Optia – Efficiency Considerations With MNC Collection
Dr. Volker Witt
![Page 2: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/2.jpg)
OPTIA™ system (CaridianBCT, Lakewood, Dnv, CO, USA)
• Introduced as the replacement for the COBE spectra system
• TPE, RCE available all over the world
• MNZ, Granulo, Leukodepletion are available in Europe
• BM processing is pending • Universal apheresis system
for therapies and collection of blood and tissue products
![Page 3: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/3.jpg)
![Page 4: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/4.jpg)
![Page 5: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/5.jpg)
Harvesting autologous PBMNC in children
• Routine procedure in many pediatric oncology therapy protocols – Neuroblastoma
– Ewing sarcoma
– Nephroblastoma
– Hepatoblastoma
– Medulloblastoma
– CNS tumours
– Lymphoma
![Page 6: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/6.jpg)
Autologous PBCS harvesting in children
Some of the problems • Extracorporeal blood volume
of the apheresis systems • Difficult venous access • Different physiological aspects
in children • Apheresis procedures in
children are accounting for less than 5% of the total number of procedures performed
• Only a few centres are specialized in pediatric apheresis.
Systems in use
• Fenwal CS 3000
– expired
• COBE spectra
– will expire 2014 in Europe
• AMICUS
– still in use
• Fresenius COMTTEC
– rarely used
• Optia
– No published data in pediatrics
![Page 7: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/7.jpg)
Apheresis procedures in the St. Anna Kinderspital from 1996 to QI-2013
0
50
100
150
200
250
300
350
AUTO DC FD DLI ERY ECP RCE TPE IA WBC XX
![Page 8: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/8.jpg)
The use of the apheresis systems in children at St. Anna Kinderspital in 2011+2012
OPTIASpectra
AMICUSothers
0
20
40
60
80
100
120
OPTIA
Spectra
AMICUS
others
![Page 9: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/9.jpg)
collection efficiency : What is efficient?
• Which cell type do I want to collect (HPC, MNC, monocytes, CD3+ T-cells,…)?
• Collecting enough cells for the poposed therapy?
• Collecting enough cells per kg bodyweight?
• Collecting all cells which are running into the apheresis system?
• Collecting all cells per liter whole blood running into the apheresis system?
![Page 10: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/10.jpg)
Factors influencing the calculation of the collection efficiency
• Cell number of the target cell population
• Total cell number
– Cells who can interfere with the target cell population
• Relation between the target cell population and the cells of no interest
• A sufficient and continous blood flow for optimal separation and harvesting
• Cell loss in the set
• A constant and valid method to describe and enumerate the target cell population
![Page 11: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/11.jpg)
Model of apheresis
BM PB AP
![Page 12: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/12.jpg)
Recruitment phenomenon
• The number of cells mobilized into the peripheral blood is even higher than the number of collected cells.
