Page 1: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Current status of the optical based SPS and LHC BGI profile monitors

J.Storey, R.Jones, S.Levasseur, H.Sandberg G.Schneider (CERN BE-BI)

IPM17, 22nd May 2017

Page 2: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Talk outline.

2

1. Design overview.

2. Experience and status of the CERN optical based:

• SPS BGI profile monitors.

• LHC BGI profile monitors.

3. Outlook.

Beam Gas Ionisation (BGI) profile monitor = IPM

Page 3: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Design overview.

3

EBe-

Vacuum

MCP

Phosphor

Prism

e-

𝜸

Magnet (B=0.2T)

Window

Optics

Inte

nsifi

er+C

amer

a

CID

MCP

30m

Camera Control

(Surface)

150m

Video Digitisation(Surface)

High Voltages (Surface)

Video Amplifier (Tunnel)

150mEGP

-2.0 kV (Cathode)

+2.0 kV (MCP in)+2.6 kV (MCPout)+6.0 kV (Phosphor)

-1.9 kV (EGP out)-1.8 kV (EGP in)

150m

CameraControl (Tunnel)

Page 4: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Field cage & MCP/Phosphor “Amplifier”.

4

NEG coated ceramic electrodes

SHV feedthroughs

EGP

Vacuum flange with

helicoflex seal

“Amplifier assembly”

MCP + Prism coated with

phosphor & Al mirror

UHV viewport

15 kV HV feedthrough

Page 5: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Vacuum chamber, optical system & camera.

5

Vacuum chamber with

viewport

Dipole magnet (0.2T) in

retracted position

Prism viewed through opening in dipole magnet

Optical system + Thermo Fisher Scientific radiation hard (1Mrad) intensified

camera (gate down to 10ns)

Page 6: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

SPS BGI profile monitors.

6

BIPMH.51634 “Horizontal” BIPMV.51734 “Vertical”

• Since 2015 same design as LHC BGI’s except no EGP calibration source.• Dipole magnet & corrector provides 0.2 T magnetic field.

Page 7: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Problems encountered in 2015/2016.

7

Failure of Thermo Scientific intensified camera readout electronics: • Not designed for operation in radiation environment.• Solution: Move electronics to lower radiation areas in the tunnel & minimise

time system is active by means of a remote relay.

Limited gain of Thermo Scientific intensified camera due to intensifier ageing: • Short term fix: Replaced camera intensifier. • Medium term: Optimise use of device to minimise degradation.• Long term: Replace with “something else”.

Failure of VME card that controls in-house HV supply: • Solution: Now use Siemens PLC based system to control HV supply.

Same for MCP & phosphor

Page 8: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

New common slow control & environmental monitoring system for CERN BGI’s.

PLC control system based on CERN standard Siemens S7 PLC, provides reliable:• Control & limits on high voltages.• Remote control of relays to turn on/off:

• Camera and camera readout boxes.• High voltage power supplies.• Video amplifiers.

• Temperature readout.• Cooling control & status (PS BGI).

Common PLC control system used for all PS, SPS & LHC BGI’s.

Based on SILECs (BE/CO) framework: • Full control through SILECs C++ client interface. • FESA classes developed.

BGI PLC based slow control system

Page 9: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Summary of 2016 run.

9

• By the end of 2016 run reasonable signal acquired from both horizontal and vertical BGI profile monitors for low & high intensity ion and proton beams.

Horizontal BGI raw image and beam profile

Vertical BGI raw image and beam profile

Relative vertical & horizontal emittance growth during the cycle

Figures from Hannes Bartosik and Alexander Huschauer (BE-ABP).

Page 10: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

SPS BGI’s: Summary of operational experience.

10

Problems encountered & solved: • Failure of Thermo Scientific intensified camera readout electronics.• Failure of legacy VME based HV control cards.• Transmission of video signal.

Known limitations: • Ageing of intensified cameras.• Ageing of the MCP.• Ageing of the phosphor.

Aims for 2017: • Establish routine operational use.

Avoid these limitations with pixel detector technology

Page 11: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

LHC BGI profile monitors.

