Page 1: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

1

LASER ABLATION

Page 2: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

2

Laser Ablation ICP-MS

CETAC LSX 500CETAC LSX 500

Thermo Finnigan ElementThermo Finnigan Elementwww.cetac.com

www.thermo.com

Page 3: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

3

“Gold can be melted and recast and is therefore virtually untraceable”

Page 4: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

4Watling et al., Spectrochim. Acta Part B 1994, 49B, 205-219.

DISTINGUISH GOLD SAMPLES BASED ON TRACE ELEMENTS?

Page 5: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

5

DEVELOPMENTS IN LASER ABLATION

GÜNTHER et al., ANAL. CHEM. 2003, 75, 341A; TrAC 2005, 24, 255. UV LASERS (266 , 213, 193 nm)HOMOGENIZED BEAM PROFILEHELIUM TRANSPORT GAS

FRACTIONATION1. VARIATION OF SIGNAL RATIO vs TIME AS DIG SINGLE PIT11.MEAS. SIGNAL RATIOS DIFFER FROM THOSE IN SAMPLE

SOLUTIONS:FLAT BOTTOM CRATERSVERTICAL SIDESSHORT PULSE (fs) LASER(RUSSO et al., ANAL. CHEM. 2002, 74, 70A).

Page 6: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

6

PARTICLE SIZE EFFECTS IN LASER ABLATION

GÜNTHER & GUILLONGJAAS 2002, 17, 831

AESCHLIMAN et al.JAAS 2003, 18, 1008

Page 7: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

7

Page 8: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

8

Particles from Ablated Y2O3 Pellet

Track length velocity ~ 27 m/s

Page 9: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

9

Pressed pelletFernald soil blank

Page 10: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

10

Page 11: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

11

Page 12: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

12

SINGLE SPOT ABLATION

Page 13: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

13

Page 14: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

14

FEMTOSECOND LASER ABLATION

RUSSO et al.JAAS 2002, 17, 1072.ANAL. CHEM. 2003, 75, 6184 2004, 76, 379.

Page 15: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

15

Femtosecond Laser Ablation

• Nanosecond laser ablation is partly a thermal process

• Differences in the vaporization properties of elements leads to elemental fractionation

• With femtosecond pulses, the ablation process occurs by a mechanism far less dependent on thermal effects. Melting is not observed with pulse widths < 1 ps.

Page 16: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

16

Crater Profiles

Crater profiles for 100 fs and 4 ns lasers after 50 pulses

Page 17: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

17

Page 18: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

18

UV fs LASER ABLATIONINGO HORNUNIV. HANNOVER

Page 19: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

19

½WP1

½WP2

TP

SHG

THG

M3

ω2ω

M4

M2

M1

½WP3

ω3ω

3ω ω3ω+4ω

M5

M6

M9

M12FHG

M7

M8 4ω

ω3ω

ω

fs-laser ablation system785 nm 196 nm

Laser Optics

Sample cell

GasMC-ICP-MS

orICP-OES

RegenerativeAmplifier

Stretcher/Compressor

Seedlaser

Pumplaser

100fs=0.0000000000001sec.

By comparison, if traveling at the speed of light for 100 fs you would onlyjust cover about 30 μm in distance.

Page 20: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

20

Page 21: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

21

Page 22: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

22

CALIBRATE LASER ABLATION?

COMPENSATE FOR MATRIX DEPENDENCEOF ABLATION PROCESS

MATCHED STANDARDS

MEAS. ANALYTE REL. TO MINOR ISOTOPEOF ELEMENT AT KNOWN CONCENTRATION

CALIBRATE REL. TO SOLUTION AEROSOL?BECKER JAAS 2001, 16, 602AESCHLIMAN JAAS 2003, 18, 872-877.

Page 23: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

23

pump

CETAC LASER ABLATION SYSTEM TSI PIEZOBALANCE

FINNIGANICP-MS

ESI NEBULIZER

Nd:YAG laser (266 nm)

camera

translation stage

argon inlet

zoom lens

impactorelectrostatic precipitator

calibration solution

waste90%

10%

20%

80%

ICP

ESA

magnet

argon inlet

Page 24: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

24

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

0 100 200 300 400 500 600 700

Time (s)

Sign

al (c

ount

s s-1

) / 1

06

60Ni+

51V+52Cr+

NIST STEEL

Solution

Ablatedsolid

Page 25: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

25

Calibration of LA-ICP-MS with Dried Solution Aerosols

• Simultaneous introduction of particles from a LA cell and desolvated aerosol particles from a micro-flow nebulizer

Stotal = Ssolid + Ssolution

= RX,solid TLAt[X]solid + RX,soln VTneb[X]soln

RX isotope-specific response factor (signal/ng X)TLA transport from LA cell (ng solid/s)t time of ablation transient (s)[X]solid concentration of isotope in solid (ng X/ng solid)V volume of solution injected to ICP (L)Tneb nebulizer efficiency[X]soln isotopic concentration in solution standard (ng X/L)

Page 26: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

26

NIST 612 Glass (13 Elements, 5 Replicates) • Particle transport from the LA cell was measured using a piezoelectric microbalance• Each replicate was generated by firing 50 laser shots per localized spot on the sample• A two-point calibration plot for each replicate was prepared and an average calculated• All elements were measured in medium resolution (R = m/Δm = 4000)