[Mayer J. Bone Marrow Transplantation, 2003, 749-757]
![Page 13: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/13.jpg)
Yield of a cell population
cells in the product [/ml] X harvest volume [ml]
[ cells x 10E06]
[ cells x 10E06/kg bw]
[Strasser E.F. Transfusion 2005, 445-452]
![Page 14: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/14.jpg)
Cell loss
(1-(postdonation cell count/predonation cell count))X(-100)[%]
pre- or postdonation cell count =
absolute number of cells in blood pre- and postdonation =
blood cell count [x10E09/L] x total blood volume [L]
[Strasser E.F. Transfusion 2005, 445-452]
![Page 15: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/15.jpg)
Cell yield per liter of processed blood volume (/L PBV)
Y ÷ PBV [/L]
Y = Yield
PBV = processed blood volume [L]
[Strasser E.F. Transfusion 2005, 445-452]
![Page 16: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/16.jpg)
Recruitment factor (RF)
(postdonation cell count + cell yield) ÷ predonation cell count
pre- or postdonation cell count =
absolute number of cells in blood pre- and postdonation =
blood cell count [x10E09/L] x total blood volume [L]
[Strasser E.F. Transfusion 2005, 445-452]
![Page 17: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/17.jpg)
Collection Efficiency (CE)
Efficiency of the procedure [%]
= cells in the harvest ÷ cells processed
Witt V. Journal of Clinical Apheresis; 2001; 161-168
![Page 18: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/18.jpg)
Collection Efficiency (CE)
a = cells/mL
b = harvest volume in mL
c = cells/mL before harvest
d = cells/mL after harvest
e = processed blood volume [mL]
absolute cell count in the harvest = a * b
processed blood volume (e) = whole blood volume processed – ACD-A volume used
Witt V. Journal of Clinical Apheresis; 2001; 161-168
![Page 19: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/19.jpg)
Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63
![Page 20: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/20.jpg)
Blood volume is the volume of blood (both red blood cells and plasma) in a person's circulatory system. A typical adult has a blood volume of approximately between 4.7 and 5 liters, with females generally having less blood volume than males. Blood volume is regulated by the kidneys. Blood volume (BV) can be calculated given the hematocrit (HCT; the fraction of blood that is red blood cells) and plasma volume (PV):
![Page 21: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/21.jpg)
Total Body Volume (TBV)
Weight in kg x BV in mL/kg = TBV
Patient Fat Thin Normal Muscular
Male 60 65 70 75
Female 55 60 65 70
Infant/Child - - 80/70 -
G. Herold Innere Medizin 2007
Tronier W., Goodfellow E., Larson K. Apheresis Math and Useful Physical constants in Principles of Apheresis Technology, 4th edition Linz W. (editor) 2010
Normal BV = 70 mL/kg bw or 1/14 part of the bodyweight
![Page 22: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/22.jpg)
Collection efficiency
![Page 23: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/23.jpg)
Collection efficiency
• Harvested cells (HPC harvested) are calculated by the formula:
Cellsharvested = cell countbuffy [cells/ml] x volumebuffy [ml]
![Page 24: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/24.jpg)
Collection efficiency
• The cells processed (HPC processed) are calculated by the forumla:
cellsstart – cellsend
2
• The processed blood volume is calculated by the formula:
Inlet volumeproc.– ACD-A used = blood volumeproc.
X processed blood volume
![Page 25: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/25.jpg)
Collection efficiency st
art
Aft
er p
rim
ing
en
d
![Page 26: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/26.jpg)
Collection efficiency
• The cells processed are calculated by the formula:
cellsstart – cellsend
2
• This formula describes the mean of cells present during the procedure.
• The cell yield is represented by the area under the curve.
• This approach reflects only an approximation to the real conditions during an apheresis procedure.
X processed blood volume
![Page 27: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/27.jpg)
Collection efficiency st
art
Aft
er p
rim
ing
en
d
A
![Page 28: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/28.jpg)
Collection efficiency st
art
Aft
er p
rim
ing
en
d
B
![Page 29: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/29.jpg)
Collection efficiency st
art
Aft
er p
rim
ing
en
d
C
![Page 30: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/30.jpg)
Collection efficiency
•A > B
•A < C
![Page 31: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/31.jpg)
Collection ratio
• CR = ((CD34+ /µl product x PV) /kg donor BW) / CD34+ pre
• This formular gives an idea about the collection efficiency of one cell populationen independently from the donor or patients weight.
![Page 32: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/32.jpg)
Collection efficiency
• Ideal times points for target cell count to calculate the „true“ collection efficiency:
Immediately before start of the apheresis
End of priming (saline and blood)
Before reinfusion
Immediately after end of the apheresis
![Page 33: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/33.jpg)
Collection efficiency
• For practical reasons target cell counts are performed only before and after apheresis procedure
• For economic reasons target cell count are performed before apheresis procedure only
![Page 34: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/34.jpg)
Collection efficiency
• How good could the target cell yield predicted by measuring the target cell population before the apheresis procedure only?
