manual presion de vapor

57
GS-100 0 plus VOC Gasoline Analyzer Operating Manual P/N MAN GS-1000 Plus VOC Version 2 300 Bammel Westfield Road • Houston, Texas 77090 • USA Telephone: (281) 580-0339 • (800) 444-8378 • FAX: (281) 580-0719 e-mail: [email protected] • web site: http://www.paclp.com

Upload: ivone

Post on 16-Apr-2015

379 views

Category:

Documents


62 download

DESCRIPTION

Petroespec

TRANSCRIPT

Page 1: manual presion de vapor

GS-1000 plus VOCGasoline Analyzer

Operating ManualP/N MAN GS-1000 Plus VOC Version 2

300 Bammel Westfield Road • Houston, Texas 77090 • USATelephone: (281) 580-0339 • (800) 444-8378 • FAX: (281) 580-0719

e-mail: [email protected] • web site: http://www.paclp.com

Page 2: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

i i

THIS MANUAL CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION…

…which is the property of PAC. This manual and all information disclosed herein shall not be used to manufacture,

construct, or fabricate the goods disclosed herein; shall not be exploited or sold; shall not be copied or otherwise

reproduced in whole or in part and shall not be revealed or disclosed to others or in any manner made public without

the express written permission of PAC.

REVISION HISTORY

Rev Description Release DateVer 1 First release –Ver 2 Updates to format and content April, 2004

Copyright ©2004, by PAC, Houston, Texas.Printed in the United States of America

All rights reserved.Contents of this publication may not be reproduced in any form

without permission of the copyright owner.04.01.04

Page 3: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

TOC-1

TABLE OF CONTENTS

SECTION 1INTRODUCTION ......................................................................................................... 1

General ...................................................................................................................... 1

Damage Claims .......................................................................................................... 1

Return Shipping ......................................................................................................... 1

Volatile Organic Compounds (VOC) ........................................................................... 1

SECTION 2GENERAL INFORMATION ........................................................................................... 3

FOR BEST RESULTS FROM YOUR GS-1000 FUEL ANALYZER ......................................... 3

Tips For Properly Maintaining Your PetroSpec GS Series Instrument ............................ 3

Factory Default Settings ............................................................................................. 5

Sample Care And Preparation ..................................................................................... 6Collecting And Storing A Retained Sample For Calibration Updating ............................... 6Collecting And Analyzing Sample For Updating Gs Series Instrument .............................. 6Updating Instruments .............................................................................................................. 6

Periodic Instrument Performance Validation by Control Charting ............................... 6

SECTION 3: PRODUCT DESCRIPTION......................................................................... 11

Product Features ...................................................................................................... 11

The Technology........................................................................................................ 11

Accessories Included ................................................................................................ 13

Parts and Accessories Available ................................................................................. 14

SECTION 4BASIC OPERATING INFORMATION ............................................................................ 15

Safety ....................................................................................................................... 15

Controls ................................................................................................................... 15

The Power Switch .................................................................................................... 15

The Display .............................................................................................................. 15

The Keypad .............................................................................................................. 15

The Power Connection ............................................................................................. 15

The RS-232 Communication Port .............................................................................. 15

The Parallel Printer Port ............................................................................................ 15

GS-1000 Front and Back Panels............................................................................... 16

Fixed Arm Sample Induction System......................................................................... 17

Filter Housing .......................................................................................................... 17

Operate/Clean Switch............................................................................................... 18

Cleaning Bottle Pressure Port ................................................................................... 18

Sample Input Port ..................................................................................................... 19

Sample Waste Port .................................................................................................... 20

Page 4: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

TOC-2

SECTION 5ANALYZING FUEL WITH THE GS-1000 PLUS VOC ...................................................... 21

Setting Up the GS-1000 ............................................................................................ 21

Sample Preparation .................................................................................................. 21

Analyzing Samples — Quick Start Using the AUTO Function .................................... 22

Results Screens ......................................................................................................... 24

Mahalanobis Distance And Extrapolated Results ........................................................ 25

SECTION 6 ANALYZER SET-UP FUNCTIONS................................................................ 27

1. PUMP ................................................................................................................... 27

2. PRINT .......................................................................................................................3. TIME ..................................................................................................................... 27

4. DEN (Density) ...................................................................................................... 28

5. VER (Version) ........................................................................................................ 28

6. VOC ..................................................................................................................... 28

7. Bias ...................................................................................................................... 29

8. UNIT .................................................................................................................... 30

Determining Weight Percent Oxygen...................................................................... 30

SECTION 7DATA MANAGEMENTFUNCTIONS ............................................................................................................. 31

Logging Analysis Results........................................................................................... 31

Entering Identification Codes ................................................................................... 31

Printing Data ........................................................................................................... 31

Erasing Data............................................................................................................. 31

Exporting Data ......................................................................................................... 32

SECTION 8USER UPDATEABLECALIBRATION .......................................................................................................... 34

SECTION 9PRODUCT MAINTENANCE ........................................................................................ 35

Cleaning the GS-1000 .............................................................................................. 35

Sample Induction Cleaning Procedure ..................................................................... 35

Replacing Fuel Filters ............................................................................................... 40

Page 5: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

TOC-3

APPENDIX 1MESSAGES ............................................................................................................... 41

Extrapolated Result ................................................................................................... 41

Printer Fault Detected, No Connection Or Paper Feed Error..................................... 41

Hydrocarbons Detected, Please Power Down ........................................................... 41

Chopper Wheel Failure Detected .............................................................................. 41

Filter Wheel Failure Detected ................................................................................... 41

Source Failure Detected ........................................................................................... 41

APPENDIX 2GASOLINE SAFETY INFORMATION ............................................................................ 45

Personal Protection Information ............................................................................... 45

Handling And Storage Precautions .......................................................................... 45

Reactivity Data.......................................................................................................... 45

Health Hazard Data .................................................................................................. 45Acute Effects of Overexposure ............................................................................................. 45Subchronic and Chronic Effects of Overexposure ............................................................. 45

First Aid And Emergency Procedures ....................................................................... 46

Fire And Explosion Data ........................................................................................... 46

Spill, Leak, And Disposal Procedures......................................................................... 46

Hazard Classification................................................................................................. 46

APPENDIX 3WARRANTY INFORMATION ...................................................................................... 47

PETROSPEC WARRANTY REGISTRATION FORM ......................................................... 49

PetroSpec Warranty Registration Form...................................................................... 49

Page 6: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

TOC-4

THIS PAGE INTENTIONALLY BLANK.

Page 7: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

1

SECTION 1INTRODUCTION

General

Thank you for selecting the PetroSpec GS-1000 Gasoline Analyzer for your fuel analysis needs. Wehave designed the instrument to be accurate, durable, and easy to use, and we are confident that ourproduct will perform to your satisfaction. Furthermore, we guarantee our product to be free of defectsin materials and workmanship for a period of one year.

This manual describes the correct operating procedure for the GS-1000 Gasoline Analyzer. The GS-1000 is a product of PetroSpec, a subsidiary of PAC.

Damage Claims

Thoroughly examine your GS-1000 and all accompanying accessories as soon as they are received.

NOTE:NOTE:NOTE:NOTE:NOTE: Insist that the carrier’s agent verify the inspection and sign the description.

Immediately notify the delivering agent of damage or loss. This notification may be given either inperson or by telephone. Written confirmation must be mailed within 48 hours. Railroads and motorcarriers are understandably reluctant to make adjustments for damaged merchandise unless it isinspected and reported promptly.

For technical support and customer service, please call 1-800-444-8378.

Risk of loss of, or damage to, merchandise remains with the buyer. It is the buyer’s responsibility tofile a claim with the carrier involved.

Contact PAC Customer Service immediately so that we may assist you.

Return Shipping

All return shipments of PetroSpec products must be prepaid. If the product is damaged, return it in theoriginal shipping container. PetroSpec will accept no liability for damage caused by improperpackaging. Address all returns to: PAC, 300 Bammel Westfield Road, Houston, TX 77090.

Please call PAC in advance, or enclose a detailed letter stating why you are returning your analyzer.

Volatile Organic Compounds (VOC)

Gasoline vapors and vehicle exhaust contain volatile organic compounds that react in the atmospherein the presence of sunlight and heat to produce ozone, a major componenet of smog. Through the 1990amendments to the Clean Air Act (CAA), Congress mandated that the Environmental ProtectionAgency (EPA) is responsible to regulate gasoline sold in certain areas to reduce vehicle emissions ofsuch compounds. Thus the EPA developed the complex model (defined by EPA 40 CFR Ch. 1 [7-1-96Edition] § 80.45 Complex emissions model), a method by which VOC content can be measured, andreduction monitored.

The GS1000 Plus VOC uses the complex model to calculate the total VOC (mg/mile) based on theoperator entered values of RVP (psig) and sulfur content (ppm), along with predicted values of oxygen(wt.%), aromatic (vol.%), and olefin content (vol.%), as well as E200 (%) and E300 (%) predictions.The total VOC value is compared to a specified baseline gasoline (defined by region, season, and phaseof the fuel being tested) to determine the percent reduction of the fuel sample.

