mechanical device to support the flexion- extension of the ... · the movement in any way. recently...
TRANSCRIPT
POLITECNICO DI MILANO
Scuola di Ingegneria Industriale e dell'Informazione
Corso di Laurea Magistrale in
Ingegneria Meccanica
Mechanical device to support the flexion-
extension of the elbow in post-stroke spasticity
Relator: Prof. Hermes Giberti
Co-relator: Eng. Matteo Malosio
Thesis of:
Giulio Flavio Luigi
Francesco Schifino
Matr. 824160
Academic year 2015 – 2016
iii
Contents
Abstract ....................................................................................................................... 1
Chapter 1 Introduction ................................................................................................. 3
Overview of the issues related to upper limb rehabilitation ............................ 3
Specific problems concerning the elbow joint in hypotonic patients .............. 7
State of art ....................................................................................................... 8
Chapter 2 Device description .................................................................................... 15
Design specifications ..................................................................................... 15
General operation of the device ..................................................................... 16
Hooking, releasing and charging mechanisms .............................................. 19
Mechanism regulation ................................................................................... 28
Transmission and regulation of the torsional spring ..................................... 32
Device components list ................................................................................. 32
Chapter 3 Device sizing ............................................................................................ 35
Genera sizing ................................................................................................. 35
Sizing of the setting mechanism .................................................................... 39
Transmissions design .................................................................................... 43
FEM analysis of the most stressed components ............................................ 50
Dynamic simulation with rigid bodies model ............................................... 62
iv
Chapter 4 Prototype realization ................................................................................ 69
Specification of the prototype ....................................................................... 69
Parts produced with CNC ............................................................................. 70
Parts produced with 3D printer ..................................................................... 71
Chapter 5 Conclusion and future developments ....................................................... 73
Bibliography ............................................................................................................. 75
v
Figure Index
Figure 1.1 LIGHTarm exoskeleton CAD .............................................................. 5
Figure 1.2 Human arm kinematic scheme and range of motion of the joints
without and with the LIGHTarm exoskeleton .............................................. 6
Figure 1.3 LIGHTarm exoskeleton kinematic scheme and joints range of motion
....................................................................................................................... 6
Figure 1.4 Human arm kinematic scheme and joints range of motion ................. 8
Figure 1.5 Automatic watch with a transparent case back, showing its eccentric
weight ............................................................................................................ 9
Figure 1.6 Magneti Marelli Kinetic energy recovery system ............................. 10
Figure 1.7 NASA Flywheel energy storage ........................................................ 11
Figure 1.8 JAS EZ elbow Orthosis ..................................................................... 12
Figure 1.9 Retractable pen mechanism ............................................................... 13
Figure 2.1 Device view with highlighted the spring-arm in yellow and the exo-
arm in orange ............................................................................................... 17
Figure 2.2 Scheme of the extension movement of the elbow with highlighted in
green the discharging zone .......................................................................... 18
Figure 2.3 Scheme of the flexion movement of the elbow with highlighted in
blue the charging zone ................................................................................ 18
Figure 2.4 View of the spring-arm with highlighted in blue the support, in red
the charging-block and in green the sledge ................................................. 20
Figure 2.5 Explosion view of the sledge ............................................................. 20
Figure 2.6 Two views of the Exo-arm with highlighted in red the superior ramp
and in blue the inferior ramp ....................................................................... 21
Figure 2.7 Schematic representation of the beginning of the hooking phase 1: the
ramp of the exo-arm (orange) enters in contact with the superior shaft of
the sledge (yellow) ...................................................................................... 22
Figure 2.8 Schematic representation of the beginning of the hooking phase 2: the
inferior shaft (yellow) starts to climb the step of the support disc (red) by
itself ............................................................................................................. 22
Figure 2.9 Schematic representation of the end of the hooking phase 2: the ramp
of the sledge (yellow) enters in contact with the inferior ramp of the exo-
arm (orange) ................................................................................................ 23
Figure 2.10 Schematic representation of the start of the releasing phase: the
inferior shaft of the sledge (yellow) comes to the ramp of the support disc
(red) ............................................................................................................. 23
Figure 2.11 Schematic representation of the releasing phase: The sledge
(yellow) starts to go down following the ramp of the support disc (red) .... 24
Figure 2.12 Schematic representation of the releasing phase: The sledge
(yellow) completed the descendent and the exo-arm (orange) is free to
come back .................................................................................................... 24
vi
Figure 2.13 View of the recharging-block with highlighted in blue the contact
plate............................................................................................................. 25
Figure 2.14 Schematic representation of the charging phase: Finished the phase
of discharge the exo-arm (orange) is free to move and comes back during
the flexion movement ................................................................................. 26
Figure 2.15 Schematic representation of the charging phase: The exo-arm
(orange) enters in contact with contact plate (blue).................................... 26
Figure 2.16 Schematic representation of the charging phase: The exo-arm
(orange) pushes the contact plate (blue) so the sledge (yellow) goes up the
ramp of the support disc (red) and goes forward the hooking point ........... 27
Figure 2.17 Schematic representation of the charging phase: The sledge (yellow)
comes in proximity of the step of the support disc (red) ............................ 27
Figure 2.18 Schematic representation of the charging phase: The sledge falls
down in the hooking point of the support disc (red) pushed by the contrast
spring .......................................................................................................... 27
Figure 2.19 View of the releasing compartment: Extreme position in extension
.................................................................................................................... 29
Figure 2.20 View of the releasing compartment: Extreme position in flexion .. 29
Figure 2.21 View of the hooking compartment: Position of minimum extension
.................................................................................................................... 30
Figure 2.22 View of the hooking sector: Position of maximum extension ........ 30
Figure 2.23 View of the charging compartment: In blue the charging-block .... 31
Figure 3.1 Example of circular ramp, prospected view on left and lateral view on
right ............................................................................................................. 36
Figure 3.2 View of spring-arm with highlighted in red the sides of the
articulated parallelogram ............................................................................ 38
Figure 3.3 Scheme for the calculation of the horizontal movement of the sledge
.................................................................................................................... 39
Figure 3.4 View of the setting mechanism for the range of discharge with
highlighted in red the articulated quadrilateral ........................................... 40
Figure 3.5 Scheme 1 for the design of the setting mechanism for the discharging
range ........................................................................................................... 41
Figure 3.6 Scheme 2 for the design of the setting mechanism for the discharging
range ........................................................................................................... 42
Figure 3.7 View of the gear transmission of the device, in grey the bull gear and
in orange the pinion .................................................................................... 44
Figure 3.8 Schematic representation of the torque spring transmission, in grey
the torque spring, in yellow the Spring-arm shaft and in red the wire who
connects the previous elements .................................................................. 46
Figure 3.9 Graphic of the length l in function of the angle α ............................. 47
Figure 3.10 Graphic of the angle β in function of the angle α ............................ 48
Figure 3.11 Graphic of the ratio r in function of the angle α .............................. 49
vii
Figure 3.12 FEM analysis of the inferior shaft first design, Von Mises Stress
arrangement ................................................................................................. 52
Figure 3.13 FEM analysis of the inferior shaft first design, Displacement
arrangement ................................................................................................. 52
Figure 3.14 FEM analysis of the inferior shaft first design, Safety Factor
arrangement ................................................................................................. 53
Figure 3.15 FEM analysis of the inferior shaft final design, Von Mises Stress
arrangement ................................................................................................. 54
Figure 3.16 FEM analysis of the inferior shaft final design, displacement
arrangement ................................................................................................. 54
Figure 3.17 FEM analysis of the inferior shaft final design, Safety Factor
arrangement ................................................................................................. 55
Figure 3.18 FEM analysis of the exo-arm superior ramp, Von Mises Stress
arrangement ................................................................................................. 56
Figure 3.19 FEM analysis of the Exo-arm superior ramp, displacement
arrangement ................................................................................................. 57
Figure 3.20 FEM analysis of the Exo-arm superior ramp, Safety Factor
arrangement ................................................................................................. 57
Figure 3.21 FEM analysis of the Exo-arm inferior ramp, Von Mises Stress
arrangement ................................................................................................. 58
Figure 3.22 FEM analysis of the Exo-arm inferior ramp, displacement
arrangement ................................................................................................. 59
Figure 3.23 FEM analysis of the Exo-arm inferior ramp, Safety Factor
arrangement ................................................................................................. 59
Figure 3.24 FEM analysis of the sledge ramp, Von Mises Stress arrangement .. 60
Figure 3.25 FEM analysis of the sledge ramp, displacement arrangement ........ 61
Figure 3.26 FEM analysis of the sledge ramp, Safety Factor arrangement ........ 61
Figure 3.27 View of the simplified model of the device used for the dynamic
simulation .................................................................................................... 62
Figure 3.28 View of the simplified model in order to see the contact between the
sledge inferior shaft and the Hooking-block in the simulation start point .. 63
Figure 3.29 Graphic of the theoretical working cycle of the device in function of
the flexion angle .......................................................................................... 65
Figure 3.30 Dynamic Simulation results: Torque transmitted to the user during
the discharging phase .................................................................................. 66
Figure 3.31 Dynamic Simulation results: Sledge elevation during the
discharging phase ........................................................................................ 67
Figure 3.32 Dynamic Simulation results: Torque transmitted in function of the
sledge elevation ........................................................................................... 68
Figure 4.1 Front view of the prototype ............................................................... 70
Figure 4.2 Rear view of the prototype ................................................................ 70
Figure 5.1 View of the device mounted on the LIGHTarm exoskeleton ............ 73
viii
Abstract
The purpose of this dissertation is to present an innovative device for upper limb
rehabilitation and, in particular, for the flexion-extension of the elbow for patients
struggling with the complete extension of the joint. The device has been designed
for LIGHT arm, the exoskeleton for rehabilitation developed by the Institute of
Industrial Technology and Automation part of the Italian National Research
Council. Its primary function is to provide a driving torque to support the
extension of the elbow in a range near the complete extension of the joint. The
energy needed for this aid is provided by the patient himself during the flexion
movement, limited to the area of complete flexion, where the patient usually
presents enough residual muscular tone. A torsional spring is used as energy
accumulator and, furthermore, the device does not require any forms of power
beyond the energy transmitted by the patient.
Index terms – Exoskeleton, rehabilitation, spring mechanism, snap mechanism,
elbow flexion-extension, torsional spring.
