non-invasive glucose determination in the human eye

8
Non-invasive glucose determination in the human eye Wolfgang Schrader a, * , Petra Meuer b , Ju ¨rgen Popp c , Wolfgang Kiefer b , Johannes-Ulrich Menzebach d , Bernhard Schrader d a Universita ¨tsaugenklinik, Universita ¨t Wu ¨rzburg, Josef-Schneider-Str. 11, D-97080 Wu ¨rzburg, Germany b Institut fu ¨r Physikalische Chemie, Universita ¨t Wu ¨rzburg, Am Hubland, D-97074 Wu ¨rzburg, Germany c Institut fu ¨r Physikalische Chemie, Friedrich-Schiller-Universita ¨t Jena, Helmholtzweg 4, D-07743 Jena, Germany d Institut fu ¨r Physikalische und Theoretische Chemie, Universita ¨t Duisburg-Essen, Soniusweg 20, D-45259 Essen, Germany Received 21 September 2004; revised 16 October 2004; accepted 18 October 2004 Dedicated to Professor Hiroaki Takahashi, Tokyo! One of the authors (B.S.) is grateful for the instruction in 1966 to perform the Normal Coordinate Analysis of Molecular Crystals and great friendship and scientific cooperation since Abstract For non-invasive in vivo glucose determinations by means of near-infrared spectroscopy, the anterior chamber of the human eye is a promising site. An optical set-up for the non-invasive glucose determination in the human eye precisely in the anterior chamber with a beam reflected from the surface of the eye lens is presented here. As the anterior chamber has a depth of 3.13G0.50 mm, the beam follows an optical path of 5.3–7.3 mm depending on the angle of incidence, which is individually constant. We will show that it is possible to acquire good concentration predictions for physiological glucose concentrations with such a long optical path. A chemometric study of NIR glucose spectra with concentrations of glucose in water of 10–350 mg/dL (0.56–1.94 mmol/L) resulted in a calibration model which was able to predict physiological glucose concentrations with a root mean square error of prediction RMSEP Test Z15.41 mg/dL. The Clarke error grid diagram shows that the model performs well according to medical impact. Using a first in vivo set-up, the precision is not sufficient for a reliable prediction of glucose concentration, especially due to the flickering of the patient’s eye and the low reflectivity of the eye lens. Therefore, we have designed a new in vivo set-up: a prototype for a self-monitoring device with controlled geometry and laser radiation at several distinct wavelengths instead of the halogen lamp as light source. This allows a far higher signal/noise ratio under much better reproducible geometrical conditions and at the same time a much smaller necessary light flux. q 2004 Elsevier B.V. All rights reserved. Keywords: NIR absorption; Aqueous humour; Glucose; Non-invasive determination 1. Introduction Diabetes mellitus has increasingly become a health threat in industrialized countries. The prevalence of diabetes mellitus has doubled within the last 30 years in these countries to about 5% of the population (in Germany, from two to four million people) [1]. As shown by the diabetes control and complications trial (DCCT), late complications occur less frequently in patients who adjust their insulin intake to their eating behaviour with multiple injections rather than keep a strict diet with only two injections per day [2]. Therefore, diabetic patients have to rely on frequent blood glucose measurements to control their blood glucose levels. Most diabetics use a self-monitoring device, which works with a small droplet of blood drawn from the fingertip. This can turn out to be quite painful in the long run, and even lead to severe sensitivity loss of the fingertips. To ensure a high patient compliance with this therapy scheme, the glucose determinations should be as painless and convenient for the patient as possible. Therefore, various approaches for minimal and non- invasive determination have been suggested [3]. 0022-2860/$ - see front matter q 2004 Elsevier B.V. All rights reserved. doi:10.1016/j.molstruc.2004.10.115 Journal of Molecular Structure 735–736 (2005) 299–306 www.elsevier.com/locate/molstruc * Corresponding author. Tel.: C49 931 20120610; fax: C49 931 20120490. E-mail address: [email protected] (W. Schrader).