Page 8: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

2

THIS PAGE INTENTIONALLY BLANK.

Page 9: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

3

SECTION 2GENERAL INFORMATION

WARNING! WARNING! WARNING! WARNING! WARNING! READ ALL SAFETY AND OPERATING INSTRUCTIONS BEFOREOPERATING

This product must not be operated near the highly combustible environment of a fuel storage tank, vent,fill opening or dispensing systems.

Fuel samples for analysis must be stored and handled in a closed container. Gasoline and gasolinevapors can create serious environmental, health, and fire hazards. It is the buyer’s responsibility tooperate this product according to the instructions found in this manual.

For Best Results From Your Gs-1000 Fuel Analyzer

Caution:Caution:Caution:Caution:Caution: Do not attempt to use a cracked, chipped, damaged, or incompatible fuel samplecontainer with the GS-1000 fuel analyzer. The GS-1000 pumps gasoline through the use ofcompressed air — use of such containers may risk damage to the fuel delivery system orcause fuel to spill.

1. Tighten the sample bottle securely to the sample input port.

2. Replace the fuel filter every 50-75 fuel samples, and make sure that the fuel filter has not slippedduring installation and that the fuel filter cap is tight.

3. Keep all tube fittings in good condition and replace any kinked tubing.

4. Always keep fuel samples covered. Evaporation will change the composition of the fuel andmay affect the measured results.

5. The instrument will give best results after a thirty-minute warm-up time.

6. Operate the instrument at temperatures between 60° and 90° F (15° and 32° C).

7. Operate the instrument on a secure, stationary surface.

Tips For Properly Maintaining Your PetroSpec GS SeriesInstrument

1. When instrument is received, turn it on, warm it up for 30 minutes, and check the performancewith a check standard. Check standards that are traceable to the factory validation of theinstrument can be purchased from PAC. BTGS-17011 Check Standard with Ethanol, or BTGS-17022 Check Standard with MTBE. Call for pricing and availability. PetroSpec recommendsthe users of this instrument periodically validate the performance of the instrument byconstructing a control chart using a check standard. The procedure for developing a controlchart is presented on page 10.

2. Be sure that the sample bottles are in good condition. Cracks could cause a premature pressurerelease that may invalidate the measurement. Chips on the top lip of the bottle not only cancause a loss of pressure, but they also can damage the gasket on the sample input port.

3. Never leave fuel in the instrument for an extended period of time. Always pump the fuel outof analyzer by running a sample with an empty bottle. A good practice to use is to flush thesystem with petroleum ether and then pump the system dry after the day’s use.

4. Change the fuel filter often. A dirty filter can diminish the flow of the instrument causinginvalid results. If the fuel filter does not seat correctly sediment can bypass the filter and clogthe sample cell.

Page 10: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

4

5. If there is a noticeable amount of particles in the fuel sample, pre-filtering the sample usinga 20-micron or less pore size filter paper may be in order.

6. Be sure that the filter housing does not leak. Fuel stains on the left side panel are an indicationthat the housing is leaking. A leak not only causes unsightly stains but the leak could also causeinvalid measurements.

7. A minimum flow rate of three milliliters is recommended for proper operation of theinstrument. To check for proper flow, first prime the instrument, (run a sample into the wastebottle) then, using a graduated cylinder as the waste vessel, run another sample. The pump timeis set at the factory for fifty seconds. If the instrument does not pump a minimum of threemilliliters in this time then the pump can be increased.

NOTE:NOTE:NOTE:NOTE:NOTE: If 3ml or more of sample flow cannot be obtained in ninety seconds or less, it mayindicate that the instrument needs service.

8. Always use clean bottles to introduce the sample into the instrument. Cross contaminationcould occur if the bottles are not clean. Sediment in the bottle can result in a clogged system.

9. Screw the red fuel fittings into the sample input port firmly and be sure not to over-tighten.A fitting that is too tight could restrict the fuel flow. Use care in tightening the fitting – crossthreading could result in a fuel and or pressure loss that could invalidate the measurement.If the threads of the RED fitting become stripped or the tubing gets kinked, please callPetroSpec. We offer replacement tubing kits: Model Number BTGS-1402 for Coaxial SideArm: Model Number BTGS-1403 for Boston Round Side Arm.

10. If you have any problems or questions please do not hesitate to call us at 281-580-0339.

Page 11: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

5

Factory Default Settings

Your GS-1000 arrives with the following factory preset defaults.

Pump Time ----------------------------------------------------------------------------- 50 Seconds

Print Format ----------------------------------------------------------------------------- Table Format

Time ----------------------------------------------------------------------------- Eastern Standard Time

Density

MeOH ------------------------------------------------------------------- .796

EtOH --------------------------------------------------------------------- .794

TBA ---------------------------------------------------------------------- .792

MTBE -------------------------------------------------------------------- .746

DIPE --------------------------------------------------------------------- .730

ETBE --------------------------------------------------------------------- .745

TAME ------------------------------------------------------------------- .755

GAS ---------------------------------------------------------------------- .742

Version ----------------------------------------------------------------------------- Revision Dependent

Demo Mode ----------------------------------------------------------------------- Off

Bias

Aromatics -------------------------------------------------------------- 0

Benzene ----------------------------------------------------------------- 0

MTBE -------------------------------------------------------------------- 0

MeOH ------------------------------------------------------------------- 0

TAME ------------------------------------------------------------------- 0

ETBE --------------------------------------------------------------------- 0

EtOH --------------------------------------------------------------------- 0

DIPE --------------------------------------------------------------------- 0

TBA ---------------------------------------------------------------------- 0

Research Octane Number --------------------------------------- 0

Motor Octane Number-------------------------------------------- 0

VOC

E200 ---------------------------------------------------------------------- 0

E300 ---------------------------------------------------------------------- 0

Volatility

T50 ----------------------------------------------------------------------- 0

T90 ----------------------------------------------------------------------- 0

Temperature ---------------------------------------------------------- Location Dependent

Unit ----------------------------------------------------------------------------- Vol%

Total Oxygen --------------------------------------------------------------------- Wt%

NOTE:NOTE:NOTE:NOTE:NOTE: The Total Oxygen values are expressed in Wt% only and cannot be changed.

Page 12: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

6

Sample Care And Preparation

Collecting And Storing A Retained Sample For Calibration Updating

1. Samples should be stored in an amber bottle with a cap that has a conical insert.

2. Bottles must be tightly capped.

3. Store retained samples in an explosion proof refrigerator at 5 degrees Celsius.

4. Opening of the samples should be avoided. The lighter components of the fuel escape everytime the cap is opened. This will change the composition of the fuel.

Collecting And Analyzing Sample For Updating Gs Series Instrument

1. Draw a single sample from the tank or pipeline.

2. Split the sample into appropriately sized containers. Retain a portion of the sample to be usedlater for updating. Submit the remaining sample for laboratory analysis. The samples must bepackaged and cared for in accordance with instructions presented in the previous section.

3. The laboratory should analyze the samples immediately to avoid weathering.

Updating Instruments

1. Bring the sample to room temperature.

2. Make sure the fuel filter is clean. Replace the filter if it is tan to brown in color.

3. Set up the instrument as described on page 22.

4. Run a check sample in the GS-1000 before adding fuels. The results for the check sample mustfall within the limits established by the instrument control chart (page 8). Repeat the analysisof check standard if the sample results fall outside the control limits. If the results for the checksample continue to fall outside the control limits call PetroSpec Customer Service. It is veryimportant to make sure the unit is operating properly before adding fuel.

5. Laboratory results must be available before the GS-1000 can be updated. The outside analysisand sample data will be used to update the instrument.

6. Use the Calibration Update Software to download the absorbance data from the samples thatare to be used for the calibration update. Refer to the Calibration Update Software Manual formore detailed instructions.

Periodic Instrument Performance Validation by ControlCharting

It is important to periodically monitor the performance of the GS-1000 and to demonstrate that thecurrent performance is statistically consistent with the historical performance of the instrument.Constructing a control chart is the best method for validating the performance of the instrument. Thissection will describe the procedure for constructing an individual differences control chart.

The control chart is used to plot the difference between the stated property value of a check standardand the property value determined by the instrument. Since the control chart is plotting the differencebetween the expected value and the predicted value, it is not necessary to begin a new control chartif the expected value for the check standard changes. For example, if the new check standard is froma different batch.

Benzene is the property used for the example described in this section. The data required forconstructing the control chart for benzene is presented in Table 1.

