Technical performance:
• Integrated with high frequency and high voltage generator, main control system, intelligent computer
system, numeric display and beam limited device.
• Designed and manufactured precisely and compactly, more flexibly and widely used; Using the technique
of inverse converter for HF and computer-controlled closed-loop adjusting System, stable high voltage
output is generated and high-quality images can be obtained.
• Using the buttons of KV and mAs to adjust the photographic parameters, display them on high-brightness
LEDs, interlock protection conditions is determined by software, which is convenient for doctors to operate.
• X-ray Tube is placed in lead cylinder that can effectively shield the gamma leakage and protect
environment and operators’ safety.
• 24 radiographic conditions are preset and can be customized according to user requirements.
• Manually controlled, controls remotely the double control photography function.
• Photographic time by microcomputer accuracy control, photographic conclusion aft engine automatic time
delay 30S, but the active control photography gap, avoids overloading the use, realizes the automatic
protection.
• Uses the microcomputer to monitor each kind of breakdown source, once breaks down, microcomputer
cut-off all outputs, and demonstrate the breakdown code, the recognition breakdown reason.
• With more suitable animal hospital bed sets.
• The X-ray machine is hanging on the support then the rapid development throws according to The X-ray
machine fuselages hangs on the support may control ≥±90° revolving, the X-ray shoe may make 90° the
turn over, is advantageous for the different angle to throw illuminates. The arm support may reciprocate
that follow-up balanced and has locks installs.
• Stand arm adjustable up and down, running balance and lock up (Option) .
Main specifications and parameters:
Power supply Voltage: AC220V± 10%, 50/60Hz ± 1Hz
Capacity: ≥5kVA; ternal resistance: ≤0.5Ω
Maximum output power P =110kV×36mA=4kW
Nominal output power 100kV, 40mA, 0.1s, 4kW
Frequency ≥40kHz
kV adjustable range 40kV~110kV, continuous regulation, stepping 1kV
mA adjustable range 36mA~70mA
mAs adjustable range 1mAs-200mAs
Exposure time adjustable range 0.04s~3.2s
X ray tube Focus: 0.6/1.5mm,
Thermal Capacity: 76kHU
Anatomical program 24 memory choices available.
Accessories Standard: manual switch, cable
Optional: mobile steel support