11

• Motivation:• Ion beams (synchrotron light not available at injection energy).• Study emittance blow-up during the ramp for proton beams.

• 4 x BGI’s: Horizontal & Vertical for B1 + Horizontal & Vertical for B2.• Installed in 2007/8, operated throughout Run 1.• Not been operationally used so far in Run 2; same problems as SPS + other issues.

LHC BGI B1 Horizontal.

Same as SPS BGI’s, except:• EGP electron calibration source.• Neon gas injection providing

10-8 mbar pressure bump.

Page 12: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Experience from LHC Run I:Work of M.Sapinski, D.Vilsmeier, B.Dehning.

12

Measured beam profile shape is deformed for very high brightness LHC beams.

Proposed solutions:

1. Interquartile method • Okay for beams sizes > 0.3 mm.

2. Stronger magnet • Need ~1T → expensive for the aperture size needed for optical based BGI.

3. Determine gyro-radius with “electron sieve” • Filter above MCP with holes of various hole diameter.

Ref. http://accelconf.web.cern.ch/AccelConf/HB2014/papers/mopab42.pdf

Page 13: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Heating problem.

13

BGI temperature vs beam intensity

beam intensity (1014 p)

tem

pera

ture

(℃)

• Temperature sensor installed on prism support.• Prism is the substrate for P46 phosphor and

aluminium mirror that reflects light towards viewport.

Study by Benoit Salvant (BE-ABP).

Page 14: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Vacuum gauge and Beam Loss Monitor (BLM) measurements close to the BGI.

14

beam intensity

LHC dipole magnet current

beam loss close to BGI

vacuum pressure close to BGI

Page 15: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Broken prism.

15

• All four LHC BGI’s removed for inspection during 2016/17 winter technical stop.

• Fused silica prism broken on all 4 x LHC BGI’s.

Aluminium prism support.Broken fused silica prism.

Page 16: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

LHC BGI power loss with NEG coated electrodes.

16Study by Thomas Kaltenbacher & Christine Vollinger (BE-RF).

Impedance spectrum in red (HFSS)

LHC beam spectrum in blue

Broadband resonances → Power loss = 53.6 W

Page 17: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

LHC BGI power loss with metal electrodes.

17

Resonance overlap with beam harmonic at 480 MHz → Power loss = 205 W power loss.

Study by Thomas Kaltenbacher & Christine Vollinger (BE-RF).

Impedance spectrum in red (HFSS)

LHC beam spectrum in blue

SPS metal electrode

Page 18: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Ageing of Phosphor, MCP & Intensifier.

18

Phosphor coating burn off. Image recorded with EGP electron source Lead ion beam

Ageing effects observed in Phosphor, MCP and Intensifier, leads to:• Inhomogeneous gain → Correction with EGP “standard candle”, however, EGP

breaks & homogeneity of source not known. • Lower sensitivity.

Solutions:• Short term - optimise operational use.• Long term - replace with “something else” ( e.g. pixel detector ).

Page 19: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Conclusion & outlook: LHC BGI’s.

19

LHC BGI’s:• Aim to re-install LHC BGI’s during 2017/18 winter shutdown.• Source of prism heating identified:• Not limiting to machine operation (slight pressure rise)• More of an issue for detector components

• Studying design changes to reduce power loss• Not obvious as this comes from cage design

• Prism breaking mechanism identified:• Working on redesign

• Current design okay for ion beams but operationally limited by profile broadening for proton beams:• Could be solved by addition of stronger magnet (~2T -> superconducting magnet)• Significant investment that will depend on future operational requirements.

Page 20: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

20

SPS BGI’s:• Reasonable signals acquired for both horizontal and vertical devices in 2016.• No signs of prism heating problem.• Aim to fulfil operational requirements this year.

Conclusion & outlook: SPS BGI’s.

Page 21: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Spare Slides

21

Page 22: IPM17 CERN SPS LHC BGI · 11 • Motivation: • Ion beams (synchrotron light not available at injection energy). • Study emittance blow-up during the ramp for proton beams. •

Original 2007 MCP / Phosphor / Prism amplifier assembly photos.

22