Page 13: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

7

TABLE 1. BENZENE CONCENTRATIONS, IN VOLUME PERCENT

FROM A PETROSPEC CHECK SAMPLE

Date Sample (n) Expected Predicted Diff. (dn) MR (|dn+1-dn|)

05/21/98 1 2.19 2.19 0.00 0.0005/28/98 2 2.19 2.21 -0.02 0.0206/03/98 3 2.19 2.19 0.00 0.0206/15/98 4 2.19 2.20 -0.01 0.0106/12/98 5 2.19 2.22 -0.03 0.0206/17/98 6 2.19 2.18 0.01 0.0406/24/98 7 2.19 2.22 -0.03 0.0407/01/98 8 2.19 2.17 0.02 0.0507/12/98 9 2.19 2.19 0.00 0.0207/19/98 10 2.19 2.20 -0.01 0.0107/31/98 11 2.19 2.24 -0.05 0.0408/08/98 12 2.19 2.20 -0.01 0.0408/15/98 13 2.19 2.23 -0.04 0.0308/22/98 14 2.19 2.23 -0.04 0.0008/30/98 15 2.19 2.23 -0.04 0.0009/06/98 16 2.19 2.22 -0.03 0.0109/13/98 17 2.19 2.16 0.03 0.0609/20/98 18 2.19 2.16 0.03 0.0009/27/98 19 2.19 2.16 0.03 0.0010/02/98 20 2.19 2.18 0.01 0.0210/09/98 21 2.19 2.16 0.03 0.0210/16/98 22 2.19 2.18 0.01 0.0210/23/98 23 2.19 2.17 0.02 0.0111/01/98 24 2.19 2.19 0.00 0.02

The expected value is the certified benzene concentration of the check sample, the predicted value isthe benzene concentration determined using the instrument, and the difference, dn, is the differencebetween the expected value and the predicted value for sample n. The value MRn is calculated by takingthe absolute value of the difference between dn and dn-1

Examples:

For the sample number 2 analyzed on 5/28/98,

d2 = 2.19 – 2.21 = -0.02and

MR2 = |d2 – d1| = | -0.02 – 0.00| = 0.02.

For the sample number 6 analyzed on 6/17/98,

d6 = 2.19 – 2.18 = 0.01and

MR6 = |d6 – d5| = |0.01 – (-0.03)| = 0.04.

The upper control limit (UCL) and lower control limit (LCL) are calculated using the equations:

UCL = dave + 2.66*MRave

LCL = dave - 2.66* MRave

Page 14: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

8

The value dave is the running average of the differences dn. For the data listed in Table 1, dave is calculatedby summing the values for d1 through d24 and dividing the sum by 24 yielding the value –0.005. Thevalue MRave is the running average of the absolute value of the difference between dn and dn-1. In thisexample, MRave is calculated by summing the values for MR1 through MR24 and dividing the sum by24 yielding the value 0.022. The values for UCL and LCL are 0.053 and –0.063, respectively.

Examples:

For the sample number 2 analyzed on 5/28/98,

UCL = [(0.00+(-0.02))/2]+2.66*[(0.00+0.02)/2] = 0.0166and

LCL = [(0.00+(-0.02))/2]-2.66*[(0.00+0.02)/2] = -0.0366.

For the sample number 6 analyzed on 6/17/98,

UCL = [(0.00+(-0.02)+0.00+(-0.01)+(-0.03)+0.01)]+2.66*[(0.00+0.02+0.02+0.01+0.02 + 0.04)/6] = 0.0405

andLCL = [(0.00+(-0.02)+0.00+(-0.01)+(-0.03)+0.01)/6]-2.66*[(0.00+0.02+0.02+0.01+0.02 + 0.04)/6] = -0.0571.

The graph in Figure 1 is a graphical representation of the data summarized in Table 1 and is called acontrol chart, specifically the Individual Differences control chart. The graph is a plot of dn versus thesample number, n. Each time the control sample is analyzed the difference dn is calculated. If thedifference value, dn, falls within the upper and lower control limits, recalculate the UCL, and LCL, andplot the new d-value.

Figure 1. Control chart for benzene concentration determination

Benzene Control Chart

-0.10

-0.08

-0.06

-0.04

-0.02

0.00

0.02

0.04

0.06

0.08

0.10

0 5 10 15 20 25 30

Sample Number

d

UCL

LCL

Page 15: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

9

If the difference, dn, falls outside the upper or lower control limit, the instrument and/or the checksample appears to be unstable, and efforts should be made to determine the cause. Check the fuel filter,and the hose connections to ensure the sample is being properly introduced into the instrument, thenre-analyze the check sample. If the result continues to fall outside the control limits contact PACCustomer Service.

When adding a check sample result to the control chart, it is important not to bias the statistics by addingmore than one result for each check sample. The interval between check samples is variable but shouldnot be longer than one week or every 50 samples, whichever interval is shorter, during the initial periodof testing. As the trend of the instrument performance is established, the interval between checksamples can be increased, however the interval should not be more than one month. The more oftenthe instrument is used, the more often a check sample should be tested.

A control chart should be constructed for every parameter reported by the GS-1000 and recorded bythe user.

Check standards can be purchased directly from PAC. Model numbers BTGS-17011 for ethanol orBTGS-17022 for MTBE. They come in packs of 5 vials. Call for price and availability.

Page 16: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

10

THIS PAGE INTENTIONALLY BLANK.

Page 17: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

11

SECTION 3: PRODUCTDESCRIPTION

Product Features

The GS-1000 Gasoline Analyzer provides a fast, accurate analysis of fuel composition in the field orlaboratory. In one simple, two-minute analysis process, the instrument reports the following:

Component Calibration Range

Research Octane Number (RON) 90-101Motor Octane Number (MON) 80-90Pump Octane Number (R+M/2) CalculatedOxygenates:MTBE 0-20 vol %ETBE 0-24 vol %Ethanol 0-15 vol %Methanol 0-8 vol %TAME 0-24 vol %DIPE 0-20 vol %TBA 0-15 vol %Total Oxygen CalculatedBenzene 0-5 vol %Total Aromatics 0-60 vol %Olefins 0-35 vol %Saturates CalculatedToluene 0-25 vol %Xylene 0-15 vol %T50 150-265 FT90 265-360 FE200 30-70%E300 75-95%Total VOC CalculatedVOC Redirection Calculated

Note:Note:Note:Note:Note: The calibration ranges stated above are the ranges found in the standard “factory”calibration. Certain ranges can be expanded higher and lower via calibration update. Pleasecall PAC at 281-580-0339 for more information.

The Technology

The GS-1000 Multi-Function Analyzer uses a mid-infrared spectroscopic analysis technique todifferentiate and quantify the individual components in a fuel sample. Every fuel component has aunique spectral absorbance pattern that contains characteristic absorbance bands. The GS-1000utilizes a proprietary set of optical band-pass filters to measure the amount of energy that is absorbedby a fuel sample at these identifying bands. The amount of light absorbed is proportional to theconcentration of that component in the fuel sample. The instrument predicts the property value of afuel sample using a model that relates the mid-infrared absorbance of the sample to the property valuefor the sample. The model is developed using the concentrations and inter-relations of the manycomponents in a set of calibration samples having defined property values.

The GS-1000 utilizes a thermoelectric temperature control system. The temperature control systemrequires an external thermal management system, which consists of a thermal bus, a fan, a thermalshield, and a heat sink. The thermal bus is mounted on the bottom of the instrument and transfersthermal energy to or from the optical system to the heat sink, which is located on the right side of theinstrument. The thermal shield covers the thermal bus on the bottom of the instrument.

Page 18: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

12

The thermal electric temperature control system may need to heat or cool the GS optical system in orderto maintain the optimum internal temperature. This heating or cooling mode is dependent on theexternal temperature. The instrument must obtain thermal equilibrium before accurateanalyses can be made, therefore, a warm up period of 30 minutes (minimum) is requiredbefore running any fuels.

The temperature control system can maintain optimum temperature within the GS-1000 whenexternal temperature is between 60 and 90 degrees F. In order to maintain this operating range, theremust be free air circulation around the thermal bus and the heat sink. It is important to keep theimmediate area around the heat sink clear of obstructions. It is equally important that the exhaust fanvents be kept clear and unobstructed.

Product Specifications

Sample Size 5 ml minimumData Output 2 lines x 40 character LCDRS-232 serial portParallel printer portDimensions 10.5W x 11.0D x 12.5H (inches)Weight 26 poundsOperating Temperature Range 60 F - 90 FPower Requirements 120/240 VAC 50/60 Hz

Property Repeatability Reproducibility Test Method

RON .0022*(X + 2.0) .013*(X + 2) Correlates to D2699MON .11 .64 Correlates to D2700R+M/2 .17 .81Total Aromatics .17*X^.25 .97*X^.25 Correlates to D1319Total Olefins .019*(X + 11.43) .076*(X + 14.89) Correlates to D1319Saturates .67 2.8Total Xylenes .1 .3*Benzene .021 + (.027*X) .121 + (.012*X) D6277*MTBE .13 .98 D5845*EtOH .13 .59 D5845*TAME .13 1.36 D5845*ETBE .15 .77 D5845*DIPE .14 .79 D5845*TBA .1 .59 D5845*MeOH .07 .37 D5845*Wt. % Oxygen .05 .3 D5845T50 Correlates to D86T90 Correlates to D86E200 Correlates to D86E300 Correlates to D86

*From ASTM Round Robin

Page 19: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

13

Accessories Included

Your GS-1000 is shipped with the following accessories:

• 120/240 VAC external power supply

• 4 large (8 oz.) sample bottles

• 4 small (4 oz.) sample bottles

• Calibration Update Manual

• Or Cleaning Cap

• 20 Teflon fuel filters

• Tubing kit

• User Calibration Software CD

• Instrument Database Diskette

• 1 Serial Cable

• GS-1000 Operating Manual

• Accessories Price List

• Certificate Of Compliance

WARNING: WARNING: WARNING: WARNING: WARNING: Only use the 120/240 VAC external power supply designated for the GS-1000series instruments.

