    Technical

    Publications

    Direction 5314625 -100

    Rev. 5

    LOGIQ e Advanced Reference Manual

    Copyright© 2008-2010 By General Electric Co.

    Operating Documentation

    GE Healthcare

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    2/212

    Regulatory Requirement  

    This product complies with regulatory requirements of the following European

    Directive 93/42/EEC concerning medical devices

    This manual is a reference for the LOGIQ e. It applies to R5.2.x, R6.x.x, and

    backward software version for the LOGIQ e ultrasound systems.

    GE Healthcare

    GE Healthcare: Telex 3797371

    P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.

    (Asia, Pacific, Latin America, North America)

    GE Ultraschall TEL: 49 212.28.02.208

    Deutschland GmbH & Co. KG FAX: 49 212.28.02.431

    Beethovenstraße 239

    Postfach 11 05 60

    D-42655 Solingen GERMANY 

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    3/212

    LOGIQ e Advanced Reference Manual i  -1

    Direction 5314625-100 Rev. 5 

    Revision History

    List of Effective Pages

    Table i-1: Reason for Change

    REV DATE REASON FOR CHANGE

    Rev. 1 Aug 1st, 2008 Preliminary Release

    Rev. 2 Mar 17st, 2009 Update Maximum Probe Temperature and FDA Acoustic OutputData

    Rev. 3 Mar 17st, 2009 Add a note:16L-RS is not available for China

    Rev. 4 Jan 28, 2010 Add new probes 6Tc - RS and 6S-RS

    Rev. 5 Nov 30th, 2010 Update the parameter of probe

    Table i-2: List of Effective Pages

    PAGE NUMBER

    REVISION

    NUMBER PAGE NUMBER

    REVISION

    NUMBER

    Title Page Rev. 5 2-1 thru 2-10 Rev. 5

    Revision History i-1 and i-2 Rev. 5 3-1 thru 3-74 Rev. 5

    Table of Contents ii-1 thru ii-2 Rev. 5 Index 1 Rev. 5

    1-1 thru 1-108 Rev. 5

    Please verify that you are using the latest revision of this document.

    Information pertaining to this document is maintained on GPC (GE

    Medical Systems Global Product Configuration). If you need to know the

    latest revision, contact your distributor, local GE Sales Representative or

    in the USA call the GE Ultrasound Clinical Answer Center at

    1 800 682 5327 or 1 262 524 5698.

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    4/212

    i -2 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    This page intentionally left blank.

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    5/212

    Table of Contents

    LOGIQ e Advanced Reference Manual ii  -1

    Direction 5314625-100 Rev. 5 

    Table of Contents

    Table of Contents - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-1

    Chapter 1 — Acoustic OutputBioeffects - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -1-2

    Concerns Surrounding the Use of Diagnostic Ultrasound - - - - - - - - - - - 1-2

    Measurement Precision and Uncertainty - - - - - - - - - - - - - - - - - - - - - - - 1-3

    FDA Acoustic Output Data - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -1-4

    Maximum output summary- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-4IEC Acoustic Output Tables - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-91

    Chapter 2 — Measurement FormulasMeasurement Formulas - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -2-2

    Formulas–Generic - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-2

    Formulas–OB - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-4

    Formulas–GYN - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-8

    Formulas–Vascular - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-10

    Chapter 3 — OB TablesOB Tables- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-2

     ASUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2

    Berkowitz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-3

    Brenner - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4

    Campbell - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4

    Eriksen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-5

    Goldstein - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-6

    Hadlock - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-7

    Hansmann - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-13

    Hellman - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-21Hill - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-21

    Hohler - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-22

    Jeanty - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-22

    JSUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-34

    Kurtz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-38

    Mayden - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-39

    Mercer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-40

    Merz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-41

    Moore - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-51

    Nelson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-51

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    6/212

    ii -2 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Osaka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-52

    Paris - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-56

    Rempen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-59

    Robinson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-64

    Tokyo - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-64

    Tokyo Shinozuka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-68

    Williams - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-74

    Yarkoni - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-74

    Index - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -1-1

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    7/212

    LOGIQ e Advanced Reference Manual 1-1

    Direction 5314625-100 Rev. 5 

    Chapter 1

     Acoustic Output

    Provides Acoustic Output information and tables, as well as possible

    bioeffects and prudent use.

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    8/212

     Acoustic Output 

    1-2 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Bioeffects

    Concerns Surrounding the Use of Diagnostic Ultrasound

    During a diagnostic ultrasound examination, high frequency sound penetrates and

    interacts with tissue in and around the area of anatomy to be imaged. Only a small

    portion of this sound energy is reflected back to the probe for use in constructing the

    image while the remainder is dissipated within the tissue. The interaction of sound

    energy with tissue at sufficiently high levels can produce biological effects (aka

    bioeffects) of either a mechanical or thermal nature. Although the generation of bioeffect

    is intentional with therapeutic ultrasound, it is generally undesired in diagnostic

    applications and may be harmful in some conditions.

    NOTE: The American Institute of Ultrasound in Medicine has published

    a document entitled "Medical Ultrasound Safety". This three

     part document covers Bioeffects and Biophysics, Prudent Use

    and Implementing ALARA.

    Ultrasound users should read the AIUM to become more familiar with Ultrasound safety.

     A copy of this document is included as part of the documentation CD (5148218).

    To contact the AIUM concerning their publications:• In the USA, by telephone at 1-800-638-5352.

    • To write them, use the following address:

     AIUM

    14750 Sweitzer Lane

    Suite 100

    Laurel, MD, USA 20707-5906

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    9/212

    Bioeffects

    LOGIQ e Advanced Reference Manual 1-3

    Direction 5314625-100 Rev. 5 

    Measurement Precision and Uncertainty

    Precision and measurement uncertainty of the system used to measure the acoustic

    output is provided in the following table. These values are determined in accordance with

    Section 6.4 of the Standard for Real-Time Display of Thermal and Mechanical Acoustic

    Output Indices on Diagnostic Ultrasound Equipment, rev 2 AIUM/NEMA 2004, and the

     Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment, rev 3

    NEMA 2004.

    Table 1-1: Measurement Precision and Uncertainty

    Center

    Frequency

    Acoustic

    Power 

    Peak

    Rarefactional

    Pressure

    Acoustic

    Intensity

    Measurement

    Uncertainty

    + 0.25% + 5.6% + 9.5% + 18.9%

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    10/212

     Acoustic Output 

    1-4 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    FDA Acoustic Output Data

    Maximum output summary

    The following tables list the typical maimum acoustic output levels achievalbe with

    LOGIQ e for all probes and operational modes. It is intended that this is information be

    useful in making ALARA decisions and selecting the most appropriate probe for the

    application. In accordance with US FDA Guidelines, the overall maximum acoustic SPTA

    intensity for LOGIQ e is limited to 720 mW/cm2 and MI is limited to 1.9. Modes for which

    TI does not exceed 0.1 are indicated by

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    11/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-5 

    Direction 5314625-100 Rev. 5 

    Maximum Probe Temperature

    In accordance with IEC 60601-2-37 requirement, the maximum lens temperature

    is limited to 43 degC with thermal phantom, while 50 degC in air.

    Table Release Notes-2: Maximum Probe Temperature (Degrees C)

    Probe

    Max Temp

    RemarkWith TMM Phan-tom* Mode In Air Mode

    4C-RS 38.8 CFM 41 CFM For LOGIQ e R5.2.x

    37.9 CFM 32.2 CFM For LOGIQ e R6.x.x

    E8C-RS 42.6 B 38.6 B

    8L-RS 40.1 CFM 38 Dop

    8C-RS 38.8 B 39.7 B

    i12L-RS 42.2 B 37.6 B For LOGIQ e R5.2.x

    40.8 CFM 37.3 CFM For LOGIQ e R6.x.x

    3S-RS 40.3 CFM 45.4 CWD

    12L-RS 37.9 CFM 36.6 Dop

    16L-RS 39.5 CFM 36.2 CFM

    9L-RS 39.2 Dop 39.7 B

    i739-RS 41.8 Dop 37.6 B

    t739-RS 42.3 CFM 36.6 Dop

    6S-RS 41.9 B 42.8 M

    6Tc-RS 41.7 CFM 35.3 CFM

    Notes:

    Lens temperature measured under following conditions per IEC 60601-2-37:

    Thermocouple was placed at the geometric center of the lens.

    Thermal phantom* placed in 33 degC water bath for non-endo probes (all probes except E8C-RS/i12LRS/

    i739/t739/6Tc).

    Thermal phantom* placed in 37 degC water bath for endo probes (E8C-RS/i12L-RS/i739/t739/6Tc).

    Probe placed upright in contact with above thermal phantom.

     Autofreeze capability disabled.

    Lens temperature monitored for 30 min.

    Measurement uncertainty: 2 degC.

    *Thermal phantom made with tissue-mimicking material as referenced in IEC60601-2-37: Appendix J.

    16L-RS is not available for China.

    6S-RS and 6Tc-RS are only for LOGIQ e R6.x.x.

