![Page 1: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/1.jpg)
1
Rebecca Haley, MD Medical Director, Cell Therapy
Amicus and Optia, the New Generation
Validation and Qualification of New Equipment for Clinical Applications
![Page 2: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/2.jpg)
2
Disclosures
None
Off Label Uses
Filgrastim used under IND for unrelated, allogeneic donations under NMDP protocol
![Page 3: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/3.jpg)
3
Today’s Apheresis Devices
Validation Experience
Collection Characteristics As Precursors To Biotech Products
Product Effect On Manufacturing Plans
Terumo Optia
Fenwal Amicus
![Page 4: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/4.jpg)
4
Today’s APHERESIS Devices
Evolution of Design
![Page 5: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/5.jpg)
5
Uses of Current Apheresis Machines - Collections
Hematopoietic Progenitor Cells, Apheresis (HPC, Apheresis)
for transplantation cell source
for CD34+ selection and/or expansion
Granulocyte collection
Mononuclear Cells (MNC)
for treatment post transplant
as starting material for other products
for dendritic cell vaccines
![Page 6: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/6.jpg)
6
Apheresis Machines as Collection/Treatment Devices and Laboratory Equipment
COBE Spectra and Fenwal CS3000 were workhorses for plasma exchange, plasma collection, red cell exchange
Both had protocols for selective blood product collection
Both could be used to selectively concentrate or refine other collected products
Ex. Red cell removal from bone marrow collections in ABO mismatched transplants
![Page 7: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/7.jpg)
7
General Rules of Design
The simpler the device:
More adaptable the device was
Products were more variable
Had fewer patient-protecting controls
Required operator skill
– Took longer to train staff
![Page 8: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/8.jpg)
8
Rules of Technologic Evolution
Devices became more sophisticated Computer programs were designed to run
specific protocols Defined protocols yield consistent products
Improvements were incremental with increasing software control and fewer operator interactions
Basic devices remained relatively stable
![Page 9: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/9.jpg)
9
Newer Devices Are Highly Engineered
Programming controls the run so that less operator intervention is needed
Higher precision in specific programs decrease adaptability
Patient protections are programmed in:
Air detectors to prevent air embolism
Limits on ACD infusion to the patient according to calculated blood volume to prevent Ca++ chelation problems
![Page 10: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/10.jpg)
10
Engineering Improvement
Spectra Optia is smaller, lighter than COBE Spectra
Kits and software packages are specific to purposes described
Amicus is smaller and lighter than CS 3000
Amicus and Spectra were developed for therapeutic plasma exchange, plasma collection
Specialized cellular collection programs were added later
![Page 11: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/11.jpg)
11
What Happens After the Retirement?
Will a single machine replace all the functions of the older machines?
What are the regulatory approvals on the newer devices?
What does each laboratory and treatment center need to do to qualify the new devices?
![Page 12: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/12.jpg)
12
Novel Uses Are Going to Different or Older Devices
There is no approved software control package in the Optia or Amicus for buffy coat refinement of bone marrow at this time
The COBE 2991 is still in service in many places
Newer devices, such as the Sepax 2 (Biosafe, Eysins, SW), can be used in labs where they are already in place – usually Cord Blood Banks
![Page 13: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/13.jpg)
13
Validation Experience
Spectra Optia and Amicus Put to the Test
![Page 14: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/14.jpg)
14
Bloodworks NW Validation Plan
A carefully controlled “in process validation” was the practical format
Lab monitoring of each step informed progress of each collection
Data collected allowed us to formulate effective collection processes going forward
Spectra Optia was validated
![Page 15: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/15.jpg)
15
Validation Plan
Study the method of operation of the machine Electronic data readouts, operational details
Decide on the “must-haves” in the new machine Safety for the donor (check alarms) Sterility of product Viability of product Yield of apheresis run
Measure what operating parameters will change Anticoagulant strategy Time of run Product volume
![Page 16: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/16.jpg)
16
Validation Run Plans
The Performance Qualification can practically be done on patients or allogeneic donors with oversight and controls
Other centers have set up plans ̶ new device is used on patients or donors where multiple collections are planned
The new and test devices alternate days
Run unstimulated research donors first to verify approximate collection efficiencies, then CD34+ stimulated subjects
Number of procedures planned: 10 to 50
![Page 17: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/17.jpg)
17
A Sample from the collection bag was taken 2 hours into collection
CD34+ count was done; lab calculated the number of mL of product needed to fulfill the collection goal Assumption: the CD34+ content would be stable throughout
collection Collection time and liters of blood processed were
determined by this calculation
ACD only anticoagulant was used as per Optia instructions – no heparin, no aspirin