For your convenience, the following accessories are available directly from PetroSpec:

• Sample bottles

• Replacement fuel filters

• Replacement tubing kits

• Air shippable carrying case

• Replacement 120/240 VAC external power supply

• Instrument Check Standards

The following page contains a complete list of GS accessories. For pricing and ordering information,please call PAC at 281-580-0339.

Page 20: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

14

Parts and Accessories Available

PETROSPEC ACCESSORIES AND PARTS

Description Model

Large 8 oz., Glass Bottles w/ Plastic Armor (Box of 5) BTGS-1301

Small 4 oz., Glass Bottles clear w/o Plastic Armor (Case of 24) BTGS-1320

Small 4 oz., French Square Bottles (Case of 24) BTGS-1325Fits Boston Round Cap

Glass Vials 20 ml., Clear w/o Plastic Armor (Case of 100) BTGS-1330Teflon Fuel Filters (Pack of 20) BTGS-1420

Tubing & Compression Fittings for Co-axial Cap (Set of 4) BTGS-1402

Tubing & Compression Fittings for Boston Round Cap (Set of 4) BTGS-1403

Power Supply & Cord for GS-360 & GS-500 BTGS-1501Power Supply & Cord for GS-1000 (120/240 VAC) BTGS-1240(Auto Switching for International Use)

Auto Cigarette Lighter Adapter (For use with GS-360 &GS-500) BTGS-1503Auto Cigarette Lighter Adapter (For use with GS-1000) BTGS-1504

Surge Protector BTGS-1510Soft-Sided Carrying Case BTGS-1702

Gasoline Analyzer Check Standard with Ethanol BTGS-17011Gasoline Analyzer Check Standard with MTBE BTGS-17022(Five 20 ml Vials) Suitable for GS-1000GS-360, GS-500, GS-1000PlusVOC series

Boston round bottle threaded ring and gasket for BTGS-BRSIAfixed arm assembly. Fits GPI 33-400 bottle thread

Page 21: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

15

SECTION 4BASIC OPERATING INFORMATION

Safety

The GS-1000 should not be operated near the highly combustible environment of an underground fuelstorage tank, vent, fill opening, or dispensing system. The GS-1000 contains a hydrocarbon sensor thatcan detect potentially explosive levels of gasoline vapors. If dangerous levels are detected, the GS-1000will automatically shut off all power to its infrared source and will alert the user to power down theinstrument.

Controls

The GS-1000 is designed to be easy to use. This section will familiarize you with the few buttons andconnections that will enable you to use the GS-1000 to its fullest capacity.

The Power Switch

The power switch turns the GS-1000 on and off. Upon powering up, the instrument completes aninitialization and self-diagnostic procedure that takes approximately five minutes. This period alsoallows the infrared filament source within the GS-1000 optical system to warm up. Although theinstrument can be used immediately after the initialization period, for best results it is recom-mended that a minimum warm-up period of 30 minutes be observed.

The Display

The GS-1000 features a 2 line x 40 character back-lit LCD display. All analysis results and menu choicesare shown on the display. When the displayed information does not fit onto one screen, the arrow keysare used to scroll through multiple screens of data.

The Keypad

The GS-1000 keypad features 12 numerical keys (left side) and 8 control keys (right side). Thenumerical keys are used primarily to enter menu choices. The control keys are used to verify menuselections, scroll through analysis results screens, and for data management.

The Power Connection

The power connection is located on the rear panel of the GS-1000. Power is supplied by the 120/240VAC external power supply shipped with the unit. Only use power supplies designated for use withthe GS-1000 or GS-1000 series of instruments.

The RS-232 Communication Port

The RS-232 serial port is located on the rear panel of the GS-1000. The RS-232 serial port allows theuser to download data from the GS-1000 to a computer. For more information see Section 7, DataManagement.

The Parallel Printer Port

The parallel printer port allows analysis reports to be printed directly from the GS-1000 to any parallelprinter compatible with the Centronics protocol.

Page 22: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

16

GS-1000 Front and Back Panels

Fixed Arm Sample Induction System

Page 23: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

17

Fixed Arm Sample Induction System

(Fixed Arm Features photo)

Filter Housing

The filter housing contains the fuel filter disc. To access the filter, simply unscrew the knurled screwand lift up on the hinged filter housing top plate. If the filter housing is dry the filter may become difficultto remove. To aid in the ease of filter removal PetroSpec recommends that a fuel sample be run on theinstrument prior to changing the fuel filter.

NOTE:NOTE:NOTE:NOTE:NOTE: Be sure that the sample bottle is disconnected from the sample input port whenreplacing the fuel filter. The pressurized fuel in the sample bottle will flow during fuel filterreplacement if the sample bottle is still attached.

Be sure to close the filter housing and tighten the knurled screw after the filter has been replaced andbefore running fuel samples.

Page 24: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

18

Operate/Clean Switch

The Fixed Arm Sample Induction System features a Operate / Clean toggle switch. The purpose of thetoggle switch is to allow the user to easily and effectively clean the sample induction system of theanalyzer. The switch must be in the up position for normal fuel sampling operations. With the switchin the up position the instrument will pressurize the fuel sample bottle to allow adequate fuel flow forthe test being performed. Placing the switch in the down position will redirect the pressure to the quickdisconnect fitting located on the left side panel of the instrument for cleaning operations. Please referto the cleaning instructions on page 35 for additional information on cleaning the instrument’s sampleinduction system.

Warning:Warning:Warning:Warning:Warning: Running fuel samples without the fuel filter being installed may result in aclogged Sample Induction System.

Cleaning Bottle Pressure Port

The pressure port is use during cleaning operations. Use the Cleaning Cap located in the accessory kitwhen cleaning operations are performed. The pressure hose must be connected to the pressure porton the side panel of the instrument during this operation. Please refer to the figure below for propercleaning set up.

Fuel analyzer set up for cleaning operation

Page 25: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

19

Sample Input Port

Fuel samples are introduced into the analyzer via the Sample Input Port. There are two styles of SampleInput Ports available from PetroSpec.

The most common option is the Coaxial Sample Input Port. The Coaxial Sample Input Port is designedto accommodate sample bottles with GPI thread sizes 22-400 or 45-400. The second option is theBoston Round Sample Input Port. This port accommodates bottles with a GPI thread size of 33-400.

Please refer to the chart below for the Sample Input Port Options and the mating sample bottles thatare available from PetroSpec.

Sample Port Type Thread Size Sample Bottles Available Part Number

Coaxial Sample Input Port 22-400 20 ml vial BTGS-1330 (Case of 100)4 oz clear bottle BTGS-1320 (Case of 24)8 oz amber bottle with armor BTGS-1301 (Case of 5)

Boston Round Sample Input Port 33-400 4 oz French Square bottle BTGS-1325 (Case of 24)

For best results, fill the sample bottle at least ¾ full of gasoline sample. Be sure to use the proper diptube for the sample bottle selected. Screw the sample bottle firmly onto the Sample Input Port but toavoid breakage do not over-tighten.

Be sure to select the correct length dip tube to match the chosen sample bottle.

Page 26: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

20

Sample Waste Port

Connect the amber waste bottle to the Sample Waste Port. To avoid bottle breakage, do not over-tighten.

Standard sample run set-up

Page 27: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

21

SECTION 5ANALYZING FUEL

WITH THE GS-1000 PLUS VOC

Setting Up the GS-1000

The GS-1000 is factory calibrated and is ready to use. Power is supplied through the 120/240VACexternal power supply. Plug the power supply into the power socket on the rear of the GS-1000 andturn the power switch to the ON position. The GS-1000 will perform a five-minute self-test andinitialization procedure, followed by a 25-minute warm-up period in which the screen will be flashing.After the warm-up period the main menu will stop flashing.

During the initialization/warm-up period, the condition of the fuel filter should be checked andreplaced if necessary (See Section 9: Product Maintenance).

NOTE:NOTE:NOTE:NOTE:NOTE: Best results will be obtained if the analyzer is allowed to warm up for at least 30minutes. If the instrument is momentarily shut down after the 25-minute warm-up hastaken place due to a fuel spill or any other reason then the 25 minute warm-up period canbe overridden simply by pressing the (1) AUTO KEY.

