metrology for lng conference 17-18 october 2013 delft, the ... · metrology for lng conference,...

36
Metrology for LNG Conference, 17-18 th Oct. 2013 Delft, the Netherlands Metrology for LNG Conference 17-18 th October 2013 Delft, the Netherlands

Upload: others

Post on 30-Oct-2019

4 views

Category:

Documents


1 download

TRANSCRIPT

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Metrology for LNG Conference

17-18th October 2013

Delft, the Netherlands

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

LNG Sampling System

Jorg Udo Gramlich

Advisor(Measurements)

OTD Maintenance Engineering

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

• Qatar / Ras Laffan / RasGas Company

• Challenge / Goal

• Design

• Golden Rules

• Issues at Site

• Way Forward

• Raman Spectroscopy

Agenda

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Qatar is a small peninsula in the

Middle East half way down the West

coast of the Arabian Gulf

Qatar has a tradition steeped in

pearl diving, but oil was discovered

in 1930

The North Field has estimated

reserves of over 900 trillion standard

cubic feet of recoverable gas

Qatar

Source: www.raslaffan.com

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Ras Laffan Industrial City

Source: www.condotte.com

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

• Located on the NE coast

• International and National Companies

• QatarGas, RasGas

• Shell, Dolphin and Oryx

• In 1997 - 39 shipments

• In 2007 - 479 shipments

• In 2012 - 3000 shipments

Ras Laffan Industrial City

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

RasGas Company Ltd.

RasGas Company Limited is the operator of Ras Laffan Liquefied Natural Gas Company Limited and other project companies. RasGas operates seven LNG trains, two sales gas units and 2 Helium Units.

Train 1 & 2: 8.1999&4.2000, combined 6.6 Mt/a

Train 3,4 & 5: 2.2004,8.2005&11.2006 combined 14.1 Mt/a

Train 6 & 7: 7.2009&2.2010, combined 15.6 Mt/a

AKG-1 & 2: 11.2006&12.2009,

approx. 2.700.000 Nm^3/h (inlet)

Helium 1 & 2: 8.2005&8.2013 approx. 4223 t/a liq. He

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

RasGas Company Ltd.

Train 7

Train 6

AKG 2

Train 5

Train 4

Train 3

AKG 1

Train 2

Train 1

He 1 & 2 (not shown on

the picture)

Source: RasGas Company Ltd.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Challenge

It is essential that the LNG is vaporized homogenously and that partial vaporization is avoided under all circumstances. Partial vaporization gives erratic reading on the GC measuring values, which effects on density and the gross calorific value calculations and lead to a quadratic impact!

E = VLNG

*ρLNG

* GCVLNG

Where

VLNG

: Volume of LNG

ρLNG

: Density of LNG

GCVLNG

: Gross Calorific Value of LNG

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Goal

• That the enthalpy rise in the LNG sample due to heat absorption throughout the Sampling Process is smaller than degree of sub-cooling to guarantee LNG come as a sub cooled liquid to the vaporizer

• To extract a sample of LNG that is truly representative

• To produce a reliable composition to calculate the density and the heating value

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Design as per ISO 8943

ISO 8943:2007 Refrigerated light hydrocarbon fluids — Sampling of liquefied natural gas — Continuous and intermittent methods

Source: ISO 8943:2007

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Sample Point Requirements

LNG must be in a sub-cooled condition at the Sampling Point. The degree of sub-cooling is ascertained by measuring the temperature and the pressure at Sampling Point and within the phase diagram the delta between the boiling point and the measured dataset will provide the information.

T LNG Sampling Point < T boiling: LNG as liquid

T LNG Sampling Point > T boiling: Vaporization of LNG (L+V)

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Sample Point

There are different ways to take the sample from the LNG line. The protruding sampler is preferred.

Sample from the periphery of the header might causes disadvantages (b).

Direct sampling from the header is preferred against sampling from a bypass line.

Source: LNG Custody Transfer Handbook, www.igu.org

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Most efficient insulation is vacuum insulation. Vacuum Jacked Hose Advantage: flexible and sealed vacuum. Disadvantage: connection points not insulated. Piping System Advantage: sealed vacuum, bayonet connection Disadvantage: not flexible

Sample Transport

Source: www.duraflex.com

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Heat Absorbtion Calculation

Heat absorption Calculation as per ISO 8943-07 : π (Ta – Ts) Q = ----------------------- X L 1 1 Do ----- +---- ln----- ha*Do 2k Di

Q : Heat absorption by sample line in Watts.

Ta : Atmospheric Temperature in K

Ts : LNG Sampling Temperature in K

d : LNG Density Kg/M³

ha : LNG pressure kPa

k : Therm. conductivity insulation material W/m²k

Do: Outside diameter insulation m

Di : Inside Diameter insulation m

L : Length of sampling line m

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Enthalphy Rise Calculation

The enthalpy rise ∆H1 in J/Kg in the LNG sample due to heat absorption through the sample line is given by the equation below : Q x 3600 ∆H1 = ------------- F

Q : Heat absorption by sample line

F : Sample flow rate

If the enthalpy rise ∆H1 < as the degree of sub cooling, no fractionation occurs.

If the enthalpy rise ∆H1 > as the degree of sub cooling, fractionation occurs.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Phase Diagram

Enthalpy of the saturated liquid

P: Pressure in kPa H: Enthalpy in J/Kg

1: Degree of Sub-Cooling 3: Saturated liquid

2: Pressure drop 4: Heat absorption sample line

Filled dot: LNG loading line

White dot: LNG at vaporizer

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Vaporizer

ISO 8943 states:

6.3.1 The heat exchange capacity of the LNG sample vaporizer shall be sufficient to gasify the whole volume of LNG which is being withdrawn for sampling.