Our Approach to Validation
![Page 18: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/18.jpg)
18
y = -0.0016x + 0.6804
0%
20%
40%
60%
80%
100%
120%
0 50 100 150 200 250
% E
ffic
ien
cy
X 103 CD34+ cells/ mL n = 65
Pre C34+ vs Collection Efficiency
![Page 19: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/19.jpg)
19
Peripheral Blood CD34+
Result
Predictive Equation
< 90x103 cells/ml 𝑮𝒐𝒂𝒍 × 𝟏. 𝟏
. 𝟒𝟎= 𝑻𝒂𝒓𝒈𝒆𝒕
≥ 90x103 cells/ml 𝑮𝒐𝒂𝒍 × 𝟏. 𝟏
. 𝟑𝟎= 𝑻𝒂𝒓𝒈𝒆𝒕
𝑻𝒂𝒓𝒈𝒆𝒕
𝑷𝒓𝒆 𝑹𝒆𝒔𝒖𝒍𝒕× 𝟏𝟎𝟎𝟎 = 𝑻𝒂𝒓𝒈𝒆𝒕 𝑽𝒐𝒍, 𝒎𝒍
Predictive Equation
![Page 20: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/20.jpg)
20
0
5000
10000
15000
20000
25000 COBESpectra
+2 HourCD34Analysis
PredictiveEquation
13.9 15 L
24 L
Average Collection Run Volumes
![Page 21: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/21.jpg)
21
Optia Validation – Lessons Learned
Predictive CD34+ counts were very valuable
Very high white-count donors, such as NMDP donors, give slightly lower collection efficiencies
Investigators and transplant centers need to order “expected plasma volume” or “minimum plasma volume” instead of “add 200 mL plasma to product”
Orders that are based on volume (xx liters) run through the device are not predictive, need predictive formula
![Page 22: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/22.jpg)
22
20%
30%
40%
50%
60%
70%
80%
90%
100%
110%
120%
9/16/2013 11/20/2013 2/24/2014 5/13/2014 7/10/2014 10/1/2014 12/16/2014
(207%)
Operator Efficiencies
Training Opportunities
![Page 23: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/23.jpg)
23
Amicus Product RBC and Platelet Content
ABO type mismatched allogeneic transplants request <20 mL ABO mismatched red blood cells (RBC)
Company instructions say to expect 15%
Experience of most centers using the device (my survey) reported average hematocrits of 8 to 12%
In a 200+ mL product this is approaching the 20 mL point of caution
Amicus is the most platelet-sparing of the devices
![Page 24: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/24.jpg)
24
Amicus Validation – Lessons Learned
Volume of blood processed/min apparently affects collection efficiency
Combination of ACD-A and heparin can be used, but higher flow rates are problematic
Platelets are spared
Hematocrits are higher than Optia products One large institution, not measuring directly,
estimated at 7 to 9%
Company information – expect 15%
Bergstaler E et al. J. Clin Apheresis 2011:26;186
![Page 25: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/25.jpg)
25
Data Display Characteristics
At the end of the procedure FACT Standards require labeling of the bag with contents before takedown and disconnection from the donor
Amicus does not provide the readout of mL in product or final volume ACD-A
Product must be removed and weighed
ACD-A must be calculated
Optia provides data display with the above information on the Graphical User Interface
![Page 26: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/26.jpg)
26
Collection characteristics as precursors to biotech products
Spectra Optia and Amicus: Collection Characteristics
![Page 27: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/27.jpg)
27
Requirements for Late Stage Clinical Trial or Licensed Products Raw Material
Qualification of personnel
Qualification process is described by contract
Continuous exchange of quality metrics between supplier and manufacturer
Quality Review of your collection operation – Site qualification – Regulatory status review of your facility – Technical assessment of IT system for speed
and privacy
Physical appearance of facility and patient comfort features
![Page 28: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/28.jpg)
28
Product and Collection Process Requirements
Example – MNCs for Dendritic Cell Manufacture
Identification of Donor/Patient
Method of collection – Regulatory qualification of
collection device – Certification of clinical grade
anticoagulants and IV solutions
# of expected MNC in product
Product labeled with excipient contents, ex. mL ACD
***Hematocrit < 4%
![Page 29: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/29.jpg)
29
Donor Experience
Differences for the Donor
Optia - lower volume
Little citrate toxicity
Sparing of platelets
Less time on the machine
![Page 30: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/30.jpg)
30
Retrospective: Predicting 40% efficiency
Requested CD34+
Volume Processed
Total CD34+
% of Goal Collected
Efficiency
435 11741 889 204 50.14
144 10573 175 122 43.56
325 14539 478 147 53.03
440 23625 400 91* 65.12
305 8612 801 262 68.90
385 14241 645 168 60.39
732 20030 822 112 50.66
40% efficiency was too conservative *poor mobilizer
![Page 31: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/31.jpg)
31
Summary
Newer apheresis devices advantages: Smaller size Easier portability More run automation specific to the procedure
selected Higher efficiency (formula adjusted to 50%, Optia) Platelet sparing Flexibility of the COBE Spectra and CS3000 is less
with successors Donor/patient safety are improved
![Page 32: Validation and Qualification of New Equipment for Clinical Applications€¦ · Amicus and Optia, the New Generation Validation and Qualification of New Equipment for Clinical Applications](https://reader030.vdocuments.us/reader030/viewer/2022041000/5ebe7676a2c61c498b2f381b/html5/thumbnails/32.jpg)
32
Bergstaler E et al. J. Clin Apheresis 2011:26;186
Brauninger S et al. Transfusion 2012;52:1137
Dannull J et.al. JCI 2013; 123(7):3135 (Biotech example)
Flommersfeld et al, Transfus. and Aph. Science 2013, in press
Ikeda K et al, Transfusion 2007;47:1234
Reinhardt P et al, Transfusion 2011;51:1321
Rox JM et al, Vox Sanguinis 2011; 101(Suppl 1):337
Steininger et al, Transfusion 2013;53:747
Vitt et al, 2012 ASFA Meeting, abstract
References