Upload: wolfgang-schrader

Post on 21-Jun-2016

223 views

Category:

Documents


6 download

TRANSCRIPT

Page 1: Non-invasive glucose determination in the human eye

Non-invasive glucose determination in the human eye

Wolfgang Schradera,*, Petra Meuerb, Jurgen Poppc, Wolfgang Kieferb,Johannes-Ulrich Menzebachd, Bernhard Schraderd

aUniversitatsaugenklinik, Universitat Wurzburg, Josef-Schneider-Str. 11, D-97080 Wurzburg, GermanybInstitut fur Physikalische Chemie, Universitat Wurzburg, Am Hubland, D-97074 Wurzburg, Germany

cInstitut fur Physikalische Chemie, Friedrich-Schiller-Universitat Jena, Helmholtzweg 4, D-07743 Jena, GermanydInstitut fur Physikalische und Theoretische Chemie, Universitat Duisburg-Essen, Soniusweg 20, D-45259 Essen, Germany

Received 21 September 2004; revised 16 October 2004; accepted 18 October 2004

Dedicated to Professor Hiroaki Takahashi, Tokyo! One of the authors (B.S.) is grateful for the instruction in 1966 to perform the Normal

Coordinate Analysis of Molecular Crystals and great friendship and scientific cooperation since

Abstract

For non-invasive in vivo glucose determinations by means of near-infrared spectroscopy, the anterior chamber of the human eye is a

promising site. An optical set-up for the non-invasive glucose determination in the human eye precisely in the anterior chamber with a beam

reflected from the surface of the eye lens is presented here. As the anterior chamber has a depth of 3.13G0.50 mm, the beam follows an

optical path of 5.3–7.3 mm depending on the angle of incidence, which is individually constant. We will show that it is possible to acquire

good concentration predictions for physiological glucose concentrations with such a long optical path. A chemometric study of NIR glucose

spectra with concentrations of glucose in water of 10–350 mg/dL (0.56–1.94 mmol/L) resulted in a calibration model which was able to

predict physiological glucose concentrations with a root mean square error of prediction RMSEPTestZ15.41 mg/dL. The Clarke error grid

diagram shows that the model performs well according to medical impact. Using a first in vivo set-up, the precision is not sufficient for a

reliable prediction of glucose concentration, especially due to the flickering of the patient’s eye and the low reflectivity of the eye lens.

Therefore, we have designed a new in vivo set-up: a prototype for a self-monitoring device with controlled geometry and laser radiation at

several distinct wavelengths instead of the halogen lamp as light source. This allows a far higher signal/noise ratio under much better

reproducible geometrical conditions and at the same time a much smaller necessary light flux.

q 2004 Elsevier B.V. All rights reserved.

Keywords: NIR absorption; Aqueous humour; Glucose; Non-invasive determination

1. Introduction

Diabetes mellitus has increasingly become a health threat

in industrialized countries. The prevalence of diabetes

mellitus has doubled within the last 30 years in these

countries to about 5% of the population (in Germany, from

two to four million people) [1]. As shown by the diabetes

control and complications trial (DCCT), late complications

occur less frequently in patients who adjust their insulin

0022-2860/$ - see front matter q 2004 Elsevier B.V. All rights reserved.

doi:10.1016/j.molstruc.2004.10.115

* Corresponding author. Tel.: C49 931 20120610; fax: C49 931

20120490.

E-mail address: [email protected]

(W. Schrader).

intake to their eating behaviour with multiple injections

rather than keep a strict diet with only two injections per day

[2]. Therefore, diabetic patients have to rely on frequent

blood glucose measurements to control their blood glucose

levels.

Most diabetics use a self-monitoring device, which

works with a small droplet of blood drawn from the

fingertip. This can turn out to be quite painful in the long

run, and even lead to severe sensitivity loss of the fingertips.

To ensure a high patient compliance with this therapy

scheme, the glucose determinations should be as painless

and convenient for the patient as possible.

Therefore, various approaches for minimal and non-

invasive determination have been suggested [3].

Journal of Molecular Structure 735–736 (2005) 299–306

www.elsevier.com/locate/molstruc

Page 2: Non-invasive glucose determination in the human eye

W. Schrader et al. / Journal of Molecular Structure 735–736 (2005) 299–306300

Non-invasive measurement sites on the human body have to

fulfill certain criteria [4]. The site should be easily

accessible at least with a quartz fibre, it should be

temperature stable or temperature controllable, it has to

contain glucose in measurable concentrations, and the

glucose concentration at that site has to have a constant

relation towards that in the blood.