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    12/212

     Acoustic Output 

    1-6 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table Key

    Table 1-1: Table Key

    FDA IEC Meaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3

    a    Acoustic Attenuation Coefficient  / Derating factor (usually 0.3 dB/cm-MHz)

     Aaprt  Aarpt -12db Output Beam Area / Active aperture area

    C MI  Normalizing Coefficient 

    Deq Deq Equivalent Aperture Diameter  / (same)

    d-6 d -6 Pulse Beam Width / Beam diameter at –6 dB

    deq d eq Equivalent Beam Diameter 

    f c ƒawf   Acoustic Working Frequency  / Center frequency

    Ipa I pa Pulse-Average Intensity 

    Ipa.3 I pa,  Attenuated Pulse-Average Intensity 

    PII I pi Pulse-Intensity Integral 

    PII.3 I pi,  Attenuated Pulse-Intensity Integral 

    ITA I ta(z ) Temporal-Average Intensity 

    ITA.3(Z) I ta,(z )  Attenuated Temporal-Average Intensity  / (at depth z)

    ISPTA(Z) I zpta(z ) Spatial-Peak Temporal-Average Intensity 

    ISPTA.3(Z) I zpta,(z )  Attenuated Spatial-Peak Temporal-Average Intensity 

    MI MI Mechanical Index  

    Wo P Output Power   / Time average acoustic power at the source

    W.3(Z) P   Attenuated Output Power  / Time average acoustic power derated to depth z

    Wo1 P 1 Bounded Output Power  / Power emitted from the central 1cm of aperture

    PII  pi Pulse Pressure Squared Integral / Pulse intensity integral 

    pr   pr  Peak-Rarefactional Acoustic Pressure / (same)

    pr.3  pr   Attenuated Peak-Rarefactional Acoustic Pressure / (same)

    PRF  prr Pulse Repetition Rate / Pulse repetition frequency

    TI TI Thermal Index   / (same)

    TIB TIB Bone Thermal Index   / (same)

    TIC TIC Cranial-Bone Thermal Index   / (same)

    TIS TIS Soft-Tissue Thermal Index   / (same)

    PD t d Pulse Duration / (same)

    x-12,y-12  X, Y -12 dB Output Beam Dimensions / (same)

    Z Z Distance from the Source to a Specified Point   / (same)

    Zsp Z b Depth for TIB / Depth at which the relevant index is maximum

    Zbp Z bp Break-Point Depth / (same)

    Zsp Z s Depth for TIS / Depth at which the relevant index is maximum

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    13/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-7 

    Direction 5314625-100 Rev. 5 

    Table 1-2: Table Key (in French 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    14/212

     Acoustic Output 

    1-8 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-3: Table Key (in French 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    15/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-9

    Direction 5314625-100 Rev. 5 

    Table 1-4: Table Key (in German 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    16/212

     Acoustic Output 

    1-10 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-5: Table Key (in German 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    17/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-11

    Direction 5314625-100 Rev. 5 

    Table 1-6: Table Key (in Italian 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    18/212

     Acoustic Output 

    1-12 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-7: Table Key (in Italian 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    19/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-13

    Direction 5314625-100 Rev. 5 

    Table 1-8: Table Key (in Brazilian Portuguese 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    20/212

     Acoustic Output 

    1-14 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-9: Table Key (in Brazilian Portuguese 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    21/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-15 

    Direction 5314625-100 Rev. 5 

    Table 1-10: Table Key (in Danish 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    22/212

     Acoustic Output 

    1-16 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-11: Table Key (in Danish 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    23/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-17 

    Direction 5314625-100 Rev. 5 

    Table 1-12: Table Key (in Latin American Spanish 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    24/212

     Acoustic Output 

    1-18 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-13: Table Key (in Latin American Spanish 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    25/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-19

    Direction 5314625-100 Rev. 5 

    Table 1-14: Table Key (in Norwegian 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    26/212

     Acoustic Output 

    1-20 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-15: Table Key (in Norwegian 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    27/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-21

    Direction 5314625-100 Rev. 5 

    Table 1-16: Table Key (in Swedish 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    28/212

     Acoustic Output 

    1-22 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-17: Table Key (in Swedish 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    29/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-23

    Direction 5314625-100 Rev. 5 

    Table 1-18: Table Key (in Finnish 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    30/212

     Acoustic Output 

    1-24 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-19: Table Key (in Finnish 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    31/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-25 

    Direction 5314625-100 Rev. 5 

    Table 1-20: Table Key (in Japanese 1)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    32/212

     Acoustic Output 

    1-26 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-21: Table Key (in Japanese 2)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    33/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-27 

    Direction 5314625-100 Rev. 5 

    Table 1-22: Table Key (in Chinese)

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    34/212

     Acoustic Output 

    1-28 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    4C-RS

    Table 1-23: Transducer Mode 4C-RS Operating Mode: B Mode (for LOGIQ e R5.2.x)

    Index Label MI

    TIS TIB

    TICscan non-scan non-sca

    nAaprt< Aaprt>1

    Global Maximum: Index Value 0.92 1.15 0.45

    IEC FDA Units

     pra  pr.3  (MPa) 1.47

    P  Wo (mW) 82.4 82.4

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 4.3

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 2.55 2.70 2.70

    Dim of Aaprt X (cm) 1.2 1.2

    Y (cm) 1.3 1.3

    td PD (µsec) 0.46

     prr  PRF (Hz) 45

     pr  at max. I pi  pr @PIImax  (MPa) 2.15

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 2.75 2.75

    FLY (cm) 5.24 5.24

    I pi at max. MI  (uJ/

    cm2)

    64.2

    2

    Frequency (MHz) 3.0 3.0 3.0

    Image Depth (cm) 7.0 7.0 7.0

    Focus (cm) 6.4 5.12 5.12

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    35/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-29

    Direction 5314625-100 Rev. 5 

    Table 1-24: Transducer Mode 4C-RS Operating Mode: M Mode (Inc. B-Mode)

    (for LOGIQ e R5.2.x)

    Index Label MI

    TIS TIB

    TIC

    scan

    non-scan

    non-

    scanAaprt<

    Aaprt

    >1

    Global Maximum: Index Value 0.70 0.88 - 0.56 1.52 1.22

    IEC FDA Units

     pra  pr.3  (MPa) 1.12

    P  Wo (mW) 69.7 - 69.7 69.7

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) 82.83

    z s  z1 (cm) 4.3

    z bp  zbp (cm) 4.3

    z b  zsp (cm) 4.3

    z  at max. I pi, zsp (cm) 4.3

    d eq(z b) deq(zsp) (cm) 3.15

    ƒawf  f c (MHz) 2.72 2.72 - 2.72 2.72 2.72

    Dim of Aaprt X (cm) 1.2 - 6 6 1.2

    Y (cm) 1.3 - 1.3 1.3 1.3

    td PD (µsec) 0.46

     prr  PRF (Hz) 519.

    07

     pr  at max. I pi  pr @PIImax  (MPa) 1.63

    d eq at max. I pi  deq@PIImax  (cm) 3.15

    Focal Length FLX

    (cm) 2.75 - 2.75 2.75

    FLY (cm) 5.24 - 5.24 5.24

    I pi at max. MI  (uJ/cm2) 37.3

    1

    Frequency (MHz) 3.0 3.0 - 3.0 3.0

    Image Depth (cm) 7.0 7.0 - 7.0 7.0

    Focus (cm) 6.4 5.12 - 5.12 5.12

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    36/212

     Acoustic Output 

    1-30 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-25: Transducer Mode 4C-RS Operating Mode: Color Flow Mode

    (for LOGIQ e R5.2.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.04 1.67 1.08

    IEC FDA Units

     pra  pr.3  (MPa) 1.39

    P  Wo (mW) 137.1 137.1

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 5.4

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 2.54 2.54 2.54

    Dim of Aaprt X (cm) 6.0 6.0

    Y (cm) 1.3 1.3

    td PD (µsec) 1.49

     prr  PRF (Hz) 8

     pr  at max. I pi  pr @PIImax  (MPa) 2.23

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 11.7 11.7

    FLY (cm) 11.9 11.9

    I pi at max. MI    (uJ/cm2) 223.50

    Frequency (MHz) 2.5 2.5 2.5

    Image Depth (cm) 18 17 17

    Vel Scale (kHz) 1.2 3.4 3.4

    ROI Preset Max Max

    PS 16 16 16

    Focus (cm) 70.4 60.8 60.8

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    37/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-31

    Direction 5314625-100 Rev. 5 

    Table 1-26: Transducer Mode: 4C-RS Operating Mode: Pulsed Doppler (Inc.B-mode)

    (for LOGIQ e R5.2.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.10 0.62 - 0.62 2.18 0.73

    IEC FDA Units

     pra  pr.3  (MPa) 1.79

    P  Wo (mW) 46.5 - 46.5 46.5

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) 139.1

    z s  z1 (cm) 3.9

    z bp  zbp (cm) 3.9

    z b  zsp (cm) 3.9 3.9

    z  at max. I pi, zsp (cm) 3.9

    d eq(z b) deq(zsp) (cm) 1.53

    ƒawf  f c (MHz) 2.84 2.46 - 2.46 2.46 2.46

    Dim of Aaprt X (cm) 1.43 - 6.0 6.0 1.43

    Y (cm) 1.3 - 1.3 1.3 1.3

    td PD (µsec) 0.51

     prr  PRF (Hz) 170

    7.6

     pr  at max. I pi  pr @PIImax  (MPa) 2.62

    d eq at max. I pi  deq@PIImax  (cm) 1.53

    Focal Length FLX (cm) - 2.5 2.5

    FLY (cm) - 3.7 3.7

    I pi aat max. MI  (uJ/

    cm2)

    63.2

    4

    Frequency (MHz) 3.3 2.5 - 2.5 2.5 2.5

    SVD (cm) 5.7 5.7 - 5.7 5.7 5.7

    Vel Scale (kHz) 1.7 3.2 - 3.2 3.2 3.2

    SV 1 2 - 2 2 2

    Image Depth(cm) 8 8 8 8 8

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    38/212

     Acoustic Output 

    1-32 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    4C-RS (continued)

    Table 1-27: Transducer Mode 4C-RS Operating Mode: B Mode (for LOGIQ e R6.x.x)

    Index Label MI

    TIS TIB

    TICscan non-scan non-sca

    nAaprt< Aaprt>1

    Global Maximum: Index Value 1.31 1.12 0.56

    IEC FDA Units

     pra  pr.3  (MPa) 1.88

    P  Wo (mW) 80.2 80.2

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 3.8

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 2.07 2.09 2.09

    Dim of Aaprt X (cm) 1.2 1.2

    Y (cm) 1.3 1.3

    td PD (µsec) 0.70

     prr  PRF (Hz) 59

     pr  at max. I pi  pr @PIImax  (MPa) 2.47

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 2.75 2.75

    FLY (cm) 5.24 5.24

    I pi at max. MI  (uJ/

    cm2)

    180.1

    1

    Frequency (MHz) CHI4.0 CHI4.0 CHI4.0

    Image Depth (cm) 7.0 7.0 7.0

    Focus (cm) 5.1 5.1 5.1

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    39/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-33