2
Chapter 1
Introduction
Overview of the issues related to upper limb rehabilitation
The upper limb provides a significant contribution to the implementation of daily
activates (Activities of Daily Living, ADL); state this, any type of impairment that
reduce the mobility of the limb will negatively affect the quality of life. Issues to
the upper limb maybe caused by deficit or damage to the musculoskeletal system
or the central nervous system. The most common clinic condition concerning
resulting after these events id the spasticity. There are several definition of
spasticity, but the one that seems to be the more widely accepted is the one of
Lance (1980) [1]:
“Spasticity is a motor disorder characterized by a velocity
dependent increase in tonic stretch reflexes (muscle tone) with
exaggerated tendon jerks, resulting from hyper-excitability of
the stretch reflexes, as one component of the upper
motoneuron syndrome”
First, following the definition, we can infer that the spasticity is identified with
“muscle hypertonicity” – if we try to move a paretic limb, we will encounter more
resistance the more is the speed at which we make the movement; second, the
definition implies that this is a condition affecting the muscles that the patient is
unable to control, since, with the damage, the brain lost the possibility to adjust
the movement in any way.
Recently Lance’s definition has been considered too strict and it has been
expanded:
“Sensory-disordered motor control, resulting from an upper
motorneuron injury, which comes as involuntary activation of
the muscles in an intermittent or continuous way”
Now spasticity means both a generic neuromotor issue – a disorder that arises
from neurological problems – and a condition that damage the performance of the
musculoskeletal system. Stated this, the spasticity causes may be:
- Infantile cerebral paralysis – persistent but non-progressive disorder of
posture and movement, due to alterations of cerebral function before the
central nervous system has completed its development;
- Stroke – pathological cerebral vascular event, resulting in acute
disturbance of brain function, focal or generalized;
Chapter 1
4
- Multiple sclerosis – chronic autoimmune demyelinating disease that
affects nerve cells making difficult the communication between the brain
and spinal cord;
- Head injuries – It occurs when an external force traumatically injures the
brain. They are the leading cause of disability worldwide;
- Spinal injuries – Injuries that may result from the interruption of the
ascending and descending nerve pathways of the spinal cord. They may
be a consequence of a trauma, of a tumor process or of ischemia and
determine a paralysis of the lower limbs and, in severe cases, even higher
ones;
- Anoxia – Pathological condition resulting in the cells suffering due to lack
of oxygen.
Scope of the rehabilitation is to involve the patient in evolving his competencies
by:
- Exploiting the remaining functional systems;
- Developing new ability to better achieve the control of himself and the
surroundings;
- Reducing the unwell perception caused by the limitation imposed by the
biological damage.
Summarizing, the function of the neurological rehabilitation is to help the recover,
at least partial, of the lost functionalities in order to allow the patient to realize the
movement typical of ADL.
The phenomenon known as Neuroplasticity is what allows this functional recover.
Neuroplasticity refers to the fact that the Control Neural System (CNS) is able to
experience structural and functional changes in response to physiological events,
environmental stimuli (ex. learning), and pathological events. Therefore, in
patients with lesions affecting the CNS, neuroplasticity represents a mechanisms
of spontaneous recovery and, if properly interpreted, can serve as rational basis to
promote targeted rehabilitative approaches.
Furthermore, recent studies have shown that training after injury can revive neural
plasticity and functional recovery in the limb [2]. The human brain is not only
sensitive to the experience, but is able to retain this plasticity after injuries
Without going into further details, we can say that the brain has the ability to
compensate brain injuries with specific mechanisms and neuroplasticity is the
biological substrate that allows the recovery considering the physiological
conditions (for more detail [3]).
Introduction
5
Concerning the rehabilitation
The exoskeleton LIGHTarm developed by the Institute of Industrial Technology
and Automation (ITIA) – which is part of the National Research Council (CNR)
– is, in its version not implemented, an exoskeleton for the rehabilitation of upper
limb. This device provides an antigravity support to the patient's arm, allowing
patients with a muscle hypotonus to perform movements in almost all the volume
of work. Its particular shape allows a good handling of the limb and, in particular,
implements various ingenious solutions thanks to the connection between the
robotic arm to the human one at the shoulder and elbow level. More details on
this can be found in the articles [4], [5], [6], [7], [8], [9], [10].
Figure 1.1 LIGHTarm exoskeleton CAD
Chapter 1
6
Figure 1.2 Human arm kinematic scheme and range of motion of the joints
without and with the LIGHTarm exoskeleton
Figure 1.3 LIGHTarm exoskeleton kinematic scheme and joints range of motion
Introduction
7
Specific problems concerning the elbow joint in hypotonic
patients
A characteristic of spasticity, which should be emphasized in this specific context,
is that the muscle tone depends on the speed of the movement, that is, it increases
with the increase of the speed of movement. It causes the so-called “Switchblade
effect” in the passive limb movements involving both the agonist and the
antagonists muscles, thus interfering with both extension and bending of the body
part [11].
In order to better understand the function of the proposed device, it is worth
emphasize that spasticity is always accompanied by a disturbance in muscle
strength and, in particular, it affects more the antigravity muscles; the practical
effect is that, in the upper limbs, we can find a predominance of the flexion spasms
(and so the flexor muscles), while in the lower limbs the spasm of extensors
prevails. This pathophysiology is the starting point to assess adequately a device
that stores energy during the muscle flexion of the limb, and then gives it in the
extension, when the muscle of the patient experiences a strength deficit.
Stated this, after some analysis done on post stroke patients, it has been shown
that in patients with muscular hypotonus there are more pronounced difficulties
in elbow extension, especially in the area close to full extension, although it is
still maintained a higher muscle tone in the flexion. For these reasons, the patient
was able to exert maximum force in the area close to the full flexion of the arm.
These analysis has made clear the need to implement a kinematic joint in the
design of the LIGHTarm exoskeleton placed at the elbow, in order to support the
elbow extension phase in the patients.
Looking at the kinematic scheme of the exoskeleton (figure1.3), we see that the
elbow joint is associated with two rotational degrees of freedom (𝐮e,1 and 𝐮e,2)
oriented perpendicularly. These form a universal joint centered in the intersection
of the two axes of rotation (Oee) in which the patient’s elbow should be located
(as shown by the figure (1.3)).
Then, a stage of preliminary study showed that the vast majority of rehabilitation
exercises for this kind of diseases are carried out keeping the flexion-extension of
the elbow at a horizontal level; then, observing the figure (1.2), the axis of rotation
(𝐮h,4) will be angled of a few degrees with respect to the 𝑥 axis of the main
reference gaming system (ground). Following this, it was decided to seek a
solution to implement by involving only the last joint.
Chapter 1
8
Figure 1.4 Human arm kinematic scheme and joints range of motion
It arises then the idea to design a mechanical device that implements the rotation
(𝐉𝒆,𝟐), in order to provide an aid to the patient in the extension of the elbow when
the axis of rotation is next to the vertical one.
State of art
The fundamental concept of the device is to accumulate energy in a certain range
of motion and then release it in a different range. There are several devices that
use the same concept for similar purposes, from the charging mechanism of
watches up to the most modern KERS (Kinetic Energy Recovery System) used in
Formula 1. The kinetic energy can be stored and maintained in different ways, as
potential energy in a spring, converted into electrical energy and then stored in a
battery as chemical, or maintained as kinetic energy using rotating disks with very
low friction bearings in a vacuum environment.
Introduction
9
We can quickly examine a few examples:
Figure 1.5 Automatic watch with a transparent case back, showing its eccentric
weight
Self-winding watch – An automatic watch is a mechanical wrist watch which has
the characteristic of recharging itself by exploiting the movement of a rotor placed
in it. The rotor movement is caused by the movement of the surface on which it
is placed, like the arm of its owner, making unnecessary any operation of manual
winding. The rotor rotates around a pivot and, moving with the oscillation of the
arm, acts on the spring that charges the mechanism. In this case the energy
accumulator is a spiral spring [12].
Chapter 1
10
Figure 1.6 Magneti Marelli Kinetic energy recovery system
KERS by Magneti Marelli – Magneti Marelli has developed an innovative
technology for Formula 1 in the field of energy recovery. This system, called
KERS (Kinetic Energy Recovery System), is in fact a system that converts
mechanical energy into electrical energy to be stored in special batteries; the pilot,
then, may decide to re-use this energy in special conditions such as straights,
overtaking, or strategic points of the circuit.
The device is connected directly to the crankshaft through a motor-generator,
which, during braking, converts the kinetic energy into electric current. This
current, through the control unit and using shielded cabling, goes to recharge the
lithium-ion batteries. In acceleration, instead, when the pilot uses the boost, the
electrical energy is taken by the batteries and, always by means of the electronic
control unit, sent to the motor-generator that rotates in the reverse direction and
transmits the accelerating force on the motor shaft [13].
Introduction
11
Figure 1.7 NASA Flywheel energy storage
FES – The Flywheel Energy Storage (FES), is an electromechanical device
capable of energy storage in the form of rotational kinetic energy.
The basic idea is to accumulate energy by rapidly rotating a flywheel. This idea
is very interesting, since they can accumulate large amounts of energy in a "small"
object (with a good specific energy capacity) compared to other types of
accumulator, such as electrochemical cells [14].
None of these devices is used in rehabilitation; in fact, when speaking of
exoskeletons for the rehabilitation, electric motors are the most common solution
used to assist the movement of the elbow. The LIGHTarm exoskeletons, instead,
relieves the patient's arm from its weight, but still leaves to the patient the load of
the inertial contributions generated during the movement. However, there are
certain types of orthoses that, through a spring, exert over the limb contrasting
forces in order to make it move towards a given position following the therapy.
To cite one example:
Chapter 1
12
Figure 1.8 JAS EZ elbow Orthosis
JAS EZ Elbow – Stress relaxation and low-load stretch are established stretching
techniques that safely and efficiently restore normal length to shortened tissues
surrounding a joint, allowing for motion and use when stiffness develops after
injury. Stress relaxation and low-load stretch can be carried out by hand (via a
therapist) or by using JAS and JAS EZ product systems. During a stress relaxation
and low-load stretch therapy session, the joint is brought to a pain-free stretched
position and held there for several minutes, to allow for the surrounding tissues to
relax and lengthen. Once this occurs the joint is stretched and held again, and
more lengthening is achieved. This process is repeated several times to complete
a JAS treatment session [15].
Introduction
13
So, there are no exoskeletons hosting kinetic energy storage devices. All the
accumulation devices for kinetic energy considered have a differentiation
between accumulation stage and release stage that is exclusively temporal. What
we are introducing with the designed mechanism is the possibility to choose one
ROM (Range of Motion) for accumulating energy, one in which keep it
accumulated, and finally one in which release it. The aim is to keep these three
ranges within an angular sector of less than 360°. A common mechanism which
is probably the closest to the device developed is the retractable pen mechanism.