Specific studies have demonstrated that NIR spec-

troscopy represents a promising tool for the non-invasive

prediction of blood glucose concentration. Diffuse reflection

measurements were performed by Robinson et al. [5] at the

finger with different instrument configurations, by Marbach

et al. [6] and Heise et al. [7] of the oral muscosa, by

Fischbacher et al. [8], Danzer et al. [9] and Muller et al. [10]

on the middle finger of the right hand and by Malin et al.

[11] on the forearm. Burmeister et al. [12] collected

transmission spectra through the tongue. Another method

for non-invasive glucose determination is the combination

of photoacoustic spectroscopy with modulated laser diodes

in aqueous solutions as reported by Spanner et al. [13].

In all these studies, limitations were cited that affect the

reliability of the method. These limitations included

sensitivity, sampling problems, time lag, calibration bias,

long-term reproducibility, and instrument noise, as well as

the fact that glucose concentrations in these body regions

are very low compared to the ones in blood [3].

Additionally, accurate non-invasive estimation of blood

glucose is limited at present by the dynamic nature of the so

far used sample sites: the skin and living tissue of the patient

[14]. Chemical, structural, and physiological variations

occur that produce dramatic changes in the optical proper-

ties of the tissue sample. Temperature variations are another

obstacle, as NIR spectra are very sensitive towards

temperature changes especially when dealing with aqueous

matrices [15]. The measurement is further complicated by

the varying background signals of other substances present

in the blood or tissue, e.g. body fat and proteins. The

analysis of in vivo spectra is quite complex as the spectra

exhibit very broad, overlapping bands.

Another possible sample location, which has not been

extensively investigated so far, is the eye. Wickstedt et al.

[16] and Borchert et al. [17] reported on Raman spectro-

scopic measurements of rabbit aqueous humour. Backhaus

et al. [18,19], Menzebach [20] and Schrader et al. [21,22]

proposed in vivo NIR transmission measurements of the

aqueous humour, where the light is reflected at the front of

the eye lens. Additionally Menzebach [20] discussed

different techniques for the realization of glucose concen-

tration measurement in the human eye, e.g. the use of optical

rotation. This has also been investigated by Cameron and

Cote [23] and Cameron et al. [24].

The aqueous humour is a liquid, which fills the anterior

chamber of the human eye. This chamber is located between

the cornea and the lens with a depth of 3.13G0.50 mm. [25].

It can be easily reached by spectroscopic means—in the

wavelength range where the cornea is transparent. Due to its

location in the orbital, it is well temperature stabilized, even

more so as the cornea with its tear film regulates the

temperature of the eye. The aqueous humour is a kind of

ultra filtrate of the human blood, and is responsible for

controlling the intra-ocular pressure and the nourishment of

the lens, as well as removing the intermediate catabolic

products from the cornea [26]. It contains glucose in

concentrations of 65–85 mg/dL (3.6–4.8 mmol/L), corre-

sponding to about 63–76% of the glucose content in the

blood [27,28].

A glucose determination in the aqueous humour is

possible when some requirements are fulfilled. The optical

path has to have an intra-individually reproducible length

and has to allow an easy non-invasive measurement.

Cameron et al. suggested a path for polarimetry measure-

ments in the anterior chamber that requires contact to the

cornea and therefore anaesthesia [24]. The path under

investigation here uses the third of the Purkinje–Sanson

images [29]. The four Purkinje–Sanson images are pro-

duced by reflections from the front and back of the cornea

and the lens (see Fig. 6).

By focusing on the reflection on the front of the eye lens,

the returning light travels twice through the anterior

chamber thus covering a distance of about 7 mm at an

angle of about 458 from the normal. For the in vitro

experiments, a 5 mm cell is used [30].

Near-infrared spectra are not very expressive for aqueous

solutions with small glucose concentrations as the spectra

are dominated by water absorption. However, an analysis

can be performed with the use of multivariate procedures.

To establish a valid chemometric model for the in vivo

determination of glucose in the aqueous humour, we

developed an in vitro calibration model and present an

experimental set-up for the in vivo non-invasive determi-

nation of glucose.