    Direction 5314625-100 Rev. 5 

    Table 1-28: Transducer Mode 4C-RS Operating Mode: M Mode (Inc. B-Mode)

    (for LOGIQ e R6.x.x)

    Index Label MI

    TIS TIB

    TIC

    scan

    non-scan

    non-

    scan

    Aapr 

    t<

    Aapr 

    t>1

    Global Maximum: Index Value 0.82 0.89 - 0.58 1.54 1.24

    IEC FDA Units

     pra  pr.3  (MPa) 1.14

    P  Wo (mW) 69.5 - 69.5 69.5

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) 4.2

    z s  z1 (cm) 4.2

    z bp  zbp (cm) 4.2

    z b  zsp (cm) 3.8

    z  at max. I pi, zsp (cm) 4.2

    d eq(z b) deq(zsp) (cm) 3.15

    ƒawf  f c (MHz) 2.08 2.11 - 2.11 2.11 2.11

    Dim of Aaprt X (cm) 1.2 - 6 6 1.2

    Y (cm) 1.3 - 1.3 1.3 1.3

    td PD (µsec) 0.72

     prr  PRF (Hz) 519.0

    7

     pr  at max. I pi  pr @PIImax  (MPa) 1.50

    d eq at max. I pi  deq@PIImax  (cm) 3.15

    Focal Length FLX

    (cm) 2.75 - 2.75 2.75

    FLY (cm) 5.24 - 5.24 5.24

    I pi at max. MI  (uJ/cm2) 48.10

    Frequency (MHz) CHI4.0 CHI4.0 - CHI4.0 CHI4.0

    Image Depth (cm) 7.0 7.0 - 7.0 7.0

    Focus (cm) 5.1 5.1 - 5.1 5.1

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    40/212

     Acoustic Output 

    1-34 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-29: Transducer Mode 4C-RS Operating Mode: Color Flow Mode

    (for LOGIQ e R6.x.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.95 1.45 1.01

    IEC FDA Units

     pra  pr.3  (MPa) 1.47

    P  Wo (mW) 109.6 109.6

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 4.6

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 2.52 2.53 2.53

    Dim of Aaprt X (cm) 6.0 6.0

    Y (cm) 1.3 1.3

    td PD (µsec) 1.49

     prr  PRF (Hz) 3

     pr  at max. I pi  pr @PIImax  (MPa) 2.21

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 11.7 11.7

    FLY (cm) 11.9 11.9

    I pi at max. MI    (uJ/cm2) 164.79

    Frequency (MHz) 2.5 2.5 2.5

    Image Depth (cm) 18 18 18

    Vel Scale (kHz) 3.4 3.4 3.4

    ROI Max Max Max

    PS 16 16 16

    Focus (cm) 70.4 70.4 70.4

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    41/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-35 

    Direction 5314625-100 Rev. 5 

    Table 1-30: Transducer Mode: 4C-RS Operating Mode: Pulsed Doppler (Inc.B-mode)

    (for LOGIQ e R6.x.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.18 0.66 - 0.66 2.06 0.78

    IEC FDA Units

     pra  pr.3  (MPa) 1.82

    P  Wo (mW) 49.2 - 49.2 49.2

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) 4.8

    z s  z1 (cm) 4.8

    z bp  zbp (cm) 4.8

    z b  zsp (cm) 4.0 4.8

    z  at max. I pi, zsp (cm) 4.8

    d eq(z b) deq(zsp) (cm) 1.53

    ƒawf  f c (MHz) 2.40 2.50 - 2.50 2.50 2.50

    Dim of Aaprt X (cm) 1.43 - 6.0 6.0 1.43

    Y (cm) 1.3 - 1.3 1.3 1.3

    td PD (µsec) 0.60

     prr  PRF (Hz) 651.

    0

     pr  at max. I pi  pr @PIImax  (MPa) 2.57

    d eq at max. I pi  deq@PIImax  (cm) 1.53

    Focal Length FLX (cm) - 2.5 2.5

    FLY (cm) - 3.7 3.7

    I pi aat max. MI  (uJ/

    cm2)

    106.

    86

    Frequency (MHz) 2.5 2.5 - 2.5 2.5 2.5

    SVD (cm) 5.7 5.7 - 5.7 5.7 5.7

    Vel Scale (kHz) 0.7 3.2 - 3.2 3.2 3.2

    SV 1 2 - 2 2 2

    Image Depth(cm) 8 8 8 8 8

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    42/212

     Acoustic Output 

    1-36 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    E8C-RS

    Table 1-31: Transducer Mode E8C-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TIC

    scan non-scan non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.24 0.56 0.5

    IEC FDA Units

     pra  pr.3  (MPa) 2.70

    P  Wo (mW) 24.9 24.9

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.5

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 4.71 4.73 4.73

    Dim of Aaprt X (cm) 0.81 0.81

    Y (cm) 0.5 0.5

    td PD (µsec) 0.32

     prr  PRF (Hz) 45

     pr  at max. I pi  pr @PIImax  (MPa) 3.41

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.81 0.81

    FLY (cm) 3.0 3.0

    I pi at max. MI  (uJ/

    cm2)

    94.14

    Frequency (MHz) THI10 THI10 THI10

    Image Depth (cm) 5.0 5.0 5.0

    Focus (cm) 2.4 1.76 1.76

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    43/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-37 

    Direction 5314625-100 Rev. 5 

    Table 1-32: Transducer Mode E8C-RS Operating Mode: M Mode (Inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scan

    non-scan

    non-

    scan

    Aaprt

    <

    Aaprt

    >1

    Global Maximum: Index Value 0.90 0.48 0.48 1.74 0.69

    IEC FDA Units

     pra  pr.3  (MPa) 1.96

    P  Wo (mW) 16.7 16.7 16.7 16.7

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.5

    z  at max. I pi, zsp (cm) 1.9

    d eq(z b) deq(zsp) (cm) 1.13

    ƒawf  f c (MHz) 4.77 4.77 4.77 - 4.77 4.77

    Dim of Aaprt X (cm) 0.81 2.5 - 2.5 0.81

    Y (cm) 0.5 0.5 - 0.5 0.5

    td PD (µsec) 0.30

     prr  PRF (Hz) 522.4

     pr  at max. I pi  pr @PIImax  (MPa) 2.51

    d eq at max. I pi  deq@PIImax  (cm) 1.13

    Focal Length FLX (cm) 0.81 2.5 2.5 0.81

    FLY (cm) 3.0 3.0 3.0 3.0

    I pi at max. MI  (uJ/cm2) 59.58

    Frequency (MHz) THI10 THI10 THI10 THI10 THI10

    Image Depth (cm) 5.0 5.0 5.0 5.0 5.0

    Focus (cm) 2.2 2.2 2.2 2.2 2.2

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    44/212

     Acoustic Output 

    1-38 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-33: Transducer Mode E8C-RS Operating Mode: Color Flow Mode

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.83 0.69 0.63

    IEC FDA Units

     pra  pr.3  (MPa) 1.84

    P  Wo (mW) 26.6 26.6

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.6

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 4.91 4.91 4.91

    Dim of Aaprt X (cm) 0.81 0.81

    Y (cm) 0.5 0.5

    td PD (µsec) 0.74

     prr  PRF (Hz) 7

     pr  at max. I pi  pr @PIImax  (MPa) 2.41

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.81 0.81

    FLY (cm) 3.0 3.0

    I pi at max. MI  (uJ/cm2)

    130.54

    Frequency (MHz) 5.0 5.0 5.0

    Image Depth (cm) 5.0 5.0 5.0

    Vel Scale (kHz) 6.6 6.6 6.6

    ROI Max Max Max

    PS 14 14 14

    Focus (cm) 2.2 2.2 2.2

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    45/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-39

    Direction 5314625-100 Rev. 5 

    Table 1-34: Transducer Mode: E8C-RS Operating Mode: Pulsed Doppler (Inc.B-mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.86 0.55 0.55 - 3.1 0.93

    IEC FDA Units

     pra  pr.3  (MPa) 1.89

    P  Wo (mW) 23.3 23.3 23.3 23.3

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.5 1.8

    z  at max. I pi, zsp (cm) 1.8

    d eq(z b) deq(zsp) (cm) 0.63

    ƒawf  f c (MHz) 4.84 4.95 4.95 - 4.95 4.95

    Dim of Aaprt X (cm) 0.81 2.5 - 2.5 0.81

    Y (cm) 0.5 0.5 - 0.5 0.5

    td PD (µsec) 0.50

     prr  PRF (Hz) 3930

    .7

     pr  at max. I pi  pr @PIImax  (MPa) 2.47

    d eq at max. I pi  deq@PIImax  (cm) 0.63

    Focal Length FLX (cm) 0.81 - 0.81

    FLY (cm) 3.0 - 3.0

    I pi at max. MI  (uJ/

    cm2)

    79.7

    0

    Frequency (MHz) 5.0 5.0 5.0 - 5.0 5.0

    SVD (cm) 2.4 2.2 2.2 - 2.2 2.2

    Vel Scale (kHz) 3.9 2.6 2.6 - 2.6 2.6

    SV 1 2 2 - 2 2

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    46/212

     Acoustic Output 

    1-40 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    8L-RS

    Table 1-35: Transducer Mode 8L-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TIC

    scan non-scan non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.36 0.67 0.57

    IEC FDA Units

     pra  pr.3  (MPa) 3.07

    P  Wo (mW) 32 32

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.4

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 5.09 4.79 4.79

    Dim of Aaprt X (cm) 3.84 3.84

    Y (cm) 0.4 0.4

    td PD (µsec) 0.25

     prr  PRF (Hz) 39

     pr  at max. I pi  pr @PIImax  (MPa) 3.91

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 3.84 3.84

    FLY (cm) 2.3 2.3

    I pi at max. MI  (uJ/

    cm2)

    78.25

    Frequency (MHz) 7 THI8 THI8

    Image Depth (cm) 7.0 3.0 3.0

    Focus (cm) 2.0 2.4 2.4

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    47/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-41