This mechanism, using a linear spring and specific geometrical constraints, allows
four a motion phases: the first in which the cartridge tip comes out, the second
which blocks the tip the reached position, the third which retract the tip of the
cartridge and then the last that locks it in the initial position.
Figure 1.9 Retractable pen mechanism
Chapter 1
14
Chapter 2
Device description
Design specifications
To be useful, the design of the mechanism should have the following
characteristics:
- Be small in size, so that it can be mounted directly on the exoskeleton. As already mentioned, the intention is to implement the last joint through
a small mechanism that can be easily installed or uninstalled on the arm.
Obviously it should not be too bulky, to not cause discomfort to the patient
during the movements, and it should not limit too much the range of
motion of the exoskeleton. So, the maximum overall dimensions that the
mechanism may have is defined by a cylinder of diameter 100 mm and
height of 20 mm centered on the selected joint.
- Do not use electricity. Not being electricity dependent allows the LIGHTarm to be installed
anywhere; to keep this characteristic the component will have to rely only
on the strength that the patient can give. The mechanism will act as simple
kinetic energy storage hosting the energy donated by the patient in the
favorable area, to be able to unleash it as an aid in the unfavorable one.
- Support the patient in the last phase of the elbow extension.
As we have already said, the ultimate purpose of the mechanism is to
provide energy with a supporting torque, which is added to the strength of
the patient; this in order to allow him to reach the full extension of the
elbow.
- Allow the patient to reload the mechanism in the area close to full
flexion.
The area in which the patient can afford to transfer energy to recharge the
mechanism is exactly the opposite end compared to where it needs to
receive it. The mechanism has to be able to keep the spring in position
while the patient's arm passes from one zone to another.
- Being as much adjustable as possible both in the width of the range of
action of the device and in the applied torque.
Chapter 2
16
The system needs to be adjustable based on the patient's needs. In fact, the
area in which each patient requires the aid of the device may vary widely
so, although the range of motion allowed by the exoskeleton for this type
of movement is 120 degrees, some patients may not be able to exploit it
all. The same issue can be found when speaking about the intensity of the
applied torque. It must therefore be able to being adjusted according to
individual needs.
- Keep the torque applied as much constant as possible. Once the necessary couple is chosen it should vary as little as possible. By
deciding to use a spring as an energy accumulator there may be the
problem that it could generate an inversely proportional torque against the
patient needs. In fact, in the zone of full extension the patient will require
the maximum torque but the spring will be near the point of minimum
torque that it can generate.
- Try to be as much gently as possible in order not to stress the arm
with excessive acceleration. A too rough trigger of the torque must be avoided since it may generate
yanks in the patient's arm.
- Once the extension is complete the patient should be able to freely
move the arm.
Except for the charging and recharging movement, the patient's arm
should not be influenced by the spring. So in the extension movement the
spring will intervene only close to the complete stretch, while in the
flexion only in the full close of the stretch. The rest of time the spring must
not be mechanically connected to the patient's arm.
General operation of the device
We now present the characteristics of the developed device and its general
operating operation, we will enter into the details of the technical solutions
implemented in the following paragraphs [21].
In its high-level operation scheme, the mechanism is based on two components
rotating around a common axis of rotation. One will have to move together with
the patient's arm and be rigidly connected to the last link of the exoskeleton, which
is in turn bound to the forearm of the patient. The second will be together with the
torsion spring with a shape that allows both the charging and the discharging.
These two components through different types of interaction between them in
different points of the work space will allow the coupling and uncoupling between
Device description
17
the exoskeleton’s arm – and consequently the patient arm – and the torsion spring.
Henceforth, we will call the component that moves together with the exoskeleton
exo-arm, while the other one connected with the spring, spring-arm.
In the following picture, the exo-arm in orange and the spring-arm in yellow.
Figure 2.1 Device view with highlighted the spring-arm in yellow and the exo-arm
in orange
We now analyze the work cycle of the device. Although the device is adjustable
both in the range of discharging and in the range of work, we set for clarity some
parameters that we are going to maintain constant from now on:
- Complete flexion 𝛼 = 0°. - Complete extension 𝛼 = 120°. - Charging zone 𝛼 = 0° ÷ 30°. - Discharging zone 𝛼 = 90° ÷ 120°. - Total working range 120°.
- Charging range 30°.
- Discharging range 30°.
Where 𝛼 is the independent coordinate representing the opening angle of the
patient's elbow.
A working cycle consist in four phases, two in which the patient’s arm interacts
with the spring (charge and discharge) and two in which the arm can move freely:
one after the complete flexion and one after the complete extension. Referring to
the figures (2.2) and (2.3), the cycle, starting from a complete flexion position,
takes place as follows:
Chapter 2
18
1. 𝜶 between 0° and 90°, Free Zone 1. The arm is free to move both in
extension and in flexion.
2. 𝜶 = 𝟗𝟎°, Hooking point. The arm comes into contact with the spring
which begins to pull in order to help the patient in the extension.
3. 𝜶 between 90° and 120°, Discharging Zone. Arm and spring are coupled
and the spring tends to extend the arm (green area).
4. 𝜶 = 𝟏𝟐𝟎°, Releasing Point. The spring leaves the arm free to move
again.
5. 𝜶 between 120° and 30°, Free Zone 2. Again the patient is free to move
without the contribution of the spring, neither in flexion nor in extension.
6. 𝜶 between 30° and 0° Charging Zone. The arm comes into contact with
the spring and by flexing the elbow the exo-arm charges the spring (blue
region). At complete flexion, the spring is fully recharged and the cycle
can start again.
Figure 2.2 Scheme of the extension movement of the elbow with highlighted in
green the discharging zone
Figure 2.3 Scheme of the flexion movement of the elbow with highlighted in
blue the charging zone
Device description
19
The hooking and releasing points are guaranteed by the particular geometry of the
support disk which, interacting with the spring-arm, allows this to hang in these
two specific points. For this reason, there are some fundamental conditions for the
cycle to function properly:
1. The arm must reach the release point to be able to be freed from the spring,
otherwise, the arms needs to return to the hooking point that also allows
the release. In this last case, however, the patient will have to work against
the spring.
2. The spring must be fully recharged in order to correctly complete the
cycle. If the complete charge is not reached, the spring will tend to
discharge again by returning to the releasing point.
Hooking, releasing and charging mechanisms
We now go more in detail regarding the interactions that take place among the
exo-arm, the spring-arm and the support disk in the hooking and releasing points.
Although so far we have spoken of the spring-arm as a single entity for simplicity
of exposition, it is composed of several pieces and it is in fact the more complex
subset of the device.
As the figure (2.4) shows, the spring-arm can be conceptually divided into three
parts:
- A support part connected to the torsion spring, to which are connected the
other parts (the blue parts).
- An adjustable part (by means of screws) used to charge the spring which
is nothing more than a small disc. This is bound to a flat surface, on which
the exo-arm acts pushing (red part).
- A part composed of an articulated parallelogram that allows the sledge to
move vertically (green part).
Chapter 2
20
Figure 2.4 View of the spring-arm with highlighted in blue the support, in red
the charging-block and in green the sledge
Focusing on the sledge, it has two side elements which support three components:
a superior shaft and a bar forming a circular ramp both interacting with the exo-
arm; a lower shaft that interacts with the support disc.
Figure 2.5 Explosion view of the sledge
Device description
21
Regarding the exo-arm, it has two circular ramps; one above and one below that
interact respectively with the superior shaft and with the ramp-bar.
Figure 2.6 Two views of the Exo-arm with highlighted in red the superior ramp
and in blue the inferior ramp
The support disc is designed to block the lower shaft with a vertical wall. A step,
with height equal to the lower shaft diameter, which blocks the spring-arm and
prevents the torsional spring to discharge. In order to overcome this step the
sledge will have to be raised by the exo-arm (note: the geometry of the
parallelogram coordinated with that of the support disk has been studied to
prevent the slide from rising alone). Once surmounted the step, the support disc
keeps the sledge at the same height until it reaches the release area in which,
through a ramp, it degrades to the initial level.
The hooking procedure can be divided into two phases:
- Phase 1. The superior ramp of the exo-arm comes into contact with the
superior shaft of the sledge, then, with the rotation of the exo-arm
following the elbow’s direction of extension, it tends to lift the sledge
vertically.
- Phase 2. The sledge reaches an altitude that allows it to rise autonomously,
it rises until the bar with the circular ramp comes into contact with the
lower ramp of the exo-arm. This way, the spring-arm is hooked to the exo-
arm and begins to pull it helping the patient in the elbow’s extension
Chapter 2
22
Figure 2.7 Schematic representation of the beginning of the hooking phase 1:
the ramp of the exo-arm (orange) enters in contact with the superior shaft of
the sledge (yellow)
Figure 2.8 Schematic representation of the beginning of the hooking phase 2:
the inferior shaft (yellow) starts to climb the step of the support disc (red) by
itself
Device description
23
Figure 2.9 Schematic representation of the end of the hooking phase 2: the
ramp of the sledge (yellow) enters in contact with the inferior ramp of the exo-
arm (orange)
The release procedure is much simpler than the previous one and is substantially
entrusted to the support disk. As said the support disc has at the point of release a
circular ramp. So, when the sledge will reach it, it would tend to decline in altitude
disengaging the exo-arm, now free to go back. The support disc then has a second
vertical wall to prevent the sledge to go beyond the release point.
Figure 2.10 Schematic representation of the start of the releasing phase: the
inferior shaft of the sledge (yellow) comes to the ramp of the support disc (red)
Chapter 2
24
Figure 2.11 Schematic representation of the releasing phase: The sledge
(yellow) starts to go down following the ramp of the support disc (red)
Figure 2.12 Schematic representation of the releasing phase: The sledge
(yellow) completed the descendent and the exo-arm (orange) is free to come
back
Device description
25
We complete the paragraph explaining the last part of the charge process for
completeness. As mentioned this stage presents many less complexity compared
to the discharging phase. It starts when the part of exo-arm opposite to the ramps
comes in contact with the contact plate of the charging block (blue in the figure
(2.13)).
Figure 2.13 View of the recharging-block with highlighted in blue the contact
plate
We see on the figures (2.14), (2.15), (2.16), (2.17), (2.18) the phases of the
charging process:
1. Once ended the releasing phase, the exo-arm is free to move and begin the
returning phase.