2. Material and methods

D(C)-Glucose for biochemistry (MERCK) was used

without further purification. Each glucose sample was

weighed separately to achieve stochastic independence of

the samples. The samples were dissolved in highly purified

water and conserved by adding 0.05% NaN3. All samples

were prepared freshly each day. Prior to the measurement,

the samples were thermostated at about 36 8C. The reference

concentrations were measured with the hexokinase method

on a Hitachi 911 analyser from Roche (instrument precision

G0.5% within a day).

The near-infrared spectra were recorded with a Vector

22/N-C interferometer system from Bruker (Germany) with

a Peltier-cooled InGaAs detector. The detector (D 427/N,

Bruker, Germany) covers a range between 4000 and

12,800 cmK1 and has a noise equivalent power NEP of !2!10K13 W/HzK1/2. The data were collected between

4000 and 12,500 cmK1 with a spectral resolution of

Page 3: Non-invasive glucose determination in the human eye

Fig. 1. Typical near-infrared spectra of aqueous glucose solutions with

concentrations in the physiological range between 10 and 350 mg/dL

(0.56–1.94 mmol/L). Only the regions 5377.2–6542.1 cmK1 and 7170.9–

11001.2 cmK1 showed a sufficient signal/noise ratio and were used for the

PLS regression.

W. Schrader et al. / Journal of Molecular Structure 735–736 (2005) 299–306 301

8 cmK1. The interferograms were collected double-sided

forward–backward.

All samples were measured in a Suprasil thermo cell

(Hellma, Germany) with a fixed path length of 5 mm. This

cell has a water-cooled jacket, which was attached to a

thermostat (MV-4, Julabo). The temperature in the sample

cell was controlled at 37.0 8C (G0.1 8C). Prior to the

measurements, the solution in the sample cell was checked

for air bubbles, which were removed if present.

Each spectrum collected was obtained as an average of

100 scans/sample with a total scan time of 60 s. Two sets of

35 samples each were prepared with varying concentrations,

ranging from 10 to 350 mg/dL (0.56–1.94 mmol/L) in

10 mg/dL steps.

Fig. 2. Differential spectra of aqua bidest (thickness of 1 mm) under various temp

were taken with a Perkin–Elmer Lambda 9 Spectrometer.

The chemometric analysis was performed as described in

the dissertation of Meuer [31] with the program package

QUANT from OPUS (Bruker, Germany) [32] using the

implemented PLS-1 algorithm and in-house developed

routines for Matlab (The Mathworks, Natick MA, USA)

[33,34]. To put the results into a medical context, the

concentration correlations are plotted in a Clarke error grid

produced with the BD Error Grid.xls program [35].

3. Results and discussion

3.1. In vitro glucose determination with 5 mm optical path

Near-infrared absorption spectra of aqueous glucose

solution (see Fig. 1) are dominated by water, and show very

little variation between different concentrations in the

physiological range.

NIR spectra of water exhibit four bands in the NIR region

[36,37], which are combinations of the fundamental

vibrations: the symmetric stretching mode n1 (ns) at

3615 cmK1, the antisymmetric stretching mode n3 (na) at

3450 cmK1 and the bending mode n2 (d) at 1640 cmK1. In the

NIR these give rise to the combination mode n1Cn3 around

7040 cmK1 (1420 nm), the n1Cn2Cn3 around 8620 cmK1

(1160 nm) and around 10,340 cmK1 (967.1 nm) the 2n1Cn3

combination mode. The combination mode n2Cn3 around

5260 cmK1 (1901 nm) is not observed due to the high sample

thickness of 5 mm, which leads to a cut off lower than

5350 cmK1 (1869 nm). The same applies to the combination

mode n1Cn3 around 7040 cmK1 (1420 nm). The position

and width of OH-vibrations are usually [40] strongly

temperature-dependent. This is shown in Fig. 2, where the

absorption spectrum of water at temperatures 32–38 8C is

recorded relative to that at 40 8C.

