOperationOperation .................................................. 9

    - Control Panel ............................................ 9

    - Operation ................................................ 10

    - Program setting .......................................... 11

    ReferenceReferenceReferenceReference ................................................. 13

    WarrantyWarrantyWarrantyWarranty .................................................. 14

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500N

  • 8/10/2019 PRP machine manual.pdf

    5/14

    - 5 -

    ContentsContentsContentsContents

    Please check all contents before installation.

    1. Centrifuge body : 1 EA

    2. Standard accessory : Swing rotor, Bucket (VB-32) : 1 SET

    3. AC Power Code : 1 EA

    4. Manual : 1 EA

    InstallationInstallationInstallationInstallation

    1. Install on place where is plane & no vibration. And apart from the wall by 30Cm.

    2. Check the outside and all contents.

    3. Connect the AC Power Cord and on rear part and connect to the grounded outlet.

    And then turn on the main power switch on rear part.

    Please use proper voltage to prevent fire, electrical accident or damage on unit.

    4. Turn right the Rotor lid to separate from rotor.

    ( : Close, : Open)

    5. Put the separated rotor to shaft on chamber.

    6. Fix the Rotor lid on rotor and fasten by turning left.

    If the lid is loose, it can be opened during operation. Therefore, check the lidone more time before operation.

    When you use swing rotor, please fasten strongly by 20mm spanner. (Not to fasten

    it only by your hands.)

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NComposition

  • 8/10/2019 PRP machine manual.pdf

    6/14

    - 6 -

    Composition of each partComposition of each partComposition of each partComposition of each part

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NComposition

    No. DescriptionDescriptionDescriptionDescription FUNCTIONFUNCTIONFUNCTIONFUNCTION

    DOORDOORDOORDOOR

    DOOR LockingDOOR LockingDOOR LockingDOOR Locking

    Air ventAir ventAir ventAir vent Vent for AIR COOLING

    ReceptacleReceptacleReceptacleReceptacle Connection to AC POWER CORD (220V)

    Connection PortConnection PortConnection PortConnection Port RS-232C INTERFACE CONNECTOR

    ChamberChamberChamberChamber

    Main power S/WMain power S/WMain power S/WMain power S/W

    CASECASECASECASE

    Front panelFront panelFront panelFront panel Consist of control part. Open when the repair is required.

    ROTOR NUTROTOR NUTROTOR NUTROTOR NUT Nut to fix the ROTOR to motor shaft

    SWING ROTORSWING ROTORSWING ROTORSWING ROTOR

    BUCKETBUCKETBUCKETBUCKET

    MotorMotorMotorMotor

    CONTROL PANELCONTROL PANELCONTROL PANELCONTROL PANEL

    RearRearRearRearCHAMBER InsideCHAMBER InsideCHAMBER InsideCHAMBER Inside

  • 8/10/2019 PRP machine manual.pdf

    7/14

    - 7 -

    Meaning of wordMeaning of wordMeaning of wordMeaning of word

    1. RPM : Rotation per minute.

    2. RCF : Relative Centrifugal Force

    3. : Total centrifugal effect. Speed which is integrated by time.

    FeatureFeatureFeatureFeature

    1. VS-5000N(5500N) is Table-top centrifuge with Brushless D.C. Motor.

    2. CHAMBER is made of Stainless steel, and body is powder coated steel plate(CR 41).

    3. 9 profile for acceleration and deceleration for motor speed.

    4. Microprocessor support 3 mode display (RPM, RCF, ) and 200 of memory is available.

    5. Self-diagnostic detect any error and indicate on display.

    6. Back-light LCD

    7. Strong STAINLESS STEEL CHAMBER, Self-diagnostic, ROTOR unbalance detector

    8. Communication through RS-232C

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NFEATURE

  • 8/10/2019 PRP machine manual.pdf

    8/14

    - 8 -

    SpecificationSpecificationSpecificationSpecification

    Type of RotorType of RotorType of RotorType of Rotor

    Rotor noRotor noRotor noRotor no Rotor / BucketRotor / BucketRotor / BucketRotor / Bucket VS-5000NVS-5000NVS-5000NVS-5000N VS-5500NVS-5500NVS-5500NVS-5500N