2. At the established point the exo-arm will come into contact with the plate,
we have already said that the plate is rigidly secured to the spring-arm with
screws.
3. Continuing its run, the exo-arm will push the plate back recharging the
spring.
4. When the patient's arm reaches full flexion the device will be in the
situation illustrated in the figure (2.18).
5. A small spring placed between the support of the spring-arm and the
sledge press this constantly downwards, which means that it will drop to
the lower level once left the step of the support disk.
6. With the sledge locked by the step of the support disk, the device is at the
starting point.
Chapter 2
26
Figure 2.14 Schematic representation of the charging phase: Finished the phase
of discharge the exo-arm (orange) is free to move and comes back during the
flexion movement
Figure 2.15 Schematic representation of the charging phase: The exo-arm
(orange) enters in contact with contact plate (blue)
Device description
27
Figure 2.16 Schematic representation of the charging phase: The exo-arm
(orange) pushes the contact plate (blue) so the sledge (yellow) goes up the
ramp of the support disc (red) and goes forward the hooking point
Figure 2.17 Schematic representation of the charging phase: The sledge (yellow)
comes in proximity of the step of the support disc (red)
Figure 2.18 Schematic representation of the charging phase: The sledge falls
down in the hooking point of the support disc (red) pushed by the contrast
spring
Chapter 2
28
Mechanism regulation
So far we have considered all the ranges of work, charge and discharge constants
for the sake of simplicity, of course, if this were true, it would violate one of the
required technical specifications, which provides for the possibility to adjust the
device in order to personalize the therapy. So, before continuing, let's make a brief
digression over the settings that can be performed on the device.
First, it should be pointed out that the angular distance to be traveled in the
recharging phase is slightly greater than the distance traveled in the discharging
phase; in fact, in order to recharge, the spring must travel the same stretch and in
addition there will be the part concerning the releasing mechanism.
Stated this, we get a first constraint on the range of adjustment, in fact, once
decided the range of discharge, the recharging one is directly derived.
Now let's see the maximum and minimum values for each range and where they
come from:
- The maximum value of the working range is 120 ° and is directly derived
from the maximum excursion that the exoskeleton arm can made together
with the patient.
- The sum of the maximum values of charge and discharge mustn’t be
higher than 120°, so, assuming for simplicity that the two ranges have the
same values, they will both have a maximum value of 60°.
- The minimum values are given substantially by the impediment caused by
the adjustment mechanism itself. Since a minimum of 20° is the value that
was established for the range (refer to section 3.2 for sizing details), we
obtain a minimum working range of 40° (20° + 20°).
- For the points of start and end of the charging and discharging range, the
mandatory constraints are the end of the working range towards the
extension for the release point and the point of maximum deflection
reached by the patient for the charging point.
We enter at this point into the merits of the design choices. So far we have treated
the settings that the device allows referring to the patient's workspace (0° to 120°).
Actually the device has a gear transmission that allows to take advantage of a
broader working space. Thanks to this, the working range is almost doubled
during the setting process, allowing a better accuracy.
From a functional point of view, we can distinguish three compartments used for
setting:
Device description
29
Releasing compartment – Used to adjust the maximum extension. It consists of
two coaxial discs arranged one above the other. Referring to the figures (2.19) and
(2.20), the first (dark blue) is bound to the device and is therefore fixed. The
second (blue light) mounts above it the releasing block (red) allowing to decide
the position of the releasing point and consequently the maximum extension.
Once set, they are locked by means of a screw on the outer ring.
Figure 2.19 View of the releasing compartment: Extreme position in extension
Figure 2.20 View of the releasing compartment: Extreme position in flexion
Chapter 2
30
Hooking compartment – Used to adjust the discharge range. It consists of two
coaxial components as before to which were added two arms connected as shown
in the figures (2.21) and (2.22); the two discs and the arms make up an articulated
parallelogram. The lower disc is the one that bears the release block while the
upper one host the hooking one. The function of the arms is to provide support to
the sledge of the arm spring to prevent it from falling in altitude and being released
from the exo-arm. Again, once chosen the hooking point and consequently the
range of charging, the discs are tightened together by a screw.
Figure 2.21 View of the hooking compartment: Position of minimum extension
Figure 2.22 View of the hooking sector: Position of maximum extension
Device description
31
Charging compartment – Used to adjust the working range. Once defined the
maximum extension point and the range of discharge, it only remains to set the
maximum flexion point. In order to do so, we shaped a coaxial disk to fit in the
block of charging and then fixed it with a screw on the spring-arm. The setting of
this component needs to be done with the spring loaded, or at least with the sledge
of the spring-arm blocked on the hooking block; this way it will be possible to
choose the point of maximum flexion by adjusting the diskette.
Figure 2.23 View of the charging compartment: In blue the charging-block
Chapter 2
32
Transmission and regulation of the torsional spring
In the previous section we focused more on the geometrical adjustments of the
device, we now turn to the dynamic adjustments. As already said several times,
the device uses a torsional spring for storing the energy; this spring is not directly
mounted on the device, but uses of a wire transmission. The main reason for this,
is the intrinsic behavior of the springs since, as we know, they have a linear trend
for the force or torque exerted as a function of elongation [16]:
𝐹 = 𝑘 ∙ 𝛥𝑙 (2.1)
This behavior is bad for the function that the spring has in the device, since torque
should be as uniform as possible. We analyzed the case in which the spring was
mounted without transmission taking into account that generally the torsional
springs allow a maximum deflection of 270°; What we have seen is that, if we
think that in order for the patient to make a 30° extension with the aid of the spring
our device accomplishes about 60° (assuming the use of spring loaded to the
maximum), the patient will get the maximum torque at the beginning of the
process (where he has less difficulty) and only 77% at the end. Moreover, the
possibility of torque adjustment in this case would be very limited.
Considered this issues it was chosen to introduce a transmission for the spring, in
order to maintain the drop of the torque to less than 10% when used in the worst
case (60° of excursion of the patient's arm). To save space and simplify the device,
it was chosen to introduce an unconventional transmission; taking advantage of
the spring geometry and considering the fact that it has always an excursion
inferior to 40° (for the actual sizing, see the paragraph 3.3), it was decided to
connect the upper arm of the spring to the shaft of the spring-arm by a cable as
shown below.
Speaking about the regulation of the torque exerted by the spring, it will be
regulated through another coaxial disk connected to the spring and place under it,
to which the lower arm of the spring will be bound once decided the setting (by
means of a screw).
Device components list
In this section we present the list of the device components. The following table
shows all the elements present in the device, or to ensure its functionality, both
those designed on purpose and the ones chosen from the catalog, such as bearings,
snap rings, screws and gears. The list is structured by dividing the components
into macro-groups that bear the names used during the discussion. The members
chosen from the catalog does not have the brand names but a summary of the
Device description
33
characteristics necessary for the installation and operation. Particular mention
goes to ball bearings, they have been chosen from the catalog just for their size,
since no wear test was needed considering neither static nor dynamic resistance,
because the loads involved are extremely low and the speed of rotation is not even
comparable to those that may undermine the component.
Device elements list divided in functional groups
Macro group Under category Component
Exo-arm Exo-arm Exo-arm
Spring-arm
Support
Shaft
Key, L 3 mm H 5 mm
Seeger, 3 mm x2
Washer, Dint 4mm Dext 12 mm
Screw, M4x10 mm
Seeger 8 mm
Contrast spring
Sledge
Internal horizontal arm
External horizontal arm
Internal sledge support
External sledge support
Inferior shaft
Superior shaft
Ramp-bar
Seeger, Dint 3 mm x3
Charging-block
Setting disc
Contact plate
Screw, M3x8 mm x2
Screw M5x8 mm
Support disc
Base Disco base
Screw, M5x8 mm x2
Releasing
compartment
Releasing disc
Releasing block
Quadrilateral arm releasing side
Screw, M2x5 mm countersunk head x3
Screw, M5x8 mm
Chapter 2
34
Hooking
compartment
Hooking disc
Hooking block
Quadrilateral arm hooking side
Screw, M2x5 mm countersunk head x3
Screw, M5x8 mm
Contour elements
Rotation supports
Ball bearing, Dint 8 mm Dext 12 mm H
3.5mm x2
Bushing, Dint 8 mm Dext 11 mm H 6mm
Gear transmission
Pinion, Dp 18 mm H 10 mm
Bull gear, Dp 36 mm H 10 mm
Transmission shaft exo side
Spring
transmission
Torque spring, Def max 270° Torque max
5000 Nmm
Wire transmission, Lmin 26 mm
Torque setting disc
Screw, M5x8 mm
Chapter 3
Device sizing
Genera sizing
After having analyzed the operation of the various parts of the device, in this
chapter we focus attention on the sizing of the various components. In particular,
in this paragraph we focus on the dimensions of the support disc and the size of
the circular ramps of the Exo-arm and the sledge. The device is largely
parameterized and the current size is due to the maximum forces that the device
must be able to support. A key note before we begin this journey concern the
technical specifications on the basis of which we went to size the components.
The maximum applicable torque which has made the sizing, 𝐶𝑚𝑎𝑥 = 2000 𝑁 ∙𝑚𝑚, is a very conservative value since this is the torque necessary for a healthy
person to move the arm horizontally [17], [18], [19], [20]. Once realized the
device will run tests to verify its performance, it is expected these will lead to a
reduction of this value. In fact, the purpose of the device is to provide an aid to
the patient not to move the limb in his place; limb that will, however, already
relieved of a part of its weight, balanced by the exoskeleton.
Sizing of the circular ramps
The geometry of these ramps is quite complex and deserves a certain deepening.
The first thing to note is that the angle of inclination of the ramp is not constant,
in fact, maintaining constant the height difference between the start and end of the
ramp and the angle of the circular sector in which the ramp develops, the angle of
inclination increases approaching the center, as seen in figure (3.1).
Chapter 3
36
Figure 3.1 Example of circular ramp, prospected view on left and lateral view on
right
Drawbacks due to this variation of the angle of inclination have been taken into
account. The superior shaft of the sledge has a convexity in the middle, in order
to come into contact with the ramp at its midpoint, and not on the inner rim where
it would touch if the shaft was simply cylindrical. We wanted to prevent this from
happening in first place for a more homogeneous distribution of the stresses on
the surface and then because a smaller angle inclination allows the ramp to raise
the sledge with less effort. In fact, greater the inclination greater will be the effort
made by the shaft to slide on the ramp.
The first step for the sizing of the ramp is the choice of the diameter of the sledge
inferior shaft, it determines the vertical translation which the sledge must fulfill.