The spectra show isosbestic points—where the tempera-

ture-dependence is zero. They may be used as fixed points to

eratures. As a reference, the water spectrum at 40 8C was chosen. Samples

Page 4: Non-invasive glucose determination in the human eye

Table 1

Isosbestic points of the NIR absorption spectrum of water, with no

temperature dependence in the NIR spectrum

cmK1 nm

8770 1140

8396 1191

7655 1305

6944 1440

5591 1790

4590 2180

W. Schrader et al. / Journal of Molecular Structure 735–736 (2005) 299–306302

determine the sample thickness, and are compiled in

Table 1.

The absorption spectrum of glucose relative to water is

shown in Fig. 3. There is a combination of the IR bands of

glucose at 3310 and 1457 cmK1 (4767 cmK1), appearing at

2120 nm, further, another combination of the IR bands at

2945 and 1457 cmK1 appearing at 2260 nm. At 1690 nm,

there is the first overtone of the IR band at 2945 cmK1, a CH

stretching vibration. Since it is less dependent on tempera-

ture changes than an OH stretching modes, Hazen et al. [15]

recommended this band as useful for glucose determination.

Finally, there is at 1560 nm the overtone of a OH-stretching

vibration at 3310 cmK1. In order to see the bands due to

glucose in the diagram, the concentration of the solutions

giving the spectra is much higher (1–10%) than the

physiological one (in the order of 0.1%).

Classical univariate calibration of glucose in water is

impossible, since we found no wavenumber where sufficient

selectivity for glucose exists. Hence, multivariate cali-

bration techniques need to be applied.

Two spectral regions, 5377.2–6542.1 cmK1, and 7170.9–

11001.2 cmK1 (see Fig. 1), showing an absorbance lower

than three absorbance units were selected for partial least

squares (PLS) regression.

Fig. 3. Absorption spectra of glucose in aqua bidest at various concentrations rela

Samples were taken with a Perkin–Elmer Lambda 9 Spectrometer.

A plot of the experimental versus the predicted values is

shown in Fig. 4. The minimum cross-validated root mean

squared error of prediction (RMSEPCV) was attained at 11

PLS factors and resulted in a cross-validated R2 of 97.87%

and a RMSEPCV of 14.44 mg/dL. The 11 factors cannot be

assigned to physical parameters. The system’s variable

parameters shift the broad unseparated NIR bands, and thus

have non-linear effects on the spectral channels. Details are

described in Ref. [31].

To determine the medical impact for in vivo glucose

measurements, the results were plotted in a Clarke error grid

[38,39] in Fig. 5. This error grid analysis offers a quick

estimation of the medical accuracy of the measurement.

Data from the test device are plotted against the results of a

reference method. A scatter diagram is established and

divided into five zones [38,39]. The five zones A–E show

varying degrees of accuracy of glucose estimation, which

correlates with an adequate or inadequate treatment.

Explicitly, zone A: no effect on clinical action; B: altered

clinical action of little or no effect on clinical outcome; C:

altered clinical action—likely to effect clinical outcome; D:

altered clinical action—could have significant medical risk;

E: altered clinical action—could have dangerous

consequences.

With one exception, all data points fall in zone A and

only one prediction is located in zone B. The calibration

model used for the prediction of the data performs well, and

the predictions result in medically uncritical deviations.

3.2. Experimental set-up for the non-invasive in vivo

determination of glucose

For a non-invasive in vivo glucose determination in the

eye, a suitable experimental set-up is needed. It uses a path

through the anterior chamber of the human eye with an

tive to the spectrum of water. The temperature was kept constant at 37 8C.

Page 5: Non-invasive glucose determination in the human eye

Fig. 4. Glucose concentration correlation plot of experimental versus

predicted glucose concentrations for the 131 spectra of the training set. The

regression was performed in the spectral ranges from 5377.2 to 6542.1 cmK1

and 7170.9 to 11001.2 cmK1.

Fig. 6. Positions of the four Purkinje–Sanson images relative to the incident

light. L light source, 1: reflection from the front of the cornea, 2: reflection

from the back of the cornea, 3: reflection from the front of the lens, 4:

reflection from the back of the lens.