Sample Preparation

All fuel samples for analysis must be in a container that is compatible with the GS-1000 fuel deliverysystem. There are two Fixed Arm Induction options. The most common option is the coaxial FixedArm Induction Port. The Coaxial Fixed Arm Induction Port is designed to accommodate samplebottles with GPI thread sizes 22-400 or 45-400. The second option is the Boston Round Fixed ArmInduction Port. The Boston Round Fixed Arm Induction Port accommodates bottles with a GPI threadsize of 33-400. A minimum sample size of 5 ml is required for analysis. However, best results areobtained with relatively full sample containers.

Standard sample run set up.

Page 28: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

22

Analyzing Samples — Quick Start Using the AUTO Function

Verify that the sample and waste fuel bottles are properly connected. Attach the appropriate sampleinlet tube, based on the size of your sample container, to the Sample Input Port (see figure on page 17).Screw sample bottle securely onto the Sample Input Port. The sample bottle must be snug but use carenot to over-tighten in order to prevent damage to the gasket, sample bottle, or both.

NOTE: Be sure that the Clean / Operate switch is in the Up position prior to running fuelsamples.

The GS-1000 plus VOC Main Menu

At the main menu, press AUTO (1). The GS-1000 will prompt the user for an expected R+M/2 value.If desired, the user may enter the rated R+M/2 value. This will be logged and printed along with theanalysis results for the purpose of comparison or identification of the sample. The entered R+M/2value has no effect on the results of the analysis. After entering the expected R+M/2 value (or EXITif the user chooses not to do so) analysis cycle will begin, which consists of approximately one minuteof pumping fuel followed by two minutes of analysis. The GS-1000 will automatically begin a completeanalysis cycle, which includes the determination of all oxygenates, benzene, total aromatics, toluene,xylene, total olefins concentrations; as well as the prediction of the octane, and distillation points (T50,T90). The weight percent oxygen, the percent of saturates, and the R+M/2 octane values aredetermined by calculation.

Upon completion, analysis results will be displayed on the screen. Use the arrow keys or the ENTERkey to scroll from one screen of analysis results to the next.

Page 29: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

23

Menu FlowchartAuto Function

THIS PAGE INTENTIONALLY BLANK.

Page 30: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

24

Results Screens

Pressing ENTER on the eighth screen will display all results again starting from the first screen. PressEXIT while on the eighth screen to exit the analysis results screen. The GS-1000 will allow the operatorto choose whether to log the results. Press NO to run another analysis without logging the currentresults. For information on logging results, see Section 7: Data Management.

Note: Be sure to pump the instrument dry after the day’s use.

Page 31: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

25

Mahalanobis Distance And Extrapolated Results

The GS-1000 uses a statistical value called the Mahalanobis distance to determine whether thepredicted property values obtained from a fuel sample result from an extrapolation of the model. Theextrapolation of the model occurs when the infrared spectrum of the fuel sample is not included withinthe spectral space spanned by the calibration fuels used to construct the model. When the Mahalanobisdistance detects that the predicted property value of the fuel sample is likely the result of anextrapolation, the instrument warns the user of the extrapolated result along with the Mahalanobisdistance number, which is displayed as M=**.*. The Mahalanobis distance warning is triggered whenthe distance value exceeds 26.3. If the absorbance values from the calibration set are normallydistributed, then the Mahalanobis distance follows the chi-squared distribution. For 16 degrees offreedom (17 filters minus one), a sample with a Mahalanobis distance of 26.3 is deemed to be outsidethe calibration set with 95% confidence. Depending on the magnitude of the Mahalanobis distance,one or more samples similar in composition should be added to the calibration set. The addition ofone sample may not eliminate the extrapolation. An outlier with a large Mahalanobis distance (>50)may not be made an inlier no matter how many samples of similar composition are added to thecalibration set. For a sample with a large Mahalanobis distance, a separate calibration set may needto be constructed. The ability of the analyzer to predict the properties of a specific fuel may improveif the sample is added to the calibration set.

See User Calibration Section 8 page 33. For detailed information, refer to the Calibration Update Manual.

Page 32: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

26

THIS PAGE INTENTIONALLY BLANK.

Page 33: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

27

SECTION 6ANALYZER SET-UP FUNCTIONS

The set-up functions allow the user to modify the GS-1000’s basic operating parameters. The correctuse of these functions allows the user to customize the operation, or modify the results of the instrument.However, improper use of these functions may negatively impact the accuracy of the GS-1000.Therefore, only qualified, trained personnel should change any of these settings.

To access the set-up functions, press SETUP (4) from the GS-1000 main menu.

The set-up menu will appear:

1. PUMP

The PUMP function allows the user to set the length of time that the fuel pump operates before analysisbegins. The GS-1000 should pump at least 3ml of fuel through the system within the selected pump-time. This sample volume will displace the previous sample. A pump-time of less than 50 seconds isnot recommended. If 50 seconds is not an adequate pump time, try replacing the fuel filter and checkingthe fuel lines for kinks before adjusting the pump time. For more information, see Section 9: ProductMaintenance.

To adjust the pump time, press PUMP (1) from the set-up menu. Enter the new pump time and pressENTER.

2. PRINT

The PRINT Set-Up function sets the format of printed analysis results. You can select either TABLE(1) or REPORT (2) format. The table format lists the sample identification code, the time and date theanalysis was run, and all analysis results. The report format prints the same information but leavesroom for a description of the sample, results from running a calibration standard, and a place for theoperator’s signature.

3. TIME

PUMP PRINT TIME DEN VER VOC BIAS UNIT(1) (2) (3) (4) (5) (6) (7) (8)

The TIME function allows the user to change the time and date. Select TIME (3) and enter the newtime and date.

Page 34: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

28

4. DEN (Density)

The GS-1000 calculates total weight percent oxygen from volume percent using the EPA methoddescribed on page 64 of the Supplemental Notice of Proposed Guidance on Establishment of ControlPeriods under Section 211 (m) of the Clean Air Act as amended and released in the first quarter of 1992.

The specific densities used in the calculation are also from this method:

Oxygenate Weight percent Oxygen Density

MTBE 18.15 0.746EtOH 34.73 0.794TAME 15.66 0.755ETBE 15.66 0.745MeOH 49.93 0.796DIPE 15.66 0.730TBA 21.59 0.792GAS ——- 0.742

For greater control over the accuracy of the total weight percent oxygen measurement, you may enteryour own values for the specific density of each oxygenate component and of the base gasoline. To dothis, select 4 (DEN) choose the desired component or gasoline, type the new three digit density value,and press ENTER. The GS-1000 always displays the EPA designated value in parentheses when thescreen is accessed.

5. VER (Version)

Selecting the version function will enable the operator to see the current firm ware version of theinstrument.

The version function also allows the user to put the GS-1000 into the demonstration mode. When theinstrument is in the DEMO mode it goes through all the motions of an analysis but only displaysdemonstrated results. The letter “d” will be displayed on the LCD while the instrument is in this mode.

Note: Note: Note: Note: Note: Be sure to disable the DEMO mode once the demonstration is completed.

6. VOC

The VOC set-up function allows the operator to disable or enable the VOC mode, set the VOC region,and set the VOC season.

Page 35: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

29

7. BIAS

The BIAS function allows the user to add or subtract an offset to the final oxygenate, xylene, toluene,total aromatic, benzene, olefin, octane, T50, or T90 analysis result. The BIAS sub-menu allows theuser to select which analysis (oxygenate, aromatic, octane, evaporation point, or VOC) is to be offset.The user may then increment, decrement, or clear existing offsets. When an oxygenate bias is entered,the letter “x” will appear on the display in the main menu. When an octane bias is entered, the letter“e” (for “engine”) will appear on the display in the main menu. Please refer to the chart below for otherletter designations and their respective biasing.

Measured Property Displayed Maximum IncrementLetter Bias (offset)

Demo mode “d” N/A N/AOxygen bias “x” +-0.5 +-0.1Xylenes bias “a” +-3 +-.1Toluene “a” +-3 +-.1Aromatic or Olefin bias “a” +-3 +-0.1Benzene “a” +-0.3 +-0.01Octane bias “e” +- 3 +-0.1VOC “v” +-5 +-1

Oxygenate biases can be set in increments of 0.1, with a maximum bias of plus or minus 0.5. Aromaticand olefin biases can be set in increments of 0.1, to a maximum of plus or minus 3.0. The benzene biascan be set in increments of 0.01, with a maximum bias of plus or minus 0.3. The RON and MON octanebiases can be set in increments of 0.1 with a maximum bias of plus or minus 3.0. The offsets are storedin protected memory and are not lost if the instrument is turned off.