    1 Swing rotor / VR-50S

    50ml1Tubesx4pcs / VB-04 5,000RPM / 4,612 x g 5,500RPM / 5,580 x g

    2 Swing rotor / VR-50S50ml2Tubes4pcs / VB-08

    4,500RPM / 3,736 x g 5,000RPM / 4,612 x g

    3 Swing rotor / VR-50S15ml4Tubes4pcs / VB-16

    4,000RPM / 2,952 x g 4,500RPM / 3,736 x g

    4 Swing rotor / VR-50S15ml8Tubes4pcs / VB-32

    4,000RPM / 2,952 x g 4,500RPM / 3,736 x g

    5 Rack Carrier/VB-40 3,500RPM / 2,260 x g 3,500RPM / 2,260 x g

    6 Micro plate rotor/VR-02M

    96well micro plate x 2ea 2,500RPM / 1,013 x g 2,500RPM / 1,013 x g

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NSpecification

    VS-5000NVS-5000NVS-5000NVS-5000N VS-5500NVS-5500NVS-5500NVS-5500N

    Max. Speed 5,000rpm 5,500rpm

    Max. Force 4,612 x g 5,580 x g

    Max. Capacity 480ml, 1532Tubes

    Standard Rotor Swing rotor (VR-50S)

    Main Controller Microprocessor

    Display LCD Display (20 Character x 2 Line)

    Program 200 Memory

    Accel/Decel Rates 1-9 Profile

    Timer 99 min. 59 sec. & Hold Run

    Cooling Device Air Cooling

    Dimensions 440 545 x 330(H)mm

    Drive System Brushless DC Motor

    Safety Devices Over/Under speed protector, Overload protector

    Unbalance detector, Temp. protector

    Power Supply 220V, 50Hz, 6A

  • 8/10/2019 PRP machine manual.pdf

    9/14

    - 9 -

    CONTROL PANELCONTROL PANELCONTROL PANELCONTROL PANEL

    No. NAMENAMENAMENAME FUNCTIONFUNCTIONFUNCTIONFUNCTION

    RPMRPMRPMRPM LED RPM mode indicator

    RCFRCFRCFRCF LED RCF mode indicator

    wwww 2222

    dtx10dtx10dtx10dtx10-5-5-5-5

    LED w 2

    dtx10-5

    mode indicator

    COOLCOOLCOOLCOOL LED Cooling mode indicator No function

    RUNRUNRUNRUN LED Running mode indicator Light on during operation

    ERRORERRORERRORERROR LED Error indicator Light on with error

    DOORDOORDOORDOOR LED Door open indicator Light on when the door open

    BRAKEBRAKEBRAKEBRAKE LED Enforced stop indicator Light on with stop by BRAKE key

    DISPLAYDISPLAYDISPLAYDISPLAY

    PROGRAMPROGRAMPROGRAMPROGRAM KEY Program mode key Mode to PROGRAMMING

    UP( )UP( )UP( )UP( ) KEY Numeric key (UP)

    DOWN( )DOWN( )DOWN( )DOWN( ) KEY Numeric key (DOWN)

    SHIFT(SHIFT(SHIFT(SHIFT( )))) KEY Cursor shift key Move to next cursor in programming

    ENTERENTERENTERENTER KEY Save the program setting

    MODEMODEMODEMODE KEY Mode selection key Shift to RPM RCF w 2

    dtx10-5

    mode

    RCLRCLRCLRCL KEY Program recall Recall setting value

    TOTAL x 10TOTAL x 10TOTAL x 10TOTAL x 10 KEY Total rotation key Show the total rotation of centrifuge

    DOORDOORDOORDOOR KEY Door open key

    COOLCOOLCOOLCOOL KEY Cooling key Not for VS-5500N & 5000N

    STARTSTARTSTARTSTART KEY

    STOPSTOPSTOPSTOP KEY

    BRAKEBRAKEBRAKEBRAKE KEY Enforced stop POWERPOWERPOWERPOWER KEY

    RPM RUN

    RCF ERROR

    COOL BRAKE

    w2dtx10

    -5 DOOR

    ON

    POWER

    PGM

    ENTER

    MODE

    TOTALx10

    RCL

    DOOR

    COOL

    START

    STOP

    BRAKE

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NOperation

  • 8/10/2019 PRP machine manual.pdf

    10/14

  • 8/10/2019 PRP machine manual.pdf

    11/14

    - 11 -

    8. Press key to start operation.

    9. After complete of setting condition, it stops automatically.

    If you want to stop manually, press or button.