CONCENTRATION (ppm) MEASURED CERTIFIED Relative Diff. (%)

Mn (55Mn+) 40.8 7.9 (39.6) 3.0Fe (56Fe+) 51.6 6.1 51 1.2Co (59Co+) 36.0 4.7 (35.5) 1.4Ni (60Ni+) 39.2 4.7 38.8 1.0Cu (63Cu+) 38.5 6.7 (37.7) 2.1Ba (138Ba+) 41.6 5.5 (41) 1.5Nd (146Nd+) 36.2 2.6 (36) 0.56Sm (147Sm+) 39.5 4.7 (39) 1.3Eu (151Eu+) 36.5 4.7 (36) 1.4Dy (161Dy+) 35.1 2.5 (35) 0.29Er (166Er+) 39.3 4.2 (39) 0.77Tl (205Tl+) 15.8 1.6 (15.7) 0.64Pb (208Pb+) 39.2 5.8 38.57 1.6

Page 27: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

27

NIST 1264a Steel (8 Elements) • 266 nm QUAD. Nd:YAG LASER, CETAC LSX-100• AVG. 30 SPOTS, TWO-POINT STD. ADDNS.• 50 SHOTS PER SPOT, MED. RES.• PARTICLE TRANSPORT MEAS. WITH MICROBALANCE

CONCENTRATION (wt %)MEAS. CERT. (INFO)

V (51V+) 0.119 0.029 0.106 Cr (52Cr+) 0.073 0.012 0.066 Co (59Co+) 0.156 0.017 0.150 Ni (60Ni+) 0.143 0.017 0.142 Cu (63Cu+) 0.248 0.040 0.250 W (184W+) 0.107 0.027 0.102 Pb (208Pb+) 0.056 0.055 0.024 Bi (209Bi+) 0.0016 0.0032 (0.0009)

Page 28: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

28

NIST 1264a Steel (8 Elements) • 193 nm ArF LASER• AVG. 3 SPOTS, TWO-POINT STD. ADDNS.• 50 SHOTS PER SPOT, MED. RES.• PARTICLE TRANSPORT MEAS. WITH MICROBALANCE

CONCENTRATION (wt %)MEAS. CERT. (INFO)

V (51V+) 0.115 0.011 0.106 Cr (52Cr+) 0.078 0.036 0.066 Co (59Co+) 0.137 0.035 0.150 Ni (60Ni+) 0.139 0.108 0.142 Cu (63Cu+) 0.277 0.201 0.250 W (184W+) 0.108 0.013 0.102 Pb (208Pb+) 0.021 0.004 0.024 Bi (209Bi+) 0.0006 0.0001 (0.0009)

Page 29: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

29

Page 30: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

30

Multivariate Analysisin LA-ICP-MS

Page 31: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

31

Laser Ablation of NIST 1224 Steel

0

500000

1000000

1500000

2000000

2500000

100 150 200 250 300 350 400

Time (s)

Sign

al (c

ount

s/s)

53Cr59Co60Ni65Cu98Mo184W

start laserablation

end laserablation

signal fromargon blank

65Cu+

59Co+60Ni+

98Mo+

184W+

53Cr+

Page 32: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

32

STEEL SAMPLES

Page 33: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

33

Principal Component Analysis

• Eigenvector decomposition of covariance matrix• Matlab toolbox developed by Eigenvector

Research, Inc.• “Traditionally” applied to IR spectra• Used to extract reduced dimension “factors” that

describe trends and similarities/dissimilarities in data from multi-component spectra

Page 34: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

34

1 2 3 4 5 6 7 8 90.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

Comparison of Stainless Steel Washers from the Same Bag(16 Elements)

95% C.I.

99% C.I.

Q R

esid

ual

Sample Number

Page 35: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

35

1 2 3 4 5 6 7 8 90

2

4

6

8

10

12

Comparison of Two Stainless Steel Washers from the Same Box(19 elements)

99% C.I.

Q R

esid

ual

Sample Number

Page 36: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

36

Chemometric Comparison of Seven Carbon Steels(25 Isotope Model)

-8

-7

-6

-5

-4

-3

-2

-1

0

1

2

3

4

5

6

7

8

9

-10 -9 -8 -7 -6 -5 -4 -3 -2 -1 0 1 2 3 4

Principal Component 3

Prin

cipa

l Com

pone

nt 2

NIST 1224NIST 1227NIST 1263aNIST 1264a_1 NIST 1264a_2 pipebarplate

Page 37: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

37

0 2 4 6 8 100

2

4

6

8

10

12

95% C.I.

99% C.I.

Comparison of SRM 139 and 139b to 139a(20 Elements)

Q R

esid

ual

Replicate

SRM139b SRM139

Page 38: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

38

1 2 3 4 5 6 7 8 90.1

1

10

Comparison of Two Strands from a Single Copper Wire(7 Elements)

95% C.I.

99% C.I.Q R

esid

ual

Sample Number

Page 39: No Slide Title · 2009. 12. 8. · (RUSSO et al., ANAL. CHEM. 2002, 74, 70A). 6 PARTICLE SIZE EFFECTS ... • Each replicate was generated by firing 50 laser shots per localized spot

39

1 2 3 4 5 61E-3

0.01

0.1

1

10Copper Wire Comparison

Q R

esid

ual

Model Wire

Wire 1 Wire 2 Wire 3 Wire 4 Wire 5 Wire 6 99% C.I.