Page 36: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

30

8. UNIT

The UNIT function allows the user to select the format in which all oxygenate analysis results aredisplayed. Results can be displayed in either WEIGHT PERCENT or VOLUME PERCENT. The weightpercent values are calculated using the density values selected in the DEN (4) set-up function. (seeabove) Total oxygen is calculated as wt% only and cannot be changed. Total aromatics, total olefins,and benzene are reported as vol. % and can not be changed. The UNIT function also allows the userto select between displaying the T50 and T90 predictions in degrees Fahrenheit or degrees Celsius.

Note: Note: Note: Note: Note: GS-1000 firmware versions 2.71 and above are programmed to display the T50 andT90 temperature values in Celsius ororororor Fahrenheit and cannot be selected in set-up.

Determining Weight Percent Oxygen

Sample weight percent oxygen is calculated using the EPA method described on page 64 of the“Supplemental Notice of Proposed Guidance on Establishment of Control Periods under Section211(m) of the Clean Air Act as Amended” released in the first quarter of 1992. This method is outlinedbelow.

The following equation describes the conversion from volume percent to weight percent oxygen ina gasoline blend.

Σ(vol%oxy * doxy * wt%oxy)wt%O=

0.742*(100 - Σ(vol%oxy)) * doxy)

where:

vol%oxy =measured volume % of each oxygenate in the fuel doxy =density of each oxygenate present: from table below wt%oxy =weight % oxygen in each oxygen present: from table below wt%O =weight % oxygen in each oxygen in the sample

FACTORS FOR CALCULATING WT% OXYGEN

Component wt%O in oxygenate Density

MTBE 18.15 0.746ETOH 34.73 0.794TAME 15.66 0.775ETBE 15.66 0.745MEOH 49.93 0.796DIPE 15.66 0.73TBA 21.59 0.792Gas 0.742

Page 37: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

31

SECTION 7DATA MANAGEMENT

FUNCTIONS

Logging Analysis Results

The GS-1000 can store up to 99 analysis results in internal memory. After analysis results aredisplayed, press EXIT and select YES, when prompted, to access the log entry function. The operatorwill then be prompted to enter an identification code under which the analysis results will be filed.

Entering Identification Codes

Identification codes may be up to eight (8) characters long and can consist of letters and/or numbers.Numbers can be entered using the numerical keys on the keypad. Letter characters can be selected usingthe following procedure:

• Press the up arrow key. The alphabet will appear on the top line of the display.

• Press the right and left arrow keys to highlight the desired letter

• Press ENTER to select the desired letter.

• Press the down arrow key.

Printing Data

Pressing PRINT (2) from the main menu will access the print functions. The GS-1000 can print eitherto an external printer, or to the display screen. When printing to an external printer, the printer mustbe connected to the instrument before the print function is selected, or an error message will appear.Be sure to remove power to the GS-1000 and the printer prior to making any connections. If poweris not removed, the printer, GS instrument or both can be damaged.

Upon selecting PRINT (2) from the main menu, the operator will have four choices:

1. PRINT LAST - will print the results of the last analysis performed, even if that analysis wasnot saved to the log.

2. PRINT LOG - will print the entire contents of the log file.

3. PRINT INDIVIDUAL - will print an individual analysis entry stored in the log. Select thedesired entry by entering the identification code (see above) or by using the up and downarrows to scroll through a list of all of the stored entries.

All print functions will print the data in the format selected in the set-up menu. For information onchanging the print format, see Section 6: page27.

Erasing Data

The ERASE function permanently deletes entries from the log file. This function is accessed bypressing ERASE (3) from the GS-1000 main menu. Once selected, you will have two choices:

ERASE INDIVIDUAL - will erase an individual analysis entry stored in the log. Select the desiredentry by entering the identification code (see above) or by using the up and down arrows to scrollthrough a list of all of the stored entries.

ERASE ALL - will erase the entire contents of the log file.

Page 38: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

32

Exporting Data

Analysis data in the log file may be downloaded as an ASCII text file from the GS-1000 to a computerusing the EXPORT function, which is accessed by pressing EXPORT (5) from the main menu.

Data is exported via the 9 pin RS-232 serial port located on the rear panel of the GS-1000. The portmust be connected to an RS-232 serial port on an IBM compatible computer that is running anystandard communication software. The host computer’s protocol must be set to the following:

• Baud rate = 9600 bits per second

• Data bits = 8

• Stop bits = 1

• Parity = none

• Flow control = Xon / Xoff

When the computer is ready to accept the data file, press EXPORT LOG (1) on the GS-1000.

Page 39: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

33

SECTION 8USER UPDATEABLE

CALIBRATION

The Calibration Update Software allows the operator to add fuels to their analyzer’s calibration matrixwithout the need to return the analyzer to PetroSpec. Your GS-1000 Fuel Analyzer will notify youwhen a fuel that you are measuring is outside of the analyzer’s factory calibration matrix. (SeeMahalanobis Distance and Extrapolated Results on page 25). The software allows the user to add thefuel to the calibration matrix.

Please refer to the Calibration Update Manual for detailed instructions.

Page 40: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

34

THIS PAGE INTENTIONALLY BLANK.

Page 41: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

35

SECTION 9PRODUCT MAINTENANCE

Cleaning the GS-1000

After using the GS-1000, pump the fuel lines dry before storing the instrument. To do this, simply attachan empty bottle to the Sample Input Port and perform an analysis function. If the instrument is to beidle for more than 48 hours, it is recommended that a hydrocarbon solvent, such as petroleum ether,be used to flush the sample cell. Note: Petroleum Ether is a suitable cleaning solvent, pump dry the fuelline when completed.

Fuel spills on the outside of the instrument should be wiped off as soon as possible with a clean rag.Most stains can be removed by wiping with a mild solvent.

Sample Induction Cleaning Procedure

Follow these step-by-step instructions in order to properly clean the Instruments plumbing andsample induction system.

WARNING:WARNING:WARNING:WARNING:WARNING: Use protective gloves and eye protection when snapping open a PetroSpecCleaning Solution Ampoule or Check Standard Ampoule. Be sure the area is wellventilated. Use the plastic-opening tool provided with each tray of Cleaning Solution.Follow the instructions below for safely opening the Cleaning Solution Ampoules. Besure to keep the tool, as it is the only one provided.

WARNING:WARNING:WARNING:WARNING:WARNING: When opening a Cleaning Solution Ampoule, be sure to slide the openingtool all the way onto the stem of the ampoule. The tool should bottom out on the bodyof the ampoule. Hold the body of the ampoule firmly in one hand. Make sure that theampoule is in the upright position. Break the stem of the ampoule, being careful to snapit away from you.

WARNING: WARNING: WARNING: WARNING: WARNING: Do not leave the Cleaning Solution Bottle Assembly unattached or unat-tended. The vapor pressure will cause the Cleaning Solution to flow through the supplytube. If the Cleaning Solution is not depleted during the cleaning operation, remove theCleaning Solution Bottle Assembly Cap and recap the bottle with a normal plastic cap forstorage.

WARNING:WARNING:WARNING:WARNING:WARNING: It is very important that the Fuel Filter be installed correctly. If the FuelIt is very important that the Fuel Filter be installed correctly. If the FuelIt is very important that the Fuel Filter be installed correctly. If the FuelIt is very important that the Fuel Filter be installed correctly. If the FuelIt is very important that the Fuel Filter be installed correctly. If the FuelFilter is not seated properly in the filter housing, sediment can bypass the Fuel FilterFilter is not seated properly in the filter housing, sediment can bypass the Fuel FilterFilter is not seated properly in the filter housing, sediment can bypass the Fuel FilterFilter is not seated properly in the filter housing, sediment can bypass the Fuel FilterFilter is not seated properly in the filter housing, sediment can bypass the Fuel Filtercausing system blockages.causing system blockages.causing system blockages.causing system blockages.causing system blockages.

Page 42: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

36

1. Back-Flush the Instrument …

a. Carefully remove any waste bottles, sample bottles, and the sample tube from theSample Inlet Port of the Instrument.

b. Carefully remove the Brass Barbed Fitting from the Waste Outlet Port using a 5/16"Wrench or Nut Driver.

c. Loosen the Thumb Screw at the top of the Sample Induction System, open the FuelFilter Access Cover, and remove the Teflon Fuel Filter. Close the Cover and tightenthe Thumb Screw.

Tip: Tip: Tip: Tip: Tip: If the Teflon Fuel Filter is dry it may stick to the Housing. Use a drop of fuel to wetthe Filter to facilitate ease of removal…

d. Remove the Cap from the Cleaning Solution Bottle Assembly. Snap open a CleaningSolution vial and pour the contents of the vial into the bottle. Screw the Cap onto thebottle and tighten.

*** See Warning A ***

e. Attach the Cleaning Solution Bottle Assembly to the instrument as shown in Figure1. Connect the Red Fitting of the Cleaning Solution Bottle Assembly to the threadedhole in the underside of the Waste Outlet Port. Connect the Quick-Connect Com-pressed Air Fitting on the yellow tube of the Cleaning Solution Bottle Assembly tothe Quick-Connect Compressed Air Fitting on the side of the instrument. Place theToggle Switch in the DOWN position.

f. Connect a Waste Bottle with Drilled Cap and a Teflon Tube with Red Fitting to theSample Inlet Port of the instrument.

g. Press AUTO (1) on your GS or MD Series Analyzer. Allow the instrument to runthrough its analysis cycle. Disregard any results that are displayed.