W. Schrader et al. / Journal of Molecular Structure 735–736 (2005) 299–306 303

intra-individually constant length using the reflection

according to the third Purkinje–Sanson image. Light

reflected there travels twice through the anterior chamber

covering an individually constant length of about 6–7 mm,

and can be used to obtain transmission spectra of the

aqueous humour. A sideways cut through the human eye

and the positions of the four Purkinje–Sanson images are

shown in Fig. 6.

Fig. 5. Clark error grid diagram of the correlation of true versus calculated

glucose concentrations for the 66 spectra of the set with ‘unknown’ samples.

Zone A: no effect on clinical action, B: altered clinical action or little or no

effect on clinical outcome, C: altered clinical action—likely to affect

clinical outcome, D: altered clinical action—could have significant medical

risk, E: altered clinical action—could have dangerous consequences. The

diagram was produced with the BD Error Grid.Xls program [35].

In vivo near-infrared spectra are recorded with a set-up

illustrated in Fig. 7(A) on a Vector 22/N-C interferometer

(Bruker, Germany) equipped with an InGaAs detector. The

spectrometer is modified to allow the connection to a quartz

fibre. The internal light source of the interferometer is

switched off. The connecting quartz fibre has a diameter of

1 mm and a length of 2 m (Optran WF 1000/1100 N,

CeramOptec, Germany). The external light source is an air-

cooled tungsten lamp (12 V/30 W, model 64261, Osram,

Germany) running with a stabilized current of 2.5 A. The in

vivo optics in front of the eye consists of four plane-convex

quartz lenses and two variable iris diaphragms (Spindler &

Hoyer, Germany). The quartz fibre is mounted in the image

plane of a reflex camera (OM-1, Olympus, Germany) to

allow a visual control of the focus plane. Additionally, two

optical filters were used: a RG 665 1.0 (Schott, Germany)

between the light source and the eye to limit the light range

reaching the eye and a RG 850 1.0 (Schott, Germany) right

in front of the quartz fibre to suppress stray light from the

visible range reaching the spectrometer. This set-up was

optimised using a ray tracing program of OpticsLab

(Science Lab Software, Carlsbad, CA).

The spot size of the light beam for illumination and

observation are regulated with variable iris diaphragms.

The focus is directed by the observer. The observer, who

is used to slit lamp examinations, adjusts the third Purkinje–

Sanson image to the centre of the camera image.

The camera is focused onto the lens surface, where the

third Purkinje–Sanson image is reflected. The light reflected

from the lens is then collected with another two plane-

convex quartz lenses, and directed into a quartz fibre with a

diameter of 1 mm. The end of the quartz fibre is mounted in

the image plane of the reflex camera. This set-up allows

the observer to focus the light exactly into the third

Page 6: Non-invasive glucose determination in the human eye

Fig. 7. (A) Ray tracing of the set-up for non-invasive measurements in the human eye. (B) Typical in vivo spectrum of the human eye collected with the

described set-up (light source 2.5 A, 12 V, with absorption filters Schott RG 665 1.0, RG 850 1.0 (not shown in the tracing diagram, resolution 8 cmK1,

32 scans).

W. Schrader et al. / Journal of Molecular Structure 735–736 (2005) 299–306304

Purkinje-Spot when the mirror is down, and to take spectra

with the mirror up.

Fig. 8 shows the result of an in vivo measurement of the

glucose concentration with an arrangement as seen in

Fig. 7(A). The diagram proves that the glucose concen-

tration can be measured with an arrangement such as

Fig. 7(A). However, the mean deviation of glucose

concentration in the eye—optically measured—and that of

the blood analysis is 28 mg/dL equivalent to about 21%.

Fig. 8. Glucose concentrations in the capillary blood and in the aqueous humour of

authors, B.S.). Values in the aqueous humour follow the blood values with a late

This is somewhat high as regards the requirements of a

reliable measurement, and is due to:

1.

the

ncy

the head holder did allow movements of the patient, and

the observer had to adjust for them during each

measurement individually, making it difficult to measure

exactly at the geometrically optimal arrangement and

2.

the reflectance of the front surface of the eye lens is only

about 0.1%.

eye during an oral glucose tolerance test in a human subject (one of the

of about 20 min. Spectra were analysed with PCR and PRESS.

Page 7: Non-invasive glucose determination in the human eye

W. Schrader et al. / Journal of Molecular Structure 735–736 (2005) 299–306 305

Both effects can be compensated for as shown in the next

paragraph.