![Page 35: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/35.jpg)
N = 7 Median Min - Max
Bodyweight 35 kg 10 – 55
Bodyweight < 20 kg 3/7 10.25, 12.7 and 19 kg
Body lenght 1.45 m 0.75 – 1.72
Body surface 1.19 m² 0.46 – 1.61
Age 11.8 years 1.03 - 20
Sex 2 female 5 male
Underlyingdiseases neuroblastoma, hepatoblastoma, ATRT, PNET, AML and 2 Ewing’s sarcoma
Patients_AUTO
N = 16 Median Min - Max
Bodyweight 30,50 8 – 89
Bodyweight < 20 kg 7/16 12.70, 10.25, 19, 17, 10.7, 12.7 and 8 kg
Body lenght (cm) 134 72 – 187
Body surface (m²) 1,07 0.4 – 2.17
Age 9,88 1,03 - 20,03
Sex 4 female 12 male
Underlying diseases 2 ATRT, AML recurrency, Hepatoblastoma, 3 Neuroblastoma, PNET, 5 Ewing's sarcoma, Retinoblastoma recurrency, germ cell tumor and Rhabdoid tumor
![Page 36: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/36.jpg)
N = 9 Median Min - Max
Bodyweight (kg) 36 27.4 – 45
Bodyweight < 20 kg 0/9 -
Body lenght (cm) 130 125 - 158
Body surface (m²) 1,12 0.99 - 1.41
Age (years) 8,65 7.12 - 20.13
Sex 3 female 6 male
Underlying diseases 8 GVHD and ALL/GVHD
Patients_ECP
![Page 37: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/37.jpg)
Parameter settings_AUTO
N=16 Median Min – Max
Total whole blood drawn (ml) 5267 1687 - 18882
Whole blood (ml/kg bw) 211 123 - 322
Time used [hh:mm] 4:15 1.44 - 6:34
Product volume (ml) 90 31 - 332
N cycles 4 2 - 20
ACD-A ratio 13 12 - 15
Collection preferency 30 20 - 40
![Page 38: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/38.jpg)
Parameter settings_ECP
N=9 Median Min – Max
Total whole blood drawn 3617 2328 - 5100
Whole blood / kg bw 98 58 - 158
Time used 3:04 2:00 - 4:23
Product volume 54 18 - 72
N cycles 3 1 - 4
ACD-A ratio 12 12
Collection preferency 30 20 - 40
![Page 39: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/39.jpg)
N=16 Median Min - Max
Hemoglobin (g/dl) 10,5 8 - 12
Hematocrit (%) 30,80 21 - 34
Platelets (G/l) 173 41 - 255
Leukocyte (G/l) 15,5 6 - 74
Lymphocytes (%) 4,35 1 - 15.4
Monocytes (%) 11,75 5.5 - 41
MNC (%) 18,55 7 - 59.8
Granulocytes (%) 81,45 40.2 - 93
CD34+ cells (%) 0,33035 0.05 - 1.39
cells/µl 46,5 18 - 270
Pre apheresis values_AUTO
![Page 40: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/40.jpg)
Pre apheresis values_ECP
N=9 Median Min - Max
Hemoglobin (g/dl) 11,4 10 - 12.1
Hematocrit (%) 33,00 30 - 36
Platelets (G/l) 158 31 - 362
Leukocyte (G/l) 4,72 1.3 - 7.07
Lymphocytes (%) 31,1 10 - 52.2
Monocytes (%) 6,2 3.7 - 10.3
MNC (%) 38,3 15.7 - 67.8
Granulocytes (%) 61,6 32.5 - 84.5
![Page 41: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/41.jpg)
N=16 Median Min - Max
Hemoglobin (g/dl) 9,5 8.2 - 12.8
Hematocrit (%) 27 23 - 35.2
Platelets (G/l) 105 30 - 167
Leukocyte (G/l) 18,8 4.9 - 69
Lymphocytes (%) 4,15 1 - 16.8
Monocytes (%) 10 2.5 - 39.6
MNC (%) 14 4 - 44.4
Granulocytes (%) 86 51 - 95.5
CD34+ cells (%) 0,21 0.01 - 1.34
cells/µl 32 0.2 - 194
Post apheresis values_AUTO
![Page 42: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/42.jpg)
Post apheresis values_ECP
N=9 Median Min - Max
Hemoglobin (g/dl) 10,2 8.6 - 11.6
Hematocrit (%) 29,00 25 - 34