2. Clean the Filter Housing …

a. Unscrew the Cap of the Cleaning Solution Bottle to relieve the pressure.

b. Loosen the Thumb Screw at the top of the Sample Induction System and open the FuelFilter Access Cover.

c. Thoroughly clean the inside of the Fuel Filter housing to remove any residue.

d. Install a Teflon Fuel Filter (for Forward-Flush only).

*** See Warning C ***

e. Close the Fuel Filter Access Cover and tighten the Thumb Screw.

3. Forward-Flush the Instrument …

a. Carefully remove the Waste Bottle connected to Sample Inlet Port of the instrument.

b. Unscrew the Red Fitting on the Cleaning Solution Bottle Assembly from the WasteOutlet Port and screw it into the threaded hole in the underside of the Sample InletPort. See Figure 2.

c. Connect the Waste Bottle with Drilled Cap and a Teflon Tube with Red Fitting to theWaste Outlet Port of the instrument.

d. Press AUTO (1) on your GS or MD Series Analyzer. Allow the instrument to runthrough its analysis cycle. Disregard any results that may be displayed.

Page 43: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

37

4. Repeat steps 1, 2 and 3 …

a. Repeat steps 1, 2 and 3 until the Cleaning Solution has been depleted.

5. Revert to Normal set-up …

a. Be sure to return the Instrument to the normal Analysis mode set-up when finishedwith the Sample System Cleaning Procedure.

b. Remove the Waste Bottle and properly dispose of its contents.

c. Disconnect the Cleaning Solution Bottle Assembly Red Fitting from the Waste OutletPort.

*** See Warning B ***

d. Disconnect the Quick-Connect Compressed Air Fitting on the yellow tube of theCleaning Solution Bottle Assembly from the Fitting on the side of the Instrument.

e. Place the Toggle Switch on the side of the Instrument in the UP position.

f. Reinstall the Brass Barbed Fitting into the threaded hole on the underside of the WasteOutlet Port.

g. Replace the Waste Bottle into the Waste Outlet Port.

6. Install a fresh Fuel Filter …

a. Loosen the Thumb Screw at the top of the Sample Induction System and open the FuelFilter Access Cover.

b. Remove any Fuel Filters used in the Sample System Cleaning procedure.

c. Thoroughly clean the inside of the Fuel Filter housing to remove any residue.

d. Install a new Fuel Filter in the filter housing. Be sure that the Fuel Filter seatscorrectly.

*** See Warning C ***

e. Close the Fuel Filter Access Cover and tighten the Thumb Screw.

7. Flush the System Induction System dry …

a. Using an empty sample bottle attached to the Sample Inlet Port and a Waste Bottleattached to the Waste Outlet Port, run one more AUTO (1) cycle to flush the systemdry.

Page 44: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

38

Back-Flush Setup

Page 45: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

39

Forward-Flush Setup

Page 46: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

40

Replacing Fuel Filters

The Teflon fuel filter should be replaced every 50-75 samples, or at any time that it appears dirty orimpairs the flow of fuel through the instrument. Replacement filters can be ordered from PACat 281-580-0339. The model number for the replacement filters is BTGS-1420.

To change the fuel filter, disconnect the fuel sample bottle from the Sample Input Port. Lift the FuelFilter Access Cover by turning the Knurled Screw counter clockwise. Remove and discard the usedfilter. Inspect the o-ring located in the filter housing cap for any nicks or cuts which could cause leaks.It may be helpful to wet the filter with a drop of gasoline in order to facilitate ease of removal.

Place a new Teflon filter membrane in the filter housing. Replace the Fuel Filter Access Cover andtighten the Knurled Screw by turning it clockwise. It is very important to make sure that the filter isseated properly. A poorly seated filter can allow the fuel to bypass the filter, which will result in a clogin the fuel intake path.

Please NotePlease NotePlease NotePlease NotePlease Note: The fuel filters are shipped in tubes with 20 filters. A clear plastic separatordisk separates each filter. Be careful to discard the separator. If the separator is installedin the fuel filter housing, poor fuel flow, or no fuel flow will result.

Replacement filters are available from PAC Customer Service, at 281-580-0339.

Page 47: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

41

APPENDIX 1MESSAGES

Extrapolated Result

This message indicates that the fuel sample tested does not fall within the calibration sample set.Accuracy of the stated properties may be in jeopardy. Use of the User Updateable Calibration softwareis recommended for fuels that present this warning. Please refer to the description of MahalanobisDistance and extrapolated results on page 25.

Printer Fault Detected, No Connection Or Paper Feed Error

This message is displayed if the GS-1000 is unable to establish communications with an externalprinter when the print functions are accessed. Check that the printer is connected securely to theGS-1000 parallel printer port with a parallel printer cable. Verify that the printer has paper, is notjammed, and is on-line. If the message persists, call PetroSpec customer service.

Hydrocarbons Detected, Please Power Down

The GS-1000 contains an internal hydrocarbon sensor that detects dangerous levels of hydrocarbonvapors, either from leaks inside the instrument or in the operating atmosphere. If this message isdisplayed, IMMEDIATELY SHUT OFF POWER TO THE INSTRUMENT. If no fuel leaks areapparent, apply power again. If the condition persists, move the instrument to a better ventilated areaand power up the instrument. If the message still persists, or if there is any evidence of a fuel leak, callPetroSpec customer service.

Chopper Wheel Failure Detected

This message indicates a malfunction of the GS-1000 optical system. Call PetroSpec customer service.

Filter Wheel Failure Detected

This message indicates a malfunction of the GS-1000 optical system. Call PetroSpec customer service.

Source Failure Detected

This message indicates a malfunction of the GS-1000 optical system. Call PetroSpec customer service.

Page 48: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

42

TABLE FORMAT PRINTOUT (DEFAULT)

GS1000 Fuel Analyzer Analysis Results

——————11:38:25am 3/19/98MeOH = 2.1 Vol%EtOH = 1.3 Vol%MTBE = 5.0 Vol%ETBE = 0.0 Vol%TAME = 1.4 Vol%DIPE = 0.0 Vol%TBA = 0.0 Vol%

WEIGHT-% OXYGEN .. .26

BENZENE =.82 Vol%AROMATICS = 26.1 Vol%OLEFINS = 12.5 Vol%TOTAL XYLENES = 7.8 Vol%TOLUENE = 5.8 Vol%SATURATES = 59.8 Vol%RON = 93.5MON = 85.1R+M/2 = 89.3EXPECTED R+M/2 = “NA”RON ENGINE BIAS = 0.0MON ENGINE BIAS = 0.0DI = 1039T50 214 F = 214 FT90 = 320 FE200 = 43%E300 = 85%VOC Season: SummerVOC Region: 1Expected RVP 10.00VOC Performance = 16.6%VOC = xxxxx mg/mile

Page 49: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

43

REPORT FORMAT PRINTOUT (SELECTABLE)

GS1000 Fuel Analyzer Analysis

Date : 11:38:25am 3/19/98

Sample NUMBER : (Demonstration Data)

DESCRIPTION: ___________________________________________________________________________________________________________________________________________________________________________________________________________________________

SAMPLE STANDARD_______ __________Species Vol% Species Vol%______ ____ ______ ____

MeOH 2.1 MeOH _______EtOH 1.3 EtOH _______MTBE 5.0 MTBE _______ETBE 0.0 ETB _______TAME 1.4 TAME _______DIPE 0.0 DIPE _______TBA 0.0 TBA _______

WEIGHT-% OXYGEN ... .26

CONCENTRATION OF BENZENE ... .82 Vol%CONCENTRATION OF OLEFINS ... 12.5 Vol%CONCENTRATION OF SATURATES ... 59.8 Vol%CONCENTRATION OF TOTAL AROMATICS .. 26.1 Vol%CONCENTRATION OF TOTAL XYLENES 7.8 Vol%CONCENTRATION OF TOLUENE 5.8 Vol. %

OCTANE MEASURED EXPECTED_______ __________ __________

OCTANE RESULT 89.3 “NA”RON 93.5MON 85.1RON ENGINE BIAS = 0.0MON ENGINE BIAS = 0.0

DI 1039T50 214 FT90 320 FE200 43%E300 85%VOC Season: Summer VOC Region: 1Expected Sulfur 339ppmExpected RVP 10.0 psigTotal VOC 1709 mg/mileVOC Reduction 16.6%

SIGNATURE: _________________________

Page 50: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

44

THIS PAGE INTENTIONALLY BLANK.

Page 51: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

45

APPENDIX 2GASOLINE SAFETY INFORMATION

Personal Protection Information

Ventilation — Use adequate ventilation to control exposure below recommended levels

Respiratory Protection — For concentrations exceeding the recommended exposure level, use NIOSH/MSHA approved air purifying respirator. When entering into or exiting from concentrations ofunknown exposure, use NIOSH/MSHA approved self-contained breathing apparatus (SCBA).