We plan to improve the set-up to make a self-monitoring

device possible. The tungsten lamp light source L in Fig. 6

will be replaced by a set of lasers, emitting at wavelengths at

suitable positions of the glucose and water spectrum:

maxima and minima as well as at isosbestic points

(Table 1). A monitor detector measuring the intensity of

the beams 1 and 2 in Fig. 6 allows the measurement at

position 3 only, when the measuring beam is optimally

adjusted. An acoustical signal will help the patient to find

this optimal beam. Having a set of lasers with spectrally

very sharp lines instead of the continuous spectrum of the

halogen lamp allows a much higher signal/noise ratio of

the signals together with much smaller light flux used for the

measurement [22,41].

4. Conclusions and outlook

It has been shown that physiological concentrations of

glucose in water can be determined from NIR spectra

collected with 5 mm optical path length. The correlation

between experimental and predicted values indicate that the

models are based on glucose specific spectral information,

and that these models are potentially reliable over time.

A suitable set-up for the in vivo non-invasive NIR

measurement of glucose in the human eye is presented.

Using this set-up, the first in vivo spectra were acquired and

the glucose concentrations determined. Even though the

medically desired prediction error of less than 10 mg/dL has

not yet been reached, the results are very promising, and

show the feasibility of in vivo determination of glucose in

the human eye. A way to improve the quality of this set-up

and to construct a self-monitoring apparatus for a diabetic

patient is described.

Acknowledgements

Support from the Erweiterte Forschungsforderung der

Universitat Wurzburg, Grant 2c/1998, and the Deutsche

Forschungsgemeinschaft, DFG Grant Schr 598/2-1, is

highly acknowledged. J. Popp highly acknowledges the

support from the Freistaat Bayern (Bayerisches Habilita-

tionsstipendium) and W. Kiefer acknowledges financial

support from the Fonds der Chemischen Industrie.

References

[1] H.U. Janka, D. Michaelis, Z. Arztl. Fortbild. Qualitaetssich 96 (2002)

159.

[2] The diabetes control and complications trial research group, JAMA

276 (1996) 1409.

[3] J. Pickup, L. McCartney, O. Rolinski, D. Birch, Br. Med. J. 319 (1999)

1289.

[4] P.G. Steffes, Diabetes Technol. Ther. 1 (1999) 129.

[5] M.R. Robinson, R.P. Eaton, D.M. Haaland, G.W. Koepp,

E.V. Thomas, B.R. Stallard, P.L. Robinson, Clin. Chem. 38 (1992)

1618.

[6] R. Marbach, T. Koshinsky, F.A. Gries, H.M. Heise, Appl. Spectrosc.

47 (1993) 874.

[7] H.M. Heise, R. Marbach, T. Koschinsky, F.A. Gries, Artif. Organs 18

(1994) 439.

[8] C. Fischbacher, K.-U. Jagemann, K. Danzer, U.A. Muller,

L. Papenkordt, J. Schuler, Fresenius J. Anal. Chem. 359 (1997) 78.

[9] K. Danzer, C. Fischbacher, K.U. Jagemann, K.J. Reichelt, LEOS

Newslett. 1988 (1998) 9.

[10] U.A. Muller, B. Mertes, C. Fischbacher, K.U. Jageman, K. Danzer,

Int. J. Artif. Organs 20 (1997) 285.

[11] S.F. Malin, T.L. Ruchti, T.B. Blank, S.N. Thennadil, S.L. Monfre,

Clin. Chem. 45 (1999) 1651.

[12] J.J. Burmeister, M.A. Arnold, G.W. Small, Diabetes Technol. Ther. 2

(2000) 5.

[13] G. Spanner, R. Niesner, Fresenius J. Anal. Chem. 354 (1996) 306.

[14] O.S. Khalil, Clin. Chem. 45 (1999) 165.

[15] K.H. Hazen, M.A. Arnold, G.W. Small, Appl. Spectrosc. 48 (1994)

477.

[16] J.P. Wicksted, R.J. Erkens, M. Motamedi, W.F. March, Appl.

Spectrosc. 49 (1995) 987.