The shaft diameter d and, consequently the height of the step h, was chosen on
the basis of the FEM analysis (for FEM analysis please refer to the section 3.4)
and the following value was chosen:
𝑑 = ℎ = 3 𝑚𝑚
This dimension is one of the parameters on which we will sizing the circular
ramps of Exo-arm. Choosing the distance ℎ given by the shaft we are going to
size the ramp imposing other two dimensions, the slope of the ramp in the medium
radius α and the angle of the circular sector at the base of the ramp 𝛽.
We suppose you draw on a plane the surface of the ramp taken in its middle range,
it will be a right triangle whose base 𝑏 is function of 𝛽 and the medium radius 𝑅𝑚
according to the following equation
𝑏 =2 ∙ 𝜋 ∙ 𝑅𝑚 ∙ 𝛽
360° (3.1)
Known 𝑏 and ℎ, the inclination angle 𝛼 is given by
Device sizing
37
𝛼 = tan−1 (ℎ
𝑏) (3.2)
We chose to use the following values for 𝛼 and 𝛽:
𝛼 = 30° 𝛽 = 10°
At this point it is possible to find the medium radius through the following relation
𝑅𝑚 =ℎ ∙ 360°
tan(𝛼) ∙ 2 ∙ 𝜋 ∙ 𝛽 (3.3)
We obtain 𝑅𝑚 = 30 𝑚𝑚
This value is very important, in fact we will use it not only to size the circular
ramps but will also serve for the setting mechanism and more generally imposes
the size that will have the entire device. The ramps width will be dimensioned
later on the basis of a stress analysis. In fact, although this dimension does not
appear in the geometrical design of the device, it is fundamental for the
functioning because both the superior and the inferior ramp must be able to
support the entire load of the spring, as we will see in section 3.4. In view of the
technical difficulties for the realization of a component with a similar geometry
we will try to maintain this dimension as small as possible.
Sizing of the sledge kinematism
The sledge that deals of the hooking and releasing of the torsional spring is
actually an articulated parallelogram. For this reason, it presents some more
project problems respect to a sledge with a simple vertical translation. Referring
to the figure (3.2), the side of the parallelogram called for simplicity sled is the
side 𝑐, it is always maintained vertical, but its mean point rotates around the mean
point of the side 𝑎 that connects the two support joints.
Chapter 3
38
Figure 3.2 View of spring-arm with highlighted in red the sides of the articulated
parallelogram
Always referring to the figure (3.2), we chose to keep the lock position of the
sledge, then at its lowest height, 𝛼 = 0 because, being the Spring-arm always
under load with a torque which pulls it in a clockwise direction (to the left in the
figure), it will tend spontaneously to go towards 𝛼 = 0 being this the point where
𝑐 is more distant from 𝑎 and the equilibrium position. So if we had chosen a
different 𝛼 value the sled was not be stable in the blocking position but it would
tend to get up by itself. The size of 𝑏 and 𝑑 (𝑏 and 𝑑 must be equal) has been
chosen in order to use all the allowed space in the device to minimize the distance
of 𝑐 from the vertical axis. In this way the approximation with a purely vertical
motion will be the best possible.
Referring now to the figure (3.3) we derive the relationship between the length 𝑏
and the distance between ℎ and the vertical axis called 𝑑, with ℎ the height of the
hooking-block step.
Device sizing
39
Figure 3.3 Scheme for the calculation of the horizontal movement of the sledge
𝑑 = 𝑏 ∙ (1 − cos(𝛼)) (3.4)
ℎ = 𝑏 ∙ 𝑠𝑖𝑛(𝛼) (3.5)
Actually, after a first sizing, b is resulted bounded for reasons due to the overall
dimensions of the device for which it was obtained 𝑏 = 12 𝑚𝑚 and then 𝑑 =0,38 𝑚𝑚 little more than 10% of ℎ then an acceptable value.
Sizing of the setting mechanism
As mentioned in section 2.4, the setting mechanism of the discharge range is an
articulated quadrilateral. For his sizing we refer to the figure which highlights in
red the four sides of the parallelogram and in blue and the circumferences of the
mean radius 𝑅𝑚 , that we calculated in the previous paragraph, and those of
maximum and minimum radius 𝑅𝑚𝑖𝑛 and 𝑅𝑚𝑎𝑥, respectively. These last two radii
are those that will determine the final length of the sledge shafts and consequently
will affect the sizing of the spring-arm and the final size of the device.
Chapter 3
40
Figure 3.4 View of the setting mechanism for the range of discharge with
highlighted in red the articulated quadrilateral
The goal that has been pursued during the dimensioning of this mechanism is to
try to reduce to the minimum the size, looking for the dimensions of the arms such
that 𝑅𝑚𝑎𝑥 is minimized.
According to the project specifications, the angular range in which it must be
possible to adjust the device about the discharging phase is [20°÷60°]. After a first
sizing of it we obtained that, for lack of space we will have a minimum angle
𝛼𝑚𝑖𝑛 = 39°. We have therefore chosen to have a maximum angle 𝛼𝑚𝑎𝑥 = 117° in order to subsequently insert a gear transmission between the device and the
exoskeleton which allows to obtain the desired range.
Actually 24° of the 39° are not due to the size of the mechanism arms but to the
hooking and releasing blocks, then the mechanism has been sized in the range
𝛼 = [15° ÷ 93°].
Known the angles 𝛼𝑚𝑖𝑛, 𝛼𝑚𝑎𝑥 and the mean radius 𝑅𝑚, we derive a system of
equations to calculate the lengths of the sides 𝑎, 𝑏, 𝑐, 𝑑 and the 𝑅𝑚𝑎𝑥 and
𝑅𝑚𝑖𝑛 radii. We know also that
Device sizing
41
𝑎 = 𝑑
𝑏 = 𝑐
𝑅𝑚 =𝑅𝑚𝑎𝑥 + 𝑅𝑚𝑖𝑛
2
(3.6)
Geometrically it is obtained that for the same length of 𝑎 is minimized 𝑏 and
consequently 𝑅𝑚𝑎𝑥 imposing that, when 𝛼 is maximum, the midpoint of side 𝑏
lies on the circumference of radius 𝑅𝑚𝑖𝑛and that 𝑏 is tangent to this as shown in
figure (3.5).
Figure 3.5 Scheme 1 for the design of the setting mechanism for the discharging
range
Note that the segment that connects the midpoint of the side 𝑏 and 𝐴 is equal to
𝑅𝑚𝑖𝑛 for construction. Moreover, always for construction this segment is
perpendicular to b.
We obtain that:
𝑅𝑚𝑖𝑛 = 𝑎 ∙ cos (𝛼𝑚𝑎𝑥4
) (3.7)
𝑏 = 2 ∙ 𝑎 ∙ sin (𝛼𝑚𝑎𝑥4
) (3.8)
Considering instead the situation in which 𝛼 is equal to 𝛼𝑚𝑖𝑛 , we have the
Chapter 3
42
situation in figure (). We can see how in this case 𝑅𝑚𝑎𝑥 is equal to the distance
between 𝐴 and 𝐶.
Figure 3.6 Scheme 2 for the design of the setting mechanism for the discharging
range
We obtain:
𝑅𝑚𝑎𝑥 = 𝑎 ∙ cos (𝛼𝑚𝑖𝑛2) + √𝑎2 ∙ (sin (
𝛼𝑚𝑖𝑛2))
2
+ 𝑏2 (3.9)
At this point we have found the following system of equations:
{
𝑅𝑚 =
𝑅𝑚𝑎𝑥 + 𝑅𝑚𝑖𝑛2
𝑅𝑚𝑖𝑛 = 𝑎 ∙ cos (𝛼𝑚𝑎𝑥4
)
𝑏 = 2 ∙ 𝑎 ∙ sin (𝛼𝑚𝑎𝑥4
)
𝑅𝑚𝑎𝑥 = 𝑎 ∙ cos (𝛼𝑚𝑖𝑛2) + √𝑎2 ∙ (sin (
𝛼𝑚𝑖𝑛2))
2
+ 𝑏2
(3.10)
Solving we find the following relations
Device sizing
43
{
𝑎 = 𝑑 =
2 ∙ 𝑅𝑚
cos (𝛼𝑚𝑎𝑥4 ) + cos (
𝛼𝑚𝑖𝑛2 ) + √(sin (
𝛼𝑚𝑖𝑛2 ))
2
+4 ∙ (sin (𝛼𝑚𝑎𝑥4 ))
2
𝑏 = 𝑐 = 2 ∙ 𝑎 ∙ sin (𝛼𝑚𝑎𝑥4
)
𝑅𝑚𝑖𝑛 = 𝑎 ∙ cos (𝛼𝑚𝑎𝑥4
)
𝑅𝑚𝑎𝑥 = 2 ∙ 𝑅𝑚 − 𝑅𝑚𝑖𝑛
(3.11)
We found the following values
𝑎 = 𝑑 = 21,1 𝑚𝑚 𝑏 = 𝑐 = 20,6 𝑚𝑚 𝑅𝑚𝑖𝑛 = 18,4 𝑚𝑚 𝑅𝑚𝑎𝑥 = 41,6 𝑚𝑚
Transmissions design
As mentioned in chapter 2 the device includes two transmissions that connect this
to the exoskeleton and to the torsional spring. A good gain in terms of space was
achieved by placing coaxially the two transmissions. Furthermore, if for the
connection between the exoskeleton and the device Exo-arm we propose a gears
transmission with straight teeth, to the side of the torsional spring we have chosen
to connect the rod of the spring to the device Spring-arm with a wire. The sizing
of this second transmission is far less immediate than the first and deserves a
detailed discussion.
Sizing of the gear transmission
Let's start with the sizing of the gear transmission. As already mentioned, we have
not been able to maintain a direct transmission due to the encumbrance of the
components.
Required range: 20° ÷ 60° Minimum value due to encumbrances in the device: 39° Maximum angle necessary for the device:
𝑎𝑛𝑔𝑑𝑒𝑣,𝑚𝑎𝑥 =60° ∙ 39°
20°= 117° (3.12)
We also get a gear ratio
Chapter 3
44
𝑟 =117°
60°= 1,95 (3.13)
For simplicity we take 𝑟 = 2
We get the following ranges
𝑎𝑛𝑔𝑒𝑥𝑜 = 19,5° − 58,5° 𝑎𝑛𝑔𝑑𝑒𝑣 = 39° − 117°
The toothed wheel with the smaller diameter in the catalog, without going to
search for particular pieces, has a pitch diameter equal to 𝐷𝑝,𝑝𝑖𝑛𝑖𝑜𝑛 = 18 𝑚𝑚.