    Direction 5314625-100 Rev. 5 

    Table 1-36: Transducer Mode 8L-RS Operating Mode: M Mode (Inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scan

    non-scan

    non-

    scan

    Aaprt

    <

    Aaprt

    >1

    Global Maximum: Index Value 1.05 0.68 0.68 - 1.10 0.98

    IEC FDA Units

     pra  pr.3  (MPa) 2.36

    P  Wo (mW) 26.4 26.4 26.4 26.4

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.4

    z  at max. I pi, zsp (cm) 1.7

    d eq(z b) deq(zsp) (cm) 0.63

    ƒawf  f c (MHz) 5.07 4.79 4.79 - 4.79 4.79

    Dim of Aaprt X (cm) 3.84 0.83 - 0.83 3.84

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 0.20

     prr  PRF (Hz) 510.9

     pr  at max. I pi  pr @PIImax  (MPa) 3.02

    d eq at max. I pi  deq@PIImax  (cm) 0.63

    Focal Length FLX (cm) 3.84 3.84 - 3.84 3.84

    FLY (cm) 2.3 2.3 - 2.3 2.3

    I pi at max. MI  (uJ/cm2) 42.72

    Frequency (MHz) 7 THI8 THI8 - THI8 THI8

    Image Depth (cm) 7.0 3.0 3.0 - 3.0 3.0

    Focus (cm) 2.0 2.4 2.4 - 2.4 2.4

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    48/212

     Acoustic Output 

    1-42 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-37: Transducer Mode 8L-RS Operating Mode: Color Flow Mode

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.09 0.58 0.50

    IEC FDA Units

     pra  pr.3  (MPa) 2.39

    P  Wo (mW) 27.8 27.8

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.4

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 4.86 4.86 4.86

    Dim of Aaprt X (cm) 3.84 3.84

    Y (cm) 0.4 0.4

    td PD (µsec) 0.64

     prr  PRF (Hz) 7

     pr  at max. I pi  pr @PIImax  (MPa) 3.02

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 3.84 3.84

    FLY (cm) 2.3 2.3

    I pi at max. MI  (uJ/cm2)

    152.00

    Frequency (MHz) 5 5.0 5.0

    Image Depth (cm) 5 5.0 5.0

    Vel Scale (kHz) 8.7 8.7 8.7

    ROI Max Max Max

    PS 16 16 16

    Focus (cm) 2.0 2.0 2.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    49/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-43

    Direction 5314625-100 Rev. 5 

    Table 1-38: Transducer Mode: 8L-RS Operating Mode: Pulsed Doppler (inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.43 0.42 0.42 - 1.79 0.66

    IEC FDA Units

     pra  pr.3  (MPa) 3.12

    P  Wo (mW) 17.2 17.2 17.2 17.2

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.2 1.3

    z  at max. I pi, zsp (cm) 1.3

    d eq(z b) deq(zsp) (cm) 0.35

    ƒawf  f c (MHz) 4.78 4.94 4.94 - 4.94 4.94

    Dim of Aaprt X (cm) 3.84 0.83 - 0.83 0.84

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 0.51

     prr  PRF (Hz) 1302

     pr  at max. I pi  pr @PIImax  (MPa) 3.80

    d eq at max. I pi  deq@PIImax  (cm) 0.58

    Focal Length FLX (cm) 3.84 3.83 - 3.84

    FLY (cm) 2.3 2.3 - 2.3

    I pi at max. MI    (uJ/cm2)

    189.52

    Frequency (MHz) 5.0 5.0 5.0 - 5.0 5.0

    SVD (cm) 2.0 2.0 2.0 - 2.0 2.0

    Vel Scale (kHz) 1.3 1.3 1.3 - 1.3 1.3

    SV 1 2 2 - 2 2

    Depth (cm) 4.0 4.0 4.0 - 4.0 4.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    50/212

     Acoustic Output 

    1-44 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    8C-RS

    Table 1-39: Transducer Mode 8C-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TIC

    scan non-scan non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.28 0.49 0.43

    IEC FDA Units

     pra  pr.3  (MPa) 2.76

    P  Wo (mW) 21.8 21.8

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.5

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 4.64 4.69 4.49

    Dim of Aaprt X (cm) 0.62 0.62

    Y (cm) 0.5 0.5

    td PD (µsec) 0.31

     prr  PRF (Hz) 45

     pr  at max. I pi  pr @PIImax  (MPa) 3.50

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.62 0.62

    FLY (cm) 2.76 2.76

    I pi at max. MI  (uJ/

    cm2)

    102.1

    8

    Frequency (MHz) THI10 THI10 THI10

    Image Depth (cm) 5.0 5.0 5.0

    Focus (cm) 2.4 2.2 2.2

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    51/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-45 

    Direction 5314625-100 Rev. 5 

    Table 1-40: Transducer Mode 8C-RS Operating Mode: M Mode (Inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.99 0.5 0.5 - 2.73 0.89

    IEC FDA Units

     pra  pr.3  (MPa) 2.14

    P  Wo (mW) 22.4 22.4 22.4 22.4

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.5

    z  at max. I pi, zsp (cm) 1.5

    d eq(z b) deq(zsp) (cm) 1.5

    ƒawf  f c (MHz) 4.67 4.67 4.67 - 4.67 4.67

    Dim of Aaprt X (cm) 0.62 2.5 - 2.5 0.62

    Y (cm) 0.5 0.5 - 0.5 0.5

    td PD (µsec) 0.32

     prr  PRF (Hz) 501.

    1

     pr  at max. I pi  pr @PIImax  (MPa) 2.72

    d eq at max. I pi  deq@PIImax  (cm) 6.2

    Focal Length FLX (cm) 0.62 2.5 - 0.62

    FLY (cm) 2.76 2.76 - 2.76

    I pi at max. MI  (uJ/

    cm2)

    55.5

    8

    Frequency (MHz) THI

    10

    THI10 THI10 - THI

    10

    Image Depth (cm) 5.0 5.0 5.0 - 5.0

    Focus (cm) 2.4 2.2 2.2 - 2,2

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    52/212

     Acoustic Output 

    1-46 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-41: Transducer Mode 8C-RS Operating Mode: Color Flow Mode

    Index Label MI

    TIS TIB

    TIC

    scan

    non-scan

    non-

    scan

    Aaprt

    <

    Aaprt

    >1

    Global Maximum: Index Value 0.80 0.42 0.32

    IEC FDA Units

     pra  pr.3  (MPa) 1.78

    P  Wo (mW) 18.3 18.3

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.5

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 4.89 4.06 4.06

    Dim of Aaprt X (cm) 0.81 0.81

    Y (cm) 0.5 0.5

    td PD (µsec) 0.74

     prr  PRF (Hz) 6

     pr  at max. I pi  pr @PIImax  (MPa) 2.29

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.81 0.81

    FLY (cm) 3.0 3.0

    I pi at max. MI  (uJ/cm2) 106.86

    Frequency (MHz) 5 4 4

    Image Depth (cm) 5.0 5 5

    Vel Scale (kHz) 6.6 6.4 6.4

    ROI Max Min Min

    PS 14 14 14

    Focus (cm) 1.98 1.98 1.98

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    53/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-47 

    Direction 5314625-100 Rev. 5 

    Table 1-42: Transducer Mode: 8C-RS Operating Mode: Pulsed Doppler (Inc. B-mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.81 0.35 0.35 - 2.67 0.59

    IEC FDA Units

     pra  pr.3  (MPa) 1.70

    P  Wo (mW) 14.8 14.8 14.8 14.8

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.8 1.98

    z  at max. I pi, zsp (cm) 1.8

    d eq(z b) deq(zsp) (cm) 0.63

    ƒawf  f c (MHz) 4.88 4.98 4.98 - 4.98 4.98

    Dim of Aaprt X (cm) 0.81 2.5 - 2.5 0.81

    Y (cm) 0.5 0.5 - 0.5 0.5

    td PD (µsec) 0.51

     prr  PRF (Hz) 393

    0.8

     pr  at max. I pi  pr @PIImax  (MPa) 2.30

    d eq at max. I pi  deq@PIImax  (cm) 0.63

    Focal Length FLX (cm) 0.81 0.81 0.81

    FLY (cm) 3.0 3.0 3.0

    I pi at max. MI  (uJ/

    cm2)

    87.9

    8

    Frequency (MHz) 5 5 5 5 5 5

    SVD (cm) 2.2 2.2 2.2 2.2 2.2 2.2

    Vel Scale (kHz) 3.9 1.3 1.3 1.3 1.3 1.3

    SV 1 3 3 3 3 3

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    54/212

     Acoustic Output 

    1-48 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    i12L-RS

    Table 1-43: Transducer Mode i12L-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TIC

    scan non-scan non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.11 0.81 0.53

    IEC FDA Units

     pra  pr.3  (MPa) 2.58

    P  Wo (mW) 31.4 31.4

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.0

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 5.44 4.83 4.83

    Dim of Aaprt X (cm) 2.5 2.5

    Y (cm) 0.7 0.7

    td PD (µsec) 0.22

     prr  PRF (Hz) 54

     pr  at max. I pi  pr @PIImax  (MPa) 3.11

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 2.5 2.5

    FLY (cm) 1.2 1.2

    I pi at max. MI  (uJ/

    cm2)

    99.7

    3

    Frequency (MHz) 6.0 THI8 THI8

    Image Depth (cm) 4.0 4.0 4.0

    Focus (cm) 1.65 1.65 1.65

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    55/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-49

    Direction 5314625-100 Rev. 5 

    Table 1-44: Transducer Mode i12L-RS Operating Mode: M Mode (Inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.84 0.81 0.81 - 0.61 0.98

    IEC FDA Units

     pra  pr.3  (MPa) 1.78

    P  Wo (mW) 28 28 28 28

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.0

    z  at max. I pi, zsp (cm) 1.2

    d eq(z b) deq(zsp) (cm) 0.71

    ƒawf  f c (MHz) 5.52 4.88 4.88 - 4.88 4.88

    Dim of Aaprt X (cm) 2.5 0.4 - 0.4 2.5

    Y (cm) 0.7 0.7 - 0.7 0.7

    td PD (µsec) 0.21

     prr  PRF (Hz) 521.