Platelets (G/l) 114 9 - 320
Leukocyte (G/l) 4,85 1.5 - 7.6
Lymphocytes (%) 36,1 8.7 - 59.4
Monocytes (%) 5,3 1.6 - 11.3
MNC (%) 45,7 14.7 - 74.1
Granulocytes (%) 54,4 25.8 - 85.3
![Page 43: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/43.jpg)
N=16 Median Min - Max
RBC (T/l) 0,25 0.1 - 0.5
Hematocrit (%) 2,65 1 - 7
Platelets (G/l) 1044,5 171 - 2330
Leukocyte (G/l) 154,5 106.7 - 275
Lymphocytes (%) 25,00 4.5 - 48.5
Monocytes (%) 35 5.5 - 58
MNC (%) 68,50 21 - 96.5
Granulocytes (%) 31,50 3.5 - 79
CD34+ cells (%) 0,86 0.35 - 4.18
cells/kg bw 6,1 2.2 - 17
Apheresis product values_AUTO
![Page 44: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/44.jpg)
Apheresis product values_ECP
N=9 Median Min - Max
RBC (T/l) 0,2 0.1 - 0.5
Hematocrit (%) 2 1 - 6
Platelets (G/l) 563 70 - 2310
Leukocyte (G/l) 50,9 0.9 - 73
Lymphocytes (%) 71,50 6 - 81.5
Monocytes (%) 14 7 - 42.5
MNC (%) 97,50 13.5 - 99.5
Granulocytes (%) 2,50 0.5 - 86.5
![Page 45: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/45.jpg)
patient Eff_CD34 Eff_lympho Eff_Ery Eff_G Eff_M Eff_Th Eff_MNC Eff_L
AUTO20110001 68,48 57,56 0,1 0,87 8,02 18,9 23,56 7,49
AUTO20110006 67,95 69,64 0,2 7,25 49,46 22,41 59,56 34,75
AUTO20110007 59,56 58,97 0,13 0,63 51,26 23 58,46 13,96
AUTO20110008 26,79 56,34 0,32 5,16 57,86 16,79 55,66 14,76
AUTO20110010 72,38 20,5 0,09 13,75 71,26 18,41 73,35 31,01
AUTO20110011 66,23 66,29 0,19 1,34 55,23 14,65 57,92 6,4
AUTO20110013 83,86 80,31 0,24 11,97 75,64 38,42 79,04 16,06
AUTO20120005 277,38 86,73 0,12 5,13 52,58 11,15 67,41 19,04
AUTO20120008 55,12 50,24 0,12 50,14 54,03 9,56 52,41 51,21
AUTO20120009 74,46 76,54 0,04 6 61,43 1,97 66,63 18,63
AUTO20120008 37,39 60,51 0,1 6,46 25,6 8,23 39 10,14
AUTO20120012 166,94 177,49 0,19 7,91 33,11 17,04 84,15 16,52
AUTO20130001 53,32 132,56 0,24 2,74 22,96 19,79 49,37 10,05
AUTO20130002 117,68 112,41 0,23 3,04 60,86 24,58 87,26 8,61
AUTO20130003 113,92 116,68 0,16 16,03 26,99 19,68 112,16 25,92
AUTO20130004 65,24 59,23 0,07 27,3 2,78 19,6 45,56 29,81
median 68,215 67,965 0,145 6,23 51,92 18,655 59,01 16,29
Collection efficiencies_AUTO
![Page 46: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/46.jpg)
Collection efficiencies_ECP
patient Eff_lympho Eff_Ery Eff_G Eff_M Eff_Th Eff_MNC Eff_L
ECP20110009 55,81 0,09 8,93 39,55 16,81 52,45 31,46
ECP20110023 32,38 0,14 0,09 16,99 0 26,97 6,62
ECP20110026 21,63 0,05 36,23 64,23 7,92 30,28 35,29
ECP20110031 15,39 0,03 0,15 28,33 0 18,84 4,69
ECP20110056 46,4 0,13 1,15 68,96 18,57 48,45 24,39
ECP20110060 16,62 0,03 0,19 29,39 1,83 19,24 5,41
ECP20110085 66,65 0,19 0,8 56,94 22,46 64,47 46,19
ECP20110088 56,84 0,12 0,2 56,15 29,16 56,48 23,56
ECP20120005 1,93 0,15 0,31 1,5 8,72 0,79 0,61
median 32,38 0,12 0,31 39,55 8,72 30,28 23,56
![Page 47: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/47.jpg)
![Page 48: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/48.jpg)
HPC IN DETAIL
![Page 49: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/49.jpg)
Correlation between CD34 positive cell yield and CD34 positive cell number in the peripheral blood_AUTO