This is the pinion and must be shrunk on the Exo-arm. Since the diameter of the
pinion is less than the minimum one of the sledge, it was possible to save space
by mounting the gear just below the height in which the Ramp-arm works (see
figure (3.7)).
Therefore, having to maintain a ratio of the gear 𝑟 = 2 the pitch diameter of the
bull gear will be 𝐷𝑝,𝑏𝑢𝑙𝑙 𝑔𝑒𝑎𝑟 = 36 𝑚𝑚 [22].
Figure 3.7 View of the gear transmission of the device, in grey the bull gear and
in orange the pinion
Device sizing
45
Design of the spring transmission
Now let’s move to the sizing of the torsional spring transmission.
Having decided that the torsional spring is coaxial to the axis of the bull gear, we
know that the distance between the spring axis of rotation and that of the Spring-
arm is imposed by the gear transmission and equal to
𝑑 =18 + 36
2= 36 𝑚𝑚 (3.14)
To maintain the torque transmitted as linear as possible during the operation of
the device, we imposed that, with the maximum discharging range of the device
had a reduction of the transmitted torque lower than the 10%. Since this is the
worst case, in case of the lower range will result a less reduction. Assuming to
choose a spring with a maximum deflection of 270 ° we obtain the following
parameters:
Maximum deflection: 𝐷𝑒𝑓𝑚𝑎𝑥 = 270° Maximum deflection admitted: 𝐷𝑒𝑓𝑚𝑎𝑥,𝑎𝑑𝑚𝑖𝑡 = 270° ∙ 10% = 27° Maximum angular range of the spring-arm: 𝛼𝑑𝑒𝑣,𝑚𝑎𝑥 = 117°
Transmission ratio: 𝑟 =117°
27°= 4.33
The diameter of the spring-arm shaft is 𝐷𝑠ℎ𝑎𝑓𝑡 = 8 𝑚𝑚, so the diameter at which
we will have to connect the wire to the spring rod will be
𝐷𝑠𝑝𝑟𝑖𝑛𝑔 = 4,33 ∙ 8 = 34,64 𝑚𝑚
Having available more space, we decided to bring the diameter 𝐷𝑠𝑝𝑟𝑖𝑛𝑔 =
40 𝑚𝑚, obtaining a transmission ratio 𝑟 = 5.
So we obtain 𝐷𝑒𝑓𝑚𝑎𝑥,𝑎𝑑𝑚𝑖𝑡 = 23.4° We dimensioned the spring and the transmission ratio as if it is constant, this is
true for normal transmissions such as the gears one sized above, it is not
absolutely true for the transmission of the spring which has the appearance shown
schematically in figure (3.8).
Chapter 3
46
Figure 3.8 Schematic representation of the torque spring transmission, in grey
the torque spring, in yellow the Spring-arm shaft and in red the wire who
connects the previous elements
The first thing that we have verified in phase of the project is that there was a
fairly wide range in which the transmission ratio was constant. First we search the
mathematical relationship that connects the rotations of shaft and spring. We refer
to the figure (3.8) calling the radius of the torsion spring at the point where the
wire is connected, 𝑅1, and the radius of the spring-arm shaft, 𝑅2. We derive the
mathematical relationship between the length 𝑠 to the rotation 𝛼, this greatly
simplifies the geometry of the problem. It is also easy to show that
𝑙 = √𝑠2 − 𝑅22 (3.15)
This is because the wire will always be tangent to the shaft and thus will be
perpendicular to the radius 𝑅2 at the contact point 𝐵. We project the segment 𝑂1𝐵
on the segment 𝑂1𝑂2 and on its traverse passing for 𝑂1. We call the two
projections 𝑠𝑦 and 𝑠𝑥 respectively. These projections are function of 𝛼 as follows:
Device sizing
47
{𝑠𝑥 = 𝑅1 ∙ sin(𝛼)
𝑠𝑦 = 𝑅1 ∙ cos(𝛼) + 𝑑 (3.16)
Where 𝑑 is the length of the segment 𝑂1𝑂2.
We obtain
𝑠 = √𝑠𝑥2 ∙ 𝑠𝑦2 = √𝑅12 ∙ (sin(𝛼))2 + (𝑅1 ∙ cos(𝛼) + 𝑑)2 (3.17)
By combining the formulas (3.15) and (3.17) we get the graph of 𝑙 in function of
𝛼 (figure (3.9)).
Figure 3.9 Graphic of the length l in function of the angle α
We call 𝛽 the rotation of the shaft and 𝑑𝑙 the difference between 𝑙 and 𝑙𝑚𝑖𝑛 where
𝑙𝑚𝑖𝑛 = 𝑙(𝛼 = 180) = 12 𝑚𝑚. We have that
𝛽 =𝑑𝑙 ∙ 360°
2 ∙ 𝜋 ∙ 𝑅2 (3.18)
Chapter 3
48
The reason why we use 𝑑𝑙 instead of 𝑙 is due to the fact that 𝑙𝑚𝑖𝑛 does not
intervene in the rotation 𝛽 because it never arrives to wind on the shaft.
Figure 3.10 Graphic of the angle β in function of the angle α
We calculate at this point as change the transmission ratio 𝑟 in function of 𝛼
𝑟 =𝛽(𝛼 + 𝑑𝛼) − 𝛽(𝛼)
𝑑𝛼 (3.19)
Device sizing
49
Figure 3.11 Graphic of the ratio r in function of the angle α
The analysis clearly shows a maximum in the transmission ratio for 𝛼 = 141,3° with 𝑟 = 5,14 (see figure (3.11)). In this area, how we can also see in the
previously graphs (figure (), figure ()), is present a range where the dependence
of 𝛽 from 𝛼 tends to be linear.
At this point it remains to calculate the effective range in which the spring-arm
moves. In fact, we will have to take into account the maximum extremes that the
spring-arm sledge can reach in all configurations. We place equal to 120° ∙ 2 =240 ° the maximum extension point, it is definitely one extreme of the range we're
looking for. We can find the other extreme setting the device with the minimum
discharge range and the minimum working range, which means in the user's
coordinates, where 0° is the maximum deflection, a discharging range of 20° and
a working range of 40° (the double the discharging range). We will have that the
second extreme will be 20° which multiplied by the reduction ratio becomes 40° then the actually range where the spring-arm operate will be
𝑑𝛽 = 240° − 40° = 200°
Dividing this value by the reduction ratio that we have seen to be around 5 in the
Chapter 3
50
peak, we get the width of the range in which the spring-arm will operate
𝑑𝛼 =200°
5= 40°
Then we are looking for a range for 𝛼 of 40° in which the transmission ratio is as
high as possible. We have done an optimization researching the maximum
average value of 𝑟 on 40° in the range of [0° ÷ 180°]. We found the following
range
𝛼 = [117,2 ÷ 157,2]
We will need to forecast two limits positioned in the locations found. The average
final ratio is
𝑟𝑚 = 5,031
As concern the maximum torque transmitted by the spring, it will have to take
account of transmission ratios. Having to obtain a maximum torque at patient's
arm of 2000 𝑁 ∙ 𝑚𝑚, this will be 1000 𝑁 ∙ 𝑚𝑚 at the device level and will rise
to 5000 𝑁 ∙ 𝑚𝑚 at the level of the torsional spring. Therefore, the researched
torsional spring will have the following characteristics:
Maximum deflection: 270° Torque at the maximum deflection: 5000 𝑁 ∙ 𝑚𝑚
Torsional stiffness: 18,52 𝑁 ∙ 𝑚𝑚/𝑑𝑒𝑔
FEM analysis of the most stressed components
We speak in this section of the FEM analysis that were conducted to size or test
the components in which the concentration of stress is greater. These components
are essentially all the elements that make up the sledge, which is undoubtedly the
critical component of the entire device since it is constantly subjected to the forces
generated by the spring. Furthermore, its components have the resistant sections
that are inferior. Other components that need to be checked are the ramps of the
exo-arm since they present the same characteristics of the sledge as state above.
Sizing of the inferior shaft of the sledge
We introduced at the beginning of the second chapter that at the core of all the
sizing procedure of the device there is the diameter of the inferior shaft of the
Device sizing
51
sledge. This has been sized by simply analyzing the finite elements in order to
seek the minimum diameter to withstand the maximum load provided by the
spring. For simplicity the shaft was assumed to be bound completely
(interlocking) to both ends and subject to load in its central part.
The finite element analysis was conducted with the Autodesk Inventor
Professional 2015 software, and the results are the following:
A first analysis was made assuming a shaft diameter of 2 mm.
Stated this value all the sizing step previously presented were followed in order
to find the length that it should have and its part subjected to the load (useful
length). The dimensions used are:
Diameter: 𝐷 = 2 𝑚𝑚
Total length: 𝐿𝑡 = 19,5 𝑚𝑚
Useful length: 𝐿𝑢 = 15,5 𝑚𝑚
2 mm are left outside to each end where the lateral supports come into contact.
The maximum stress situation is the one in which the spring is charged with the
total torque required, equal to 𝐶 = 1000 𝑁 ∙ 𝑚𝑚. Furthermore, by the sizing, it
resulted a medium radius of 𝑅𝑚 = 20 𝑚𝑚; Therefore, the medium strength
applied is equal to
𝐹 = 𝐶/𝑅𝑚 = 50 𝑁.
In figures (3.12), (3.13), (3.14) are reported the images with the results obtained
in terms of Von Mises stress, displacement and Safety Factor, respectively (the
component was assumed to be made of steel) [23].
Chapter 3
52
Figure 3.12 FEM analysis of the inferior shaft first design, Von Mises Stress
arrangement
Figure 3.13 FEM analysis of the inferior shaft first design, Displacement
arrangement
Device sizing
53
Figure 3.14 FEM analysis of the inferior shaft first design, Safety Factor
arrangement
As easily predictable the greater part of the stress is at the interlocking point, and
it has a maximum value of Von Mises of 302,5 𝑀𝑝𝑎. This generates a minimum
Security Factor of 0,68. Even if this model is very conservative a value of 0,68 is
not acceptable and, in general, any value inferior to 1 would be not acceptable. So
it was decided to use a security value for the shaft of 1,5.
The test was then repeated with a diameter value of 3mm.