    0

     pr  at max. I pi  pr @PIImax  (MPa) 2.15

    d eq at max. I pi  deq@PIImax  (cm) 0.71

    Focal Length FLX (cm) 2.5 2.5 - 2.5

    FLY (cm) 1.2 1.2 - 1.2

    I pi at max. MI    (uJ/

    cm2)

    45.2

    1

    Frequency (MHz) 6.0 THI8 THI8 THI8 THI8

    Image Depth (cm) 4.0 4.0 4.0 4.0 4.0

    Focus (cm) 1.65 1.65 1.65 1.65 1.65

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    56/212

     Acoustic Output 

    1-50 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-45: Transducer Mode i12L-RS Operating Mode: Color Flow Mode

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.69 0.80 0.53

    IEC FDA Units

     pra  pr.3  (MPa) 3.79

    P  Wo (mW) 31.5 31.5

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.0

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 5.02 4.12 4.12

    Dim of Aaprt X (cm) 2.5 2.5

    Y (cm) 0.7 0.7

    td PD (µsec) 0.71

     prr  PRF (Hz) 6

     pr  at max. I pi  pr @PIImax  (MPa) 4.43

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 2.5 2.5

    FLY (cm) 1.2 1.2

    I pi at max. MI    (uJ/cm2)

    563.81

    Frequency (MHz) 5.0 4 4

    Image Depth (cm) 4.0 4.0 4.0

    Vel Scale (kHz) 5.6 6.0 6.0

    PS 16 16 16

    Focus (cm) 1.5 1.5 1.5

    ROI Max Max Max

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    57/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-51

    Direction 5314625-100 Rev. 5 

    Table 1-46: Transducer Mode: i12L-RS Operating Mode: Pulsed Doppler (inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.46 0.52 0.52 - 2.23 0.71

    IEC FDA Units

     pra  pr.3  (MPa) 3.27

    P  Wo (mW) 23.2 23.2 23.2 23.2

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.1 1.1

    z  at max. I pi, zsp (cm) 1.1

    d eq(z b) deq(zsp) (cm) 0.71

    ƒawf  f c (MHz) 5.00 4.24 4.24 - 4.24 4.24

    Dim of Aaprt X (cm) 2.5 0.8 - 0.8 2.5

    Y (cm) 0.7 0.7 - 0.7 0.7

    td PD (µsec) 0.50

     prr  PRF (Hz) 1302

     pr  at max. I pi  pr @PIImax  (MPa) 3.97

    d eq at max. I pi  deq@PIImax  (cm) 0.71

    Focal Length FLX (cm) 2.5 0.8 - 2.5

    FLY (cm) 1.2 1.2 - 1.2

    I pi at max. MI    (uJ/cm2)

    259.40

    Frequency (MHz) 5.0 4.0 4.0 - 4.0 4.0

    SVD (cm) 1.6 1.6 1.6 - 1.6 1.6

    Vel Scale (kHz) 1.3 1.0 1.0 - 1.0 1.0

    SV 1 3 3 - 3 3

    Depth (cm) 4.0 4.0 4.0 - 4.0 4.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    58/212

     Acoustic Output 

    1-52 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    12L-RS

    Table 1-47: Transducer Mode 12L-RS Operating Mode: B Mode (for LOGIQ e R5.2.x)

    Index Label MI

    TIS TIB

    TIC

    scan non-scan non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.82 0.84 0.54

    IEC FDA Units

     pra  pr.3  (MPa) 2.01

    P  Wo (mW) 30.4 30.4

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.5

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 5.96 5.96 5.96

    Dim of Aaprt X (cm) 0.5 0.5

    Y (cm) 0.4 0.4

    td PD (µsec) 0.21

     prr  PRF (Hz) 43

     pr  at max. I pi  pr @PIImax  (MPa) 2.73

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 1.5 1.5

    FLY (cm) 2.0 2.0

    I pi at max. MI  (uJ/

    cm2)

    50.6

    9

    Frequency (MHz) 7.0 7.0 7.0

    Image Depth (cm) 4.0 4.0 4.0

    Focus (cm) 1.8 1.8 1.6

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    59/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-53

    Direction 5314625-100 Rev. 5 

    Table 1-48: Transducer Mode 12L-RS Operating Mode: M Mode (Inc. B-Mode)

    (for LOGIQ e R5.2.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.83 0.81 0.81 0.83 1.15

    IEC FDA Units

     pra  pr.3  (MPa) 2.03

    P  Wo (mW) 27.6 27.6 27.6 27.6

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.5

    z  at max. I pi, zsp (cm) 1.2

    d eq(z b) deq(zsp) (cm) 0.51

    ƒawf  f c (MHz) 5.99 5.94 5.94 - 5.94 5.94

    Dim of Aaprt X (cm) 0.5 3.8 - 3.8 0.5

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 0.15

     prr  PRF (Hz) 526.

    67

     pr  at max. I pi  pr @PIImax  (MPa) 2.78

    d eq at max. I pi  deq@PIImax  (cm) 0.51

    Focal Length FLX (cm) 1.5 1.5 - 1.5

    FLY (cm) 2.0 2.0 - 2.0

    I pi at max. MI  (uJ/

    cm2)

    39.3

    9

    Frequency (MHz) 7.0 7.0 7.0 - 7.0

    Image Depth (cm) 4.0 4.0 4.0 - 4.0

    Focus (cm) 1.8 1.8 1.8 - 1.8

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    60/212

     Acoustic Output 

    1-54 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-49: Transducer Mode 12L-RS Operating Mode: Color Flow Mode

    (for LOGIQ e R5.2.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.10 1.74 1.0

    IEC FDA Units

     pra  pr.3  (MPa) 2.53

    P  Wo (mW) 55.9 55.9

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.5

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 6.54 6.54 6.54

    Dim of Aaprt X (cm) 0.5 0.5

    Y (cm) 0.4 0.4

    td PD (µsec) 0.55

     prr  PRF (Hz) 11

     pr  at max. I pi  pr @PIImax  (MPa) 3.55

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.6 0.6

    FLY (cm) 1.3 1.3

    I pi at max. MI    (uJ/cm2) 184.91

    Frequency (MHz) 6.7 6.7 6.7

    Image Depth (cm) 5 5.0 5.0

    Vel Scale (kHz) 5.0 10.4 10.4

    Penet Max Max Max

    ROI 12 18 18

    PS 1.4 1.6 1.6

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    61/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-55 

    Direction 5314625-100 Rev. 5 

    Table 1-50: Transducer Mode: 12L-RS Operating Mode: Pulsed Doppler (Inc. B-mode)

    (for LOGIQ e R5.2.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.71 0.35 0.35 - 2.29 0.51

    IEC FDA Units

     pra  pr.3  (MPa) 1.44

    P  Wo (mW) 10.8 10.8 10.8 10.8

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.8 1.8

    z  at max. I pi, zsp (cm) 1.8

    d eq(z b) deq(zsp) (cm) 0.55

    ƒawf  f c (MHz) 5.11 6.66 6.66 - 6.66 6.66

    Dim of Aaprt X (cm) 0.5 3.8 - 3.8 0.5

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 0.47

     prr  PRF (Hz) 130

    2.5

     pr  at max. I pi  pr @PIImax  (MPa) 2.24

    d eq at max. I pi  deq@PIImax  (cm) 0.55

    Focal Length FLX (cm) 1.1 1.1 - 1.1

    FLY (cm) 1.1 1.1 - 1.1

    I pi at max. MI  (uJ/

    cm2)

    41.2

    2

    Frequency (MHz) 5.0 6.7 6.7 - 6.7 6.7

    SVD (cm) 1.6 1.8 1.8 - 1.8 1.8

    Vel Scale (kHz) 1.3 1.7 1.7 - 1.7 1.7

    SV 1 2 2 - 2 2

    Depth 4.0 4.0 4.0 - 4.0 4.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    62/212

     Acoustic Output 

    1-56 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    12L-RS (continued)

    Table 1-51: Transducer Mode 12L-RS Operating Mode: B Mode (for LOGIQ e R6.x.x)

    Index Label MI

    TIS TIB

    TICscan non-scan non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.09 0.89 0.56

    IEC FDA Units

     pra  pr.3  (MPa) 2.60

    P  Wo (mW) 34.2 34.2

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.2

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 5.72 5.98 5.98

    Dim of Aaprt X (cm) 0.5 0.5

    Y (cm) 0.4 0.4

    td PD (µsec) 0.11

     prr  PRF (Hz) 41

     pr  at max. I pi  pr @PIImax  (MPa) 3.31

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 1.5 1.5

    FLY (cm) 2.0 2.0

    I pi at max. MI  (uJ/

    cm2)

    54.48

    Frequency (MHz) CHI12 CHI12 CHI12

    Image Depth (cm) 4.0 4.0 4.0

    Focus (cm) 1.4 1.4 1.4

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    63/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-57 

    Direction 5314625-100 Rev. 5 

    Table 1-52: Transducer Mode 12L-RS Operating Mode: M Mode (Inc. B-Mode)

    (for LOGIQ e R6.x.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.10 0.95 0.95 0.91 1.22

    IEC FDA Units

     pra  pr.3  (MPa) 2.63

    P  Wo (mW) 29.1 29.1 29.1 29.1

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.2

    z  at max. I pi, zsp (cm) 1.5

    d eq(z b) deq(zsp) (cm) 0.51

    ƒawf  f c (MHz) 5.71 5.94 5.94 - 5.94 5.94

    Dim of Aaprt X (cm) 0.5 3.8 - 3.8 0.5

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 0.13

     prr  PRF (Hz) 526.7

     pr  at max. I pi  pr @PIImax  (MPa) 3.33

    d eq at max. I pi  deq@PIImax  (cm) 0.51

    Focal Length FLX (cm) 1.5 1.5 - 1.5

    FLY (cm) 2.0 2.0 - 2.0

    I pi at max. MI  (uJ/

    cm2)