6.3.2 The sample vaporizer shall be so constructed that the heavier components of the LNG shall not remain in the vaporizer.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Pre - Vaporization

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Partial Vaporization

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Vaporizer

The vaporizer is mainly a simple

construction, but the essential part for the

phase change from liquid to gasouse

phase.

Important is the efficency of the heat

transfer from the heater element to sample

to archive a flash vaporisation of the

sample.

If the heater element with the sample coil is

placed in a liquid bath, the efficiency of the

heat transfer will be optimized.

As per our experience a heated regulator is

not fit for purpose. Source: www.asap.nl

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

The heated regulator is fit to vaporize liquids like LPG or condensates.

•only heated on the area where liquid enters the regulator.

•the surface where the liquid is in contact with hot metal is small.

In case heat transfer between gas and heater element is not sufficient fractional vaporization might take place

Heated Regulator

Source: www.goreg.com

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Accumulator

The accumulator is used to restructure the LNG by averaging the sample in the accumulator.

ISO 8943 mention a accumulator, but there is no volume of the accumulator specified.

Normally the volume of the accumulator is between 2 to 6 liter. delay time

A accumulator becomes more efficient if it is not just a vessel in the sample line. A good vaporization is the key to reduce the accumulator size.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Pressure Regulator

The pressure regulator should keep the pressure in the fast loop stable to keep the flow rate constant.

Based on the sample pressure drop and composition, a Joule Thompson effect should be taken in consideration. heated regulator.

The gas sample transport line can be heat traced, depending on dew point of the gas. Source: www.goreg.com

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

LNG Sampling “Golden Rules”

1. The sample needs to be taken off from a point, where the LNG is in the sub cooled state. The location needs to selected carefully to avoid stratification of the sample

2. The sample transport lines need to be insulated and the flow need to be sufficient to avoid that the heat leaks causing partial vaporization of the LNG

3. The vaporizer power need to be sufficient to flash vaporize the LNG

4. ISO 8943:07 needs to be followed. ISO 8943 defines the requirements on the samp-ling system for refrigerated hydrocarbon fluids

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

• Location:(following API requirements) Probes should be mounted vertically at the top of a straight run of horizontal pipe. Probe should be at least 5 diameter downstream of a disturbance. Probe should not be mounted in any death end section of pipe.

• Situation in Unit 71 The probe is located close to a disturbance The flow through the bypass can not be proven as there is no flow indicator available. The bypass may be a death end.

Issues on Offsite Unit 71

Source: RasGas Company Ltd.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Issues on Offsite Unit 71

• For accurate analysis sample must be conditioned from initial low temperature to gas without partial vaporization or loss of product.

• Important to ensure that LNG remains in sub cooled condition during sampling and ambient heat input minimized.

• Insulation on sample take off tube to vaporizer cabinet is not sufficient to keep sample in sub cooled temperature and vaporization occurs before vaporizer.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

• An isolation valve located close to vaporizer box, cannot be cooled by small sample flow of LNG and is a huge hot spot.

• Vaporizer is heated pressure regulator and whose

design duty with LNG is unknown. • The accumulator causes a delay of approximately

30 min.

Issues on Offsite Unit 71

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Sample Take Off Point

This picture shows the 1/8” sample line coming out of the massive foam insulation. To keep the sample liquid, a temperature below -150°C needs

to be maintained. The rubber hose insulation can not provide the required insulation. The low sample flow is not supporting the LNG to remain in sub cooled stage.

Issues on Offsite Unit 71

Source: RasGas Company Ltd.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Liquid Sample Transfer

A mechanical pressure gauge is connected to liquid sample transfer line. The pipe to the gauge is insufficiently insulated.

The impulse line to the pressure gauge must be considered as a dead end. The pipe will be a transfer line for thermal energy and will lead to vaporization of LNG.

Issues on Offsite Unit 71

Source: RasGas Company Ltd.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Liquid Sample Transfer

A cryogenic needle valve is located on sample transfer line from take off point to heated regulator. The massive body of the valve will accumulate a huge amount of thermal energy, low flow of LNG sample will never be able to cool down the needle valve.

Issues on Offsite Unit 71

Source: RasGas Company Ltd.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Liquid Sample Transfer

The picture shows the heated regulator. The poor insulated line from left side is LNG sample line. Insulation is still required to avoid a gasification before the vaporizer.

Issues on Offsite Unit 71

Source: RasGas Company Ltd.

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

A Millions Dollar Question!

Source: www.mustangeng.com

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

To improve the sample quality several steps required:

•Sample probe improvement

•Shut off valve close to the take off point

•Vacuum jacked sample line

•Improvement of the vaporizer

Operational Technical Department did a study on the available sample probes

•Omnia 8943 Sampler (LNG jacked sampler)

•Introducing a new analyzer technique: Raman spectroscopy Scientific Instruments

Way Forward

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

Raman Spectroscopy

Raman spectroscopy is a

spectroscopic technique that use

scattering for detection of the

concentration and component.

Advantage: measures the LNG without

vaporization, probe directly hocked up

to the LNG line

Disadvantage: limited to measurement

up to C5 and detection limit at 200ppm

Not suitable for measurements

following ISO 6976, Natural Gas –

Calculation of calorific value, Density

and relative density.

Source: www.scientificinstruments.com

Metrology for LNG Conference, 17-18th Oct. 2013 Delft, the Netherlands

© RasGas Company Limited 2013

www.rasgas.com