    10. After the rotor stop entirely, press key and open the door.

    During rotation, door is locked and not opened.

    11. Turn off the power switch.

    Program settingProgram settingProgram settingProgram setting

    1. When you press KEY, following is showed.

    Pn : Program number (1 ~ 9)

    Rn : Rotor number (Refer to Type of Rotor on 8page)

    Sf : Start factor (Acceleration, 1~9 profile)

    Bf : Brake factor (Deceleration, 1~9 profile)

    RPM : Revolutions per minute (Can be changed to RCF by MODE key)

    TIME : Run time (1sec ~ 99min )

    TEMP : Setting Temperature (SV)

    Pn : 1 Rn : 01 Sf : 1 Bf : 1Pn : 1 Rn : 01 Sf : 1 Bf : 1Pn : 1 Rn : 01 Sf : 1 Bf : 1Pn : 1 Rn : 01 Sf : 1 Bf : 1

    RPM : 00000 00:00RPM : 00000 00:00RPM : 00000 00:00RPM : 00000 00:00 + 00+ 00+ 00+ 00

    START

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NOperation

    STOP

    BRAKE

    PGM

    DOOR

  • 8/10/2019 PRP machine manual.pdf

    12/14

    - 12 -

    2. Pn setting. (Increasing value is 1 with pressing Key, 1~9 is available)

    3. Rn setting. (Shift the cursor to Rn, and select rotor number by key)

    Refer to Type of RotorType of RotorType of RotorType of Rotor on 8 page.

    4. Sf setting. (Shift the cursor to Sf, and select Sf profile by Key)

    1~9 profile is available.

    5. Bf setting. (Shift the cursor to Bf, and select Bf profile by Key)

    1~9 profile is available.

    6. RPM setting. (Shift the cursor to the needed RPM unit, select RPM by key)

    RPM shall not be exceed Max. RPM value of rotor. (Refer to Type of rotor on 8page)

    7. TIME setting. (Shift the cursor to 00:00, select time value by key)

    (99min. 59sec. or infinite. Infinite setting is **:** - below 0 point)

    8. After setting, press Key to save its value.

    9. In case of power failure, the door is not opened by key. In that case, find small &

    white button on right side and pull out it. Then, the door is opened manually.

    PGM

    ENTER

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NOperation

    DOOR

  • 8/10/2019 PRP machine manual.pdf

    13/14

    - 13 -

    ReferenceReferenceReferenceReference

    1. If you want to review setting value during operation of a program, just press key.

    It is turned to normal status by Key.

    .

    2. To check TOTAL RPM, press Key. It is turned to normal status by pressing key

    one more time.

    3. This model has no refrigerating system. ( key on CONTROL PANEL has no function.)( key on CONTROL PANEL has no function.)( key on CONTROL PANEL has no function.)( key on CONTROL PANEL has no function.)

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NReference

    RCL

    ENTER

    TOTALx10

    TOTALx10

    COOL

  • 8/10/2019 PRP machine manual.pdf

    14/14

    WARRANTY & SERVICE

    VS-5000N,5500N is warranted by VISION SCIENTIFIC CO., LTD for a period of one year

    after the date of purchase.

    The warranty applies only to defects due to the fault of product workmanship or materials,

    and is not valid for the defects resulting from accidental damage, misuse or neglect and in

    case of alternations or repair carried out by an unauthorized person.

    Service under the warranty is provided only upon the presentation of a declaration form.

    Service is available in all countries where the product is officially distributed by VISION

    Scientific Co., Ltd. If you need further information or have any difficulties, please contact

    your supplier or directly VISION SCIENTIFIC CO., LTD.

    Address : 201-7, Samjeong-dong, Ojeong-gu, Bucheon-city Kyeonggi

    Province, Korea

    TEL : 82-32-672-3091 FAX : 82-32-672-3098E-Mail: [email protected]

    - 14 -

    VS-5000N,5500NVS-5000N,5500NVS-5000N,5500NVS-5000N,5500NWARRANTY