    55.52

    Frequency (MHz) CHI12 CHI12 CHI12 - CHI12

    Image Depth (cm) 4.0 4.0 4.0 - 4.0

    Focus (cm) 1.4 1.4 1.4 - 1.4

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    64/212

     Acoustic Output 

    1-58 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-53: Transducer Mode 12L-RS Operating Mode: Color Flow Mode

    (for LOGIQ e R6.x.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.70 1.86 1.15

    IEC FDA Units

     pra  pr.3  (MPa) 4.32

    P  Wo (mW) 68.2 68.2

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.0

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 6.49 4.97 4.97

    Dim of Aaprt X (cm) 0.5 0.5

    Y (cm) 0.4 0.4

    td PD (µsec) 0.58

     prr  PRF (Hz) 11

     pr  at max. I pi  pr @PIImax  (MPa) 5.41

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.6 0.6

    FLY (cm) 1.3 1.3

    I pi at max. MI    (uJ/cm2) 394.86

    Frequency (MHz) 6.7 5 5

    Image Depth (cm) 4.0 4.0 5.0

    Vel Scale (kHz) 0.8 3.8 3.8

    Penet Max Max Max

    ROI 16 16 16

    PS 1.2 1.2 1.2

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    65/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-59

    Direction 5314625-100 Rev. 5 

    Table 1-54: Transducer Mode: 12L-RS Operating Mode: Pulsed Doppler (Inc. B-mode)

    (for LOGIQ e R6.x.x)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.30 0.41 0.41 - 1.75 0.54

    IEC FDA Units

     pra  pr.3  (MPa) 2.90

    P  Wo (mW) 13.8 13.8 13.8 13.8

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.0 1.0

    z  at max. I pi, zsp (cm) 1.0

    d eq(z b) deq(zsp) (cm) 0.55

    ƒawf  f c (MHz) 4.94 6.65 6.65 - 6.65 6.65

    Dim of Aaprt X (cm) 0.5 3.8 - 3.8 0.5

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 0.51

     prr  PRF (Hz) 130

    2.1

     pr  at max. I pi  pr @PIImax  (MPa) 3.44

    d eq at max. I pi  deq@PIImax  (cm) 0.55

    Focal Length FLX (cm) 1.1 1.1 - 1.1

    FLY (cm) 1.1 1.1 - 1.1

    I pi at max. MI    (uJ/

    cm2)

    202.

    93

    Frequency (MHz) 5.0 6.7 6.7 - 6.7 6.7

    SVD (cm) 1.2 0.8 0.8 - 0.8 0.8

    Vel Scale (kHz) 1.3 1.7 1.7 - 1.7 1.7

    SV 1 2 2 - 2 2

    Depth 4.0 4.0 4.0 - 4.0 4.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    66/212

     Acoustic Output 

    1-60 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    3S-RS

    Table 1-55: Transducer Mode 3S-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TIC

    scan non-scan non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.64 0.69 1.24

    IEC FDA Units

     pra  pr.3  (MPa) 2.4

    P  Wo (mW) 82.1 82.1

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 3.8

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 1.86 1.76 1.76

    Dim of Aaprt X (cm) 0.9 0.9

    Y (cm) 1.2 1.2

    td PD (µsec) 0.8

     prr  PRF (Hz) 18

     pr  at max. I pi  pr @PIImax  (MPa) 3.07

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.9 0.9

    FLY (cm) 4.8 4.8

    I pi at max. MI  (uJ/

    cm2)

    252.

    12

    Frequency (MHz) THI

    3.6

    THI

    3.8

    THI

    3.8

    Image Depth (cm) 25.0 7.0 7.0

    Focus (cm) 5.76 5.12 5.12

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    67/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-61

    Direction 5314625-100 Rev. 5 

    Table 1-56: Transducer Mode 3S-RS Operating Mode: M Mode (Inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.64 0.69 - 0.69 2.98 1.39

    IEC FDA Units

     pra  pr.3  (MPa) 1.62

    P  Wo (mW) 82.1 - 82.1 82.1

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) 2.2

    z s  z1 (cm) 2.2

    z bp  zbp (cm) 2.2

    z b  zsp (cm) 3.9

    z  at max. I pi, zsp (cm) 2.2

    d eq(z b) deq(zsp) (cm) 1.48

    ƒawf  f c (MHz) 1.86 1.76 - 1.76 1.76 1.76

    Dim of Aaprt X (cm) 0.9 - 1.9 1.9 0.9

    Y (cm) 1.2 - 1.2 1.2 1.2

    td PD (µsec) 0.8

     prr  PRF (Hz) 520.

    8

     pr  at max. I pi  pr @PIImax  (MPa) 3.07

    d eq at max. I pi  deq@PIImax  (cm) 1.48

    Focal Length FLX (cm) 0.9 - 1.9 0.9

    FLY (cm) 4.8 - 4.8 4.8

    I pi at max. MI    (uJ/

    cm2)

    252.

    12

    Frequency (MHz) THI

    3.6

    THI

    3.8

    - THI

    3.8

    THI

    3.8

    Image Depth (cm) 25.0 7.0 - 7.0 7.0

    Focus (cm) 5.76 5.12 - 5.12 5.12

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    68/212

     Acoustic Output 

    1-62 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-57: Transducer Mode 3S-RS Operating Mode: Color Flow Mode

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.24 1.23 2.13

    IEC FDA Units

     pra  pr.3  (MPa) 2.46

    P  Wo (mW) 146.8 146.8

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 4.7

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 2.0 1.76 1.76

    Dim of Aaprt X (cm) 0.9 0.9

    Y (cm) 1.2 1.2

    td PD (µsec) 1.29

     prr  PRF (Hz) 12

     pr  at max. I pi  pr @PIImax  (MPa) 3.4

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.9 0.9

    FLY (cm) 4.8 4.8

    I pi aat max. MI    (uJ/cm2)

    537.84

    Frequency (MHz) 2.0 1.8 1.8

    Image Depth (cm) 15.0 17.0 17.0

    Vel Scale (kHz) 4.1 3.8 3.8

    ROI Mid Mid Mid

    PS 8 5 5

    Focus (cm) 5.44 4.8 4.8

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    69/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-63

    Direction 5314625-100 Rev. 5 

    Table 1-58: Transducer Mode: 3S-RS Operating Mode: Pulsed Doppler (Inc. B-mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.27 0.53 0.53 - 2.27 0.77

    IEC FDA Units

     pra  pr.3  (MPa) 2.33

    P  Wo (mW) 44.2 44.2 44.2 44.2

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 3.6 4.2

    z  at max. I pi, zsp (cm) 4.2

    d eq(z b) deq(zsp) (cm) 1.44

    ƒawf  f c (MHz) 2.11 2.51 2.51 - 2.51 2.51

    Dim of Aaprt X (cm) 0.9 1.9 - 1.9 0.9

    Y (cm) 1.2 1.2 - 1.2 1.2

    td PD (µsec) 0.85

     prr  PRF (Hz) 113

    2.2

     pr  at max. I pi  pr @PIImax  (MPa) 3.03

    d eq at max. I pi  deq@PIImax  (cm) 1.44

    Focal Length FLX (cm) 0.9 - 0.9

    FLY (cm) 4.8 - 4.8

    I pi at max. MI  (uJ/

    cm2)

    232.

    46

    Frequency (MHz) 2.2 2.6 2.6 - 2.6 2.6

    SVD (cm) 4.8 4.8 4.8 - 4.8 4.8

    Vel Scale (kHz) 1.1 2.7 2.7 - 2.7 2.7

    SV 1 2 2 - 2 2

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    70/212

     Acoustic Output 

    1-64 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-59: Transducer Mode: 3S-RS Operating Mode: Continueous Wave Doppler 

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.13 - 0.42 - 3.7 1.0

    IEC FDA Units

     pra  pr.3  (MPa) 0.18

    P  Wo (mW) - 44.2 44.2 44.2

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 4.2 4.2

    z  at max. I pi, zsp (cm) 4.2

    d eq(z b) deq(zsp) (cm) 1.44

    ƒawf  f c (MHz) 2.002 - 2.002 - 2.002 2.002

    Dim of Aaprt X (cm) - 1.9 - 1.9 0.44

    Y (cm) - 1.2 - 1.2 1.2

    td PD (µsec) -

     prr  PRF (Hz) -

     pr  at max. I pi  pr @PIImax  (MPa) 0.21

    d eq at max. I pi  deq@PIImax  (cm) 1.44

    Focal Length FLX (cm) 1.9 - 0.9

    FLY (cm) 0.35 - 0.35

    I pi at max. MI  (uJ/cm2)

    -

    Frequency (MHz) 2 - 2 - 2 2

    SVD (cm) 3.6 - 3.6 - 3.6 3.6

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    71/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-65 

    Direction 5314625-100 Rev. 5 

    9L-RS

    Table 1-60: Transducer Model 9L-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TIC

    scan non-scan non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.81 0.67 0.57

    IEC FDA Units

     pra  pr.3  (MPa) 1,70

    P  Wo (mW) 32 32

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.4

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 4.41 4.74 4.74

    Dim of Aaprt X (cm) 3.84 3.84

    Y (cm) 0.4 0.4

    td PD (µsec) 0.24

     prr  PRF (Hz) 30

     pr  at max. I pi  pr @PIImax  (MPa) 2.11

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 3.84 3.84

    FLY (cm) 2.3 2.3

    I pi at max. MI  (uJ/

    cm2)