Recapping:
Diameter: 𝐷 = 3 𝑚𝑚
Total length: 𝐿𝑡 = 27,65 𝑚𝑚
Useful length: 𝐿𝑢 = 23,2 𝑚𝑚
Applied strength: 𝐹 = 1000/30 = 33,33 𝑚𝑚
The results with these values are shown in figures (3.15), (3.16), (3.17).
Chapter 3
54
Figure 3.15 FEM analysis of the inferior shaft final design, Von Mises Stress
arrangement
Figure 3.16 FEM analysis of the inferior shaft final design, displacement
arrangement
Device sizing
55
Figure 3.17 FEM analysis of the inferior shaft final design, Safety Factor
arrangement
We can see that at an increase of the diameter corresponds an increase in the arm
and a decrease in the strength if the torque is kept constant. For this reason, the
effort value grows significantly with the variation of the diameter and the security
values results in 2,39, clearly superior to the 1,5 required. It has to be taken into
consideration however that the device will be realized with the 2 mm diameter if
the torque actually used were to be inferior.
The superior shaft bears the same characteristics of the inferior one except for the
small convexity at the center, so we avoid to report its analysis because not
interesting. Moreover, the superior shaft is the pin for the parallelogram and needs
two Seeger to bound the axial translation. The commercial Seeger have a diameter
minimum of 3 mm and this represents a limitation to the willingness of look for
inferior diameter.
Sizing of the circular ramps of the exo-arm
When we described the geometrical sizing of the circular ramps in paragraph 3.1,
we didn’t stet any limitation for their width. In fact, as state above, the width is
limited only by the load that the component needs to withstand. Realistically any
value inferior to the distance between the two supports that prevents the ramp
from flexing would be good. It’s worth reminding that the distance mentioned
does not depends over the width of the ramps.
Starting with the superior ramp, the values used are:
Chapter 3
56
Width of the ramp: 𝑏 = 10 𝑚𝑚
Torque applied: 𝐶 = 1000 𝑁 ∙ 𝑚𝑚
Average arm: 𝐿 = 30 𝑚𝑚
Average strength: 𝐹 = 𝐶/𝐿 = 33,33 𝑁
If the sledge was to move only vertically, this analysis wouldn’t be necessary,
since the superior ramp would have to face only the rolling friction strength
created by the contact between the inferior shaft and the hooking block (plus a
little strength generated by the spring pushing the sledge against the support disc).
However, being the sledge part of an articulated parallelogram, the superior ramp
to raise needs to face the strength generated by the torsional spring. So, we will
use the maximum torque in this analysis.
The results:
Figure 3.18 FEM analysis of the exo-arm superior ramp, Von Mises Stress
arrangement
Device sizing
57
Figure 3.19 FEM analysis of the Exo-arm superior ramp, displacement
arrangement
Figure 3.20 FEM analysis of the Exo-arm superior ramp, Safety Factor
arrangement
We see that, in contrast with the case of the inferior shaft of the sledge in which
we were interested in the safety factor to avoid the failure of the component, in
this case we are interested in the downward displacement of the ramp. In fact, an
excessive displacement would prevent the sledge to raise enough to overcome the
step of the hooking block before the ramp ends. In any case we see that, already
Chapter 3
58
in the first attempt, with a 10 mm width, the displacement is less than one tenth
of a millimeter.
With a similar situation, the inferior sledge width is enough to allow a good
operation even with larger displacement.
We report the results obtained by the FEM analysis for completeness:
Figure 3.21 FEM analysis of the Exo-arm inferior ramp, Von Mises Stress
arrangement
Device sizing
59
Figure 3.22 FEM analysis of the Exo-arm inferior ramp, displacement
arrangement
Figure 3.23 FEM analysis of the Exo-arm inferior ramp, Safety Factor
arrangement
Chapter 3
60
Check of the sledge ramp
We conclude with the verification of the third component that is the sledge ramp
that coupling with the exo-arm allows to this to be lulled by the spring-arm. Also
in this case we use the maximum spring torque. The width of this ramp is
irrelevant for the load to failure and displacement analysis, for this reason, the
width of this ramp is kept equal to those of the exo-arm ramps. The static strength
verification relates in this case the bar on which the ramp has been obtained. The
bar has the following dimensions:
Length: 𝐿 = 27,65 𝑚𝑚
Width: 𝐵 = 4,4 𝑚𝑚
Height: 𝐻 = 4 𝑚𝑚
We check then if with these values of first attempt the component presents
acceptable values of stress and displacement.
Figure 3.24 FEM analysis of the sledge ramp, Von Mises Stress arrangement
Device sizing
61
Figure 3.25 FEM analysis of the sledge ramp, displacement arrangement
Figure 3.26 FEM analysis of the sledge ramp, Safety Factor arrangement
Chapter 3
62
We get a maximum Von Mises stress value equal to 46,9 MPa, which corresponds
to a Safety Factor of 4,41 well beyond the minimum value sought and finally a
maximum displacement of 0.002 mm. The component is so well designed and do
not presents risks of failure.
Dynamic simulation with rigid bodies model
In this section we analyze the dynamic behavior of the system. It has been created
implementing a multi-body model of rigid body of the device using a software for
simulation: Autodesk Inventor Professional 2015.
Let's start with the description of model, since, to reduce the computational burden
of the simulations, it was generated a simplified version of the device. It provides:
- The set of components called in general the support disc was bound to a
range of discharging of 60°.
- A simplified version of the spring-arm, where all the rigidly connected
parts have been modeled as a single component.
- The exo-arm that results unchanged except for the lack of the gear
transmission that is not comprehended in the simulation.
Figure 3.27 View of the simplified model of the device used for the dynamic
simulation
Device sizing
63
Figure 3.28 View of the simplified model in order to see the contact between the
sledge inferior shaft and the Hooking-block in the simulation start point
In the figure it has been removed the upper arm of the parallelogram and the outer
side one of the sledge in order to allow a better view of the lock of the lower shaft
by the component called hooking-block.
What we see in the two figures is the starting point of the dynamic simulation. In
this simulation we verify the correct developing of the discharging phase, in
which the exo-arm is connected with the spring-arm until the releasing point at
the end of the phase.
In this simulation too, we have worked assuming the most critical situation that is
when the spring is at maximum charge. Furthermore, although the device
functions primarily with small inclinations with respect to the horizontal plane,
the gravitational strength has not been considered; in fact, in the real version, there
will be a small spring with the aim to always keep in contact the lower shaft of
the sledge with the support disc.
The purpose of the dynamic analysis was also to evaluate the force required by
the spring and, to do so, it was assumed that it should be sufficient to ensure
operation even in case of a 90° inclination over the horizontal plane. A higher
inclination is physically impossible and was not taken into account.
The analysis focused solely on the discharge phase, in fact, the other phases of
operation are extremely simple, and it was not considered necessary to invest
more time in their analysis.
On the contrary, the discharge phase is actually complex and deserves to be
investigated deeply to be sure of its correct operation.
We report the final set parameters, they originate from many tests that have
guaranteed the operation of the device and an improvement of its design even
before it was actual manufactured:
- Constant torque applied to the spring-arm: 𝐶 = 1000 𝑁 ∙ 𝑚𝑚.
- Force applied to the sledge: 𝐹 = 3 𝑁.
Chapter 3
64
It was also taken into account the dissipated forces due to the rubbing in the
parallelogram hinges; the value assumed is:
- Damping: 𝐷 = 5 𝑁 ∙ 𝑚𝑚 ∙ 𝑠/𝑑𝑒𝑔
In the simulation, in order to verify the forces acting on the exo-arm, this has a
rotation speed set to:
- Rotation speed: 𝜔 = 30 𝑑𝑒𝑔/𝑠
- Simulation time: 𝑡 = 2 𝑠
Our main interest in the multi body simulation is to seek torques developed during
the course of the simulation on the exo-arm. This, in fact, is mechanically
connected to the patient's arm through the exoskeleton and then the forces acting
on it will directly be transmitted to the user. Although ideally the trend of the
torque transmitted to the entire work cycle should resemble the one in the figure,
we know that even in the hooking phase the patient will have to exert a torque
enough to allow the sledge to rise. The evaluation of this torque at an analytical
level can be obtained by applying the principle of conservation of energy as
follows:
If the sledge was free to translate, we have calculated that the distance it would
travel would be: 𝑑 = 0.38 𝑚𝑚; however, since it is bound, this distance will be
traveled by the spring-arm and it will be equal to:
𝛼 =𝑑 ∙ 360°
2 ∙ 𝜋 ∙ 𝑅𝑚= 0,73° (3.20)
This formula has been used considering the torque of the torsion spring to be
constant. This hypothesis can be established since the transaction takes place in
few tenths of rotation degree.
Knowing the transmission ratio between the spring and the arm, 𝑟 = 5, the
rotation beta of the spring is equal to:
𝛽 =𝛼
5= 0,146° (3.21)
So the energy to be transmitted to the spring is:
𝐸𝑢 = 5000 ∙ 𝛽 = 730 𝐽 (3.22)
Device sizing
65
To this we need to add the dissipated energy due to the friction over the lower
shaft following the rolling friction model:
𝐹𝐻 = 𝑁 ∙ 𝑓𝑣 (3.23)
With N standing for the transversal load equal to 33.33N and 𝑓𝑣 standing for the
rolling friction coefficient (in the case of steel-steel friction) equal to 5 ∙ 10−3
𝐹𝐻 = 0,33 𝑁. To this we need to add the 3 N given by the contrast spring.
Assuming the movement of the sledge in the phase before it rises of 2 mm, the
energy dissipated by these two contributions are:
𝐸𝑑 = (3 + 0,33) ∙ 2 = 6,66 𝐽 (3.24)
On the other side, the patient employs the big part of the movement over the ramp
(width 10°) in order to lift the sledge; assuming that 7° are needed before the rise,
the actual ones for the patient are 3.5°. This means that the torque exert by the
patient should be:
𝐶𝑚 =𝐸𝑢 + 𝐸𝑑3,5°
= 210 𝑁 ∙ 𝑚𝑚 (3.25)
Figure 3.29 Graphic of the theoretical working cycle of the device in function of
the flexion angle
Chapter 3
66
We now report the results of the simulation in order to see if the calculations are
realistic.
Figure 3.30 Dynamic Simulation results: Torque transmitted to the user during
the discharging phase
The graph presents the trend of the torque exerted on the patient's arm as a
function of the independent coordinate alpha that is the advancement of the arm
from the starting point presented in the figure. In red is reported the average of
each value for each interval of 1°. You can see how, in the first phase, the
transmitted torque is negative since it is exercised by the patient. This phase lasts
approximately 4 °, then there is a zero torque space in which the sledge raises by
itself, but the ramp of the sledge is not yet come into contact with the ramp of the
lower exo-arm. After that the hooking occurs and the spring starts to drag the
patient's arm and the transmitted torque has a downward trend with the
advancement. Finally, after about 26° it is released and the torque falls to zero.