y = 15,09x - 20,97 R² = 0,609
0
50
100
150
200
250
300
0 5 10 15 20
CD
34
_P
B_
vo
r
CD 34_10E06_kg
![Page 50: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/50.jpg)
Correlation between Eff_CD34 and Leukocyte count pre apheresis in peripheral blood_AUTO
y = 0.0019x + 29.404 R² = 0.3182
0
50
100
150
200
250
300
0 20000 40000 60000 80000
Eff
_C
D3
4
Leuko_PB_vor
![Page 51: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/51.jpg)
Correlation between CE CD34 positive cells and whole blood volume processed_AUTO
y = 0.3507x + 11.216 R² = 0.1306
0
50
100
150
200
250
300
0 100 200 300 400
Eff
_C
D34
WBV_kg BW
![Page 52: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/52.jpg)
Correlation between CE CD34 positive cells and number of cycles performed_AUTO
y = 2.9646x + 65.498 R² = 0.0814
0
50
100
150
200
250
300
0 5 10 15 20 25
Eff
_C
D3
4
No cycle
![Page 53: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/53.jpg)
Correlation between Eff_CD34 and collection preference in the peripheral blood_AUTO
0
20
40
60
80
100
120
140
160
180
200
20 30 40
Eff
_C
D3
4
Collection preferency
± Standarddeviation
![Page 54: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/54.jpg)
IN MONONUCLEAR CELLS =>
![Page 55: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/55.jpg)
Correlation between pre apheresis MNC and MNC_kg_buffy_AUTO
y = 5,3006x + 2338,6 R2 = 0,215
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
0.00 200.00 400.00 600.00 800.00 1000.00
MN
C_
vo
r
MNC_kg_buffy
![Page 56: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/56.jpg)
Whole blood processed / kg bw 1 = HPC and 2 = ECP
0
50
100
150
200
250
300
1 2
WB
V_
kg
BW
target
± Standardabweichung
![Page 57: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/57.jpg)
Number of cycles processed 1 = HPC and 2 = ECP
0
1
2
3
4
5
6
7
0 bis 22 bis 44 bis 8 8 bis12
12 bis15
15 bis20
20 bis25
Nu
mb
er
of
cyc
les
No cycle
1 2
target
![Page 58: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/58.jpg)
MNC count in blood pre 1 = HPC and 2 = ECP
0
1000
2000
3000
4000
5000
6000
7000
1 2
MN
C_
vo
r
target
± Standardabweichung
U – Test: p = 0.012
![Page 59: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/59.jpg)
MNC % in the product 1 = HPC and 2 = ECP
0
20
40
60
80
100
120
1 2
MN
C%
_b
uff
y
target
± Standardabweichung
U – Test: p = 0.0035
![Page 60: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/60.jpg)
MNC content in the product 1 = HPC and 2 = ECP
0
20000
40000
60000
80000
100000
120000
140000
160000
1 2
MN
C_
bu
ffy
target
± Standardabweichung
U – Test: p = 0.001
![Page 61: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/61.jpg)
MNC content in the product 1 = HPC and 2 = ECP
0
100
200
300
400
500
600
1 2
MN
C_
kg
_b
uff
y
target
± Standardabweichung
U – Test: p = 0.00049
![Page 62: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/62.jpg)
MNC collection ratio 1 = HPC and 2 = ECP
0
50
100
150
200
250
300
350
400
450
500
1 2
CR
MN
C
target
± Standardabweichung
U – Test: p = 0.95
![Page 63: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/63.jpg)
CAN WE PREDICT ANYTHING?