    21.58

    Frequency (MHz) 7 THI8 THI8

    Image Depth (cm) 7.0 3.0 3.0

    Focus (cm) 2.2 2.4 2.4

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    72/212

     Acoustic Output 

    1-66 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-61: Transducer Model 9L-RS Operating Mode: M Mode (Inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scan

    non-scan

    non-

    scan

    Aaprt

    <

    Aaprt

    >1

    Global Maximum: Index Value 0.81 0.62 0.62 - 0.94 0.98

    IEC FDA Units

     pra  pr.3  (MPa) 1.69

    P  Wo (mW) 26.6 26.6 26.6 26.6

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.4

    z  at max. I pi, zsp (cm) 1.7

    d eq(z b) deq(zsp) (cm) 0.63

    ƒawf  f c (MHz) 4.40 4.78 4.78 - 4.78 4.78

    Dim of Aaprt X (cm) 3.84 0.83 - 0.83 3.84

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 0.24

     prr  PRF (Hz) 510.9

     pr  at max. I pi  pr @PIImax  (MPa) 2.10

    d eq at max. I pi  deq@PIImax  (cm) 0.63

    Focal Length FLX (cm) 3.84 3.84 - 3.84 3.84

    FLY (cm) 2.3 2.3 - 2.3 2.3

    I pi at max. MI  (uJ/cm2) 22.06

    Frequency (MHz) 7 THI8 THI8 - THI8 THI8

    Image Depth (cm) 7.0 3.0 3.0 - 3.0 3.0

    Focus (cm) 2.0 2.4 2.4 - 2.4 2.4

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    73/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-67 

    Direction 5314625-100 Rev. 5 

    Table 1-62: Transducer Model 9L-RS Operating Mode: Color Flow Mode

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.39 0.58 0.50

    IEC FDA Units

     pra  pr.3  (MPa) 3.06

    P  Wo (mW) 27.8 27.8

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.9

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 4.86 4.89 4.89

    Dim of Aaprt X (cm) 3.84 3.84

    Y (cm) 0.4 0.4

    td PD (µsec) 0.77

     prr  PRF (Hz) 4

     pr  at max. I pi  pr @PIImax  (MPa) 4.21

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 3.84 3.84

    FLY (cm) 2.3 2.3

    I pi at max. MI  (uJ/cm2)

    327.87

    Frequency (MHz) 5 5.0 5.0

    Image Depth (cm) 6 5.0 5.0

    Vel Scale (kHz) 5.6 8.7 8.7

    ROI Max Max Max

    PS 16 16 16

    Focus (cm) 2.4 2.0 2.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    74/212

     Acoustic Output 

    1-68 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-63: Transducer Model 9L-RS Operating Mode: Pulsed Doppler (Inc. B Mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.95 0.36 0.36 - 1.79 0.66

    IEC FDA Units

     pra  pr.3  (MPa) 2.12

    P  Wo (mW) 15.3 15.3 15.3 15.3

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 2.2 2.1

    z  at max. I pi, zsp (cm) 2.1

    d eq(z b) deq(zsp) (cm) 0.35

    ƒawf  f c (MHz) 4.96 4.96 4.96 - 4.96 4.96

    Dim of Aaprt X (cm) 3.84 0.83 - 0.83 0.84

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 1.18

     prr  PRF (Hz) 1302

     pr  at max. I pi  pr @PIImax  (MPa) 2.86

    d eq at max. I pi  deq@PIImax  (cm) 0.58

    Focal Length FLX (cm) 3.84 0.83 - 3.84

    FLY (cm) 2.3 2.3 - 2.3

    I pi at max. MI  (uJ/cm2)

    224.32

    Frequency (MHz) 5.0 5.0 5.0 - 5.0 5.0

    SVD (cm) 2.0 2.4 2.0 - 2.0 2.0

    Vel Scale (kHz) 1.3 2.6 1.3 - 1.3 1.3

    SV 1 2 2 - 2 2

    Depth (cm) 4.0 6.0 4.0 - 4.0 4.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    75/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-69

    Direction 5314625-100 Rev. 5 

    t739-RS

    Table 1-64: Transducer Model t739-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TIC

    scan non-scan non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.11 0.84 0.54

    IEC FDA Units

     pra  pr.3  (MPa) 2.62

    P  Wo (mW) 30.4 30.4

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.4

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 5.55 5.55 5.55

    Dim of Aaprt X (cm) 2.5 2.5

    Y (cm) 0.7 0.7

    td PD (µsec) 0.21

     prr  PRF (Hz) 48

     pr  at max. I pi  pr @PIImax  (MPa) 3.43

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 1.5 1.5

    FLY (cm) 2.0 2.0

    I pi at max. MI  (uJ/

    cm2)

    69.68

    Frequency (MHz) 8.0 8.0 8.0

    Image Depth (cm) 4.0 4.0 4.0

    Focus (cm) 1.6 1.6 1.6

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    76/212

     Acoustic Output 

    1-70 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-65: Transducer Model t739-RS Operating Mode: M Mode (Inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scan

    non-scan

    non-

    scan

    Aaprt

    <

    Aaprt

    >1

    Global Maximum: Index Value 1.12 0.68 0.68 - 0.83 1.15

    IEC FDA Units

     pra  pr.3  (MPa) 2.64

    P  Wo (mW) 23.6 23.6 23.6 23.6

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.4

    z  at max. I pi, zsp (cm) 1.2

    d eq(z b) deq(zsp) (cm) 0.51

    ƒawf  f c (MHz) 5.50 5.55 5.55 - 5.55 5.55

    Dim of Aaprt X (cm) 2.5 0.4 - 0.4 2.5

    Y (cm) 0.7 0.7 - 0.7 0.7

    td PD (µsec) 0.22

     prr  PRF (Hz) 518.1

    2

     pr  at max. I pi  pr @PIImax  (MPa) 3.45

    d eq at max. I pi  deq@PIImax  (cm) 0.51

    Focal Length FLX (cm) 1.5 1.5 - 1.5

    FLY (cm) 2.0 2.0 - 2.0

    I pi at max. MI  (uJ/cm2) 70.20

    Frequency (MHz) 8.0 8.0 8.0 - 8.0

    Image Depth (cm) 4.0 4.0 4.0 - 4.0

    Focus (cm) 1.6 1.6 1.6 - 1.6

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    77/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-71

    Direction 5314625-100 Rev. 5 

    Table 1-66: Transducer Model t739-RS Operating Mode: Color Flow Mode

    Index Label MI

    TIS TIB

    TIC

    scan

    non-scan

    non-

    scan

    Aaprt

    <

    Aaprt

    >1

    Global Maximum: Index Value 1.37 1.74 1.0

    IEC FDA Units

     pra  pr.3  (MPa) 3.05

    P  Wo (mW) 55.9 55.9

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.3

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 4.94 4.92 4.92

    Dim of Aaprt X (cm) 2.5 2.5

    Y (cm) 0.4 0.4

    td PD (µsec) 0.73

     prr  PRF (Hz) 8

     pr  at max. I pi  pr @PIImax  (MPa) 5.78

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.6 0.6

    FLY (cm) 1.3 1.3

    I pi at max. MI    (uJ/cm2) 347.99

    Frequency (MHz) 5 5.0 5.0

    Image Depth (cm) 6 5.0 5.0

    Vel Scale (kHz) 5.6 8.7 8.7

    ROI Middle Max Max

    PS 12 16 16

    Focus (cm) 1.4 1.6 1.6

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    78/212

     Acoustic Output 

    1-72 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-67: Transducer Model t739-RS Operating Mode: Pulsed Doppler (Inc. B Mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.40 0.34 0.34 - 2.29 0.46

    IEC FDA Units

     pra  pr.3  (MPa) 3.10

    P  Wo (mW) 10.1 10.1 10.1 10.1

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.9 1.8

    z  at max. I pi, zsp (cm) 1.8

    d eq(z b) deq(zsp) (cm) 0.55

    ƒawf  f c (MHz) 4.91 4.01 4.01 - 4.01 4.01

    Dim of Aaprt X (cm) 2.5 0.8 - 0.8 2.5

    Y (cm) 0.7 0.7 - 0.7 0.7

    td PD (µsec) 0.52

     prr  PRF (Hz) 1302.

    5

     pr  at max. I pi  pr @PIImax  (MPa) 4.20

    d eq at max. I pi  deq@PIImax  (cm) 0.55

    Focal Length FLX (cm) 3.84 0.83 - 3.84

    FLY (cm) 2.3 2.3 - 2.3

    I pi at max. MI  (uJ/

    cm2)

    272.5

    3

    Frequency (MHz) 5.0 4.0 4.0 - 4.0 4.0

    SVD (cm) 2.3 2.3 2.3 - 2.3 2.3

    Vel Scale (kHz) 1.3 1.0 1.0 - 1.0 1.0

    SV 1 5 5 - 5 5

    Depth (cm) 4.0 4.0 4.0 - 4.0 4.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    79/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-73

    Direction 5314625-100 Rev. 5 

    i739-RS

    Table 1-68: Transducer Mode i739-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TICscan non-scan non-sca

    nAaprt< Aaprt>1

    Global Maximum: Index Value 1.05 0.81 0.53

    IEC FDA Units

     pra  pr.3  (MPa) 2.35

    P  Wo (mW) 31.4 31.4

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.2

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 5.02 4.48 4.48

    Dim of Aaprt X (cm) 2.5 2.5

    Y (cm) 0.7 0.7

    td PD (µsec) 0.25

     prr  PRF (Hz) 48

     pr  at max. I pi  pr @PIImax  (MPa) 2.90

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 2.5 2.5

    FLY (cm) 1.2 1.2

    I pi at max. MI    (uJ/

    cm2)

    48.5

    0

    Frequency (MHz) 6.0 THI8 THI

    8

    Image Depth (cm) 4.0 4.0 4.0

    Focus (cm) 1.65 1.65 1.65

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    80/212

     Acoustic Output 

    1-74 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-69: Transducer Mode i739-RS Operating Mode: M Mode (Inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.06 0.72 0.72 - 0.81 0.98

    IEC FDA Units

     pra  pr.3  (MPa) 2.37

    P  Wo (mW) 28 28 28 28

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.2

    z  at max. I pi, zsp (cm) 1.2

    d eq(z b) deq(zsp) (cm) 0.71

    ƒawf  f c (MHz) 5.02 4.48 4.48 - 4.48 4.48

    Dim of Aaprt X (cm) 2.5 0.4 - 0.4 2.5

    Y (cm) 0.7 0.7 - 0.7 0.7

    td PD (µsec) 0.24

     prr  PRF (Hz)   518

    .1

     pr  at max. I pi  pr @PIImax  (MPa) 2.92

    d eq at max. I pi  deq@PIImax  (cm) 0.71

    Focal Length FLX (cm) 2.5 2.5 - 2.5

    FLY (cm) 1.2 1.2 - 1.2

    I pi at max. MI  (uJ/

    cm2)