Eye Protection — Use safety glasses with side shields and face shield for splash protection.

Skin Protection — Use gloves resistant to the materials being used (Viton, nitrile, neoprene). Use full-body, long sleeved garments to prevent skin contact.

Handling And Storage Precautions

Do not get in eyes, on skin or clothing. Do not breathe vapors, mist, or fumes. Do not swallow. Maybe aspirated into lungs. Wear protective equipment and/or garments described above if exposureconditions warrant. Wash thoroughly after handling. Launder contaminated clothing before reuse.Use only with adequate ventilation.

Store in a well-ventilated area. Store in tightly closed container. Bond and ground during transfer.Keep away from heat, sparks, and flames.

Reactivity Data

Stability — Stable

Incompatibility (Material to Avoid) — Oxygen and strong oxidizing agents.

Hazardous Decomposition Products — Carbon oxides, lead fumes and various hydrocarbons whenburned.

Health Hazard Data

Acute Effects of Overexposure

Eye — May cause mild irritation, with stinging and redness of the eyes.

Skin — May cause mild irritation. Repeated or prolonged contact may cause defatting of the skin,resulting in dermatitis.

Inhalation — May cause headache, nausea, weakness, sedation, and unconsciousness

Ingestion — May cause irritation to intestines. If swallowed, may be aspirated resulting in inflamma-tion and possible fluid accumulation in the lungs.

Subchronic and Chronic Effects of Overexposure

Unleaded gasoline has produced kidney cancer in male rats only. No comparable kidney disease isknown to occur in humans.

Gasoline’s generally contain benzene which has been designated a carcinogen by the NationalToxicology Program (NTP), the International Agency for Research on Cancer (IARC), and theOccupational Safety and Health Administration (OSHA). Benzene may produce blood changes whichinclude reduced platelets, red blood cells, and white blood cells. Also, aplastic anemia, and acutenonlymphotic leukemia. Benzene has produced fetal death in laboratory animals and caused chromo-some changes in humans and mutation changes in cells of other organisms.

Laboratory animals have exhibited a higher degree of narcosis when exposed to both butane andbutylene (additive effect), than the degree of narcosis exhibited following exposure to butane orbutylene alone.

Isopentane did not produce kidney damage in a subchronic oral laboratory study or in a subchronicinhalation exposure to 4500 ppm and 1000 ppm or a 50/50 mixture of isobutane and isopentane.

Page 52: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

46

First Aid And Emergency Procedures

Eye — Flush eyes with running water for at least fifteen minutes. If irritation or adverse symptomsdevelop, seek medical attention.

Skin — Wash skin with soap or water for at least fifteen minutes. If irritation or adverse symptomsdevelop, seek medical attention.

Inhalation — Remove from exposure. If breathing is difficult, give oxygen. If breathing ceases,administer artificial respiration followed by oxygen. Seek immediate medical attention.

Ingestion — Do not induce vomiting. Seek immediate medical attention.

Fire And Explosion Data

Flash point — 35F (37C)

Flammable Limits (% by volume in air) — LEL — 1.5; UEL — 7.6

Fire Extinguishing Media — Dry chemical, foam, or carbon dioxide (CO2)

Special Fire Fighting Precautions — Evacuate area of all unnecessary personnel. Wear appropriatesafety equipment for fire conditions including NIOSH/MSHA self-contained breathing apparatus(SCBA). Shut off source, if possible. Water, fog or spray may be used to cool exposed containers andequipment. Do not spray water directly on fire; product will float and could re-ignite on surface ofwater.

Fire and Explosion Hazards — Carbon oxides and various hydrocarbons formed when burned.Gasoline’s containing Tetraethyl Lead will form lead fumes when burning. Highly flammable vaporswhich are heavier than air may accumulate in low areas and/or spread along ground away fromhandling site. Flashback along vapor trail may occur.

Spill, Leak, And Disposal Procedures

Precautions required if material is released or spilled:

Evacuate area of all unnecessary personnel. Wear protective equipment and/or garments describedabove if conditions warrant. Shut off source, if possible, and contain spill. Protect from ignition. Keepout of water sources and sewers. Absorb in dry, inert material. Transfer to disposal drums using non-sparking equipment.

Waste Disposal (Insure conformity with all applicable disposal regulations)

Incinerate or otherwise manage in a RCRA permitted waste management facility.

Hazard Classification

Gasoline meets the following hazard definitions as outlined by the Occupational Safety and HealthHazard Communication Standard (29 CFR Section 1910.1200):

• Flammable Liquid

• Health Hazard (Section F)

Page 53: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

47

APPENDIX 3WARRANTY INFORMATION

A. Limitations of Liability

PetroSpec warrants that the GS-1000 Plus VOC will be free from defects in materials and workmanshipfor a period of one (1) year from the date of invoice. If, during the warranty period, the product isdetermined by us to be defective, we and/or our representative will repair or replace the product at nocost to the purchaser.

PetroSpec shall not be responsible for any expenses incurred by the user.

This warranty applies only to products that are operated in accordance with PetroSpec specifications.This warranty will not apply to any product that has been subject to misuse, negligence or accident;or misapplied; or used in violation of product manuals, instructions, or warnings; or modified orrepaired by unauthorized persons; or improperly installed.

B. Inspection

It is the responsibility of the purchaser to inspect the product promptly after receipt and to notifyPetroSpec at its Texas office in writing of any claims, including claims of breach of warranty, withinthirty (30) days of discovery of the facts upon which the claim is based. Your failure to give writtennotice of a claim within the time period shall be deemed to be a waiver of such claim.

C. Limitations of Remedy and Warranty

The provisions in paragraph A are our sole obligation and exclude all other remedies or warranties,expressed or implied, including warranties of merchantability and fitness for a particular purpose,whether or not purposes or specifications are described herein. We further disclaim any responsibilitywhatsoever to the purchaser or to any other person for injury to person or damage to or loss of propertyor value caused by any product which has been subjected to misuse, negligence, or accident; ormisapplied; or used in violation of product manuals, instructions or warnings; or modified or repairedby unauthorized persons; or improperly installed.

D. Limitations of Damages

Under no circumstances shall PetroSpec be liable for any incidental, consequential or specificdamages, losses or expenses arising from this contract or its performance in connection with the useof, or inability to use, our product for any purpose whatsoever.

E. Limitation of Actions

No action, regardless of form, arising from this contract may be commenced more than one year afterthe cause of action has accrued, except an action for nonpayment.

F. Collateral Promises

There are no representations, warranties, or conditions, expressed or implied, statutory or otherwiseexcept those herein contained, and no agreement or waivers collateral hereto shall be binding on eitherparty unless in writing and signed by the purchaser and accepted by PetroSpec at its Houston office.

G. Interpretation

Rights and liabilities arising out of any contract with PetroSpec shall be determined under the UniformCommercial Code as enacted in Texas.

Page 54: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

48

THIS PAGE INTENTIONALLY BLANK.

Page 55: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

49

PETROSPEC WARRANTYREGISTRATION FORM

PetroSpec Warranty Registration Form

To assure warranty coverage of your PetroSpec Fuel Analyzer, and to allow us to keep you informedof product updates and servie advisories, please fill-in the information below and FAX to:

PAC/PetroSpec at 281-580-0339

or mail to:

PAC/PetroSpec, 300 Bammel Westfield Road, Houston, TX 77090

Company Name: ______________________________________________________________

Key Contact: ______________________________________________________________

Address (Location of Analyzer): ______________________________________________________________

______________________________________________________________

______________________________________________________________

Phone Number: ______________________________________________________________

FAX Number: ______________________________________________________________

Model: ______________________________________________________________

Serial Number: ______________________________________________________________

Purchased From: ______________________________________________________________

Purchase Date: ______________________________________________________________

Keep a copy of this warranty for your records.

Thank you for your business.PetroSpec

GS-1000 Plus VOC

Page 56: manual presion de vapor

PETROSPEC GS-1000 PLUS VOC | OPERATING M A N U A L

50

THIS PAGE INTENTIONALLY BLANK.

Page 57: manual presion de vapor

Installation Problem Report

Code: QF 8.02 Rev. D Date: 8/24/04 Page 1 of 1

Company: Installation Date:

Contact: Installed By:

Address: PAC Representative:

Sales Order #:

City: St: Zip:

Country: Territory:

Phone: Fax: Fax or e-mail complete report to Service Manager

fax: 281-580-0719 e-mail: [email protected]

E-mail:

Purchase Date S/N Model # Install Date

Problem(s) Model: S/N:

Part # Description of Problem

Model: S/N:

Part # Description of Problem

Model: S/N:

Part # Description of Problem

Type of Problem: Appearance Broken Electronic Missing Off Spec Wrong Assembly Electric Leak Mechanical Non-function