![Page 64: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/64.jpg)
Formel A CD34+ 106/kg = CD34+/µl PB / 10
• Simple formular
• Is only valid in aphereses procedures standardized for blood volume processed, stable collection efficiency (=> parameter settings!)
![Page 65: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/65.jpg)
Formel A CD34+ 106/kg = CD34+/µl PB / 10
y = 1.509x - 2.0975 R² = 0.6096
0
5
10
15
20
25
30
0 5 10 15 20
Fo
rme
l A
CD 34_10E06_kg
![Page 66: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/66.jpg)
Formel A CD34+ 106/kg = CD34+/µl PB / 10
y = 7.014ln(x) - 31.585 R² = 0.6052
-15
-10
-5
0
5
10
0 100 200 300
CD
34
/kg
- A
Eff_CD34
![Page 67: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/67.jpg)
Formel B C = [(AxD):Bx0.5]:1000
• A = CD34+ cells in peripheral blood [1003/ml]
• B = bodyweight [kg]
• C = CD34+ cells/kg bw in the product
• D = processed volume
• Estimation of an efficiency of 50%
![Page 68: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/68.jpg)
Formel B C = [(AxD):Bx0.5]:1000
y = 1.0115x + 0.1793 R² = 0.5085
0
2
4
6
8
10
12
14
16
18
20
0 5 10 15 20
Fo
rme
l B
CD 34_10E06_kg
![Page 69: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/69.jpg)
Formel B C = [(AxD):Bx0.5]:1000
y = 5.6977ln(x) - 24.851 R² = 0.6935
-12
-10
-8
-6
-4
-2
0
2
4
6
8
10
0 100 200 300
CD
34
/kg
- B
Eff_CD34
![Page 70: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/70.jpg)
WE CAN CALCULATE THE BLOOD VOLUME NEEDED TO BE PROCESSED
![Page 71: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/71.jpg)
D = [(C:0.5)xB:A]x1000
• A = CD34+ cells in peripheral blood [1003/ml]
• B = bodyweight [kg]
• C = CD34+ cells/kg bw in the product
• D = processed volume
• Estimation of an efficiency of 50%
![Page 72: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/72.jpg)
COLLECTION EFFICIENCY IN PLATELETS =>
![Page 73: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/73.jpg)
Platelet count in blood pre / CE 1 = HPC and 2 = ECP
0
10
20
30
40
50
60
70
0 100 200 300 400
CE
pla
tele
ts
Platelet count pre
1 2
target
y = -0.0594x + 27.137 R² = 0.1603
0
10
20
30
40
50
60
70
0 200 400
CE
pla
tele
t
Platelet count pre
![Page 74: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/74.jpg)
Correlation between platelets collection efficiency and number of cycles_AUTO
y = 0.8442x + 11.377 R² = 0.3657
0.00
5.00
10.00
15.00
20.00
25.00
30.00
35.00
40.00
45.00
0 5 10 15 20 25
Eff_
Th
No cycle
![Page 75: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/75.jpg)
Correlation between platelets collection efficiency and number of cycles_ECP
y = 4.9335x - 2.5341 R² = 0.1888
0.00
5.00
10.00
15.00
20.00
25.00
30.00
35.00
0 1 2 3 4 5
Eff_
Th
No cycle
![Page 76: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/76.jpg)
Platelet collection ratio versus number of cycle performed
y = 11.336ln(x) + 4.8427 R² = 0.3762
0
5
10
15
20
25
30
35
40
45
50
0 5 10 15 20 25
CR
pla
tele
ts
No cycle
![Page 77: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/77.jpg)
Side effects
• Hypocalcemia in a 10 kg child (Ca++ 0.86 mmol/l) complete restitutio after Ca 10% i.v.