    49.2

    7

    Frequency (MHz) 6.0 THI8 THI8 THI8 THI8

    Image Depth (cm) 4.0 4.0 4.0 4.0 4.0

    Focus (cm) 1.65 1.65 1.65 1.65 1.65

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    81/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-75 

    Direction 5314625-100 Rev. 5 

    Table 1-70: Transducer Mode i739-RS Operating Mode: Color Flow Mode

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.40 0.62 0.53

    IEC FDA Units

     pra  pr.3  (MPa) 3.09

    P  Wo (mW) 31.5 31.5

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.1

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 4.89 4.08 4.08

    Dim of Aaprt X (cm) 2.5 2.5

    Y (cm) 0.7 0.7

    td PD (µsec) 0.74

     prr  PRF (Hz) 4

     pr  at max. I pi  pr @PIImax  (MPa) 3.71

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 2.5 2.5

    FLY (cm) 1.2 1.2

    I pi at max. MI  (uJ/cm2)

    254.41

    Frequency (MHz) 5.0 4 4

    Image Depth (cm) 4.0 4.0 4.0

    Vel Scale (kHz) 5.6 6.0 6.0

    ROI Max Max Max

    PS 16 16 16

    Focus (cm) 1.5 1.5 1.5

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    82/212

     Acoustic Output 

    1-76 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-71: Transducer Mode: i739-RS Operating Mode: Pulsed Doppler (Inc. B-mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.32 0.55 0.55 - 2.23 0.71

    IEC FDA Units

     pra  pr.3  (MPa) 2.92

    P  Wo (mW) 23.4 23.4 23.4 23.4

    min of [P a(z s, I ta,a(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 2.0 2.0

    z  at max. I pi,a zsp (cm) 2.0

    d eq(z b) deq(zsp) (cm) 0.71

    ƒawf  f c (MHz) 4.89 4.99 4.99 - 4.99 4.99

    Dim of Aaprt X (cm) 2.5 0.8 - 0.8 2.5

    Y (cm) 0.7 0.7 - 0.7 0.7

    td PD (µsec) 0.53

     prr  PRF (Hz) 130

    2

     pr  at max. I pi  pr @PIImax  (MPa) 4.01

    d eq at max. I pi  deq@PIImax  (cm) 0.71

    Focal Length FLX (cm) 2.5 0.8 - 2.5

    FLY (cm) 1.2 1.2 - 1.2

    I pi aat max. MI  (uJ/

    cm2)

    245.

    66

    Frequency (MHz) 5.0 5.0 5.0 - 5.0 5.0

    SVD (cm) 2.3 2.3 1.6 - 1.6 1.6

    Vel Scale (kHz) 1.3 1.3 1.3 - 1.3 1.3

    SV 1 3 3 - 3 3

    Depth (CM) 4.0 4.0 4.0 - 4.0 4.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    83/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-77 

    Direction 5314625-100 Rev. 5 

    16L-RS(The probe is not not available for China)

    Table 1-72: Transducer Mode 16L-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TICscan non-scan non-sca

    nAaprt< Aaprt>1

    Global Maximum: Index Value 0.85 1.16 0.65

    IEC FDA Units

     pra  pr.3  (MPa) 2.53

    P  Wo (mW) 29.6 29.6

    min of [P a(z s, I ta,a(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 1.0

    z  at max. I pi,a zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 8.59 7.67 7.67

    Dim of Aaprt X (cm) 0.5 0.5

    Y (cm) 0.4 0.4

    td PD (µsec) 0.13

     prr  PRF (Hz) 36

     pr  at max. I pi  pr @PIImax  (MPa) 3.40

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 1.3 1.3

    FLY (cm) 1.9 1.9

    I pi aat max. MI    (uJ/

    cm2)

    27.8

    1

    Frequency (MHz) 12.0 10.0 10.0

    Image Depth (cm) 4.0 4.0 4.0

    Focus (cm) 1.4 1.4 1.4

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    84/212

     Acoustic Output 

    1-78 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-73: Transducer Mode 16L-RS Operating Mode: M Mode (Inc. B-Mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.85 1.28 1.28 - 0.59 1.31

    IEC FDA Units

     pra  pr.3  (MPa) 2.52

    P  Wo (mW) 28.1 28.1 28.1 28.1

    min of [P a(z s, I ta,a(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.0

    z  at max. I pi,a zsp (cm) 0.9

    d eq(z b) deq(zsp) (cm) 0.46

    ƒawf  f c (MHz) 8.59 7.70 7.70 - 7.70 7.70

    Dim of Aaprt X (cm) 0.5 3.8 - 3.8 0.5

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 0.12

     prr  PRF (Hz) 522.

    5

     pr  at max. I pi  pr @PIImax  (MPa) 3.40

    d eq at max. I pi  deq@PIImax  (cm) 0.46

    Focal Length FLX (cm) 1.3 1.3 - 1.3

    FLY (cm) 1.9 1.9 - 1.9

    I pi aat max. MI  (uJ/

    cm2)

    28.3

    1

    Frequency (MHz) 12.0 10.0 10.0 - 10.0

    Image Depth (cm) 4.0 4.0 4.0 - 4.0

    Focus (cm) 1.4 1.4 1.4 - 1.4

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    85/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-79

    Direction 5314625-100 Rev. 5 

    Table 1-74: Transducer Mode 16L-RS Operating Mode: Color Flow Mode

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 1.10 1.36 0.77

    IEC FDA Units

     pra  pr.3  (MPa) 3.26

    P  Wo (mW) 35.6 35.6

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 0.8

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 8.09 8.11 8.11

    Dim of Aaprt X (cm) 0.5 0.5

    Y (cm) 0.4 0.4

    td PD (µsec) 0.40

     prr  PRF (Hz)   8

     pr  at max. I pi  pr @PIImax  (MPa)   4.08

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.8 0.8

    FLY (cm) 1.5 1.5

    I pi at max. MI  (uJ/cm2)

    161.56

    Frequency (MHz) 8.0 8.0 8.0

    Image Depth (cm) 4.0 4.0 4.0

    Vel Scale (kHz) 1.9 0.7 0.7

    ROI Max Max Max

    PS 14 18 18

    Focus (cm) 1.2 1.2 1.2

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    86/212

     Acoustic Output 

    1-80 LOGIQ e Advanced Reference Manual  

    Direction 5314625-100 Rev. 5 

    Table 1-75: Transducer Mode: 16L-RS Operating Mode: Pulsed Doppler (Inc. B-mode)

    Index Label MI

    TIS TIB

    TIC

    scannon-scan

    non-

    scanAaprt< Aaprt>1

    Global Maximum: Index Value 0.98 0.84 0.84 - 1.75 0.74

    IEC FDA Units

     pra  pr.3  (MPa) 2.86

    P  Wo (mW) 15.3 15.3 15.3 15.3

    min of [P a(z s, I ta,a(z s)] [(W.3(Z1),ITA.3(z1)]) -

    z s  z1 (cm) -

    z bp  zbp (cm) -

    z b  zsp (cm) 1.0 1.3

    z  at max. I pi,a zsp (cm) 1.3

    d eq(z b) deq(zsp) (cm) 0.48

    ƒawf  f c (MHz) 8.07 9.98 9.98 - 9.98 9.98

    Dim of Aaprt X (cm) 0.5 3.8 - 3.8 0.5

    Y (cm) 0.4 0.4 - 0.4 0.4

    td PD (µsec) 0.52

     prr  PRF (Hz) 124

    0.1

     pr  at max. I pi  pr @PIImax  (MPa) 3.67

    d eq at max. I pi  deq@PIImax  (cm) 0.48

    Focal Length FLX (cm) 1.3 1.3 - 1.3

    FLY (cm) 1.2 1.2 - 1.2

    I pi aat max. MI  (uJ/

    cm2)

    151.

    32

    Frequency (MHz) 8.0 10.0 10.0 - 10.0 10.0

    SVD (cm) 1.6 1.6 1.6 - 1.6 1.6

    Vel Scale (kHz) 1.2 4.5 4.5 - 4.5 4.5

    SV 1 2 2 - 2 2

    Depth (CM) 4.0 4.0 4.0 - 4.0 4.0

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    87/212

    FDA Acoustic Output Data

    LOGIQ e Advanced Reference Manual 1-81

    Direction 5314625-100 Rev. 5 

    6S-RS (This probe is only for LOGIQ e R6.x.x)

    Table 1-76: Transducer Mode 6S-RS Operating Mode: B Mode

    Index Label MI

    TIS TIB

    TICscan non-scan non-sca

    nAaprt< Aaprt>1

    Global Maximum: Index Value 1.64 0.69 1.24

    IEC FDA Units

     pra  pr.3  (MPa) 2.85

    P  Wo (mW) 82.1 82.1

    min of [P (z s, I ta,(z s)] [(W.3(Z1),ITA.3(z1)])

    z s  z1 (cm)

    z bp  zbp (cm)

    z b  zsp (cm) 2.7

    z  at max. I pi, zsp (cm)

    d eq(z b) deq(zsp) (cm)

    ƒawf  f c (MHz) 3.01 2.93 2.93

    Dim of Aaprt X (cm) 0.9 0.9

    Y (cm) 1.2 1.2

    td PD (µsec) 0.49

     prr  PRF (Hz) 29

     pr  at max. I pi  pr @PIImax  (MPa) 3.76

    d eq at max. I pi  deq@PIImax  (cm)

    Focal Length FLX (cm) 0.9 0.9

    FLY (cm) 4.8 4.8

    I pi at max. MI  (uJ/

    cm2)

    311.15

    Frequency (MHz) THI6 THI6 THI6

    Image Depth (cm) 20 20 20

    Focus (cm) 3.2 5.12 5.12

  • 8/9/2019 Logie e BT11 Advanced Ref Manual.pdf

    88/212