[17] M.S. Borchert, M.C. Storrie-Lombardi, J.L. Lambert, Diabetes Tech.

Ther. 1 (1999) 145.

[18] J. Backhaus, D. Bocker, B. Schrader, W. Schrader, H.U. Menzebach,

E. Schmidt, Vorrichtung zur in vivo Bestimmung einer optischen

Eigenschaft des Kammerwassers des Auges, European Patent Nr.

93119838, Feb 24 1999.

[19] J. Backhaus, D. Bocker, B. Schrader, W. Schrader, H.U. Menzebach,

E. Schmidt, Device for the in vivo determination of an optical

property of the aqueous humor of the eye, US Patent 5535743, Jul 16

1996.

[20] J.U. Menzebach, Nichtinvasive Diagnostik mit Hilfe der Bestimmung

optischer Eigenschaften der Augenvorderkammerfluessigkeit, Bes-

timmung der Blutglucose beim Diabetiker, Dissertation Universitat

Essen, Essen, 1996.

[21] W.F. Schrader, Invest. Ophthalmol. Vis. Sci. 39/4 (1998) S926.

[22] B. Schrader, W. Schrader, Verfahren und Vorrichtung zur auto-

matischen Messung der Konzentration von Komponenten des

Augenvorderkammerwassers sowie Verwendung derselben,

Deutsches Patent Nr. 19830541, Jun 22 2004.

[23] B.D. Cameron, G.L. Cote, IEEE Trans. Biomed. Eng. 44 (1997) 1221.

[24] B.D. Cameron, H.W. Gorde, B.E. Satheesan, G.L. Cote, Diabetes

Tech. Ther. 1 (1999) 135.

[25] W. Haigis, Biometrie bei komplizierten Ausgangssituationen. In 9.

Kongr. d. Deutsch. Ges. f. Intraokularlinsen Implant, Springer Verlag,

Berlin, 1996. p. 17.

[26] R. Moses, Adler’s Physiology of the Eye, 8th ed., Mosby Company,

Saint Louis, 1987.

[27] S. Pohjola, Acta Ophthalmologica Suppl. 88 (1966) 1.

[28] W.F. Schrader, R. Grajewski, P. Meuer, Invest. Ophthalmol. Vis. Sci.

41 (2000) 77.

[29] H. Helmholtz, Handbuch der physiologischen Optik (Allgemeine

Enzyklopadie der Physik Bd. 9), Leipzig, 1867.

[30] K.H. Hazen, M. Arnold, G.W. Small, Appl. Spectrosc. 52 (1998)

1597.

[31] P. Meuer, Spektroskopische Untersuchungen an Kammerwasser: in

vitro und in vivo, Thesis, Julius-Maximillians-University, Wurzburg,

2003.

[32] Bruker Analytik GmbH, Multivariate Kalibration: Ein praktischer

Leitfaden zur Methodenentwicklung in der quantitativen Analytik,

Ettlingen, 1998.

Page 8: Non-invasive glucose determination in the human eye

W. Schrader et al. / Journal of Molecular Structure 735–736 (2005) 299–306306

[33] K. Baumann, H. Albert, M.v. Korff, J. Chemometrics 16 (2002)

339.

[34] K. Baumann, H. Albert, M.v. Korff, J. Chemometrics 16 (2002) 351.

[35] The BD Consumer Health Care Error Grid (13th Febr 2002), fttp

BDCHC.com, 2002.

[36] K. Buijs, G.R.J. Choppin, Chem. Phys. 39 (1963) 2035.

[37] G.R.J. Choppin, J.R.J. Downey, J. Chem. Phys. 56 (1972) 5899.

[38] W.I. Clarke, D. Cox, L.A. Gonder-Frederick, W. Carter, S.L. Pohl,

Diabetes Care 10 (1987) 622.

[39] J.L. Parkes, S.L. Slatin, S. Pardo, B.H. Ginsberg, Diabetes Care 2000; 23.

[40] Infrared and Raman Spectroscopy, in: B. Schrader (Ed.),, VCH,

Weinheim, 1995.

[41] M.N. Ediger, R.J. Landry, D.H. Sliney, Diabetes Technol. Ther. 1

(1999) 239.