• Leackage in the set => change of set, procedure was finished with 30 min delay
![Page 78: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/78.jpg)
Engraftment data
patient disease bw transplanted engrafted
AUTO20110001 ATRT 40,00 y y
AUTO20110006 AML recurrency 54,50 y y
AUTO20110007
Hepatoblastom
/hepatozelluläres
Karzinom 12,70
n n.a.
AUTO20110008 NB 10,25 n n.a.
AUTO20110010 PNET zentral 19,00 n n.a.
AUTO20110011 Ewing Sarkom 35,00 n n.a.
AUTO20110013 Ewing Sarkom 51,30 n n.a.
So far 2 / 7 patients are treated with Mega therapy and autologous rescue, both engrafted regulary.
![Page 79: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/79.jpg)
Triggering first chamber flush
• „Less operator input was required, although 26% of Spectra Optia apheresis procedures required triggering of the first chamber flush.” [1]
1 Brauninger Transfusion 2011
![Page 80: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/80.jpg)
• Flow pauses were 40% less frequent with the Spectra Optia (Table 2), likely by virtue of electronic interface management and an algorithm that automatically restarts the flow if pressure recovers quickly.
1 Brauninger Transfusion 2011
![Page 81: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/81.jpg)
1 Brauninger Transfusion 2011
![Page 82: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/82.jpg)
1 Brauninger Transfusion 2011
![Page 83: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/83.jpg)
Allogeneic stem cell apheresis outcomes with the COBE Spectra MNC
and the Spectra Optia v.5.0. (A) Collection efficiency for CD34+ cells is
negatively correlated with preapheresis CD34+ cell counts.
Preapheresis circulating CD34+ cells counts (X-axis) are plotted over
the corresponding collection efficiency CE1 (Y-axis) for this donor. (
,—) Results from the Spectra Optia; (, - - -) data fromthe COBE Spectra
MNC.
![Page 84: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/84.jpg)
Spectra Optia apheresis procedures are associated with better
collection efficiency in less well-mobilized donors. Preapheresis
circulating CD34+ cell counts were stratified as indicated. CE1
(mean + SEM) for each group is drawn to the Y-axis. Significant
differences between the groups. ( grey) Spectra Optia apheresis
procedures; (black ) COBE Spectra MNC collections.
![Page 85: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/85.jpg)
• PLT attrition was calculated as (1—postapheresis PLT count/preapheresis PLT count) x 100.
![Page 86: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/86.jpg)
Spectra Optia PLT attrition correlates with preapheresis PLT counts.
Preapheresis PLT counts (X-axis) are plotted over the corresponding
value for PLT attrition (Y-axis). ( ,—) Results from the Spectra Optia; (, -
- -) data fromthe COBE Spectra MNC.Markedly lower PLT attrition with
the Spectra Optia was observed particularly in donors with fewer PLTs.
![Page 87: Optia – Efficiency Considerations With MNC Collection...Manzone et.al. • JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY • 2007 • 55-63 . Blood volume is the volume of blood (both red](https://reader033.vdocuments.us/reader033/viewer/2022042023/5e7a95f3b10f2721275877c4/html5/thumbnails/87.jpg)
Conclusion
• Safe and efficient for HPC cell harvest in young children and adolescents
• Good correlation of PB HPC cell count and yield and easily predictable
• The same is found for MNC
• The number of cycles processed might be an independent factor for collection efficiency, not only for white blood cells
• Loss of platelets seems to be dependent on the number of cycles necessary