The peaks present in the graph during the discharging phase are probably due to
the interaction between the various rods of the adjusting quadrilateral, which have
created problems to the resolver during the computation.
Results:
Average value in the range [0° ÷ 4°]: −500 𝑁 ∙ 𝑚𝑚
Device sizing
67
Average value in 'interval [5° ÷ 25°]: 2000 𝑁 ∙ 𝑚𝑚
We report the performance of the elevation of the lower shaft of the sledge,
function of the independent coordinate 𝛼
Figure 3.31 Dynamic Simulation results: Sledge elevation during the discharging
phase
In the last graph that we present we can see how the transmitted torque to the exo-
arm changes in function of the elevation of the sledge. In particular, we see that
in the range between 2 and 3 mm no torque is applied to the arm. In fact, in this
zone, the sledge starts to autonomously rise and covers the little gap caused by
the movement of the superior ramp of the exo-arm and hooks to the lower one.
Chapter 3
68
Figure 3.32 Dynamic Simulation results: Torque transmitted in function of the
sledge elevation
Chapter 4
Prototype realization
Specification of the prototype
Once concluded the virtual prototyping phase, it was decided to design and realize
a prototype to effectively verify proper operation and feasibility. Being the
designed device very small, it has not been possible to reproduce it to its actual
size due to the lack in accuracy of the machines available at the institute; it was
therefore decided to develop a prototype with a larger size.
The prototype has followed the steps for dimensioning shown in the previous
chapter, however, the starting sizes were chosen arbitrarily in order to be easily
realized with the means available. Furthermore, the project was adapted in order
to use some mechanical parts, such as bush and bearings, which were already
present in the workshop.
Finally, since we wanted to verify only the good dynamics of the mechanism, it
was decided not to implement the adjustment mechanisms. In particular, it was
decided to create a support disc in a unique block with a fix and representative
settings of 30° for the charging and discharging range, and 120° for the working
range. Consequently, also the charging block of the spring-arm will have only one
mounting position instead of the entire expected range.
For the material, having available a 3D printer, it was decided to 3D print most of
the pieces and made with to machine tools (ones with numerical control) only
those that needed a more resistant material or with sizing tolerances more tight.
Assuming to use the same torsional spring, the loads present will be the same,
while the resistant surfaces will be higher; for this reason, it was decided to keep
valid the analysis of the efforts already made to the device without going in further
details.
We report now the list of the components used for the construction of the
prototype.
Chapter 4
70
Figure 4.1 Front view of the prototype
Figure 4.2 Rear view of the prototype
Parts produced with CNC
In this section we are going to preset the prototype parts that needs to be made of
metal by machine tools; these parts will be the ones more subject to efforts.
All the sledge components will be realized through chip removal as these are the
parts most stressed part of the entire structure.
The side supports of the sledge, and the two horizontal arms will be made of
aluminum since they do not require a particular resistance but a good dimensional
accuracy.
Instead, the two shafts will be made of brass to prevent the aluminum-aluminum
contact to seize up the mechanism. In fact, we remind that the aluminum-
aluminum contact is often cause of blocks due to the creation of micro-welds in
Prototype realization
71
response of the sliding of the two surfaces.
With regard to the ramp of the sledge, it will be hybrid; it will have a brass core
connected to the inner side arm, while the rest of the bar including the ramp will
be realized in 3D molded plastic.
The support shaft and the key will also be made of aluminum or steel.
Finally, the spring-arm is made of aluminum.
Parts produced with 3D printer
The parts made by 3D printing will instead be the support disc, in its version as a
single piece, charging support and block for the spring-arm is (also in one piece),
and finally the pedestal that will serve not only as support to the device but also
as installation set for the torsional spring.
We want to clarify that the parts produced as a single block will be created in the
geometry of the device in order to detect any issues that may be encountered
during operation due to them. Think, for example, to the arms of the adjustment
quadrilateral, which must support the slide during the discharge phase; if they are
not leveled, they could give tremors during the motion, especially in the joint
areas.
Prototype components list
Macro group Component Production Material
Exo-arm Exo-arm CNC Aluminum
Spring-arm
Main shaft CNC Aluminum
Key CNC Aluminum
Support 3D printing Plastic
Contrast spring Manual Steel
Horizontal arm x2 CNC Aluminum
Internal sledge support CNC Aluminum
External sledge support CNC Aluminum
Inferior shaft CNC Brass
Superior shaft CNC Brass
Ramp-bar core CNC Brass
Ramp-bar body 3D printing Plastic
Bushing Dint 3 mm Dext
5mm L 4 mm x2 CNC Aluminum
Chapter 4
72
Seeger 4 mm x3 Buying Steel
Screw, M3x6 mm x2 Buying Steel
Screw, M4x10 mm Buying Steel
Washer, Dint 4mm
Dext 12 mm Buying Steel
Support disc
Support disc 3D printing Plastic
Pedestal 3D printing Plastic
Screw, M5x8 mm x2 Buying Steel
Contour
elements
Bearing x2 Buying Steel
Bushing Buying Plastic
Torque spring Buying Steel
Chapter 5
Conclusion and future developments
Figure 5.1 View of the device mounted on the LIGHTarm exoskeleton
In conclusion, we can say that the presented device meets all the required
specifications satisfactorily, we have to make a few final comments on what has
been done and what will be the next phases of development.
We can see the device designed as a very simple and functional basic mechanism,
to which were added the accessory parts, such as transmissions and the setting
quadrilateral, to make it suitable to the required specifications. It presents some
improvable solutions for the proposed use or not needed in different situations. In
the future, we will check if the device characteristics match the patient's effective
needs, in which case we will continue with the improvement of the device up to
arrive at a real commercial product.
Chapter 5
74
Bibliography
[1] Duncan E Wood, JH Burridge, FM Van Wijck, C McFadden, RA Hitchcock,
AD Pandyan, A Haugh, JJ Salazar-Torres, and Ian D Swain. “Biomechanical
approaches applied to the lower and upper limb for the measurement of
spasticity: a systematic review of the literature”. Disability & Rehabilitation,
27(1-2):19-33. 2005.
[2] Riccardo Secoli. “Controllo di robot per la riabilitazione dell'arto superiore
di pazienti post-stroke” PhD thesis, Scuola di dottorato di Ingegneria Industriale,
Meccatronica e Sistemi Industriali Università degli studi di Padova. 2010.
[3] Randolph J Nudo, Birute M Wise, Frank SiFuentes, and Garrett W Milliken.
“Neural substrates for the effects of rehabilitative training on motor
recovery after ischemic infarct”. Science, 272(5269):1791-1794. 1996.
[4] G. Spagnuolo. “Esoscheletro per la riabilitazione dell'arto superiore
cinematicamente ottimizzato per I movimenti di spalla e gomito”. Tesi di laurea
magistrale. A.A. 2012/2013
[5] M. Malosio, N. Pedrocchi, F. Vicentini, L. Molinari Tosatti. “Shoulder
adaptable and elbow singularity-free exoskeleton”.
[6] M. Malosio, N. Pedrocchi, F. Vicentini, L. Molinari Tosatti. “Analysis of
elbow-joints misalignment in upper-limb exoskeleton”.
[7] M. Malosio, N. Pedrocchi, F. Vicentini, L. Molinari Tosatti, M. Caimmi, F.
Molteni. “Manipulators: Theory, types and applications, Chapter 1 Robotics in
rehabilitation – Part II: Design of devices and mechanisms”. Nova science
publishers, Inc. 2007.
[8] M. Malosio, G. Spagnuolo, M. Caimmi, J. C. Dalberto, N. Pedrocchi, F.
Vicentini, L. Molinari Tosatti. “LIGHTarm: a highly adaptable gravity-
compensated exoskeleton for upper-limb rehabilitation and ADL assistance”.
2014.
[9] A.Scano, G. Spagnuolo, M. Caimmi, A. Chiavenna, M. Malosio, G. Legnani,
L. Molinari Tosatti. “Static and dinamic characterization of the LIGHTarm
exoskeleton for rehabilitation”.
Bibliografy
76
[10] N. Jarrassé, T. Proietti, V. Crocher, J. Robertson, A. Sahbani, G. Morel, A.
Roby-Brami. “Robotic exoskeletons: A perspective for the rehabilitation of arm
coordination in stroke patients”. Frontiers in human neuroscience. 2014.
[11] R. Irene. “La mobilizzazione come prevenzione del dolore alla spalla nel
paziente emiplegico”. Tesi di laurea. A.A. 2014/2015
[12] https://it.wikipedia.org/wiki/Orologio_automatico
[13] http://www.magnetimarelli.com/it/eccellenze/eccellenze-tecnologiche/kers
[14] https://it.wikipedia.org/wiki/Volano_(batteria)
[15] http://www.jointactivesystems.com/For-Patients/How-JAS-Works.aspx
[16] N. Bachschmid, S. Bruni, A. Collina, B. Pizzigoni, F. Resta, A. Zasso.
“Fondamenti di meccanica teorica e applicata, seconda edizione”. McGraw-Hill.
2010.
[17] L. Almeida, D. M. Corcos, Z. Hasan. “Horizontal-Plane Arm Movements
with Direction Reversals Performed by Normal Individuals and Individuals with
Down Syndrome”. 2000.
[18] A. Murray, Garth R. Johnson. “A study of the external forces and moments
at the shoulder and elbow while performing everyday tasks”. 2004.
[19] N. V. Dounskaia, C. J. Ketcham, G. E. Stelmach. “Influence of
Biomechanical Constraints on Horizontal Arm Movements”. Human Kinetics
Publishers, Inc. 2002.
[20] A.J. Bennett, J. M. Hollerbach, Y. Xu, I. W. Hunter. “Time-varying stiffness
of human elbow joint during cyclic voluntary movement”. Experimental Brain
Research, 1992.
[21] G. Legnani. “Robotica industriale”. Casa editrice Ambrosiana. 2003.
[22] P. Davoli, M. Filippini, C. Goria, A. Lo Conte. “Lezioni sugli organi delle
macchine”. Maggioli Editore. 2011.
[23] M. Giglio, M. Gobbi, S. Miccoli, M. Sangirardi. “Costruzione di macchine”.
McGraw-Hill, 2010.